Ohaus AS User Manual

5 (2)

Ohaus Corporation

29 Hanover Road Florham Park NJ 07932-0900

ANALYTICAL Standard

Electronic Balances

AS Series

Instruction Manual

NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND TO COMPLY WITH THE LIMITS FOR A CLASS A DIGITAL DEVICE, PURSUANT TO PART 15 OF THE FCC RULES.

THESE LIMITS ARE DESIGNED TO PROVIDE REASONABLE PROTECTION AGAINST HARMFUL INTERFERENCE WHEN THE EQUIPMENT IS OPERATED IN A COMMERCIAL ENVIRONMENT. THIS EQUIPMENT GENERATES, USES, AND CAN RADIATE RADIO FREQUENCY ENERGY AND, IF NOT INSTALLED AND USED IN ACCORDANCE WITH THE INSTRUCTION MANUAL, MAY CAUSE HARMFUL INTERFERENCE TO RADIO COMMUNICATIONS. OPERATION OF THIS EQUIPMENT IN A RESIDENTIAL AREA IS LIKELY TO CAUSE HARMFUL INTERFERENCE IN WHICH CASE THE USER WILL BE REQUIRED TO CORRECT THE INTERFERENCE AT HIS OWN EXPENSE.

THIS DIGITAL APPARATUS DOES NOT EXCEED THE CLASS A LIMITS FOR RADIO NOISE EMISSIONS FROM DIGITAL APPARATUS AS SET OUT IN THE INTERFERENCE-CAUSING EQUIPMENT STANDARD ENTITLED “DIGITAL APPARATUS”, ICES-003 OF THE DEPARTMENT OF COMMUNICATIONS.

CET APPAREIL NUMERIQUE RESPECTE LES LIMITES DE BRUITS RADIOELECTRIQUES APPLICABLES AUX APPAREILS NUMERIQUES DE CLASSE A PRESCRITES DANS LA NORME SUR LE MATERIEL BROUILLEUR : “APPAREILS NUMERIQUES”, NMB-003 EDICTEE PAR LE MINISTRE DES COMMUNICATIONS.

Unauthorized changes or modifications to this equipment are not permitted.

2

TABLE OF CONTENTS

 

INTRODUCTION ..................................................................................................

5

DESCRIPTION .....................................................................................................

5

UNPACKING ........................................................................................................

6

INSTALLATION ....................................................................................................

7

Environment ..................................................................................................

7

Draft Shield and Spill Ring ............................................................................

7

Pan ................................................................................................................

8

AC Adapter ....................................................................................................

8

Leveling the Balance .....................................................................................

8

OPERATION ........................................................................................................

9

Turning the Balance ON ................................................................................

9

Warm Up .......................................................................................................

9

Moveable FineRange TM Model (AS260D) ......................................................

9

Checking Calibration ...................................................................................

10

Weighing ....................................................................................................

10

Zero/Tare ....................................................................................................

11

USING MENUS TO CONFIGURE THE BALANCE ...........................................

12

CALIBRATION MENU ........................................................................................

13

Calibration Menu Protection ........................................................................

13

Calibration Weights .....................................................................................

13

Span Calibration ..........................................................................................

14

Linearity Calibration ....................................................................................

15

End ..............................................................................................................

16

USER MENU ......................................................................................................

17

User Menu Protection .................................................................................

17

Reset to Factory Defaults ............................................................................

18

Averaging Level ..........................................................................................

19

Stability Range ............................................................................................

20

Auto-Zero ....................................................................................................

20

End .............................................................................................................

21

3

SETUP MENU ....................................................................................................

22

Setup Menu Protection ................................................................................

22

Type Approved Balance ..............................................................................

23

Lockswitch ...................................................................................................

24

End ..............................................................................................................

25

MENU LOCK-OUT PROTECTION .....................................................................

26

TYPE APPROVED BALANCE SEALING ...........................................................

27

CARE AND MAINTENANCE ..............................................................................

28

TROUBLESHOOTING .......................................................................................

28

Error Codes .................................................................................................

29

SERVICE INFORMATION .................................................................................

30

SPECIFICATIONS .............................................................................................

31

PARTS INFORMATION .....................................................................................

32

REPLACEMENT PARTS ...................................................................................

32

LIMITED WARRANTY ........................................................................................

33

4

INTRODUCTION

This manual covers installation, operation and troubleshooting for the Ohaus Analytical Standard balances, Models AS60, AS120, AS200 and AS260D. Type approved Analytical Standard balances, Models AS60E, AS120E and AS200E are setup to conform to OIML, EC and U.S. regulations. To insure proper operation of the balance, please read this manual completely.

