1. INTRODUCTION
1.1 Safety Precautions
Please follow these safety precautions:
•Verify that the AC Adapter input voltage matches the local AC power supply.
•Use the balance only in dry locations.
•Do not operate the balance in hostile environments.
•Do not drop loads on the platform.
•Do not place the balance upside down on the platform or platform mounting cone.
•Service should be performed only by authorized personnel.
1.2Controls
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
No |
Yes |
||||||
|
|
|
|
|
|
|
|
|||
Button: |
Functions: |
|
|
|
|
|
|
|||
O/T - On |
Short Press: |
Turns balance on, sets display to zero |
||||||||
Off |
Long Press: |
Turns balance off |
||||||||
Yes |
Short press (Menu): |
Selects or accepts setting |
||||||||
Unit |
Short Press: |
Steps through active units and modes |
||||||||
Menu |
Long Press: |
Enters Menu |
||||||||
No |
Short press (Menu): |
Steps through available settings |
||||||||
|
Long press (Menu): |
Exit menu or abort out of menu item |
||||||||
Short Press: |
Sends data |
|||||||||
Cal |
Long Press: |
Initiates Span Calibration |
||||||||
2. INSTALLATION |
|
|
|
|
|
|
|
|
|
|
2.1 Package Contents |
|
|
|
|
|
|
||||
0.1g and 0.01g Models |
0.001g and 0.0001g Models |
|||||||||
Instruction Manual |
Instruction Manual |
|||||||||
Power Adapter |
|
|
|
|
Power Adapter |
|||||
Balance |
|
|
|
|
Balance |
|||||
Pan |
|
|
|
|
Pan |
|||||
Pan Support |
|
|
|
|
Glass Doors and Panels |
|||||
Wind-Ring (InCal models only) |
Warranty Card |
|||||||||
Warranty Card |
|
|
|
|
|
|
|
|
|
|
Pioneer™ |
EN-2 |
2.2 Install Components
0.1g and 0.01g Models
1) Install Wind-Ring (InCal |
2) Install Pan Support |
3) Install Pan |
only) |
|
|
0.001g and 0.0001g Models |
|
|
1) Install Side Doors - Insert fully into Top Frame then down over retainer.
2) Install Panels - Insert bottom edge in groove then press until locked.
3) Install Top Door |
4) Install Pan |
EN-3 |
Pioneer™ |
2.3 Level Balance
Level the balance on a firm, steady surface. Avoid locations with excessive air current, vibrations, heat sources or rapid temperature changes.
2.4 Connect Power
For use with CSA Certified (or equivalent approved) power source, which must have a limited circuit output.
2.5 Initial Calibration
Without InCal – Power on the balance by pressing 0/T. Press and hold Print/Cal until [CAL] is displayed. The display flashes the calibration mass needed. To select the alternate calibration weight press No. Put the calibration mass on the pan. The display flashes [BUSY], then [CLEAr Pan]. Remove the mass. When calibration is complete, [DONE] is displayed.
InCal – Press and hold Print/Cal until [CAL] is displayed. The display flashes [BUSY], when calibration is
complete, [DONE] is displayed.
Note: Calibrations should be performed after a warm up time of 60 minutes.
3. Operation
Count, APW Optimization, Percentage or specific units of measure must be activated in the MODE or UNIT menus if they are not initially available.
3.1 Weighing Mode
Repeatedly press Unit until the desired unit icon is displayed.
Press 0/T to zero the balance and then place objects to be weighed on the pan.
– Use the Count mode to count parts of uniform weight.
To access Count Mode, press Unit until the display shows [Count].
Establish an Average Piece Weight (APW) – Each time a new type of part is to be counted, the nominal weight of one piece (APW) must be established using a small quantity of pieces.
With [CLr.APwW] displayed, press No to use the previously saved APW, or press Yes to establish a new APW. The display indicates the number of pieces to be used to establish the new APW. If a different sample size is preferred, press No until the desired sample size is displayed (5, 10, 20, 50 or 100). Put the specified number of pieces on the pan. Press Yes to accept new APW or No to abort.
