Pulsar Measurement Pulsar MicroFlow-i Loop-Powered Veloctiy Sensor Product Manual

MicroFlow-i
Instruction Manual
2
PULSAR MEASUREMENT
3
MicroFlow-i (SECOND EDITION REV 1)
April 2021 Part Number M-MFi-S-002-1P
COPYRIGHT
© Pulsar Measurement, 2017 -20. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language in any form without the written permission of Pulsar Process Measurement Limited.
WARRANTY AND LIABILITY
Pulsar Measurement guarantee for a period of 2 years from the date of delivery that it will either exchange or repair any part of this product returned to Pulsar Process Measurement Limited if it is found to be defective in material or workmanship, subject to the defect not being due to unfair wear and tear, misuse, modification or alteration, accident, misapplication, or negligence.
DISCLAIMER
Pulsar Measurement neither gives nor implies any process guarantee for this product and shall have no liability in respect of any loss, injury or damage whatsoever arising out of the application or use of any product or circuit described herein.
Every effort has been made to ensure accuracy of this documentation, but Pulsar Measurement cannot be held liable for any errors.
Pulsar Measurement operates a policy of constant development and improvement and reserves the right to amend technical details, as necessary.
The MicroFlow-i shown on the cover of this manual is used for illustrative purposes only and may not be representative of the actual MicroFlow-i supplied.
CONTACT
For technical support, please contact: Europe: supporteurope@pulsarmeasurement.com
Outside Europe: supportnorthamerica@pulsarmeasurement.com
If you have any comments or suggestions about this product, please contact:
Europe: europe@pulsarmeasurement.com Outside Europe: northamerica@pulsarmeasurement.com Pulsar Measurement website: www.pulsarmeasurement.com
United States
11451 Belcher Road South Largo, FL 33773 888-473-9546
Canada
16456 Sixsmith Drive Long Sault, Ont. K0C 1P0 855-300-9151
United Kingdom
Cardinal Building, Enigma Commercial Centre
Sandy’s Road, Malvern
WR14 1JJ 00 44 (0)1684 891371
MICROFLOW-I INSTRUCTION MANUAL
4
CONTENTS
Chapter 1: Start Here… ........................................................................................................... 6
About this Manual ............................................................................................................... 6
About the MicroFlow-i ....................................................................................................... 7
Benefits ..................................................................................................................................... 8
Functional Description ....................................................................................................... 8
Product Specification .......................................................................................................... 9
EU Certificate of Conformity ......................................................................................... 11
Chapter 2 Installation ........................................................................................................... 12
Unpacking ............................................................................................................................ 12
Power Supply Requirements......................................................................................... 12
Cable Screen and Earth Requirements ..................................................................... 12
Information specific to Hazardous Area Installation ........................................... 13
Location ................................................................................................................................ 18
Mounting .............................................................................................................................. 18
Angled Mounting Bracket.............................................................................................. 19
MicroFlow-i Sensor ........................................................................................................... 20
Terminal Connection Details ........................................................................................ 25
Chapter 3 Locating The MicroFlow-i .............................................................................. 26
Preparation for Operation ............................................................................................. 30
General Maintenance ...................................................................................................... 30
Sensor Maintenance. ....................................................................................................... 31
Chapter 4 Understanding How The MicroFlow-i Works ........................................ 33
MicroFlow-I HART compatible ..................................................................................... 33
4-20 mA Device ................................................................................................................. 33
Chapter 5 MicroFlow-i PC .................................................................................................. 34
Software installation ........................................................................................................ 34
Connecting to MicroFlow-i HART PC ........................................................................ 35
Flow Tab ................................................................................................................................ 37
Configuration ...................................................................................................................... 38
Parameter Get/Set ............................................................................................................ 38
PULSAR MEASUREMENT
5
MicroFlow-i Manual Setting ......................................................................................... 39
4 – 20mA Trim .................................................................................................................... 40
Trace Tab .............................................................................................................................. 41
Comport ................................................................................................................................ 43
Language .............................................................................................................................. 43
Tools menu .......................................................................................................................... 43
Info .......................................................................................................................................... 44
Exit ........................................................................................................................................... 45
Chapter 6 Parameter Listing and Description ............................................................ 45
Parameter Access .............................................................................................................. 45
Device Information Parameters ................................................................................... 45
Processing Parameters .................................................................................................... 46
MicroFlow-i Firmware Upgrade ................................................................................... 50
MicroFlow-i HART Firmware Upgrade ...................................................................... 52
Chapter 7 Troubleshooting ................................................................................................ 54
Chapter 8 Disposal ................................................................................................................ 55
Notes .......................................................................................................................................... 56
MICROFLOW-I INSTRUCTION MANUAL
6
CHAPTER 1: START HERE…
Congratulations on your purchase of a Pulsar MicroFlow-i. This quality system has been developed over many years and represents the latest in high technology flow monitoring.
It has been designed to give you years of trouble-free performance, and a few minutes spent reading this operating manual will ensure that your installation is as simple as possible.
About this Manual
It is important that this manual is referred to for correct installation and operation. There are various parts of the manual that offer additional help
or information as shown.
Tips
Look for t h i s i con throu g ho u t y o ur Pulsar M e a s u r ement manual to find hel p f u l information a n d answers t o f r eq uently asked q u estions .
Additional Information
References
See Also
References to other parts of the manual.
Additional Information
At various parts of the manual, you will find sections like this that explain specific things in more detail.
PULSAR MEASUREMENT
7
About the MicroFlow-i
The MicroFlow-i is a 2-wire loop-powered non-contacting velocity sensor, with hazardous area approval. The sensor provides reliable flow velocity measurements using short pulses of micro-waves, which are transmitted by an enclosed antenna. When reflected off a moving surface, the signal experiences a shift in frequency characteristics. The reflected signal is captured by the on-board microprocessor via the antenna and analysed to determine the velocity.
The MicroFlow-i’s compact and versatile design makes it easy to install and ideal for confined spaces. Housed in a robust enclosure coupled with a fully encapsulated microwave antenna, it provides an effective fit-and-forget flow velocity solution.
