Metrohm 719S User Manual

0 (0)

CH-9101 Herisau/Switzerland

E-Mail info@metrohm.com

Internet www.metrohm.com

719 S Titrino

Program version 5.719.0021

Instructions for Use

8.719.1103

03.2002 / pkl

 

 

 

Table of contens

Table of contents

 

1

Introduction ..........................................................................................................

1

 

1.1

Instrument description...........................................................................................

1

 

1.2

Controls and parts.................................................................................................

2

2

Manual operation..................................................................................................

6

 

2.1

Keypad ..................................................................................................................

6

 

2.2

Principle of data input............................................................................................

7

 

2.3

Text input ...............................................................................................................

8

 

2.4

Tutorial ...................................................................................................................

9

2.4.1

Entering data, setting the dialog language....................................................

9

2.4.2

Development of a method, titration of an acid.............................................

11

2.4.3

Storage and loading of methods .................................................................

16

2.4.4

pH calibration...............................................................................................

17

2.4.5

Statistics, acid capacity of drinking water....................................................

18

2.5

Configuration, key <CONFIG>...........................................................................

21

2.6

Selection of the mode, key <MODE> ................................................................

24

2.7

Parameters, key <PARAM> ...............................................................................

25

2.7.1

Parameters for SET ......................................................................................

25

2.7.2

Parameters for MEAS...................................................................................

34

2.7.3

Parameters for CAL......................................................................................

36

2.8

Result calculations...............................................................................................

39

2.9

Statistics calculation ............................................................................................

41

2.10

Common variables ..............................................................................................

43

2.11

Data output..........................................................................................................

44

2.12

Method memory, keys <USER METH> .............................................................

47

2.13

Calibration data, key <CAL.DATA> ...................................................................

49

2.14

Current sample data, key <SMPL DATA> .........................................................

50

2.15

Silo memory for sample data ..............................................................................

51

3 Operation via RS232 Interface...........................................................................

55

3.1

General rules .......................................................................................................

55

3.1.1

Call up of objects .........................................................................................

56

3.1.2

Triggers ........................................................................................................

57

3.1.3

Status messages .........................................................................................

58

3.1.4

Error messages............................................................................................

60

3.2

Remote control commands.................................................................................

63

3.2.1

Overview.......................................................................................................

63

3.2.2

Description of the remote control commands .............................................

76

3.3

Properties of the RS 232 Interface.......................................................................

95

3.3.1

Handshake...................................................................................................

95

3.3.2

Pin Assignment ............................................................................................

99

3.3.3

What can you do if the data transfer does not work? ................................

101

4 Error messages, Troubleshooting...................................................................

103

4.1

Error and special messages..............................................................................

103

4.2

Diagnosis...........................................................................................................

106

4.2.1

General.......................................................................................................

106

4.2.2

Procedure...................................................................................................

106

719 S Titrino

Table of contents

 

4.2.3

Equipment required: ..................................................................................

107

 

4.2.4

Diagnosis steps .........................................................................................

107

 

4.3

Initialise and test RAM.......................................................................................

116

 

4.4

Releasing a locked spindle with inserted Exchange Unit .................................

117

5

Preparations .....................................................................................................

118

 

5.1

Setting up and connecting the instruments ......................................................

118

 

5.1.1

Titrino with Stirrer or Titration Stand...........................................................

118

 

5.1.2

Connection of a printer ..............................................................................

119

 

5.1.3

Connection of a balance............................................................................

120

 

5.1.4

Connection of a Sample Changer .............................................................

121

 

5.1.5

Connection of a recorder...........................................................................

122

 

5.1.6

Connection of a computer.........................................................................

123

 

5.2

Connection of electrodes, preparing titration vessel ........................................

124

6

Appendix...........................................................................................................

127

 

6.1

Technical specifications ....................................................................................

127

 

6.2

Pin assignment of the "Remote" socket ............................................................

129

 

6.2.1

Lines of the "Remote" socket during the titration.......................................

131

 

6.2.2

Possible configurations of the activate pulse in SET and CAL..................

132

 

6.3

User methods....................................................................................................

133

 

6.3.1

General ......................................................................................................

133

 

6.3.2

"Tit.HCl" ......................................................................................................

134

 

6.3.3

"Tit.NaOH" ..................................................................................................

135

 

6.3.4

"p+m_Val"..................................................................................................

136

 

6.3.5

"Br_No".......................................................................................................

137

 

6.4

Validation / GLP.................................................................................................

138

 

6.5

Warranty and conformity ...................................................................................

139

 

6.5.1

Warranty.....................................................................................................

139

 

6.5.2

EU Declaration of conformity.....................................................................

140

 

6.5.3

Certificate of Conformity and System Validation .......................................

141

 

6.6

Scope of delivery and ordering designations ...................................................

142

Index

.........................................................................................................................

 

145

719 S Titrino

1.1 Instrument description

1 Introduction

1.1 Instrument description

The 719 S Titrino is a titrator to perform fast and precise endpoint titrations. Ready-to-start methods for the most common applications are stored in the internal method memory. The operator is free to modify and overwrite this methods or to create and store his own titration sequences.

