Klein Tools VDV500-123 User manual

Page 1
ENGLISH
INSTRUCTION MANUAL
Probe-Pro Tracing Probe
• EASY-TO­UNDERSTAND STATUS LEDs
• REPLACEABLE PROBE TIP
* When used with VDV500-063 Toner-Pro
VDV500-123
ESPAÑOL pág. 9 FRANÇAIS p. 17
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
ENGLISH
GENERAL SPECIFICATIONS
The Klein Tools VDV500-123 Probe-Pro is a professional-series tone tracer, featuring an inductive probe with speaker for amplification, and LED light for use in dark spaces. It also features a headphone jack for use in extreme noise environments.
Operating Altitude: 6562 ft. (2000 m) maximum
Relative Humidity: 10% – 90% non-condensing
Operating Temp: 14° to 122°F (-10° to 50°C)
Storage Temp: -4° to 140°F (-20° to 60°C)
Dimensions: 1.75" x 8.88" x 1.13" (44 x 226 x 29 mm)
Weight: 5.7 oz. (161.6 g) including batteries
Battery Type: 4 x 1.5V AAA Alkaline
Battery Life: Active: 25 hours
Standby/Storage: 3 years Auto-Power Off: After 10 minutes of inactivity
Specifications subject to change.
FEATURE DETAILS
P1
Replaceable Inductive Polymer Tip (VDV999-068)
P2
Worklight
P3
Worklight On/Off Button
P4
Signal Strength Indicator
P5
Power On/Off Button
P6
Volume Increase Button
P7
Volume Decrease Button
P8
Battery Status Indicator
P9
Speaker
P10
Lanyard Slot
P11
"-" (Negative) Terminal
P12
"+" (Positive) Terminal
P13
Terminal Release Buttons
P14
Verification Indicator
P15
Battery Cover
P16
Battery Cover Screw (#2 Phillips)
P17
3.5mm Headphone Jack*
* CAUTION: Excessive volume can cause permanent hearing damage.
Use as low a volume as possible.
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
P3
FEATURE DETAILS
P1
P2
P4
P5
P6
P7
P9
P10
SYMBOLS ON TESTER
Warning or Caution Always wear approved eye
protection
Do NOT use on energized circuits
Read instructions
Conformité Européenne -
Conforms with European Economic Area directives
UKCA: UK Conformity Assessment
This symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal
P16
P8
P15
P17
P13
P11 P12
P14
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
ENGLISH
WARNINGS
To ensure safe operations and service of the instrument, follow these instructions. Failure to observe these warnings can result in re, electric shock, severe injury or death.
The Probe-Pro is designed for use on extra-low voltage cabling
systems (less than 60 volts) for testing when NOT energized. DO NOT use instrument if wet, as it could pose a shock hazard.
DO NOT use instrument if damaged in any way.
Turn off instrument before attempting to replace batteries.
The battery door must be in place and secure before you operate the
instrument. DO NOT open the case, other than the battery compartment.
OPERATING INSTRUCTIONS
READ ALL INSTRUCTIONS BEFORE OPERATING AND RETAIN INSTRUCTIONS FOR FUTURE REFERENCE
TRACING PAIRED WIRES (FIG. 1)
1. Connect tone generator (VDV500-063 Toner-Pro recommended, sold seperately) to the wires of the pair to be traced, following the tone generator's instructions.
2. Turn the Probe-Pro on by pressing the Power On/Off button P5.
3. At the far end of the cable, spread the wires apart at least 2" (51 mm), if possible.
4. Use the Probe-Pro to scan the cable’s wire pairs. Move the Probe-Pro's tip
slowly across the wires (FIG. 1). The Probe-Pro’s volume will increase
P1
as it approaches the toned pair. When the Probe-Pro’s volume is high over the first wire, low in the middle (between) the two wires, and high over the second wire, you have located the pair of wires you are tracing. Use the
Volume Increase
P6 and
Volume Decrease
P7 buttons to adjust
the volume.
FIG. 1
VDV500-063 (sold seperately)
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2"
(51 mm)
Page 5
OPERATING INSTRUCTIONS
VERIFYING TONED PAIRS (FIG. 2)
1. Connect tone generator (VDV500-063 Toner-Pro recommended, sold seperately) to the cable to be traced, following the tone generator's instructions.
2. On the other end of the cable, remove approximately 2 mm (3/32") of the outer jacket of each of the wires believed to be the toned pair.
3. Insert the wire connected to the positive test clip of the tone generator to the
"+" (Positive) Terminal clip of the tone generator to the terminals, press their respective wire(s) and release (FIG. 2). The
and the wire curently in the negative test
,
P12
"-" (Negative) Terminal
Terminal Release buttons Verification Indicator
. To access the
P11
P13
will illuminate
P14
, insert the
when the correct pair is inserted.
FIG. 2
2 mm
(3/32")
Press
TRACING NON-PAIRED WIRES (FIG. 3)
1. Connect tone generator (VDV500-063 Toner-Pro recommended, sold seperately) to the wires to be traced, following the tone generator's instructions.
2. Turn the Probe-Pro on by pressing the Power On/Off button P5.
3. At the far end of the cable, spread the wires at least 2" (51 mm) apart, if possible.
4. Use the Probe-Pro to scan the cable’s wire pairs. Move the Probe-Pro's tip P1 slowly across the wires. The Probe-Pro’s volume will increase as it approaches the toned wire.
FIG. 3
VDV500-063 (sold seperately)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2"
(51 mm)
5
Page 6
ENGLISH
OPERATING INSTRUCTIONS
USING THE PROBE'S WORKLIGHT
The Probe-Pro has a worklight P2 to aid in illuminating dark or low-light work areas. Press the Worklight On/Off button P3 to turn the light on and off.
MAINTENANCE
BATTERY REPLACEMENT
When the Battery Status Indicator P8 blinks, the batteries must be replaced.
1. Turn off instrument(s) before attempting to replace batteries.
2. Loosen screw
3. Remove and properly dispose of four 1.5V AAA batteries.
4. Install new batteries (note proper polarity).
5. Replace battery cover and fasten securely with screw.
To avoid risk of electric shock, do not operate while battery door
on battery cover
P16
.
P15
is removed.
PROBE-PRO TIP REPLACEMENT (KLEIN CAT. NO. VDV999-068)
The tip
P1 of the Probe-Pro is replaceable if damaged. To remove and
replace tip:
1. Turn tip 1/4 turn and pull gently to remove.
2. Insert new tip with key in proper orientation and push gently.
3. Rotate 1/4 turn to lock into place.
CLEANING
Be sure instrument is turned off and wipe with a clean, dry lint-free cloth.
Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the GENERAL SPECIFICATIONS section, allow the equipment to return to normal operating conditions before using.
Do not place equipment and its accessories in the trash. Items must be properly disposed of in accordance with local regulations. Please see or additional information.
6
instrument
DISPOSAL / RECYCLE
is not in use for a prolonged
WARRANTY