Ohaus RANGER COMPACT COUNTING User Manual

0 (0)
Ohaus RANGER COMPACT COUNTING User Manual

Ranger RC Compact Scales

Instruction Manual

Ranger RC

 

Table of contents

Table of contents

 

 

 

Page

1

Introduction .....................................................................................

5

1.1

Safety instructions .............................................................................

5

1.2

Description ......................................................................................

6

1.3

Putting into operation ......................................................................

10

2

Operation ......................................................................................

12

2.1

Switching on and off .......................................................................

12

2.2

Zeroing / Zero point correction ..........................................................

12

2.3

Simple weighing .............................................................................

12

2.4

Weighing with tare ..........................................................................

13

2.5

Calling up the gross weight and tare value .........................................

14

2.6

Displaying weight values with a higher resolution ...............................

14

2.7

Displaying the capacity available ......................................................

14

2.8

Dynamic weighing ..........................................................................

14

2.9

Printing results ...............................................................................

15

2.10

Switching scales .............................................................................

15

2.11

Cleaning ........................................................................................

16

3

Counting .......................................................................................

17

3.1

Counting parts into a container .........................................................

17

3.2

Counting parts out of a container ......................................................

17

3.3

Counting with variable reference quantity ...........................................

18

3.4

Counting with minimum accuracy ....................................................

18

3.5

Reference optimization ....................................................................

18

3.6

Counting with automatic reference determination ................................

19

3.7

Counting with two scales .................................................................

19

4

Settings in the menu ......................................................................

21

4.1

Operating the menu ........................................................................

21

4.2

Overview .......................................................................................

23

4.3

Scale settings (SCALE) ....................................................................

26

4.4

Application settings (APPLICATION) ..................................................

28

4.5

Terminal settings (TERMINAL) ..........................................................

30

4.6

Configuring interfaces (COMMUNICATION) .........................................

31

4.7

Diagnosis and printing out of the menu settings (DIAGNOS)

................34

5

Interface description ......................................................................

35

5.1

OHAUS interface commands ............................................................

35

5.2

SICS interface commands ................................................................

37

5.3

TOLEDO Continuous mode ..............................................................

40

6

Event and error messages ..............................................................

42

7

Technical data and accessories ......................................................

44

7.1

Technical data ................................................................................

44

7.2

Accessories ....................................................................................

49

3

Table of contents

 

Ranger RC

8

Appendix ......................................................................................

50

8.1

Information for certified scales in EC countries ...................................

50

8.2

Safety checks .................................................................................

50

8.3

Table of Geo Values ........................................................................

51

8.4

Sample protocols ...........................................................................

54

9

Index ............................................................................................

55

4

Ranger RC

Introduction

1Introduction

1.1Safety instructions

CAUTION!

Do not use Ranger RC in hazardous areas!

Our product range includes special devices for hazardous areas.

DANGER!

Electric shock hazard!

Always pull out the mains plug before any work on the device.

DANGER!

Electric shock hazard if the mains cable is damaged!

Check the mains cable for damage regularly and replace it immediately if it is damaged.

On the rear side of the device, maintain a clearance of at least 3 cm in order to prevent the mains cable bending too much.

CAUTION!

On no account open the device!

The warranty is void if this stipulation is ignored. The device may only be opened by authorized persons.

Call OHAUS Service.

5

Introduction

Ranger RC

CAUTION!

Handle the compact scale with care. The scale is a precision instrument.

When the weighing pan has been removed, never clean the area under the load plate holder with a solid object!

Do not put excessive loads on the scale. Avoid banging the weighing pan.

Disposal

Observe the valid environmental regulations when disposing of the scale. If the device has a rechargeable battery:

The battery contains heavy metals and therefore must not be disposed of with normal waste.

Observe the local regulations for disposing of environmentally hazardous materials.

Note Use with foodstuffs

Parts coming into contact with foodstuffs have smooth surfaces and are easy to clean. The materials used do not splinter and are free of harmful substances.

With foodstuffs, it is recommended to use the supplied protective cover. Clean the protective cover regularly and carefully.