Data exchange with a PC is possible with the Metrodata VESUV Software and with Metrodata TiNet Software complete remote control, data acquisition and evaluation via PC is enabled.

719 S Titrino

1

 

1.2 Controls and parts

1.2 Controls and parts

Front view of instrument:

2

719 S Titrino

1.2 Controls and parts

1Exchange Unit

2Display

3Setting of display contrast

4Controls the dosing rate during manual dosing with <DOS> and subsequent filling

5Control keys and indicator lamps on the Titrino

Key <DOS>

Dosing key. Dispensing is performed as long as <DOS> is being

 

pressed. Used e.g. to prepare the Exchange Unit. The dispensing

 

rate can be set with potentiometer (4).

Key <STOP/FILL>

- Stops procedures, e.g. titrations, conditioning.

 

- Filling after manual dosing with <DOS>.

Key <START>

Starts procedures, e.g. titrations, conditioning.

 

Identical with key <START> of the separate keypad.

Indicator lamps:

 

"Statistics"

Lamp is on when the "statistics" function (calculation of mean and

 

standard deviation) is on.

"Silo"

Lamp is on when silo memory (for sample data) is on.

719 S Titrino

3

 

Metrohm 719S User Manual

1.2 Controls and parts

Rear view of instrument:

4

719 S Titrino

1.2 Controls and parts

6RS232 interface

for the connection of printer, balance or a computer

7Remote lines

(input/output)

for the connection of the Remote Box, Sample Changers, robots etc.

8Connection of electrodes and temperature sensor

2 high-impedance measuring inputs for pH and U measurements (Ind I/ Ind II). They can either be used separately or for differential potentiometry, see page 124.

Important: If you work with both measuring inputs in the same vessel, the same reference electrode must be used.

1 measuring input for polarized electrodes (Pol).

1 measuring input for PT100 or Pt1000 temperature sensor.

9Connection for separate keypad

10Analog output for the connection of a recorder

11Connection for power cable

With power supplies where the voltage is subject to severe HF disturbances, the Titrino should be operated via an additional power filter, e.g. Metrohm 615 model.

12Mains switch

13Earthing socket

14Connection for stirrer

728 Magnetic Stirrer, 802 Rod Stirrer, 703 or 727 Ti Stand Supply voltage: 9 VDC (I 200 mA)

15Display of the set mains voltage

Before switching on for the first time, check that the set mains voltage matches the voltage of your power supply. If this is not the case, disconnect mains cable and change voltage.

16Rating plate

with fabrication, series and instrument number

719 S Titrino

5

 

2.1. Keypad

2 Manual operation

2.1 Keypad

CONFIG Configuration.

PARAM Parameters. SMPL DATA Sample data.

STATISTICS ON/OFF switching of statistics calculations of consecutive determination, see page 41.

MEAS/HOLD ON/OFF switching of measurements between titrations and hold during titrations.

SILO

ON/OFF switching of silo memory for

 

sample data, see page 51.

CAL.DATA

Calibration data, see page 49.

C-FMLA

Calculation values, see page 40.

DEF

Formulas, data output, see page 39ff.

USER METH Management of internal method

 

memory, see page 47.

PRINT

Printing of reports, see page 44.

REPORTS

Result output.

MODE

Mode selection, see page 24.

<,>

Keys for text input.

SELECT

Selection of special values (dialog

 

marked with ":")

CLEAR

Clears values, set special values.

ENTER

Stores values.

STOP

Stops methods.

QUIT

Quits inquiries, waiting times, printing.

START

Starts methods.

The third functions (inscriptions in the triangle) on the keys of the keypad are used for formula entry, see page 39.

6

719 S Titrino

 

2.2 Principle of data input

2.2Principle of data input

If you press a key you will find the corresponding menu in the display.

Example key <CONFIG>:

configuration >peripheral units

>peripheral units

 

send to:

IBM

configuration >auxiliaries

SET pH

********

 

 

In the first line you see where you are: You pressed key <CONFIG> and you are now in the menu "configuration".

In our example you are in the menu "configuration" on the title ">peripheral units". By pressing <CONFIG> you can move to the other titles of this menu.

If a dialog text is marked with ">", it contains a group of inquiries. You branch into this group by pressing <ENTER>.

Example inquiries of "peripheral units": The first line indicates again where you are.

If a dialog text of an inquiry is marked with ":", you can select a value with the key <SELECT>.

A value is stored with <ENTER> and the cursor moves to the next inquiry.

Repeated pressing of <ENTER> moves you through the inquiries of the group ">peripheral units“, after the last inquiry of this group you leave this group and return to the level above.

The next group of the menu "configuration“ appears: ">auxiliaries“

With key <QUIT> you leave an inquiry or a group of inquiries, it always moves you one level up.

In this example you leave the menu "configuration" by pressing <QUIT> and return to the display of titration mode and the chosen method.

