MDA Scientific midas Instructions Manual

1
TO X I C , F L A M M A B L E
A N D O X Y G E N G A S D E T E C T O R
O P E R A T I N G I N S T R U C T I O N S
2
TABLE OF CONTENTS
1 Table of contents
2 General description
5
3 Product overview
5
3.1 Main chassis 6
3.1.1 Display module 6
3.1.2 Pump module 7
3.1.3 Sensor cartridge chamber 7
3.2 Mounting bracket assembly 7
3.2.1 Mounting bracket 7
3.2.2 Terminal module 7
3.3 Sensor cartridge 8
3.3.1 Biased sensor cartridges 8
3.4 Cover 8
4 Default configuration 9
5 Installation 9
5.1 Mounting and location of detector 10
5.2 Mechanical installation 11
5.3 Sample and exhaust tubing calculations 12
5.4 In-line filters 13
5.5 Local detector option 14
5.6 Electrical installation 15
5.7 Electrical connections 17
5.8 Refitting the main chassis 18
5.9 Installing the sensor cartridge 19
6 Detector start up procedures 19
7 General operation 21
7.1 Normal operation mode
21
7.1.1 Resetting alarms, faults and maintenance faults 22
7.2 Review mode
22
7.2.1 Review mode menu overview
23
7.3 Overview of set-up, calibration and test mode 24
7.3.1 Set-up menu overview
24
7.3.2 Calibration menu overview ‘
CAL’ 26
7.3.3 Test menu overview ‘
tESt’ 27
8 Detailed procedures for navigating review, set-up, calibration and test mode submenus 28
8.1 Review mode 28
8.1.1 Review software ‘SW’ 28
8.1.2 Review alarms ‘
ALm’ 29
8.1.3 Review faults ‘
FLt’ 29
8.1.4 Review calibration ‘
CAL’ 29
8.1.5 Review date and time ‘timE’ 29
8.1.6 Review detector address ‘
nEt’ 30
8.1.7 Review event log ‘
Hi St’ 30
© 2004 Zellweger Analytics Issue 1 12/2004 MIDAS-A-001
3
8.2 Set-up, calibration and test modes 30
8.2.1 Set-up menu ‘
8.2.2 Set alarms ‘
8.2.3 Set faults ‘
8.2.4 Set calibration interval ‘
8.2.5 Set date and time ‘timE’
8.2.6 Set address ‘
8.2.7 Set pass code ‘
8.3 Calibration menu ‘
8.3.1 Zero calibration ‘
8.3.2 Span calibration ‘
8.3.3 Flow calibration ‘
SEt’ 31
ALm’ 31
FLt’ 34
CAL’ 34
35
nEt’ 35
PWd’ 36
CAL’ 36
0CAL’ 36
SPAn’ 37
FLoW’ 38
8.3.4 mA calibration ‘mA 4-20’ 38
8.4 Test menu ‘
8.4.1 Bump test ‘
8.4.2 Alarm/fault test ‘
8.4.3 Inhibit state ‘
tESt’ 39
bUmP’ 39
Si m’ 39
I nH’ 40
9 Routine maintenance 41
9.1 Sensor Cartridge Replacement 41
9.1.1 Sensor cartridge fitting/replacement 41
9.2 Pump replacement 43
9.3 Re-assembling the detector 46
9.4 Filter replacement 46
10 Pyrolyzer module option 47
10.1 Fitting the pyrolyzer module 48
10.2 Re-assembling the MIDAS
®
detector 49
11 Analog module 51
11.1 Fitting the analog module 51
®
11.2 Re-assembling the MIDAS
detector 52
12 Trouble shooting and fault diagnosis 53
13 REFLEX
®
54
14 Internal web server 54
14.1 Physical network components 54
14.2 Internet settings 54
14.3 Running the web browser 54
15 Typical installation topologies 56
15.1 Conventional installation 57
15.2 Modbus/TCP installation 57
15.3 Power over Ethernet (PoE) installation 57
16 Ordering information 58
®
16.1 MIDAS
16.2 MIDAS
16.3 MIDAS
transmitter 58
®
pyrolyzer 58
®
analog input module 58
4
16.4 MIDAS® plug in sensor cartridges (standard warranty) 59
16.5 MIDAS
®
plug in sensor cartridges (extended warranty) 60
16.6 MIDAS
®
complete gas detector kits 61
16.7 Accessories and spares 61
17 General specifications 62
18 Calibration and bump testing 63
19 Warranty statement 67
20 Software menu state charts 69
20.1 Top level 69
20.2 Review mode 70
20.3 Review software info, alarm, fault and gas calibration 71
20.4 Review date/time and network 72
20.5 Review event log 73
20.6 Set-up mode 74
20.7 Set-up alarm, fault and gas calibration 75
20.8 Set-up date/time and network 76
20.9 Set-up pass code 77
20.10 Calibration mode 78
20.11 Calibration gas zero and span 79
20.12 Calibration - Flow calibration 80
20.13 Calibration - 4-20 mA 81
20.14 Test mode 82
20.15 Test bump, alarm/fault simulation 83
20.16 Test inhibit 84
21 Contact details 85
5
2 GENERAL DESCRIPTION
The MIDAS
®
gas detector is an extractive gas sampling system that draws a sample locally or from a remote point to a sensor cartridge that is located inside the detector’s chassis. A wide range of toxic, flammable and oxygen gas sensor cartridges are available that enable detection of gases used or generated in the Semiconductor and other manufacturing industries.
MIDAS
®
is wall mounted and displays gas concentration, alarm, fault and status information via its backlit LCD and LEDs. A simple to use 4-button keypad located under the display provides the facility to set-up, review, operate and make changes to the detector’s configuration.
MIDAS
®
has flexible power and communications capabilities built in as standard. These include 3 on board relays, 0-22 mA analog output, Modbus/TCP outputs for signal and service connectivity as well as the innovative Power over Ethernet (PoE) protocol that enables a single Ethernet connection to be made for all power, control and communication requirements.
3 PRODUCT OVERVIEW
The MIDAS
®
gas detector comprises of 4 parts: the main chassis, the mounting bracket assembly, the sensor
cartridge and the unit cover. Diagram 1 details the MIDAS
®
general arrangement. Additionally, there is an
optional Pyrolyzer module required for the detection of NF
3
and an optional Analog module for the connection of
external 4-20 mA devices. Please refer to section 10 and 11 respectively for details of these options.
Main chassis
Unit cover
Sensor cartridge
Mounting bracket assembly
Diagram 1. MIDAS® general arrangement exploded view
6
1
2
1
3
1
5
1
4
1
6
1
7
1
8
1
21
1
20
1
9
1
10
1
11
1
12
1
13
1
14
1
15
1
16
1
17
1
19
1
18
1
1
3.1 Main Chassis
The main chassis comprises of the display module, pump assembly, and plug in sensor cartridge chamber.
Diagram 2. Main chassis
Pump module
Display module
Sensor cartridge chamber
3.1.1 Display module
The display module is located at the front of the main chassis and consists of a large alphanumerical and graphical backlit LCD, 3 LED indicators and a 4-button keypad. Under normal operation the LCD and LEDs display gas concentration, alarm and system status. In set-up, review, calibration and test modes, the LCD shows the relevant menu options. These menus are simply navigated using the ‘
’ up, ‘
’ down, ‘
’ accept and ‘X’
cancel buttons.
1. Red alarm LED
2. Normal operation icon
3. Review mode icon
4. Green power LED
5. Set-up mode icon
6. Calibration mode icon
7. Yellow Fault LED
8. Test mode icon
9. Gas concentration and message display area
10. Displayed units
11. Pass code icon
12. Accept button
13. Address icon
14. Network icon
15. Inhibit icon
16. Fault icon
17. Up button
18. Alarm level 1 icon
Alarm level 2 icon (For flammable and toxic) O
2
deficiency alarm icon
O
2
excess alarm icon
19. Cancel button
20. Zero
and Span
calibration icons
21. Flow indicator
Diagram 3. MIDAS
®
display module layout
7
3.1.2 Pump Module
The pump module is located at the back of the main chassis. It draws the gas sample from the inlet port located at the bottom of the mounting bracket assembly via an inline filter to the sensor cartridge chamber located at the front of the main chassis. The sample is then exhausted via the exhaust port located at the bottom of the mounting bracket assembly. The pump and filter assemblies are designed for easy replacement. For replacement details refer to sections 9.2 and 9.4 respectively.
3.1.3 Sensor Cartridge Chamber
The sensor cartridge chamber is located at the front of the main chassis below the display module. The plug in sensor cartridge is fitted into this area which makes the electrical connection between the sensor cartridge and the rest of the electronics as well as providing the chamber where the sensor cartridge is exposed to the sampled gas. For details of fitting sensor cartridge refer to section 5.9.
3.2 Mounting Bracket Assembly
The mounting bracket assembly comprises of the detector mounting bracket, the terminal module, the gas sample inlet and outlet ports, the cable/conduit entry and Ethernet communications socket.
Diagram 4. Mounting bracket assembly
Mounting bracket Terminal module
Cable entry
Gas inlet port
Ethernet/PoE socket
Gas outlet port
3.2.1 Mounting Bracket
The metal mounting bracket has two slots that allow the detector to be easily mounted to a wall using two suitable screw fixings. For further details of mounting the detector refer to section 5.
3.2.2 Terminal Module
The terminal module is located on the mounting bracket. All electrical connections to MIDAS
®
are made via this module. Wire entry to the terminal module area is via the PG16 cable entry/conduit entry located at the bottom of the mounting bracket assembly.
8
3.3 Sensor Cartridge
A wide range of toxic, flammable or oxygen sensor cartridges can be fitted to MIDAS
®
. The plug in sensor cartridges are fitted in the sensor cartridge chamber at the front of the main chassis. To access the chamber the unit cover is removed by unscrewing the thumbscrew located at the front of the detector. The pre calibrated smart sensor cartridges can easily be fitted or replaced as they simply plug into the detector without the need for any tools. The sensor cartridge is firmly held in place by two locking tabs.
Diagram 5. Sensor cartridge
Front Back
Connector
Locking tabs
3.3.1 Biased Sensor Cartridges
Some sensor cartridges are shipped with a battery powered electrical supply in order to keep the cell effectively ‘warmed up’ and ready to monitor once installed in the MIDAS
®
unit. Bias cells are supplied for TEOS, NO and
O
3
. The bias circuit is removed just before insertion into the MIDAS
®
system and the sensor cartridge is thus ready
for effective gas detection.
Should a bias voltage not be applied (e.g. during a power failure) then the cell will take a longer time to recover before effective gas detection can take place. The longer the loss of applied power, the longer the recovery time. Refer to the relevant sensor cartridge data sheet for information on each sensor cartridge.
In order to avoid the risk of loss of gas detection due to unforeseen power loss, we recommend that a power management solution such as uninterruptible power supplies, battery back ups etc. are used.
NOTE: Sensor warranty is void if the sensor cartridge is opened by unauthorized user.
3.4
Cover
The cover provides environmental protection and fits over the top, front and sides of the main chassis. The front panel has viewing windows for the LCD, LEDs and sensor cartridge fitted in the sensor cartridge chamber. Underneath the LCD window are the 4 push buttons used for navigating the detector’s software menus. The cover is easily removed to allow access to the chassis by unscrewing the thumbscrew on the front panel and pulling the cover forwards off the main chassis.
Diagram 6. MIDAS
®
cover
LED windows
LCD window
Push buttons
Sensor cartridge viewing window
Thumbscrew
9
4 DEFAULT CONFIGURATION
5 INSTALLATION
For ease of installation MIDAS
®
has been designed to allow the installation of the mounting bracket assembly and terminal module separately from the other parts of the detector. The detector location and hard wiring can therefore be completed before fitting the detector’s main chassis and sensor cartridge.
WARNING
MIDAS
®
is designed for installation and use in indoor safe area non-explosive atmospheres. Installation must be in accordance with the recognised standards of the appropriate authority in the country concerned. Prior to carrying out any installation ensure local regulations and site procedures are followed.
As standard, the MIDAS® gas detector is factory configured as below:
Table 1. MIDAS
®
default configuration
Current source with:
Full Scale (FS)
Lowest Alarm Level
(LAL)
Lower Detectable
Limit (LDL)
Alarm Relay 1
Alarm Relay 2
Fault Relay
Latching
Pass code
0 mA
1.0 mA
3.0 mA
4.0 to 20.0 mA
2.0 mA
Fault (open circuit)
Fault
Maintenance Fault
Gas reading (normal operation)
Inhibit (test mode)
21.0 mA Over range
Toxic Gas Flammable Gas
Typically 4 x Toxic
Threshold Limit (TLV)
100% Lower Explosive
Limit (LEL)
1
Oxygen
25% Volume (v/v)
Typically 1/2 TLV 10% LEL 5% v/v
Typically 0.4 TLV 9% LEL 0% v/v
The LDL is the minimum level that is reliably distinguishable from zero.
Alarm level 1 of 1/2 TLV 10% LEL 23.5% v/v (Rising)
Normally de-energized, energizes on alarm.
Contact Normally Open (NO), closes on alarm.
Alarm level 2 of TLV 20% LEL 19.5% v/v (Falling)
Normally de-energized, energizes on alarm.
Contact Normally Open (NO), closes on alarm.
Normally energized, de-energizes on fault.
Contact Normally Open (NO).
Non latching. Alarm and fault relays automatically reset when reading falls
below alarm thresholds or fault clears.
No pass code set.
®
1
MIDAS
detectors are not ETL approved for monitoring in or sampling from classified areas above 25% LEL
10
5.1 Mounting and Location of Detector
The MIDAS
®
gas detector has an integral mounting bracket assembly that is easily mounted to a suitable vertical
surface such as a wall, tool housing, mounting plate on a pole etc.
Nitrogen Trifluoride
NF3
FS: 4-40 PPM PN: MIDAS-X-HFX SN: 2564-0243 Expires: Nov2006
zellweger analytics
z
zellweger analytics
z
GAS DETECTOR
midas
5.91 [150.00]
4.72 [120.00]
2.48 [63.00]
Diagram 7. MIDAS® outline dimensions
Drill Template
Drill 2 x M4 holes
58.50mm (2.3”)
11
Below are some considerations when installing the MIDAS® Gas Detector:
1. Mount the detector on a surface that is flat, firm and suitable for its size and weight.
2. Use the drill template supplied to drill the holes for the fixings.
3. Use fixings appropriate for the surface being mounted to.
4. Ensure the head size of fixings used will not snag the terminal PCB.
5. Consider the conduit/cable weight and its stress on the installation.
6. Position the detector so that it can be easily accessed.
7. Position the detector so that it is at a suitable height (normally eye level) for the display to be clearly seen.
8. Take into consideration the space required to remove the detector’s cover and locking/unlocking the sensor cartridge locking clips.
9. Take into consideration the space required for sample inlet and exhaust tubing (for remote monitoring), and for the inlet filter (for local monitoring).
10. Take into consideration the space required for cable or conduit access.
5.2 Mechanical Installation
The following steps and diagrams show how to separate the mounting bracket assembly from the main chassis and mount it on a vertical flat surface.
1. Unscrew the thumbscrew located on the front panel.
2. Remove the cover by pulling it forwards off the main chassis.
3. Unscrew the two retaining screws located at the bottom front of the chassis.
4. Holding the mounting bracket assembly with one hand use the other to carefully pull the main chassis forwards to disconnect it from the mounting bracket assembly.
5.
Using the drill template provided drill two holes 58.50 mm vertically apart for 2 x round head M4 fixing screws.
6. Partially screw the fixings into the mounting surface.
7. Place the mounting bracket assembly over the screws so they pass through the mounting holes and then slide down to locate in the slots.
8. Tighten the screws to secure the mounting bracket assembly.
Diagram 8. Mechanical installation
12
5.3 Sample and exhaust tubing calculations
The following tables show the flow rate, tubing length, transport time, and maximum pressure and vacuum at the inlet and exhaust points.
Table 2. Inlet sample specifications
Inlet Sample Specifications:
Maximum
Tubing Length, m (ft)
30 (100) 20 (66) 10 (33)
0
Transport Time (sec)
28 19 10 1
Sample Point Vacuum (Negative pressure)
-25.4 cm H
2
O (-10 in H2O) Maximum
Flow rate, cc/min.
500 (Flow is constant)
1
Tubing OD, mm (in)
6.35 (0.25)
Tubing ID, mm (in)
3.18 (0.125)
1
NOTE: The flow rate is electronically maintained at 500 cc/min
Table 3. Outlet sample specifications
Outlet Sample Specifications:
Maximum
Tubing Length, m (ft)
30 (100) 20 (66) 10 (33)
0
Back Pressure at Exhaust Point (Excluding tubing)
20.3 cm H
2
O (8 in H2O) Maximum
Tubing OD, mm (in)
6.35 (0.25)
Diagram 8. Mechanical installation
ATTACHING BRACKET TO SCREWS ON A WALL
mounting bracket
screws
13
5.4 In-line filters
External filters must be used to protect the tubing from contamination. Use particulate filter part number 0780248 for normal gases and 1830-0055 for corrosive gases. Replace the filter every 3 months. Refer to the table below for specific gases.
Target Gas
Recommended
maximum sample
line (m)
End of line
particulate filter
(P/N 0780248)
End of line particulate
Teflon filter
(1830-0055)
Ammonia
10 YES **
Arsine
30 YES **
Boron Trichloride
5* NO YES
Boron Trifluoride
5* NO YES
Bromine
10 NO YES
Carbon Dioxide
30 YES **
Carbon Monoxide
30 YES **
Chlorine
10 NO YES
Chlorine Dioxide
10 NO YES
Chlorine Trifluoride
5* NO YES
Diborane
30 YES **
Dichlorosilane
5* NO YES
Disilane
10 YES **
Fluorine
10 NO YES
Germane
30 YES **
Hydrogen (% LEL)
30 YES **
Hydrogen (ppm)
30 YES **
Hydrogen Bromide
5* NO YES
Hydrogen Chloride
5* NO YES
Hydrogen Cyanide
10 YES **
Hydrogen Fluoride
5* NO YES
Hydrogen Selenide
30 YES **
Hydrogen Sulfide
30 YES **
Methane (% LEL)
30 YES **
Nitrogen Dioxide
30 NO YES
Nitrogen Oxide
30 YES **
Nitrogen Trifluoride
30 YES **
Oxygen Proficiency & Deficiency
30 YES **
Ozone
5 NO YES
Phosphine
30 YES **
Phosphorous Oxychloride
5* NO YES
Table 4. In-line filter recommendations
14
Target Gas
Recommended
maximum sample
line (m)
End of line
particulate filter
(P/N 0780248)
End of line particulate
Teflon filter
(1830-0055)
Silane
30 YES **
Silane (low level)
30 YES **
Sulfur Dioxide
30 NO YES
Sulfur Tetrafluoride
5* NO YES
Tetra Ethyl Ortho Silicate
30 YES **
Tungsten Hexafluoride
5* NO YES
5.5 Local Detector Option
The MIDAS
®
gas detector can also be used to monitor for gas at the location of the detector. To do this an inline filter is simply connected to the sensor cartridge gas inlet port. The external dust filter part number is 780248 for normal gases and 1830-0055 for corrosive gases. The area around the detector is then being monitored as opposed to a sample being drawn from a remote location.