Perfect for remote monitoring or when there is no mains power available,
due to its low power consumption. To access flow velocity data, use Pulsar’s
MicroFlow-i HART PC or HART protocol. Overall, the device has outstanding stability, accuracy and repeatability.
MICROFLOW-I INSTRUCTION MANUAL
8
Benefits
Combined with the ease of programming there are many other benefits to using the MicroFlow, these are listed below:
Non-contact velocity measurement.
Loop powered and ideal for remote monitoring.
Low-power consumption.
Hazardous area approval (Ex ia) for Zone 0.
‘See through’ capability for enclosed plastic pipes and concrete (not
reinforced) channels.
Suitable for abrasive and aggressive materials.
Resilient to liquid vapour, wind or temperature.
No minimum liquid level required for measurement
Maintenance free.
No interruption to operational flow.
Can be installed in addition to existing in process contact measuring
devices.
Easily setup using MicroFlow-i HART PC.
The sensor is capable of monitoring flow of very shallow liquid,
mitigating the constraint associated with in-liquid techniques.
Functional Description
Based upon state-of-the-art micro-wave sensing technology, the MicroFlow­i integrates Pulsar’s innovative and award-winning DSP (Digital Signal Processing) platform, offering you a cost-effective solution with excellent reliability and performance.
The Pulsar MicroFlow-i has been designed to provide maintenance free performance. The MicroFlow-i is also intrinsically safe (Ex ia) for Zone 0. The MicroFlow-i can be mounted via a rear 1” BSP to M20 thread adapter (supplied with unit). The MicroFlow-i is loop powered and gives a 4 to 20mA output which is proportional to the flow velocity. A ‘fault condition’ alarm will give out a reading of either 3.8mA or 22mA.
The 2-wire sensor can either be used in digital HART mode or as 4-20mA loop powered device. The MicroFlow-i loop powered version can be set up using a HART modem with either proprietary HART software such as Pact ware or Pulsar MicroFlow-i HART PC software.
Boot up time from power up to stable reading: cold boot 8.5 seconds, warm boot 4 seconds (if within 12 hours from last start up).
PULSAR MEASUREMENT
9
Product Specification
PHYSICAL
Sensor body material
Valox 357U
Mounting Connection
Via 1” BSP or NPT back mounted thread, 45º Angled mounted bracket (optional) and M20 thread adaptor.
Mounting angle
45º optimum and mounted at the center line of the channel, with clear uninterrupted flow.
Sensor dimensions
Diameter 90mm x height 140mm (3,54” x 5.51”)
Weight
0.7kg (1.54 lbs)
Transducer Cable Extensions
2-core screened
Maximum Separation
Up to 1000m (3,280 ft)
ENVIRONMENTAL
Enclosure protection
IP68
Max & min temperature (electronics)
-20º to +60ºC (-4ºF to +140ºF)
APPROVALS
CE and Radar Approval
See EU Declaration of Conformity
ATEX Approval
CML 16ATEX2331X, IECEx CML 16.0105X II 1 G Ex ia IIC T4 Ga II 1 D Ex ia IIIC T1350C Da
PERFORMANCE
Velocity Range
0.2 – 6 m/s (0.66 ft/s – 19.7 ft/s)
Accuracy
Greater of +/- 0.5% or 0.05 m/s (0.1 ft/s)
Optimum Installation
Install at an angle of 45° in line with flow. For more information, see ‘Locating the MicroFlow-i sensor’ section of this manual.
Max. channel width per sensor
1.5m
Radar
K-Band (ISM)
Transmit Power
<15 dBM
Beam width
20º inclusive
OUTPUTS
Communication
HART compatible, 4-20mA loop powered.
PROGRAMMING
PC programming
Via RS232
Programmed data integrity
Via non-volatile RAM, plus backup
PC Setup and monitoring Software
MicroFlow PC – compatible with Win 7 (latest service pack), Win 8, Win 10
MICROFLOW-I INSTRUCTION MANUAL
10
SUPPLY
Operating voltage
10 – 28 VDC
Power Consumption
Start-up = 20mA. Average current = 60µA per hour when one velocity measurement is performed every 15 minutes
Pulsar Measurement operates a policy of constant development and improvement and reserve the right to amend technical details, as necessary.
PULSAR MEASUREMENT
11
EU Certificate of Conformity
MICROFLOW-I INSTRUCTION MANUAL
12
CHAPTER 2 INSTALLATION
Unpacking
Power Supply Requirements
The MicroFlow-i can operate from a DC supply of 10 to 28V. In all cases the unit typically consumes 0.52W of power, and will typically draw less than 22mA.
Cable Screen and Earth Requirements
The MicroFlow-i comes with a fitted integral cable, if this cable is extended then the total capacitance must not exceed the limits for the voltage applied, and the hazardous area it is installed in.
The HART modem and PC or HART programmer are used to set up the operation parameters for the MicroFlow-i sensor. The output can either be as a 4-20mA current giving flow velocity or as a digital signal via a HART master.
The MicroFlow-i cable is a twisted pair with an overall cable screen. The twisted pair are red and black, connect the red to +ve and the black to the ­ve. The screen should go to signal ground.
Important Information
All shipping cartons should be opened carefully. When using a box cutter, do not plunge the blade deeply into the box, as it could potentially cut or scratch equipment components. Carefully remove equipment from each carton, checking it against the packing list before discarding any packing material. If there is any shortage or obvious shipping damage to the equipment, report it immediately to Pulsar Process Measurement Limited.
Important Information
If the equipment is installed or used in a manner not specified in this manual, then the protection provided by the equipment may be impaired
PULSAR MEASUREMENT
13
Information specific to Hazardous Area Installation
(Reference European ATEX Directive 2014/34/EU)
The following instructions apply to equipment covered by certificate number CML 16ATEX2331X
1. The equipment may be used with flammable gases and vapours with
apparatus groups IIC, IIB, and IIA with temperature classes; T4 ambient temperature range -20oC to +60oC.
2. The equipment is only certified for use in ambient temperatures in the
range -20oC to +60oC and should not be used outside this range.