719 S Titrino

7

 

2.3. Text input

2.3 Text input

 

Example storing a method:

<USER METH>

Press key <USER METH>, the group ">recall

 

method" appears.

user methods

 

>recall method

 

 

 

<USER METH> <ENTER>

>store method method name:

<CLEAR>

<<>

>store method

method name: ABCDEFG

<QUIT>

<ENTER>

>store

method

 

method

name:

text

 

 

 

Choose ">store method" by pressing <USER METH> and press <ENTER>.

The name of the method which is currently in the working memory is displayed.

Delete this name with <CLEAR>.

Open the "text writing mode" with key <<>.

Select the character marked by the blinking cursor with the Keys <<> and <>>, confirm it with <ENTER>. Select the next character...

When you confirmed the last character, i.e. your name is complete, you quit the text writing mode with <QUIT>.

Confirm now the name with <ENTER>.

If your text fills the whole text field, just press <ENTER> to store the text.

During text input you can correct typing errors with <CLEAR>:

<CLEAR> deletes the characters one by one.

If you wish to modify an existing name (e.g. if you have names like Text 1, Text 2, Text 3), do not delete the existing name before you start the text input mode. Proceed then as follows:

1.Press <USER METH>, place the cursor to ">store method" and press <ENTER>.

2.Open the text writing mode directly: Press key <<>.

3.<CLEAR> now deletes the characters one by one or you can add additional characters.

8

719 S Titrino

 

2.4 Tutorial

2.4 Tutorial

This short operating course teaches you to work quick and efficient with the 719 S Titrino, by means of the most important applications.

Set up your Titrino and connect the peripheral devices needed, see chapter 5.

2.4.1 Entering data, setting the dialog language

We can thus make a start and first take a look at the fundamentals of the entry of data. We change the dialog language.

SET pH

********

 

 

<CONFIG>

configuration >peripheral units

<CONFIG>

configuration >auxilliaries

<ENTER>

Switch on the Titrino. It is now in the standby state, it shows you the active mode and method name.

Press the key <CONFIG>, the display shows:

This is the title of the group "peripheral units". This group contains various inquiries about peripheral units.

Press again <CONFIG>. You see the title of a new group of inquiries.

This "auxiliaries" group contains the inquiry for the dialog language.

Pressing the <ENTER> key takes you to the inquiries of the group "auxiliaries". Note the ">" sign. All titles of inquiry groups are prefixed by this sign.

>auxilliaries

 

This is the first inquiry of the "auxiliaries" group: the

 

selection of the dialog language.

dialog:

english

 

 

 

<SELECT>

>auxilliaries

dialog: español

<ENTER>

>ajustes varios

fecha 2002-03-13

You select the various dialog languages with the <SELECT> key. Press <SELECT> repeatedly until "español" appears in the display.

Note the sign ":". It appears if the values can be selected with the key <SELECT>.

Accept the new "value" with <ENTER>.

The next inquiry "fecha" (date) of the group "ajustes varios" (auxiliaries) is shown.

You can open this inquiry by pressing <ENTER> too and go through all the inquiries of this group this way.

719 S Titrino

9

 

2.4. Tutorial

<QUIT>

configuration >ajustes varios

Because this inquiry follows no colon ":" the value can't be selected by <SELECT>, the date "fecha" has to be entered with the numeric keys.

Exit the inquiries with <QUIT>. You are one level higher in the "configuration" menu showing the title "ajustes varios" (auxiliaries).

<QUIT>

 

Press <QUIT> once again to exit the "configuration"

 

 

menu and return to the standby state.

 

 

All the dialog texts will now be displayed in Spanish.

SET pH

********

 

 

If you prefer English as the displayed dialog lan-

 

 

guage, proceed as before and select "English".

 

 

10

719 S Titrino

 

2.4 Tutorial

2.4.2 Development of a method, titration of an acid

Selection of the mode

 

<MODE>

 

 

mode

 

mode

SET

 

 

 

<ENTER>

 

 

SET pH

********

SET:

pH

 

 

4 × <SELECT>

SET pH

********

 

 

Now set the end point.

Press <MODE> repeatedly until "SET" appears in the display. For a description of the SET mode see page 30ff.

Confirm "SET" with <ENTER>

Now select the measured quantity: Press <SELECT> repeatedly until "pH" appears again in the display.

Confirm the measured quantity "pH" with <ENTER>.

Entry of the end point and the control parameters

<PARAM>

 

• Press <PARAM>. The display shows:

 

 

 

parameters

 

Control parameters for EP1

>SET1

 

 

 

 

<ENTER>

 

 

 

 

 

>SET1

 

 

EP at pH

OFF

 

 

 

• Set the end point to pH = 7.00.

 

 

>SET1

 

EP at pH

7.00

 

 

 

 

 

<ENTER>

 

 

>SET1

 

dynamics

OFF

 

 

 

 

>SET1

 

dynamics

3

 

 

 

<ENTER>

 

 

>SET1

 

max.rate

10.0 ml/min

 

 

 

<ENTER>

Dynamics means the control range in which the adding of the titrant is controlled in dependence on the measured value (OFF: largest control range, i.e. slow titration).

