Notice: This document contains references to
Agilent. Please note that Agilent’s Test and
Measurement business has become Keysight
Technologies. For more information, go to
www.keysight.com.
No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
N9330-90011
Edition
Edition 2, July 2014
Printed in China
Published by:
Keysight Technologies
No 116 Tianfu 4th street
Chiengdu, 610041 China
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Power on the N9330B 16
Making the first measurement 17
South Korea Class A EMC Declaration 22
Contents
3Making Measurements
Selecting a Measurement Mode 24
Calibrate the N9330B before making measurements 25
Calibration Mode 25
Types of Calibrators 26
Measuring Return Loss/SWR/Cable Loss 29
Perform a basic return loss measurement 29
Adjusting the measurement resolution 30
Using Marker/Limit 30
Measuring Distance to Fault (DTF) 31
Perform a DTF measurement 31
Adjusting the measurement resolution 32
Using Marker/Limit 32
Page 4
Measured distance 33
Making an Average Power Measurement 34
Making a Basic Average Power Measurement 35
Figure 3-1. Connection with base station 36
4Using Functions
Using Markers and Pass/Fail Limit Lines 40
Using markers 40
Using Pass/Fail Limit Lines 41
Processing Traces 43
Trace Operation 43
Trace overlap 44
Saving and Recalling a file 45
Save a trace data 45
Save an instrument setup 46
Copy a screen 46
Editing a file name 47
Recall a trace data 47
Recall an instrument setup 48
Viewing system statistics 49
Check system status 49
Running a self test 49
Some Helpful Hints 50
Preset the tester 50
Adjust amplitude scale 51
Using single sweep 51
Enabling interference immunity 51
Quick access to freq/dist setups 52
Connector Care 52
System Setups 58
Setting screen resolution 58
Setting power manager 58
Choosing a unit 58
Setting date and time 59
Firmware upgrade 59
Key light setting 59
Setting the brightness of key light 60
Light Setting 60
Units 84
Language 84
Clock 84
Status 85
Self Test 85
E Calib Test 85
Intelligent 85
License Option 86
Key Light 86
Key Light Edit 86
Light Set 86
6Instrument Messages
Error Message List 88
Warning List 89
7Menu Map
FREQ/DIST 92
AMPTD 93
MEAS/VIEW 94
SYS 95
LIMIT 96
Marker 97
Documentation Conventions:
1. A pair of curly brackets { } indicates a softkey, for example
{Start} refers to the Start softkey.
2. A pair of square brackets [ ] indicates a hardkey,
for example, [ENTER] refers to the ENTER hardkey.
3. “DUT” refers to a device under test.
Page 7
1Overview
Overview
1
Page 8
Overview
Introduction
Introduction
Keysight has designed the N9330B handheld cable and
antenna tester (25 MHz to 4 GHz) to simplify the job of
installing and maintaining cable and antenna systems. In
addition to measuring SWR, Return Loss and Cable
Loss, the N9330B also performs Distance-to-Fault (DTF)
measurements that assist in determining cable defect
locations, It is also useful in identifying maintenance
necessary to maintain peak performance of your cable
and antenna systems.
Basic measurement functionality
Basic measurement functionality of the N9330B
includes:
•SWR
• Return loss
•Cable loss
• DTF (SWR and return loss)
• Power Meter (option PWM)
2
Page 9
Optimized Usabilities
The Keysight N9330B has the following features:
• 6.5-inch TFT colorful LCD screen (640×480 pixels)
enables you to read the scans easily and clearly
outdoors.
• Rechargeable lithium-ion battery supports a
minimum of four hours continuous operation when
fully charged.
• Print-screen button for quick screen capture and easy
file saving onto the memory stick.
• Auto file naming makes editing file names fast and
easy.
• The light sensor adjusts the display brightness
according to the environment to save power.
• USB interface for easy and fast file saving to PC or
memory stick.
• User definable power saving mode.
• Back-lit keys make you find the right keys in darkness
easily.
• Arc-shaped handles and a rugged rubber casing
ensure a comfortable and firm hold.
Overview
Introduction
3
Page 10
Overview
N9330B
Handheld Cable & Antenna Tester
25 MHz - 4.0 GHz
ENTER
1ABC 2DEF 3 GHI
4JKL 5 MNO 6PQR
7
STU8VWX9YZ_
0SAVE
MODE
FREQ/
DIST
AMPTD
MEAS/
VIEW
SYS
HOLD/
RUN
LIMIT MARKER
CAL ESC/CLR
AUTOSCALE
RECALL
F1
F2
F3
F4
F5
F6
F7
PRT SCR
12
9
2
1
16
10
11
5
4
3
6
7
8
14
15
13
Front Panel Overview
Front Panel Overview
CaptionBrief Description
1 Standby switchTurns the tester On or Off
2 Function hardkeysIncluding hardkeys: Mode, FREQ/DIST,
3 Preset hardkeyReturns the tester to a known state
4 Enter hardkeyConfirms a selection or configuration
5 CAL hardkeyEnters a calibration procedure
6 ESC/CLR hardkeyEscapes from a procedure or clears
7 Save and Recall keysFor file operation
8 Auto scale hardkeySets the amplitude borders automatically
9 Print screen hardkeyFor quick screen capture
10 KnobSelects an option or edits a number
11 Brightness hardkeyAdjusts the screen brightness
12 BuzzerBeeps differently on different actions
13 Light sensorAdjusts brightness automatically
14 SoftkeysIndicates current menu functions
15 ScreenDisplays measured traces and status
16 Strap handleFor hand carry
4
AMPTD, MEAS/VIEW, SYS, Hold/Run, Limit
and Marker
characters or numbers
Page 11
Top Panel Overview
Ext. Power
Charging
11-25VDC
RF OUT 50
PC
23dBm 50VDC MAX
55W MAX
8
1
2
3
CaptionBrief Description
1 RF Output Connector
Outputs swept signal to DUT
(N-type)
2 USB interface Type B (Host) Connects to a PC
3 USB interface Type A
Connects to a USB memory stick
(Device)
4 LED indicator Lights (On) when external DC power
is connected to the tester
5 LED indicator (Charging)Lights (On) when the battery is
charging
6External DC power
connector
Connects to external DC power, for
use with either the AC-DC charger-adapter or automotive 12V DC
adapter
7 LAN interfaceReserved for future expansion
8Tilt stand (rear of
instrument)
Allows user to lean instrument back
for better viewing when placed on
flat surface
5
4
Overview
Top Panel Overview
6
7
5
Page 12
Overview
ISM1-A
ICES/NMB-001
Instrument Markings
Instrument Markings
The CE mark shows that the product complies with all
relevant European Legal Di rectives .
The CSA mark is a registered trademark of the Canadian
Standards Association.
All Level 1, 2 or 3 electrical equipment offered for sale in
Australia and New Zealand by Responsible Suppliers must be
marked with the Regulatory Compliance Mark.
This symbol is an Industrial Scientific and Medical Group 1
Class A product (CISPR 11, Clause 4)
The ISM device complies with Canadian Interference-Causing
Equipment Standard-001.
Cet appareil ISM est conforme à la norme NMB-001 du
Canada.
The instruction manual symbol: indicates that the user must
refer to specific instructions in the manual.
The standby symbol is used to mark a position of the
instrument power switch.
Indicates the time period during which no hazardous or toxic
substance elements are expected to leak or deteriorate
during normal use. Forty years is the expected useful life of
the product.