3. Installation shall be carried out in accordance with the applicable code
of practise by suitably trained personnel.
4. Repair of the equipment shall be carried out in accordance with the
acceptable code of practise.
5. Certification as detailed in drawing number D-804-1200 (Ex ia).
6. If the equipment is likely to encounter aggressive substances, then it is
the responsibility of the user to take suitable precautions that prevent it from being adversely affected, thus ensuring that the type of protection is not compromised.
Aggressive Substances: e.g., such as acidic liquids or gases that may attack metals or solvents that may affect polymeric materials.
Suitable precautions: e.g., regular checks as part of routine inspections or establishing from the materials data sheet that it is resistant to specific chemicals.
7. The certificate number has an ‘X’ suffix that indicates that the following
special condition of certification applies. In the case of the MicroFlow-i sensor, due to the housing and labels being non-conductive plastic care needs to be taken with regards to electrostatic charge. The equipment shall not be installed if the conditions are conductive to the build-up of electrostatic charge. Additionally, the equipment should only be cleaned with a damp cloth.
MICROFLOW-I INSTRUCTION MANUAL
14
Limitations on use
1. The MicroFlow-i must be routinely inspected to avoid the build-up of
dust layers when installed in Zone 21 & Zone 22.
2. Electrostatic hazard – The MicroFlow-i must only be wiped with a damp
or antistatic cloth.
3. The outer enclosure is made from Valox357U, consider the performance
of this material with respect to chemicals that may be present in the hazardous area.
Hazardous Area Specific Power Supply and barrier Requirements
The MicroFlow-i operates from a DC supply of 10 – 28VDC and will typically draw less than 22mA.
When connecting the sensor to the PC software with a HART modem, it is recommended to use a supply of at least 18VDC.
The MicroFlow-i cable is a twisted pair with overall cable screen. The twisted pair is red and black, connect the red to +ve and the black to -ve. The screen should go to signal ground.
Important Information
In the case of the Ex ia version, the sensor must be powered from an approved I.S. safety barrier or approved I.S. power supply with the following limits:
Uo <= 28V, Io <= 162mA, Po <= 1.03W
PULSAR MEASUREMENT
15
Typical wiring for the MicroFlow-i:
MICROFLOW-I INSTRUCTION MANUAL
16
Important Information
If the equipment is installed or used in a manner not specified in this manual, then the protection provided by the equipment may be impaired.
PULSAR MEASUREMENT
17
Loop Resistance
For two wire operation, the maximum cable resistance allowable can be calculated from the graph below. For example, if an dBi Series Intelligent Transducer was supplied from 24V, the maximum total cable resistance is 600 ohms, for a typical 77 ohm /km cable this would mean a maximum cable length of 600/77 = 7.8km, remember this total cable resistance, so this figure must be divided by 2 to give 3.9km max distance.
Allowable total cable resistance for
Microflow-i
Total Cable resistance
Supply Voltage
MICROFLOW-I INSTRUCTION MANUAL
18
Location
When choosing a location to mount the bracket and MicroFlow, note the following:
For optimum readings, it is recommended that the MicroFlow should
be positioned at a 45° angle to the surface of the material being measured.
The mounting surface must be vibration-free.
The ambient temperature should be between -20ºC and 60ºC.
Mounting
For the best results it is highly recommended to use the MicroFlow-i angled bracket, which is available from your Pulsar distributor, which will ensure that the MicroFlow-i can be correctly installed. The bracket can be used if the MicroFlow-i sensor is to be mounted on its own in a specific location or
along with a Pulsar dBi or dB level transducer. Please see ‘Angled Mounting Bracket’ for dimensions of this bracket, please note that the
drawing is not to scale.
Mounting sensor to a bracket’ illustrates how you would install the sensor to one of Pulsar’s angled brackets.
The MicroFlow-i angled bracket should be mounted by marking and drilling the holes suitable for fixing your screws/bolts (length to suit your application) and then fixing the bracket in place.
Important Information
All electronic products are susceptible to electrostatic shock, so follow proper grounding procedures during installation.
PULSAR MEASUREMENT
19
Angled Mounting Bracket
Part number: dBA0008MF
45
o
Side view
176mm
22mm
R20mm
12mm
100mm
90mm
130mm
Top view
MICROFLOW-I INSTRUCTION MANUAL
20
R20mm
4 off 12mm
100mm
60mm
20mm
20mm
100mm
60mm
26mm
Rear view
Important Information
Using a spirit level, ensure that the Mounting Bracket is level before attaching the sensor to the bracket.
PULSAR MEASUREMENT
21
MicroFlow-i Sensor
The dimensions of the sensor body are shown below in Fig.1 and Fig.2:
DON’T FORGET
90 mm
1” BSP
78 mm
39mm
140 mm
73 mm
M i
45°
Surface of measurement
Green ‘Dot’ to be
positioned on top of sensor.
To obtain the most accurate results, ensure the device is mounted at a 45° angle to the surface being measured.
MICROFLOW-I INSTRUCTION MANUAL
22
Mounting sensor to a bracket
Mounting the sensor to the 45° angled bracket is done via an adapter and M20 nut, as shown in Fig.3:
For correct installation, we recommend that the adapter is threaded on the cable, and carefully screwed onto the MicroFlow-i before fitting to the bracket. This will reduce the risk of any ’twisting’ in the cable.
Ensure that the sensor is tightened, and the dot is in the correct position.
Bracket
Adapter
Fig.3
Cable M20 Nut
Green dot to be
displayed on top
Important Information
When fixing the sensor to the adapter, ensure that care is taken when pulling the sensor cable through. And once attached to the adapter, and the M20 nut is tightened, the green dot should be central to the movement of flow as shown in Fig.4.
PULSAR MEASUREMENT
23
Care should be taken not to overtighten the sensor when everything is connected, as seen in Fig.5 below, as this could cause damage to the housing.
M
Central to the movement
of flow in channel.