Count – Place the quantity to be counted on the pan.
APW Optimization – Since the weight of each piece varies slightly, APW Optimization may be used to increase the
Pioneer™ |
EN-4 |
accuracy of the count. The balance automatically recalculates the Average Piece Weight when the number of parts on the pan is less than three times the original sample size. The display shows [APwW.OPT] each time the APW is optimized.
3.3 Percent Mode – Use this mode to measure the weight of a sample as a percentage of a reference weight. To access the Percent Mode, press Unit until the display shows [PErCEnt].
Establish a new Reference Weight - With Clear reference [CLr.rEF] displayed, Press No to use the previously saved Reference Weight. Press Yes to establish a new Reference Weight. Put the reference sample on the pan and press Yes to accept or No to abort.
Percent – Place the object(s) to be compared to the reference weight on the pan.
3.4 Weigh Below Feature
Remove Weigh Below Cover |
Attach wire or string to Hook |
Suspend sample |
4. SETTINGS
4.1 Menu Navigation
|
|
|
|
|
|
.END. |
|
|
.CAL. |
.SETUP. |
.MODE. |
.UNIT. |
.PRINT. |
.RS232. |
.LOC. |
|
InCal* |
AZSM |
COUNT |
GRAM |
OUT STA |
BAUD |
CAL |
|
SPAN |
FILTER |
APWOPT |
MILLIGRAM |
A. PRINT |
PARITY |
SETUP |
|
LINEAR** |
A.TARE |
PERCENT |
KILOGRAM |
CONTENT |
HAND.SH |
MODE |
|
CAL ADJ |
GRAD |
END.MOD |
OUNCE |
LINE FO |
END.R232 |
UNIT |
|
END.CAL |
GBL RST |
|
POUND |
END.PRT |
|
|
|
|
END.SET |
|
NOT ALL |
|
|
|
|
|
|
UNITS |
|
|
RS232 |
|
|
|
|
NO |
SHOWN |
|
|
|
* |
InCal models only |
|
|
|
|
|
|
|
|
|
|
|
|
||
** Not available in InCal models |
YES |
END.UNIT |
|
|
END.LOCK |
||
|
|
|
|
|
Enter Menu – When the balance is on, press and hold Unit/Menu until [mMENU] appears. Release the button and the Calibrate [.CAL.] menu will display.
Menu Navigation – Select menus, menu items and settings through use of the Yes and No buttons. Solid arrows point to the content displayed when Yes is pressed, Dashed lines when No is pressed. Changing Settings – To select the displayed Setting, press Yes. To move to the next Setting, press No.
Exit Menu – When [.END.] is displayed, press Yes to exit the menu function, or press No to return to the Cal menu. Note: Press and hold No at any time to exit quickly.
EN-5 |
Pioneer™ |
4.2 Calibration Menu [.CAL.]
InCal or Span calibration should be performed daily and when the room temperature changes. InCal [InCal] calibrates the balance using an internal mass.
Span Calibration [SpaN] uses two weight values: zero and a weight between 50% and 100% of the capacity of the balance.
Linearity calibration [LIN] uses three weight values zero, 50% of capacity and full capacity. Generally this calibration is not required unless testing shows that the linearity error exceeds the Linearity tolerance in the Specification table. (Not available in InCal models)
Calibration Adjust [CAL Adj] allows adjustment to the result of the internal calibration by +/- 99 divisions. (InCal models only)
4.3 Setup Menu [.setup.]
Automatic Zero-Setting [azsmM] – Environmental changes can cause the display to drift. The Automatic ZeroSetting Mechanism (AZSM) is designed to keep the balance set at zero, despite these slight changes. (OFF, SET.5d, SET 1d, SET 2d, SET 5d)
Filter [filter] – Use the low setting (SET LO) when environmental disturbances are not present. Use the medium setting (SET MED) for normal environments. Use the high setting (SET HI) when vibrations or air currents are present.
