1. INTRODUCTION

1.1 Safety Precautions

Please follow these safety precautions:

Verify that the AC Adapter input voltage matches the local AC power supply.

Use the balance only in dry locations.

Do not operate the balance in hostile environments.

Do not drop loads on the platform.

Do not place the balance upside down on the platform or platform mounting cone.

Service should be performed only by authorized personnel.

1.2Controls

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

Yes

 

 

 

 

 

 

 

 

Button:

Functions:

 

 

 

 

 

 

O/T - On

Short Press:

Turns balance on, sets display to zero

Off

Long Press:

Turns balance off

Yes

Short press (Menu):

Selects or accepts setting

Unit

Short Press:

Steps through active units and modes

Menu

Long Press:

Enters Menu

No

Short press (Menu):

Steps through available settings

 

Long press (Menu):

Exit menu or abort out of menu item

Print

Short Press:

Sends data

Cal

Long Press:

Initiates Span Calibration

2. INSTALLATION

 

 

 

 

 

 

 

 

 

 

2.1 Package Contents

 

 

 

 

 

 

0.1g and 0.01g Models

0.001g and 0.0001g Models

Instruction Manual

Instruction Manual

Power Adapter

 

 

 

 

Power Adapter

Balance

 

 

 

 

Balance

Pan

 

 

 

 

Pan

Pan Support

 

 

 

 

Glass Doors and Panels

Wind-Ring (InCal models only)

Warranty Card

Warranty Card

 

 

 

 

 

 

 

 

 

 

Pioneer

EN-2

2.2 Install Components

0.1g and 0.01g Models

1) Install Wind-Ring (InCal

2) Install Pan Support

3) Install Pan

only)

 

 

0.001g and 0.0001g Models

 

 

1) Install Side Doors - Insert fully into Top Frame then down over retainer.

2) Install Panels - Insert bottom edge in groove then press until locked.

3) Install Top Door

4) Install Pan

3.2 Count Mode

EN-3

Pioneer

2.3 Level Balance

Level the balance on a firm, steady surface. Avoid locations with excessive air current, vibrations, heat sources or rapid temperature changes.

2.4 Connect Power

For use with CSA Certified (or equivalent approved) power source, which must have a limited circuit output.

2.5 Initial Calibration

Without InCal – Power on the balance by pressing 0/T. Press and hold Print/Cal until [CAL] is displayed. The display flashes the calibration mass needed. To select the alternate calibration weight press No. Put the calibration mass on the pan. The display flashes [BUSY], then [CLEAr Pan]. Remove the mass. When calibration is complete, [DONE] is displayed.

InCal – Press and hold Print/Cal until [CAL] is displayed. The display flashes [BUSY], when calibration is

complete, [DONE] is displayed.

Note: Calibrations should be performed after a warm up time of 60 minutes.

3. Operation

Count, APW Optimization, Percentage or specific units of measure must be activated in the MODE or UNIT menus if they are not initially available.

3.1 Weighing Mode

Repeatedly press Unit until the desired unit icon is displayed.

Press 0/T to zero the balance and then place objects to be weighed on the pan.

– Use the Count mode to count parts of uniform weight.

To access Count Mode, press Unit until the display shows [Count].

Establish an Average Piece Weight (APW) – Each time a new type of part is to be counted, the nominal weight of one piece (APW) must be established using a small quantity of pieces.

With [CLr.APwW] displayed, press No to use the previously saved APW, or press Yes to establish a new APW. The display indicates the number of pieces to be used to establish the new APW. If a different sample size is preferred, press No until the desired sample size is displayed (5, 10, 20, 50 or 100). Put the specified number of pieces on the pan. Press Yes to accept new APW or No to abort.

Count – Place the quantity to be counted on the pan.

APW Optimization – Since the weight of each piece varies slightly, APW Optimization may be used to increase the

Ohaus PIONEER ANALYTICAL AND PRECISION BALANCES User Manual

Pioneer

EN-4

accuracy of the count. The balance automatically recalculates the Average Piece Weight when the number of parts on the pan is less than three times the original sample size. The display shows [APwW.OPT] each time the APW is optimized.

3.3 Percent Mode – Use this mode to measure the weight of a sample as a percentage of a reference weight. To access the Percent Mode, press Unit until the display shows [PErCEnt].

Establish a new Reference Weight - With Clear reference [CLr.rEF] displayed, Press No to use the previously saved Reference Weight. Press Yes to establish a new Reference Weight. Put the reference sample on the pan and press Yes to accept or No to abort.

Percent – Place the object(s) to be compared to the reference weight on the pan.