Diagram 9. Local gas detector option
MIDAS-T-001 transmitter installed with in line particulate filter for local ambient monitoring mode
* Recommend to keep the sample lines as short as possible where the RH condition at the sample point is high (above <50%RH); there will be some sample loss due to absorption onto the sample line. ** These gases can use either filter.
Table 4. In-line filter recommendations (cont.)
15
5.6 Electrical Installation
The terminals used are suitable for conductors of 24 to 14 AWG (0.5 to 1.8mm Dia.). The use of 16 AWG (1.5 mm Dia.) conductors is recommended.
NOTE: When powered by Power over Ethernet (PoE) the 3 on board relays do not require separate 24 VDC to be energized
If Power over Ethernet (PoE) is used to power the device, then 24 VDC power must not also be connected to the device, (or conversely if 24 VDC is used to power the MIDAS
®
, then electrical power via the Ethernet port must not be applied). Failure to observe this requirement may cause damage to the gas detection system and will not be covered by the standard warranty.
When connecting the wires ensure that the power switch is in the off position.
Diagram 11 shows the terminal module layout and terminal identification as well as the jumper locations. Diagnostic LED lamps are positioned on the internal main chassis for ease of viewing. These indicate that Ethernet communication activity is active. In some versions of MIDAS
®
the rear mounted Power over Ethernet port
has two LED status lamps; note that these are not utilized and do not illuminate.
24V DC
RELAYS
OUT
COM
EXT INT
CABLE GLAND
WIRES
RJ45 CONNECTOR
Access for the electrical wires to the terminal module is made via the PG16 cable gland located at the bottom of the mounting bracket assembly. The cable gland can be removed and replaced with a suitable conduit fitting if required. The wire routing of a typical installation is shown in the diagram below.
Diagram 10. Typical wire routing
ANALOG OUTPUT
ALM 1
ALM 2
LT
F
24V DC IN
0V DC COM
16
Diagram 11. MIDAS® terminal layout and identification
See relay function table below
Display Description Relay 1 Relay 2 Relay 3
1FLt
Instrument Fault
Only
Alarm 1 Alarm 2 Instrument Fault
2Flt
Separate Fault
Relays
Any Alarm Maintenance Fault Instrument Fault
CmbF
Combined Fault
Relay
Alarm 1 Alarm 2 Any Fault
NOTE: EARTHING REQUIREMENTS
If the MIDAS® unit metal chassis is not connected directly to a metal surface for earthing purposes, an additional earth wire will be required. Connect a wire via the PG16 gland to the dedicated earth tag (screw terminal) located on the bottom bracket and connect the other end of the wire to a dedicated external earthing point.
If Power over Ethernet (PoE) power supply is being used, shielded CAT5 ethernet cable is recommended.
Please ensure that your wiring avoids earth ground loops that may affect the performance of your equipment.
17
5.7 Electrical Connections
Diagram 14. Example MIDAS® and 5704 connection
Diagram 15. Example MIDAS® and 5701 connection
Diagram 16. Generic example MIDAS
®
4-wire isolated output
Diagram 17. Generic Example MIDAS
®
3-wire sourcing output
Diagram 12. Example MIDAS® and Touchpoint 1 connection
Diagram 13. Example MIDAS® and Touchpoint 1 connection
®
MIDAS
can be powered by either 24 VDC via traditional discrete wiring or by approximately 48 VDC delivered through the Ethernet cable from a PoE source. In either case the 4-20 mA analog output can be used. This can be configured for sinking, sourcing or isolated operation. Below are some schematic diagrams of typical electrical connection configurations.
18
Diagram 18. Generic example MIDAS
®
3-wire sinking output
5.8 Refitting the Main Chassis
The main chassis can be refitted to the mounting bracket assembly using the following steps.
1. Align the PCB at the top rear of the main chassis with the connector located at the top of the mounting bracket assembly
2. At the same time align the two tubes at the bottom rear of the main chassis with the two tubes located on
the bottom of the mounting bracket assembly.
3. Slide the chassis backwards on the mounting bracket assembly so that the PCB and connector and tubes
engage simultaneously. (See diagram below).
4. Ensure the PCB, connector and tubes are fully engaged by firmly pushing the main chassis horizontally
backwards on the mounting bracket assembly (WARNING: DO NOT PUSH ON THE LCD AS THIS MAY CAUSE DAMAGE).
5. Align the two fixing screws located at the bottom of the chassis with the screw threads on the mounting
bracket assembly.
6. Tighten the screws to secure the chassis to the mounting bracket assembly.
Diagram 19. Refitting the chassis.
Diagram 19. Generic Example MIDAS® sourcing 4-20 mA with PoE power
Diagram 20. Generic example MIDAS
®
isolated 4-20 mA
output with PoE power
19
5.9 Installing the Sensor Cartridge
The MIDAS
®
sensor cartridge is supplied separately and needs to be fitted to the detector’s main chassis. The following steps and diagrams detail the procedure for installing the sensor cartridge for the first time. This procedure is carried out with the power off and the detector cover removed.
1. Remove sensor cartridge from packaging
2. Check the part number and type of sensor cartridge is correct.
3. Check the activate by date.
4. Align the pins at the top of the sensor cartridge with the socket in the sensor cartridge chamber.
5. Carefully push the sensor cartridge into the sensor cartridge chamber until fully home.
6. Lock the sensor cartridge in place using the tabs either side of the sensor cartridge to lock the sensor cartridge to the main chassis.
7. Switch the power switch located on the terminal module to the ‘on’ position.
8. Refit the detector’s cover by aligning the slots either side with the locating tabs on the mounting bracket assembly.
9. Push the cover horizontally until home.
10. Tighten the thumbscrew located on the front panel.
Diagram 20. Installing the sensor cartridge
6 DETECTOR START UP PROCEDURES
WARNING
Prior to carrying out any work ensure local and site procedures are followed. Ensure that the associated control panel is inhibited so as to prevent false alarms.
The following procedure should be followed carefully and only performed by suitably trained personnel.
1. Ensure the detector is wired correctly according to sections 5.6 and 5.7.
2. Ensure that the correct sensor cartridge is fitted. (If the cartridge has not been stored at room temperature, allow one hour for equilibration.)
3. Ensure the on/off switch on the mounting bracket assembly is in the on position.
4. Apply power to the system.
5. After the start up routine the detector will display normal operating mode as shown in section 7.
6. If using a multi gas sensor cartridge refer to section 8.2.2 to ensure the correct gas id code is selected.
7. Allow the detector to stabilize using the table over to determine the maximum sensor cartridge warm up time.
m
idas
G
a
s Se
n
s
o r
C
a
rt
r
i d
g
e
N
IT
R
OG
E
N
TR
I
F
LU
OR
I
D
E
N
F
3
P
N
:
M
I
D
A
S-S-
H
F
X
SN
: 02
3
4
4
56
6
7
0
Us
e
b
y
:
07
-
09
-
2
006
ze
ll
w
e
g
er
a
na
l
y
t
i
c
s
HOW CARTRIDGE IS FITTED
chassis
cartridge
m
i d
a
s
G
a s
S
en so
r
C
a rt
r i d
g e
NITROG
E N
T
R I F
LU
O R
I DE
N
F3
P
N
:
M
I
D A
S-
S
-H FX
SN
:
02
3 44
5
6 6
70
U se
by
:
0 7
­09
-20
0
6
z
el
l w
eg
e
r
a
n a
l y t
i cs
slots
tighten
HOW COVER IS FITTED
tabs
cartridge fitted
20
Sensor Cartridge Family
Sensor Cartridge Part No.
Gas Name
Maximum
Warm-up
Time
(minute)
Ammonia
MIDAS-S-NH3
Ammonia 10
Hydrides
MIDAS-S-ASH
Arsine 20
MIDAS-S-HYD Diborane, Germane
20
MIDAS-S-PH3
Phosphine 20
MIDAS-S-SHX Silane (20 ppm), Disilane
20
MIDAS-S-SHL Silane (2 ppm)
20
MIDAS-S-HSE Hydrogen Selenide
20
Mineral Acid (HF)
MIDAS-S-HFX
Hydrogen Fluoride, Boron Trifluoride, Nitrogen Trifluoride, Tungsten Hexafluoride
20
MIDAS-S-SF4 Sulfur Tetrafluoride, Chlorine Trifluoride
20
Mineral Acid (HCl)
MIDAS-S-HCL
Hydrogen Chloride, Boron Trichloride, Dichlorosilane, Hydrogen Bromide
20
MIDAS-S-POC Phosphorous Oxychloride
20
Halogens
MIDAS-S-HAL Chlorine, Fluorine
10
MIDAS-S-BR2 Bromine, Chlorine Dioxide
20
Carbon Dioxide
MIDAS-S-CO2 Carbon Dioxide
20
Carbon Monoxide
MIDAS-S-COX Carbon Monoxide
10
Hydrogen
MIDAS-S-H2X
Hydrogen 10
Hydrogen Cyanide
MIDAS-S-HCN Hydrogen Cyanide
10
Hydrogen Sulfide
MIDAS-S-H2S Hydrogen Sulfide
10
Nitric Oxide
MIDAS-S-NOX Nitric Oxide
10
Nitrogen Dioxide
MIDAS-S-NO2 Nitrogen Dioxide
10
Oxygen
MIDAS-S-O2X
Oxygen 30
Ozone
MIDAS-S-O3X
Ozone 10
Sulfur Dioxide
MIDAS-S-SO2 Sulfur Dioxide
10
TEOS
MIDAS-S-TEO
TEOS 10
Flammables
MIDAS-S-LEL Hydrogen, Methane
30
Table 5. Sensor cartridge warm up times
21
7 GENERAL OPERATION
After applying power to the detector, the display will go through a start up test routine illuminating in sequence all the LEDs, icons and digits of the display. The display will show the message ‘WAIT’ and ‘LOAD’ as it checks for cartridge data. It will then display the message ‘WARM’ until the sensor cartridge to reaches operating temperature. When complete the detector will enter normal monitoring mode indicated by the ‘ ’ icon on the display cycling through three states (2 rings, 3 rings, 4 rings). The measured gas concentration will be shown on the display and transmitted on the 4-20 mA output. The green LED will flash once every second indicating power and the sample flow rate indicator will be displayed. If monitoring is interrupted due to a fault, a test or calibration process or a user requested inhibit, the ‘ ’ icon will flash. For details of fault and maintenance fault codes refer to section 12.
Final Start Up Routine Screen Normal Operation
7.1 Normal Operation Mode
In this mode the detector displays gas concentration, alarm, fault and status information via its backlit LCD and front panel LEDs. Typical normal operation display and output states are shown below. See Section 12 for a full list of fault codes.
NOTE: The examples below are for a linear 4-20 mA output over a full scale range of 2 ppm. The current output for a given gas concentration will be different for other full scale ranges (linear 4 mA = 0 % full scale to 20 mA = 100 % full scale).
See section 12 for a full list of fault codes.
Operational State Relay status 4-20 mA output
(for 2ppm range)
LEDs Display*
Normal operation Alarm relay 1
de-activated
Alarm relay 2 de-activated
Fault relay activated
4 mA Green slow flash
Alarm 1 Alarm relay 1
activated
Alarm relay 2 de-activated
Fault relay activated
6 mA
Green slow flash
Red on
Table 6. Normal operation display and output states.
22
Table 6. Normal operation display and output states. (cont.)
Operational State Relay status 4-20 mA output
(for 2ppm range)
Alarm 2 Alarm relay 1
de-activated
Alarm relay 2
8 mA
activated
Fault relay activated
Greater than full scale
Alarm relay 1 activated
Alarm relay 2
21 mA
activated
Fault relay activated
Inhibit Alarm relay 1
de-activated
Alarm relay 2
2 mA Green slow flash
de-activated
Fault relay activated
Low flow rate Alarm relay 1
de-activated
Alarm relay 2
3 mA
de-activated
LEDs Display*
Green slow flash
Red fast flash
Green slow flash
Red fast flash
Green slow flash
Yellow on
Fault relay activated
Low flow fault Alarm relay 1
de-activated
Alarm relay 2 de-activated
Fault relay
1.0 mA
Green slow flash
Yellow fast flash
de-activated
7.1.1 Resetting Alarms, Faults and Maintenance Faults
The alarm relay function of MIDAS
®
can be set to latching (see set-up mode) so that when an alarm occurs the associated relay remains activated even if the gas reading has dropped below the alarm level. To reset the latched alarm relay press the ‘X’ cancel button for 2 seconds. If the gas level is still above the alarm point you can reset the associated relay but it will activate again after the elapse of any alarm on delay that has been set.
The fault relay function can also be set to latching. It can be reset by pressing the ‘X’ cancel button only if the fault condition has cleared. The fault relay cannot be reset if the fault condition is still present.
The maintenance faults that are displayed on the LCD follow the latching or non-latching function set for the fault relay. If the fault relay operation is set to non-latching, the display will automatically clear a maintenance fault message when the maintenance fault condition is cleared. If set to latching the maintenance fault message can be cleared by pressing the ‘X’ cancel button only if the maintenance fault condition has cleared.
7.2 Review Mode
The detector settings can be reviewed safely without the possibility to make changes by using review mode. To select review mode press the ‘
’ up or ‘’ down button once. The review mode icon ‘ ’ will show on the
display and the first review mode menu icon is displayed.
23
Review
Submenu
OK
OK
OK
OK
OK
OK
Software
SW
Displays software
revision
Displays
software
checksum
Press or to select next
menu or X
to exit
Alarms
Displays
sensor
cartridge
ID (X) and gas ID code (Y) in format
(X -Y)
Level 1 (L1)
rising (U) or
falling (d)
alarm type
followed by set point then same for level
(L2) alarm
Alarm time delay (secs)
Alarm relays
latching (L) or
non latching
(nL)
Alarm relays
normally
energized (nE)
or normally
de-energized
(nd)
Press or to select next
menu or X
to exit
Faults
Fault relay
configuration
Fault relay
latching (L) or
non latching
(nL)
Fault relay
normally
energized (nE)
or normally
de-energized
(nd)
Press or to select next menu or X
to exit
Calibration
Days
remaining
to next
calibration
due
Year of last
calibration
Month and
day of last
calibration
Press or to select next
menu or X
to exit
Date/Time
timE
Year
Date
Time
Press or to select next
menu or X
to exit
Address
Auto address detect (AU Y)
or manual set
(AU n)
IP address
segments
(x4)
Sub net
address
segments
(x4)
Press or to select next
menu or X
to exit
Event Log
Number
of dots on
display shows
number of
viewable
events (0-7)
Date of first
event
Time of event
Icon and
event code
Repeat for
next event or
X to exit
▲ ▼
OK
OK
OK
OK
OK
OK
Table 7. Review menu overview.
7.2.1 Review Mode Menu Overview
The menu is simply navigated by using the ‘’ up and ‘’ down buttons to select the required menu, and then using the ‘
’ accept button to enter that submenu and scroll through to view the settings. The ‘X’ cancel button can be used to exit the submenu and allow selection of a different submenu, or can be pressed again to exit to normal operating mode. When in review mode the unit will automatically return to the main normal operation status display if either an alarm level is exceeded or no button is pressed for 60 seconds.
The table below shows an overview of the different review menus and how they are navigated. For a detailed step-by-step instruction of how to review the detector setting in review mode refer to section 8.1.
24
7.3 Overview of Set-up, Calibration and Test Mode
WARNING: Set-up, calibration and test modes are intended for use by trained personnel or service engineers only. Access to these modes can be pass code protected by following the procedure in section
8.2.7.
Set-up, calibration and test modes are used to make setting changes, calibrate and test the detector. To select set-up, calibration and test mode press and hold the ‘
’ up button or ‘’ down button for a second. The unit will automatically go to the main normal operation status display from setup/calibration/test menus (but not from inside a setup/calibration/test function) if no button is pressed for 5 minutes or if an alarm level is exceeded.
PASS CODE: If a pass code has been set the display will show 0000 with the first 0 flashing. Use the ‘
’ up or ‘’ down buttons to set the first digit of the pass code. Press ‘’ to enter the first digit. The second digit will then flash. Repeat the process until all four pass code digits have been entered. Please record your pass code in a separate archive that can be securely retrieved. Failure to be able to retrieve your pass code may lead to delays in gaining access to all the protected functions in each MIDAS
®
unit. If an incorrect code is entered the display will show ‘Err’ and return to the normal operation mode. If a pass code is forgotten contact your local Zellweger Analytics service department.
After successfully entering the pass code (if set) the first menu ‘
SET’ set-up icon will show on the display.
The ‘
CAL’ calibration or ‘ tESt’ test menu can also be selected using the ‘’ up or ‘’ down buttons.
Press the ‘
’ accept button to enter the selected menu or the ‘X’ cancel button to return to normal operation
mode.
7.3.1 Set-up Menu Overview
The set-up menu allows changes to be made to the detector alarm, fault, calibration interval, date/time and digital address settings. The menu is simply navigated using the ‘’ up and ‘’ down buttons to select the required submenu and then using ‘
’ accept button to enter that submenu. The ‘’ up and ‘’ down buttons
are used to make changes to the selected setting and are confirmed using the ‘
’ accept button. The ‘X’ cancel button can be used to exit the submenu and allow selection of a different submenu, or can be pressed again to exit to the main set-up, calibration and test menu. Pressing the ‘X’ cancel button again returns the detector to normal operating mode.
NOTE: All settings in a submenu are accepted when the ‘
’ accept button is pressed after the last submenu setting. This saves the changes and is indicated by displaying ‘UPdt’ on the LCD. If however the ‘X’ cancel button is pressed at any time before the changes are accepted, this will cause any changes to be cancelled in that particular submenu.
The table opposite shows an overview of the set-up submenus and how they are navigated. For a detailed step-by-step instruction of how to change the detector settings using the set-up menu refer to section 8.2.
25
Table 8. Set-up menu overview.
