Olympus BX61 User Manual [ru]

4 (1)

INSTRUCTIONS

BX61

MOTORIZED SYSTEM

MICROSCOPES

This instruction manual is for the Olympus Motorized System Microscopes Models BX61 and BX62. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope, we recommend that you study this manual thoroughly before operating the microscope. Retain this instruction manual in an easily accessible place near the work desk for future reference.

A X 7 6 1 1

BX61

CONTENTS

Correct assembly and adjustments are indispensable to allow the microscope manifest its full performance. If you are going to assemble the microscope by yourself, please read chapter 7, “ASSEMBLY” (pages. 29 to 32).

IMPORTANT — Be sure to read this section for safe use of the equipment. —

1-3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

NOMENCLATURE

 

 

 

 

4-5

 

 

 

 

 

 

2

TRANSMITTED LIGHT BRIGHTFIELD OBSERVATION PROCEDURE

6-7

 

 

 

 

 

 

 

 

 

 

3

USING THE CONTROLS

 

 

8-21

 

 

 

 

 

 

 

 

 

3-1

Base .................................................................................................................................................................................................................

 

 

 

 

8-10

 

 

1

Voltage Indication

2

Transmitted/Reflected Light Switch

 

 

 

3

Light Preset Switch

4

Using the Filters

 

 

 

3-2

Focusing Block...............................................................................................................................................................................

 

 

 

 

11-12

 

 

1

Replacing the Focus Adjustment Knob

2

F/C button

 

 

 

3

Stage UP/DOWN Buttons

4

Stage Escape Button

 

 

3-3

Stage ............................................................................................................................................................................................................

 

 

 

 

13-15

 

 

1

Placing the Specimen

2 Adjusting the X/Y-Axis Knob Tension

 

 

 

3

Rotating the Stage

 

4 Adjusting the Stage Height

 

 

3-4

Observation Tube.......................................................................................................................................................................

 

 

 

 

16-18

 

 

1

Adjusting the Interpupillar Distance 2

Adjusting the Diopter

 

 

 

3

Using the Eye Shades

4

Using the Eyepiece Micrometer Disk

 

 

 

5

Selecting the Light Path of the Trinocular Tube

 

 

 

6

Adjusting the Tilt

 

 

 

 

 

 

3-5 Condenser ..........................................................................................................................................................................................

 

 

 

 

19-20

 

 

1

Centering the Condenser 2 Compatibility of Objectives and Condensers

 

 

3-6

Immersion Objectives...................................................................................................................................................................

 

 

 

21

 

3-7

Objectives with Correction Collar ...............................................................................................................................

 

 

21

 

 

 

 

 

4

TROUBLESHOOTING GUIDE

 

 

22-24

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

SPECIFICATIONS

 

 

 

 

25-26

 

 

 

 

 

 

6

OPTICAL CHARACTERISTICS <<UIS2 (UIS) Series>>

27-28

 

 

 

 

 

 

 

 

 

7ASSEMBLY — See this section for the replacement of the light bulb. — 29-32

 

...................................................................PROPER SELECTION OF THE POWER SUPPLY CORD

33-34

This device complies with the requirements of directive 98/79/EC concerning in vitro diagnostic medical devices. CE marking means the conformity to the directive.

NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

FCC WARNING: Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

BX61

IMPORTANT

Each of these microscopes employs a UIS2/UIS (Universal Infinity System) optical design, and should be used only with modules designed for the BX2 series (which belong to the Olympus BX series microscopes) and objectives/eyepieces for the UIS series.

For the applicable module, please consult Olympus or the catalogues. Less than optimum performance may result if inappropriate accessories are used.

Configuration of Instruction Manuals

Since these microscopes are expandable to a variety of systems, separate instruction manuals are provided so that the user has to read only the manuals according to the user’s own system.

Manual Name

Main Contents

BX61

Transmitted light brightfield observation procedure

 

 

BX-UCB/U-HSTR2

Functions of the Control Box (incorporating the power supply)

 

and Hand Switch

BX2 Software for PC (CD-ROM)

Methods of PC Control of Functions

 

 

BX-RFAA

Reflected light fluorescence observation procedure

 

 

BX-RLAA

Reflected light brightfield/darkfield observation

 

 

U-UCD8A

Motorized universal condenser

 

 

U-FWT/FWR/FWO

Motorized filter wheels

 

 

SAFETY PRECAUTIONS

1.

After the equipment has been used in an observation of a specimen that

 

is accompanied with a potential of infection, clean the parts coming in

 

contact with the specimen to prevent infection.