3.4 Weigh Below Feature

Remove Weigh Below Cover

Attach wire or string to Hook

Suspend sample

4. SETTINGS

4.1 Menu Navigation

 

 

 

 

 

 

.END.

 

 

.CAL.

.SETUP.

.MODE.

.UNIT.

.PRINT.

.RS232.

.LOC.

 

InCal*

AZSM

COUNT

GRAM

OUT STA

BAUD

CAL

 

SPAN

FILTER

APWOPT

MILLIGRAM

A. PRINT

PARITY

SETUP

 

LINEAR**

A.TARE

PERCENT

KILOGRAM

CONTENT

HAND.SH

MODE

 

CAL ADJ

GRAD

END.MOD

OUNCE

LINE FO

END.R232

UNIT

 

END.CAL

GBL RST

 

POUND

END.PRT

 

PRINT

 

 

END.SET

 

NOT ALL

 

 

 

 

 

 

UNITS

 

 

RS232

 

 

 

NO

SHOWN

 

 

 

*

InCal models only

 

 

 

 

 

 

 

 

 

 

 

** Not available in InCal models

YES

END.UNIT

 

 

END.LOCK

 

 

 

 

 

Enter Menu – When the balance is on, press and hold Unit/Menu until [mMENU] appears. Release the button and the Calibrate [.CAL.] menu will display.

Menu Navigation – Select menus, menu items and settings through use of the Yes and No buttons. Solid arrows point to the content displayed when Yes is pressed, Dashed lines when No is pressed. Changing Settings – To select the displayed Setting, press Yes. To move to the next Setting, press No.

Exit Menu – When [.END.] is displayed, press Yes to exit the menu function, or press No to return to the Cal menu. Note: Press and hold No at any time to exit quickly.

EN-5

Pioneer

4.2 Calibration Menu [.CAL.]

InCal or Span calibration should be performed daily and when the room temperature changes. InCal [InCal] calibrates the balance using an internal mass.

Span Calibration [SpaN] uses two weight values: zero and a weight between 50% and 100% of the capacity of the balance.

Linearity calibration [LIN] uses three weight values zero, 50% of capacity and full capacity. Generally this calibration is not required unless testing shows that the linearity error exceeds the Linearity tolerance in the Specification table. (Not available in InCal models)

Calibration Adjust [CAL Adj] allows adjustment to the result of the internal calibration by +/- 99 divisions. (InCal models only)

4.3 Setup Menu [.setup.]

Automatic Zero-Setting [azsmM] – Environmental changes can cause the display to drift. The Automatic ZeroSetting Mechanism (AZSM) is designed to keep the balance set at zero, despite these slight changes. (OFF, SET.5d, SET 1d, SET 2d, SET 5d)

Filter [filter] – Use the low setting (SET LO) when environmental disturbances are not present. Use the medium setting (SET MED) for normal environments. Use the high setting (SET HI) when vibrations or air currents are present.

Auto Tare [A-tare] – The initial item placed on the balance is assumed to be a container so it is zeroed out. The next item is then weighed. When the pan is cleared the balance resets, waiting for a container. (SET OFF, SET ON)

Grad [Grad] – Select the readability displayed. Reducing the readability may be needed for approval. (SET 1d, SET [1]d, SET 10d)

Global Reset [Gbl rst] – Resets all settings to factory default values. (RESET)

4.4 Mode Menu [.mMode.]

Count Mode [Count] - [SET ON or OFF]

Average Piece Weight (APW) Optimization [aPwW.OPT] - (SET ON, SET OFF) Percent Mode [PerCent] - (SET ON, SET OFF)

4.5 Unit Menu [.Unit.]

The Unit menu is used to enable or disable a specific unit. (SET ON, SET OFF) The unit is indicated by a small character next to Unit in the display (g = grams). The default setting is Grams SET ON and all other units SET OFF.

T-Units – When unit [t] displays; press Yes to show the T-Unit settings; SET OFF, SET TT (Taiwan Tael), SET TH (Hong Kong Tael), SET TS (Singapore Tael), SET TO (Tola) or SET TI (Tical).

M-Units – When unit [m] is displayed, press Yes to show the M-Unit settings; SET OFF, SET MO (Momme) or SET ME (Mesghal).

Custom Unit – Custom Unit (C) is used to create a unit of measure not provided with the balance. The Custom Unit is defined by a factor, a multiplier (E) and a least significant digit (LSD). The balance will use this to convert grams to a custom unit of measure. (Example: 1 gram = 0.257206 Avoirdupois Dram, using a 4100g x 0.01g balance)

To create a custom unit, press Yes when unit [c] is displayed.