Korea Certification indicates this equipment is Class A
suitable for professional use and is for use in electromagnetic
environments outside of the home.
The symbol indicates this product complies with the WEEE
Directive (2002/96/EC) marking requirements and you must
not discard this equipment in domestic household waste.
Do not dispose in domestic household waste. To return
unwanted products, contact your local Keysight office.
6
Page 13
2Getting Started
Getting Started
7
Page 14
Getting Started
Checking the Shipment
Checking the Shipment
Check the shipment and order list when you receive the
shipment.
• Inspect the shipping container for damages.
Signs of damage may include a dented or torn
shipping container or cushioning material that
indicates signs of unusual stress or compacting.
• Carefully remove the contents from the shipping
container, and verify if the standard accessories and
your ordered options are included in the shipment.
For any question or problems, contact Keysight
Technologies for consultant and service at:
http://www.keysight.com/find/assist
8
Page 15
Safety Considerations
WARNING
WARNING
WARNING
WARNING
WARNING
Keysight has designed and tested the N9330B in
accordance with IEC Publication 61010-1 Safety
Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use, and the tester is supplied in
a safe condition.
Read the following safety notices carefully before you
start to use this tester set to ensure safe operation and
to maintain the product in a safe condition.
Personal injury may result if the tester’s cover is removed. There
are no operator-serviceable parts inside. Always contact
Keysight qualified personnel for service. Disconnect the product
from all voltage sources while it is being opened.
This product is a Safety Class I tester. The main plug should be
inserted in a power socket outlet only if provided with a
protective earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
Electrical shock may result when cleaning the tester with the
power supply connected. Do NOT attempt to clean internally.
Use a clean soft cloth to clean the outside case only.
Getting Started
Safety Considerations
Danger of explosion if the battery is incorrectly replaced.
Replace only with the same or equivalent type recommended.
Do NOT dispose of batteries in a fire. Do NOT place batteries in
the trash. Batteries must be recycled or disposed of properly.
Always use the three-pin AC power cord supplied with this
product. Failure to ensure adequate earth grounding by not
using this cord may cause personal injury and product damage.
9
Page 16
Getting Started
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
Safety Considerations
The VxWorks operating system requires full conformity to USB 1.1
or USB 2.0 standards from a USB disk. Not all the USB disk are
built that way. If you have problems connecting a particular USB
disk, please reboot the analyzer before inserting the next.
Use USB disk in FAT or FAT32 format, with only one partition for
data saving and loading with the tester.
If left the battery in tester unused, a fully charged battery will
discharge itself over time.
Never use a damaged or worn-out adapter or battery.
The charging process will be intermitted if the battery
temperature exceed the safety value. This process will restart
when the temperature declines.
If you are charging the batteries internally, even while the tester is
powered off, the tester may become warm. To avoid overheating,
always disconnect the tester from the AC adapter before storing
the tester into the soft carrying case.
Temperature extremes will affect the ability of the battery to
charge. Allow the battery to attain ambient operating temperature
before use or charging.
Storing a battery in extreme hot or cold places will reduce the
capacity and lifetime of a battery. Battery storage is
recommended at 25
Always power on the N9330B before connecting an electronic
calibrator.
o
C.
The average continuous power input is limited not to exceed +23
dBm, DC voltage to 50 VDC. Instrument damage may result if
these precautions are not followed.
10
Page 17
Getting Started
Safety Considerations
Environmental Requirements
Keysight Technologies has designed this tester for use
under the following conditions:
• Operating temperature:
o
C to 40oC (using AC-DC adapter)
0
o
C to +50oC (using battery)
–10
•Storage temperature: –40
• Battery temperature: 0
• Humidity: < 95% (40
o
C to 45oC
o
C)
o
C to +70oC
Electrical Requirements
This tester allows the use of either a lithium battery pack
(internal), AC-DC adapter shipped with the tester, or
optional Automotive +12VDC adapter for its power
supply.
Electrostatic Discharge (ESD) Precautions
This tester was manufactured in an ESD protected
environment. This is because most of the semiconductor
devices used in this tester are susceptible to damage by
static discharge.
Depending on the magnitude of the charge, device
substrates can be punctured of destroyed by contact or
mere proximity of a static charge. The result can cause
degradation of device performance, early failure, or
immediate destruction. ElectroStatic charges are
generated in numerous ways, such as simple contact,
separation of materials, and normal motions of persons
working with static sensitive devices.
When handling or servicing equipment containing static
sensitive devices, adequate precautions must be taken
to prevent device damage or destruction.
Only those who are thoroughly familiar with industry
accepted techniques for handling static sensitive
devices should attempt to service circuitry with these
devices.
11
Page 18
Getting Started
Power Requirements
Power Requirements
The AC power supplied must meet the following
requirements
Voltage:100 VAC to 240 VAC
Frequency:47 to 63 Hz
Power:Maximum 63 W
The AC/DC power supply charger adapter supplied with
the analyzer is equipped with a three-wire power cord, in
accordance with international safety standards. This
power cord grounds the analyzer cabinet when it is
connected to an appropriate power line outlet. The
power cord appropriate to the original shipping location
is included with the analyzer.
Various AC power cables are available from Keysight that
are unique to specific geographic areas. You can order
additional AC power cords are correct for use in different
areas. The AC Power Cords table provides a lists of the
available AC power cords, the plug configurations, and
identifies the geographic area in which each cable is
typically used.
The detachable power cord is the product disconnecting
device. It disconnects the main AC circuits from the DC
supply before other parts of the product. The front-panel
switch is only a standby switch and do not disconnect
instrument from AC LINE power.
:
12
Page 19
AC Power Cord
230V 15A
250V 16A
Getting Started
Power Requirements
Plug TypeCable Part
Number
8120-1703BS 1363/AOption 900
250V 10A
8120-0696 AS 3112:2000 Option 901
250V 10A
8120-1692IEC 83 C4Option 902
250V 16A
8120-1521 CNS 10917-2
125V 10A
8120-2296SEV 1011Option 906
125V 10A
8120-4600SABS 164-1 Option 917
8120-4754JIS C8303Option 918
125V 15A
8120-5181SI 32Option 919
a
Plug
Description
/NEMA 5-15P
For use in
Country & Region
United Kingdom,
Hong Kong,
Singapore, Malaysia
Australia, New
Zealand
Continental Europe,
Korea, Indonesia,
Italy, Russia
Option 903
Unite States, Canada,
Taiwan, Mexico
Switzerland
South Africa, India
Japan
Israel
8120-8377GB 1002Option 922
China
250V 16A
13
Page 20
Getting Started
CAU-CAUTION
Working with Batteries
Working with Batteries
Full charge the battery before first using theanalyzer.
Installing a Battery
StepNotes
1 Open the battery coverUse a phillips type screwdriver,
2 Insert the batteryObserve correct battery polarity
3 Close the battery coverPush the cover closed, then
loosen the retaining screw, then
pull the battery cover open.
orientation when installing.
re-fasten the cover with the
retaining screw.
Viewing Battery Status
Determine the battery status in either of the following
methods:
• Check the battery icon in the lower right corner of the
front panel screen: it indicates the approximate level
of charge.
• The LED on the battery indicates the remaining
battery power.
•Press [SYS] > {More} > {Status} to check the current
battery voltage.
14
Page 21
Getting Started
CAU-CAUTION
Working with Batteries
Charging a Battery
You may charge the battery both in the tester and in the
external battery charger (option BCG).