Set up
Submenu
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
Alarms
Change gas
ID code (only
for multi
gas sensor
cartridge)
Set level 1
(L1) rising (U)
or falling (d)
alarm type. Set alarm set point. Repeat
for level (L2)
alarm
Set alarm
time delay
(X-X secs)
Set alarm
relays latching
(L) or non
latching (nL)
Set alarm
relays
normally
energized (nE)
or normally
de-energized
(nd)
Press or to select
next menu or
X for set-up,
calibration
and test
menu
Faults
Set fault relay configuration
(1FLt, 2FLt,
CmbF)
Set fault relay
latching (L) or
non latching
(nL)
Set fault relay
normally
energized (nE)
or normally
de-energized
(nd)
Press
or to select
next menu or X for set-up,
calibration and
test menu
Calibration
Set calibration
interval
(0-365 days)
Press or
to select
next menu or
X for set-up,
calibration
and test menu
Date/Time
timE
Set date format (dd: mm or mm:
dd)
Set year
(2003-2030)
Set month
(1-12)
Set day (1-31)
Set hours
(00-23)
Set minutes
(0-59)
Press
or to select next menu or X for set-up, calibration
and test
menu
Set Address
Set auto
address detect
(AU Y) or
manual set
(AU n)
If AU Y,
address is
detected,
then returns
to set-up,
calibration
and test
menu. If AU n, set the first part of the IP
address (0-255)
Repeat setting
address for 2nd, 3rd and 4th segments
of IP address.
Set the 4 sub net addresses (0-255) using same method
Press or
to select
next menu or
X for set-up,
calibration
and test menu
Set Pass
Code
Press or
to set pass
code 1st digit
Press or
to set pass
code 2nd
digit
Press or
to set pass
code 3rd
digit
Press or
to set pass
code 4th digit
Repeat to
confirm pass
code
Pass code
saved if both
entries are
the same
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
26
7.3.2 Calibration Menu Overview ‘ CAL’
The calibration menu allows calibration of the detector zero, span, flow and 4-20 mA. The menu is simply navigated using the ‘’ up and ‘’ down buttons to select the required submenu and then using ‘’ accept button to enter that submenu. The ‘
’ up and ‘’ down buttons are used to make any changes to a selected
setting and are confirmed using the ‘
’ accept button. The ‘X’ cancel button can be used to exit the submenu and allow selection of a different submenu, or can be pressed again to exit to the main set-up, calibration and test menu. Pressing the ‘X’ cancel button again returns the detector to normal operating mode.
The table below shows an overview of the calibration submenus and how they are navigated. For a detailed step-by-step instruction of how to change the detector calibration settings refer to section 8.2.
Calibration
submenu
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
Zero
Icon flashes alerting user to prepare to
apply zero gas
Level 1 (L1)
rising (U) or
falling (d) alarm type followed by
set point then
same for level
(L2) alarm
If ok ‘’PASS’’
is displayed. If
fail fault code
is displayed.
Press or to select
next menu or
X for set-up,
calibration and
test menu
Span
Select gas ID code of calibration
gas (for multi
gas ID sensor
cartridges only)
Select if
humidified
‘HUm’ or
dry ‘drY’
calibration gas
Adjust value
to display span gas
concentration
being used
Display goes
steady and
displays span
gas reading.
The dots
indicate span
progress. If
ok ‘’PASS’’ is
displayed. If
fail fault code
is displayed
Press or
to select
next menu or
X for set-up,
calibration and
test menu
Flow
Icon flashes and display
shows ‘0’
indicating that
flow zero will
be set
Unit counts down from
10 to 0 and sets flow zero. Display shows
1st set point
target flow rate.
Use or to
make reading
on external
flow meter
350cc/min +/-
50cc/min
Use or
to change the
flashing display
to the actual
reading from
the external
flow meter
Unit counts
down from 10
to 0 and sets
1st set point.
Icon flashes
and displays
2nd set point target. Repeat process to set.
Press or
to select
next menu or
X for set-up,
calibration and
test menu
4-20 mA
4 mA is
displayed
indicating
analog output
should be
4 mA
Adjust until
analog
output is
4 mA
20 mA is displayed
indicating
analog output
should be
20 mA
Adjust until
analog output
is 20 mA
Press or
to select
next menu or
X for set-up,
calibration and
test menu
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
Table 9. Calibration menu overview.
27
7.3.3 Test Menu Overview ‘ tEST’
The test menu is used to test the detector gas reading using bump test gas, and for simulation of alarm and fault display and output operation. The test menu also contains the detector Inhibit facility. Use the ‘
’ up and ‘
down buttons to select either bump test or alarm/fault test. Press the ‘
’ accept button to enter that submenu.
The table below shows an overview of the test submenus and how they are navigated. For a detailed step-by-step instruction of how to test the detector operation refer to section 8.2.
Test
Submenu
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
Bump
Apply bump
test gas and display shows measured gas concentration with all alarm
outputs inhibited
Press ‘X’ to exit
to test menu
Press or to select next
submenu or X
for to return
to the set-up,
calibration and
test menu
Alarm/fault
Display shows ‘Sim’ and the
A1 ‘
’ symbol.
Select either
A1, A2 ‘
’ or
Fault ‘
’ for
simulation
Display shows
‘SuRE’.
Display flashes
‘on’ and
simulates the selected A1, A2 or Fault display
and output.
Press x to exit
and select
another
simulation or x
again to return
to test submenu
Press or to select next
sub menu or
X for set-up,
calibration and
test menu
Inhibit
Press or to select ALm,
ALm-Ft, ALL
or nonE inhibit
state
Set inhibit
timeout period
UPdt is
displayed and
unit enters
selected inhibit
state. Display
returns to
submenu
Press or
to select next sub menu or
X for set-up,
calibration and
test menu
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
OK
▲ ▼
Table 10. Test Submenu
28
8 DETAILED PROCEDURES FOR NAVIGATING REVIEW, SET-UP, CALIBRATION
AND TEST MODE SUBMENUS
The following sections provide step-by-step procedures for navigating all the modes and submenus.
8.1 Review Mode
Review mode allows the settings of the detector to be reviewed safely without any changes being made. Review mode consists of 7 submenus as shown in the table below:
Review submenu
Icon Settings Displayed
Software
SW Revision and checksum
Alarms
Sensor cartridge and gas ID codes, Alarm 1 and alarm 2 configuration and set points, time delay, latching/non latching, energized/de-energized
Faults
Configuration, Latching/non latching, energized/de-energized
Calibration
Days left to next calibration due and date of last calibration
Time/Date
timE Current year, date and time
IP address
Auto address selection on/off, IP address, sub net mask values.
Event log
View the date, time and details of the last seven alarm, fault and maintenance fault events
Table 11. Review mode submenus.
Review mode allows the settings of the detector to be reviewed safely without any changes being made. Review mode consists of 7 submenus as shown in the table below:
To select review mode from normal operation, press the ‘
’ up or ‘’ down button once. The icon will be
displayed along with the first submenu icon. Press the ‘
’ up or ‘’ down buttons to select a different submenu
icon.
8.1.1 Review Software ‘SW’
1. Select the software ‘SW’ submenu and press ‘
’ to accept.
2. The software version number is displayed.
3. Press ‘
’ to display the software checksum
4. Press ‘
’ to return to step 1.
5. Repeat steps 1 through 4 to view the information again or press ‘
’ up or ‘’ down to select another
submenu.
6. Press ‘X’ to return to normal operation
29
8.1.2 Review Alarms ‘ ALm’
1. Select the alarms ‘
Alm’ submenu and press ‘’ to accept.
2. The sensor cartridge ID (X) and the set gas ID code (Y) are displayed in the format (X – Y). (See section 8.2.2 for a list of sensor cartridge and gas ID codes).
3. Press ‘
’ to display if the level 1 alarm (L1) type is rising (U) or falling (d).
4. Press ‘
’ to display the alarm 1 value along with the A1 icon‘’.
5. Press ‘
’ to display if the level 2 alarm (L2) type is rising (U) or falling (d).
6. Press ‘
’ to display the alarm 2 value along with the A2 icon ‘ ’.
7. Press ‘
’ to display the alarm on delay (seconds).
8. Press ‘
’ to display if the alarm relays are set to latching (L) or non latching (nL).
9. Press ‘
’ to display if the alarm relays are set to normally energized (nE) or normally de-energized (nd).
10. Press ‘
’ to return to step 1.
11. Repeat steps 1 through 7 to view the settings again or press ‘
’ up or ‘’ down to select another submenu.
12. Press ‘X’ to return to normal operation.
8.1.3 Review Faults ‘
FLt’
1. Select the faults ‘
FLt’ submenu and press ‘’ to accept.
2. The fault relay configuration (1FLt, 2 FLt or CmbF) is displayed. (See section 8.2.3 for details of fault relay configurations).
3. The fault relay latching (L) or non latching (nL) setting is displayed.
4. Press ‘
’ to display the fault relay normally energized (nE) or normally de-energized (nd) setting.
5. Press ‘
’ to return to step 1.
6. Repeat steps 1 through 4 to view the settings again or press ‘
’ up or ‘’ down to select another submenu.
7. Press ‘X’ to return to normal operation.
8.1.4 Review Calibration ‘
CAL’
1. Select the review calibration ‘
CAL’ submenu and press ‘’ to accept.
2. The number of days remaining to the next calibration due date is displayed.
3. Press ‘
’ to display the year and press ‘’ again to display the month and day of the last calibration.
4. Press ‘
’ to return to step 1.
5. Repeat steps 1 through 4 to view the settings again or press ‘
’ up or ‘’ down to select another submenu.
6. Press ‘X’ to return to normal operation.
8.1.5 Review Date and Time ‘
timE’
1. Select the review date and time ‘timE’ submenu and press ‘
’ to accept.
2. The current year setting is displayed.
3. Press ‘
’ to display the current month and day
4. Press ‘
’ again to display the current time.
5. Press ‘
’ to return to step 1.
6. Repeat steps 1 through 5 to view the settings again or press ‘
’ up or ‘’ down to select another submenu.
7. Press ‘X’ to return to normal operation.
30
8.1.6 Review Detector Address ‘ nET’
1. Select the review address ‘
nEt’ submenu and press ‘’ to accept.
2. ‘AU Y’ or ‘AU n’ is displayed depending on if auto address detection is on (AU Y) or off (AU n).
3. Press ‘
’ to display the first part of the IP address preceded by the letter ‘A’.
4. The dot on the upper left indicates that the first part of the IP address is being displayed.
5. Press ‘
’ to display the second part of the address. The two dots on the upper left indicate the 2nd portion of
the address.
6. Press ‘
’ again to display the third part and again to display the last part of the address.
7. Press ‘
’ and the first part of the sub net mask values will be displayed preceded by the letter ‘n’.
8. Press ‘
’ to scroll through the second, third and forth sub net values. The number of dots in the upper left
indicate which part of the sub net address is being viewed.
9. Press ‘
’ to return to step 1.
10. Repeat steps 1 through 9 to view the settings again or press ‘
’ up or ‘’ down to select another submenu.
11. Press ‘X’ to return to normal operation.
8.1.7 Review Event Log ‘
Hi St’
1. Select the review event log ‘
Hi St’ submenu.
2. The number of logged data events available is indicated by the number of dots (0-7) shown on the left of the display.
3. Press ‘
’ to view the date of the last recorded event.
4. Press ‘
’ to display the time of the event.
5. Press ‘
’ again to display the relevant alarm icon and event code.
6. Repeat to view next logged event details.
7. Press ‘X’ to return to normal operation.
Note: More detailed event log information can be viewed using the web browser feature. Refer to section 14 for further details
8.2 Set-up, Calibration and Test Modes
WARNING: Set-up, calibration and test modes are intended for use by trained personnel or service engineers only. Access to these modes can be pass code protected by following the procedure in section 8.2.7.
Set-up, calibration and test modes are used to make setting changes, calibrate and test the detector. To select set-up, calibration or test mode press and hold the ‘
’ up button or ‘’ down button for a second. The unit will automatically go to the main normal operation status display from setup/calibration/test menus (but not from inside a setup/calibration/test function) if no button is pressed for 5 minutes or if an alarm level is exceeded.
PASS CODE: If a pass code has been set the display will show 0000 with the first 0 flashing. Use the ‘
’ up
or ‘
’ down buttons to set the first digit of the pass code. Press ‘’ to enter the first digit. The second digit will then flash. Repeat the process until all four pass code digits have been entered. If an incorrect code is entered the display will show Err and return to the normal operation mode. If a pass code is forgotten contact your local Zellweger Analytics service department.
Zellweger Analytics is not responsible for any costs associated with the recovery of the passcode in the event that the user has forgotten or cannot access their unique passcode. It is strongly advised that the user records all pass codes with the instrument serial number in a secure and separate location to the MIDAS
®
unit. Delays in
recovering the lost passcodes may be experienced and are not the responsibility of Zellweger Analytics.
NOTE: Should the passcode be activated, the unit will continue to detect gas and indicate maintenance and instrument faults.
After successfully entering the pass code (if set) the first menu ‘ SET’ set-up icon will show on the display. The ‘
CAL’ calibration or ‘ tESt’ test menu can also be selected using the ‘’ up or ‘’ down buttons. Press the ‘
accept button to enter the selected menu or the ‘X’ cancel button to return to normal operation mode.
31
8.2.1 Set-up Menu ‘ SEt’
The set-up menu allows the settings of the detector to be changed. The set-up menu consists of 6 submenus as shown in the table below.
Table 12. Set-up mode submenus.
Set-up submenu
Icon Changeable settings
Alarms
Gas ID, Alarm 1 and alarm 2 configuration, set points, time delay, latching/non latching, energized/de-energized
Faults
Configuration, Latching/non latching, energized/de-energized
Calibration
Calibration interval (days)
Time/Date
timE Date format mm:dd or dd:mm, current year, month, day, hours, minutes
IP address
Auto address selection on/off, IP address, sub net mask values.
Pass code
Set pass code
NOTE: All settings in a submenu are accepted when the ‘’ accept button is pressed after the last submenu setting. This saves the changes and is indicated by displaying ‘UPdt’ on the LCD. If however the ‘X’ cancel button is pressed at any time before the changes are accepted, this will cause any changes to be cancelled in that particular submenu.
To select the set-up menu from normal operation, press the ‘
’ up button for a few seconds. Enter the pass code
(if set). Use the ‘
’ up or ‘’ down buttons to select the set-up menu ‘ ’ icon and press the ‘’ accept button.
8.2.2 Set Alarms ‘
ALm’
1. Use the ‘
’ up or ‘’ down buttons to select the set alarms ‘ ALm’ submenu and press ‘’ to accept.
2. The flashing gas id code is displayed along with the gas cylinder and alarms icon ‘
’.
3. Use the ‘
’ up or ‘’ down buttons to change the gas ID number (only applicable on multi gas sensor
cartridges- see table below).
4. Press ‘’ to accept.
5. The flashing level 1 (L1) alarm type is displayed (U) rising or (d) falling.
6. Use the ‘’ up or ‘’ down buttons to change the alarm type.
7. The flashing alarm 1 value is displayed along with the icon ‘
’.
8. Use the ‘
’ up or ‘’ down buttons to change the value.
9. Press ‘’ to accept.
10. The flashing level 2 (L2) alarm type is displayed (U) rising or (d) falling.
11. Use the ‘’ up or ‘’ down buttons to change the alarm type.
12. Press ‘’ to display the flashing alarm 2 value along with the icon ‘ ’.
13. Use the ‘’ up or ‘’ down buttons to change the A2 value.
14. Press ‘
’ to display the flashing alarm on time delay (seconds).
32
15. Use the ‘’ up or ‘’ down buttons to change the alarm on time delay (seconds).
16. Press ‘
’ to display the flashing alarm relay latching (L) or non latching (nL) setting.
17. Use the ‘
’ up or ‘’ down buttons to change between the settings.
18. Press ‘
’ to display the flashing alarm relays normally energized (nE) or normally de-energized (nd) setting.
19. Use the ‘
’ up or ‘’ down buttons to change between the settings.
20. Press ‘
’ to update all the changes (UPdt displayed) and return to step 1.
21. Press ‘
’ up or ‘’ down to select another submenu.
22. Press ‘X’ to return to set-up, calibration and test menu selection.
23. Press ‘X’ again to return to normal operation.
Note: When replacing a single gas sensor cartridges with the same type single gas sensor cartridge, no change of gas confirmation is required. When changing a multi gas sensor cartridge with the same type multi gas sensor cartridge, the new sensor cartridge will assume the same previously set gas ID and will not request a change gas confirmation.
Table 13. Set-up mode submenus.
Sensor cartridge Part No.
MIDAS-S-NH3
MIDAS-S-ASH
MIDAS-S-HYD
MIDAS-S-PH3
MIDAS-S-SHX
MIDAS-S-SHL
MIDAS-S-HSE
MIDAS-S-HFX
MIDAS-S-SF4
MIDAS-S-HCL
Gas Name Range Sensor
cartridge ID
Ammonia 0-100 ppm 01 Not applicable
Arsine 0-0.2 ppm 02 Not applicable
Diborane 0-0.4 ppm 03 1
Germane 0-0.8 ppm 03 2
Phosphine 0-1.2 ppm 04 Not applicable
Silane(20) 0-20 ppm 05 1
Disilane 0-20 ppm 05 2
Silane(2) 0-2 ppm 06 Not applicable
Hydrogen Selenide
Hydrogen Fluoride
Boron Trifluoride
Nitrogen Trifluoride
Tungsten Hexafluoride
Sulfur Tetrafluoride
Chlorine Trifluoride
Hydrogen Chloride
Boron Trichloride
Dichlorosilane 0-8 ppm 10 3
Hydrogen Bromide
0-0.4 ppm 07 Not applicable
0-12 ppm
0-8 ppm 08 2
0-40 ppm 08 3
0-12 ppm
0-0.8 ppm 09 1
0-0.8 ppm 09 2
0-8 ppm
0-8 ppm 10 2
0-8 ppm
08 1
08 4
10 1
10 4
Gas ID
33
MIDAS-S-POC
1
MIDAS
®
detectors are not ETL approved for monitoring in or sampling from classified areas above 25% LEL
MIDAS-S-HAL
MIDAS-S-BR2
MIDAS-S-CO2
MIDAS-S-COX
MIDAS-S-H2X
MIDAS-S-HCN
MIDAS-S-H2S
MIDAS-S-NOX
MIDAS-S-NO2
MIDAS-S-O2X
MIDAS-S-O3X
MIDAS-S-SO2
MIDAS-S-TEO
MIDAS-S-LEL
Phosphorous
0-0.8 ppm 11 Not applicable
Oxychloride
Chlorine 0-2 ppm 12 1
Fluorine 0-4 ppm 12 2
Bromine 0-0.4 ppm 13 1
Chlorine
0-0.4 ppm 13 2
Dioxide
Carbon
0-2% Vol 14 Not applicable
Dioxide
Carbon
0-100ppm 15 Not applicable
Monoxide
Hydrogen 0-1000 ppm 16 Not applicable
Hydrogen
0-20 ppm 17 Not applicable
Cyanide
Hydrogen
0-40 ppm 18 Not applicable
Sulfide
Nitric Oxide 0-100 ppm 19 Not applicable
Nitrogen
0-12 ppm 20 Not applicable
Dioxide
Oxygen 0-25% Vol 21 Not applicable
Ozone 0-0.4 ppm 22 Not applicable
Sulfur Dioxide 0-8 ppm
23 Not applicable
TEOS 0-40 ppm 24 Not applicable
Hydrogen 0-100% LEL
Methane 0-100% LEL
1
1
25 1
25 2
34
8.2.3 Set Faults ‘ FLt’
The set-up menu allows the settings of the detector to be changed. The set-up menu consists of 6 submenus as shown in the table below.