 

· Moving this product is accompanied with the risk of dropping the

 

specimen. Be sure to remove the specimen before moving this

 

product.

 

 

· In case the specimen is damaged by erroneous operation, promptly

 

take the infection prevention measures.

 

· The product becomes unstable if its height is increased by an accessory

 

mounted on it. In this case, take anti-toppling measures to prevent the

 

specimen from being dropped when the product topples down.

2.

Install the microscope on a sturdy, level table or bench so as not to block

 

the air vents on the underside of the base.

 

Do not place the microscope on a flexible surface, as this could result in

 

blocking the air vents and cause overheating or a fire.

3.

If a foreign object is caught during motorized focusing operation, there

 

will be an error in the focusing block and the motorized focusing opera-

 

tion will be suspended.

 

 

Recovery procedure

 

 

· If there is no error in motorized operation, the caught object can be

 

removed by turning the focusing knob.

 

· If there is an error in motorized operation, the focusing knob becomes

 

inoperable. Disassemble the relevant modules to remove the caught

 

object. Replace the relevant modules afterward.

 

· Turn off the power and then on again. The system will restart unless

 

there is a malfunction in the motor.

4.

To activate temporary stop during focusing operation, turn the focusing

 

knob (or dial) on the microscope frame (in either direction) or press any of

 

the FOCUS control buttons ( ,

, F/C and ESC) (except during PC

 

communication).

 

When the main switch of the BX-UCB control box is set to “ I ” (ON), the focusing operation starts automatically as part of initialization (this operation consists of temporary lowering and then returning to the original position of the stage). If any of the operations mentioned above is performed, an emergency stop will also occur. In this case, set the main

switch to “ ” (OFF) and then “ I ” (ON) again.

1

Fig. 1

|³

Fig. 2

Safety Symbols

 

5.

When lowering the stage by pressing the stage DOWN button @, be

 

 

careful not to have your hand caught between the bottom ² of the con-

 

 

denser and the base.

 

 

6.

The surfaces of the lamp housing on the rear of the microscope will

 

 

become extremely hot during operation. When installing the microscope,

 

 

make sure to allow ample free space (10 cm or more) around and in

 

 

particular above the lamp housing.

 

7.

When installing the microscope, route the power cord away from the

 

 

lamp housing. Should the power cord come in contact with the hot lamp

 

 

housing, the power cord could melt and cause electric shock.

²

8.

To avoid potential shock hazards and burns when replacing the light

@

bulb, set the main switch ³ to “ ” (OFF) then disconnect the power cord

 

 

from the wall outlet in advance. Whenever you replace the bulb during

 

 

use or right after use, allow the lamp housing | and bulb to cool before

 

 

touching. (Fig. 2)

 

 

 

Designated

12V100WHAL (PHILIPS 7724)

 

 

haiogen bulbs:

12V50WHAL-L (LIFE JC)

 

 

#The microscope also incorporate a fuse (this should be

 

 

replaced by the manufacturer or an authorized agent).

 

9.

Always use the power cord provided by Olympus. If no power cord is

 

 

provided, please select the proper power cord by referring to the section

 

 

“PROPER SELECTION OF THE POWER SUPPLY CORD” at the end of

 

 

this instruction manual. If the proper power cord is not used, product

 

10.

safety performance cannot be warranted.

 

Always ensure that the grounding terminal of the microscope and that

 

 

of the wall outlet are properly connected. If the equipment is not grounded,

 

 

Olympus can no longer warrant the electrical safety performance of the

 

11.

equipment.

 

 

Never insert metallic objects into the air vents of the microscope frame

 

 

as this could result in electrical shock, personal injury and equipment

 

 

damage.

 

 

12.

When using the U-FWR motorized filter wheel for reflected light observa-

 

 

tion, remove the push ring driver from the position where it is attached by

magnetism in advance to prevent it from dropping.

If the driver drops in a certain place, an electric shock and/or burns may result.

The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment in the safest possible manner.

Symbol

Explanation

 

 

Indicates that the surface becomes hot, and should not be touched with bare hands.

Before use, carefully read the instruction manual. Improper use could result in personal injury to the user and/or damage to the equipment.

lIndicates that the main switch is ON.

Indicates that the main switch is OFF.

Warnings

Warning engraving/stickers are placed at parts where special precaution is required when handling and using the microscope. Always heed the warnings.