Connect the automotive adapter to the IT power outlet of your
automobile (with option 1DC) for battery recharging.
You may recharge the battery in the N9330B while the
tester is operating or when it is turned off.
1 Install the battery in the tester.
2 Plug in the correct AC/DC adapter and switch external
power on.
3 The green LED indicating external power and
charging status lights should turn on, indicating that
the battery is charging.
When the battery is fully charged, the green charging
LED turns off.
During charging and discharging, the battery voltage,
current, and temperature are monitored. If any of the
monitored conditions exceed their safety limits, the
battery will terminate any further charging or
discharging until the error condition is corrected.
The charging time for a fully depleted battery, is
approximately four hours.
15
Page 22
Getting Started
CAU-CAUTION
Using the tilt stand
Your first 10 minutes with the N9330B
Your first 10 minutes with the N9330B
The N9330B is fitted with a TFT screen, which is
viewable under all lighting conditions.
Do not apply excessive RF or DC signals to the tester RF OUTPUT
connector. The maximum damage RF input level is +23 dBm (or
50 VDC signal input).
+
Power on the N9330B
Pressing the standby button switches on the tester.
The tester then boots up and runs an internal self-test
that includes the following items:
• internal temperature
• battery voltage (if battery powered)
• battery capacity (if battery powered)
16
Page 23
Your first 10 minutes with the N9330B
Getting Started
Preparation for Use
Use [SYS] hardkey to set or check the system
settings of the tester.
Setting up the N9330B
Press {Power Manager} to enable or unable the power
saving mode.
Pressing [SYS] > {Power Manager (On)} turns off the LCD
screen and key light after a pre-defined idle time set by
pressing {Power Manager} > {Edit}.
Press [SYS] > {More} > {More} > {Key Light} to turn on or
turn off the key light.
Press [SYS] > {Language} to select a language
displayed on the screen.
Press [SYS] > {Clock} to set the time and date.
Check system status
Pressing [SYS] > {More} > {Status} shows the primary
information and current status of the tester.
Making the first measurement
Four easy steps:
1 Select a measurement mode.
2 Set the desired measurement range, or select a
frequency range from the list of signal standards.
3 Calibrate the tester.
4 Connect the DUT (Device Under Test) to the tester RF
OUT connector being careful not to apply a signal
which exceeds the maximum
signal levels.
17
Page 24
Getting Started
Your first 10 minutes with the N9330B
Step 1. Selecting a measurement mode
Pressing [MODE] brings up a check box for you to choose
a measurement mode from the following five modes:
• SWR (Standing Wave Ratio)
• Return loss vs. Frequency
•Cable loss
•DTF-SWR
• DTF-Return loss
• Power Meter
Example: selecting the return loss measurement
1Press [MODE] to bring up the mode selection check
box
2 Use the rotary control knob to select the Return Loss
vs. Frequency
3Press [ENTER] to confirm your selection.
A measurement mode indicator – FREQ-RL will be
displayed in the top middle part of the screen.
"Selecting a Measurement Mode" on
See
page 24
for more information.
18
Step 2. Setting a measurement range
Pressing [FREQ/DIST] brings up a submenu for setting
the measurement range.
If you previously chose a frequency domain
measurement, the submenu allows you to set the start
and stop frequency, or select a pre-defined frequency
range from the list of signal standards.
If you previously chose a DTF measurement, the
submenu allows you to set the start and stop
displayed distance, and other cable measurement
related parameters.
Page 25
Your first 10 minutes with the N9330B
Getting Started
Example: setting a frequency range from 1
GHz to 2 GHz
1 Press {Start} > [1000] > [ENTER] to set the start
frequency to 1 GHz
2 Press {Stop} > [2000] > [ENTER] to set the stop
frequency to 2 GHz
The start frequency of 1000.0 (in MHz) displays at the
left lower corner of screen, and the stop frequency
2000.0 displays at the right lower corner of the screen.
Step 3. Calibrating the N9330B
The N9330B must be calibrated before making an actual
measurement (SWR/Return loss/Cable loss/DTF) using
the selected frequency span or the full span of the
instrument. The N9330B offers two calibration modes:
• FullSpan Cal
Performs an Open-Short-Load calibration across the
full frequency range. The tester allows you to modify
the frequency range for measurement after the
FullSpan Cal is performed.
• SelectedSpan Cal
SelectedSpan
mode. Performs an Open-Short-Load calibration
across a user-defined frequency range. The tester
requires re-calibration if the Selected Frequency Span
is changed.
Key access: [SYS] > {Cal Mode}
Example: Performing a calibration
Assuming the Selected Frequency Span is from 1 to 2
GHz for return loss measurement and the T-combo
calibrator tool is being used.
1 Press [CAL] to bring up a dialog box that will guide
you through the calibration. The normal calibration
order is open, short, and load.
2 Fasten the OPEN end of the T-combo to the RF OUT
connector on the top panel of the tester, then press
[ENTER] to initiate the open calibration.
calibrationisthedefault calibration
19
Page 26
Getting Started
Your first 10 minutes with the N9330B
3 Fasten the SHORT end of the T-combo to the RF OUT
connector, then press [ENTER] to initiate the short
calibration.
4 Fasten the LOAD end of the T-combo to the RF OUT
connector, then press [ENTER] to initiate the load
calibration.
To abort a calibration, press the [ESC/CLR] button.
For more details of calibration, please refer to
"Calibrate the N9330B before making
measurements" on page 25
Step 4. Connecting DUT to RF Output port
Example: Measuring return loss
After you have completed the calibration procedure
above. Refer to the following instructions to make a
return loss measurement.
• Connect the cable under test to the RF OUT
connector on the top panel of the tester.
• The tester immediately performs the measurement
and displays a trace on the screen.
If further data analysis is necessary, refer to:
"Using Markers and Pass/Fail Limit
Lines" on page 40
Traces" on page 43
.
, "Processing
.
20
Page 27
Your first 10 minutes with the N9330B
Getting Started
Contact Keysight Technologies
Keysight maintains sales and service offices around the
world to provide you with complete support for your
handheld cable and antenna tester. In any
correspondence or telephone conversations, refer to
your handheld cable and antenna tester by its product
number and the full serial number.
Go to http://www.keysight.com/find/assist for help with:
• product selection, configuration, and purchases.
• technical application assistance, and consulting.
• rental options, and refurbished equipment.
• repair, calibration, education and training.
If you do not have access to the internet, call the
appropriate number shown below. Or contact local
Keysight Technologies Sales and Service Office.
United States(tel) 800 829 4444
Canada(tel) 877 894 4414
Latin America(tel) +1 (305) 269 7500
China(tel) 800 810 0189
Korea(tel) 080 769 0800
Japan(tel) +81 426 56 7832
Taiwan(tel) 0800 047 866
Europe(tel) +31 20 547 2111
Australia(tel) 1 800 629 485
Other Asia Pacific
Countries
(fax) 800 829 4433
(fax) 800 746 4866
(fax) 800 820 2816
(fax) 080 769 0900
(fax) +81 426 56 7840
(fax) 0800 286 331
(fax) +61 (3) 9210 5947
(tel) +65 6375 8100
(fax) +65 6755 0042
Email: tm_ap@keysight.com
21
Page 28
Getting Started
South Korea Class A EMC Declaration
South Korea Class A EMC Declaration
This equipment is Class A suitable for professional use
and is for use in electromagnetic environments outside
of the home.