Auto Tare [A-tare] – The initial item placed on the balance is assumed to be a container so it is zeroed out. The next item is then weighed. When the pan is cleared the balance resets, waiting for a container. (SET OFF, SET ON)
Grad [Grad] – Select the readability displayed. Reducing the readability may be needed for approval. (SET 1d, SET [1]d, SET 10d)
Global Reset [Gbl rst] – Resets all settings to factory default values. (RESET)
4.4 Mode Menu [.mMode.]
Count Mode [Count] - [SET ON or OFF]
Average Piece Weight (APW) Optimization [aPwW.OPT] - (SET ON, SET OFF) Percent Mode [PerCent] - (SET ON, SET OFF)
4.5 Unit Menu [.Unit.]
The Unit menu is used to enable or disable a specific unit. (SET ON, SET OFF) The unit is indicated by a small character next to Unit in the display (g = grams). The default setting is Grams SET ON and all other units SET OFF.
T-Units – When unit [t] displays; press Yes to show the T-Unit settings; SET OFF, SET TT (Taiwan Tael), SET TH (Hong Kong Tael), SET TS (Singapore Tael), SET TO (Tola) or SET TI (Tical).
M-Units – When unit [m] is displayed, press Yes to show the M-Unit settings; SET OFF, SET MO (Momme) or SET ME (Mesghal).
Custom Unit – Custom Unit (C) is used to create a unit of measure not provided with the balance. The Custom Unit is defined by a factor, a multiplier (E) and a least significant digit (LSD). The balance will use this to convert grams to a custom unit of measure. (Example: 1 gram = 0.257206 Avoirdupois Dram, using a 4100g x 0.01g balance)
To create a custom unit, press Yes when unit [c] is displayed.
Factor – The Factor (F) is a value from 0.1000000 to 1.999999. When the Factor is displayed, the first digit is flashing. Press Yes to accept its value and activate the next digit, or No to edit. When editing, press No until the desired value appears, then press Yes to accept. Repeat until all digits have been accepted. When the Factor flashes on the display, press Yes to accept or No to re-edit. (Example: F = 0.257206)
Multiplier (E) – The settings are, [E 0] (Fx1), [E 1] (Fx10), [E 2] (Fx100), [E 3] (Fx1000), [E-3] (F/1000), [E-2] (F/100), and [E-1] (F/10). Press No to display the next setting, Yes to accept. (Example: E = 0).
Note: The multiplier selections are limited when the capacity in grams is exceeded.
Pioneer™ |
EN-6 |
LSD – The Least Significant Digit (LSD) is the number of displayed divisions (d) by which the weight is incremented. The values are 1d, 2d, 5d, 10d, 100d or 0.5d. Press No to go to the next setting, press Yes to accept. (Example LSD = 1d) Note: LSD options are limited if the readability in grams is exceeded.
The example custom unit will display [0.26 C] when 1g is placed on the pan.
4.6 Print Menu [.Print.]
Output Stable [Out.Stab] – Data will only be sent when the Stable indicator is on. This setting works with manual button pressing or continuous and interval Auto Print. (SET ON, OFF)
Auto Print [A.Print] – Data will be continuously sent when [Continu] is set. Interval [Inter] sends data every 1 to 3600 seconds. When Stable [Stable] will send data when the balance detects a new stable reading. This can be a weight value only [LOad] or it can also include a stable zero [L+zero]. [OFF] disables automatic printing.
Content [CONteNt] – The content in the data transmission can be modified. Each of the following settings can be set on or off. Number Only [numMber] will only send the numeric result. Balance ID [Bal. ID] will add the Balance
serial number for traceability purposes. Reference [refer] will add reference information relevant to the current mode. GLP [GLP] will send additional items to allow proper documentation of laboratory results.