Warning engraving

Lamp housing

position

(Warning against high temperature)

 

 

 

 

 

 

 

2

1Getting Ready

Fig. 3

BX61

1.A microscope is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact.

2.Do not use the microscope where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations. (For the operating conditions, refer to chapter 5, “SPECIFICATIONS”.)

3.When moving the microscope, remove the modules that may drop including the specimen from the microscope in advance, hold it by the hand hook on the rear and the base as shown on the left (Fig. 3) and carry carefully (Weight: approx. 16 kg).

#Before transporting the microscope, be sure to remove the specimen, eyepieces and color glass filter so that they will not drop. Also remove other modules attached to the microscope because they increase the system weight.

#Damage to the microscope will occur if you grasp it by the stage, focus adjustment knob or binocular section of the observation tube.

4.The BX61 series can be used with up to two intermediate attachments including a U-CA magnification changer, U-EPA2 eyepoint adjuster, etc. For restrictions when using two intermediate attachments, make sure to read the instruction manuals provided with the respective intermediate attachments.

2Maintenance and Storage

1.To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe gently using a piece of cleaning paper (or clean gauze).

If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute alcohol.

!Since the absolute alcohol is highly flammable, it must be handled carefully.

Be sure to keep it away from open flames or potential sources of electrical sparks –– for example, electrical equipment that is being switched on or off, which could cause ignition of a fire.

Also remember to always use it only in a well-ventilated room.

2.Do not attempt to use organic solvents to clean the microscope components other than the glass components. To clean them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.

3.Do not disassemble any part of the microscope as this could result in malfunction or reduced performance.

4.When not using the microscope, set the main switch to “” (OFF), wait until the lamp housing has cooled down, and cover the microscope with a dust cover before storage.

5.When disposing of the microscope, check the regulations and rules of your local government and be sure to observe them.

3Caution

If the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the equipment may also be damaged. Always use the equipment as outlined in this instruction manual.

The warning, caution and other notes in this manual use the following symbols.

!: Indicates that failure to follow the instructions in the warning could result in bodily harm to the user and/or damage to equipment (including objects in the vicinity of the equipment).

# : Indicates that failure to follow the instructions could result in damage to equipment. } : Indicates commentary (for ease of operation and maintenance).

4Intended use

This instrument has been designed to be used to observe magnified images of specimens in routine and research applications.Do not use this instrument for any purpose other than its intended use.

3

1 NOMENCLATURE

}If you have not yet assembled the microscope, read chapter 7, “ASSEMBLY” (pages 29 to 32).

Transmitted light specification models

Light path selector knob (Page 17)

Interpupillary distance adjustment scale (Page 16)

Diopter adjustment ring (Page 16)

Specimen holder (Page 13)

Aperture iris diaphragm ring (Page 20)

Filter mount (Page 9)

Field iris diaphragm ring (Page 19)

Brightness adjustment buttons (Page 8)

Condenser height adjustment screw (Page 19)

Allen screwdriver

(accommodation position)

Lamp voltage indicator

Transmitted/reflected light switch (Page 8)

PHOTO indicator (Page 8)

Y-axis knob (Page 14)

X-axis knob (Page 14)

Optional filter knob

Filter knobs (Page 9)

ND6, ND25 and LBD from the front to the rear

4

BX61

Base

Focus adjustment knob (Page 11)

Stage escape button

 

 

 

 

 

UCB connector

 

(Page 12)

 

 

 

 

 

 

 

 

Light Preset switch (Page 8)

 

 

Stage UP button (Page 12)

 

 

F/C button (Page 11)

 

 

 

 

 

 

Stage DOWN button (Page 12)

 

Lamp ON-OFF switch

 

 

 

 

 

 

 

 

 

 

 

 

 

Reflected light specification models * For the reflected illuminators and reflected lamp housings, please also refer to their instruction manuals.

Motorized reflected light illuminator*

·Motorized fluorescence illuminator (BX-RFAA)

·Motorized reflected light illuminator (BX-RLAA)

Reflected lamp housing*

·100 W mercury burner housing (U-LH100HG)

·100 W mercury apo-burner housing (U-LH100HGAPO)

·75 W xenon lamp housing (U-LH75XEAPO)

·100 W halogen lamp housing (U-LH100-3)

·50 W metal hydride lamp housing (U-LH50MH)

5

TRANSMITTED LIGHT BRIGHTFIELD OBSERVATION PROCEDURE

 

 

 

(Controls Used)

Set the main switch to “ I ” (ON).

