Keysight Technologies N9330B User Manual

Page 1
Keysight N9330B Handheld Cable & Antenna Tester
Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com.
User’s Guide
Page 2

Notices

© Keysight Technologies, Inc. 2009-2014
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
N9330-90011
Edition
Edition 2, July 2014
Printed in China
Published by: Keysight Technologies
No 116 Tianfu 4th street Chiengdu, 610041 China
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software
documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Page 3
1 Overview
Introduction 2
Basic measurement functionality 2
Front Panel Overview 4
Top Panel Overview 5
Instrument Markings 6
2 Getting Started
Checking the Shipment 8
Safety Considerations 9
Electrical Requirements 11 Electrostatic Discharge (ESD) Precautions 11
Power Requirements 12
Working with Batteries 14
Installing a Battery 14 Viewing Battery Status 14
Your first 10 minutes with the N9330B 16
Power on the N9330B 16 Making the first measurement 17
South Korea Class A EMC Declaration 22

Contents

3 Making Measurements
Selecting a Measurement Mode 24
Calibrate the N9330B before making measurements 25
Calibration Mode 25 Types of Calibrators 26
Measuring Return Loss/SWR/Cable Loss 29
Perform a basic return loss measurement 29 Adjusting the measurement resolution 30 Using Marker/Limit 30
Measuring Distance to Fault (DTF) 31
Perform a DTF measurement 31 Adjusting the measurement resolution 32 Using Marker/Limit 32
Page 4
Measured distance 33
Making an Average Power Measurement 34
Making a Basic Average Power Measurement 35 Figure 3-1. Connection with base station 36
4 Using Functions
Using Markers and Pass/Fail Limit Lines 40
Using markers 40 Using Pass/Fail Limit Lines 41
Processing Traces 43
Trace Operation 43 Trace overlap 44
Saving and Recalling a file 45
Save a trace data 45 Save an instrument setup 46 Copy a screen 46 Editing a file name 47 Recall a trace data 47 Recall an instrument setup 48
Viewing system statistics 49
Check system status 49 Running a self test 49
Some Helpful Hints 50
Preset the tester 50 Adjust amplitude scale 51 Using single sweep 51 Enabling interference immunity 51 Quick access to freq/dist setups 52 Connector Care 52
System Setups 58
Setting screen resolution 58 Setting power manager 58 Choosing a unit 58 Setting date and time 59 Firmware upgrade 59 Key light setting 59 Setting the brightness of key light 60 Light Setting 60
Page 5
5Key Reference
AMPTD 62
Top 62
CAL 64
Bypass Electronic Calibrator 64
Distance 66
Start 66 Stop 66 DTF Config 67 Cable Loss 68 Velocity Factor 68 Cable 69 Window 69
Enter and ESC/CLR 70
Enter 70 ESC/CLR 70
Frequency 71
Start 71 Stop 71 Signal Standard 71
Limit 72
Limit 1 72 Limit 2 72 Limit Beep 73
Marker 74
Marker 1...6 74 Marker All off 75
Meas/View 76
Resolution 76 Single 76 IIM (Interference Immunity) 76 Trace Overlap 77
Mode 78
Power Meter 78
System 84
Cal Mode 84 Power Manager 84
Page 6
Units 84 Language 84 Clock 84 Status 85 Self Test 85 E Calib Test 85 Intelligent 85 License Option 86 Key Light 86 Key Light Edit 86 Light Set 86
6 Instrument Messages
Error Message List 88
Warning List 89
7 Menu Map
FREQ/DIST 92
AMPTD 93
MEAS/VIEW 94
SYS 95
LIMIT 96
Marker 97
Documentation Conventions:
1. A pair of curly brackets { } indicates a softkey, for example {Start} refers to the Start softkey.
2. A pair of square brackets [ ] indicates a hardkey, for example, [ENTER] refers to the ENTER hardkey.
3. “DUT” refers to a device under test.
Page 7
1 Overview

Overview

1
Page 8
Overview

Introduction

Introduction
Keysight has designed the N9330B handheld cable and antenna tester (25 MHz to 4 GHz) to simplify the job of installing and maintaining cable and antenna systems. In addition to measuring SWR, Return Loss and Cable Loss, the N9330B also performs Distance-to-Fault (DTF) measurements that assist in determining cable defect locations, It is also useful in identifying maintenance necessary to maintain peak performance of your cable and antenna systems.

Basic measurement functionality

Basic measurement functionality of the N9330B includes:
•SWR
• Return loss
•Cable loss
• DTF (SWR and return loss)
• Power Meter (option PWM)
2
Page 9

Optimized Usabilities

The Keysight N9330B has the following features:
• 6.5-inch TFT colorful LCD screen (640×480 pixels) enables you to read the scans easily and clearly outdoors.
• Rechargeable lithium-ion battery supports a minimum of four hours continuous operation when fully charged.
• Print-screen button for quick screen capture and easy file saving onto the memory stick.
• Auto file naming makes editing file names fast and easy.
• The light sensor adjusts the display brightness according to the environment to save power.
• USB interface for easy and fast file saving to PC or memory stick.
• User definable power saving mode.
• Back-lit keys make you find the right keys in darkness easily.
• Arc-shaped handles and a rugged rubber casing ensure a comfortable and firm hold.
Overview
Introduction
3
Page 10
Overview
N9330B
Handheld Cable & Antenna Tester
25 MHz - 4.0 GHz
ENTER
1ABC 2DEF 3 GHI
4JKL 5 MNO 6PQR
7
STU8VWX9YZ_
0SAVE
MODE
FREQ/ DIST
AMPTD
MEAS/ VIEW
SYS
HOLD/ RUN
LIMIT MARKER
CAL ESC/CLR
AUTO­SCALE
RECALL
F1
F2
F3
F4
F5
F6
F7
PRT SCR
12
9
2
1
16
10
11
5
4
3
6
7
8
14
15
13

Front Panel Overview

Front Panel Overview
Caption Brief Description
1 Standby switch Turns the tester On or Off 2 Function hardkeys Including hardkeys: Mode, FREQ/DIST,
3 Preset hardkey Returns the tester to a known state 4 Enter hardkey Confirms a selection or configuration 5 CAL hardkey Enters a calibration procedure 6 ESC/CLR hardkey Escapes from a procedure or clears
7 Save and Recall keys For file operation 8 Auto scale hardkey Sets the amplitude borders automatically 9 Print screen hardkey For quick screen capture 10 Knob Selects an option or edits a number 11 Brightness hardkey Adjusts the screen brightness 12 Buzzer Beeps differently on different actions 13 Light sensor Adjusts brightness automatically 14 Softkeys Indicates current menu functions 15 Screen Displays measured traces and status 16 Strap handle For hand carry
4
AMPTD, MEAS/VIEW, SYS, Hold/Run, Limit and Marker
characters or numbers
Page 11

Top Panel Overview

Ext. Power
Charging
11-25 VDC
RF OUT 50
PC
23 dBm 50 VDC MAX
55 W MAX
8
1
2
3
Caption Brief Description
1 RF Output Connector
Outputs swept signal to DUT
(N-type) 2 USB interface Type B (Host) Connects to a PC 3 USB interface Type A
Connects to a USB memory stick
(Device) 4 LED indicator Lights (On) when external DC power
is connected to the tester
5 LED indicator (Charging) Lights (On) when the battery is
charging
6External DC power
connector
Connects to external DC power, for use with either the AC-DC char­ger-adapter or automotive 12V DC
adapter 7 LAN interface Reserved for future expansion 8Tilt stand (rear of
instrument)
Allows user to lean instrument back
for better viewing when placed on
flat surface
5
4
Overview
Top Panel Overview
6
7
5
Page 12
Overview
ISM1-A
ICES/NMB-001

Instrument Markings

Instrument Markings
The CE mark shows that the product complies with all relevant European Legal Di rectives .
The CSA mark is a registered trademark of the Canadian Standards Association.
All Level 1, 2 or 3 electrical equipment offered for sale in Australia and New Zealand by Responsible Suppliers must be marked with the Regulatory Compliance Mark.
This symbol is an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 4)
The ISM device complies with Canadian Interference-Causing Equipment Standard-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.
The instruction manual symbol: indicates that the user must refer to specific instructions in the manual.
The standby symbol is used to mark a position of the instrument power switch.
Indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
Korea Certification indicates this equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
The symbol indicates this product complies with the WEEE Directive (2002/96/EC) marking requirements and you must not discard this equipment in domestic household waste.
Do not dispose in domestic household waste. To return unwanted products, contact your local Keysight office.
6
Page 13
2 Getting Started

Getting Started

7
Page 14
Getting Started

Checking the Shipment

Checking the Shipment
Check the shipment and order list when you receive the shipment.
• Inspect the shipping container for damages. Signs of damage may include a dented or torn shipping container or cushioning material that indicates signs of unusual stress or compacting.
• Carefully remove the contents from the shipping container, and verify if the standard accessories and your ordered options are included in the shipment.
For any question or problems, contact Keysight Technologies for consultant and service at:
http://www.keysight.com/find/assist
8
Page 15

Safety Considerations

WARNING
WARNING
WARNING
WARNING
WARNING
Keysight has designed and tested the N9330B in accordance with IEC Publication 61010-1 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, and the tester is supplied in a safe condition.
Read the following safety notices carefully before you start to use this tester set to ensure safe operation and to maintain the product in a safe condition.
Personal injury may result if the tester’s cover is removed. There are no operator-serviceable parts inside. Always contact Keysight qualified personnel for service. Disconnect the product from all voltage sources while it is being opened.
This product is a Safety Class I tester. The main plug should be inserted in a power socket outlet only if provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
Electrical shock may result when cleaning the tester with the power supply connected. Do NOT attempt to clean internally. Use a clean soft cloth to clean the outside case only.
Getting Started
Safety Considerations
Danger of explosion if the battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Do NOT dispose of batteries in a fire. Do NOT place batteries in the trash. Batteries must be recycled or disposed of properly.
Always use the three-pin AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause personal injury and product damage.
9
Page 16
Getting Started
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
Safety Considerations
The VxWorks operating system requires full conformity to USB 1.1 or USB 2.0 standards from a USB disk. Not all the USB disk are built that way. If you have problems connecting a particular USB disk, please reboot the analyzer before inserting the next.
Use USB disk in FAT or FAT32 format, with only one partition for data saving and loading with the tester.
If left the battery in tester unused, a fully charged battery will discharge itself over time.
Never use a damaged or worn-out adapter or battery. The charging process will be intermitted if the battery
temperature exceed the safety value. This process will restart when the temperature declines.
If you are charging the batteries internally, even while the tester is powered off, the tester may become warm. To avoid overheating, always disconnect the tester from the AC adapter before storing the tester into the soft carrying case.
Temperature extremes will affect the ability of the battery to charge. Allow the battery to attain ambient operating temperature before use or charging.
Storing a battery in extreme hot or cold places will reduce the capacity and lifetime of a battery. Battery storage is recommended at 25
Always power on the N9330B before connecting an electronic calibrator.
o
C.
The average continuous power input is limited not to exceed +23 dBm, DC voltage to 50 VDC. Instrument damage may result if these precautions are not followed.
10
Page 17
Getting Started
Safety Considerations

Environmental Requirements

Keysight Technologies has designed this tester for use under the following conditions:
• Operating temperature:
o
C to 40oC (using AC-DC adapter)
0
o
C to +50oC (using battery)
–10
•Storage temperature: –40
• Battery temperature: 0
• Humidity: < 95% (40
o
C to 45oC
o
C)
o
C to +70oC

Electrical Requirements

This tester allows the use of either a lithium battery pack (internal), AC-DC adapter shipped with the tester, or optional Automotive +12VDC adapter for its power supply.