Set the range to 3. I.e. 3 pH units before the set EP controlled dosing starts.

Maximum possible titration rate inside the control range.

719 S Titrino

11

 

2.4. Tutorial

 

 

Minimum titration rate inside the control range.

>SET1

 

 

 

min.rate

25.0µl/min

 

 

• Quit the inquiry with <QUIT>.

2 × <QUIT>

For the titration put a Exchange Unit with c(NaOH) = 0.1 mol/L on the Titrino and rinse the tubing and the buret Tipp with <DOS>. Fill the buret again with <STOP/FILL>.

Plug a combined pH glass electrode into measuring input 1 (Ind I).

Pipette 2 mL c(HCl) = 0.1 mol/L into your titration vessel, dilute with ca. 20 mL dist. water. Put a magnetic stirrer in the titration vessel and place buret tip and electrode in the probe, see page 126 for their arrangement.

Start the stirrer and press <START>

During the titration the first line of the display shows the current measured value and the volume already dispensed. The second line shows a "control bar", which indicates the control deviation of the current measured value from the set end point.

pH 2.67

0.351 ml

#======

 

 

 

After the titration the end point volume and its pH value is displayed.

With this end point a result can be calculated:

Calculation of the result: formula entry

<DEF>

def >formula

<ENTER>

RS?

<1>

RS1=

With this equivalence point a result can be calculated.

Press the <DEF> key.

Press <ENTER> to move on to the formula entry. The display now shows "RS?".

Press "1", i.e. the first formula.

You now can enter a formula. Note here the top inscription on the keys of the keypad and the numbers. The following symbols can be used:

EP# EP's with 1-digit number, e.g. EP1.

RS# Previously calculated results, e.g. RS1 in the second formula.

C## Calculation constant with 2-digit number, e.g. C01. C00 is reserved for the sample size. For the meaning of the calculation constants see page 40.

12

719 S Titrino

 

2.4 Tutorial

RS1=EP1*C01*C02/C00

<ENTER>

Calculate the content of your hydrochloric acid in g/L: RS1=EP1*C01*C02/C00

End point*conc.(titrant)*molecular weight/sample size To correct a formula, delete it with <CLEAR>.

Confirm the formula with <ENTER>.

>formula

 

You may enter a text for the result output, see page

RS1 text

RS1

 

 

8.

 

 

>formula

 

RS1 decimal places

2

 

 

 

 

>formula

 

RS1 unit

%

 

 

<SELECT>

 

<ENTER>

 

 

 

RS?

 

 

 

2 x <QUIT>

 

Enter the desired number of decimal places for the result.

Select the desired unit g/L with <SELECT>, or type a text as unit, see page 8.

The Titrino prompts for the calculation of the next result.

Quit the formula entry by pressing <QUIT> twice.

In order to be able to calculate the result, enter the calculation constants used in the formula.

Entry of the calculation constants

<C-FMLA>

 

 

Press <C-FMLA>.

 

 

 

The constants which have been used in the formula

C-fmla

 

 

 

are requested: input with digit keys, confirm with

>C01

0.0

 

 

 

 

<ENTER>.

0.1

 

 

 

C01: Concentration of your titrant= 0.1 mol/L.

<ENTER>

 

 

Enter 0.1.

 

 

 

C02: Molar mass of HCl = 36.47 g/mol

C-fmla

 

 

 

 

Enter 36.47.

>C02

0.0

 

 

 

 

 

36.47

 

 

 

<ENTER>

 

 

 

Now only the constant C00 is missing to perform the calculation. This constant is reserved for the sample size. It can be entered directly.

719 S Titrino

13

 

2.4. Tutorial

Entry of the sample size

4 x <SMPL DATA>

smpl

data

 

smpl

size

1.0 g

 

 

 

Press <SMPL DATA> repeatedly until "sample size" appears in the display.

<2>

 

Enter 2.

<ENTER>

 

 

 

 

 

smpl data

 

 

smpl unit:

g

 

 

 

 

<SELECT>

 

Use <SELECT> to select the unit "mL" and confirm

<ENTER>

 

the new value with <ENTER>.

 

 

The result is now calculated and can be displayed in place of the end point. If your method already includes a formula at the end of the titration, the calculated result is displayed directly after the titration. As we have entered the formula later, we now have to select the result display:

Display of the result

 

<SELECT>

 

 

SET pH

********

>display results

Press <SELECT> repeatedly until ">display results" appears in the display.

 

<ENTER>

Press <ENTER> to move to the result display.

 

 

 

SET pH

********

 

RS1

3.71 g/l

 

 

 

 

If you have a printer connected, you probably wish to have the curve and a result report printed out automatically at the end of a titration.

To install a printer, see page 119.

14

719 S Titrino

 

2.4 Tutorial

Print reports

3 x <DEF>

Press <DEF> repeatedly until the display shows:

def >report

<ENTER>

report:

3 x <SELECT> <;>

report:curve;full

<ENTER>

<QUIT>

Press <ENTER> to move to the definition of reports.