Line Format [LinE Fo] – Single line format [single] will put all the data in one line separating each output with a comma (,). Multi line format [mMulti] will put each data output on a new line. Multi +4 [mM-4LF] will add 4 line spaces between each output.
^^^^^^^^^^^^^^^^^^^^^
|
|
|
Multi Line Format with 4 line feeds |
|
|
|
Line Feed-2 |
|
|
|
Line Feed-3 |
|
|
|
Line Feed-4 |
User ID: ………………........ |
|
|
GLP (ON) |
Bal ID: 123456789 |
|
|
Balance ID (ON) |
Proj ID: …………………… |
|
|
GLP (ON) |
Time: ……:…….:……. |
|
|
GLP (ON) |
Date: …../ ….. / ….. |
|
|
GLP (ON) |
120.01 g |
|
|
Result |
^^^^^^^^^^^^^^^^^^^^^
4.7 RS232 Menu [.rs232.]
Baud [BAUD] – The RS232 baud rate can be set to 600, 1200, 2400, 4800, 9600 and 19200. Parity [Parity] – Parity can be sent to 7 bits-even parity [7 eveN], 7bits-odd parity [7 odd],
7bits-no parity [7 No] or 8bit-no parity [8 No].
Handshake [Handsh] – Handshake can be set to off [Off], X on – X off [on-off], or hardware [HardwWr].
4.8 Lock Menu [.LOC.]
When a Lock Menu item is SET ON the indicated menu cannot be changed.
[Loc Cal] – Calibration, [Loc SEt] – Setup, [Loc.mMod] – Mode, [Loc.Unit] – Unit, [Loc Prt] – Print, [Loc 232] – RS232.
4.9 Sealing Access to the Balance Settings
The Menu Lock switch prevents changes to the Lock Menu. The switch can be secured using paper seals, wire seals or plastic ties.
Un-Locked |
Locked with Plastic Tie |
EN-7 |
Pioneer™ |
5.0MAINTENANCE
5.1Troubleshooting
Symptom |
Possible Cause |
Remedy |
|
|
|
Cannot turn on |
No power to balance |
Verify connections and voltage. |
Poor accuracy |
Improper calibration |
Perform calibration |
|
Unstable environment |
Move balance to suitable location |
Cannot calibrate |
Unstable environment |
Move the balance to suitable location |
|
Incorrect calibration masses |
Use correct calibration masses |
Cannot access mode |
Mode not enabled |
Enter menu and enable mode |
Cannot access unit |
Units not enabled |
Enter menu and enable units |
Err 5 |
Average Piece Weight too small |
Add additional samples |
Err 7.0 |
Time out |
|
Err 8.1 |
Pan has load during power on |
Remove weight from pan and re-zero. |
Err 8.2 |
Pan was removed prior to power on |
Install pan and re-zero. |
Err 8.3 |
Weight on pan exceeds capacity |
Remove weight from the pan |
Err 8.4 |
Pan was removed during weighing |
Re-install pan |
Err 9.5 |
Factory calibration data corrupted |
Contact the authorized dealer |
Err 9.8 |
Factory calibration data corrupted |
Perform calibration |
Error 53 |
EEPROM Checksum error |
Contact the authorized dealer |
REF Err |
Reference Weight is too small |
Add additional samples |
LOWREF |
Reference Weight is too low for accurate |
Add additional samples or continue to weigh |
|
parts counting or percent weighing. |
with less accurate results. |
5.2 Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. Please visit our web site, www.ohaus.com to locate the Ohaus office nearest you.