Factor – The Factor (F) is a value from 0.1000000 to 1.999999. When the Factor is displayed, the first digit is flashing. Press Yes to accept its value and activate the next digit, or No to edit. When editing, press No until the desired value appears, then press Yes to accept. Repeat until all digits have been accepted. When the Factor flashes on the display, press Yes to accept or No to re-edit. (Example: F = 0.257206)

Multiplier (E) – The settings are, [E 0] (Fx1), [E 1] (Fx10), [E 2] (Fx100), [E 3] (Fx1000), [E-3] (F/1000), [E-2] (F/100), and [E-1] (F/10). Press No to display the next setting, Yes to accept. (Example: E = 0).

Note: The multiplier selections are limited when the capacity in grams is exceeded.

Pioneer

EN-6

LSD – The Least Significant Digit (LSD) is the number of displayed divisions (d) by which the weight is incremented. The values are 1d, 2d, 5d, 10d, 100d or 0.5d. Press No to go to the next setting, press Yes to accept. (Example LSD = 1d) Note: LSD options are limited if the readability in grams is exceeded.

The example custom unit will display [0.26 C] when 1g is placed on the pan.

4.6 Print Menu [.Print.]

Output Stable [Out.Stab] – Data will only be sent when the Stable indicator is on. This setting works with manual button pressing or continuous and interval Auto Print. (SET ON, OFF)

Auto Print [A.Print] – Data will be continuously sent when [Continu] is set. Interval [Inter] sends data every 1 to 3600 seconds. When Stable [Stable] will send data when the balance detects a new stable reading. This can be a weight value only [LOad] or it can also include a stable zero [L+zero]. [OFF] disables automatic printing.

Content [CONteNt] – The content in the data transmission can be modified. Each of the following settings can be set on or off. Number Only [numMber] will only send the numeric result. Balance ID [Bal. ID] will add the Balance

serial number for traceability purposes. Reference [refer] will add reference information relevant to the current mode. GLP [GLP] will send additional items to allow proper documentation of laboratory results.

Line Format [LinE Fo] – Single line format [single] will put all the data in one line separating each output with a comma (,). Multi line format [mMulti] will put each data output on a new line. Multi +4 [mM-4LF] will add 4 line spaces between each output.

^^^^^^^^^^^^^^^^^^^^^

 

 

 

Multi Line Format with 4 line feeds

 

 

 

Line Feed-2

 

 

 

Line Feed-3

 

 

 

Line Feed-4

User ID: ………………........

 

 

GLP (ON)

Bal ID: 123456789

 

 

Balance ID (ON)

Proj ID: ……………………

 

 

GLP (ON)

Time: ……:…….:…….

 

 

GLP (ON)

Date: …../ ….. / …..

 

 

GLP (ON)

120.01 g

 

 

Result

^^^^^^^^^^^^^^^^^^^^^

4.7 RS232 Menu [.rs232.]

Baud [BAUD] – The RS232 baud rate can be set to 600, 1200, 2400, 4800, 9600 and 19200. Parity [Parity] – Parity can be sent to 7 bits-even parity [7 eveN], 7bits-odd parity [7 odd],

7bits-no parity [7 No] or 8bit-no parity [8 No].

Handshake [Handsh] – Handshake can be set to off [Off], X on – X off [on-off], or hardware [HardwWr].

4.8 Lock Menu [.LOC.]

When a Lock Menu item is SET ON the indicated menu cannot be changed.

[Loc Cal] – Calibration, [Loc SEt] – Setup, [Loc.mMod] – Mode, [Loc.Unit] – Unit, [Loc Prt] – Print, [Loc 232] – RS232.

4.9 Sealing Access to the Balance Settings

The Menu Lock switch prevents changes to the Lock Menu. The switch can be secured using paper seals, wire seals or plastic ties.

Un-Locked

Locked with Plastic Tie

EN-7

Pioneer

5.0MAINTENANCE

5.1Troubleshooting

Symptom

Possible Cause

Remedy

 

 

 

Cannot turn on

No power to balance

Verify connections and voltage.

Poor accuracy

Improper calibration

Perform calibration

 

Unstable environment

Move balance to suitable location

Cannot calibrate

Unstable environment

Move the balance to suitable location

 

Incorrect calibration masses

Use correct calibration masses

Cannot access mode

Mode not enabled

Enter menu and enable mode

Cannot access unit

Units not enabled

Enter menu and enable units

Err 5

Average Piece Weight too small

Add additional samples

Err 7.0

Time out

 

Err 8.1

Pan has load during power on

Remove weight from pan and re-zero.

Err 8.2

Pan was removed prior to power on

Install pan and re-zero.