With <SELECT> you select the individual repot blocks. Use a ";" as delimiter between the report blocks to print more than one block. If you wish to print out a full result report, enter "full".

Confirm the entry with <ENTER> and quit the inquiries with <QUIT>.

Now print your reports by pressing <PRINT><REPORTS><ENTER>. Your printout will look like the following:

'fr

 

 

 

Identification of the report type (fr=full report)

719 S Titrino

18200

719.0021

Device type, device number and program version

date 2002-03-13

 

time 17:34

1

Initial pH value, mode and method name

pH(init)

2.40

SET pH

********

smpl size

2

ml

 

 

EP1

2.032

ml

8.14

Volume and pH value of EP1

RS1

3.71

g/l

 

Calculated result

---------------

 

 

 

 

 

 

 

The development of your method is now complete. Before we store it in the method memory, you should check it again. Prepare a fresh sample and restart the titration with <START>.

If everything appears to be all right, you can now store the method in the method memory.

719 S Titrino

15

 

2.4. Tutorial

2.4.3 Storage and loading of methods

Now store the method you have just developed in the method memory.

Storage of a method

2 × <USER METH>

user methods >store method

Press <USER METH> repeatedly until the title ">store method" appears in the display.

<ENTER>

>store method

method name: ********

<ENTER>

Open the inquiry with <ENTER>.

Enter an identifier, e.g. 1 or Acid. For text input see page 8.

 

 

The method now runs under the name "Acid". It is ready

SET pH

Acid

to titrate.

 

 

 

 

 

If you have a printer connected, you can print out the contents of your method memory. Key sequence: <PRINT><USER METH><ENTER>

Stored methods can be loaded at any time.

Loading a method from the method memory

<USER METH>

• Press <USER METH>. The display shows the title

 

">recall method".

user methods

 

>recall method

 

 

 

<ENTER>

>recall method

method name: ********

<SELECT> or

direct entry

<ENTER>

SET pH

Acid

 

 

Open the inquiry <ENTER>

You can select the desired method with <SELECT> (it shows all methods of the method memory one after the other) or by directly enter the method name.

Load the method with <ENTER>

The method is ready to titrate.

16

719 S Titrino

 

2.4 Tutorial

2.4.4 pH calibration

For end-point titrations, however, where titration is performed to a fixed, specified pH value, a calibration should be performed.

Selection of the calibration mode, CAL

<MODE>

 

Press <MODE> repeatedly until "CAL" appears in

 

 

the display and confirm the mode with <ENTER>.

mode

 

 

 

mode

CAL

 

 

 

 

<ENTER>

CAL

********

pH(as) 7.00 slope

1.000

 

 

Calibration procedure

The instrument is ready for a 2-point calibration. The second display line shows the current calibration data for measuring input 1.

<START>

CAL

 

cal.temp.

25.0°C

<ENTER>

 

 

25.0°C

buffer 1

pH

7.00

 

 

 

 

<ENTER>

 

 

 

 

 

 

25.0°C

buffer 2

pH

4.00

 

 

 

<ENTER> oder <STOP>

Immerse your electrode in the first buffer and start the calibration procedure.

Inquiry of calibration temperature.

Enter the current temperature. If you have a temperature sensor connected, this inquiry is skipped.

pH value of the first buffer.

Enter the pH value of the buffer at your calibration temperature.

The voltage of the first buffer is measured. When the measured value has met the set drift criterion, the measurement is stopped and the pH value of the second buffer is requested.

Now enter the pH value of the second buffer. If you require a 1-point calibration, you can also terminate the calibration with <STOP>.

CAL

********

At the end of the calibration, the calibration data ob-

pH(as) 6.89 slope

0.985

tained are displayed: asymmetry pH and slope.

 

 

 

The calibration data can be viewed at any time under the <CAL.DATA> key. Our calibration data are stored under ">input 1".

The calibration report can be printed out at any time with the key sequence: <PRINT><CAL.DATA><ENTER>

719 S Titrino

17

 

2.4. Tutorial

2.4.5 Statistics, acid capacity of drinking water

Let us now determine the acid capacity of drinking water. For this, the SET (set endpoint titration) mode is used to titrate to pH = 4.3.

Load the method "Acid", the one that you just have stored in the method memory, see page 16 .

Set the end point pH = 4.3 and dynamics 3 as you did with your first example (key <PARAM>). If you still feel unfamiliar with that, see page 11 again.

Now alter the calculation formula before the titration (keys <DEF>,<ENTER> and <1>). Delete the existing formula with <CLEAR> and enter the following formula:

 

RS1=EP1*C01*C02

 

with an accuracy of

RS1 decimal places

2

the unit

RS1 unit:

 

mmol/L

and the calculation constants

C01

1

(concentration of titrant ×10)

(key <C-FMLA>)

C02

4

(factor for the sample size 100 mL/25 mL)

If you have a printer connected to your Titrino, you can print out the titration parameters with the keys

<PRINT><PARAM><ENTER>.