DESCRIPTION

Ohaus Analytical Standard series balances are precision weighing instruments, designed to provide years of service with virtually no maintenance. The Analytical Standard series is constructed using a die-cast aluminum base finished with a durable epoxy powder paint which is resistant to commonly used acids, contains a one piece solid-state precision electronics PC board and a seven digit LCD display which is 0.6 inches in height. All Analytical Standard series electronic balances are factory set to measure in grams. To prevent measurements from being affected by air currents, a draft shield is mounted to the balance. A stainless-steel spill ring is removable for cleaning in the event of accidental spills. Power is supplied through an AC Adapter which is available in five configurations for world-wide usage. Accessories include: an RS232 interface kit which allows printing of results through an external computer, an RS232 interface cable with a print switch, a security device and calibration weights.

ANALYTICAL STANDARD Balance

5

UNPACKING

Your Analytical Standard balance was shipped with the following items:

a pan

an AC power adapter

a draft shield and spill ring

this instruction manual

your warranty card

It is recommended to save the carton and packing material for storing, transporting the balance or returning it for service.

Analytical Standard

Balance

Pan

AC Adapter

 

Draft Shield and

 

 

Spill Ring

 

 

6

Ohaus AS User Manual

INSTALLATION

Environment

The balance should always be used in an environment which is free from excessive air currents, corrosives, vibration, and temperature or humidity extremes. These factors will affect displayed weight readings.

DO NOT install the balance:

Next to open windows or doors causing drafts or rapid temperature changes.

Near air conditioning or heat vents.

Near vibrating, rotating or reciprocating equipment.

Near magnetic fields or equipment that generates magnetic fields.

On an unlevel work surface.

Draft Shield and Spill Ring

To install the draft shield and spill ring:

1.Position the draft shield on top of balance as shown (approximately degrees counterclockwise from base of the balance).

2.Looking down through the top of shield, line up the hole in the bottom of the shield with the hole in the ance weighing mechanism.

3.Put the draft shield down, fitting holes, and turn the shield clockwise until it locks into place.

4.Make sure that the draft shield firmly locked in place.

5.Install the stainless steel spill ring inside of the draft shield with the raised surface facing up and correctly oriented.

CAUTION:

Never remove the draft shield with the pan in place.

7

Pan

Place the pan into the hole in the weighing

 

mechanism.

PAN

SPILL RING

AC Adapter

Plug the molded connector of the AC Adapter into the receptacle at the rear of the balance. Plug the AC Adapter into a convenient ac outlet. When power is applied to the balance, it will begin a 60 second self test cycle. During this time, if the balance is turned ON, the display will count down from 60 and display the word CHEC.

Leveling the Balance

The balance is equipped with a level indicator on the rear and two adjustable leveling feet. Adjust the leveling feet until the bubble appears in the center circle of the indicator.

LEVELING FEET (2)

AC ADAPTER

RECEPTACLE

LEVEL INDICATOR

8

OPERATION

Turning the Balance ON

With no load on the pan, switch the bal-

ance ON by pressing

. When first

switched ON, all segments of the display should be on as shown in the illustration.

This display check will be displayed briefly, then the model number of the balance followed by a software revision number.

Warm Up

Before initally using the balance, allow time for it to adjust to changes in environment. The balance need only be plugged in to warm up. Recommended warm up period is 30 minutes.

Moveable FineRangeModel (AS260D)

The AS260D offers both a fine range (lower capacity/higher readability) and a coarse range (higher capacity/lower readability). When first turned on, the balance is in the fine range. It remains in this range until the weight on the pan exceeds the fine range capacity. When weight on the pan is greater than the fine range capacity, the balance switches to the coarse range.

For balance with Moveable FineRange, please note:

If the weight of the object on the platform exceeds the limit of the Moveable FineRange, the balance will automatically change to the coarse range until either:

1. The load (excluding tare) is reduced to below the limit of the fine range.

-OR-

2. is momentarily pressed, which tares the balance and reactivates the fine

range.

9

Checking Calibration

Before using the balance, it should be calibrated. The balance has been calibrated before shipment, however, calibration is influenced by factors such as:

Variations in the earths gravitational field at different latitudes of the world.

Rough handling.

Changes in work location.

Height above sea level.

Environmental conditions.

To check the balance’s calibration, place a known mass on the center of the pan and read the displayed weight.

If the displayed weight differs from the known mass by more than acceptable limits, refer to the Calibration Menu and the Specifications at the rear of the manual.

Weighing

1. Press

to zero the display.

STABILITY

INDICATOR

2.Place the object(s) or material to be weighed on the pan.