Fig.4
Fig.5
305mm
215mm
205mm
MICROFLOW-I INSTRUCTION MANUAL
24
When the sensor is mounted on the bracket with dB series level transducer, you can position the bracket in a way where the MicroFlow-i is obtaining measurements following the direction of the flow, or facing the direction of flow. Fig.6 and Fig.7 illustrate the correct positioning of the devices according to the direction of flow:
Direction of flow
Fig.6
Green dot to be
displayed on top
Direction of flow
Fig.7
Green dot to be
displayed on top
PULSAR MEASUREMENT
25
Terminal Connection Details
Power
The MicroFlow-i operates from a DC voltage supply of 10-28 volts, and
should be installed and connected in accordance with ‘Chapter 2–Power
supply requirements’ and the instructions below. When wiring the
MicroFlow-i the cable used is of a multicore design. The wire coding is shown below for ease of use when connecting the MicroFlow-i to monitor flow velocity:
You can also use Pulsar’s MicroFlow-i HART PC software to interrogate and obtain readings from the device. Please refer to Chapter 6 MicroFlow-i HART PC for more details on how to use it.
MicroFlow-i cable
110 mm
Heat Shrink
Power +10-28 VDC (Red cable) 0 Volts (Black cable) Screen (Green cable)
Important Information
If the equipment is installed or used in a manner not specified in this manual, then the protection provided by the equipment may be impaired.
MICROFLOW-I INSTRUCTION MANUAL
26
CHAPTER 3 LOCATING THE MICROFLOW-I
For optimum accuracy install the MicroFlow-i where the flow is not turbulent. An ideal location for the sensor is in the centre of a long straight channel. Vertical drops, baffles, curves, or junctions can cause the velocity profile to be distorted.
To obtain optimum results in readings, the MicroFlow-i sensor requires surface ripples to be present to provide a reliable measurement. If there are not any surface ripples within the sensors field of view under ant flow conditions, then an alternative measurement point should be found, or ripples created.
Locating the MicroFlow-i
If the width of the channel exceeds 1.5m. Multiple MicroFlow-i sensors
are required to be used in conjunction with the Ultimate Controller with firmware version 1.5.2 and later.
When a level measurement transducer is used and mounted on the
same bracket as the MicroFlow-i. Consider the ‘blanking’ distance of the transducer when mounting the bracket for the application.
Ensure that there is an unobstructed path between the sensor and the
surface of the moving liquid being measured.
Position the sensor at a height of 250mm above maximum liquid level
or up to two times the channel width from the minimum liquid level, whichever is greater. As shown in Fig.8 below:
A = Minimum 250mm above Maximum Liquid level.
B = Maximum 3m above Minimum Liquid Level.
Fig.8
B
A
Max Liquid
Level
Min Liquid Level
MicroFlow-i
PULSAR MEASUREMENT
27
In the instance where there are vertical drops, baffles, curves or junctions. The sensor should then be positioned on a straight part of the channel, at a minimum distance that is at least five times the width of the channel before a bend, to obtain optimal velocity readings.
Examples of positioning the sensor are shown in diagrams 1 - 4 below:
Important Information
If the maximum water height exceeds 2x width of the channel, OR If the maximum water height in channel is less than 50% of the width of the channel. Please consult Pulsar product support for assistance.
Distance
Distance
Diagram 1
Distance
Diagram 2
MICROFLOW-I INSTRUCTION MANUAL
28
Diagram 3
Distance
Distance Distance
Distance
Diagram 4
PULSAR MEASUREMENT
29
Installing the Sensor
The Sensor housing is designed to accept a 1-inch BSP female
threaded coupler for pipe or conduit.
Ensure that the sensor is positioned at a 45° angle to the surface of the
measured application, as this will achieve the best results. An example of this is shown in the diagram below:
When using one of the Pulsar Angled Mounting Brackets, secure the
bracket using the correct size screws/bolts/U-bolts (12mm diameter thread).
Tighten the sensor into the bend, and then screw the thread adapter
onto the bracket using the nut supplied. CAUTION: extreme over- tightening may crack or cause damage to the unit or bracket.
Position the sensor so the green ‘dot’ that is shown on the label
between the two green arrows is on the top of the device, and is central to the movement of flow in the channel.
If desired, the sensor cable may be extended up to a maximum of
100m. Install the sensor cables in grounded metal conduit. Do not run-in cable trays or duct banks with variable frequency drives or other high voltage sources.
When extending cable for a MicroFlow-i sensor, ensure that all wiring
in any junction /connector boxes are correct
When connecting multiple MicroFlow-i sensors for use with the
Ultimate controller. Ensure that all wiring in any junction /connector box is correct.
45°
Surface of measurement
Green dot to be
displayed on top
MICROFLOW-I INSTRUCTION MANUAL
30
The sensor can be positioned to obtain readings following the
direction of the flow. Or it can be positioned to measure the oncoming flow as seen below:
Preparation for Operation
Before switching on, ensure that the following is checked:
MicroFlow-i is mounted correctly and is secure. All wiring is correct to a Pulsar controller. The power supply is correctly installed.
General Maintenance
There are no user serviceable parts inside MicroFlow-i. If you experience any problems with the unit, then please contact Pulsar Process Measurement for advice. To clean the equipment, wipe with a damp cloth. Do not use any solvents on the enclosure.
MicroFlow-i obtaining readings
following the direction of the
flow.
MicroFlow-i obtaining readings from the oncoming direction of the flow.
PULSAR MEASUREMENT
31
Sensor Maintenance.
Procedure for the Removal of a MicroFlow-i Sensor
From time to time, it may be necessary to remove the sensor for cleaning or maintenance purposes, the below procedure is to ensure that this is done carefully with regard for the health and safety of the operator involved, and without damage to the sensor.
If it is necessary to replace the sensor for any reason, the following procedure should be followed:
Disconnect the power supply to the sensor.
Remove any fittings that the sensor may be attached to, allowing the
cable to be pulled through with ease.
If using a Pulsar angled bracket, the sensor head is mounted onto the
end of the bend via its process fitting, this should be carefully unscrewed in an anticlockwise direction. Ensure that the sensor is not dropped or knocked as this can cause damage to the unit.
When replacing the sensor, care must be taken not to over tighten the
unit as this can cause damage. Hand tight is sufficient.