Electrostatic Discharge (ESD) Precautions

This tester was manufactured in an ESD protected environment. This is because most of the semiconductor devices used in this tester are susceptible to damage by static discharge.
Depending on the magnitude of the charge, device substrates can be punctured of destroyed by contact or mere proximity of a static charge. The result can cause degradation of device performance, early failure, or immediate destruction. ElectroStatic charges are generated in numerous ways, such as simple contact, separation of materials, and normal motions of persons working with static sensitive devices.
When handling or servicing equipment containing static sensitive devices, adequate precautions must be taken to prevent device damage or destruction.
Only those who are thoroughly familiar with industry accepted techniques for handling static sensitive devices should attempt to service circuitry with these devices.
11
Page 18
Getting Started

Power Requirements

Power Requirements
The AC power supplied must meet the following requirements
Voltage: 100 VAC to 240 VAC
Frequency: 47 to 63 Hz
Power: Maximum 63 W
The AC/DC power supply charger adapter supplied with the analyzer is equipped with a three-wire power cord, in accordance with international safety standards. This power cord grounds the analyzer cabinet when it is connected to an appropriate power line outlet. The power cord appropriate to the original shipping location is included with the analyzer.
Various AC power cables are available from Keysight that are unique to specific geographic areas. You can order additional AC power cords are correct for use in different areas. The AC Power Cords table provides a lists of the available AC power cords, the plug configurations, and identifies the geographic area in which each cable is typically used.
The detachable power cord is the product disconnecting device. It disconnects the main AC circuits from the DC supply before other parts of the product. The front-panel switch is only a standby switch and do not disconnect instrument from AC LINE power.
:
12
Page 19
AC Power Cord
230V 15A
250V 16A
Getting Started
Power Requirements
Plug Type Cable Part
Number
8120-1703 BS 1363/A Option 900
250V 10A
8120-0696 AS 3112:2000 Option 901
250V 10A
8120-1692 IEC 83 C4 Option 902
250V 16A
8120-1521 CNS 10917-2
125V 10A
8120-2296 SEV 1011 Option 906
125V 10A
8120-4600 SABS 164-1 Option 917
8120-4754 JIS C8303 Option 918
125V 15A
8120-5181 SI 32 Option 919
a
Plug Description
/NEMA 5-15P
For use in Country & Region
United Kingdom, Hong Kong, Singapore, Malaysia
Australia, New Zealand
Continental Europe, Korea, Indonesia, Italy, Russia
Option 903 Unite States, Canada,
Taiwan, Mexico
Switzerland
South Africa, India
Japan
Israel
8120-8377 GB 1002 Option 922
China
250V 16A
13
Page 20
Getting Started
CAU-CAUTION

Working with Batteries

Working with Batteries
Full charge the battery before first using the analyzer.

Installing a Battery

Step Notes
1 Open the battery cover Use a phillips type screwdriver,
2 Insert the battery Observe correct battery polarity
3 Close the battery cover Push the cover closed, then
loosen the retaining screw, then pull the battery cover open.
orientation when installing.
re-fasten the cover with the retaining screw.

Viewing Battery Status

Determine the battery status in either of the following methods:
• Check the battery icon in the lower right corner of the front panel screen: it indicates the approximate level of charge.
• The LED on the battery indicates the remaining battery power.
•Press [SYS] > {More} > {Status} to check the current battery voltage.
14
Page 21
Getting Started
CAU-CAUTION
Working with Batteries

Charging a Battery

You may charge the battery both in the tester and in the external battery charger (option BCG).
Connect the automotive adapter to the IT power outlet of your automobile (with option 1DC) for battery recharging.
You may recharge the battery in the N9330B while the tester is operating or when it is turned off.
1 Install the battery in the tester.
2 Plug in the correct AC/DC adapter and switch external
power on.
3 The green LED indicating external power and
charging status lights should turn on, indicating that the battery is charging.
When the battery is fully charged, the green charging LED turns off.
During charging and discharging, the battery voltage, current, and temperature are monitored. If any of the monitored conditions exceed their safety limits, the battery will terminate any further charging or discharging until the error condition is corrected.
The charging time for a fully depleted battery, is approximately four hours.
15
Page 22
Getting Started
CAU-CAUTION
Using the tilt stand

Your first 10 minutes with the N9330B

Your first 10 minutes with the N9330B
The N9330B is fitted with a TFT screen, which is viewable under all lighting conditions.
Do not apply excessive RF or DC signals to the tester RF OUTPUT connector. The maximum damage RF input level is +23 dBm (or
50 VDC signal input).
+

Power on the N9330B

Pressing the standby button switches on the tester. The tester then boots up and runs an internal self-test that includes the following items:
• internal temperature
• battery voltage (if battery powered)
• battery capacity (if battery powered)
16
Page 23
Your first 10 minutes with the N9330B
Getting Started

Preparation for Use

Use [SYS] hardkey to set or check the system settings of the tester.
Setting up the N9330B
Press {Power Manager} to enable or unable the power saving mode. Pressing [SYS] > {Power Manager (On)} turns off the LCD screen and key light after a pre-defined idle time set by pressing {Power Manager} > {Edit}.
Press [SYS] > {More} > {More} > {Key Light} to turn on or turn off the key light.
Press [SYS] > {Language} to select a language displayed on the screen.
Press [SYS] > {Clock} to set the time and date.
Check system status
Pressing [SYS] > {More} > {Status} shows the primary information and current status of the tester.

Making the first measurement

Four easy steps:
1 Select a measurement mode.
2 Set the desired measurement range, or select a
frequency range from the list of signal standards.
3 Calibrate the tester.
4 Connect the DUT (Device Under Test) to the tester RF
OUT connector being careful not to apply a signal which exceeds the maximum signal levels.
17
Page 24
Getting Started
Your first 10 minutes with the N9330B
Step 1. Selecting a measurement mode
Pressing [MODE] brings up a check box for you to choose a measurement mode from the following five modes:
• SWR (Standing Wave Ratio)
• Return loss vs. Frequency
•Cable loss
•DTF-SWR
• DTF-Return loss
• Power Meter
Example: selecting the return loss measurement
1Press [MODE] to bring up the mode selection check
box
2 Use the rotary control knob to select the Return Loss
vs. Frequency
3Press [ENTER] to confirm your selection.
A measurement mode indicator – FREQ-RL will be displayed in the top middle part of the screen.
"Selecting a Measurement Mode" on
See
page 24
for more information.
18
Step 2. Setting a measurement range
Pressing [FREQ/DIST] brings up a submenu for setting the measurement range.
If you previously chose a frequency domain measurement, the submenu allows you to set the start and stop frequency, or select a pre-defined frequency range from the list of signal standards.
If you previously chose a DTF measurement, the submenu allows you to set the start and stop displayed distance, and other cable measurement related parameters.
Page 25
Your first 10 minutes with the N9330B
Getting Started
Example: setting a frequency range from 1
GHz to 2 GHz
1 Press {Start} > [1000] > [ENTER] to set the start
frequency to 1 GHz
2 Press {Stop} > [2000] > [ENTER] to set the stop
frequency to 2 GHz
The start frequency of 1000.0 (in MHz) displays at the left lower corner of screen, and the stop frequency
2000.0 displays at the right lower corner of the screen.
Step 3. Calibrating the N9330B
The N9330B must be calibrated before making an actual measurement (SWR/Return loss/Cable loss/DTF) using the selected frequency span or the full span of the instrument. The N9330B offers two calibration modes:
• FullSpan Cal Performs an Open-Short-Load calibration across the full frequency range. The tester allows you to modify the frequency range for measurement after the FullSpan Cal is performed.
• SelectedSpan Cal SelectedSpan mode. Performs an Open-Short-Load calibration across a user-defined frequency range. The tester requires re-calibration if the Selected Frequency Span is changed.
Key access: [SYS] > {Cal Mode}
Example: Performing a calibration
Assuming the Selected Frequency Span is from 1 to 2 GHz for return loss measurement and the T-combo calibrator tool is being used.
1 Press [CAL] to bring up a dialog box that will guide
you through the calibration. The normal calibration order is open, short, and load.
2 Fasten the OPEN end of the T-combo to the RF OUT
connector on the top panel of the tester, then press [ENTER] to initiate the open calibration.
calibration is the default calibration
19
Page 26
Getting Started
Your first 10 minutes with the N9330B
3 Fasten the SHORT end of the T-combo to the RF OUT
connector, then press [ENTER] to initiate the short calibration.
4 Fasten the LOAD end of the T-combo to the RF OUT
connector, then press [ENTER] to initiate the load calibration.
To abort a calibration, press the [ESC/CLR] button.
For more details of calibration, please refer to
"Calibrate the N9330B before making measurements" on page 25
Step 4. Connecting DUT to RF Output port
Example: Measuring return loss
After you have completed the calibration procedure above. Refer to the following instructions to make a return loss measurement.
• Connect the cable under test to the RF OUT connector on the top panel of the tester.
• The tester immediately performs the measurement and displays a trace on the screen.
If further data analysis is necessary, refer to:
"Using Markers and Pass/Fail Limit Lines" on page 40 Traces" on page 43
.
, "Processing
.
20
Page 27
Your first 10 minutes with the N9330B
Getting Started