3.Wait for the stability indicator to appear before reading the weight.

10

Zero/Tare

When weighing material or objects that must be held in a container, O/T enables you to store the container weight in the balance’s memory, separate from the weight of the material in the container.

1.Place an empty container on the pan. Its weight will be displayed.

2. Press

.

The display will show zero and the container’s weight will be stored in memory.

3.Add material to the container. As material is added, its net weight will be displayed.

4.Removing the container and material from the pan will cause the balance to display the container’s weight as a negative number.

The weight remains in balance memory

until

is pressed again.

11

USING MENUS TO CONFIGURE THE BALANCE

Analytical Standard balances contain four display menus which enable you to calibrate and configure the balance for your specific operating requirements.

Calibration Menu: Used to calibrate the balance for span or linearity.

User Menu: Used to adapt the balance to environmental conditions.

Setup Menu: Used to enable or disable different balance features.

Print Menu: Used to configure the RS232 Interface.

Functions not allowed on verified balances have shaded backgrounds.

CAL

 

USER

 

 

SETUP

 

 

 

SPAN

 

 

RESET

 

 

LFT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes, No

 

 

On,

Off

 

 

 

 

LIN

 

 

AL

 

 

LOCSW

 

 

 

 

 

 

 

 

END

 

 

 

 

 

 

 

 

 

 

 

 

0, 1, 2, 3

 

 

 

 

 

 

 

Calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STB

 

 

 

 

 

User

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Setup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

.5d, 1d, 2d, 5d

 

 

 

 

 

Print

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AUTO-0

 

 

 

 

 

End

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

END

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OFF , .5d ,

1d, 3d

 

 

 

 

 

 

 

 

 

 

PRINT

NOTE:

PRINT menu appears only when RS232 option is installed.

END

To access a menu, press and hold

until desired menu appears, then release it.

Original factory default settings are shown in boldface type.

Use these buttons to step through menus and select submenus:

next selection

select

displayed item

 

12

CALIBRATION MENU

Analytical Standard balances can be calibrated in two ways: Span calibration or Linearity calibration. Span calibration resets the balance’s weighing range using two weight values: zero and a weight value at or near the balance’s capacity. Linearity calibration minimizes deviation between actual and displayed weights within the balance’s weighing range. Three weight values are used: zero, a weight value within the balances weighing range, and a weight value at or near the balance’s specified capacity. The following table shows the sequence in which submenus appear on the Calibration menu.

CALIBRATION MENU TABLE

Selects span calibration.

Selects linearity calibration.

Used to exit the Calibration menu.

Calibration Menu Protection

Calibration may be locked out to prevent unauthorized personnel from changing calibration. To lock out calibration menu, refer to the section titled Menu Lock-Out Protection.

NOTE: If calibration has been locked out, you will not be able to access it.

Calibration Masses

Before beginning calibration, make sure masses are on hand. If you begin calibration and realize masses are not available, either turn the balance off, or go through the procedure without masses. The balance will use previously stored calibration data. Calibration should be performed as necessary to ensure accurate weighing. Masses required to perform the procedures are listed in the adjacent table.

CALIBRATION MASSES

 

LINEARITY

SPAN ONLY

MODEL

MASSES

MASS

AS60

20g,

50g

50g

AS120

50g,

100g

100g

AS260D

100g, 200g

200g

AS200

100g, 200g

200g

Masses must meet or exceed ASTM Class 1 Tolerance. Calibration masses are available as accessories.

13

Span Calibration

1. Press and hold

until CAL is

displayed, then release it. Balance will display SPAN.

2.

Press

to start the Span calibra-

 

tion procedure.

3.

When

is released, C 0g will

 

be displayed indicating that no weight

 

should be on the pan.

4.

Press

. The display will show

-C- momentarily, followed by the value of the weight which must be placed on the pan. Do not disturb the balance when -C- is displayed. Disturbances will result in improper calibration.

5.Place the required weight on the pan

and press

. The display will

show -C- momentarily while the balance recalibrates.

6.When the weight on the pan is displayed along with the current unit indicator, the balance is recalibrated.

14

Linearity Calibration

1. Press and hold

until CAL is

displayed, then release it. Balance will display SPAN.

2.

Press

and the display will

 

show LIN.

 

3.

Press

to start the Linearity

 

Calibration Procedure.

4.

When

is released, C 0g will be

 

displayed, indicating that no weight

 

should be in the pan.

5.

Press

. The display will show

-C- momentarily, followed by the value of the weight which must be placed on the pan. Do not disturb the balance when -C- is displayed. Disturbances will result in improper calibration.

6.Place the required weight on the pan.