@Main switch “ I ” (ON)

²Lamp ON-OFF switch ON

 

 

 

 

 

 

 

 

 

 

Select the transmitted light and adjust the

³Transmitted/reflected light switch

brightness.

|Brightness adjustment buttons

 

 

 

 

 

 

 

 

 

 

Engage the LBD filter in light path.

ƒLBD filter knob

 

 

 

 

 

 

 

 

 

 

(Page)

(P. 8) (P. 8)

Select the light path (when a trinocular tube is used).

Place the specimen on the stage.

Engage the 10X objective in the light path.

Bring the specimen in focus.

Adjust the interpupillary distance.

Adjust the diopter.

Adjust the light axis.

Adjust the aperture iris and field iris diaphragms.

Engage the objective to be used in the light path and bring the specimen in focus.

…Light path selector knob

(P. 17)

†Specimen holder

(P. 13)

‡X/Y-axis knobs

(P. 14)

ŠRevolving nosepiece button

 

‰Stage UP/DOWN button, focus

(P. 12)

adjustment knob

 

‹Binocular tube

(P. 16)

ŒDiopter adjustment ring

(P. 16)

™Condenser height adjustment screw

(P. 19)

šCondenser centering screws

(P. 19)

›Aperture iris diaphragm ring

(P. 20)

œField iris diaphragm ring

(P. 19)

ŠRevolving nosepiece button

 

‰Stage UP/DOWN button, focus

(P. 12)

adjustment knob

 

Engage the required filters.

ŸFilters

(P. 9)

Adjust the brightness.

|Brightness adjustment buttons

(P. 8)

Start observation.

6

Olympus BX61 User Manual

BX61

Use the Allen screwdriver.

Œ

³

š

Use the Allen screwdriver.

œ

|

²

Hand Switch

U-HSTR2

ƒ

Ÿ

@

 

Control Box*

 

BX-UCB

 

* Set the Z-board or AF-board. For the setup

 

and installation of the Z-board, refer to

 

page 30. For the information regarding

Š

the AF-board, refer to the instruction

 

manual of the AF unit.

} Make a photocopy of the observation procedure pages and post it near your microscope.

7

USING THE CONTROLS

3-1 Base

³

1

Voltage Indication

(Fig. 4)

 

 

1.

Press the Brightness adjustment button @ to increase the voltage and

 

 

 

make illumination brighter.

 

 

 

 

Pressing the button ² makes the illumination darker.

 

 

 

2.

The numerals on the left of the lamp voltage LEDs ³ indicate the

 

 

 

approximate voltage.

 

 

 

#The LED may sometimes go out (temporarily) when the motor is

² @ |

 

driving. Even in this case, there is no fluctuation in illumination

 

voltage.

 

Fig. 4

2

Transmitted/Reflected Light Switch

 

 

(Fig. 4)

The illumination can be switched between the transmitted light and reflected light by pressing the transmitted/reflected light switch |.

: Reflected light illumination

: Transmitted light illumination

The indicator of the selected illumination lights.

ƒ

Fig. 5

3

Light Preset Switch

(Fig. 5)

}The Light Preset switch sets the light intensity voltage to a voltage suitable for color photography* (the factory default is 9 V) regardless of the current setting of the light intensity adjustment buttons.

* Achieved by engaging the built-in LBD filter.

Setting the Desired Brightness (Figs. 4 & 5)

1.Press the Light Preset switch ƒ The PHOTO indicator … lights up.

2.Press one of the light intensity adjustment buttons @ ² to set the desired brightness.

3.Press the Light Preset switch again. The indictor turns off, the brightness returns to the original brightness but the setting made above is stored in memory.

4.Hereafter, pressing the Light Preset switch sets the brightness to the in-

tensity value set in step 2 above.

#Be sure to set the light intensity voltage back to 9 V when color photography is to be performed.

}When the Light Preset switch ƒ is pressed, a note sound (short beep) is generated at the 9 V position.

4

Using the Filters

(Figs. 6 - 11)

}You can place a filter in the light path with one of the following four methods.

·Push in the knob of a built-in filter to engage it in the light path. (Page 9)

·Place a filter on the filter mount on the base and engage in the light path. (Page 9)

·Insert a filter in the U-FC filter cassette, attach it on the filter mount and engage it in the light path by sliding the filter lever. (Page 9)

·Insert a filter in the U-FWT filter wheel, attach it on the filter mount and engage it in the light path by motorized operation (using the PC). (Refer to the separate instruction manual.)

8