22
Page 29
Making Measurements
3Making Measurements
23
Page 30
Making Measurements
Selecting “Return Loss”
measurement
Power Meter
Selecting a Measurement Mode
Selecting a Measurement Mode
Assuming you wish to make a return loss measurement,
you need to:
•Press the [MODE] hardkey to call up the mode
selection check box.
• Use the front-panel knob to select a
measurement type from the Mode Select list.
•Press the [ENTER] hardkey to confirm your
selection.
The icon indicates the following frequency domain
measurement:
•SWR
• Return loss
•Cable loss
The icon indicates the following Distance to Fault
(DTF) measurement:
•SWR
• Return loss
• Power Meter
24
Page 31
Calibrate the N9330B before making measurements
Making Measurements
Calibrate the N9330B before making
measurements
Before making a measurement (SWR/Return loss/Cable
loss/DTF) in a specified frequency range or the full span,
you must calibrate your N9330B.
It is important to keep the calibration frequency range as
close as possible to the actual sweep frequencies
necessary for the measurements. Calibrating over a
larger frequency range than the actual measurement
range will reduce the accuracy of the test results. For
example, using a 1 GHz calibration range when the
measurement range is only few kHz.
Performing one calibration is acceptable provided the
calibration frequency range selected incorporates all
intended measurement frequencies. If the frequency
span is changed, a new calibration must be performed
for that measurement. A new calibration must be
performed if any components used in the calibration
setup are changed. For example, a short coaxial cable
used to connect the tester to the DUT.
Calibration Mode
The N9330B offers two calibration modes:
• FullSpan Cal
Runs an open-short-load calibration across the full
frequency span (25 MHz to 4 GHz).
A FullSpan Cal ignores the frequency changes during
measurements.
• SelectedSpan Cal
Performs an open-short-load calibration across a
user-defined frequency span. A SelectedSpan Cal
requires a re-calibration of the tester, after changing
the measurement frequency range. Normally, the
selected span should be set as close to the actual
swept frequency range.
Key access: [SYS] > {Cal Mode}
25
Page 32
Making Measurements
Calibrate the N9330B before making measurements
Types of Calibrators
Two types of calibrator tools are available from Keysight
for use with the tester:
• T-combo calibrator
This is a T-shaped mechanical calibrator. It consists of
a calibrated open, short and 50 load. Using the
T-combo calibrator will result in a more accurate
calibration than when using the an Electronic
calibrator.
• Electronic calibrator
The Electronic calibrator enables you to perform a
one-step calibration process. It automatically
performs an open, short, and 50W load calibration.
The Electronic calibrator is a faster but less accurate
than when using a T-combo mechanical calibrator.
26
Page 33
Calibrate the N9330B before making measurements
Making Measurements
Calibrating procedure introduction
Choose either the T-combo or the Electronic calibrator
to perform a calibration process. Different combinations
of calibration mode and calibrator may benefit the
measurement accuracy differently. (Assuming the
measurement resolution remains unchanged):
•Using SelectedSpan calibration mode and the T-combo
calibrator provides a more accurate calibration.
• Selecting FullSpan calibration mode and electronic
calibrator is the simplest, most convenient calibration.
The tester default setting is the SelectedSpan calibration
mode. Keysight recommends using the SelectedSpan
calibration mode for best results.
It is important to keep the calibration frequency range as
close as possible to the actual sweep frequencies
necessary for the measurements. Calibrating over a
larger frequency span than the actual measurement
range will reduce the accuracy. For example, using a 1
GHz calibration range when the measurement range is
only few kHz.
Calibrate with an electronic calibrator
The tester needs about 35 sec. to record the data of an
Electronic calibrator the first time a new Electronic
calibrator is used with a specific tester.
After the initial use of an Electronic calibrator, the tester
requires approximately 12 sec. to perform the automated
open-short-load calibration.
The Electronic calibrator performs an automatic,
open-short-load consecutively:
1Press [CAL] to bring up a dialog box for calibration
guidance.
2 Connect the Electronic calibrator to the RF OUT
connector on the top panel of the tester. Connect the
USB cable between the electronic calibrator and the
tester USB port. The correct USB cable is supplied
with an Electronic calibrator.
27
Page 34
Making Measurements
NOTE
Calibrate the N9330B before making measurements
3Press [ENTER] to start calibration.
Never connect an electronic calibrator to the tester until after
powering on the N9330B.
Calibrate with a T-combo calibrator
Using a T-combo calibrator is a three step process
requiring the user to manually change the
open-short-load components. The default order of the
calibration process is open-short-load.
1Press [CAL] to bring up the dialog box for calibration
guidance.
2 Connect the “Open” component of the T-combo
calibrator to the RF OUT connector of the tester.
Press [ENTER] to start the open calibration.
3 Connect the “Short” component of the T-combo
calibrator to the RF OUT connector of the tester.
Press [ENTER] to start the short calibration.
4 Connect the “Load” component of the T-combo
calibrator to the RF OUT connector of the tester.
Press [ENTER] to start the load calibration.
A “Calibrated” indicator will be displayed on the upper
left corner of the screen when the calibration is
completed.
Using User Cal Mode A User Cal Mode is an optional
selection when using a mechanical calibrator. The User
Cal Mode allows the user to choose one cal standard
from the open-short-load to perform a customized
calibration.
For example, pressing [CAL] > {User Cal Mode} > {Open} initiates a calibration
using the Open calibration standard only.
28
Page 35
Measuring Return Loss/SWR/Cable Loss
NOTE
Making Measurements
Measuring Return Loss/SWR/Cable Loss
Return loss is the measurement of the signal reflection
characteristics of a cable and antenna system. A return
loss measurement is useful in detecting problems in the
antenna feedline system and the antenna itself. A
portion of the incident power is reflected back to the
source from transmission line faults as well as from the
antenna itself. The ratio of the reflected voltage to the
incident voltage is called the reflection coefficient. The
reflection coefficient is a complex number, meaning it
has both magnitude and phase information. In
S-parameter terms, return loss is referred to as an S11
measurement.
SWR and cable loss measurement derive from return
loss measurement.
Test signals can cause interference. When testing cables attached
to an antenna, the test signal is radiated from the antenna. Verify
that the signal used for the test CANNOT cause interference to
other antennas.
Perform a basic return loss measurement
When measuring return loss, SWR, or cable loss with
your N9330B, follow the same operation procedure as
shown below.
To make a return loss (versus. frequency) measurement
over a frequency range of 50 to 400 MHz, perform the
following steps:
1Press the [MODE] hardkey to call up a measurement
mode menu. Rotate the front-panel knob to select a
measurement mode (for this example, Return Loss).
Press the [ENTER] hardkey to confirm your selection.
For more detailed information, refer to "Selecting a
Measurement Mode" on page 24.
2Press the {Start} softkey to call up the start frequency
menu. Press {50} > [ENTER] to set the start frequency
to 50 MHz.
29
Page 36
Making Measurements
Measuring Return Loss/SWR/Cable Loss
3Press the {Stop} softkey to call up the stop frequency
menu. Press {400} > [ENTER] to set the stop frequency
to 400 MHz.
4Press the [CAL] hardkey to call up the calibration
process menu. Follow the instructions for performing
the calibration. For more information about
calibration, refer to "Calibrate the N9330B before
making measurements" on page 25.
A “Calibrated” indicator will be displayed on the upper
left corner of the screen when the calibration is
completed.
5 Remove the calibrator and connect the DUT to the RF
OUT connector of the tester. The tester will then
measure the return loss of the DUT.