Contact Keysight Technologies

Keysight maintains sales and service offices around the world to provide you with complete support for your handheld cable and antenna tester. In any correspondence or telephone conversations, refer to your handheld cable and antenna tester by its product number and the full serial number.
Go to http://www.keysight.com/find/assist for help with:
• product selection, configuration, and purchases.
• technical application assistance, and consulting.
• rental options, and refurbished equipment.
• repair, calibration, education and training.
If you do not have access to the internet, call the appropriate number shown below. Or contact local Keysight Technologies Sales and Service Office.
United States (tel) 800 829 4444
Canada (tel) 877 894 4414
Latin America (tel) +1 (305) 269 7500
China (tel) 800 810 0189
Korea (tel) 080 769 0800
Japan (tel) +81 426 56 7832
Taiwan (tel) 0800 047 866
Europe (tel) +31 20 547 2111
Australia (tel) 1 800 629 485
Other Asia Pacific Countries
(fax) 800 829 4433
(fax) 800 746 4866
(fax) 800 820 2816
(fax) 080 769 0900
(fax) +81 426 56 7840
(fax) 0800 286 331
(fax) +61 (3) 9210 5947
(tel) +65 6375 8100 (fax) +65 6755 0042
Email: tm_ap@keysight.com
21
Page 28
Getting Started

South Korea Class A EMC Declaration

South Korea Class A EMC Declaration
This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
22
Page 29

Making Measurements

3 Making Measurements
23
Page 30
Making Measurements
Selecting “Return Loss” measurement
Power Meter

Selecting a Measurement Mode

Selecting a Measurement Mode
Assuming you wish to make a return loss measurement, you need to:
•Press the [MODE] hardkey to call up the mode selection check box.
• Use the front-panel knob to select a measurement type from the Mode Select list.
•Press the [ENTER] hardkey to confirm your selection.
The icon indicates the following frequency domain measurement:
•SWR
• Return loss
•Cable loss
The icon indicates the following Distance to Fault (DTF) measurement:
•SWR
• Return loss
• Power Meter
24
Page 31

Calibrate the N9330B before making measurements

Making Measurements
Calibrate the N9330B before making measurements
Before making a measurement (SWR/Return loss/Cable loss/DTF) in a specified frequency range or the full span, you must calibrate your N9330B.
It is important to keep the calibration frequency range as close as possible to the actual sweep frequencies necessary for the measurements. Calibrating over a larger frequency range than the actual measurement range will reduce the accuracy of the test results. For example, using a 1 GHz calibration range when the measurement range is only few kHz.
Performing one calibration is acceptable provided the calibration frequency range selected incorporates all intended measurement frequencies. If the frequency span is changed, a new calibration must be performed for that measurement. A new calibration must be performed if any components used in the calibration setup are changed. For example, a short coaxial cable used to connect the tester to the DUT.

Calibration Mode

The N9330B offers two calibration modes:
FullSpan Cal Runs an open-short-load calibration across the full frequency span (25 MHz to 4 GHz). A FullSpan Cal ignores the frequency changes during measurements.
SelectedSpan Cal Performs an open-short-load calibration across a user-defined frequency span. A SelectedSpan Cal requires a re-calibration of the tester, after changing the measurement frequency range. Normally, the selected span should be set as close to the actual swept frequency range.
Key access: [SYS] > {Cal Mode}
25
Page 32
Making Measurements
Calibrate the N9330B before making measurements

Types of Calibrators

Two types of calibrator tools are available from Keysight for use with the tester:
T-combo calibrator This is a T-shaped mechanical calibrator. It consists of a calibrated open, short and 50 load. Using the T-combo calibrator will result in a more accurate calibration than when using the an Electronic calibrator.
Electronic calibrator The Electronic calibrator enables you to perform a one-step calibration process. It automatically performs an open, short, and 50W load calibration. The Electronic calibrator is a faster but less accurate than when using a T-combo mechanical calibrator.
26
Page 33
Calibrate the N9330B before making measurements
Making Measurements

Calibrating procedure introduction

Choose either the T-combo or the Electronic calibrator to perform a calibration process. Different combinations of calibration mode and calibrator may benefit the measurement accuracy differently. (Assuming the measurement resolution remains unchanged):
•Using SelectedSpan calibration mode and the T-combo calibrator provides a more accurate calibration.
• Selecting FullSpan calibration mode and electronic calibrator is the simplest, most convenient calibration.
The tester default setting is the SelectedSpan calibration mode. Keysight recommends using the SelectedSpan calibration mode for best results.
It is important to keep the calibration frequency range as close as possible to the actual sweep frequencies necessary for the measurements. Calibrating over a larger frequency span than the actual measurement range will reduce the accuracy. For example, using a 1 GHz calibration range when the measurement range is only few kHz.
Calibrate with an electronic calibrator
The tester needs about 35 sec. to record the data of an Electronic calibrator the first time a new Electronic calibrator is used with a specific tester.
After the initial use of an Electronic calibrator, the tester requires approximately 12 sec. to perform the automated open-short-load calibration.
The Electronic calibrator performs an automatic, open-short-load consecutively:
1Press [CAL] to bring up a dialog box for calibration
guidance.
2 Connect the Electronic calibrator to the RF OUT
connector on the top panel of the tester. Connect the USB cable between the electronic calibrator and the tester USB port. The correct USB cable is supplied with an Electronic calibrator.
27
Page 34
Making Measurements
NOTE
Calibrate the N9330B before making measurements
3Press [ENTER] to start calibration.
Never connect an electronic calibrator to the tester until after powering on the N9330B.
Calibrate with a T-combo calibrator
Using a T-combo calibrator is a three step process requiring the user to manually change the open-short-load components. The default order of the calibration process is open-short-load.
1Press [CAL] to bring up the dialog box for calibration
guidance.
2 Connect the “Open” component of the T-combo
calibrator to the RF OUT connector of the tester. Press [ENTER] to start the open calibration.
3 Connect the “Short” component of the T-combo
calibrator to the RF OUT connector of the tester. Press [ENTER] to start the short calibration.
4 Connect the “Load” component of the T-combo
calibrator to the RF OUT connector of the tester. Press [ENTER] to start the load calibration.
A “Calibrated” indicator will be displayed on the upper left corner of the screen when the calibration is completed.
Using User Cal Mode A User Cal Mode is an optional selection when using a mechanical calibrator. The User Cal Mode allows the user to choose one cal standard from the open-short-load to perform a customized calibration.
For example, pressing [CAL] > {User Cal Mode} > {Open} initiates a calibration using the Open calibration standard only.
28
Page 35

Measuring Return Loss/SWR/Cable Loss

NOTE
Making Measurements
Measuring Return Loss/SWR/Cable Loss
Return loss is the measurement of the signal reflection characteristics of a cable and antenna system. A return loss measurement is useful in detecting problems in the antenna feedline system and the antenna itself. A portion of the incident power is reflected back to the source from transmission line faults as well as from the antenna itself. The ratio of the reflected voltage to the incident voltage is called the reflection coefficient. The reflection coefficient is a complex number, meaning it has both magnitude and phase information. In S-parameter terms, return loss is referred to as an S11 measurement.
SWR and cable loss measurement derive from return loss measurement.
Test signals can cause interference. When testing cables attached to an antenna, the test signal is radiated from the antenna. Verify that the signal used for the test CANNOT cause interference to other antennas.

Perform a basic return loss measurement

When measuring return loss, SWR, or cable loss with your N9330B, follow the same operation procedure as shown below.
To make a return loss (versus. frequency) measurement over a frequency range of 50 to 400 MHz, perform the following steps:
1Press the [MODE] hardkey to call up a measurement
mode menu. Rotate the front-panel knob to select a measurement mode (for this example, Return Loss). Press the [ENTER] hardkey to confirm your selection. For more detailed information, refer to "Selecting a
Measurement Mode" on page 24.
2Press the {Start} softkey to call up the start frequency
menu. Press {50} > [ENTER] to set the start frequency to 50 MHz.
29
Page 36
Making Measurements
Measuring Return Loss/SWR/Cable Loss
3Press the {Stop} softkey to call up the stop frequency
menu. Press {400} > [ENTER] to set the stop frequency to 400 MHz.
4Press the [CAL] hardkey to call up the calibration
process menu. Follow the instructions for performing the calibration. For more information about calibration, refer to "Calibrate the N9330B before
making measurements" on page 25.
A “Calibrated” indicator will be displayed on the upper left corner of the screen when the calibration is completed.
5 Remove the calibrator and connect the DUT to the RF
OUT connector of the tester. The tester will then measure the return loss of the DUT.

Adjusting the measurement resolution

The tester has three resolution settings, 521, 261, or 131 points. Press the [MEAS/VIEW] hardkey, then the {Resolution} softkey to choose the resolution setting desired, then press the [ENTER] hardkey to enter the selection.
The higher the resolution, the more accurate the measurement results. However, a higher resolution setting will require a longer sweep time.

Using Marker/Limit

For further trace analysis, refer to:
"Using markers" on page 40.
"Using Pass/Fail Limit Lines" on page 41
"Trace Operation" on page 43.
For performing an isolated return loss measurement, refer to:
"Using single sweep" on page 51.
For minimizing the interference, refer to:
"Enabling interference immunity" on page 51.
30
Page 37
Making Measurements

Measuring Distance to Fault (DTF)

Measuring Distance to Fault (DTF)
For this measurement, an internally generated signal emits from the RF OUT port of the tester and is transmitted to the cable and antenna system (DUT). The tester receives and processes the signal power reflected from faults and imperfections.
In performing a DTF measurement, the tester uses frequency domain reflectometry. The transmitted and reflected signals contain information about the distance to the faults. This information is used to determine the physical distance to the faults. The tester displays the physical distance to the probable faults, corrected for cable loss and the velocity propagation factor of the cable.

Perform a DTF measurement

To make a DTF-return loss measurement over a frequency range of 50 to 400 MHz, perform the following steps:
1Press [FREQ/DIST] > {Start} > 50 > [ENTER] to set the
start frequency to 50 MHz.
2Press {Stop} > 400 > [ENTER] to set the stop frequency
to 400 MHz.
3Press the [MODE] hardkey to select a measurement
mode. Rotate the front-panel knob to highlight and select the DTF-Return loss, and press [ENTER] to confirm your selection.
4Press [CAL] to bring up the calibration process
guidance. Follow the instructions to perform a calibration. For more information about calibration, refer to "Calibrate the N9330B before making
measurements" on page 25.
5 Remove the calibrator and connect the cable under
test (DUT) to the RF OUT connector on the top panel of the tester. The DTF measurement starts and the results are displayed on the screen. A peak in the horizontal trace indicates a defective location.
31
Page 38
Making Measurements
NOTE
Measuring Distance to Fault (DTF)
For a more precise measurement, enter the cable loss and velocity factor information.
The measured distance and the displayed distance can be different. The measured distance refers to the distance the tester has measured according to the frequency ranges you specified. You can check the measured distance by pressing {DTF Config} The displayed distance refers to that part of the entire measured distance chosen. The displayed distance may be manually set by pressing [FREQ/DIST] > {Start}, and {Stop}.