Replace damaged or very dirty protective cover immediately.

1.2Description

This user manual applies to the following types of compact scales:

Compact scale RC..S with strain gauge weighing cell

Compact scale RC..M with Monobloc

The compact scales are available in a small and large size in various capacities and resolutions.

The power supply is carried out via a built-in power supply device, an internal rechargeable battery with an external mains adapter or an external battery.

One of the following options can also be ordered:

Additional interface RS232 or RS485

Ethernet interface

USB interface

Digital I/O

Analog second scale interface

6

Ranger RC Introduction

 

1.2.1

Overview

1

Display

 

2

Scale specifications

 

3Load plate

4Adjustable foot

5Keys

1Power supply

connection

2 Optional interface

3RS232 interface

1

2

3

7

Introduction

Ranger RC

1.2.2Display

1 7-segment display, 7 digits, with decimal point

2Active interface

3 Symbol for displaying gross and net values

4Active scale

5Weighing range display

6 Battery charge level; only present on scales with a battery

7Weight units

8Selected reference quantity

9 Symbols for optimizing the average piece weight

10Symbol for dynamic weighing

11Graphic display of the weighing range

12Stability monitor (goes out when a stable weight value is reached)

13Sign

14Identification for changed or calculated weight values, e.g. higher resolution, minimum weight not reached

8

Ranger RC Introduction

1.2.3

Keypad

 

 

Main functions

 

Key

Function in operating mode

Function in the menu

 

Switching device on / off, abort

To the last menu item –End-

 

Setting scale to zero

Scrolling back

 

Taring scale

Scrolling forward

 

Transfer key

Activating menu item

 

Long key press: Calling up menu

Accepting selected setting

 

Additional functions

 

Key

Function

 

Switching between gross and net weight; displaying tare specification

Calling up additional information, e.g. gross weight, average piece weight, higher

resolution ...

Switching the scale

Switching between weight value and number of pieces

Determining average piece weight from 10 pieces

Determining average piece weight from any number of pieces

9

Introduction

Ranger RC

1.3Putting into operation

1.3.1Selecting or changing the location

The correct location is crucial to the accuracy of the weighing results!

Select a stable, vibration-free and if possible a horizontal location.

The ground must be able to safely bear the weight of the fully loaded scale. Observe the following environmental conditions:

No direct sunlight

No strong drafts

No excessive temperature fluctuations

Aligning the scale

Only scales that have been aligned precisely horizontally provide accurate weighing results. The certified scales have a spirit level to simplify alignment.

Turn the adjustable feet of the scale until the spirit level’s air bubble is inside the inner circle.

Major geographical The manufacturer adjusts each scale to the local gravity conditions (GEO value). In location changes the event of major geographical location changes, this setting must be adjusted by a service technician. Certified scales must also be recertified observing the national certification regulations. These steps are not necessary for scales with an internal cali-

bration weight.

10

Ranger RC

Introduction

1.3.2Connecting the power supply CAUTION!

Before connecting the scale to the mains, check whether the voltage value printed on the rating plate corresponds with the local mains voltage.

Never connect the device if the voltage value printed on the rating plate is different to the local mains voltage.

Plug the mains plug into the socket.

After connection, the device performs a self-test. When the zero display appears, the device is ready to weigh.

Calibrate the device in order to obtain the greatest possible precision, see Section 4.3.2.

Note Partially certified scales (scales with first-level certification) must be certified by an authorised body or by the OHAUS Service.

Call OHAUS Service.

Scales with a built-in battery can work independently from the mains for approximately 30 hours in normal operation. A prerequisite for this is that the background

lighting is switched off and that no peripheral devices are connected.

The device automatically switches to battery operation as soon as the mains supply

is interrupted. When the mains supply is restored, the device automatically switches

back to mains operation.

The battery symbol indicates the present charging level of the battery. 1 segment corresponds to approx. 25 % capacity. When the symbol flashes the battery must be charged (min. 4 hours). The charging period is extended if work is continued during charging. The battery is protected against overcharging.

Note The battery’s charging capacity can be reduced under continuous mains operation.