Adjusting the measurement resolution
The tester has three resolution settings, 521, 261, or 131
points. Press the [MEAS/VIEW] hardkey, then the
{Resolution} softkey to choose the resolution setting
desired, then press the [ENTER] hardkey to enter the
selection.
The higher the resolution, the more accurate the
measurement results. However, a higher resolution
setting will require a longer sweep time.
Using Marker/Limit
For further trace analysis, refer to:
"Using markers" on page 40.
"Using Pass/Fail Limit Lines" on page 41
"Trace Operation" on page 43.
For performing an isolated return loss measurement,
refer to:
"Using single sweep" on page 51.
For minimizing the interference, refer to:
"Enabling interference immunity" on page 51.
30
Page 37
Making Measurements
Measuring Distance to Fault (DTF)
Measuring Distance to Fault (DTF)
For this measurement, an internally generated signal
emits from the RF OUT port of the tester and is
transmitted to the cable and antenna system (DUT). The
tester receives and processes the signal power reflected
from faults and imperfections.
In performing a DTF measurement, the tester uses
frequency domain reflectometry. The transmitted and
reflected signals contain information about the distance
to the faults. This information is used to determine the
physical distance to the faults. The tester displays the
physical distance to the probable faults, corrected for
cable loss and the velocity propagation factor of the
cable.
Perform a DTF measurement
To make a DTF-return loss measurement over a
frequency range of 50 to 400 MHz, perform the following
steps:
1Press [FREQ/DIST] > {Start} >50 > [ENTER] to set the
start frequency to 50 MHz.
2Press {Stop} > 400 > [ENTER] to set the stop frequency
to 400 MHz.
3Press the [MODE] hardkey to select a measurement
mode. Rotate the front-panel knob to highlight and
select the DTF-Return loss, and press [ENTER] to
confirm your selection.
4Press [CAL] to bring up the calibration process
guidance. Follow the instructions to perform a
calibration. For more information about calibration,
refer to "Calibrate the N9330B before making
measurements" on page 25.
5 Remove the calibrator and connect the cable under
test (DUT) to the RF OUT connector on the top panel
of the tester. The DTF measurement starts and the
results are displayed on the screen. A peak in the
horizontal trace indicates a defective location.
31
Page 38
Making Measurements
NOTE
Measuring Distance to Fault (DTF)
For a more precise measurement, enter the cable loss
and velocity factor information.
The measured distance and the displayed distance can be
different.
The measured distance refers to the distance the tester has
measured according to the frequency ranges you specified. You
can check the measured distance by pressing
{DTF Config}
The displayed distance refers to that part of the entire measured
distance chosen. The displayed distance may be manually set by
pressing [FREQ/DIST] > {Start}, and {Stop}.
Adjusting the measurement resolution
The tester has three resolution settings, 521, 261, or 131
points. Press the [MEAS/VIEW] hardkey, then the
{Resolution} softkey to choose the resolution setting
desired, then press the [ENTER] hardkey to enter the
selection.
The higher the resolution, the more accurate the
measurement results. However, a higher resolution
setting will require a longer sweep time.
Using Marker/Limit
For further trace analysis, refer to:
"Using markers" on page 40.
"Using Pass/Fail Limit Lines" on page 41.
"Trace Operation" on page 43.
For getting an isolated return loss signal, refer to:
"Using single sweep" on page 51.
For minimizing the interference, refer to:
"Enabling interference immunity" on page 51.
32
Page 39
Making Measurements
Measuring Distance to Fault (DTF)
Measured distance
It is not always understood how frequency range affects
measured distance and resolution, and may appears
counter-intuitive. This section is intended to help
simplify understanding the DTF measurement.
In the following equations:
Resolution refers to the number of points that are
measured and used in the displayed trace. Pressing
[MEAS/VIEW] > {Resolution} enables you to change the
resolution setting.
) refers to the velocity factor of a cable, its
•(V
t/Vc
transmission rate relative to light. The velocity factor
may be set by pressing {Velo Factor}.
• F2 and F1 refer to the stop and start frequency (MHz).
The measured distance (in meters) of a DTF
measurement is determined by the following equation:
Distance= (Resolution x 150 x V
= 3 x 10
8
V
c
From this equation, you can adjust the measurement:
• To Increase the measured distance:
• Increase the number of points,
or
• Reduce the frequency span.
• To decrease the measured distance:
• Reduce the number of points,
or
• Increase the frequency span.
) / (F2–F1)
t/Vc
33
Page 40
Making Measurements
CAU-CAUTION
Making an Average Power Measurement
Making an Average Power Measurement
Average power measurements provide a key metric in
transmitter performance.
Base station transmit power must be set accurately to
achieve optimal coverage in wireless networks. If the
transmit power is set too high due to inaccurate power
measurements, undesired interference can occur. If the
transmit power is set too low, coverage gaps or holes
may occur. Either case may affect system capacity and
may translate into decreased revenue for service
providers.
Average power can be measured for the channel of
interest while the base station is active. All other
channels should be inactive. Average power is a
broadband measurement. If other signals are present the
analyzer will also measure their power contributions.
The maximum power for the RF IN port and the RF OUT port
of the analyzer is +20 dBm. The maximum power for the
Power Sensor port is +24 dBm. When directly coupled to a
base station, the test set can be damaged by excessive power
applied to any of these three ports.
To prevent damage in most situations when directly coupling
an analyzer to a base station, use a high power attenuator
between the analyzer and the BTS.
The analyzer supports the U2000 Series USBpower
sensors.
The U2000 Series USB power sensors do not need
manual calibration and zero routines performed.
Calibration and zeroing are performed without removing
the power sensor from the source, through internal
zeroing. With internal zeroing of U2000 Series USB
power sensors, there is no need to disconnect the
sensor or power-off the DUT. The U2000 Series do not
require 50 MHz reference signal calibration, allowing the
factory calibration to ensure measurement accuracy. For
best accuracy, users are recommended to perform
34
Page 41
Making an Average Power Measurement
NOTE
NOTE
external zeroing for input signals below -30 dBm for
best accuracy.
Making Measurements
If you suspect other signals may be present, it is
recommended that you turn off all the other channels and
measure average power only on the signal of interest.
Another option is to measure channel power (which is less
accurate), that filters out all other channels (signals). You can
measure channel power for CDMA using the CDMA Analyzer
or CDMA Over Air tool. For other modulation formats, use
their respective analyzers (that is, GSM, 1xEV-DO, or
W-CDMA) or measure channel power using either the
spectrum analyzer or the Channel Scanner tool.
Connect the power meter as close as possible to the power
amplifier/duplexer output. Do not use a coupled port. Sensors
may not be as accurate at the power levels provided by
coupled ports.
Making a Basic Average Power Measurement
To make an average power measurement, connect the
power sensor and cable, zero and calibrate the meter,
before making a measurement.
Zeroing of the Power Meter will occur automatically:
• Every time the Power Meter function is used.
• When a 5 degree C. change in instrument
temperature occurs.
• Whenever the power sensor is changed.
• Every 24 hours (min.).
• Before measuring low level signals -for example,
10 dB above the lowest specified power the power
sensor is capable of.
Calibrating the Power Meter every time you cycle the
power on and off.
In most situations, you can press {Zero} to complete the
two steps (zero and cal) together.
35
Page 42
Making Measurements
Making an Average Power Measurement
To Make a Basic Average Power
Measurement
You can follow the steps below to make a basic average
power measurement.