Err 8.3

Weight on pan exceeds capacity

Remove weight from the pan

Err 8.4

Pan was removed during weighing

Re-install pan

Err 9.5

Factory calibration data corrupted

Contact the authorized dealer

Err 9.8

Factory calibration data corrupted

Perform calibration

Error 53

EEPROM Checksum error

Contact the authorized dealer

REF Err

Reference Weight is too small

Add additional samples

LOWREF

Reference Weight is too low for accurate

Add additional samples or continue to weigh

 

parts counting or percent weighing.

with less accurate results.

5.2 Service Information

If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. Please visit our web site, www.ohaus.com to locate the Ohaus office nearest you.

5.3 Accessories

Security device

76288-01

 

Auxiliary Display

PAD7

 

Density Determination Kit

Contact Ohaus

(0.1mg and 1mg models only)

Printer – Thermal

Contact Ohaus

 

Printer - Impact

Contact Ohaus

 

Cable Kit – Thermal Printer

Contact Ohaus

 

Cable Kit - Impact Printer

Contact Ohaus

 

Data collection Software

SW12W

 

Pioneer

EN-8

6. TECHNICAL DATA

Ambient conditions – The technical data is valid under the following ambient conditions: Ambient temperature: 10° C to 30° C

Relative humidity: 15 % to 80 % at 31° C non-condensing, decreasing linearly to 50% at 40° C Height above sea level: Up to 2000 m

Operability is assured at ambient temperatures between 5° C and 40° C

Power – AC adapter, Balance power input 8-14.5 VAC, 50/60Hz 4VA or 8-20 VDC, 4W Protection – Protected against dust and water, Pollution degree: 2, Installation category: Class II

6.1 Drawings

0.01g and 0.1g Models

0.0001g and 0.001g Models

EN-9

 

 

 

 

 

 

 

 

 

 

 

Pioneer

6.2 Specifications

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODEL

PA64*

PA114*

PA214*

PA213*

PA413*

PA512*

PA2102*

PA4102*

PA4101*

Capacity (g)

65

 

110

 

210

210

410

510

2100

 

4100

 

4100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Readability (g)

 

0.0001

 

 

0.001

 

0.01

 

 

 

0.1

Repeatability (sd)

 

0.0001

 

 

0.001

 

0.01

 

 

 

0.1

Linearity (g)

 

± 0.0002

 

± 0.002

 

± 0.02

 

 

 

± 0.1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Tare Range

 

 

 

 

 

To capacity by subtraction

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Stabilization

 

 

 

 

 

 

3 seconds

 

 

 

 

 

Span cal mass (g)

50 or

 

50 or

 

100 or

100 or

200 or

200 or

1000 or

 

2000 or

 

2000 or

 

60

 

100

 

200

200

400

500

2000

 

4000

 

4000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lin. cal masses (g)

25,

 

50,

 

100,

100,

200,

250,

1000,

 

2000,

 

2000,

 

50

 

100

 

200

200

400

500

2000

 

4000

 

4000

 

 

 

 

 

 

 

 

 

 

 

 

 

Pan size (in / cm)

 

3.5 / 9 dia.

 

4.7 / 12 dia.

 

7.1 / 18 dia.

 

 

 

 

 

 

 

 

 

 

 

Net Weight (lb/kg)

 

10.1 / 4.6 (*11.2 / 5.2)

 

 

10 / 4.5

(*11.1 / 5.1)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

*C= Internal calibration

6.3 Communication

6.3.1 Commands

The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed weight. The balance will return “ES” for invalid commands.

Command

Function

IP

Immediate Print of displayed weight.

 

 

P

Print displayed weight (uses Stable ON/OFF menu settings).

 

 

CP

Continuous Print.

xP

Interval Print x = Print Interval (1-3600 sec)

 

 

T

Same as pressing Zero Key.

ON

Turns balance ON.

 

 

OFF

Turns balance OFF.

PSN

Show Serial Number.

 

 

PV

Version: Print product name, software revision and LFT ON (if LFT is set ON).

 

 

PU

Print current mode/unit

x#

Set PC ref wt (x) in grams

 

 

P#

Print PC ref wt

x%

Set % ref wt (x) in grams

 

 

P%

Print % ref wt

6.3.2 RS232 (DB9) Pin Connections

Pin 2: Balance transmit line (TxD) Pin 3: Balance receive line (RxD) Pin 5: Ground signal (GND)

Pin 7: Clear to send (hardware handshake) (CTS) Pin 8: Request to send (hardware handshake) (RTS)

GND

 

RxD

 

 

 

TxD

5 1

9 6

CTS

RTS