To maintain the charging capacity, after a maximum of 4 weeks discharge the battery completely before recharging it.

11

Operation

Ranger RC

2Operation

2.1Switching on and off

Switching on

Press ON/OFF.

The scale conducts a display test. When the weight display appears, the scale is ready to weigh.

Switching off

Press ON/OFF.

Before the display goes out, -OFF- appears briefly.

2.2Zeroing / Zero point correction

Zeroing corrects the influence of slight changes on the load plate.

Manual 1. Unload scale.

2.Press ZERO.

The zero display appears.

Automatic In the case of scales that cannot be certified, the automatic zero point correction can be deactivated in the menu or the amount can be changed.

As standard, the zero point of the scale is automatically corrected when the scale is unloaded.

2.3Simple weighing

1.Place weighing sample on scale.

2.Wait until the stability monitor goes out.

3.Read weighing result.

12

Ranger RC

Operation

2.4Weighing with tare

2.4.1Taring

Place the empty container on the scale and press TARE. The zero display and the symbol NET appear.

The tare weight remains saved until it is cleared.

2.4.2Clearing the tare

Unload scale and press TARE.

The symbol NET goes out, the zero display appears.

If A.CL-tr is activated in the menu, the tare weight is automatically cleared as soon as the scale is unloaded.

2.4.3Automatic taring

Prerequisite

A-tArE is activated in the menu, the symbol T flashes in the display. Place the container or packaging material on the scale.

The packaging weight is automatically saved as the tare weight, the zero display and the symbol NET appear.

2.4.4Chain tare

Prerequisite

The tare function CHAIn.tr is activated in the menu.

With this function it is possible to tare several times if, for example, cardboard is placed between individual layers in a container.

1.Place the first container or packaging material on the scale and press TARE.

The packaging weight is automatically saved as the tare weight, the zero display and the symbol NET appear.

2.Weigh the weighing sample and read/print out the result.

3.Place the second container or packaging material on the scale and press TARE again.

The total weight on the scale is saved as the new tare weight. The zero display appears.

4.Weigh the weighing sample in the second container and read/print the result.

5.Repeat the last two steps for other containers.

13

Operation

Ranger RC

2.5Calling up the gross weight and tare value

1.Press G/N/T.

The gross weight is displayed for 2 seconds.

2.Press G/N/T again while the gross weight is displayed.

The tare value is displayed. After a few seconds, the scale changes back to the net weight display.

2.6Displaying weight values with a higher resolution

Press Info.

The current weight value is displayed for 2 seconds in a higher resolution. The scale then changes back to the normal resolution.

2.7Displaying the capacity available

The scale has a graphic display of the scale capacity available. The bar indicates how many per cent of the scale capacity is already occupied and what capacity is still available. In the example, approx. 65 % of the scale capacity is occupied.

2.8

Dynamic weighing

 

 

With the dynamic weighing function, it is possible to weigh restless weighing samples

 

such as live animals. If this function is activated, the symbol

appears in the dis-

 

play.

 

 

With dynamic weighing, the scale calculates the mean value from 56 weighing oper-

 

ations within 4 seconds.

 

With manual start

Prerequisite

 

 

AVErAGE -> MAnuAL is selected in the menu.

 

 

The weighing sample must be heavier than 5 scale divisions.

 

 

1.

Place the weighing sample on the scale and wait until it has stabilized.

 

2.

Press PRINT to start dynamic weighing.

 

 

 

During dynamic weighing, horizontal segments appear in the display, and the

 

 

dynamic result is then displayed with the symbol *.

 

 

3.

Unload the scale to be able to start a new dynamic weighing operation.

14

Ranger RC Operation

With automatic start Prerequisite

AVErAGE -> AUtO is selected in the menu.

The weighing sample must be heavier than 5 scale divisions.

1.Place the weighing sample on the scale.

The scale starts the dynamic weighing automatically.

During dynamic weighing, horizontal segments appear in the display, and the dynamic result is then displayed with the symbol *.

2.Unload the scale to be able to perform a new dynamic weighing operation.