1Press [Preset] to perform a factory preset.
2Press [MODE]> {Power Meter} > [ENTER] to turn on
the power meter.
3 Zero and calibrate the meter. Press {Zero} to make a
Zero operation of the power sensor followed by a
calibration operation.
4 Connect the power sensor to the power Ref 50 MHz
port. The analyzer supports the U2000 Series power
sensors.
5 Connect the external attenuation, if required, with the
power sensor used.
6 Connect the analyzer to the base station. Connect the
power sensor to the signal to be measured. A typical
BTS connection is after the output of the power
amplifier and duplexer. The connection requires a 2 ft.
cable and two Type-N barrel connectors, in addition
to the power sensor and external attenuator.
Figure 3-1 Connection with base station
36
Page 43
Making an Average Power Measurement
Making Measurements
Setting Power Meter Resolution
You can choose from four levels of Power Meter
resolution. Higher resolutions provide more accuracy but
slow the measurement speed.
1Press [Preset] to perform a factory preset.
2Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3Press {Meas Disp} > {Resolution}. Select 1,2,3 or 4. The
four options represent different resolution as follow:
• [1] = 1
• [2] = 0.1
• [3] = 0.01
• [4] = 0.001
SettingthePowerMeter’sTop and Bottom
End-Points
Setting the end points (Disk Range) close to expected
measurement value changes the sensitivity of the Power
Meter scale resolution. This is an advantage in viewing
small changes in power. However, this will not affect the
overall range of the sensor.
1Press [Preset] to perform a factory preset.
2Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3Press {Meas Disp} > {Disp Range} to access the
end-points menu.
4Press {Top} and enter the maximum scale value
desired using the numeric keypad. Then press {dBm}
to complete the setup.
5Press {Bottom} and enter the minimum scale value
desired using the numeric keypad. Then press {dBm}
to complete the setup.
37
Page 44
Making Measurements
Making an Average Power Measurement
Setting the Power Meter’s Upper and Lower
Limits
The internal Power Meter can be configured to detect
when a measurement has failed a user predefined upper
and lower limits.
1Press [Preset] to perform a factory preset.
2Press [MODE] > {Power Meter} > [ENTER] to turn on
the power meter.
3Press {Meas Setup} > {Limits} to access the limits
menu.
4Press {Limits} to activate the limits function. Each time
the softkey is pressed, the selected option changes.
5Press {Upper Limits} and enter the high limit value
using the numeric keypad. Then press {dBm} to
complete the setup.
6Press {Lower Limits} and enter the low limit value
using the numeric keypad. Then press {dBm} to
complete the setup.
38
Page 45
4Using Functions
Using Functions
39
Page 46
Using Functions
NOTE
Using Markers and Pass/Fail Limit Lines
Using Markers and Pass/Fail Limit Lines
Markers and limit lines are used for measurement trace
analysis.
Using markers
Markers are used to track the specified points on a trace.
You may apply up to six markers and visually track up to
six points on each trace. Each of the six markers has the
following settings:
•On
Activates a Marker and places a vertical, colored bar,
with the Marker number ranging from M1 to M6 to the
foot of the bar.
•Off
Deactivates a marker and erases the Marker bar.
•Edit
Brings up a dialog box and allows you to set the position of a marker.
Rotating the knob will move the position of a marker.
• Marker to Peak
Places the active marker on the highest indicated
value of the trace.
•Marker to Minimum
Places the active marker on the lowest indicated
value of the trace.
•Delta
Is associated with the reference marker (Marker 1 is
the default reference marker). Therefore, Marker 1
must be applied before creating the delta. Delta
displays the value difference between the reference
marker position and the delta marker position. Only
one delta marker can be associated with a given
reference marker.
40
Page 47
Using Markers and Pass/Fail Limit Lines
Using Functions
General steps for adding a marker
Take adding Marker 1 for example:
StepNote
1 Press [Marker] hardkey to call up
a submenu of markers
2 Press {Mark 1} softkey to enter a
submenu of Marker 1
3 Press {On} softkey to activate
Marker 1
4 Press {Edit} softkey to set the
position of the Marker 1
Marker 1 sets its position
refer to its previous position
Rotating the front-panel
knob also positions the
marker
Using Pass/Fail Limit Lines
Limit lines will alert the user to a measurement that
exceeds the upper limit value of a trace. Limits are set in
dB, and define the location of the limit lines displayed on
the screen.
The tester applies upper limits only. Limit lines are used
as pass/fail indicators:
• If the amplitude of a measured trace is below the limit
line, a green Pass indicator is displayed in the upper
right corner of the screen.
• If the amplitude of a trace is above a limit line, a red
Fail indicator is displayed in the upper right corner of
the screen. A limit fail beep sound can be turned on
for an additional trace fail indicator.
Two types of limit lines
•Single limit
A single limit value is applied over the full frequency
range, 25 MHz to 4 GHz. A maximum of two single
limit lines may be applied.
41
Page 48
Using Functions
Using Markers and Pass/Fail Limit Lines
• Multiple limit segment
After setting a limit segment, the limit line is valid only
for your selected frequency/distance range and
amplitude range. A maximum of five limit segments
may be applied.
Setting up a limit line
Refer to the following example to add a limit line.
Example: adding limit 1
StepNote
1 Press [Limit] hardkey to access a
submenu of limit settings.
2 Press {Multi Limit} softkey to access
the multi limit segments submenu.
3 Press {Limit 1} softkey to access a
submenu of the limit 1.
4Press {On} softkey to activate a limit
line.
4 Press [Edit] softkey to set the posi-
tion of a limit line.
5Press
{Prev segment/Next segment}
to continue setting specifications for
a multi-limit trace.
Use this feature if setting
multiple limit values on a
trace.
choosing the specific segment number.
This will set the limit line
at previously configured
limit value.
To adjust the limit line to
your specifications.
For multi limit traces.
42
Page 49
Processing Traces
NOTE
Trace overlap and trace operation are used in the tester
for post trace processing. Very useful in comparing
measurements to identify differences and faults.
Trace Operation
Trace operation mathematically adds or subtracts the
current measured trace to a previously saved trace. This
will provide information that may indicate when a cable
or antenna system is degrading in performance and
would require maintenance.i
The previously saved trace must have the same setup, with consistent measurement parameters as the current trace. For example, the same mode, resolution, and frequency/distance range.
Example: A saved a trace (data) has been entered into
the local memory of the tester. The trace was measured
under return loss mode in GSM full band (880 MHz to
960 MHz), with 521 data points.
Perform the following steps for a trace operation:
1 Press [MEAS/VIEW] > {Trace Operation} to access the
check box for selecting a math operation mode:
• Trace – Memory
Current active trace mathematically subtracts the
trace from local memory of the tester.
• Trace + Memory
Current active trace mathematically adds the trace
from local memory of the tester.
2 Rotate the knob to select a mode and press [ENTER].
3 The tester brings up a list of existing trace data in the
local memory the tester.
4 Rotate the knob to select a trace that has the same
measurement mode, resolution and frequency range,
and press [ENTER].
The tester displays the mathematically calculated trace
on its screen.
Using Functions
Processing Traces
43
Page 50
Using Functions
NOTE
Processing Traces
Trace overlap
The tester allows a previously saved trace to be recalled
onto the currently displayed screen. The two displayed
traces are identified by different colors.
Trace overlap is a real-time tool for quick comparisons
between the current trace and a previously saved trace.
The tester cannot save the overlaid traces into either
local memory or external USB memory stick.