5.3 Accessories
Security device |
76288-01 |
|
Auxiliary Display |
PAD7 |
|
Density Determination Kit |
Contact Ohaus |
(0.1mg and 1mg models only) |
Printer – Thermal |
Contact Ohaus |
|
Printer - Impact |
Contact Ohaus |
|
Cable Kit – Thermal Printer |
Contact Ohaus |
|
Cable Kit - Impact Printer |
Contact Ohaus |
|
Data collection Software |
SW12W |
|
Pioneer™ |
EN-8 |
6. TECHNICAL DATA
Ambient conditions – The technical data is valid under the following ambient conditions: Ambient temperature: 10° C to 30° C
Relative humidity: 15 % to 80 % at 31° C non-condensing, decreasing linearly to 50% at 40° C Height above sea level: Up to 2000 m
Operability is assured at ambient temperatures between 5° C and 40° C
Power – AC adapter, Balance power input 8-14.5 VAC, 50/60Hz 4VA or 8-20 VDC, 4W Protection – Protected against dust and water, Pollution degree: 2, Installation category: Class II
6.1 Drawings
0.01g and 0.1g Models
0.0001g and 0.001g Models
EN-9 |
|
|
|
|
|
|
|
|
|
|
|
Pioneer™ |
|
6.2 Specifications |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||
MODEL |
PA64* |
PA114* |
PA214* |
PA213* |
PA413* |
PA512* |
PA2102* |
PA4102* |
PA4101* |
||||
Capacity (g) |
65 |
|
110 |
|
210 |
210 |
410 |
510 |
2100 |
|
4100 |
|
4100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Readability (g) |
|
0.0001 |
|
|
0.001 |
|
0.01 |
|
|
|
0.1 |
||
Repeatability (sd) |
|
0.0001 |
|
|
0.001 |
|
0.01 |
|
|
|
0.1 |
||
Linearity (g) |
|
± 0.0002 |
|
± 0.002 |
|
± 0.02 |
|
|
|
± 0.1 |
|||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Tare Range |
|
|
|
|
|
To capacity by subtraction |
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Stabilization |
|
|
|
|
|
|
3 seconds |
|
|
|
|
|
|
Span cal mass (g) |
50 or |
|
50 or |
|
100 or |
100 or |
200 or |
200 or |
1000 or |
|
2000 or |
|
2000 or |
|
60 |
|
100 |
|
200 |
200 |
400 |
500 |
2000 |
|
4000 |
|
4000 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Lin. cal masses (g) |
25, |
|
50, |
|
100, |
100, |
200, |
250, |
1000, |
|
2000, |
|
2000, |
|
50 |
|
100 |
|
200 |
200 |
400 |
500 |
2000 |
|
4000 |
|
4000 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Pan size (in / cm) |
|
3.5 / 9 dia. |
|
4.7 / 12 dia. |
|
7.1 / 18 dia. |
|
|
|||||
|
|
|
|
|
|
|
|
|
|||||
Net Weight (lb/kg) |
|
10.1 / 4.6 (*11.2 / 5.2) |
|
|
10 / 4.5 |
(*11.1 / 5.1) |
|
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
*C= Internal calibration
6.3 Communication
6.3.1 Commands
The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed weight. The balance will return “ES” for invalid commands.
Command |
Function |
IP |
Immediate Print of displayed weight. |
|
|
P |
Print displayed weight (uses Stable ON/OFF menu settings). |
|
|
CP |
Continuous Print. |
xP |
Interval Print x = Print Interval (1-3600 sec) |
|
|
T |
Same as pressing Zero Key. |
ON |
Turns balance ON. |
|
|
OFF |
Turns balance OFF. |
PSN |
Show Serial Number. |
|
|
PV |
Version: Print product name, software revision and LFT ON (if LFT is set ON). |
|
|
PU |
Print current mode/unit |
x# |
Set PC ref wt (x) in grams |
|
|
P# |
Print PC ref wt |
x% |
Set % ref wt (x) in grams |
|
|
P% |
Print % ref wt |
6.3.2 RS232 (DB9) Pin Connections
Pin 2: Balance transmit line (TxD) Pin 3: Balance receive line (RxD) Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS) Pin 8: Request to send (hardware handshake) (RTS)
GND |
|
RxD |
|
||
|
|
TxD
5 1
9 6
CTS
RTS