Mitutoyo 103-192 User manual

3
1
2
A
A
B
C
B
3
C
2
1
x3 - x5
0.01mm
--- 7 mm
--- 0.37 mm+
7.37 mm
--- 6 mm
--- 0.21 mm
--- 0.003 mm
+
6.213 mm
0 5
45
40
35
30
+1 μm
+2 μm
0
8
6 4 2
5
30
25
20
15
10
1 2 3
0.001mm
NG
OK
NG
NG
OK
NG
外 側 マイクロメータ
No. 99MAA001M3
ユーザーズマニュアル
40ºC/104ºF 5ºC/41ºF
25 mm (1”)
ACM BMB BMD BMS CHM CPM GMA GMB LMM M110 M120 M125 M220 M225 M310 M317 M320 M325 M327
OM OMC OMS PMU SPM TMC TMS VM
Outside Micrometer
►Safety Precautions
To ensure operator safety, use this instrument in conformance with the directions and specications
given in this User's Manual. Non-conformance to the instructions may compromise safety.
►Notes on Export Regulations
Do not commit an act which could, directly or indirectly, violate any law or regulation of Japan, your country, or any other international treaty, relating to the export or re-export of any commodities.
en
CAUTION
• Always handle the sharp measuring faces of this product with care to avoid injury.
IMPORTANT
Be sure to perform reference point adjustment by following points to notice below.
-
Do not perfom the adjustment at sites where the temperature will change abruptly. Thermally stabilize the instrument
sufciently at room temperature.
Wipe oil and dust off the both measuring faces and master gage using a gauze or chamois skin.
-
-
Perform reference point setting and measurement in the same posture of the product. Use a constant force device like a ratchet stop to apply a constant measuring force.
-
-
Use a master gage like a setting standard which is periodically calibrated.
After use, take corrosion prevention measures. Corrosion can cause the caliper to malfunction.
KEEP THIS MANUAL FOR FUTURE REFERENCE. For more information, refer to http://manual.mitutoyo.co.jp.
-------------------------------------------------------------------------------------------------------------
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com