2.9Printing results

If a printer or computer is connected to the scale, the weighing results can be printed out or sent to a computer.

Press PRINT.

The display contents are printed out and transferred to the computer. See Section 8.4 for sample protocols.

2.10Switching scales

If a second scale or a weighing platform is connected, e. g. via the optional analog second scale interface, the currently active scale is shown in the display.

The second scale can be operated in exactly the same way as the first scale. Press Scale Select.

The display changes from one scale to the other.

15

Operation

Ranger RC

2.11Cleaning

CAUTION!

Electric shock hazard!

Before cleaning with a damp cloth, pull out the mains plug to disconnect the unit from the power supply.

CAUTION!

When the weighing pan has been removed, never clean the area under the load plate holder with a solid object!

This could damage the weighing cell.

Other cleaning information:

Use damp cloths.

Do not use any acids, alkalis or strong solvents.

Do not clean using a high-pressure cleaning unit or under running water.

If very dirty, remove the weighing pan, protective cover (if present) and adjustable feet and clean these items separately.

Follow all the relevant instructions regarding cleaning intervals and permissible cleaning agents.

16

Ranger RC

Counting

3Counting

The Ranger RC compact scales have additional functions for piece counting.

The relevant settings in the menu are described in Section 4.4.1.

3.1Counting parts into a container

1.Place the empty container on the scale and press TARE. The container is tared and the zero display appears.

2.Place 10 reference parts on the scale and press Sample 10. -or-

 

 

Place the number of pieces displayed above the key Sample Size on the scale

 

 

and press Sample Size.

 

 

The scale determines the average piece weight and then shows the number of

 

 

pieces.

 

3.

Add more parts to the container until the required number of pieces is reached.

Note

The average piece weight remains saved in the factory setting until a new average

 

 

piece weight is determined.

With Count Weight it is possible to switch betweenthe number of pieces and the weight.

When the number of pieces is displayed, it is possible to display the average piece weight with Info, i. e. the weight of a single reference part, for 2 seconds.

If A.CL-APW ON is set in the menu, the average piece weight is automatically cleared after each counting operation. The average piece weight must be determined again for the next counting operation.

If ACCurCY ON is set in the menu, the accuracy achieved is briefly shown after the number of pieces is determined.

3.2Counting parts out of a container

1.Place the full container on the scale and press TARE. The container is tared and the zero display appears.

2.Remove 10 reference parts and press Sample 10.

-or-

Remove the number of pieces displayed above the key Sample Size and press

Sample Size.

The scale determines the average piece weight and then shows the number of pieces removed, together with a minus sign.

3.Remove more parts from the container until the required number of pieces is reached.

17

Counting

Ranger RC

3.3Counting with variable reference quantity

If VAr-SPL ON is set in the menu, it is possible to select from 5 preset reference quantities via Sample Size.

Press Sample Size as often as necessary until th e display above the key has changed to the desired reference quantity.

The rest of the counting process is as described earlier.

3.4Counting with minimum accuracy

The item Min.rEFW in the menu allows to preset a minimum accuracy of 97.5 %, 99.0 % or 99.5 %. On the basis of this, the scale calculates the minimum reference weight necessary to reach the defined accuracy.

1.Place the reference parts on the scale and press Sample 10 or Sample Size.

2.If the average piece weight is not sufficient to ensure the desired accuracy, Add x PCS appears.

3.Add the displayed number of pieces.

The scale then automatically determines the average piece weight with the larger reference quantity.

The rest of the counting process is as described earlier.

3.5Reference optimization

The greater the reference quantity, the more accurately the scale determines the number of pieces.

3.5.1Automatic reference optimization

rEF.OPt -> AUtO must be set in the menu for this. The symbol Auto Opt appears in the display.

1.Place the reference parts on the scale and press Sample 10 or Sample Size.

2.Place additional reference parts, max. the same number as for the first reference determination, on the scale.

The scale automatically optimises the average piece weight with the larger number of reference parts.

The rest of the counting process is as described earlier.

Note The reference optimization can be performed several times.

18