Mount an Exchange Unit with c(HCl) = 0.1 mol/L as titrant. Add 25 mL drinking water as sample and start the titration with <START>. During the titration the first line of the display shows the measured value and dispensed volume. The "c" in pHc shows that the electrode assembly has been calibrated. The second line shows a "control bar", which indicates the control deviation of the current measured value from the set end point.

pHc 6.34

0.426 ml

#======

 

 

 

If the titration runs too slow or too fast, you can change the control parameters during the titration. If you wish to titrate faster, change the following parameters:

dynamics: lower value, the control range becomes smaller (Attention: possibility of titrating over the set EP)

max.rate: bigger value

min.rate: bigger value

You will find further details of the control parameters on page 31.

After the titration, end point volume and pH value at the end point are displayed.

18

719 S Titrino

 

2.4 Tutorial

If the previous titration has run to your satisfaction, you can start thinking about performing statistics calculations. Have you already added a new sample to the titration vessel? If you are no longer certain, you can find out immediately with <MEAS/HOLD>.

Rapid measurement between titrations

Press <MEAS/HOLD>. The pH value of your sample is displayed. You can stop the measurement with a second <MEAS/HOLD>.

Statistics calculations

Now switch on the statistics calculations. Press <STATISTICS>. The "Statistics" LED is on. Duplicate determinations are now performed.

• Perform 2 titrations.

After the second titration you receive a printout with statistics calculations:

'fr

 

 

 

719 S Titrino

18200

719.0021

date 2002-03-13

time 17:34

3

pHc(init)

6.29

SET pH

Acid

EP1

0.0628

ml

4.26

m value

5.02

mmol/l

 

 

mean(2)

+/s

s/%

m value

5.04

0.028 mmol/l 0.56

==============

If you have no printer connected, you can view the mean value and standard deviation:

Press <SELECT> until "display mean" appears in the display.

With <ENTER> you obtain the mean value.

A second <ENTER> displays the number of single values which have been used for the mean calculation.

If you press <SELECT> again, the display "display std.deviation" appears. Once again, you can view this value, by pressing <ENTER>.

Perhaps you have noticed that the two values differ too greatly? In any case, we shall perform a third determination with the same sample. The result of this determination will be another value which is incorporated in the statistics calculations.

Addition of more determinations for the statistics calculations

5 × <PARAM>

 

• Press <PARAM> until the display shows:

 

 

 

 

parameters

 

 

 

>statistics

 

 

 

 

 

 

 

<ENTER>

 

 

 

 

 

• Statistics calculations can be switched on either with

>statistics

 

 

the <STATISTICS> key or in the inquiry "status:".

status:

ON

 

 

We shall leave them switched on and proceed to the

<ENTER>

 

 

next inquiry.

719 S Titrino

 

 

 

 

 

19

 

 

2.4. Tutorial

 

 

Mean value of 2 single determinations.

>statistics

 

 

Enter "3" to include an additional determination.

mean:

n= 2

 

 

 

 

<3>

 

 

 

<ENTER>

 

 

 

2 × <QUIT>

 

Quit the inquiry.

Perform another titration.

You can now decide which of the results is an "outlier". You can delete it from the statistics calculation.

Deleting a result from the statistics calculation

5 × <PARAM>

 

• Press <PARAM> until the display shows:

 

 

 

 

parameters

 

 

 

>statistics

 

 

 

 

 

• Select with <ENTER> the inquiry of the result table

3 × <ENTER>

 

 

 

 

"res.tab:"

 

 

 

 

>statistics

 

 

 

res.tab:

original

 

 

 

• To delete a single result with index n from the result

2 × <SELECT>

 

 

 

 

table, press <SELECT> so that "delete n" appears.

>statistics

 

 

 

 

 

res.tab:

delete n

 

 

 

 

<ENTER>

 

 

 

 

 

• Enter index n of the result you wish to delete; in our

>statistics

 

 

 

 

example the second:

delete

n=

1

 

 

 

 

<2>

 

 

 

<ENTER>

 

 

2 × <QUIT>

 

• Quit the inquiry by pressing <QUIT>.

 

 

 

Mean value and standard deviation are recalculated

 

 

 

and can be viewed in the display.

With <PRINT><REPORTS><ENTER> you can print a fresh report.

With <PRINT> <STATISTICS><ENTER> you can print a statistics overview, in this report the deleted result is marked with a " " in the report.

20

719 S Titrino

 

 

 

 

 

 

2.5 Configuration, key <CONFIG>

 

 

 

 

 

 

2.5 Configuration, key <CONFIG>

 

 

 

 

 

 

 

 

 

 

Key <CONFIG> serves to enter device specific data.

 

CONFIG

 

 

The set values apply to all modes.

 

 

 

 

peripheral units:

 

configuration

 

 

 

 

 

Selection of printer, balance and the curve at the ana-

 

>peripheral units

 

 

 

 

 

log output.

 

 

 

 

 

 

 

 

 

 

 

 

auxiliaries:

 

 

 

 

 

 

e.g. setting of dialog language, date, time, etc.

