Ohaus D60BR, D60VL, D60VR, D60BL, D30VR User Manual

...
0 (0)
Ohaus D60BR, D60VL, D60VR, D60BL, D30VR User Manual

Defender3000 Series Base

Instruction Manual

Defender 3000 Series Base EN-1

INTRODUCTION

This manual covers installation, and maintenance instructions for the Ohaus Defender TM 3000 Series Base. Please read this manual completely before installation and operation.

SAFETY PRECAUTIONS

For safe and dependable operation of this product, please comply with the following precautions:

Operate the base only under ambient conditions specified in these instructions

Ensure that the load cell cable does not pose an obstruction or tripping hazard

Do not operate the base in hazardous environments or unstable locations

Do not drop loads on the base

Do not lift the base by the top frame; always lift from the bottom frame when moving the base

Service should only be performed by authorized personnel

INSTALLATION

Unpacking

Unpack and inspect the product to make sure that all components have been included. The package includes the following:

Base Weighing Pan Warranty Card Instruction Manual When purchased as a complete Defender 3000 Series Scale, will also include:

3000 Series Indicator Column Assembly

Assembly

Weighing Pan

Place the weighing pan securely over the rubber load pads on the top frame of the base.

Wiring Connections

Connect the load cell cable to an indicator using the wiring codes in Table 1 below. Refer to the indicator manual for information about load cell connections, setup and calibration.

Note: When purchased as a Defender Series Scale, the base is already pre-wired to the 3000 Series Indicator.

TABLE 1. SIX WIRE LOAD CELL CONNECTION.

FUNCTION

WIRE COLOR

+ Excitation

Green

– Excitation

Black

+ Signal

Red

– Signal

White

+ Sense

Blue

– Sense

Brown

Shield

Yellow

Selecting the Location

To ensure accuracy, proper performance and safety, locate and operate the base on a stable, level surface. Avoid locations with rapid temperature changes or excessive dust, air currents, vibrations, electromagnetic fields or heat.

Level the base by adjusting the four leveling feet until the bubble in the level indicator (located at the rear of the base) is centered. A wrench may be needed to loosen the locking nut above each leveling foot. When the base is level, retighten the locking nuts up against the base to lock each foot into place.

Note: Ensure that the base is level each time its location is changed.

CORRECT INCORRECT