15

7. Press

. The display will show

-C- momentarily, then C followed by the next weight to be placed on the pan.

8.Place the required weight on the pan,

then press

. The display will

show -C- momentarily, while the bal-

ance recalibrates.

9. When the weight on the pan is dis-

played along with the current indica-

tor, the balance is recalibrated.

End

If you have entered the Calibration menu and do not wish to calibrate the balance, use END to return to normal weighing operations.

Repeately press

until End is dis-

played.

 

 

Press

, when released, the balance

will returned to normal weighing opera-

tions.

16

USER MENU

The User menu is used to adapt the balance to environmental conditions. It contains submenus which enable you to reset the balance to factory default settings or to select specific range settings. Access to the User menu can be disabled using the Lock out switch. The following table shows the sequence in which submenus appear on the User menu.

USER MENU TABLE

Sets all submenus below to original factory default settings. Reset does not appear if menu has been locked out.

Specifies the averaging level.

Specifies the desired stability range.

Sets Auto-Zero threshold.

Used to exit the Setup menu and store your selections.

User Menu Protection

The User menu may be locked out to prevent unauthorized personnel from changing the settings. To lock out the User menu, refer to the section titled Menu Lock-Out Protection.

NOTE: If -SAFE- is displayed, the User menu has been locked out. Settings may be viewed but not changed. See the Menu Lock-Out Protection section to enable it for making changes.

To access the User menu, press and hold

until USER is displayed, then re-

lease it.

17

To access a submenu:

 

1.

Repeatedly press

until the de-

 

sired submenu is displayed.

2.

Press

to select the displayed

 

submenu.

 

 

NOTE: You must use END to store any changes you make to the User menu.

The following sections describe each item on the User menu in detail.

Reset to Factory Defaults

This submenu enables you to reset all User menu selections to the factory default settings outlined in the adjacent table.

To reset to factory defaults:

1.Access the RESET submenu.

2. Press

to change the setting.

3.Select YES to reset settings or, no to leave current settings.

4. Press

to accept the displayed

setting.

USER MENU

FACTORY DEFAULTS

Averaging Level

AL 1

Stability Range

1d

Auto-Zero Tracking

.5d

 

 

 

 

 

 

 

 

18

Averaging Level

Averaging level compensates for vibration or excessive air currents. During operation, the balance continually takes weight readings from the weighing cell. Successive readings are then digitally processed to achieve a stabilized display. Use this submenu to specify how much processing you need to obtain stable results.

NOTE: Averaging level does not affect balance accuracy.

Select one of four averaging levels using the adjacent table as a guide.

To view or change the averaging level:

1.Access the AL submenu to display the current setting.

2.

Press

to change the setting.

3.

Press

to accept the displayed

 

setting.

 

4.

When

is released, AL will be

 

displayed again and the Setup menu

 

will be returned.

AVERAGING LEVEL

AL 0 reduced stability, fastest stabilization time

AL 1 normal stability, normal stabilization time

AL 2 more stability, slow stabilization time

AL 3 maximum stability, slowest stabilization time

19

Stability Range

The stability range specifies how much a displayed weight may change while the stability indicator remains ON. When displayed weight changes beyond the allowable range, the stability indicator turns OFF indicating an unstable condition. Analytical Standard balances permit you to select one of four stability ranges (in divisions) as shown in the table.

When the RS232 interface is configured to print stable data only, the stability range also governs data output. Displayed data will only be output if it is within the selected stability range.

To view or change the stability range:

1.Access the Stb submenu to display the current setting.

2.

Press

to change the setting.

3.

Press

to accept the displayed

 

setting.

 

4.

When

is released, Stb will be

 

displayed again and the Setup menu

 

will be returned.

Auto-Zero

Auto-Zero minimizes the effects of temperature changes and shift on the zero reading. By defining a threshold level in divisions, the balance maintains the zero display until the threshold is exceeded. This submenu permits you to select one of three threshold levels, or turn the feature OFF. Auto-Zero only functions when the display reads zero.

STABILITY RANGE

.5d smallest range: stability indicator is ON only when displayed weight is within .5 divisions

1d reduced range

2d normal range

5d largest range: stability indicator is ON even though displayed weight changes slightly

20

To view or change the Auto-Zero setting:

1.Access the Auto-0 submenu to display the current setting.

2.

Press

to change the setting.

3.

Press

to accept the displayed

 

setting.

 

4.

When

is released, Auto-0 will

be displayed again and the User menu will be returned.

End

You must use END to exit the User menu.

Changes you make in the User menu are only stored in memory if you use END.

1.To exit the User menu and store your

settings, press

when End is

displayed.