Ensure that the green dot has been positioned in the direction of the
flow of the vessel, this will obtain optimum velocity readings that are
available. Please refer to ‘Fig.3’ in ‘Chapter 2-Installing the sensor
section on how to do this.
When the sensor has been replaced for the application, and re-
connected to a controller and the power re-applied, the unit will re­initialise and will begin the process of acquiring flow and velocity readings.
When refitting or replacing a sensor, ensure that it is installed in
accordance with ‘Chapter 2-Mounting a sensor to a bracket.
When replacing the sensor for a new one, you can use MicroFlow-i
HART PC to upload any changed parameters from your old sensor to the new sensor. Allowing continued use of the sensor which is setup
Important Information
Before attempting to remove the sensor for cleaning or any other purpose, the power to the unit should be disconnected.
MICROFLOW-I INSTRUCTION MANUAL
32
for your application.MicroFlow protocol. Allowing you to continue use
without the need to change any parameters. The correct PPE should be worn on site when performing maintenance on the system, if in doubt please contact your site Health and Safety Officer for further advice.
To remove the sensor, first loosen any glands, couplers etc. that may connect it to the Mounting Bracket, and then lift the sensor body onto a safe place. Taking care not to damage the sensor housing or cable whilst performing this removal.
The sensor can then be safely inspected for damage or material fouling and can be carefully cleaned with a damp cloth to remove any foreign debris.
Important Information
Care should be taken when removing and cleaning to not damage the device.
PULSAR MEASUREMENT
33
CHAPTER 4 UNDERSTANDING HOW THE MICROFLOW-I WORKS
To initially set up the MicroFlow-i (if required) to obtain readings from your application, it is advised to do this using MicroFlow-i HART PC.
Once the MicroFlow-i has been powered up, you can now begin to obtain flow velocity from the sensor.
MicroFlow-I HART compatible
The MicroFlow-i uses a unique protocol which enables it to be compatible to any HART controller that it is connected to, the HART address is ‘0’.
4-20 mA Device
The MicroFlow-i can be used as an ‘out of the box’ 4 – 20mA device, this is the case, by default when you switch the sensor on. The 4 – 20mA output is proportional to the flow velocity being seen, with respect to the maximum range of the sensor up to 6 m/s (19.7 ft/s).
To enable the MicroFlow-I to work successfully, ensure that the power is disconnected prior to a connection being made
Important Information
Care should be taken when removing and cleaning to not damage the device.
MICROFLOW-I INSTRUCTION MANUAL
34
CHAPTER 5 MICROFLOW-I PC
The software that accompanies the MicroFlow-i can be used to access and change all parameters via the HART modem, test, obtain and record trace readings from a sensor. To utilise the features of the software, a USB HART modem which is available from Pulsar (as shown in Fig.9 below), is required to communicate between the sensor and your machine.
Software installation
Insert the CD that accompanies the MicroFlow-i sensor into your PC/Laptop. If you do not have a CD drive on your laptop, then this software can be downloaded from the Pulsar Process Measurement website:
https://pulsarmeasurement.com/downloads/software/ and choose to
download MicroFlow-i HART PC from the list which will then be downloaded onto your machine. When setting up the HART modem for the first use, you will need to install the necessary drivers for it to operate correctly with your OS. These drivers can be installed from your installation disc. Run the MicroFlow-i HART PC Setup.exe from the CD, or from the downloaded file from our website to install the MicroFlow-i HART PC software correctly, which will automatically install the necessary drivers for your HART modem. A shortcut icon will be created onto your Desktop for ease of use. Double click on this icon to launch the program.
Important Information
For optimum installation insert your HART modem to your PC whilst installing MicroFlow-i HART PC, where your PC should then assign the modem to a Com Port. Or if you have an internet connection, upon inserting the USB of the HART modem for the first time, Windows updates should install this onto your
computer for you. Once the driver’s installation has completed, remove the
HART modem from your machine. Your modem is ready to be used when a connection to MicroFlow-i PC is required.
PULSAR MEASUREMENT
35
Connecting to MicroFlow-i HART PC
When connecting to the software using the HART modem, power to the sensor can be obtained from a 10-28VDC source.
To enable the MicroFlow-i to work correctly please ensure that the sensor is wired as shown in the picture below:
To enable the sensor to work correctly, the power (red) cables from the supply and to the MicroFlow-i must be wired in series with the 250 Ohm resistor (as seen in the above picture).
Once the sensor is connected to the HART modem, insert the USB into the PC/laptop ready for use. You can now use MicroFlow-i HART PC to change settings of the sensor to suit your application.
To make a connection to the MicroFlow-i sensor, supply power to the sensor
and then select the button which will change to ‘Scanning’ and the following pop-up windows will appear during an initial connection:
Power (+) to MicroFlow-i (Red cable)
0V to MicroFlow-i (Black cable)
Power (+) from supply (Red cable)
0V from supply (Black cable)
250 Ohm resistor
Important Information
Ensure that power to the sensor is switched ‘OFF’ before removing cables and
any connection is made.
Connector
MICROFLOW-I INSTRUCTION MANUAL
36
Once the MicroFlow-i parameters have been retrieved successfully, the status condition will then appear as:
PULSAR MEASUREMENT
37
If the sensor fails to connect to the software, please check that the Comport has been correctly set (details of how to do this is outlined in this chapter). After launching the program and connecting to the MicroFlow-i, the device’s parameters will be automatically extracted once a connection is established.
Once connected you can now view the information received from the sensor on the Software’s individual tabs: Flow, Configuration and Trace.
appears after a connection is established, selecting this will
then disconnect the MicroFlow-i.
Flow Tab
The above picture is an example of the default screen of MicroFlow-i HART PC, which is the Flow tab. The dial on the left displays the linear flow velocity of a maximum up to 6m/s (19.7 ft/s). While the numerical display on the right shows the velocity flow rate and the current mA reading.
Signal Strength
The signal strength gives an indication of the actual strength of the velocity signal detected. A level of higher than 70% is desired for reliable operation.