1. Use the ‘’ up or ‘’ down buttons to select the set faults ‘ FLt’ submenu and press ‘’ to accept.
2. The flashing fault relay configuration (1FLt, 2 FLt or CmbF) is displayed.
3. Use the ‘
’ up or ‘’ down buttons to change the configuration.
(See the table below for details of fault relay configurations)
4. Press ‘’ to display the flashing fault relay latching (L) or non latching (nL) setting is displayed.
5. Use the ‘
’ up or ‘’ down buttons to change between the settings.
6. Press ‘
’ to display the flashing fault relay normally energized (nE) or normally de-energized (nd) setting.
7. Use the ‘
’ up or ‘’ down buttons to change between the settings.
8. Press ‘
’ to update all the changes (UPdt displayed) and return to step 1.
9. Press ‘
’ up or ‘’ down to select another submenu.
10. Press ‘X’ to return to set-up, calibration and test menu selection.
11. Press ‘X’ again to return to normal operation.
Table 14. Fault relay configuration options.
Fault Relay Configuration Relay 1 Relay 2 Relay 3
Instrument Fault Only (1FLt) Alarm 1 Alarm 2 Instrument Fault
Separate Fault Relays (2FLt) Any Alarm Maintenance Fault Instrument Fault
Combined Fault Relay (CmbF) Alarm 1 Alarm 2 Any Fault
8.2.4 Set Calibration Interval ‘ CAL’
1. Select the set calibration interval ‘
CAL’ submenu and press ‘’ to accept.
2. The flashing display shows the number of days interval after a calibration that a calibration due maintenance fault will be displayed.
3. Use the ‘
’ up or ‘’ down buttons to change the number of days. (If the calibration interval is set below
‘001’ then ‘OFF’ will be displayed and no calibration interval will be activated and no reminders displayed)
4. Press ‘
’ to update the change (UPdt displayed) and return to step 1.
5. Press ‘
’ up or ‘’ down to select another submenu.
6. Press ‘X’ to return to set-up, calibration and test menu selection.
7. Press ‘X’ again to return to normal operation
35
8.2.5 Set Date and Time ‘timE’
1. Select the set date and time ‘timE’ submenu and press ‘
’ to accept.
2. The flashing display will show the current date format mm:dd or dd:mm.
3. Use the ‘
’ up or ‘’ down buttons to change the date format.
4. Press ‘
’ to display the flashing current year setting.
5. Use the ‘
’ up or ‘’ down buttons to change the year setting (between 2003 and 2030).
6. Press ‘
’ to display the flashing current month setting.
7. Use the ‘
’ up or ‘’ down buttons to change the month setting (between 1 and 12)
8. Press ‘
’ to display the flashing current day setting.
9. Use the ‘
’ up or ‘’ down buttons to change the day setting (between 1 and 31)
10. Press ‘
’ to display the flashing hours of current time.
11. Use the ‘
’ up or ‘’ down buttons to change the hours setting (between 00 and 23).
12. Press ‘
’ to display the flashing minutes of current time.
13. Use the ‘
’ up or ‘’ down buttons to change the minutes setting (between 00 and 59).
14. Press ‘
’ to update the changes (UPdt displayed) and return to step 1.
15. Press ‘
’ up or ‘’ down to select another submenu.
16. Press ‘X’ to return to set-up, calibration and test menu selection.
17. Press ‘X’ again to return to normal operation.
8.2.6 Set Address ‘ nET’
1. Select the set address ‘
nEt’ submenu and press ‘’ to accept.
2. A flashing ‘n’ or ‘Y’ is displayed depending on if auto address detection is on (AU Y) or off (AU n).
3. If automatic address detection (AU Y) is selected then the detector will automatically appoint a valid IP
address and return to the submenu.
4. If manual address setting (AU n) is selected press ‘
’ to display the flashing first part of the IP address.
The dot on the upper left indicates that the first part of the IP address is being displayed.
5. Use the ‘
’ up or ‘’ down buttons to change the address setting (between 0 and 255)
6. Press ‘
’ to display the second part of the address. The two dots on the upper left indicate the 2nd portion of
the address.
7. Use the ‘’ up or ‘’ down buttons to change the address setting (between 0 and 255).
8. Repeat for the third and forth parts of the IP address.
9. Press ‘’ to display the flashing first part of the sub net mask value.
10. Use the ‘’ up or ‘’ down buttons to change the address setting (between 0 and 255).
11. Press ‘’ accept and repeat for the second, third and forth sub net values. The number of dots in the upper
left indicates which part of the sub net address is being viewed.
12. Press ‘
’ to update the changes (UPdt displayed) and return to step 1.
13. Press ‘’ up or ‘’ down to select another submenu.
14. Press ‘X’ to return to set-up, calibration and test menu selection.
15. Press ‘X’ again to return to normal operation.
NOTE: If the IP address is changed the new settings will not be implemented until the MIDAS
®
unit is
powered down and back on again.
36
8.2.7 Set pass code ‘ PWd’
1. Select the set pass code ‘
PWd’ submenu and press ‘’ to accept.
2. Press ‘
’ up or ‘’ down to set the first pass code value.
(Note holding down the button will increase the increment speed).
3. Press ‘’ to enter the first value and move to setting the second value.
4. Press ‘
’ up or ‘’ down to set the second value.
5. Press ‘
’ to enter and repeat for the third and forth values in the passcode.
6. To confirm the pass code re enter it again using the same procedure.
7. The new pass code will be saved after the last entry if the two entered passcodes are the same.
NOTE: Pass codes can be set between 0001 and 9999. Setting the pass code to 0000 will switch off the pass code. If a pass code is forgotten contact your local Zellweger Analytics service department. In the event that a pass code is forgotten by the user, Zellweger Analytics is not liable for any costs associated with the recovery process nor for any inconvenience incurred while the user is unable to access protected settings. Please ensure secure records are kept for all pass codes implemented on MIDAS
®
units.
8.3 Calibration Menu ‘ CAL’
The calibration menu allows the calibration settings of the detector to be changed. The calibration menu comprises of 4 submenus as shown in the table below.
Table 15. Calibration mode submenus.
To select the calibration menu from normal operation, press the ‘’ up button for a few seconds. Enter the pass code (if set). Use the ‘
’ up or ‘’ down buttons to select the calibration menu ‘ ’ icon and press the ‘’ accept button.
8.3.1 Zero Calibration ‘
0CAL’
1. Use the ‘’ up or ‘’ down buttons to select the zero calibration ‘ 0CAL’ submenu and press ‘’ to accept.
2. The zero calibration icon starts flashing in order to tell the user to prepare to apply zero gas.
3. The inhibit ‘
’ icon is also displayed indicating that no alarm outputs will be generated during this process.
4. Press ‘’ to confirm when ready and the icon goes steady.
5. Apply the zero gas (or ambient air).
6. The dots on the left of the display indicate progress to a successful zero.
7. The display will show the zero gas reading and if measured to be stable for an appropriate period of time
will display ‘PASS’.
8. If the zero calibration is unsuccessful then the display will show an error code (see table oppersite)
9. Press ‘
’ to exit.
10. Press ‘’ up or ‘’ down to select another submenu.
11. Press ‘X’ to return to set-up, calibration and test menu selection.
12. Press ‘X’ again to return to normal operation.
Calibration submenu
Icon Calibration setting
Zero Set detector zero
Span Set detector span
Flow Calibrate detector sample flow rate
mA Calibrate detector analog output
37
8.3.2 Span Calibration ‘ SPAn’
1. Use the ‘
’ up or ‘’ down buttons to select the span calibration ‘ SPAn’ submenu and press ‘’ to accept.
2. The span calibration icon starts flashing in order to tell the user to prepare to apply span gas.
3. The inhibit ‘
’ icon is also displayed indicating that no alarm outputs will be generated during this process.
4. Use the ‘
’ up or ‘’ down buttons to select the gas ID code of calibration gas (for multi gas ID sensor
cartridges only) and press ‘¸’ to accept. (Refer to section 8.2.2 for details of gas ID codes).
5. Use the ‘
’ up or ‘’ down buttons to select if humidified ‘HUm’ or dry ‘drY’ calibration gas is being used.
6. Use the ‘
’ up or ‘’ down buttons to change the value to the concentration of span calibration gas being
used.
7. Press ‘
’ to confirm when ready and the icon goes steady.
8. Apply the span gas.
9. The dots on the left of the display indicate progress to a successful span.
10. The display will show the span gas reading and if measured to be stable for an appropriate period of time
will display ‘PASS’.
11. If the span calibration is unsuccessful then the display will show an error code (see below).
12. Press ‘
’ to exit.
13. Press ‘
’ up or ‘’ down to select another submenu.
14. Press ‘X’ to return to set-up, calibration and test menu selection.
15. Press ‘X’ again to return to normal operation
NOTE: The change in the gas bottle icon’s contents gives an indication as to the relative stability of the gas reading. The arrows and bars inside the cylinder indicate whether it is rising or falling as appropriate (see table below).
Table 16. Calibration codes
Code Code meaning
PASS Successful Calibration
FL:0H Zero Calibration Timeout - Over Range
FL:0L Zero Calibration Timeout - Under Range
FL:0U Zero Calibration Timeout - Unstable
FL:SH Span Calibration Timeout - Over Range
FL:SL Span Calibration Timeout - Under Range
FL:SU Span Calibration Timeout - Unstable
Table 17. Calibration stability icons
Stable
Over-Range
Unstable-Rising
Unstable-Flat
Unstable-Falling
Under-Range
38
8.3.3 Flow Calibration ‘ FLoW’
1. Use the ‘
’ up or ‘’ down buttons to select the flow calibration ‘ FLoW’ submenu and press ‘’ to accept.
2. The flow icon with the dot at the bottom starts flashing in order to tell the user that the unit is ready to read
the zero flow offset.
3. The inhibit ‘
’ icon is also displayed indicating that no alarm outputs will be generated during this process.
4. Press ‘
’ to confirm and the icon goes steady and a count down from 10 is shown as the zero flow offset
is read.
5. The display will show the flashing flow icon with the dot 1/4 ways up to tell the user to set the set point
1flow (350 cc/min).
6.
Press ‘’ up or ‘’ down keys to adjust the reading on the external flow meter to the setpoint 1 target value.
(NOTE: The reading must be +/- 50 cc/min of target to be accepted).
7. Press ‘
’ to confirm and the target value flashes.
8. Press ‘
’ up or ‘’ down to enter the actual value read on the external flow meter.
(NOTE: Holding down the button will increase the increment speed).
9. Press ‘
’ to confirm and the icon goes steady and a count down from 10 is shown as the setpoint 1
value is read.
10. The display will show the flashing flow icon with the dot 3/4 ways up to tell the user to set the set point
2 flow (650 cc/min).
11.
Press ‘’ up or ‘’ down keys to adjust the reading on the external flow meter to the set point 2 target value.
(NOTE: The reading must be +/- 50 cc/min of target to be accepted).
12. Press ‘
’ to confirm and the target value flashes.
13. Press ‘
’ up or ‘’ down to enter the actual value read on the external flow meter.
(NOTE: Holding down the button will increase the increment speed).
14. Press ‘
’ to confirm and the icon goes steady and count down from 10 is shown as the setpoint 2 value
is read.
15. If successful the display shows UPdt (update) and the flow calibration is complete.
16. Press ‘
’ up or ‘’ down to select another submenu.
17. Press ‘X’ to return to set-up, calibration and test menu selection.
8.3.4 mA Calibration ‘mA 4-20’
1. Use the ‘
’ up or ‘’ down buttons to select the mA calibration ‘ mA4-20’ submenu and press
’ to accept.
2. The display shows 4.00 mA indicating that the analog output should be reading 4 mA
3. The inhibit ‘ ’ icon is also displayed indicating that no alarm outputs will be generated during this process.
4. Use the ‘
’ up or ‘’ down buttons to adjust the analog output to read 4.00 mA.
5. Press ‘
’ to accept.
6. The display will show 20.00 indicating that the analog output should be reading 20 mA
7. Use the ‘’ up or ‘’ down buttons to adjust the analog output to read 20.00 mA.
8. Press ‘’ to update the changes (UPdt displayed) and return to step 1.
9. Press ‘’ up or ‘’ down to select another submenu.
10. Press ‘X’ to return to set-up, calibration and test menu selection.
11. Press ‘X’ again to return to normal operation.
39
8.4 Test Menu ‘ tEST’
The test menu is used to test the detector gas reading using bump test gas, and for simulation of alarm and fault display and output operation (relay, analog and digital). The test submenu also includes the detector inhibit facility. The test menu comprises of 3 submenus as shown in the table below.
Test submenu Icon Test
Bump Bump test detector with inhibited alarm outputs
Alarm/Fault
Test the alarm and fault display and output operation
Inhibit Put the unit into/out of an inhibit state and set
inhibit time out
Table 18. Test mode submenus.
To select the test menu from normal operation, press the ‘’ up button for a few seconds. Enter the pass code (if set). Use the ‘
’ up or ‘’ down buttons to select the test menu ‘ ’ icon and press the ‘’ accept button.
8.4.1 Bump Test ‘ bUmP’
1. Use the ‘
’ up or ‘’ down buttons to select the bump test ‘ bUmP’ submenu and press ‘’ to accept.
2. The inhibit ‘
’ icon is also displayed indicating that no alarm outputs will be generated during this process.
3. Apply the bump test gas and the display will show the measured gas concentration.
4. Remove the bump test gas and allow the detector reading to return to zero.
5. Press ‘X’ to exit.
6. Press ‘
’ up or ‘’ down to select another submenu.
7. Press ‘X’ to return to set-up, calibration and test menu selection.
8. Press ‘X’ again to return to normal operation.
8.4.2 Alarm/Fault Test ‘
Si m’
1. Use the ‘
’ up or ‘’ down buttons to select the alarm/fault test ‘ Si m’ submenu and press ‘
to accept.
2. The display shows ‘Si m’ and the A1 ‘
’ icon.
3. Use the ‘
’ up or ‘’ down buttons to select A1 ‘’, A2 ‘ ’ or Fault ‘ ’ for test simulation.
4. Press ‘’ to select and the displays ‘SurE’ to indicate that the next step will activate the selected output
(relay, analog and digital)
5. Press ‘
’ and the display flashes ‘on’ indicating that the selected output is activated
6. Press ‘X’ to return to step 2 and select a different output for test.
7. Press ‘X’ to exit.
8. Press ‘
’ up or ‘’ down to select another submenu.
9. Press ‘X’ to return to set-up, calibration and test menu selection.
10. Press ‘X’ again to return to normal operation.
40
8.4.3 Inhibit State ‘ I nH’
1. Use the ‘
’ up or ‘’ down buttons to select the inhibit ‘ I nH’ submenu and press ‘’ to accept.
2. The display flashes ‘nonE’ indicating there is no inhibit currently set.
3. Use the ‘
’ up or ‘’ down buttons to select alarm, alarm and fault or all (Alm, AL-Ft or ALL) output inhibit
states. See table below for details of inhibit states.
4. Press ‘
’ to accept the selected inhibit state.
5. Press ‘
’ up or ‘’ down to set the inhibit timeout (between 0 minutes and 4 hrs- default 30 minutes).
6. Press ‘
’ to accept (UPdt displayed).
7. The selected outputs will be inhibited until the inhibit timeout has elapsed.
NOTE: If the inhibit timeout elapses before the inhibit state is set back to ‘nonE’ the maintenance fault
code M12 will be displayed.
8. To take the unit out of inhibit, select the inhibit ‘ ’ I nH submenu and press ‘’ to accept.
9. Use the ‘
’ up or ‘’ down buttons to select ‘nonE’
10. Press ‘
’ to return to the submenu selection.
11. Press ‘
’ up or ‘’ down to select another submenu.
12. Press ‘X’ to return to set-up, calibration and test menu selection.
13. Press ‘X’ again to return to normal operation.
Table 19. Inhibit states.
Inhibit state Display Function
None nonE No functions are inhibited.
Alarms Inhibited
ALm Alarm events will be detected, but alarm outputs (relays, 4-20 mA current loop
and Ethernet) will be disabled.
Alarms and Faults Inhibited
AL-Ft Alarm and fault events will be detected, but alarm and fault outputs (relays, 4-20
mA current loop and Ethernet) will be disabled.
Full Inhibit
ALL All monitoring functions inhibited. No monitoring is performed and no alarms or
faults (except for Inhibit Timeout) will be reported.
41
9 ROUTINE MAINTENANCE
MIDAS
®
is a fully serviceable product designed with modular components that can be readily replaced by trained
service personnel so as to minimize the time that the gas detector is not available.
External in-line air filters should be replaced every three months or more frequently if the system is sampling in environments that have high levels of particulate matter or very acidic / wet atmospheres. Similarly the internal particulate filter should be replaced once a year or more frequently if the sample lines are prone to heavy contamination.
Every sensor cartridge is shipped with a 12 month warranty and an extended 2 year warranty is also available for purchase. All sensor cartridges are factory calibrated to traceable national standards before shipment to the end user.
Note that testing or calibrating with the wrong (incorrect, out of date, non-traceable) calibration gases, calibration equipment, methods or operating conditions can actually damage the sensor cartridge’s lifetime and alter the calibration adversely. Only qualified calibration technicians should attempt to calibrate the MIDAS
®
gas
detector.
NOTE: For details regarding sensor cartridge calibration and bump testing method refer to section 18.
The internal pump module is designed to operate for a minimum of 18 - 24 months and it is recommended that this pump module (part number MIDAS-A-007) be replaced every 2 years.
9.1 Sensor Cartridge Replacement
To avoid the possibility of unwanted alarms or faults Zellweger Analytics recommend that the sensor cartridge be replaced without power to the MIDAS
®
unit. If fitting a sensor cartridge to a unit that is powered, please refer to
section 8.4.3 to inhibit the detectors outputs.
9.1.1 Sensor Cartridge Fitting/Replacement
1. Unscrew the thumbscrew located on the front panel and remove the cover by pulling it forwards off the main chassis (see diagram 22).