Adjusting the measurement resolution

The tester has three resolution settings, 521, 261, or 131 points. Press the [MEAS/VIEW] hardkey, then the {Resolution} softkey to choose the resolution setting desired, then press the [ENTER] hardkey to enter the selection.
The higher the resolution, the more accurate the measurement results. However, a higher resolution setting will require a longer sweep time.

Using Marker/Limit

For further trace analysis, refer to:
"Using markers" on page 40.
"Using Pass/Fail Limit Lines" on page 41.
"Trace Operation" on page 43.
For getting an isolated return loss signal, refer to:
"Using single sweep" on page 51.
For minimizing the interference, refer to:
"Enabling interference immunity" on page 51.
32
Page 39
Making Measurements
Measuring Distance to Fault (DTF)

Measured distance

It is not always understood how frequency range affects measured distance and resolution, and may appears counter-intuitive. This section is intended to help simplify understanding the DTF measurement.
In the following equations:
Resolution refers to the number of points that are measured and used in the displayed trace. Pressing [MEAS/VIEW] > {Resolution} enables you to change the resolution setting.
) refers to the velocity factor of a cable, its
•(V
t/Vc
transmission rate relative to light. The velocity factor may be set by pressing {Velo Factor}.
• F2 and F1 refer to the stop and start frequency (MHz).
The measured distance (in meters) of a DTF measurement is determined by the following equation:
Distance= (Resolution x 150 x V
= 3 x 10
8
V
c
From this equation, you can adjust the measurement:
• To Increase the measured distance:
Increase the number of points,
or
Reduce the frequency span.
• To decrease the measured distance:
Reduce the number of points,
or
Increase the frequency span.
) / (F2–F1)
t/Vc
33
Page 40
Making Measurements
CAU-CAUTION

Making an Average Power Measurement

Making an Average Power Measurement
Average power measurements provide a key metric in transmitter performance.
Base station transmit power must be set accurately to achieve optimal coverage in wireless networks. If the transmit power is set too high due to inaccurate power measurements, undesired interference can occur. If the transmit power is set too low, coverage gaps or holes may occur. Either case may affect system capacity and may translate into decreased revenue for service providers.
Average power can be measured for the channel of interest while the base station is active. All other channels should be inactive. Average power is a broadband measurement. If other signals are present the analyzer will also measure their power contributions.
The maximum power for the RF IN port and the RF OUT port of the analyzer is +20 dBm. The maximum power for the Power Sensor port is +24 dBm. When directly coupled to a base station, the test set can be damaged by excessive power applied to any of these three ports.
To prevent damage in most situations when directly coupling an analyzer to a base station, use a high power attenuator between the analyzer and the BTS.
The analyzer supports the U2000 Series USB power sensors.
The U2000 Series USB power sensors do not need manual calibration and zero routines performed. Calibration and zeroing are performed without removing the power sensor from the source, through internal zeroing. With internal zeroing of U2000 Series USB
power sensors, there is no need to disconnect the sensor or power-off the DUT. The U2000 Series do not
require 50 MHz reference signal calibration, allowing the factory calibration to ensure measurement accuracy. For
best accuracy, users are recommended to perform
34
Page 41
Making an Average Power Measurement
NOTE
NOTE
external zeroing for input signals below -30 dBm for best accuracy.
Making Measurements
If you suspect other signals may be present, it is recommended that you turn off all the other channels and measure average power only on the signal of interest. Another option is to measure channel power (which is less accurate), that filters out all other channels (signals). You can measure channel power for CDMA using the CDMA Analyzer or CDMA Over Air tool. For other modulation formats, use their respective analyzers (that is, GSM, 1xEV-DO, or W-CDMA) or measure channel power using either the spectrum analyzer or the Channel Scanner tool.
Connect the power meter as close as possible to the power amplifier/duplexer output. Do not use a coupled port. Sensors may not be as accurate at the power levels provided by coupled ports.

Making a Basic Average Power Measurement

To make an average power measurement, connect the power sensor and cable, zero and calibrate the meter, before making a measurement.
Zeroing of the Power Meter will occur automatically:
• Every time the Power Meter function is used.
• When a 5 degree C. change in instrument temperature occurs.
• Whenever the power sensor is changed.
• Every 24 hours (min.).
• Before measuring low level signals -for example, 10 dB above the lowest specified power the power sensor is capable of.
Calibrating the Power Meter every time you cycle the power on and off.
In most situations, you can press {Zero} to complete the two steps (zero and cal) together.
35
Page 42
Making Measurements
Making an Average Power Measurement
To Make a Basic Average Power Measurement
You can follow the steps below to make a basic average power measurement.
1Press [Preset] to perform a factory preset.
2Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3 Zero and calibrate the meter. Press {Zero} to make a
Zero operation of the power sensor followed by a calibration operation.
4 Connect the power sensor to the power Ref 50 MHz
port. The analyzer supports the U2000 Series power sensors.
5 Connect the external attenuation, if required, with the
power sensor used.
6 Connect the analyzer to the base station. Connect the
power sensor to the signal to be measured. A typical BTS connection is after the output of the power amplifier and duplexer. The connection requires a 2 ft. cable and two Type-N barrel connectors, in addition to the power sensor and external attenuator.
Figure 3-1 Connection with base station
36
Page 43
Making an Average Power Measurement
Making Measurements
Setting Power Meter Resolution
You can choose from four levels of Power Meter resolution. Higher resolutions provide more accuracy but slow the measurement speed.
1Press [Preset] to perform a factory preset.
2Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3Press {Meas Disp} > {Resolution}. Select 1,2,3 or 4. The
four options represent different resolution as follow:
[1] = 1
[2] = 0.1
[3] = 0.01
[4] = 0.001
Setting the Power Meter’s Top and Bottom End-Points
Setting the end points (Disk Range) close to expected measurement value changes the sensitivity of the Power Meter scale resolution. This is an advantage in viewing small changes in power. However, this will not affect the overall range of the sensor.
1Press [Preset] to perform a factory preset.
2Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3Press {Meas Disp} > {Disp Range} to access the
end-points menu.
4Press {Top} and enter the maximum scale value
desired using the numeric keypad. Then press {dBm} to complete the setup.
5Press {Bottom} and enter the minimum scale value
desired using the numeric keypad. Then press {dBm} to complete the setup.
37
Page 44
Making Measurements
Making an Average Power Measurement
Setting the Power Meter’s Upper and Lower Limits
The internal Power Meter can be configured to detect when a measurement has failed a user predefined upper and lower limits.
1Press [Preset] to perform a factory preset.
2Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3Press {Meas Setup} > {Limits} to access the limits
menu.
4Press {Limits} to activate the limits function. Each time
the softkey is pressed, the selected option changes.
5Press {Upper Limits} and enter the high limit value
using the numeric keypad. Then press {dBm} to complete the setup.
6Press {Lower Limits} and enter the low limit value
using the numeric keypad. Then press {dBm} to complete the setup.
38
Page 45
4 Using Functions

Using Functions

39
Page 46
Using Functions
NOTE

Using Markers and Pass/Fail Limit Lines

Using Markers and Pass/Fail Limit Lines
Markers and limit lines are used for measurement trace analysis.

Using markers

Markers are used to track the specified points on a trace. You may apply up to six markers and visually track up to six points on each trace. Each of the six markers has the following settings:
•On Activates a Marker and places a vertical, colored bar, with the Marker number ranging from M1 to M6 to the foot of the bar.
•Off Deactivates a marker and erases the Marker bar.
•Edit Brings up a dialog box and allows you to set the posi­tion of a marker.
Rotating the knob will move the position of a marker.
• Marker to Peak Places the active marker on the highest indicated value of the trace.
•Marker to Minimum Places the active marker on the lowest indicated value of the trace.
•Delta Is associated with the reference marker (Marker 1 is the default reference marker). Therefore, Marker 1 must be applied before creating the delta. Delta displays the value difference between the reference marker position and the delta marker position. Only one delta marker can be associated with a given reference marker.
40
Page 47
Using Markers and Pass/Fail Limit Lines
Using Functions
General steps for adding a marker
Take adding Marker 1 for example:
Step Note
1 Press [Marker] hardkey to call up
a submenu of markers
2 Press {Mark 1} softkey to enter a
submenu of Marker 1
3 Press {On} softkey to activate
Marker 1
4 Press {Edit} softkey to set the
position of the Marker 1
Marker 1 sets its position refer to its previous position
Rotating the front-panel knob also positions the marker

Using Pass/Fail Limit Lines

Limit lines will alert the user to a measurement that exceeds the upper limit value of a trace. Limits are set in dB, and define the location of the limit lines displayed on the screen.
The tester applies upper limits only. Limit lines are used as pass/fail indicators:
• If the amplitude of a measured trace is below the limit line, a green Pass indicator is displayed in the upper right corner of the screen.
• If the amplitude of a trace is above a limit line, a red Fail indicator is displayed in the upper right corner of the screen. A limit fail beep sound can be turned on for an additional trace fail indicator.
Two types of limit lines
•Single limit A single limit value is applied over the full frequency range, 25 MHz to 4 GHz. A maximum of two single limit lines may be applied.
41
Page 48
Using Functions
Using Markers and Pass/Fail Limit Lines
• Multiple limit segment After setting a limit segment, the limit line is valid only for your selected frequency/distance range and amplitude range. A maximum of five limit segments may be applied.
Setting up a limit line
Refer to the following example to add a limit line. Example: adding limit 1
Step Note
1 Press [Limit] hardkey to access a
submenu of limit settings.
2 Press {Multi Limit} softkey to access
the multi limit segments submenu.
3 Press {Limit 1} softkey to access a
submenu of the limit 1.
4Press {On} softkey to activate a limit
line.
4 Press [Edit] softkey to set the posi-
tion of a limit line.
5Press
{Prev segment/Next segment} to continue setting specifications for a multi-limit trace.
Use this feature if setting multiple limit values on a trace.
choosing the specific seg­ment number.
This will set the limit line at previously configured limit value.
To adjust the limit line to your specifications.
For multi limit traces.
42
Page 49

Processing Traces

NOTE
Trace overlap and trace operation are used in the tester for post trace processing. Very useful in comparing measurements to identify differences and faults.