 

 

2. When

is released, the balance

will be returned to normal weighing operations.

AUTO ZERO

OFF turns Auto-Zero OFF

.5d sets threshold to .5 divisions

1d sets threshold to 1 division

3d sets threshold to 3 divisions

21

SETUP MENU

The Setup menu enables you to retain program balance parameters once they have been set. Access to the Setup menu can be disabled using the Lock out switch. The following table shows the sequence in which submenus appear on the Setup menu.

SETUP MENU TABLE

Sets balance for type approved operation.

Enables individual or all menus to be locked out.

Used to exit the Setup menu and store your selections.

Setup Menu Protection

The Setup menu may be locked out to prevent unauthorized personnel from changing settings. To lock out the Setup menu, refer to the section titled Menu Lock-Out Protection.

NOTE: If -SAFE- is displayed, the Setup menu has been locked out. Settings may be viewed but not changed. See the Menu Lock-Out Protection section to enable it for making changes.

To access the Setup menu, press and

hold

until SETUP is displayed, then

release it.

22

To access a submenu:

 

1.

Repeatedly

press

until the

 

desired submenu is displayed.

2.

Press

to select the displayed

 

submenu.

 

 

NOTE: You must use END to store any changes you make to the Setup menu.

The following sections describe each item on the Setup menu in detail.

Type Approved Balance

This parameter can be set to ON or OFF. Selecting ON automatically sets certain parameters for type approved requirements.

1.Access the LFT submenu.

2.The display will show the current status (ON/OFF). If ON is selected, Auto-Zero Tracking will be limited to 0.5d or OFF and Lock switch default becomes Setup.

3.

Press

to change the status.

4.

Press

to accept the status.

5.

When

is released, the display

will show LFT.

23

Lockswitch

Lockswitch enables you to lock out one or more menu selections. Each menu can be individually locked YES, or unlocked NO. Set the appropriate balance functions, and then decide which functions of the balance are to be locked. The Calibration, User, and Setup menus can be individually locked by selecting YES for the appropriate menu, and then locked by the switch located under the right hand side of the control panel.

1.Access the LOCSW submenu. When

is released, the LOCSW

submenu is displayed.

2.To access one or more menus, press

to select the calibration menu, -CAL- is displayed.

 

NOTE: Pressing

changes the

 

selection to the other menus.

 

3.

To select a YES or NO, press

.

 

NOTE: The

switch acts as a

 

toggle and can select either YES or

 

NO.

 

 

 

4.

To

confirm

your selection,

 

press

again. The display indi-

cates the last menu you were in.

24

5.To lock out the other menus, press

and repeat the procedure in

steps 3 and 4.

End

You must use END to exit the Setup menu. Changes you make in the Setup menu are only stored in memory if you use END.

1.To exit the Setup menu and store

your settings, press

when END

is displayed.

 

 

2. When

is released, the balance

will be returned to normal weighing operations.

25

MENU LOCK-OUT PROTECTION

Access to the Calibration, User, and Setup menus, can be disabled using the lock out switch located under the right side of the balance, near the display.

1.Turn the display off and unplug the power cord.

2.Slide the balance toward you, with the front over the edge of a table. (You can also turn the balance on its left side, but if you do, you MUST remove the pan and spill ring first!)

3.Locate hole under display where switch is located.

4.Using a small screwdriver, slide the switch forward for LOCKED or back for UNLOCKED.

5.Plug in the power cord and turn on the balance.

LOCK OUT

SWITCH

LOCK

26

TYPE APPROVED BALANCE SEALING

Analytical Standard Balances, Models AS60E, AS120E and AS200E may be sealed for type approved applications. Factory sealed balances include a bracket, lead seal and wire and security screw as shown in the figure below.

Type approved balances are Class I devices, consult local Weights and Measures officials to determine scaling requirements. If the balance needs to be sealed in the field, a wrench has been provided.

After the balance has been set up properly and the menus are locked out (LFT menu selection set ON), proceed as follows to seal the balance:

1.Turn OFF and unplug the balance. Remove the Draft Shield, Spill Ring and Pan.

2.Turn the balance over and locate the access hole to the lock out switch.

3.Remove the existing screw next to the access hole and discard.

4.Place the Bracket supplied over the access hole and secure with the Security Screw supplied. Tighten with the Wrench supplied.

5.Obtain the Lead Seal and wire. Pass the wire through the Security Screw and the eyelet on the Bracket as shown in the illustration.

6.Crimp the lead seal tightly. Turn the balance over and reinstall items removed.

SECURITY SCREW

LEAD AND WIRE SEAL

BRACKET

BASE OF BALANCE

27

Loading...
+ 62 hidden pages