MICROFLOW-I INSTRUCTION MANUAL
38
Stability
The stability level is an indication of the consistency of flow reading, as derived from the combination of signal strength and statistical fluctuations of the flow reading. A high level of stability would indicate optimum condition, and relatively healthy flow within the channel and indicate higher levels of repeatability and better flow tracking. The stability does not necessarily indicate the accuracy of the flow reading, as this is subject to calibration.
Configuration
Parameter Get/Set
This feature allows you to obtain and change parameter values from the list available. Further information of these parameters and their values can be found in Chapter 7 – Parameter Listing and Descriptions.
PULSAR MEASUREMENT
39
Default:
When selected, this will display the current parameters default value. Selecting ‘Set’ will change the parameters value to that of its default value.
Set:
Allows the value of the selected parameter to be changed. Select the
desired parameter from the drop down ‘Parameter’ box, enter the required
value and select ‘Set’. The value box will turn green when the change of parameter value has been successfully carried out.
MicroFlow-i Manual Setting
The option for manually setting the parameters can be done by utilising the Manual Setting terminal, as shown below:
This feature is also available when the sensor has been set into ‘Program mode’, where all if the MicroFlow-i parameters can be queried by entering the parameter address in the Parameter field and then clicking on the ‘Get’
button. The MicroFlow-i will then reply with the current value of that parameter.
Selecting ‘Get’ will read and display the current value of the MicroFlow-i parameter selected. When ‘selected, within the terminal box you will see a
command appear, example <</p102 as the software queries the chosen parameter. The unit will reply with the parameter and its current value, example: ==/p102:120
Selecting ‘Set’ allows the value of the selected MicroFlow-i parameter to be changed, both the address and value must be entered before clicking on the button. <</p102:100 will show the parameter number and the value you have changed in the manual setting box. And ==/p102:140 will appear when the change of the parameter and its value is complete. The appropriate access levels are required for setting parameters.
MICROFLOW-I INSTRUCTION MANUAL
40
The values need to be valid and within the allowed range for that parameter
for the change to be successful. Please refer to ‘Chapter 7 Parameter listing and Descriptions’ in this manual, on the parameters available and
the range each one can be set between. If a parameter has been incorrectly
set, choose the parameter from the options in ‘Parameter Get/Set’ and select
the default button to display the default value for that parameter. Pressing set will revert the parameter to its default value.
4 – 20mA Trim
This feature is only available to use when the sensor has been set to ‘Run mode’, the options will then no longer be greyed out and will now be
available to use.
If the device connected to the mA output is out of calibration, and is unable to be calibrated. Then the low and high current levels can be calibrated by
altering ‘Set 4mA’ (Low Trim) and ‘Set 20mA’ (High Trim). It is advised that
to carry out calibration, that a Digital Multi Meter (DMM) is used, so that 4mA or 20mA respectively are shown on the device.
Calibrating 4mA
To calibrate the 4mA, type in ‘4’ in the ‘Constant mA’ value box and press ‘Send’. If the low value reading is incorrect, then it can be trimmed using the ‘Set 4mA’ option. Enter the offset directly into the value box and then select the ‘Set 4mA’ button to enable you to get the expected result
(reading).
Calibrating 20mA
To calibrate the 20mA, type in ‘20’ in the ‘Constant mA’ value box and press ‘Send’. If the high value reading is incorrect, then it can be trimmed using the ‘Set 20mA’ option. Enter the offset directly into the value box and then select ‘Set 20mA’ button to enable you to get the expected result (reading).
PULSAR MEASUREMENT
41
Resume mA
When calibration is complete, to release the constant mA reading select
‘Resume mA’ which will return you to normal measurement/mA output
readings. If this is not selected when calibration is complete, then your mA output will not be updated.
Trace Tab
Upon opening the trace tab, the information at the bottom of the screen will scroll through Measurement in progress, updating mA, Updating Trace. Once the trace has updated the diagnostic traces from the sensor will be displayed. You are also able to see the current velocity measurement obtained by the sensor.
The vertical axis is related to the signal strength, while the horizontal axis is related to the detected flow.
MICROFLOW-I INSTRUCTION MANUAL
42
When a signal is received, the ‘green square indicator’ will move down from 10 on the vertical axis. The further the ‘green square indicator’ moves from 10, the larger the strength of the received signal. During flow
condition, the ‘green square indicator’ will normally be within the range of
30 to 70. Under no-flow conditions the ‘green square indicator’ will remain at 10.
When flow is detected, the ‘red circle indicator’ will move horizontally to indicate flow detection. The ‘red circle indicator’ should stay at 0 under no-
flow conditions, and move along the horizontal axis when flow is detected. Under flow condition, the line should still approach and flatten out at just
below the signal index 10 line. If the trace remains flat or the indicators struggle to remain stable during flow, then increase the ‘Gain’ (P102) setting.
You can zoom in by clicking and holding the mouse to draw a box over the region to zoom in on. Zooming out can be done by right clicking on the chart and selecting “Undo All Zoom”.
Record Traces
To record traces and the current velocity, select the button and enter a site ID when prompted to do so. The record button will then change to green whilst the software is recording the data from the sensor. A new trace will be recorded when Updating Trace…. is shown in the status box.
Replay Traces
To replay trace and velocity data, the sensor needs to be disconnected from the software. Select the button and a prompt will appear to
choose the data file you wish to play. The replay button will change to green whilst the software is replaying the data. During replay select the replay button again to stop.
Record
Replay
Important Information
Traces will only be recorded whilst the trace screen is displayed. Upon leaving the trace screen, a message will appear to inform you of this. Returning to the trace screen will resume recording of the trace data.
PULSAR MEASUREMENT
43
Comport
If MicroFlow-i PC fails to connect to the sensor, the status condition will appear as below:
You will need to change the communications port that is being used. To do this, select ‘Comport’ from the top menu bar and the Comm Port setup window will appear, as seen below:
From the drop-down box select the Comport that has been assigned to your HART modem, then press select and then close. Comport assignments can be viewed in the Device Manager area of Microsoft Windows.
Your Comport information will then appear at the bottom of the MicroFlow-i
HART PC: .