Trace Operation

Trace operation mathematically adds or subtracts the current measured trace to a previously saved trace. This will provide information that may indicate when a cable or antenna system is degrading in performance and would require maintenance.i
The previously saved trace must have the same setup, with con­sistent measurement parameters as the current trace. For exam­ple, the same mode, resolution, and frequency/distance range.
Example: A saved a trace (data) has been entered into
the local memory of the tester. The trace was measured under return loss mode in GSM full band (880 MHz to 960 MHz), with 521 data points.
Perform the following steps for a trace operation:
1 Press [MEAS/VIEW] > {Trace Operation} to access the
check box for selecting a math operation mode:
Trace – Memory
Current active trace mathematically subtracts the trace from local memory of the tester.
Trace + Memory
Current active trace mathematically adds the trace from local memory of the tester.
2 Rotate the knob to select a mode and press [ENTER].
3 The tester brings up a list of existing trace data in the
local memory the tester.
4 Rotate the knob to select a trace that has the same
measurement mode, resolution and frequency range, and press [ENTER].
The tester displays the mathematically calculated trace on its screen.
Using Functions
Processing Traces
43
Page 50
Using Functions
NOTE
Processing Traces

Trace overlap

The tester allows a previously saved trace to be recalled onto the currently displayed screen. The two displayed traces are identified by different colors.
Trace overlap is a real-time tool for quick comparisons between the current trace and a previously saved trace. The tester cannot save the overlaid traces into either local memory or external USB memory stick.
The previously saved trace must have consistent measurement parameters with the current trace. For example, the same mode, resolution, and frequency/distance range.
Example: To compare the current measurement trace to a saved trace (data) in the local memory of the tester. They were both measured in the return loss mode, GSM full band (880 MHz to 960 MHz), resolution 521 points.
Perform the following steps to do a trace overlap operation:
1Press [MEAS/VIEW] > {Trace Overlap} > {Select Trace}
to display a list of trace data stored in the local memory of the tester.
2 Use the front-panel knob to select a trace and press
[ENTER].
The tester will display the saved trace, while continuously making measurement of the DUT (current measured trace). The two traces are displayed on the screen simultaneously:
Trace one is displayed as a static trace and is green
Trace two is yellow in color and is repeatedly
in color. It is the saved trace results recalled from the local memory of the tester.
scanned. It is the currently measured trace result.
44
Page 51

Saving and Recalling a file

N9330B allows you to save trace data, instrument setups, or the displayed screen graphic information into the local memory of the tester. With a USB memory stick inserted, you can also save trace data and screen copies into an external memory devices.
File Format Max. #
Trace data *. data 200 Instrument setting *.data 15 Screen copy *.jpg Stored onto USB disk
Trace data includes both the measurement data and the instrument setup information. Trace data and instrument setup files are saved with a “.data” file extension.

Save a trace data

Trace data includes the measured data and the current instrument setups. Save Data does not include limit lines. Limit lines can be saved using the Save Setup feature.
•Pressing [Save] > {Save Data} allows you to save the trace data into local memory ONLY. It also activates the auto file naming editor and automatically saves the current trace data with a name that is consecutive to name of the latest saved file. This mode does NOT allow you to edit the file name. For example, if the previously saved file name is “Chicago_site1”, the current trace data will be automatically named as “Chicago_site2”.
• Pressing [Save] > {Save Data as} allows you to:
select a destination where a trace data is saved:
Local or USB.
edit a user defined file name.
See "Editing a file name" on page 47.
Using Functions
Saving and Recalling a file
(Internal memory)
45
Page 52
Using Functions
NOTE
Saving and Recalling a file

Save an instrument setup

The tester will save instrument setups to the local memory. The maximum number of instrument setups that can be saved is 15.
Pressing [Save] > {Save Setup} saves the current instrument settings to the tester’s local memory. The tester will record the following instrument settings:
•calibration mode records FullSpan or SelectedSpan calibration mode.
•measurement mode records measurement modes with their ancroyn: RL, SWR, CL, DTF-RL, or DTF-SWR.
• parameters records frequency/distance range, and date and time. Limit lines, if activated, will be saved using Save Setup.

Copy a screen

The Keysight N9330B offers two methods to copy a screen:
•Press [PRT SCR] to get a quick screen capture and save the file onto an external USB memory stick ONLY. Pressing the button also activates the auto file naming editor and automatically saves the screen shot with a name that is auto-sequential to the number of the latest saved file, from 1 to 65533.
•Pressing [Save] > {Copy Screen} allows you to save a screen shot into local memory or external USB memory stick that will be automatically assigned with a default file name, from 1 to 10.
Insert a memory stick into the USB connector before copying screen.
46
Page 53
Using Functions
Saving and Recalling a file
Both of the methods allow you to save the currently displayed screen in JPEG format. Each screen copied is an 8-bit colored, 640×480 resolution graph. Copy screen will include any limit lines, if shown on screen. However, a screen shot is not editable.

Editing a file name

A valid file name may consist of capitalized letters (A to Z), decimal numbers (0 to 9), and symbols (dot and underscore).
Pressing [Save] > {Save Data as} accesses a dialog box with the last file name used. Press [ESC/CLR] to erase the current file name, before editing or creating a new file name.
Input a new file name Use the numerical keypad to input characters, numbers, and symbols.
• A single press on a numerical button inputs a decimal number.
• Consecutive press on a numerical button selects a letter or symbol.

Recall a trace data

To recall a trace data perform the following steps:
1Pressing [Recall] > {Recall Data} accesses a dialog box
for selecting the trace data to be recalled.
2 Rotate the knob to select a trace data.
3Press [ENTER] hardkey to confirm the selection.
Trace data can be deleted from the trace data list in local memory by:
•Press {Delete Trace} softkey to delete a trace data.
•Press {Delete All} to delete all the trace data from the list.
47
Page 54
Using Functions
Saving and Recalling a file

Recall an instrument setup

The following steps show the steps to recall a stored instrument setup:
1Press [Recall] > {Recall Setup} to bring up a dialog box
for selecting which instrument setup is to be recalled.
2 Rotate the knob to select the desired trace data.
3Press the [ENTER] hardkey to confirm your
selection.
48
Page 55

Viewing system statistics

Check system status

Pressing [SYS] > {More} > {Status} shows the primary information and current status of the tester, including:
• mode number, current firmware version, product serial number, main board serial number, and option status.
• current calibration mode and status, calibration temperature range, current internal temperature, current main battery capacity, and external power input status.

Running a self test

Pressing [SYS] > {More} > {Self Test} triggers the tester to run a self test on the following items:
•Temperature
• RTC battery voltage
•RF PLL
•LO PLL
Using Functions
Viewing system statistics
49
Page 56
Using Functions

Some Helpful Hints

Some Helpful Hints

Preset the tester

Pressing the green button (the [Preset] hardkey) returns the tester parameters/settings to a known default state.
N9330B Default Settings
Settings Default
measurement mode return loss (vs.freq) start frequency 25 MHz stop frequency 4000 MHz signal standard custom start distance 0 m stop distance 9 m calibration mode SelectedSpan calibration status Uncalibrated calibrator indicator mechanic trace resolution 521 point marker(s) all Off single/multiple limit(s) all Off limit beep Off active [FREQ/DIST] FREQ active [HOLD/RUN] RUN single sweep Off IIM (interference immunity) Off velocity factor 0.5 windowing none unit meter language Depends on previous
date format YYYY-MM-DD
language setting
50
Page 57
Using Functions
NOTE
Some Helpful Hints

Adjust amplitude scale

Adjusting the amplitude scale can improve the clarity and view ability of the amplitude of a trace. the following methods are for adjusting the amplitude scale:
•Pressing [Auto Scale] hardkey automatically adjust the amplitude scale to best fit the current trace view.
•Pressing [AMPTD] > {Top}, or [AMPTD] > {Bottom} manually set the amplitude scale.

Using single sweep

Using single sweep allows you to get an isolated return loss signal. In a single sweep, the tester sweeps from the start to the stop frequency/distance for only one time, and holds the trace static.
Key access: [MEAS/View] > {Single}

Enabling interference immunity

Other signals can interfere with Return Loss/SWR measurement. You can minimize the effect of the interference on the measurement by activating interference immunity in the tester.
Key access: [MEAS/VIEW] > {IIM}
Use of interference immunity increases the measurement time. Interference immunity should be used if a known interference signal exists or the Return Loss/SWR measurement displays suspicious characteristics, such as a spike or rapid movements in the noise floor.
51
Page 58
Using Functions
CAU-CAUTION
Some Helpful Hints

Holding a measurement

The user may hold (pause) a measurement.
•Pressing [HOLD/RUN] hardkey pauses (holds) a measurement, and displays the last trace scan.
•Pressing the [HOLD/RUN] hardkey again will restart the measurement.

Quick access to freq/dist setups

Pressing [FREQ/DIST] allows the user to quickly access to frequency or distance setup submenu:
• When the measurement mode is SWR, return loss or cable loss, pressing [FREQ/DIST] hardkey accesses a submenu for the frequency range configuration.
• When the measurement mode is DTF-SWR, or DTF-return loss, pressing [FREQ/DIST] hardkey accesses a submenu for distance and DTF measurement configuration.

Connector Care

Connectors quality is a critical link for a precision measurement. RF connectors are manufactured to extremely precise tolerances and must be used and maintained with care to protect the measurement accuracy and repeatability.
Taking proper care of cables and connectors will protect your tester’s ability to make accurate measurements. Inaccurate measurements often result of improper connections, dirty, or damaged connectors.
Worn, damaged, out of tolerance, or dirty connectors degrade the accuracy and repeatability of RF measurements.
Always take proper electrostatic precautions before touching the center conductor of any connector, or cable that is connected to the tester.
52
Page 59
Using Functions
CAU-CAUTION
Some Helpful Hints
Repeatability
If two identical measurements are made with the tester, the differences in the measured data should be so small that they do not affect the value of the measurement. Connector repeatability may be a cause of poor measurement results. Connector repeatability problems may be affected by:
• Dirty or damaged connectors
• Connections that have been made without using proper torque techniques (this applies when connections to the tester have been removed, then reconnected).
Extending the life of your cables and connectors
• Minimize the bending of test cables. A single sharp bend can damage a RF cable.
• Avoid repeated connection and disconnection of all RF cable connections.
• Do not bend cables at or near the RF connectors. This may affect the RF performance.
• Inspect all connectors for dirt, nicks, and other signs of damage or wear before connection. A bad connector can ruin a good connector instantly.
• Clean dirty connectors. Dirt and foreign matter can cause poor electrical connections, affect RF measurements, and may permanently damage the connectors.
Never exceed the recommended connector torque values when attaching cables or damage may result.
Proper Connector Torque
Connector Torque
cm-kg
Type-N 52 508 45
3.5 mm 9.2 90 8
SMA 5.7 56 5
Torque N-cm
Torque in-lbs
53
Page 60
Using Functions
Some Helpful Hints