2. Ensure the power switch on the terminal module is in the off position.
3. Remove the old sensor cartridge from the unit (if fitted) by unclipping the two sensor cartridge locking tabs located either side of the sensor cartridge and using them to firmly pull the sensor cartridge out (see diagram 23).
4. Fit the new sensor cartridge by aligning the pins at the top of the sensor cartridge with the socket in the sensor cartridge chamber.
5. Carefully push the sensor cartridge into the sensor cartridge chamber until fully home and lock in place using the tabs either side of the sensor cartridge (see diagram 24).
6. Switch the power switch on the terminal module to the ‘on’ position.
7. Refit the detector’s cover by aligning the slots either side with the locating tabs on the mounting bracket assembly and pushing the cover horizontally until home. Tighten the thumbscrew located on the front panel (see diagram 25).
NOTE: If replacing a sensor cartridge with a different gas type sensor cartridge the display will scroll the message ‘CHANGE GAS?’ If you are changing the sensor cartridge gas type press ‘
’ to accept. If not, fit the correct sensor cartridge. The pass code (if set) must be entered to change sensor cartridge gas type. To set the correct gas for a multi gas type sensor cartridge refer to section 8.2.2.
42
Diagram 22. Removing detector cover Diagram 23. Removing sensor cartridge
Diagram 24. Fitting/replacing sensor cartridge Diagram 25. Refitting detector cover
REMOVING COVER
mounting
bracket
chassis
cover
loosen
m
idas
G a
s Se
n
s o
r C
a
rt r
i d
g e
N IT
R OG
E
N
TR
I F LU
OR
I D
E
N
F 3
P
N
: M
I D
A
S
-S
­H
F
X
SN
: 02
3
4 4
56
6
7
0
U
s e
b y
:
07
­09
­2
0
06
ze
ll w
e
g
er
a
na
l
yt
i c s
chassis
sensor
m
idas
G
a s
Se
n s
o r
C a
rt
r i d
g
e
N IT
R OG
E N
TR
I F LU
OR
I D
E
N
F
3
P
N
:
M
I D
A
S-S-
H
F
X
SN
: 02
3
4
4 56
6
7
0
Us
e
b
y
: 07
­09
­2
006
ze
ll
w
e g
er
a
na
l
y
t i c
s
HOW CARTRIDGE IS FITTED
chassis
cartridge
m
i d
a
s
G
a s
S
en
so r
C
a
rt
r i d
g e
NITROG
E N
T
R I F
LU
O
R
I DE
N
F3
P
N
:
M
I
D A
S-
S
-H FX
SN
:
02
3 44
5
6
6
70
U se
by
:
0
7
­09
-20
0
6
z
el
l w
eg
e
r
a
n
a
l y t
i cs
slots
tighten
HOW COVER IS FITTED
tabs
cartridge fitted
43
9.2 PUMP REPLACEMENT
The pump module has been designed to allow easy replacement. New pump modules are supplied with new springs, bracket and tubing pre-assembled for quick release / replacement
Diagram 26. Location of pump module
The following procedure should be followed carefully and only performed by suitably trained personnel.
1. Isolate the power to the detector.
2. Unscrew the thumbscrew located on the front panel.
3. Remove the cover by pulling it forward off the main chassis.
4. Unscrew the two retaining screws located at the bottom front of the chassis.
5. Pull the main chassis forward to disconnect it from the mounting bracket assembly.
Pump module
44
6. Remove the 4 pump module screws.
7. Remove the two fixing clips and disconnect the tubes at the manifold.
45
8. Slide the pump module out and disconnect the connector from the pcb.
9. Fit the new pump module following the steps above in reverse order.
46
9.3 Re Assembling the Detector
1. Align the PCB at the top rear of the main chassis with the connector located at the top of the mounting bracket assembly.
2. At the same time align the two tubes at the bottom rear of the main chassis with the two tubes located on the bottom of the mounting bracket assembly.
3. Slide the chassis backwards on the mounting bracket assembly so that the PCB, connector and tubes engage simultaneously.
4. Ensure the PCB, connector and tubes are fully engaged by firmly pushing the main chassis horizontally backwards on the mounting bracket assembly (WARNING: DO NOT PUSH ON THE LCD AS THIS MAY CAUSE DAMAGE).
5. Align the two fixing screws located at the bottom of the chassis with the screw threads on the mounting bracket assembly.
6. Tighten the screws to secure the chassis to the mounting bracket assembly.
7. Switch the power switch on the terminal module to the ‘on’ position.
8. Refit the detector’s cover by aligning the slots either side with the locating tabs on the mounting bracket assembly.
9. Push the cover horizontally until home.
10. Tighten the thumbscrew located on the front panel.
9.4 Filter Replacement
The internal filter has been designed for easy replacement.
The following procedure should be followed carefully and only performed by suitably trained personnel.
1. Isolate the power to the detector.
2. Unscrew the thumbscrew located on the front panel.
3. Remove the cover by pulling it forwards off the main chassis.
4. Unscrew the two retaining screws located at the bottom front of the chassis.
5. Pull the main chassis forward to disconnect it from the mounting bracket assembly.
6. Locate the filter access slot in the side of the main chassis.
Main chassis
Filter
Filter access slot
Diagram 30. Filter location
47
7. Carefully disconnect both sides of the filter from the pump manifold.
8. Remove the old filter and replace with a new filter ensuring that the filter connectors are fully engaged in the manifold ports and that the filter is the correct way round (arrow pointing downwards).
Diagram 31. Filter orientation
Filter connector
Manifold ports
New filter correct orientation
10 PYROLYZER MODULE OPTION
The pyrolyzer module option is installed underneath the standard MIDAS through the pyrolyzer to the sensor cartridge. The pyrolyzer converts Nitrogen Trifluoride (NF sample into Hydrogen Fluoride (HF) by means of pyrolysis. The HF can then measured by the sensor cartridge and the concentration displayed as the equivalent reading in ppm NF availability of other detectable gases using the pyrolyzer module.
®
gas detector. The air sample is drawn
. Contact Zellweger Analytics for the
3
) present in the air
3
Diagram 32. Pyrolyzer module attached to MIDAS
®
gas detector
MIDAS
®
gas detector
Pyrolyzer module (MIDAS-T-00P)
WARNING: To maintain stated sensor performance when using the pyrolyzer, ensure the constant ambient temperature and humidity levels of the installation point do not exceed the maximum operating temperature and humidity range. This product can tolerate temporary changes in temperature and humidity, but constant exposure to higher levels of temperature and humidity will require more frequent bump testing or calibration to confirm working specification.
48
10.1 Fitting the Pyrolyzer Module
1. Isolate the power to the detector.
2. Unscrew the captive thumbscrew located on the front panel.
3. Remove the cover by pulling it forwards off the main chassis.
4. Unscrew the two captive screws located at the bottom front of the chassis.
5. Pull the main chassis forwards to disconnect it from the mounting bracket assembly.
6. Thread the connector and ribbon cable from the pyrolyzer through the rectangular access in the bottom of the mounting bracket.
7. Plug the connector into the socket (con5) at the bottom left of the terminal board.
Diagram 33. Connecting the Pyrolyzer module
8. Offer the pyrolyzer module up underneath the mounting bracket ensuring that the fitting at the top rear of the pyrolyzer mates with the sample inlet port at the bottom of the mounting bracket.
9. Align the three screw mounting bosses on the top of the pyrolyzer with the three screw holes in the mounting bracket.
10. Insert and tighten the three screws provided.
Rectangular access at bottom of bracket
Mounting screws (x3)
Socket
Connector
Pyrolyzer module
Power switch
Thermal module
49
Gas sample inlet as NF
3
To
MIDAS sample
inlet as HF
Pyrolyzer Option
super-heater
Gas sample
chamber
Needle valve (Heater flow adjustment
)
Diagram 34. Pyrolyzer main components
10.2 Re-Assembling the MIDAS® Detector
1. Align the PCB at the top rear of the main chassis with the connector located at the top of the mounting bracket assembly.
2. At the same time align the two tubes at the bottom rear of the main chassis with the two tubes located on the bottom of the mounting bracket assembly.
3. Slide the chassis backwards on the mounting bracket assembly so that the PCB and connector and tubes engage simultaneously.
4. Ensure the PCB, connector and tubes are fully engaged by firmly pushing the main chassis horizontally backwards on the mounting bracket assembly (WARNING: DO NOT PUSH ON THE LCD AS THIS MAY CAUSE DAMAGE).
5. Align the two fixing screws located at the bottom of the chassis with the screw threads on the mounting bracket assembly.
6. Tighten the screws to secure the chassis to the mounting bracket assembly.
7. Fit sensor cartridge MIDAS-S-HFX into the sensor cartridge chamber.
8. Switch the power switch on the terminal module to the ‘on’ position.
9.
Refit the detector’s cover by aligning the slots either side with the locating tabs on the mounting bracket assembly.
10. Push the cover horizontally until home.
11. Tighten the thumbscrew located on the front panel.
12. After applying power ensure that the sensor cartridge gas id code 08-03 is set. Refer to section 8.2.2 for details of how to set the gas id code.
The MIDAS
®
gas detector automatically detects the connection of the pyrolyzer module and provides all the power
and signaling for the device.
Diagram 35. Pyrolyzer detail
Gas sample inlet as NF3
Exhaust Tubing
To
MIDAS sample
inlet as HF
Jumper Teflon tubing
between pyrolyzer and
MIDAS (included with
pyrolyzer option)
Mounting holes
for M3 screws (3)
Terminal for Power and communication
connection
Pyrolyzer Option
Freon filter
50
Diagram 36 shows in a simple form how the gas sample is drawn through the pyrolyzer module by the pump (located at the end of the gas path) and is first routed to the pyrolyzer via the freon filter before being resent to the gas sensor cartridge, where the gas measurement is taken. The sample continues via the flow-meter through the dust filter and is finally exhausted from the instrument. Adjustment of the gas flow through the instrument is done automatically. To perform a flow calibration refer to section 8.3.3.
Diagram 36. Pyrolyzer configuration gas flow









 





Diagram 37 shows in diagrammatic form how the gas sample flows through the various components. It shows how the instrument controls, monitors and measures the gas sample flow and signals any fault conditions to the user via the LCD screen, e.g. due to flow problems caused by pump failure, or failure of the pyrolyzer unit, etc.
Diagram 37. Pyrolyzer option flow diagram
Pyrolyzer Gas Sensor
Sample Inlet 500 ml/min
Pyrolyzer Flow rate
50 ml/min
Voltage/
Current
Sensin
g
Driver
Heater
Heater Controller
Pyrolyzer Option
heater voltage/current sensing system to detec
t
Heater Controller utilizes
element failure and out of
conditions such as heater
temperature range.
Bypass flow
Flow Controller
Pump Drive
Pressure Sensin
g
Pyro Fault Flow Fault
Sensor
Pressure
Pump
Exhaus
t
Sample
Flow Controller utilizes differential pressure over laminar flow element to monitor flow conditions
.
Elemen
t
Laminar Flow
Gas Sample Flow
Signal/Control Flow
Dust Filter
Particulat
e
Freon/
Fi
lter
MIDA
S Transmitter
Main
Control PC
B
450 ml/min
51
11 ANALOG INPUT MODULE
MIDAS
®
can be fitted with an analog module (MIDAS-T-00A) that allows it to be used as a control ‘hub’ for other remotely connected 4-22 mA analog detection devices such as other gas detectors, flow sensor cartridges and other industrial monitoring apparatus.
NOTE: Not all analog devices will be compatible with the MIDAS
®
Analog Input module. Zellweger Analytics
is not responsible for ensuring the compatibility of the MIDAS
®
transmitter and analog input module with other manufacturers’ analog output devices. Please contact your local Zellweger Analytics representative prior to ordering this option.
Diagram 38. Analog module configuration
Any MIDAS® detector can be fitted with the optional analog input module that accepts the analog signal from a remote device. MIDAS
®
can then be configured to display the remote system’s analog reading on the local LCD,
trigger relays and provide a digital output via Modbus/TCP Ethernet.
For full installation and wiring instructions please refer to the separate MIDAS
®
Analog Input module Quick Start
Guide (MIDAS-A-023) available on request from Zellweger Analytics.
11.1 Fitting the Analog Module
1. Isolate the power to the detector.
2. Unscrew the thumbscrew located on the front panel.
3. Remove the cover by pulling it forwards off the main chassis.
4. Unscrew the two retaining screws located at the bottom front of the chassis.
5. Pull the main chassis forwards to disconnect it from the mounting bracket assembly.
6. Thread the connector and ribbon cable from the analog module through the rectangular access in the bottom of the mounting bracket.
7. Plug the connector into the socket (con5) at the bottom left of the terminal board.
8. Offer the analog module up underneath the mounting bracket ensuring that the three screw mounting bosses on the top of the module align with the three screw holes in the mounting bracket.
9. Insert and tighten the three screws provided.
52
Diagram 39. Analog module connection.
Rectangular access at bottom of bracket
Mounting screws (x3)
Socket
Connector
Analog module
11.2 RE-ASSEMBLING THE MIDAS
®
DETECTOR
1. Align the PCB at the top rear of the main chassis with the connector located at the top of the mounting bracket assembly.
2. At the same time align the two tubes at the bottom rear of the main chassis with the two tubes located on the bottom of the mounting bracket assembly.
3. Slide the chassis backwards on the mounting bracket assembly so that the PCB, connector and tubes engage simultaneously.
4. Ensure the PCB connector and tubes are fully engaged by firmly pushing the main chassis horizontally backwards on the mounting bracket assembly (WARNING: DO NOT PUSH ON THE LCD AS THIS MAY CAUSE DAMAGE).
5. Align the two fixing screws located at the bottom of the chassis with the screw threads on the mounting bracket assembly.
6. Tighten the screws to secure the chassis to the mounting bracket assembly.
7. Fit the dummy sensor (optional accessory MIDAS-A-013) as per Section 9.1.1
8. Switch the power switch on the terminal module to the ‘on’ position
9. Refit the Detector’s cover by aligning the slots either side with the locating tabs on the mounting bracket assembly.
10. Push the cover horizontally until home.
11. Tighten the thumbscrew located on the front panel.
53
12 TROUBLE SHOOTING AND FAULT DIAGNOSIS
General trouble shooting guide and specific fault code table.
Table 20. Fault code descriptions
Fault code Description Condition Recovery
m10
Over range. A large concentration has
been detected. The MIDAS
®
requires an independent confirmation that the gas hazard is gone.
Supply known clean air to the MIDAS
®
and clear this fault.
m11
Calibration expires soon.
The user specified calibration interval has elapsed.
Perform zero and span calibrations. Increase span calibration period.
m12
Cartridge expires soon.
Cartridge is old and will expire soon.
Replace the cartridge with a new cartridge.
m13
Flow low.
MIDAS
®
is no longer able to regulate flow.
Check filters and pump.
m14
Interferent present.
An interferent is degrading the ability of
the MIDAS
®
to detect gas.
Check application.
m15
Temperature near limit.
Temperature within 2 Celsius of limit.
Check installation environment.
m16
BaseLine fault. Sensor baseline has
drifted.
Check for background gas concentration, temperature or humidity fluctuations. Perform zero calibration. Replace cartridge.
m17
Inhibit timeout. Transmitter has been in
inhibit mode too long.
Resume monitoring or increase timeout value.
F40
Sensor overdosed. Sensor has been exposed
to high gas concentrations for long periods.
Replace cartridge.
F41
BaseLine fault. Sensor baseline has
drifted.
Check for background gas concentration, temperature or humidity fluctuations. Perform zero calibration. Replace cartridge.
F42
Calibration expired.
Too long since last calibration.
Replace or calibrate the cartridge.
F43
Cartridge expired. Cartridge is too old. Replace cartridge.
F44
Cell failure. Cartridge has failed
Reflex™ check.
Replace cartridge.
F45
Stabilization timeout.
Cartridge has failed to stabilize.
If temperature or humidity shocks exist, precondition the cartridge. Check for background gas concentration. Replace cartridge.
F46
Cartridge analog failure.
Various reasons. Replace cartridge.
F47
Cartridge memory invalid.
Checksum error. Replace cartridge.
F48
Cartridge absent. No communications. Reseat cartridge. Replace cartridge.
F49
Cartridge wrong type.
Cartridge type found to be incorrect after boot-up.
Replace cartridge.
F80
Temperature limits Exceeded.
Temperature is outside limits
Check installation environment.
F81
Flow fail. Flow < 70% of nominal
for 15 seconds.
Check filters. Check for kinked tubing, Replace pump.
F82
Excessive electrical noise.
Internal electronics repeatedly noisy.
Check grounding of MIDAS
®
chassis. Check termination of cable shields.
Relo
cate the MIDAS® further from noise sources. Add ferrite inductors to cables.
F83
Pyrolyzer fail. Pyrolyzer fails to heat. Check electrical connection to pyrolyzer. Replace heater. Replace pyrolyzer.
F84
Misc. transmitter fault.
Transmitter is defective.
Service or replace MIDAS
®
.
54
13 REFLEX®
MIDAS
®
uses patented Zellweger Analytics technology to continuously monitor the health check status of specific electrochemical cells and alert the user if a cell enters a variety of fault conditions (such as open or short circuit etc.) which would leave the cell unable to detect gas and raise an appropriate alarm signal.
REFLEX® overcomes this unseen failure mode by applying periodically a special electronic pulse to the cell and reviewing the ‘echo’ from the cell as it responds to the applied signal. If the cell is deteriorating within certain pre-set limits based on the received signals then MIDAS
®
will decrease the REFLEX® sampling interval in order
to establish the actual viability of the cell. Within a relatively short time, MIDAS
®
will be able to alert the user via fault codes that the electrochemical cell is likely to be requiring replacement and is possibly unable to correctly detect gas.
REFLEX® is not required for pellistors or oxygen electrochemical cells as these sensor cartridges provide alternative electronic means to indicate open circuits and other sensor cartridge damage issues.
14 INTERNAL WEB SERVER
The MIDAS
®
gas detector unit utilizes an Ethernet port with the TCP/IP protocol as standard. The MIDAS® can function as a HTML web page server and these web pages can be viewed on external computer equipment (PC, PocketPC etc.). These web pages replicate the user interface on the front panel of the MIDAS
®
in a more flexible and friendly format for diagnostic and data entry purposes. The web pages also contain additional features not available via the keypad.
This procedure explains how to view web pages for a single MIDAS
®
which is connected to a single PC only.
In this example the MIDAS
®
is separately powered by 24 VDC. Of course it is possible to connect hundreds of
MIDAS
®
units to an Ethernet network which is part of the end user’s process LAN and more information may be
required from the local IT department and Zellweger Analytics as appropriate.