Language
This menu allows you to select the language you wish to view the information on the software in the language option selected.
Tools menu
This menu option allows you to choose the following options:
Convert Logs to CSV
This allows you to export a saved data file (.MLG only) into CSV format. And once opened in Excel, the information recorded in this file can be viewed and each section is labelled at the top of the row: Date, Time, Velocity
COM 8 1200 8 Odd
MICROFLOW-I INSTRUCTION MANUAL
44
(m/s), Signal Strength and Stability. This information from the MicroFlow­i can then be placed into a chart/graph format.
Change Work Folder
This will allow you to select the default folder for where the trace data files are to be saved to.
Run HART Bootloader Control
This function allows you to reprogram the sensor with HART firmware, and should only be accessed by service personnel or under the guidance of Pulsar service engineers.
Run Bootloader Control
This function allows you to reprogram the sensor with MicroFlow-i firmware, and should only be accessed by service personnel or under the guidance of Pulsar service engineers.
Set Program mode
This allows you to interrogate the sensor and make changes to its parameters, and view traces from the sensor. Whilst in program mode, the mA output will automatically change to high (20mA) allowing the change of parameters to be completed quickly and efficiently. However, the MicroFlow-i will continue to monitor and display any change in velocity and mA readings on the Flow tab.
Set Run mode
This mode is used once the MicroFlow-i has been set up in program mode. It is also the default mode that the sensor is in when a connection is made for the first time to a device, or after a power failure. In run mode no parameters can be changed, and all current readings are reported, and you are able to adjust the 4 – 20mA Trim.
Info
This menu option allows you to choose the following options:
PC Software
This allows you to view the current version of MicroFlow-i HART PC being used.
MicroFlow-i Firmware
When connected to a MicroFlow-i sensor, the information regarding the firmware. Hardware and serial number currently connected is displayed here.
PULSAR MEASUREMENT
45
Exit
Disconnects the sensor (leaving it in program mode) and closes MicroFlow-i HART PC.
CHAPTER 6 PARAMETER LISTING AND DESCRIPTION
Parameter Access
All the MicroFlow-i parameters have factory default values which the user receives upon first use, or when the sensor is reset. The parameters consist of two main types: Device information and Processing parameters.
Device parameters are read only and cannot be set by the user, such as the serial number of the sensor. Processing parameters can be queried and set,
there is also an associated access level required for setting, either ‘User’ or ‘Service’. Upon connection to a MicroFlow-i, the access level is defaulted to ‘User’.
The parameters are always stored and entered as whole numbers, please refer to the individual parameter for its individual value range.
Device Information Parameters
These parameters are read-only.
NAME
PARAMETER
NO.
DESCRIPTION
Serial Number
240 & 241
Registers 240 & 241 form a 32-bit number. Register 240 = Most significant 16-bit. Register 241 = Least significant 16-bit.
Firmware ID
261
A number associated with the version of firmware that is currently programmed onto the sensor.
Hardware ID
262
A number associated with the version of hardware that is currently used with the sensor
MICROFLOW-I INSTRUCTION MANUAL
46
Processing Parameters
These parameters relate to the detection and processing of the flow velocity signal.
NAME
PAR
NO.
OPTIONS
DEFAULT
NOTES
Gain
102
100 – 1600
120
Used to enter a fixed gain (sensitivity) value, a higher number means higher gain.
Quick Response Damping Factor
104
10 -999
750
When P113=1, this damping factor will be applied. When P113=0, measurement will only be damped by P109 the main damping factor.
Cal Factor
108
1 – 500
100
In percentage terms, default is 100, for 100%. 50% would half and 200% would double.
High/Main Damping
109
0 – 28
0
This sets the damping factor by adding to P104. A higher number represents more damping.
Step Response Mode
113
0 = Off 1 = On
1
When turned off, no damping bypass will be performed.
Low Damping Persist
115
1 – 100
1
This is the number of measurements that the sensor acquires, before switching into damping mode.
High Damping Persist
116
1 – 100
1
This is the number of measurements that the sensor acquires, before switching into Step Response Mode.
No Signal Persist
117
1 – 100
2
This is the number of non­measurement to persist before the sensor sees a state of no signal.
PULSAR MEASUREMENT
47
NAME
PAR NO.
OPTIONS
DEFAULT
NOTES
Signal Present Persist
118
1 -100
2
This is the number of measurements that need to persist for there to be a signal present state.
Minimum Flow Cutoff
120
10 – 4000
150
In units of mm/s velocity for user input, to set the minimum flow velocity present.
Response
125
0 = Instant
1 = Moderate
2 = Damped
Instant
Changing this to ‘Instant’
will automatically calibrate parameters in the sensor to track measurements faster. This is recommended for
Pumped flow. ‘Damped’ is
recommended when there is natural flow, as measurements will be tracked at a slower pace.
Multipoint Calibration @ 0.2 m/s
152
5 – 500
100
In percentage terms, default is 100, for 100%. 50% would half and 200% would double.
Multipoint Calibration @ 0.4 m/s
153
5 – 500
100
In percentage terms, default is 100, for 100%. 50% would half and 200% would double.
Multipoint Calibration @ 0.8 m/s
154
5 – 500
100
In percentage terms, default is 100, for 100%. 50% would half and 200% would double.
Multipoint Calibration @ 1.5 m/s
155
5 – 500
100
In percentage terms, default is 100, for 100%. 50% would half and 200% would double.
Multipoint Calibration @ 3.0 m/s
156
5 – 500
100
In percentage terms, default is 100, for 100%. 50% would half and 200% would double.
MICROFLOW-I INSTRUCTION MANUAL
48
PULSAR MEASUREMENT
49
NAME
PAR NO.
OPTIONS
DEFAULT
NOTES
Multipoint Calibration @ 5.0 m/s
157
5 – 500
100
In percentage terms, default is 100, for 100%. 50% would half and 200% would double.
Multipoint calibration mode
158
0 = Off 1 = On
1
Turns multipoint calibration on and off
Simulated flow mode
288
0 = Off 1 = On
Off
Switch simulated flow on or off.