 

 

 

 

RS232 settings:

 

 

 

 

RS parameters for the COM interface.

 

 

 

 

common variables:

 

 

 

 

Values of common variables.

 

 

 

 

The display texts of the Titrino are shown to the left.

 

 

 

 

The values are the default values.

 

 

 

 

 

 

 

>peripheral units

 

 

Settings for peripheral units

 

send to:

IBM

Selection of printer (Epson, Seiko, Citizen, HP, IBM) at

 

 

 

 

the Titrino Rs232 interface.

 

 

 

 

"Epson", for Epson

 

 

 

 

"Seiko", e.g. for DPU-414

 

 

 

 

"Citizen", e.g. for iDP 562 RS, Custom DP40-S4N

 

 

 

 

"HP" e.g. for Desk Jet types. Place curves always at the

 

 

 

 

beginning of a page as you cannot have them over 2

 

 

 

 

pages.

 

 

 

 

 

 

"IBM" for all printers with IBM character set Table 437

 

 

 

 

and IBM graphics, as well as for the data transmission

 

 

 

 

to a computer with Metrodata software TiNET or

 

 

 

 

VESUV.

 

 

 

balance:

Sartorius

Selection of balance (Sartorius, Mettler, Mettler AT,

 

 

 

 

AND, Precisa)

 

 

 

 

Sartorius:

Models MP8, MC1

 

 

 

 

Mettler:

Models AM, PM, AX, MX, UMX and

 

 

 

 

 

balances with 011, 012, or 016 interface

 

 

 

 

Mettler AT:

Model AT

 

 

 

 

AND:

Models ER-60, 120, 180, 182, FR-200, 300

 

 

 

 

 

and FX-200, 300, 320

 

 

 

 

Precisa:

Models with RS232C interface

 

record:

U

Selection of the curve for the output at the analog

 

 

 

 

output (U, dU/dt, V, dV/dt, U(rel), T

 

 

 

 

U:.

Voltage

 

 

 

 

dU/dt:

Measured value drift

 

 

 

 

V:

Volume

 

 

 

 

dV/dt

Volume drift

 

 

 

 

U(rel):

Control deviation with SET

 

 

 

 

T:

Temperature with MEAS T

719 S Titrino

21

 

2.5. Configuration, key <CONFIG>

>auxiliaries

 

General settings

dialog:

english

Selection of dialog language (english, deutsch,

 

 

francais, español, italiano, portugese, svenska)

date

2001-03-13

Current date (YYYY-MM-DD)

 

 

Format: Year-month-day, entry with leading zeros.

time

08:13

Current time (HH-MM)

 

 

Format: Hours-minutes, entry with leading zeros.

run number

0

Current run number for result output (0...9999)

 

 

The sample number is set to 0 when the instrument is

 

 

switched on and incremented on every determination.

auto start

OFF

Automatic starts of titrations. (1...9999, OFF)

 

 

Number of automatic starts ("number of samples").

 

 

Used for instrument interconnections in which the

 

 

external instrument does not initiate a start. Not

 

 

advisable in connections with Sample Changers.

start delay

0 s

Start delay (0...999999 s)

 

 

Delay time after start of methods. Abort start delay time

 

 

with <QUIT>.

dev.label.

 

Individual identification of devices (up to 8 ASCII

 

 

characters)

 

 

Will be printed in the result report, see page 45.

program

719.0021

Display of program version

>RS232 settings

 

Settings of RS232 interface

 

 

see also page 95ff.

baud rate:

9600

Baud rate (300, 600, 1200, 2400, 4800, 9600)

data bit:

8

Data bit (7, 8)

stop bit:

1

Stop bit (1, 2)

parity:

none

Parity (even, odd, none)

handshake:

HWs

Handshake (HWs, HWf, SWline, SWchar, none)

 

 

see page 95ff.

RS control:

ON

Control via RS232 interface (ON, OFF)

 

 

"OFF" means that the receipt of commands via the

 

 

RS232 interface is blocked. Data output is possible.

22

719 S Titrino

 

 

 

2.5 Configuration, key <CONFIG>

>common variables

 

Values of the common variables

C30

0.0

Common variables C30...C39 (0.. ± 999 999)

etc.

 

The values of all common variables are displayed. For

 

 

creating of common variables see page 43.

719 S Titrino

23

 

2.6. Selection of the mode, key <MODE>

2.6 Selection of the mode, key <MODE>

MODE ;

>

Press key <MODE> until the desired mode is displayed and confirm with <ENTER>.

Select the measured quantity pH, U, Ipol, Upol, (T) with <SELECT> and confirm it also with <ENTER>.

The following modes can be selected:

SET: Set Endpoint Titration.

CAL: pH Calibration.

MEAS: Measuring.

These standard modes are loaded with a set of default parameters and are ready to work (only with SET an end point has to be set).