14.1 Physical Network Components
The Ethernet 100BaseT physical network is intended to connect computers to other computers through hubs. For this reason, a MIDAS
®
cannot communicate directly with a PC using a standard cable. This problem can be overcome by using a special “cross-over” Ethernet cable or by using a ‘hub’ or ‘switch’ and two standard (straight through wire connections) Ethernet cables. An example of a cross-over cable is the Belkin A3X126 family. An example of an Ethernet switch is the Linksys SD205. These or similar parts are commonly available from local IT equipment suppliers.
14.2 Internet Settings
Communications requires knowledge of the IP address and netmask of both the MIDAS® and the PC. Unless the MIDAS® is connected to a large network with a DHCP server, the addresses must be set manually. To view or set the IP address of the MIDAS
®
, see section 8.2.6. The procedure for setting the IP address of a PC is
detailed in the following example using a PC loaded with Microsoft™ Windows XP and Internet Explorer version
5.0 or higher as illustrated by Diagram 40. In the usual case the netmask for both computers should be set to
255.255.255.0. The most-significant three bytes of the IP address must be identical and the least-significant byte must be unique. For example, if the IP address of the MIDAS
®
was set to 169.254.60.1 then an appropriate IP
address for the PC would be 169.254.60.42.
14.3 Running the Web Browser
Start Microsoft™ Internet Explorer or similar web browser. Set the URL to “http://xxx.xxx.xxx.xxx” where the “xxx” fields are replaced with the IP address of the MIDAS
®
. A screen similar to Diagram 41. should be displayed. The
status and configuration of the MIDAS
®
can be viewed and set by clicking on the appropriate links.
55
Diagram 40 - IP Address setting in Windows XP
Diagram 41 - Sample MIDAS
®
web page
56
15 TYPICAL INSTALLATION TOPOLOGIES
MIDAS
®
has very flexible installation options to allow the user to select the one most suitable for their specific
application.
15.1 Conventional Installation
57
15.2 Modbus/TCP Installation
15.3 Power over Ethernet (POE) Installation
58
16 ORDERING INFORMATION
This section contains details of how to order complete MIDAS
®
detector and sensor cartridge kits, separate
transmitters and sensor cartridges as well as spares and accessories.
16.1 MIDAS
®
Transmitter
Description Part no.
MIDAS
®
transmitter. Complete with manual and quick start guide. (1) PG16 gland fitting for power in and relays out, 1/4” O.D. x 1/8” I.D. sample inlet tubing (10’) [3m], 3/16” I.D. exhaust tubing (10’) [3m], and a sample line tubing duct adapter. Order a sensor cartridge separately for each transmitter.
MIDAS-T-001
16.2 MIDAS® Pyrolyzer
Description Part no.
MIDAS
®
Pyrolyzer Module. Pyrolyzer MIDAS® transmitters provide detection of NF3 specifically. Unit includes a snap on pyrolyzer module that operates with and universal MIDAS
®
Extractive Transmitter (order separately) and (1) Freon in line filter. Units are powered via direct connection to transmitter unit. Consult ZA for additional gases that may become available via pyrolysis. Pyrolyzer includes Pyrolyzer Quick Start Guide and installation screws. Order the NF
3
cartridge (MIDAS-S-HFX)
separately.
MIDAS-T-00P
16.3 MIDAS® Analog INPUT Module
Description Part no.
MIDAS
®
Analog Input Module. For use with the MIDAS® gas detector for connection of external 4-20 mA inputs from other 4-20 mA analog transmitter devices to be displayed on the local MIDAS
®
display; activation of the 3 on board relays and transmission of the signal output via 0-22 mA and TCP Modbus Ethernet output. In this configuration the transmitter cannot be configured to measure gas directly with a separate MIDAS
®
cartridge, which should not be installed.
MIDAS-T-00A
59
16.4 MIDAS® Plug in Sensor Cartridges (Standard Warranty)
The plug in sensor cartridges for the MIDAS
®
Gas Transmitter are sold separately with a 1 year (12 month) standard warranty. For extended warranty sensor cartridges see the section 16.5. Some sensor cartridges can be configured to detect more than 1 target gas. Details of the gases, ranges and part numbers for the standard warranty sensor cartridges available are listed below.
Description Range Part no.
Ammonia 0-100 ppm MIDAS-S-NH3
Arsine 0-0.2 ppm MIDAS-S-ASH
Boron Trichloride 0-8 ppm MIDAS-S-HCL
Boron Trifluoride 0-8 ppm MIDAS-S-HFX
Bromine 0-0.4 ppm MIDAS-S-BR2
Chlorine 0-2 ppm MIDAS-S-HAL
Chlorine Dioxide 0-0.4 ppm MIDAS-S-BR2
Chlorine Trifluoride 0-0.8 ppm MIDAS-S-SF4
Carbon Dioxide 0-2.0% vol MIDAS-S-CO2
Carbon Monoxide 0-100 ppm MIDAS-S-COX
Diborane 0-0.4 ppm MIDAS-S-HYD
Dichlorosilane 0-8 ppm MIDAS-S-HCL
Disilane 0-20 ppm MIDAS-S-SHX
Fluorine 0-4 ppm MIDAS-S-HAL
Germane 0-0.8 ppm MIDAS-S-HYD
Hydrogen* 0-1000 ppm MIDAS-S-H2X
Hydrogen 0-100% LEL
Hydrogen Bromide 0-8 ppm MIDAS-S-HCL
Hydrogen Chloride 0-8 ppm MIDAS-S-HCL
Hydrogen Cyanide 0-20 ppm MIDAS-S-HCN
Hydrogen Fluoride 0-12 ppm MIDAS-S -HFX
Hydrogen Selenide 0-0.4 ppm MIDAS-S-HSE
Hydrogen Sulfide 0-40 ppm MIDAS-S-H2S
Methane 0-100% LEL
Nitric Oxide 0-100 ppm MIDAS-S-NOX
Nitrogen Dioxide 0-12 ppm MIDAS-S-NO2
Nitrogen Trifluoride 0-40 ppm MIDAS-S-HFX
Oxygen 0-25% v/v MIDAS-S-O2X
Ozone 0-0.4 ppm MIDAS-S-O3X
Phosphine 0-1.2 ppm MIDAS-S-PH3
Phosphorous Oxychloride 0-0.8 ppm MIDAS-S-POC
Silane 0-20 ppm MIDAS-S-SHX
Silane low level 0-2 ppm MIDAS-S-SHL
Sulfur Dioxide 0-8 ppm MIDAS-S-SO2
Sulfur Tetrafluoride 0-0.8 ppm MIDAS-S-SF4
TEOS Tetraethyl Orthosilicate 0-40 ppm MIDAS-S-TEO
Tungsten Hexafluoride 0-12 ppm MIDAS-S-HFX
1
1
MIDAS-S-LEL
MIDAS-S-LEL
1
MIDAS® detectors are not ETL approved for monitoring in or sampling from classified areas above 25% LEL
60
16.5 MIDAS® Plug in Sensor Cartridges (Extended Warranty)
The plug in sensor cartridges for the MIDAS
®
Gas Transmitter are available with an extended 2 year (24 month) warranty. Some sensor cartridges can be configured to detect more than 1 target gas. Details of the gases, ranges and part numbers for the extended warranty sensor cartridges available are listed below.
Description Range Part no.
Ammonia 0-100 ppm MIDAS-E-NH3
Arsine 0-0.2 ppm MIDAS-E-ASH
Boron Trichloride 0-8 ppm MIDAS-E-HCL
Boron Trifluoride 0-8 ppm MIDAS-E-HFX
Bromine 0-0.4 ppm MIDAS-E-BR2
Chlorine 0-2 ppm MIDAS-E-HAL
Chlorine Dioxide 0-0.4 ppm MIDAS-E-BR2
Chlorine Trifluoride 0-0.8 ppm MIDAS-E-SF4
Carbon Dioxide 0-2.0% vol MIDAS-E-CO2
Carbon Monoxide 0-100 ppm MIDAS-E-COX
Diborane 0-0.4 ppm MIDAS-E-HYD
Dichlorosilane 0-8 ppm MIDAS-E-HCL
Disilane 0-20 ppm MIDAS-E-SHX
Fluorine 0-4 ppm MIDAS-E-HAL
Germane 0-0.8 ppm MIDAS-E-HYD
Hydrogen 0-1000 ppm MIDAS-E-H2X
Hydrogen 0-100% LEL
Hydrogen Bromide 0-8 ppm MIDAS-E-HCL
Hydrogen Chloride 0-8 ppm MIDAS-E-HCL
Hydrogen Cyanide 0-20 ppm MIDAS-E-HCN
Hydrogen Fluoride 0-12 ppm MIDAS-E-HFX
Hydrogen Selenide 0-0.4 ppm MIDAS-E-HSE
Hydrogen Sulfide 0-40 ppm MIDAS-E-H2S
Methane 0-100% LEL
Nitric Oxide 0-100 ppm MIDAS-E-NOX
Nitrogen Dioxide 0-12 ppm MIDAS-E-NO2
Nitrogen Trifluoride 0-40 ppm MIDAS-E-HFX
Oxygen 0-25% v/v MIDAS-E-O2X
Ozone 0-0.4 ppm MIDAS-E-O3X
Phosphine 0-1.2 ppm MIDAS-E-PH3
Phosphorous Oxychloride 0-0.8 ppm MIDAS-E-POC
Silane 0-20 ppm MIDAS-E-SHX
Silane low level 0-2 ppm MIDAS-E-SHL
Sulfur Dioxide 0-8 ppm MIDAS-E-SO2
Sulfur Tetrafluoride 0-0.8 ppm MIDAS-E-SF4
TEOS Tetraethyl Orthosilicate 0-40 ppm MIDAS-E-TEO
Tungsten Hexafluoride 0-12 ppm MIDAS-E-HFX
1
1
MIDAS-E-LEL
MIDAS-E-LEL
1
MIDAS® detectors are not ETL approved for monitoring in or sampling from classified areas above 25% LEL
61
16.6 MIDAS® Complete Gas Detector Kits
A complete kit including a universal MIDAS
®
transmitter (MIDAS-T-001) and a selected MIDAS® sensor cartridge can be ordered as a combined package. Each sensor cartridge is supplied with an extended 2 year warranty. Sensor cartridge and gas detector are packaged separately for ease of installation. Note that to detect NF
3
a
separate pyrolyzer module (MIDAS-T-00P) must also be ordered.
Complete detector kit and sensor cartridge description Kit part no.
Ammonia 0-100 ppm kit MIDAS-K-NH3
Arsine 0-0.2 ppm kit MIDAS-K-ASH
Boron Trichloride 0-8 ppm, Dichlorosilane 0-8 ppm, Hydrogen Bromide 0-8 ppm, Hydrogen Chloride 0-8 ppm MIDAS-K-HCL
Boron Trifluoride 0-8 ppm, Hydrogen Fluoride 0-12 ppm, Nitrogen Trifluoride 4-40 ppm and Tungsten Hexafluoride 0-12 ppm kit
Bromine 0-0.4 ppm and Chlorine Dioxide 0-0.4 ppm kit MIDAS-K-BR2
Chlorine 0-2 ppm and Fluorine 0-4 ppm kit MIDAS-K-HAL
Carbon Dioxide 0-2.0% v/v kit MIDAS-K-CO2
Carbon Monoxide 0-100 ppm kit MIDAS-K-COX
Diborane 0-0.4 ppm and Germane 0-0.8 ppm kit MIDAS-K-HYD
Hydrogen 0-1000 ppm kit MIDAS-K-H2X
Hydrogen 0-100% LEL and Methane 0-100% LEL kit
1
Hydrogen Cyanide 0-20 ppm kit MIDAS-K-HCN
Hydrogen Selenide 0-0.4 ppm kit MIDAS-K-HSE
Hydrogen Sulfide 0-40 ppm kit MIDAS-K-H2S
Nitric Oxide 0-100 ppm kit MIDAS-K-NOX
Nitrogen Dioxide 0-12 ppm kit MIDAS-K-NO2
Oxygen 0-25% v/v kit MIDAS-K-O2X
Ozone 0-0.4 ppm kit MIDAS-K-O3X
Phosphine 0-1.2 ppm kit MIDAS-K-PH3
Phosphorous Oxychloride 0-0.8 ppm kit MIDAS-K-POC
Silane 0-20 ppm and Disilane 0-20 ppm kit MIDAS-K-SHX
Silane low level 0-2 ppm kit MIDAS-K-SHL
Sulfur Tetrafluoride 0-0.8 ppm and Chlorine Trifluoride 0-0.8 ppm kit MIDAS-K-SF4
Sulfur Dioxide 0-8 ppm kit MIDAS-K-SO2
TEOS 0-40 ppm kit MIDAS-K-TEO
1
MIDAS® detectors are not ETL approved for monitoring in or sampling from classified areas above 25% LEL
MIDAS-K-HFX
MIDAS-K-LEL
16.7 Accessories and Spares
Description Part no.
Sample Line Tubing Duct Adapter
Heater Element for NF
Replacement Pump Assembly for MIDAS
Replacement Internal Air Filter for MIDAS
Pyrolyzing Module MIDAS-A-006
3
®
Gas Detector MIDAS-A-007
®
Gas Detector MIDAS-A-009
Replacement RJ45 restraining strap MIDAS-A-010
PoE Ethernet Injector MIDAS-A-011
PoE 24 point Ethernet Hub MIDAS-A-012
Dummy Sensor (for use with analog module) MIDAS-A-013
IPA Scrubber Kit for CO applications
Replacement IPA Filter
Freon Filter for NF
Pyrolyzing Gas Detector 1830-0027
3
Flexible Conduit 21” - length
Flexible Conduit 27” - length
Flexible Conduit 36” - length
®
Operating Instructions and Quick Start Guide for MIDAS
Gas Detector - English MIDAS-A-001
End of line particulate filter
1283K1090
1283K2220
1830-0080
0235-0128
0235-0163
0310-2055
0780248
62
17 General Specifications
Physical
Size (unit with Sensor cartridge) 120 mm(H) x 63 mm(W) x 145 mm (D) (4.72 x 2.48 x 5.71 in)
Weight (unit with Sensor cartridge) 0.8 kg (1.76 lb)
Pyrolyzer Dimensions
Size 70 mm (H) x 63 mm (W) x 80 mm (D) (2.75 x 2.48 x 3.15 in)
Weight 0.41 kg (0.9 lb)
Power Requirements:
Operating Voltage 24 V Nominal +10 to -15%
Operating Voltage with Power over Ethernet 36-57 Vdc via PoE, 48 V Nominal
Power Consumption:
Transmitter unit < 5 W
With pyrolyzer < 12.95 W
Outputs:
Visual Alarm, power and fault LEDs plus LCD with all gas readings and events.
Relays Alarm1, Alarm2, Fault Relays (3) rated 1.0 A @ 30Vdc or 0.5 A @ 125 Vac, configurable as normally
open or closed, latched or unlatched.
Analog 2 wire Isolated 0-22 mA, Sink or Source
Digital Communications Modbus / TCP Ethernet / Power over Ethernet (PoE)
Service Port RS232C / PPP protocol
Certification & Specification:
European CE Marked for sale in European Community
Meets EN 50270:1999 (Type 2) and EN55011:2000
Environmental ETL approved UL61010B-1 and CSA-C22.2 No. 1010.1-92
Performance Designed to meet UL2075 (pending 2004)
Electrical IEEE 802.3af-2003
Response Times:
Typically Toxic T90 < 30 seconds
Catalytic T90 < 10 seconds
Transport System:
Flow Rate 500 cc / min
Transport Time 2 – 30 seconds maximum
Performance
LDL < LAL LAL = ½ TLV (typically 12% FSD) FSD = Typically 4 x TLV
Sampling Distance:
Tubing Length Up to 30 m (100 feet) with FEP tubing
Ambient Point In line air filter required
Exhaust Length Up to 30 m (100 feet)
Operating Temperature:
Unit with sensor cartridge 0°C to 40°C (32°F to 104°F)
Tubing Requirement:
Sample 6.35 mm OD (1/4”) x 3.18 mm (1/8”), FEP, 30 m (100 feet) maximum dependent on gas type
Exhaust 6.35 mm (1/4”) OD x 4.76 mm (3/16”), FEP, 30 m
Wiring Requirement:
4- 20mA 2 wire, 14 AWG maximum
Digital CAT5 cable or equivalent; RJ45 connector
63
Gas Concentration Display & Interface:
Instrument 4-digit alphanumeric display with separate units, flow rate bar graph and other icon driven indicators.
Remote Option for PC / PDA internet browser access via Ethernet or other Bus system
Warranty:
Transmitter Unit 1 year
Sensor cartridge 1 year standard, 2 years with extended warranty program
Expected lifetime of pump 2 years
Mounting:
Case Material:
4 button interface membrane keypad
Wall mounted using pre-drilled holes on chassis
Painted steel
18 Calibration and Bump Testing
®
All MIDAS approved calibration methods to a proven ISO 9000 quality controlled procedure. The MIDAS design is very robust and resistant to long term drift; in fact in a normal operating environment it is possible to extend the calibration interval with MIDAS calibrations). This design feature is another element that supports MIDAS
sensor cartridges are pre-calibrated by Zellweger Analytics using traceable gas standards and
®
up to 24 months (subject to local requirements concerning
®
as a long term cost effective gas
®
sensor cartridge
detection solution.
NOTE: It is the sole responsibility of each user to determine their own calibration and bump test schedule based on their own safety assessments and understanding of local requirements.
Calibration of an electrochemical sensor cartridge is a relatively complex discipline to master and it is possible that in less ideal field conditions the calibration can be far less accurate than under laboratory conditions and therefore substantial inaccuracies can be introduced. In the absence of any formal calibration policy by the end user and due to the low drift and long lifetime of the MIDAS
®
equipment, it is acceptable that the original factory
calibrations can be retained where appropriate thereby ensuring a high degree of accuracy of calibration
Care must be taken to ensure that the correct materials for an extractive application are obtained from reputable suppliers and are of proven quality and composition. Inappropriate calibration equipment will lead to under ­or over-presenting the wrong concentration of gas to the sensor cartridge during the calibration period. Likewise contaminated or inadequately flushed tubing and other gas-wet surfaces can also introduce errors into the calibration process. Only qualified, trained personnel should attempt to perform gas calibrations; contact your local Zellweger Analytics Service Representative for further details on calibration services
Bump tests are also used as a quick, low cost and approximate means to present a controlled gas release to the sensor cartridge in order to verify that the transmitter does respond accordingly. Bump testing is popular as typical experiences show that a smaller list of cheaper and easier to handle gases can be used to provide functional tests on a wider range of gas types. In the absence of a formal calibration policy from the end user, Zellweger Analytics would recommend a minimum annual schedule of at least one ‘bump test’ per sensor with the appropriate test gas (see details below).