Using signal standards

There are 88 signal standards available in the tester. They can be accessed and selected for use in measurement setups.
The following is a list of common signal standards with their uplink, downlink and full span ranges.
Signal Standard Start freq.
(MHz)
Bluetooth US/Europe uplink 2400.0 2484.0 Bluetooth US/Europe downlink 2400.0 2484.0 Bluetooth US/Europe full band 2400.0 2484.0
Bluetooth Japan uplink 2472.0 2497.0 Bluetooth Japan downlink 2472.0 2497.0 Bluetooth Japan full band 2472.0 2497.0
C 450 P uplink 453 464 C 450 P downlink 463 474 C 450 P fullband 453 474
C 450 SA uplink 465 470 C 450 SA downlink 455 460 C 450 SA fullband 455 470
CDMA China 12 uplink 872 915 CDMA China 12 downlink 917 960 CDMA China 12 fullband 872 960
Cellular uplink 824 849 Cellular downlink 869 894
Cellular fullband 824 894 Cellular 700 MHz uplink 776 794 Cellular 700 MHz downlink 746 764
Cellular 700 MHz fullband 746 794 DCS GSM 1800 uplink 1710 1785 DCS GSM 1800 downlink 1805 1880
DCS GSM 1800 fullband 1710 1880 Digital Multimedia Broadcasting (DMB) 2592.5 2692.5 GSM 900 uplink 880 915
GSM 900 downlink 925 960
Stop freq. (MHz)
54
Page 61
Using Functions
Some Helpful Hints
Signal Standard Start freq.
(MHz)
GSM 900 fullband 880 960 IEEE 802 11 FH uplink 2402 2495 IEEE 802 11 FH downlink 2402 2495
IEEE 802 11 FH fullband 2402 2495 IEEE 802 11 DS uplink 2412 2484 IEEE 802 11 DS downlink 2412 2484
IEEE 802 11 DS fullband 2412 2484 IEEE 802 11 B G uplink 2400 2484 IEEE 802 11 B G downlink 2400 2484
IEEE 802 11 B G fullband 2400 2484 IMT 2000 UMTS WCDMA uplink 1920 1980 IMT 2000 UMTS WCDMA downlink 2110 2170
IMT 2000 UMTS WCDMA fullband 1920 2170 ISM 2 4 GHz uplink 2400 2484 ISM 2 4 GHz downlink 2400 2484
ISM 2 4 GHz fullband 2400 2484 JTACS/NTAC Japan ARIB uplink 887 925 JTACS/NTAC Japan ARIB downlink 832 870
JTACS/NTAC Japan ARIB fullband 832 925 NMT 411 uplink 411 420 NMT 411 downlink 421 430
NMT 411 fullband 411 430 NMT 450 uplink 450 460
NMT 450 downlink 460 470 NMT 450 fullband 450 470 NMT 450 20 kHz CDMA2000 uplink 451 484
NMT 450 20 kHz CDMA2000 downlink 461 494 NMT 450 20 kHz CDMA2000 fullband 451 494 NMT 450 25 kHz CDMA2000 uplink 411 458
NMT 450 25 kHz CDMA2000 downlink 421 468 NMT 450 25 kHz CDMA2000 fullband 411 468 NMT 900 MATS E uplink 890 915
NMT 900 MATS E downlink 935 960 NMT 900 MATS E fullband 890 960
Stop freq. (MHz)
55
Page 62
Using Functions
Some Helpful Hints
Signal Standard Start freq.
(MHz)
PCS GSM 1900 uplink 1850 1910 PCS GSM 1900 downlink 1930 1990 PCS GSM 1900 fullband 1850 1990
PCS Korea uplink 1750 1780 PCS Korea downlink 1840 1870 PCS Korea fullband 1750 1870
PDC 800 uplink 898 940 PDC 800 downlink 843 885 PDC 800 fullband 843 940
PDC 1500 uplink 1525 1549 PDC 1500 downlink 1477 1501 PDC 1500 fullband 1477 1549
PHS uplink 1895 1918 PHS downlink 1895 1918 PHS fullband 1895 1918
SMR 800 uplink 806 821 SMR 800 downlink 851 866 SMR 800 fullband 806 866
SMR 1500 uplink 1453 1465 SMR 1500 downlink 1501 1513 SMR 1500 fullband 1453 1513
TACS/ETACS uplink+C22 872 915 TACS ETACS downlink 917 960
TACSETACS fullband 872 960 Tetra uplink 380 430 Tetra downlink 380 430
Stop freq. (MHz)
56
Page 63
Using Functions
Some Helpful Hints

Using cable folder

In DTF measurement, you are required to input the velocity factor and the average cable loss (dB/m). In case that you do not remember the velocity factor and the average cable loss of the cable-under-test, use the cable folder to find the cable type.
When you are making a DTF measurement, pressing {Cable} brings up a dialog box for frequency band selection first. You should select any of the following categories:
• Standard (measure in 1000 MHz)
• Standard (measure in 2000 MHz)
• Standard (measure in 2500 MHz)
•Customized cables
Press [ENTER] to confirm the selection. Then the tester displays a cable list in the frequency band. Each category of cables is listed with cable type velocity factor and average cable loss.
If you wish to create customized cable library, you need to use N9330B Post Analysis Tool to send the customized cable list to the tester.
57
Page 64
Using Functions
NOTE

System Setups

System Setups

Setting screen resolution

Three types of screen resolution choices are available:
• 521
• 261
• 131
The higher resolution you set, the more precise trace you get, but longer sweep time you need to spend.
Key access: [MEAS/VIEW] > {Resolution}

Setting power manager

The tester will turn off the LCD screen and key light after non-use of from 1 minute to 60 minutes, user defined by pressing {Edit}, if power manager is turned on.
The edited length of time will be noneffective if Power Manager is set to “off “ that the tester will turn off the display and lights never after the defined idle time.
Key access: [SYS] > {Power Manager}

Choosing a unit

Two types of distance units are available:
• meter
•foot
Key access: [SYS] > {Units}
The change of unit is displayed at the left lower corner of the screen.
58
Page 65
Using Functions
NOTE
NOTE
System Setups

Setting date and time

Key access: [SYS] > {Clock}
The date and time display on the top line of the screen.

Firmware upgrade

The N9330B provides an easy and fast access for firmware update.
When updating firmware, make sure there will be a constant power supply for at least 10 minutes.
Refer to the following steps to upgrade the firmware:
1 Download firmware upgrade package from website.
2Extract and copy N9330B.bin file to the ROOT
directory of USB memory stick.
3 Insert this USB stick to the USB port on the top panel
of N9330B.
4Press {SYS} > {Upgrade} to upgrade the firmware.
5 Cycle the power to N9330B when notice “Upgrade
Complete. Effective after Restart.”
When upgrade N9330B firmware from some old version, unrec­ognized fonts will be displayed instead of this message. please restart N9330B as well.
6 When unit powered on, press
{SYS} > {More} > {Statues} to check the firmware version.

Key light setting

The key light of the tester can be turned on or be turned off. The choice is based on personal preference or using conditions.
Key access: [SYS] > {More} > {Key Light}
59
Page 66
Using Functions
System Setups

Setting the brightness of key light

The user can adjust the brightness of key light. for different conditions.
Key access: [SYS] > {More} > {More} > {Key Light Edit}

Light Setting

The N9330B offers two brightness adjusting modes. Addressing “Auto” for adjusting brightness automatically by the light sensor, while “Man” for adjusting by manually operation.
Key access:
[SYS] > {More} > {More} > {More} > {Light Set}
60
Page 67
5 Key Reference

Key Reference

61
Page 68
Key Reference

AMPTD

AMPTD
Sets the displayed amplitude range. Use the numerical keypad to input the amplitude value or rotate the knob to increase or decrease the amplitude value. The tester saves the previous modified amplitude range and recalls it when the tester is powered on.
The displayed amplitude range is different when the tester is in different mode:
Return loss
Cable loss
SWR (Standing Wave Ratio)
DTF_Return loss
DTF_SWR
Top
Sets the displayed top amplitude value.
Tester will automatically revise your input by a minimum amplitude increment, if you previously input an improper top amplitude.
Key access: [AMPTD] > {Top}
In this mode, the default unit dB does not display on the screen.
In this mode, the default unit dB does not display on the screen.
In this mode, the default unit dB does not display on the screen.
62
Page 69
Key Reference
CAU-CAUTION

Bottom

Sets the displayed bottom amplitude.
Tester will automatically revise your input by a minimum amplitude increment, if you previously input improper bottom amplitude.
Key access: [AMPTD] > {Bottom}
Damage to the tester may result when the input power is greater than +23 dBm or 50 VDC.
AMPTD
63
Page 70
Key Reference
CAL
CAL
Activates the calibration process for the tester. A calibration process performs open, short and load (OSL) calibrations. You can either follow the instructions from the tester to run a calibration or customize a calibration.
You can select either an electronic calibrator or a T-combo mechanical calibrator to perform an open-short-load calibration.
T-combo calibrator
Electronic calibrator
When a calibration is completed, a “Calibrated” indicator displays on the left top corner the screen.
Refer to "Calibrate the N9330B before making
measurements" on page 25, for more information.
Is a T-shaped mechanical calibrator. The three parts of the T-combo are the open, short and 50 W load. The T-combo calibrator is a more accurate calibration than the electronic calibrator and will improve the measurement range of the tester.
The electronic calibrator enables you to perform a one-connection calibration process. It performs an open, short, and 50 W load calibration consecutively. The electronic calibrator is faster but less accurate than the mechanical calibrator.

Bypass Electronic Calibrator

The electronic calibrator is the default setting for the tester. However, a user may choose to use either the to electronic calibrator, or T-combo mechanical calibrator. When using the T-combo mechanical calibrator, the user will bypass the default electronic calibrator process by pressing [CAL] > {Bypass E Calib}. The tester will then access the manual open-short-load calibration.
Key access: [CAL] > {Bypass E Calib}
64
Page 71
Key Reference
NOTE

User Cal Mode

The user may define a custom calibration process instead of using the tester default, open-short-load combination. Pressing [Cal] > {User Cal Mode} to select the calibration item for your customized calibration.
User defined calibration accepts only the T-combo calibrator. The electronic calibrator is not compatible with any user defined calibration mode. A User Cal requires the user to perform an open-short-load calibration process.
Key access: [CAL] > {User Cal Mode}
CAL
65
Page 72
Key Reference

Distance

Distance
When you are using the DTF (Distance to Fault) mode, pressing [FREQ/DIST] brings up a submenu of setting DTF parameters, instead of frequency range. The tester saves the latest defined DTF parameters and recalls them when you power on the tester next time.