Simulated flow velocity mm/s
289
0 – 4000
0
This sets the flow velocity in mm/s, which can be used to simulate velocity and mA readings.
Sensor Reset
297
7 = Default
0
This resets all MicroFlow-i parameters to their default values. (It does not affect any mA Trims applied).
Both P834 and P835 are used for spanning the mA output, by default the 4­20 mA output span is 0 – 6 m/s. Below are some examples of the parameters use:
Example 1: 0 – 3m/s span over 4 – 20 mA. Set P835 to 3 (m/s). Example 2: 0 – 6m/s span over 4 – 12mA. Set P835 to 12 (m/s). Example 3: 0 – 6m/s span over 20 – 4 mA. Set P834 to 6 (m/s) and
set P835 to 0 (m/s).
Example 4: 0 – 6m/s over 8 – 20 mA. Set P834 to -2 (m/s).
Important Information
Traces will only be recorded whilst the trace screen is displayed. Upon leaving the trace screen, a message will appear to inform you of this. Returning to the trace screen will resume recording of the trace data.
If using the MicroFlow-i on an application where Flow type is changed to ‘Fast’, this will reduce the damping to obtain better results. If there is natural flow and ‘Slow’ is chosen, then damping is increased to obtain better results.
MICROFLOW-I INSTRUCTION MANUAL
50
MicroFlow-i Firmware Upgrade
Firmware upgrades are performed in ‘Bootloader’ mode, and require service
level access. Note: It is advised that reprogramming is carried out by Pulsar service personnel. Also, all flow measurement and output operation will be halted when the sensor is in the ‘Bootloader’.
Please follow these steps to reprogram a sensor MicroFlow-i firmware:
1. Please ensure that the sensor is connected to a PC via the HART
modem. Prior to reprogramming, you will need the MicroFlow-i firmware file, and the COM port number of the modem.
2. Establish a connection to the sensor.
3. You will require Service access to Run Bootloader Control.
4. Select Tools > Run Bootloader Control, which will then open the
Bootloader control interface window as seen in Fig.10.
5. Once connected click ‘Load Hex file’, and choose the firmware file
you are using to upgrade the sensor.
6. Click ‘Erase-Program-Verify’, and wait for the process to finish.
7. When the ‘Verification Successful’ message is displayed, click ‘Run
firmware’ and the Bootloader Interface will close.
8. Select the ‘Search’ button on the software, and the sensor will
reconnect to MicroFlow-i HART PC with the new HART firmware installed.
Important Information
It is important to know that it takes approximately 60 minutes to reprogram a sensor with MicroFlow-i firmware successfully.
PULSAR MEASUREMENT
51
Important Information
Do not switch off supply to the unit or close the software down during the firmware upgrade.
MICROFLOW-I INSTRUCTION MANUAL
52
MicroFlow-i HART Firmware Upgrade
1. Please ensure that the sensor is connected to a PC via the HART
modem. Prior to reprogramming, you will need the HART firmware file, and the COM port number of the modem.
2. Establish a connection to the sensor.
3. You will require Service access to Run HART Bootloader Control.
4. Select Tools > Run HART Bootloader Control, which will then open the
Bootloader control interface window as seen in Fig.11.
5. Once connected click File’, and choose the firmware file you are using
to upgrade the sensor.
6. Click ‘Re-Program’, turn the sensor off and on again when prompted to
do so by the Bootloader control and then wait for the process to finish.
7. When the ‘Program File OK now rebooting’ message is displayed, you
can close the HART Bootloader and reconnect to the PC software.
8. Select the ‘Search’ button on the software, and the sensor will
reconnect to MicroFlow-i HART PC with the new HART firmware installed.
PULSAR MEASUREMENT
53
Important Information
Do not switch off supply to the unit or close the software down during the firmware upgrade.
MICROFLOW-I INSTRUCTION MANUAL
54
CHAPTER 7 TROUBLESHOOTING
This section describes many common symptoms, with suggestions as to what to do. If the issue persists, please contact your local Pulsar distributor.
SYMPTOM
WHAT TO DO
Currently velocity reads zero, but you know that there is flow movement.
Ensure that the sensor is mounted correctly in accordance with Chapter 2.
Current velocity reading is higher/lower than known movement of flow
Check that the sensor is mounted correctly. Adjust the Gain (P102) or Damping (P109) parameters to obtain correct measurements.
Unable to connect to MicroFlow-i HART PC
Check that wiring to the modem is correct. Close MicroFlow-i HART PC software, cycle power on the sensor and re-open the software. Ensure that the correct Com port has been set and retry connection.
Sensor remains at a high mA out reading (20mA)
Ensure that Simulation mode (P288) is set to 0 = Off.
Ensure that ‘Resume mA’ has been selected
if in Run mode.
Important Information
If you experience any issues with our equipment, then please contact your local Pulsar Distributor for assistance.
PULSAR MEASUREMENT
55
CHAPTER 8 DISPOSAL
Incorrect disposal can cause adverse effects to the environment. Dispose of the device components and packaging material in accordance
with regional environmental regulations including regulations for electrical \ electronic products.
Transducers
Remove power, disconnect the Transducer, cut off the electrical cable and dispose of cable and Transducer in accordance with regional environmental regulations for electrical \ electronic products.
Controllers
Remove power, disconnect the Controller, and remove battery (if fitted). Dispose of Controller in accordance with regional environmental regulations for electrical \ electronic products.
Dispose of batteries in accordance with regional environmental regulations for batteries.
EU WEEE Directive Logo
This symbol indicates the requirements of Directive 2012/19/EU regarding the treatment and disposal of waste from electric and electronic equipment.
MICROFLOW-I INSTRUCTION MANUAL
56
NOTES
PULSAR MEASUREMENT
57
www.pulsarmeasurement.com
S U P P O R T @ P U L S A R M E A S U R E M E N T . C O M
Copyright © 2020 Pulsar Measurement Ltd. Registered Address: 1 Chamberlain Square CS, Birmingham B3 3AX Registered No.: 3345604 England & Wales Rev 1.0
Loading...