The risks of bump testing are that it may reduce sensor cartridge lifetime and the accuracy of the calibration if applied incorrectly or too frequently. Testing sensor cartridges with non target gases can also lead to errors and drift related effects.
®
sensor cartridges have their calibration and bump test gases documented in data sheets that are readily
MIDAS available from Zellweger Analytics.
WARNING: To maintain stated sensor performance when using the pyrolyzer, ensure the constant ambient temperature and humidity levels of the installation point do not exceed the maximum operating temperature andhumidity range. This product can tolerate temporary changes in temperature and humidity, but constant exposure to higher levels of temperature and humidity will require more frequent bump testing or calibrationto confirm working specification.
64
The following table indicates the recommended calibration or bump test gas for the MIDAS® sensor cartridges:
Please consult with your Zellweger Analytics Service Representative for approved methods and materials before commencing calibrations and bump tests.
Gas Name Chemical
Ammonia NH
Arsine AsH
Boron Trichloride BCl
Boron Trifluoride
Bromine Br
Carbon Dioxide
Formula
3
3
3
BF
3
2
CO
2
Range Sensor
Cartridge Part Number
0-100 ppm MIDAS-S-NH3 Ammonia Ammonia
0-0.2 ppm MIDAS-S-ASH Arsine Phosphine
0-8.0 ppm MIDAS-S-HCL Hydrogen
0-8.0 ppm MIDAS-S-HFX Hydrogen
0-0.4 ppm MIDAS-S-BR2 Chlorine Chlorine
0-2.0% vol MIDAS-S-CO2 Carbon Dioxide Carbon Dioxide
Carbon Monoxide CO 0-100 ppm MIDAS-S-COX Carbon Monoxide Carbon Monoxide
Chlorine Cl
Chlorine Dioxide
Chlorine Trifluoride
Diborane B2H
Dichlorosilane H2SiCl
Disilane Si2H
Fluorine F
Germane GeH
Hydrogen (% LEL)
Hydrogen (ppm)
Hydrogen Bromide
Hydrogen Chloride
Hydrogen Cyanide
Hydrogen Fluoride
Hydrogen Selenide
Hydrogen Sulfide
Methane (% LEL) CH
Nitrogen Dioxide
Nitrogen Oxide
Nitrogen Trifluoride
Oxygen Proficiency & Deficiency
Ozone O
Phosphine PH
Phosphorous Oxychloride
Silane SiH
Silane (low level)
Sulfur Dioxide
Sulfur Tetrafluoride
Tetra Ethyl Ortho Silicate
Tungsten Hexafluoride
1
MIDAS® detectors are not ETL approved for monitoring in or sampling from classified areas above 25% LEL
2
ClO
2
ClF
3
6
2
6
2
4
H
2
H
2
HBr 0-8.0 ppm MIDAS-S-HCL Hydrogen
HCl 0-8.0 ppm MIDAS-S-HCL Hydrogen
HCN 0-20 ppm MIDAS-S-HCN Hydrogen
HF 0-12 ppm MIDAS-S-HFX Hydrogen
H2Se 0-0.4 ppm MIDAS-S-HSE Silane Hydrogen Sulfide
H2S 0-40 ppm MIDAS-S-H2S Hydrogen Sulfide Hydrogen Sulfide
4
NO
2
NO 0-100 ppm MIDAS-S-NOX Nitric Oxide Nitric Oxide
NF
3
O
2
3
3
POCl
3
4
SiH
4
SO
2
SF
4
TEOS 0-40 ppm MIDAS-S-TEO TEOS Carbon Monoxide
WF
6
0-2.0 ppm MIDAS-S-HAL Chlorine Chlorine
0-0.4 ppm MIDAS-S-BR2 Chlorine Chlorine
0-0.8 ppm MIDAS-S-SF4 Hydrogen
0-0.4 ppm MIDAS-S-HYD Diborane Phosphine
0-8.0 ppm MIDAS-S-HCL Hydrogen
0-20 ppm MIDAS-S-SHX Silane Hydrogen Sulfide
0-4.0 ppm MIDAS-S-HAL Chlorine Chlorine
0-0.8 ppm MIDAS-S-HYD Diborane Phosphine
0-100% LEL
1
MIDAS-S-LEL Hydrogen Hydrogen
0-1000 ppm MIDAS-S-H2X Hydrogen Hydrogen
0-100% LEL
1
MIDAS-S-LEL Hydrogen Hydrogen
0-12 ppm MIDAS-S-NO2 Nitrogen Dioxide Nitrogen Dioxide
0-40 ppm MIDAS-S-HFX Hydrogen
0-25% vol MIDAS-S-O2X Oxygen Air
0-0.4 ppm MIDAS-S-O3X Ozone Chlorine
0-1.2 ppm MIDAS-S-PH3 Phosphine Phosphine
0-0.8 ppm MIDAS-S-POC Hydrogen
0-20 ppm MIDAS-S-SHX Silane Hydrogen Sulfide
0-2.0 ppm MIDAS-S-SHL Silane Hydrogen Sulfide
0-8.0 ppm MIDAS-S-SO2 Sulfur Dioxide Sulfur Dioxide
0-0.8 ppm MIDAS-S-SF4 Hydrogen
0-12 ppm MIDAS-S-HFX Hydrogen
Calibration Gas Recommended Bump Test Gas
Chloride
Fluoride
Fluoride
Chloride
Chloride
Chloride
Cyanide
Fluoride
Fluoride
Chloride
Fluoride
Fluoride
Chlorine
Chlorine
Chlorine
Chlorine
Chlorine
Chlorine
Carbon Monoxide
Chlorine
Chlorine
Chlorine
Chlorine
Chlorine
65
Bump Testing MIDAS
®
A bump test is not intended to be an accurate calibration, but is used as a confidence check to ensure that the gas detection system is functional. Since some of the target gases are difficult to handle, bump testing often makes use of cross sensitivities so that more convenient gases can be used.
Equipment
The following items are required:
Test gas (see table)
Special regulator for extractive use (if using field calibration cylinder) or suitable sample bag (if not sampling directly from cylinder)
Tubing (suitable for gas being applied)
Preparation
Most of the test gases are highly toxic. It is essential that personnel using these toxic gases be trained in their use. Ensure that the test gas will be vented safely.
Applying test gas may cause alarm, fault or maintenance fault indications on the 4-20 mA loop, relays or digital outputs. Before starting a test, ensure that suitable steps have been taken to prevent these indications from triggering unwanted actions. The MIDAS
®
has a special mode for bump testing where the alarm outputs are
inhibited (see section 8.4.1).
To test extractive systems we recommend the use of a gas sample bag and allow the extractive unit to draw a sample of gas at its normal working flow rate. Suitable sample bags are available from SKC Inc. (www.skcinc.com). Special regulators for extractive applications to use with field calibration cylinders are also available.
If non-PTFE-lined tube has to be used, response times may be longer than expected, especially when using chlorine. In this case, it is especially important to keep the tubing short.
The pyrolyzer should be tested in the same way as an extractive system, but using NF is not available, chlorine can be used at a concentration of 5 ppm. If using chlorine, the freon filter must be
as the test gas. If NF3
3
removed from the unit before the test and replaced after the test.
NOTE: A test with chlorine operates the electrochemical cell and the unit’s electronics, but does not test the pyrolyzer module itself.
Target Gas Formula Cartridge
Ammonia NH
Arsine AsH
Boron Trichloride
Boron Trifluoride
Bromine Br
Carbon Dioxide
Carbon Monoxide
Chlorine Cl
Chlorine Dioxide
Chlorine Trifluoride
Diborane B2H
Dichlorosilane SH2Cl
Disilane Si2H
Fluorine F
Germane GeH
Hydrogen H2(ppm) MIDAS-S -H2X 1,000 Hydrogen 1000 500 120
Hydrogen H2(%LEL) MIDAS-S-LEL 100% LEL Methane 500 120
3
3
BCl
3
BF
3
2
CO
2
CO MIDAS-S-COX 100 Carbon
2
ClO
2
ClF
3
6
2
6
2
4
Part No.
MIDAS-S-NH3 100 Ammonia 50 500 120
MIDAS-S-ASH 0.2 Phosphine 0.5 500 120
MIDAS-S-HCl 8 Chlorine 5 500 120
MIDAS-S-HFX 8 Chlorine 5 500 120
MIDAS-S-BR2 0.4 Chlorine Low 500 120
MIDAS-S-CO2 2% v/v Carbon Dioxide 2% v/v 500 120
MIDAS-S-HAL 2 Chlorine 2 500 120
MIDAS-S-BR2 0.4 Chlorine Low 500 120
MIDAS-S-SF4 0.8 Chlorine Low 500 120
MIDAS-S-HYD 0.4 Phosphine 0.5 500 120
MIDAS-S-HCl 10 Chlorine 5 500 120
MIDAS-S-SHX 20 Hydrogen Sulfide < 25 500 120
MIDAS-S-HAL 4 Chlorine 5 500 120
MIDAS-S-HYD 0.8 Phosphine 0.5 500 120
Range (ppm) Recommended
Bump Test Gas
Monoxide
Conc. (ppm) Flow Rate
100 500 120
(cc/min)
Max. Time
66
Target Gas Formula Cartridge
Hydrogen Bromide
Hydrogen Chloride
Hydrogen Cyanide
Hydrogen Fluoride
Hydrogen Selenide
Hydrogen Sulfide
Methane CH
Nitric Oxide
Nitrogen Dioxide
Nitrogen Trifluoride
Oxygen O
Ozone O
Phosphine PH
Phosphorus Oxychloride
Silane SiH
Silane SiH
Sulfur Dioxide
Sulfur Tetrafluoride
HBr MIDAS-S-HCl 8 Chlorine 5 500 120
HCl MIDAS-S-HCl 8 Chlorine 5 500 120
HCN MIDAS-S-HCN 20 Sulfur Dioxide 15 500 120
HF MIDAS-S-HFX 12 Chlorine 5 500 120
H2Se MIDAS-S-HSE 0.4 Hydrogen Sulfide Low 500 120
H2S MIDAS-S -H2S 40 Hydrogen Sulfide 25 500 120
4
NO MIDAS-S-NOX 100 Nitric Oxide 50 500 120
NO
2
NF
3
2
3
3
POCl
3
4
4
SO
2
SF
4
Part No.
MIDAS-S-LEL 100% LEL Methane 500 120
MIDAS-S-NO2 12 Nitrogen Dioxide 10 500 120
MIDAS-S-HFX 40 Chlorine 5 500 120
MIDAS-S-O2X 25% v/v Air 500 120
MIDAS-S-O3X 0.4 Chlorine Low 500 120
MIDAS-S-PH3 1 Phosphine 0.5 500 120
MIDAS-S-POC 0.8 Chlorine Low 500 120
MIADS-S-SHX 2 Hydrogen Sulfide < 25 500 120
MIDAS-S-SHL 20 Hydrogen Sulfide < 25 500 120
MIDAS-S-SO2 8 Sulfur Dioxide 8 500 120
MIDAS-S-SF4 0.8 Chlorine Low 500 120
TEOS TEOS MIDAS-S-TEO 40 Carbon
Tungsten Tetrafluoride
WF
6
MIDAS-S-HFX 12 Chlorine 5 500 120
Range (ppm) Recommended
Bump Test Gas
Monoxide
Conc. (ppm) Flow Rate
(cc/min)
Max. Time
100 500 120
Notes
1. If Phosphine is not available, Hydrogen Sulfide can be used to test these sensors. Use a concentration of 2 ppm. Important: this should only be done once to avoid the danger of poisoning the sensor.
2. The ambient Oxygen level is typically 20.9% by volume. If there is a serious problem with an Oxygen sensor, the reading will not be 20.9%. As an extra check, the sensor can be breathed on or temporarily blocked. In either case the reading should decrease.
3. The flow rate is regulated by the Midas unit.
All test gases must be in a balance of air, not Nitrogen.
As an alternative, the actual target gas can always be used instead of a cross-sensitive gas.
Higher concentrations than the recommended ones can be used if necessary, but extra time may be needed for the reading to recover to zero.
IMPORTANT: Do not exceed the recommended concentrations or application times where Hydrogen Sulfide is used to test Silane. Doing this can damage the cell.
67
Finishing
Wait for the gas reading to return to normal levels. This may take some time, especially if concentrations above the recommended levels have been used or if the test gas has been applied for an extended period of time.
Ensure that any latched faults or alarms that have been generated by the testing have been cleared.
Make sure that the gas detection system is fully restored to its normal operating state and that any control system overrides used during the testing have been removed.
The alarms generated by the testing will have created entries in the transmitter’s history log.
Troubleshooting
If the test does not produce a satisfactory result, check the following points.
1. Check the gas cylinder concentration and type are correct
2. Check the expiration date of the cylinder.
3. Check that there is sufficient gas left in the cylinder.
4. Check that there are no leaks on the test system.
5. Make sure that the tubing length is as short as possible and that there are no blockages.
6. Make sure that the transmitter is not displaying a fault before the test is started.
19 Warranty Statement
All products are designed and manufactured to the latest internationally recognized standards by Zellweger Analytics under a Quality Management System that is certified to ISO 9001.
As such, this instrument (including the pump) is warranted under proper use, to the original end-user purchaser, against defects in materials and workmanship within 24 months from the date of first turn-on. Separate warranty conditions apply to the sensor cartridges limited as indicated below. During this period, Zellweger Analytics will repair or replace defective parts on an exchange basis, F.O.B. to approved service centers on a global basis.
This warranty does not cover damage caused by accident, abuse, abnormal operating conditions or extreme poisoning of the sensor cartridge.
Defective goods must be returned by the Buyer at the Buyer’s own expense to Zellweger Analytics premises accompanied by a detailed report stating the nature of the defect. Returned goods must detail the Service Event Number (SE#) clearly on the package and the Buyer shall obtain an SE# by contacting Zellweger Analytics in advance.
If no such report is included then Zellweger Analytics reserves the right to charge an investigative fee (prices available upon request) before any repair or replacement is performed.
Zellweger Analytics shall not be liable for any loss or damage whatsoever or howsoever occasioned which may be a direct or indirect result of the use or operation of the Contract Goods by the Buyer or any Party.
This warranty covers the gas detector and parts sold to the Buyer only by authorized distributors, dealers and representatives as appointed by Zellweger Analytics. A warranty claim will only be accepted if a proof of purchase is submitted and all conditions obtained within this Warranty are met. When in the opinion of Zellweger Analytics, a warranty claim is valid, Zellweger Analytics will repair or replace the defective product free of charge and send it or any replacement back.
The initial warranty period is not extended by virtue of any works carried out there under.
The decision to repair or replace parts shall be determined by Zellweger Analytics.
68
Sensor Cartridge Warranties
All sensor cartridges are supplied with a standard 12 months (1 year) warranty; sensor cartridges purchased with an extended warranty for 24 months (2 year) cover period have the same conditions as the standard cartridge conditions as detailed below but for the longer time period of 24 months.
1. The oxygen sensor cartridge is guaranteed to operate satisfactorily for one year from the date of sale (i.e. the date of shipment) and will be replaced at no charge within that period only if it will not provide a correct reading after calibration by authorized service personnel. Physical or chemical damage, resulting from exposure to improper elements, is not covered.
2. The combustible (LEL) sensor cartridge is guaranteed to operate satisfactorily for one year from the date of sale (i.e. from date of shipment). If it will not provide a correct reading after calibration by authorized service personnel, it will be replaced at no charge within that period. This guarantee is void if it has been contaminated by some unusual substance, including but not limited to, water and/or other liquids, oily compounds, corrosives, silicones, lead vapors, extremely high concentrations of combustible gases, and various particulates which may inhibit gas flow to the sensor cartridge element.
3. The toxic sensor cartridges are guaranteed to operate satisfactorily for one year from the date of sale (i.e. the date of shipment). They will be replaced at no charge within that period only if they will not provide a correct reading after calibration by authorized service personnel, and only if the sensor cartridge membranes exhibit no physical or chemical damage resulting from exposure to improper substances.
Note that the MIDAS
®
sensor cartridge stores an electronic history of each individual sensor cartridge’s
manufacture, warranty and calibration activity that is accessible by authorized Service personnel.
Conditions and Exclusions
To maintain this warranty, purchaser must perform maintenance and appropriate testing as prescribed in the instrument operation manual, including prompt replacement or repair of defective parts and such other necessary calibration, maintenance and repair as may be required in the reasonable judgment of Zellweger Analytics. Normal wear and tear, and parts damaged by abuse, misuse, negligence or accidents are specifically excluded from the warranty.
Purchaser acknowledges that, notwithstanding any contrary term or provision in the purchaser’s purchase order or otherwise, the only warranty extended by Zellweger Analytics is the express warranty contained herein. Purchaser further acknowledges that there are no other warranties expressed or implied, including without limitation, the warranty of merchantability or fitness for a particular purpose; that there are no warranties which extend beyond the description of the face hereof; that no oral warranties, representations, or guarantees of any kind have been made by Zellweger Analytics, its distributors or the agents of either of them, that in any way alter the terms of this warranty; that Zellweger Analytics and its distributors shall in no event be liable for any consequential or incidental damages, including but not limited to injury to the person or property of the purchaser or others, and from other losses or expenses incurred by the purchaser arising from the use, operation, storage or maintenance of the product covered by the warranty; that Zellweger Analytics liability under this warranty is restricted to repair or replacement of defective parts at Zellweger Analytics sole option; and that Zellweger Analytics neither assumes nor authorizes any other person to assume for it any other warranty. The warranty shall be void if serial numbers affixed to the products are removed, obliterated or defaced.