Start

Sets the displayed start distance. You can pre-define the default unit to be either meters or feet by pressing [SYS] > {Units}. The minimum distance increment is 0.1 m or
0.1 foot.
Key access: {Start}

Stop

Sets the displayed stop distance. Stop distance is always greater than the start distance. The maximum stop distance is calculated from the following equation:
(Resolution * 150 * V
In this equation,
• Resolution means the current resolution of the trace.
•V
•V
•F
Key access: {Stop}
) / (F2–F1)
t/Vc
Pressing [MEAS/VIEW] > {Resolution} enables you to select resolution setups: 131, 261 or 521 points.
means the velocity factor of a cable.
t /Vc
means velocity of light.
c
means the stop frequency and F1 refers to the start
2
frequency. The offset between F2 and F1 should always be greater than a tenth of current resolution (Resolution/10). For example, if the current resolution is 521, then F
should be greater than 52.1 MHz.
2–F1
66
Page 73
Key Reference

DTF Config

Lists and edits the DTF parameters.
Key access: {DTF Config}
Pressing {DTF Config} brings up a check list, showing the parameters in a DTF measurement:
Start Distance When you rotate knob to highlight this parameter and press [ENTER], the displayed start distance becomes editable. This is equivalent to pressing [FREQ/DIST] > {Start} in DTF mode.
Stop Distance When you rotate knob to highlight this parameter and press [ENTER], the displayed stop distance becomes editable. This is equivalent to pressing [FREQ/DIST] > {Stop} in DTF mode.
Start Frequency When you rotate knob to highlight this parameter and press [ENTER], the start frequency becomes editable. This is equivalent to pressing [FREQ/DIST] > {Start} when in non-DTF mode. Changing frequency range changes the maximum measured and displayed distance.
Stop Frequency When you rotate knob to highlight this parameter and press [ENTER], the stop frequency becomes editable. This is equivalent to pressing [FREQ/DIST] > {Stop} when in non-DTF mode. Changing frequency range changes the maximum measured and displayed distance. Refer to more information.
Resolution When you rotate knob to highlight this parameter and press [ENTER], the trace resolution becomes editable. This is equivalent to pressing [Meas/View] > {Resolution}. Changing the resolution changes the maximum measured and displayed distance.
"Distance" on page 66, for
Distance
67
Page 74
Key Reference
Distance
Windowing
Cable
Veloci ty Factor
Cable Loss

Cable Loss

Sets the average loss of a transmission cable, with minimum increment is 0.001. Use either dB/m or dB/foot as its unit.
Key access: {Cable Loss}
When you rotate knob to highlight this parameter and press [ENTER], the windowing menu is presented. This is equivalent to pressing {Window} in DTF mode.
When you rotate knob to highlight this parameter and press [ENTER], a cable library is brought up. This is equivalent to pressing {Cable} in DTF mode.
When you rotate knob to highlight this parameter and press [ENTER], the velocity factor becomes editable. This is equivalent to pressing {Velocity Factor} in DTF mode.
When you rotate knob to highlight this parameter and press [ENTER], the cable loss becomes editable. This is equivalent to pressing {Cable Loss} in DTF mode.

Velocity Factor

Sets the velocity factor of a transmission cable. Velocity factor has no unit. It refers to the rate of a signal’s transmission speed in a particular cable versus its transmission speed in vacuum. Velocity factor ranges from 0.01 to 1.
Key access: {Velocity Factor}
68
Page 75
Key Reference

Cable

The cable library consists of four categories of cables:
Standard cables below 1000 MHz
Standard cables below 2000 MHz
Standard cables below 2500 MHz
Customized cables
Key access: {Cable}
In each category, commonly used cables are listed with cable type, velocity factor, and cable loss. If you need to customize cables, use the N9330B Post Analysis Tool (PC software) to send customized cable list to the tester.

Window

• Adds window to traces. Five options of windowing include:
Rectangular windowing
Hamming windowing
3-sample Blackman windowing
4-sample Blackman windowing
None
Key access: {Window}
Distance
69
Page 76
Key Reference

Enter and ESC/CLR

Enter and ESC/CLR
[ENTER] and [ESC/CLR] are the frequently used keys on the tester.

Enter

• Terminates date entry and enters into the analyzer a numerical value keyed from numeric keypad on the front panel.
• Terminates filename entries.
Key access: [ENTER]

ESC/CLR

• Clears data entries.
• Exits from a functional block.
Key access: [ESC/CLR]
70
Page 77

Frequency

NOTE
Sets the measurement frequency range. The default unit is MHz and the tester provides this unit only. The tester saves the latest defined frequency range and recalls the frequency setup when you power on the tester next time.

Start

Sets the start frequency. The minimum increment is 0.1 MHz. Use numerical keypad to input the frequency and press [ENTER] to confirm your input. Or rotate the knob to increase or decrease the frequency. You can also press {Left} or {Right} to move the cursor to the digit you want to edit on, then rotate the knob to modify the value.
Key access: [FREQ/DIST] > {Start}

Stop

Sets the stop frequency. The minimum increment is 0.1 MHz.
Key access: [FREQ/DIST] > {Stop}

Signal Standard

Includes a frequency range list of several signal standards. When you select a signal standard from the list, the tester automatically sets the start and stop frequency according to the frequency of the signal standard you just selected. Refer to
standards" on page 54
Key access: [FREQ/DIST] > {Signal Standard}
When in DTF mode, pressing [FREQ/DIST] brings up a submenu of setting the distance range, instead of frequency range.
Key Reference
Frequency
"Using signal
for more information.
71
Page 78
Key Reference

Limit

Limit
Sets the pass/fail bar for traces. Two individual single limit lines and single limit line with five multiple limit segments are selectable. The tester also provides a limit beep function to alert a failure of limit judgement. If traces are below a limit or a limit segments, the tester judges the measurement as “passed”, if the traces are above the limit bars, the tester judges the measurement as “failed”.

Limit 1

Pressing [LIMIT] > {Limit 1} brings up a sub menu for editing a single limit applied horizontally across the displayed bandwid th.
The amplitude range of a single limit varies according to the current measurement mode. Refer to
"AMPTD" on page 62 for more information.
Key access: [LIMIT] > {Limit 1}

Limit 2

Pressing [Limit] > {Limit 2} brings up a sub menu for editing a single limit that applies a second horizontal limit line across the displayed bandwidth.
The amplitude range of a single limit varies according to the current measurement mode. Refer to
"AMPTD" on page 62 for more information.
Key access: [LIMIT] > {Limit 2}
72
Page 79
Key Reference
NOTE

Multiple Limit

Pressing {Multi Limit} brings up a submenu for editing multiple limit segments. Up to five individual limit segments may be applied. Each segment requires a defined frequency and amplitude value of the start and end point for the segment.
Limit segments may not overlap with each other and segments must have a horizontal component, that is. they cannot be vertical.
Key access: [LIMIT] > {Multi Limit}

Limit Beep

The tester emits a beeping sound when the trace is above the limit bar to indicate the measurement fails as defined in the limit setups.
Key access: [LIMIT] > {Limit Beep}
Limit
73
Page 80
Key Reference

Marker

Marker
Provides frequency and amplitude readout for a particular point or multiple points on a trace. The tester provides 6 markers, with 6 different colors. You can turn on or off any one of them, or turn off all of them at them same time.
Marker readout displays at the lower side of the screen, in a format of (frequency, amplitude). The tester always uses MHz as its default frequency unit and dB as its default amplitude unit except for SWR measurement.

Marker 1...6

Provides editable markers, labeled from Marker1 to Marker6. Two methods of placing a marker onto a particular point on a trace:
• Turn on a marker and rotate the knob to move the
• Turn on a marker and press [Edit] to manually set the
Key access: [MARKER] > {Marker 1...6} > {Edit}
Delta
Calculates and reads out the frequency and amplitude offsets of Markers, comparing to Marker1. When using delta function, the tester keeps the marker on the trace as it is, but the readout changes to the frequency and amplitude offsets comparing to the Marker1.
Key access: [MARKER] > {Marker 2...6} > {Delta}
marker to the point.
marker to a specified point.
74
Marker to Peak
Places a marker to the point that has the highest amplitude on the trace.
Key access: [MARKER] > {Marker 1...6} > {Marker to Peak}
Page 81
Marker to Minimum
Places a marker to the point that has the lowest amplitude on the trace.
Key access:
[MARKER] > {Marker
1...6} > {Marker to Minimum}

Marker All off

Turn off all the active markers.
Key access: [MARKER] > {Marker All off}
Key Reference
Marker
75
Page 82
Key Reference

Meas/View

Meas/View
Five functional blocks are available under [MEAS/VIEW]:
• Setting the trace resolution
• Enabling the single sweep
• Performing mathematical operation on two traces
• Performing trace overlap comparison
• Enabling interference immunity

Resolution

Three types of trace resolution are selectable:
• 131 points
• 261 points
• 521 points
Key access: [MEAS/VIEW] > {Resolution}

Single

Toggles the single sweep between on and off. When single sweep is off, the tester sweeps continuously.
Key access: [MEAS/VIEW] > {Single}

IIM (Interference Immunity)

Toggles the interference immunity between on and off. When IIM is on, the sweep speed becomes slower than normal.
Key access: [MEAS/VIEW] > {IIM}
76
Page 83
Key Reference

Trace Operation

Mathematically compares current trace with a pre-stored trace. Trace operation requires both measurements be made using the same parameters, mode, resolution, and frequency/distance range.
Pressing {Trace Operation} brings up a check box for you to choose a trace operation mode:
•Off The tester turns off trace operation function.
•Trace – Memory Current trace subtracts pre-stored trace.
• Trace + Memory Current trace adds pre-stored trace.
Key access: [MEAS/VIEW] > {Trace Operation}

Trace Overlap

Overlaps currently displayed measurement trace with a previously-stored trace, using a different color. Trace Overlap requires both measurements be made using the same parameters, mode, resolution, and frequency/distance range.
Key access: [MEAS/VIEW] > {Trace Overlap}
Meas/View
77
Page 84
Key Reference

Mode

Mode
Accesses the check box menu for selecting the measurement mode from the following five options:
• Standing wave ratio (SWR)
• Return loss
•Cable loss
• Distance-to-Fault (DTF)
• Power Meter
Key access: [MODE]