The previously saved trace must have consistent
measurement parameters with the current trace. For example, the
same mode, resolution, and frequency/distance range.
Example: To compare the current measurement trace
to a saved trace (data) in the local memory of the tester.
They were both measured in the return loss mode, GSM
full band (880 MHz to 960 MHz), resolution 521 points.
Perform the following steps to do a trace overlap
operation:
to display a list of trace data stored in the local
memory of the tester.
2 Use the front-panel knob to select a trace and press
[ENTER].
The tester will display the saved trace, while
continuously making measurement of the DUT (current
measured trace). The two traces are displayed on the
screen simultaneously:
• Trace one is displayed as a static trace and is green
• Trace two is yellow in color and is repeatedly
in color. It is the saved trace results recalled from
the local memory of the tester.
scanned. It is the currently measured trace result.
44
Page 51
Saving and Recalling a file
N9330B allows you to save trace data, instrument
setups, or the displayed screen graphic information into
the local memory of the tester. With a USB memory stick
inserted, you can also save trace data and screen copies
into an external memory devices.
FileFormatMax. #
Trace data*. data200
Instrument setting*.data15
Screen copy*.jpgStored onto USB disk
Trace data includes both the measurement data and the
instrument setup information. Trace data and instrument
setup files are saved with a “.data” file extension.
Save a trace data
Trace data includes the measured data and the current
instrument setups. Save Data does not include limit
lines. Limit lines can be saved using the Save Setup
feature.
•Pressing [Save] > {Save Data} allows you to save the
trace data into local memory ONLY. It also activates
the auto file naming editor and automatically saves
the current trace data with a name that is consecutive
to name of the latest saved file.
This mode does NOT allow you to edit the file name.
For example, if the previously saved file name is
“Chicago_site1”, the current trace data will be
automatically named as “Chicago_site2”.
• Pressing [Save] > {Save Data as} allows you to:
• select a destination where a trace data is saved:
Local or USB.
• edit a user defined file name.
See "Editing a file name" on page 47.
Using Functions
Saving and Recalling a file
(Internal memory)
45
Page 52
Using Functions
NOTE
Saving and Recalling a file
Save an instrument setup
The tester will save instrument setups to the local
memory. The maximum number of instrument setups
that can be saved is 15.
Pressing [Save] > {Save Setup} saves the current
instrument settings to the tester’s local memory. The
tester will record the following instrument settings:
•calibration mode
records FullSpan or SelectedSpan calibration mode.
•measurement mode
records measurement modes with their ancroyn: RL,
SWR, CL, DTF-RL, or DTF-SWR.
• parameters
records frequency/distance range, and date and time.
Limit lines, if activated, will be saved using Save
Setup.
Copy a screen
The Keysight N9330B offers two methods to copy a
screen:
•Press [PRT SCR] to get a quick screen capture and
save the file onto an external USB memory stick ONLY.
Pressing the button also activates the auto file naming
editor and automatically saves the screen shot with a
name that is auto-sequential to the number of the
latest saved file, from 1 to 65533.
•Pressing [Save] > {Copy Screen} allows you to save a
screen shot into local memory or external USB
memory stick that will be automatically assigned with
a default file name, from 1 to 10.
Insert a memory stick into the USB connector before copying
screen.
46
Page 53
Using Functions
Saving and Recalling a file
Both of the methods allow you to save the currently
displayed screen in JPEG format. Each screen copied is
an 8-bit colored, 640×480 resolution graph. Copy
screen will include any limit lines, if shown on screen.
However, a screen shot is not editable.
Editing a file name
A valid file name may consist of capitalized letters (A to
Z), decimal numbers (0 to 9), and symbols (dot and
underscore).
Pressing [Save] > {Save Data as} accesses a dialog box
with the last file name used. Press [ESC/CLR] to erase
the current file name, before editing or creating a new
file name.
Input a new file name Use the numerical keypad to
input characters, numbers, and symbols.
• A single press on a numerical button inputs a decimal
number.
• Consecutive press on a numerical button selects a
letter or symbol.
Recall a trace data
To recall a trace data perform the following steps:
1Pressing [Recall] > {Recall Data} accesses a dialog box
for selecting the trace data to be recalled.
2 Rotate the knob to select a trace data.
3Press [ENTER] hardkey to confirm the selection.
Trace data can be deleted from the trace data list in
local memory by:
•Press {Delete Trace} softkey to delete a trace data.
•Press {Delete All} to delete all the trace data from the
list.
47
Page 54
Using Functions
Saving and Recalling a file
Recall an instrument setup
The following steps show the steps to recall a stored
instrument setup:
1Press [Recall] > {Recall Setup} to bring up a dialog box
for selecting which instrument setup is to be recalled.
2 Rotate the knob to select the desired trace data.
3Press the [ENTER] hardkey to confirm your
selection.
48
Page 55
Viewing system statistics
Check system status
Pressing [SYS] > {More} > {Status} shows the primary
information and current status of the tester, including:
• mode number, current firmware version, product
serial number, main board serial number, and option
status.
• current calibration mode and status, calibration
temperature range, current internal temperature,
current main battery capacity, and external power
input status.
Running a self test
Pressing [SYS] > {More} > {Self Test} triggers the tester
to run a self test on the following items:
•Temperature
• RTC battery voltage
•RF PLL
•LO PLL
Using Functions
Viewing system statistics
49
Page 56
Using Functions
Some Helpful Hints
Some Helpful Hints
Preset the tester
Pressing the green button (the [Preset] hardkey) returns
the tester parameters/settings to a known default state.
N9330B Default Settings
SettingsDefault
measurement modereturn loss (vs.freq)
start frequency25 MHz
stop frequency4000 MHz
signal standardcustom
start distance0 m
stop distance9 m
calibration modeSelectedSpan
calibration statusUncalibrated
calibrator indicatormechanic
trace resolution521 point
marker(s)all Off
single/multiple limit(s)all Off
limit beepOff
active [FREQ/DIST]FREQ
active [HOLD/RUN]RUN
single sweepOff
IIM (interference immunity)Off
velocity factor0.5
windowingnone
unitmeter
languageDepends on previous
date formatYYYY-MM-DD
language setting
50
Page 57
Using Functions
NOTE
Some Helpful Hints
Adjust amplitude scale
Adjusting the amplitude scale can improve the clarity
and view ability of the amplitude of a trace. the following
methods are for adjusting the amplitude scale:
•Pressing [Auto Scale] hardkey automatically adjust the
amplitude scale to best fit the current trace view.
•Pressing [AMPTD] > {Top}, or [AMPTD] > {Bottom}
manually set the amplitude scale.
Using single sweep
Using single sweep allows you to get an isolated return
loss signal. In a single sweep, the tester sweeps from the
start to the stop frequency/distance for only one time,
and holds the trace static.
Key access: [MEAS/View] > {Single}
Enabling interference immunity
Other signals can interfere with Return Loss/SWR
measurement. You can minimize the effect of the
interference on the measurement by activating
interference immunity in the tester.
Key access: [MEAS/VIEW] > {IIM}
Use of interference immunity increases the measurement time.
Interference immunity should be used if a known interference
signal exists or the Return Loss/SWR measurement displays
suspicious characteristics, such as a spike or rapid movements in
the noise floor.
51
Page 58
Using Functions
CAU-CAUTION
Some Helpful Hints
Holding a measurement
The user may hold (pause) a measurement.