Contact Zellweger Analytics
For all information on ordering, spares and other technical issues please contact the Zenter Customer Center:
Zellweger Analytics – Zenter USA Zellweger Analytics – Zenter Europe & Asia
400 Sawgrass Corporate Pkwy. Wilstrasse 11-U11
Suite 100 CH-8610 Uster
Sunrise, FL 33325 USA
Switzerland
Tel: +1 954 514 2700 + 41 (0) 1 943 4300
Toll Free: +1 800 538 0363 +41 (0) 1 943 4398
Fax: +1 954 514 2784
sales@zelana.co.uk
Email: sales@zelana.com
Website: www.zelana.com
69
20 SOFTWARE MENU STATE CHARTS
20.1 Top Level
Midas Simulator State Chart
Monitoring
Status
Review
Mode
Mode
Selection
-
Setup
Mode
Selection
-
Calibratio
n
Mode
Selection -
Test
Cancel
Up (1 Sec -
No
Passcode)
Down (1 Sec ­No Passcode
)
Down
Up
Down
Up
Down
Down
Up
Cancel
Cancel
Accept
Top Level
Accept (Check)
Up
Down
Cancel (X
)
Monitor
Reset Alarms
and Faults (2 Sec Hold)
Additional Button Functions
Setup Mode
Up
Accept
Calibration
Mode
Accept
Test
Mode
Enter
Passcode
Up (1 Sec -
With Passcode)
Down (1 Sec -
With Pa
sscode)
Cancel
Accept
(With Correct
Pa
sscode)
Accept
(With Incorrect Passcode)
Test - Inhibit
Enter
Passcode
Cancel
Accept
(With Correct
Passcode)
Accept
(With Incorrect Passcode)
Accept (1 sec -
No
Passcode)
Accept
(1
sec - With Passcode)
70
20.2 Review Mode
Midas Simulator State Chart
Function
Selection
-
Show Alar
m
Function
Selection -
Show
Fault
Cancel
Down
Up
Review
Function
Selection
-
Show
Network
Down
Up
Accept
Cancel
Down
Up
Review - Network
Function
Selection
-
Show
Da
te/Time
Down
Review - Date/Time
Accept
Monitoring
Status
Function
Selection
-
Show
Even
t Log
Accept
Review - Event Log
Up
Down
Cancel
Cancel
Down
Up
Cancel
Function
Selection
-
Show
Gas Ca
l
Up
Down
Up
Cancel
Accept
Review - Alarm
Accept
Review - Fault
Accept
Review - Gas
Calibration
Function
Selection
-
Show
Software
In
fo
Down
Up
Accept
Review - Software
In
fo
Cancel
71
20.3 Review Software Info, Alarm, Fault and Gas Calibration
Midas Simulator State Chart
Function
Selection
-
Show Alar
m
Function
Selection
-
Show Fault
Show
Alarm Rela
y
Off
State
Show Alarm
La
tching
Show Fault
Latching
Show Fault
Relay Of
f
State
Accept
Accept
Accept
Accept
Accept
Accept
Review - Software Info, Alarm, Fault & Gas Calibration
Show
Alarm
2
Setpoint
Show
Alarm
1
Setpoint
Accept
Accept
Show Alarm
Dela
y
Accept
Function
Selection
-
Show
Gas Ca
l
Show Last
Gas Ca
l
Month An
d
Date
Show Days
To Next
Gas
Ca
l
Show Last
Gas Ca
l
Year
Accept
Accept
Accept
Accept
Show
Alarm
1
Type
Accept
Accept (Check)
Up
Down
Cancel (X
)
Review
Returns to Normal
Operation - Monitoring
Goes
to next Function Selection state. See
Review
state chart for Function Selection
sequence.
Additional Button Functions
Goes to previous Function Selection state. See Review
state chart for Function Selection
sequence
.
Show
Alarm
2
Type
Show
Ga
s
Accept
Accept
Function
Selection
-
Show
Software
In
fo
Show Software Revision
Show Software
Checksum
Accept
Accept
Accept
Show Fault
Rela
y
Configura
tion
Accept
72
20.4 Review Date/Time and Network
Midas Simulator State Chart
Review - Date/Time & Networ
k
Function
Selection
-
Show
Networ
k
Show
Subnet
4th Digi
t
Show
Subnet
3rd Digi
t
Accept
Accept
Accept
Show IP
Addres
s
1s
t Digit
Show
IP Addres
s
Auto
/
Manual
Accept
Show
Subnet
2nd Digit
Show
Subnet
1st Digi
t
Accept
Accept
Accept
Show IP
Addres
s
3rd Digi
t
Show IP
Addres
s
2nd Digi
t
Accept
Accept
Show IP Addres
s
4th Digit
Accept
Function
Selection
-
Show
Date/Time
Show
Time
Show
Month an
d
Date
Accept
Accept
Accept
Show
Year
Accept
Accept (Check)
Up
Down
Cancel (X
)
Review
Retu
rns to Normal Operation - Monitoring
Goes
to next Function Selection state. See
Review
state chart for Function Selection
sequence
.
Additional Button Functions
Goes to previous Function Selection state. See Review
state chart for Function Selection
sequence
.
73
20.5 Review Event Log
Midas Simulator State Chart
Review
- Event Log
Function
Selection
-
Show
Even
t Log
Show
Event Data
Show
Even
t Time
Accept
(After Last
Even
t)
Accept
Accept
(Go To
Firs
t
Event)
Show
Event Month
and Date
Accept
Accept
(Before Last
Event - Go To
Next
Event)
Notes:
1)
Events are in chronological order starting with the newest event.
2)
Maximum number of e vents is 7.
3) At Function Selec tion - Show Event Log, the dots from the flow meter show the number of events available.
4) When showing even
t data, the dots from the flow meter show the index of the current event.
Accept (Check)
Up
Down
Cancel (X
)
Review
Returns to Normal
Operation - Monitor ing
Goes
to next Function Selection state. See
Review
state chart for Function Selection
sequence
.
Additional Button Functions
Goes to previous Function Selection state. See Review
state chart for Function Selection
sequence
.
74
20.6 Set-up Mode
Midas Simulator State Chart
Mode
Selection
-
Se
tup
Function
Selection
-
Set Alarm
Function
Selection
-
Set Faul
t
Cancel
Accept
Down
Up
Setup
Function
Selection
-
Se
t
Passcode
Down
Up
Accept
Cancel
Down
Up
Setup - Passcode
Function
Selection
-
Set
Network
Down
Up
Setup - Network
Accept
Cancel
Cancel
Function
Selection
-
Set Ga
s
Calibratio
n
Down
Up
Cancel
Function
Selection
-
Se
t
Da
te/Time
Setup - Date/Time
Accept
Down
Up
Cancel
Accept
Setup - Alarm
Accept
Setup - Fault
Accept
Setup - Gas
Calibratio
n
Accept (Check)
Up
Down
Cancel (X
)
Setup
When
at Function Selection, 1 second hold goes
to
Function Selection - Zero Gas Calibration in
Calibration.
When
at Function Selection, 1 second hold goes
to
Function Selection - Bump Test in Test.
Additional Button Functions
75
20.7 Set-up Alarm, Fault and Gas Calibration
Midas Simulator State Chart
Function
Selection
-
Set Alarm
Function
Selection
-
Se
t Fault
Set Alarm Relay Of
f
Stat
e
Set Alarm
Latching
Set Fault
Rela
y
Configuratio
n
Set Fault La
tching
Accept
Accept
Cancel
Cancel
Cancel
Accept
Accept
Cancel
Accept
Setup - Alarm, Fault & Gas Calibration
Set Alarm 2
Setpoint
Set Alarm 1
Se
tpoint
Accept
Accept
Cancel
Cancel
Set Alarm
Dela
y
Accept
Function
Selection
-
Se
t Gas
Calibration
Set Gas
Calibratio
n
Period
Cancel
Accept
Accept
Set Alarm 1
Type
Cancel
Accept
Cancel
Accept (Check)
Up
Down
Cancel (X
)
Setup
Save
new set of values if last step in sequence.
Abort changes.
When
setting value, increments or toggle value
(1 second hold for fas ter increment).
When
setting value, decrements or toggle value
(1 second hold for fas ter decrement).
Additional Button Functions
Set Alarm 2
Type
Set Gas
Accept
Accept
Cancel
Cancel
Set Fault Relay Of
f
State
Accept
Cancel
Accept
76
20.8 Set-up Date/Time and Network
Midas Simulator State Chart
Setup - Date/Time & Networ
k
Function
Selection
-
Set
Network
Set Subnet
4th Digi
t
Set Subnet
3rd Digi
t
Accept
Accept
Cancel
Cancel
Accept
Set IP
Addres
s
1s
t Digit
Set
IP Address
Auto
/
Manual
Accept
(Man.)
Cancel
Cancel
Set Subnet
2nd Digi
t
Set Subnet
1st Digi
t
Accept
Accept
Cancel
Accept
Set IP
Addres
s
3rd Digi
t
Set IP
Addres
s
2nd Digit
Accept
Accept
Cancel
Set IP
Addres
s
4th Digi
t
Accept
Cancel
Cancel
Cancel
Accept
(Auto)
Function
Selection
-
Se
t
Date/Tim
e
Set Minutes
Set Hours
Accept
Accept
Cancel
Cancel
Accept
Set Year
Set Date
Format
Accept
Cancel
Cancel
Accept
Set MonthSet Date
Accept
Accept
Cancel
Cancel
Accept (Check)
Up
Down
Cancel (X
)
Setup
Save
new set of values if last step in sequence.
Abort changes.
When
setting value, increments or toggle value
(1 second hold for faster increment).
When
setting value, decrements or toggle value
(1 second hold for faster decrement).
Additional Button Functions
77
20.9 Set-up Pass Code
Midas Simulator State Chart
Setup - Passcode
Function
Selection
-
Set
Passcode
Repeat
Passcode
4th Digi
t
Repeat
Passcode
3rd Digi
t
Accept
Accept
Cancel
Cancel
Accept
Set
Pass
code
2nd Digi
t
Set
Passcode
1s
t Digit
Accept
Cancel
Cancel
Repeat
Passcode
2nd Digi
t
Repeat
Passcode
1st Digi
t
Accept
Accept
Cancel
Accept
Set
Passcode
4th Digi
t
Set
Pass
code
3rd Digi
t
Accept
Accept
Cancel
Cancel
Cancel
Accept (Check)
Up
Down
Cancel (X
)
Setup
Saves new passcode after last entry if the two ente
red passcodes are the same.
Abort changes.
When
setting value, increments value (1 second
hold
for faster increment).
When
setting value, decrements value (1 second
hold
for faster decrement).
Additional Button Functions
78
20.10 Calibration Mode
Midas Simulator State Chart
Mode
Selection
-
Calibration
Function
Selection
-
Zero
Gas
Calibration
Cancel
Accept
Calibration
Function
Selection
-
Span
Gas
Calibration
Cancel
Down
Up
Down
Up
Accept (Check)
Up
Down
Cancel (X
)
Calibration
When
at Function Selection, 1 second hold goes
to
Function Selection - Bump Test in Test.
When
at Function Selection, 1 second hold goes
to
Function Selection - Set Alarm in Setup.
Additional Button Functions
Function
Selection
-
4-20
mA
Calibra
tion
Function
Selection
-
Flow
Ca
libration
Down
Up
Down
Up
Accept
Calibration - 4-20 mA
Calibra
tion
Accept
Calibration - Span Gas
Calibra
tion
Accept
Calibration - Zero Gas
Calibration
Cancel
Cancel
Accept
Calibration - Flow
Ca
libration
79
20.11 Calibration Gas Zero and Span
Midas Simulator State Chart
Function
Selection
-
Zero Ga
s
Calibration
On Calibration
Success or
Failur
e
Cancel
Accept
Calibration - Gas Zero & Span Calibration
Accept (Check)
Up
Down
Cancel
(X)
Calibration
Abort changes
When
setting value, increments value (1 second
hold fo
r faster increment).
When
setting value, decrements value (1 second
hold fo
r faster decrement).
Additional Button Functions
Zero
Calibration
Prepare Zero
Ga
s
Accept
Cancel
Cancel
Show Gas
Calibration
Result
Accept
Stable
Over-Range
Unstable -Rising
Unstable - Flat
Unstable - Falling
Unde
r-Range
Calibration Status
Indicators In
Gas
Bottle Icon (based
on
last 5 seconds)
Current Settings:
Max Delta for Stability = 2% of Full Scale Ze
ro Gas Range = +/- 10% of Full Scale
Span
Gas Range = Target Conc +/- 20% of Full Scale
Stable
Time Required For Calibration Success = 15 seconds
Calibration Timeout = 60 second
s
Gas Calibration Result Codes:
PASS Successful Calibration FL:0H Zero Calibration Timeout - Over-Rang
e
FL:0L Zero Calibration Timeout - Under-Rang
e FL:0U Zero Calibration Timeout - Unstable FL:SH Span Calibration Timeout - Over-Range FL:SL Span Calibration Timeout - Unde
r-Range
FL:SU Span Calibration Timeou
t - Unstable
Function
Selection
-
Span
Gas
Calibra
tion
Span
Calibra
tion
Select Cal
Ga
s
On Calibration
Success or
Failur
e
Accept
Cancel
Cancel
Zero Cal
Required
Messag
e
Accept (if Zero Cal
Not Re
quired)
Cancel
Show Gas
Calibra
tion
Result
Accept
Cancel
Accept (if Zero Cal
Required)
Accept
At Calibration Result state, saves new calibration values
if calibration was successful.
Select Cal
Gas Type
Enter Span
Gas Conc
Accept
Accept
Cancel
Cancel
80
20.12 Calibration - Flow Calibration
Midas Simulator State Chart
Function
Selection
-
Zero Ga
s
Calibra
tion
After A/D Readings
are co
mpleted
Accept
Read Flow Zero Offset
Prepare For
Calibratio
n
Accept
Cancel
Cancel
Calibration - Flow Calibration
Enter
Setpoi
nt 2
Flow
Adjust
Setpoint
1
Flow
Adjust
Setpoi
nt 2
Flow
Enter
Setpoint
1
Flow
Read A/D For
Setpoint
1
Read A/D For
Setpoi
nt 2
Accept (Check)
Up
Down
Cancel
(X)
Calibration
Abort changes
When
adjusting flow, increases flow (simulated
on flow
slider). When entering flow, increments
value
(1 second hold for faster increment).
When
adjusting flow, decreases flow (simulated
on flow
slider). When entering flow, decrements
value
(1 second hold for faster decrement).
Additional Button Functions
Accept
Accept
Accept
Accept
After A/D Readings
are co
mpleted
After A/D Readings
are comple
ted
Cancel
Cancel
Cancel
Cancel
Cancel
Cancel
Notes:
1)
Setpoint 1 values:
Target Flow
: 350 cc/min Minimum Flow: 300 cc/min Maximum Fl ow
: 400 cc/min
2)
Setpoint 2 values:
Target Flow
: 650 cc/min Minimum Flow: 600 cc/min Maximum Fl ow
: 700 cc/min
3)
For “Adjust Flow” states, flow would be adjust to approximately
the target flow ra
te using the arrow buttons and would be
measured
on an external flowm eter.
4)
For “Enter Flow” states, the actual flow shown on the external
flowme
ter would be entered.
5)
For “Read Flow” states, the display will count down the read
process. Final values
would be saved after the Read Setpoint 2
Flow state.
81
20.13 Calibration - 4-20 mA
Midas Simulator State Chart
Calibration
- 4-20 mA Calibration
Accept (Check)
Up
Down
Cancel (X
)
Calibration
Abort changes.
When
adjusting, increases output level (1 second
hold
for continuous increase).
When
adjusting, decrease s output level (1 second
hold
for continuous decrease).
Additional Button Functions
Saves new calibration values after last adjustment
.
Function
Selection
-
4-20
mA
Calibra
tion
Adjust
4 mA
Adjust
20
mA
Cancel
Accept
Accept
Cancel
Accept
82
20.14 Test Mode
Midas Simulator State Chart
Mode
Selection
-
Test
Function
Selection
-
Bump
Test
Cancel
Accept
Test
Accept (Check)
Up
Down
Cancel (X
)
Test
When
at Function Selection, 1 second hold goes
to
Function Selection - Set Alarm in Setup.
When
at Function Selection, 1 second hold goes
to
Function Selection - Zero Gas Calibration in
Calibration.
Additional Button Functions
Function
Selection
-
Alarm/Faul
t
Simulation
Down
Up
Down
Up
Accept
Test - Bump Test
Accept
Test - Alarm/Fault
Simulation
Cancel
Function
Selection
-
Inhibi
t
Accept
Test - Inhibit
Down
Up
83
20.15 Test Bump, Alarm/Fault Simulation
Midas Simulator State Chart
Function
Selection
-
Bump Test
Bump
Test
Cancel
Accept
Test - Bump Test & Alarm/Fault Simulation
Function
Selection
-
Alarm/Fault
Simulatio
n
Request
Alarm
1
Simulatio
n
Cancel
Accept
Request
Alarm
2
Simulation
Request
Faul
t
Simulation
Confirm Alarm
1
Simulatio
n
Confirm Alarm
2
Simulation
Confirm
Faul
t
Simulation
Alarm 1
Simulatio
n
Active
Alarm 2
Simulation
Active
Fault
Simulation
Active
Cancel
Cancel
Accept
Accept
Cancel
Cancel
Accept
Accept
Cancel
Cancel
Accept
Accept
Cancel
Cancel
Down
Up
Down
Up
Down
Up
84
20.16 Test Inhibit
Midas Simulator State Chart
Test
- Inhibit
Accept (Check)
Up
Down
Cancel (X
)
Test
Abort changes
When
setting value, increments value (1 second
hold fo
r faster increment).
When
setting value, decrements value (1
second
hold for faster decr ement).
Additional Button Functions
At Set Inhibit Timeout state or Set Inhibit Level state
(if Inhibit is None), saves new values.
Function
Selection
-
Bump
Test
Set Inhibit
Leve
l
Cancel (if entered
from Test
Menu)
Accept
Set Inhibit
Timeout
Cancel (if entered
from
Test Menu)
Monitoring
Status
Cancel (if entered
directly
from
Moni
toring Status)
Cancel (if entered
directly from
Monitoring Status
)
Accept (if Inhibit Level
not set to None
)
Accept (if entered
directly
from
Moni
toring Status)
Accept (if entered
from Test Menu)
Accept (if entered
directly from
Monitoring Status
and Inhibi
t Level
set to None
)
Accept
(if entere
d
from
Test
Menu
and
Inhibit Leve
l
se
t to None)
Start here if
en
tered directly
from
Monitoring Status
The inhibit levels are:
a) None - no functions are inhibited.
b) Alarms Inhibited - alarm events will be detected, but alarm outputs (relays, 4-20 mA current loop and Ethernet) will be disabled.
c) Alarms and Faults Inhibited - alarm and fault events will be detected, but alarm and fault outputs (relays, 4-20 mA current loop and Ethernet) will be disabled.
d) Full Inhibit - all monitoring functions inhibited. No monitoring is performed and not alarms or faults (except for Inhibit Timeout) will be reported.
85
21 Contact Details
Contact your local Zellweger Analytics representative or visit Zellweger Analytics’ web site at: http://www.zelana.com
Specifications subject to change without notification.
MDA Scientific and MIDAS
®
are registered trademarks of Zellweger Analytics, Inc.
For Emergency (after hours) Service, call: +1 (847) 634-2840
PLEASE NOTE:
While every effort has been made to ensure accuracy in this publication, no responsibility can be accepted for errors or omissions. Data may change, as well as legislation, and you are strongly advised to obtain copies of the most recently issued regulations, standards, and guidelines. This publication is not intended to form the basis of a contract.
© 2004 Zellweger Analytics
© 2004 Zellweger Analytics Issue 1.1 2/2005 MIDAS-A-001
Loading...