Power Meter

Use the arrow keys to highlight the mode.
Select this mode, and press [ENTER] to make a measurement using the power meter function. When selected, the Power Meter softkey menu will be displayed and many of the instrument hardkeys will be disabled as they are not necessary for this mode.
Key access: [MODE] > {Power Meter}
Standing wave ratio
Return loss
78
Zero
Access the associated Zeroing submenu.
Zero
Make a Zero operation of the power sensor.
The U2000 Series USB Power Sensors have their sen­sor calibration tables stored in non-volatile memory which means that the reference calibration factor is automatically loaded upon power-up.
Zero Type
Page 85
Key Reference
Toggles the zeroing type of the Power Sensor between internal and external. When this type is set to EXT, pressing Zeroing will bring up a message which indicates the external signal need be pulled out. This setting is not affected by the preset in Power Meter menu.
Key access: [MODE] > {Power Meter} > {Zero}
Mode
Meas Disp
Pressing this key accesses the Meas Disp submenu for defining the display range and resolution while in Power Meter mode.
• Disp Range
Access the associated Disp Range submenu.
Auto Range
Toggles the auto range function between On and Off. When the Auto Range is set to On, the Top and Bot­tom value of the power meter’s display range will be set automatically according to the measurement data:
Top = Max. data + 0.5
Bottom = Min. data - 0.5
The default setting is Off.
Top
Set the top value of the power meter’ display range.
Range from (Bottom + 0.01) to 200 dBm.
The default value is 20 dBm.
Bottom
Set the bottom value of the power meter’ display range.
Range from -200 to (Top - 0.01) dBm.
The default value is -70 dBm.
•Resolution
Access the associated Resolution submenu.
79
Page 86
Key Reference
Mode
• Disp Mode
•Chart View
Key access: [MODE] > {Power Meter} > {Meas Disp}
When the measurement suffix is W or %, this parame­ter represents the number of significant digits. When the measurement suffix is dB or dBm, 1 through 4 represents 1, 0.1, 0.01, and 0.001 dB respectively.
The default value is 3.
Change the display mode between Meter and Chart.
The default mode is Meter.
Access the Chart View submenu. This soft key is dis­abled when the measurement is in process or if the display mode is set to Meter.
This submenu is used to view the measurement chart.
Home
View the first 21 points of current measurement chart.
End
View the lastest 21 points of current measurement chart.
Prev
View the previous points of the chart.
Next
View the next points of the chart.
80
Meas Setup
Accesses the Meas Disp submenu which is used to configure the mesurement settings.
•Freq
Sets a frequency between 1 kHz and 90.0 GHz using the numeric keypad, arrow key or knob. If the fre­quency does not correspond directly to a frequency in the sensor calibration table, the power sensor calcu­lates the calibration factor using linear interpolation.
Page 87
Key Reference
•Power Range
Accesses the Power Range submenu.
Auto Range
Enables and disable auto ranging. When auto rang­ing is ON, the power sensor selects the best measur­ing range for the measurement. When auto ranging is OFF, the power sensor remains in the current range and may be manually set by the user.
Range
Selects the Lower or Upper of power ranges when the Auto Range is Off.
• Limits
Accesses the Limits submenu.
Limits
Enables or disables the limit function.
Upper Limit
Sets a value of the upper test limit for the current measurement. The units used are dependent on the current setting of power unit. When the measured power is greater than the value specified, a failure will be reported.
Lower Limit
Sets the lower test limit value for the current mea­surement. The units used are dependent on the cur­rent setting of power unit. When the measured power is less than the value specified, a failure will be reported.
Limit Beep
Enables or disables the limit beep.
• Rel/Offset
Accesses the Rel/Offset submenu.
Rel
Enables the comparison of a measurement result to a reference value and sets a reference value on the cur­rent measured data.
Mode
81
Page 88
Key Reference
Mode
•Average
•Meas Interval
Key access: [MODE] > {Power Meter} > {Meas Setup}
Offset
Sets a value of the display offset on the current mea­surement, and enable or disable the display offset. The display offset is applied to the measurement sig­nal after any math calculation.
Accesses the Average submenu.
Average
Enables or disables the Average function.
Avg Count
Sets a value of the filter length and to set the average count value to be Auto or Manual.
The power sensor automatically sets the number of readings averaged together to satisfy most power measurements averaging requirements. The number of average readings depends on the resolution and the power level settings the power sensor is currently using.
Step Detect
Enables or disables step detection. In AUTO filter mode, the average of the last four values entered into the filter is compared to the entire filter average. If the difference between the two averages is greater than 12.5%, the digital filter is cleared. The filter will then start storing new measurement data. This fea­ture shortens the filter time required when the input power changes substantially. Note that the result appears to settle faster, although true settling to the final value is unaffected.
Sets the interval time between two measurements. The default value is 100 ms.
82
Page 89
Key Reference
Mode
Continue/Pause
Continues or pauses the current measurement. When the measurement is in process, this softkey is displayed as “Pause”; when the measurement is paused, the softkey is displayed as “Continue”.
Key access: [MODE] > {Power Meter} > {Continue/Pause}
Preset
Sets the power meter softkey menu options to the preset conditions. For example, the upper and lower limits are set to their preset limit values.
Key access: [MODE] > {Power Meter} > {Preset}
83
Page 90
Key Reference

System

System
Personalizes the tester and manages system setups.

Cal Mode

Toggles calibration mode between full span calibration and selected span calibration.
Key access: [SYS] > {Cal Mode}

Power Manager

Pressing [SYS] > {Power Manager (on) } enables the power saving mode. The tester will turn off the LCD screen and key light after non-use of from 1 minute to 60 minutes, user defined by pressing [SYS] > {Power Manager} > {Edit}. Pressing any key on the front panel, except the power switch, brings the tester back to its previous state immediately.
Key access: [SYS] > {Power Manager}

Units

Toggles the default setting unit of distance between meter and foot.
Key access: [SYS] > {Units}

Language

Clock

84
Enables you to select a regional language displayed on the screen. Choices are: English, French, German, Italian, Spanish, Portuguese, Russian, Korean, Japanese, Simplified Chinese and Traditional Chinese.
Key access: [SYS] > {Language}
Accesses the date and time settings. Key access: [SYS] > {Clock}
Page 91
Key Reference

Upgrade

Enables the firmware upgrade process.
1 Copy the upgrade application onto the root level of a
USB memory stick.
2 Insert the USB memory stick into the USB
interface of the tester on its top panel.
3Press [SYS] > {Upgrade} to initiate firmware upgrade.
4 The tester will promotes a message to indicate
firmware upgrading is in process.
5 After upgrading completes, power off the tester then
reboot it.
Key access: [SYS] > {Upgrade}

Status

Shows the systematic information of the tester.
Key access: [SYS] > {Status}

Self Test

Runs a self test on temperature and hardware.
Key access: [SYS] > {Self Test}
System

E Calib Test

Verifies the communication between the electronic calibrator and the tester.
Key access: [SYS] > {E Calib Test}

Intelligent

Toggles the ability to identify the calibration loads (Open/Short/Load) from each other between on and off.
Key access: [SYS] > {Intelligent}
85
Page 92
Key Reference
System

Fast Meas

Toggles the fast sweep between on and off.
Key access: [SYS] > {Fast Meas}

License Option

Brings up a window for inputting a license code for an option.
Key access: [SYS] > {License Opt}

Key Light

The key light of the tester can be turned on or be turned off resting with using conditions.
Key access: [SYS] > {Key Light}

Key Light Edit

The brightness of the key light can be adjusted based on personal preference.
Key access: [SYS] > {Key Light Edit}.

Light Set

The Keysight N9330B provides two brightness adjusting modes. Addressing “Auto” for adjusting brightness automatically by the light sensor, while “Man” for manually operation.
Key access: [SYS] > {Light Set}
86
Page 93

Instrument Messages

6 Instrument Messages
87
Page 94
Instrument Messages

Error Message List

Error Message List
Error Code
-310 System error Failed to write to file.
-311 System error Failed to load system file.
-312 System error Failed to delete file.
-313 System error Failed to read file.
-314 System error Failed to save file.
525 Read Ext. power voltage error Failed to read Ext. power
526 RTC battery low voltage RTC battery Low voltage.
527 Read RTC battery error Failed to read RTC battery.
528 Read main battery error Failed to read main battery.
529 Temperature sensor
530 Read RTC time error Failed to read RTC time.
531 Read Ecal serial number error Failed to read electronic
701 RF PLL error RF PLL unlocked.
702 LO PLL error LO PLL unlocked.
703 RF and LO error RF and LO PLL unlocked.
Error Message Description
voltage.
communication error
Failed to temperature sensor communication.
calibrator serial number.
88
Page 95
Instrument Messages

Warning List

No. Warning Message
1 Open segment error. Overlapped with other segment(s).
Continue in 3 seconds.
2 Open segment error. The start is greater than the stop.
Continue in 3 seconds.
3 Edit segment error. The start is greater than the stop.
Continue in 3 seconds.
4 Edit segment error. Overlapped with other segment(s).
Continue in 3 seconds.
5 Trace operation error. Incompatible traces. Continue in 3
seconds.
6 Trace overlap error. Incompatible traces. Continue in 3
seconds.
7 Edit segment error. Segment cannot be vertical. Continue
in 3 seconds.
8 File name already exists, overwrite it? Press ENTER to
overwrite. Press ESC to abort.
Warning List
9 File name too long. Max. 20 char allowed. Press
ENTER/ESC to abort.
10 Calibration invalid. Frequency changed. Continue in 3
seconds.
11 Calibration will be invalidated. Temperature out of range.
Press ENTER/ESC to abort.
12 Cannot open segment. Segment cannot be vertical.
Continue in 3 seconds.
13 Save file error. File name cannot be null. Press ENTER/ESC
to abort.
89
Page 96
Instrument Messages
Warning List
No. Warning Message
14 Save trace error. Memory is full. Continue in 3 seconds.
15 Cannot modify the value now Continue in 3 seconds.
16 Changing CAL mode invalidates calibration. Press ENTER
to continue Press ESC to abort.
17 Changing frequency invalidates calibration. Press ENTER
to continue Press ESC to abort.
18 Delete this trace? Press ENTER to continue. Press ESC to
abort.
19 Delete all traces? Press ENTER to continue. Press ESC to
abort.
20 Freq range is changed. BW is below min. BW. Continue in
3 seconds.
21 Cannot upgrade. There isn't application in the U-disk.
Press ENTER/ESC to abort.
22 Upgrade complete. Press ENTER/ESC to abort.
90
Page 97
7 Menu Map

Menu Map

91
Page 98
Menu Map

FREQ/DIST

Start
Home
Signal
Standard
Stop
End
Page down
Page up
FREQ/DIST
FREQ/DIST
92
Page 99

AMPTD

AMPTD
Top
Bottom
Menu Map
AMPTD
93
Page 100
Menu Map

MEAS/VIEW

Resolution
IIM
ON OFF
Trace
Overlap
Trace
Operation
Single
ON OFF
Return
Select Trace
Off
On
Delete All
Delete Trace
Bottom
Page Down
Page Up
Top
Trace +
Memory
Trace –
Memory
Off
MEAS/VIEW
MEAS/VIEW
94
Loading...