24

719 S Titrino

 

2.7 Parameters, key <PARAM>

2.7 Parameters, key <PARAM>

The key <PARAM> is used for the entry of values that PARAM determine the modes. Values marked with "cond." are

accessible during the conditioning in the SET mode. "**titr." means that these values can be changed during the titration. They influence the ongoing determination. Other values can only be changed in the inactive state.

The display texts of the Titrino are shown to the left. The values are the default values.

2.7.1 Parameters for SET

 

 

 

 

SET1, SET2:

 

parameters

 

 

 

>SET1

 

 

Control parameters for EP1 and EP2.

 

 

 

 

titration parameters

 

 

 

 

 

 

 

 

are valid for the global titration sequence.

 

 

 

 

stop conditions:

 

 

 

 

Parameters for the termination of the titration.

 

 

 

 

statistics:

 

 

 

 

Calculation of mean values and standard deviation,

 

 

 

 

see also page 41.

 

 

 

 

preselections:

 

 

 

 

ON/OFF of various auxiliary functions such as auto-

 

 

 

 

matic requests after the start and activate pulse.

 

 

 

 

 

 

>SET1

 

 

Control parameters for EP1 or EP2, resp.

 

EP at pH

OFF

Preset EP1 (input range depends on the measured

 

**titr.

 

 

quantity:

 

 

 

 

 

pH:

0...±20.00, OFF

 

 

 

 

U, Ipol:

0...±2000 mV, OFF

 

 

 

 

Upol:

0...±200.0 µA, OFF)

 

 

 

 

<CLEAR> sets "OFF".

 

 

 

 

If EP1 is "OFF", no further inquiries under SET1 appear.

 

dynamics

OFF

Distance from EP where constant dosing should stop

 

**titr.

 

 

and controlling begins. (control range, input range

 

 

 

 

depends on the measured quantity:

 

 

 

 

pH:

0.01...20.00, OFF

 

 

 

 

U, Ipol:

1...2000 mV, OFF

 

 

 

 

Upol:

0.1...200.0 µA, OFF)

 

 

 

 

<CLEAR> sets "OFF".

 

 

 

 

"OFF" means largest control range, i.e. low titration.

 

 

 

 

Outside the control range, dispensing is performed

 

 

 

 

continuously, see also page 31.

 

 

 

 

 

 

719 S Titrino

25

 

2.7. Parameters, key <PARAM>

max.rate

10.0 ml/min

Maximum dosing rate (0.01...150 mL/min, max.)

**titr.

 

 

<CLEAR> sets "max.".

 

 

 

This parameter determines primarily the addition rate

 

 

 

outside the control range, see also page 31.

 

 

 

The maximum rate depends on the Exchange Unit:

 

 

 

Exchange Unit

max.

 

 

 

 

5 mL

15 mL/min

 

 

 

 

10 mL

30 mL/min

 

 

 

 

20 mL

60 mL/min

 

 

 

 

50 mL

150 mL/min

min.rate

25.0 µl/min

Minimum dosing rate (0.01...999.9 µL/min)

**titr.

 

 

This parameter determines the addition rate right at the

 

 

 

start and the end of the titration, see also page 42.

 

 

 

This parameter influences the titration speed and

 

 

 

therefore its accuracy very strongly: A smaller min.rate

 

 

 

results in a slower titration.

stop crit:

 

drift

Type of stop criteria (drift, time)

**titr.

 

 

Titration stops if EP is and stop drift is reached. (1...999

stop drift

 

20 µl/min

**titr.

 

 

µL/min)

 

t(delay)

 

10 s

Titration stops if there is no dosing during t(delay).

**titr.

 

 

(0...999 s, INF)

 

 

 

 

<CLEAR> sets "INF"

 

 

 

Switch off when the end point is reached and the set

 

 

 

time after the last dispensing has elapsed.

 

 

 

If "INF" is set, an inquiry regarding the stop time

 

 

 

appears.

 

 

 

 

If t(delay) is "INF"

 

stop time

 

OFF s

Stop after a time (0...999999 s, OFF)

**titr.

 

 

<CLEAR> sets "OFF".

 

 

 

Stop after the set time after the start of the titration.

 

 

 

"OFF" means no stop, i.e. titration for an "infinitely" long

 

 

 

time.

 

 

 

>titration parameters

General titration parameters

titr.direction:

auto

Direction is set automatically (+, -, auto)

 

 

 

auto: The direction is set automatically by the Titrino

 

 

 

+:

(sign [Ufirst

- EP]).

 

 

 

Direction of higher pH, higher voltage (more

 

 

 

 

"positive"), larger currents.

 

 

 

- :

Direction of lower pH, lower voltage, smaller

 

 

 

 

currents.

 

 

 

 

The titration direction is fixed if two EP's are set. In this

 

 

 

case, an input for titration direction has no meaning.

pause 1

 

0 s

Pause 1 (0...999999 s)

**titr.

 

 

Waiting time, e.g. for the equilibration of the electrode

 

 

 

after the start. The waiting time can be aborted with

 

 

 

<QUIT>.

 

 

 

 

 

 

 

26

719 S Titrino

 

Loading...
+ 122 hidden pages