•Pressing [HOLD/RUN] hardkey pauses (holds) a
measurement, and displays the last trace scan.
•Pressing the [HOLD/RUN] hardkey again will restart
the measurement.
Quick access to freq/dist setups
Pressing [FREQ/DIST] allows the user to quickly access
to frequency or distance setup submenu:
• When the measurement mode is SWR, return loss or
cable loss, pressing [FREQ/DIST] hardkey accesses a
submenu for the frequency range configuration.
• When the measurement mode is DTF-SWR, or
DTF-return loss, pressing [FREQ/DIST] hardkey
accesses a submenu for distance and DTF
measurement configuration.
Connector Care
Connectors quality is a critical link for a precision
measurement. RF connectors are manufactured to
extremely precise tolerances and must be used and
maintained with care to protect the measurement
accuracy and repeatability.
Taking proper care of cables and connectors will protect
your tester’s ability to make accurate measurements.
Inaccurate measurements often result of improper
connections, dirty, or damaged connectors.
Worn, damaged, out of tolerance, or dirty connectors
degrade the accuracy and repeatability of RF
measurements.
Always take proper electrostatic precautions before touching
the center conductor of any connector, or cable that is
connected to the tester.
52
Page 59
Using Functions
CAU-CAUTION
Some Helpful Hints
Repeatability
If two identical measurements are made with the tester,
the differences in the measured data should be so small
that they do not affect the value of the measurement.
Connector repeatability may be a cause of poor
measurement results. Connector repeatability problems
may be affected by:
• Dirty or damaged connectors
• Connections that have been made without using
proper torque techniques (this applies when
connections to the tester have been removed, then
reconnected).
Extending the life of your cables and
connectors
• Minimize the bending of test cables. A single sharp
bend can damage a RF cable.
• Avoid repeated connection and disconnection of all
RF cable connections.
• Do not bend cables at or near the RF connectors. This
may affect the RF performance.
• Inspect all connectors for dirt, nicks, and other signs
of damage or wear before connection. A bad
connector can ruin a good connector instantly.
• Clean dirty connectors. Dirt and foreign matter can
cause poor electrical connections, affect RF
measurements, and may permanently damage the
connectors.
Never exceed the recommended connector torque values
when attaching cables or damage may result.
Proper Connector Torque
ConnectorTorque
cm-kg
Type-N5250845
3.5 mm9.2908
SMA5.7565
Torque
N-cm
Torque
in-lbs
53
Page 60
Using Functions
Some Helpful Hints
Using signal standards
There are 88 signal standards available in the tester.
They can be accessed and selected for use in
measurement setups.
The following is a list of common signal standards with
their uplink, downlink and full span ranges.
Signal StandardStart freq.
(MHz)
Bluetooth US/Europe uplink2400.02484.0
Bluetooth US/Europe downlink2400.02484.0
Bluetooth US/Europe full band2400.02484.0
Bluetooth Japan uplink2472.02497.0
Bluetooth Japan downlink2472.02497.0
Bluetooth Japan full band2472.02497.0
C 450 P uplink453464
C 450 P downlink463474
C 450 P fullband453474
C 450 SA uplink465470
C 450 SA downlink455460
C 450 SA fullband455470
CDMA China 12 uplink872915
CDMA China 12 downlink917960
CDMA China 12 fullband872960
In DTF measurement, you are required to input the
velocity factor and the average cable loss (dB/m). In
case that you do not remember the velocity factor and
the average cable loss of the cable-under-test, use the
cable folder to find the cable type.
When you are making a DTF measurement, pressing
{Cable} brings up a dialog box for frequency band
selection first. You should select any of the following
categories:
• Standard (measure in 1000 MHz)
• Standard (measure in 2000 MHz)
• Standard (measure in 2500 MHz)
•Customized cables
Press [ENTER] to confirm the selection. Then the tester
displays a cable list in the frequency band. Each
category of cables is listed with cable type velocity
factor and average cable loss.
If you wish to create customized cable library, you need
to use N9330B Post Analysis Tool to send the
customized cable list to the tester.
57
Page 64
Using Functions
NOTE
System Setups
System Setups
Setting screen resolution
Three types of screen resolution choices are available:
• 521
• 261
• 131
The higher resolution you set, the more precise trace you
get, but longer sweep time you need to spend.
Key access: [MEAS/VIEW] > {Resolution}
Setting power manager
The tester will turn off the LCD screen and key light after
non-use of from 1 minute to 60 minutes, user defined by
pressing {Edit}, if power manager is turned on.
The edited length of time will be noneffective if Power
Manager is set to “off “ that the tester will turn off the display
and lights never after the defined idle time.
Key access: [SYS] > {Power Manager}
Choosing a unit
Two types of distance units are available:
• meter
•foot
Key access: [SYS] > {Units}
The change of unit is displayed at the left lower corner of
the screen.
58
Page 65
Using Functions
NOTE
NOTE
System Setups
Setting date and time
Key access: [SYS] > {Clock}
The date and time display on the top line of the screen.
Firmware upgrade
The N9330B provides an easy and fast access for
firmware update.
When updating firmware, make sure there will be a constant
power supply for at least 10 minutes.
Refer to the following steps to upgrade the firmware:
1 Download firmware upgrade package from website.
2Extract and copy N9330B.bin file to the ROOT
directory of USB memory stick.
3 Insert this USB stick to the USB port on the top panel
of N9330B.
4Press {SYS} > {Upgrade} to upgrade the firmware.
5 Cycle the power to N9330B when notice “Upgrade
Complete. Effective after Restart.”
When upgrade N9330B firmware from some old version, unrecognized fonts will be displayed instead of this message. please
restart N9330B as well.
6 When unit powered on, press
{SYS} > {More} > {Statues} to check the firmware
version.
Key light setting
The key light of the tester can be turned on or be turned
off. The choice is based on personal preference or using
conditions.
Key access: [SYS] > {More} > {Key Light}
59
Page 66
Using Functions
System Setups
Setting the brightness of key light
The user can adjust the brightness of key light. for
different conditions.
The N9330B offers two brightness adjusting modes.
Addressing “Auto” for adjusting brightness automatically
by the light sensor, while “Man” for adjusting by
manually operation.
Key access:
[SYS] > {More} > {More} > {More} > {Light Set}
60
Page 67
5Key Reference
Key Reference
61
Page 68
Key Reference
AMPTD
AMPTD
Sets the displayed amplitude range. Use the numerical
keypad to input the amplitude value or rotate the knob
to increase or decrease the amplitude value.
The tester saves the previous modified amplitude range
and recalls it when the tester is powered on.
The displayed amplitude range is different when the
tester is in different mode:
• Return loss
• Cable loss
• SWR (Standing Wave Ratio)
• DTF_Return loss
• DTF_SWR
Top
Sets the displayed top amplitude value.
Tester will automatically revise your input by a minimum
amplitude increment, if you previously input an improper
top amplitude.
Key access: [AMPTD] > {Top}
In this mode, the default unit dB does not display on
the screen.
In this mode, the default unit dB does not display on
the screen.
In this mode, the default unit dB does not
display on the screen.
62
Page 69
Key Reference
CAU-CAUTION
Bottom
Sets the displayed bottom amplitude.
Tester will automatically revise your input by a minimum
amplitude increment, if you previously input improper
bottom amplitude.
Key access: [AMPTD] > {Bottom}
Damage to the tester may result when the input power is
greater than +23 dBm or 50 VDC.
AMPTD
63
Page 70
Key Reference
CAL
CAL
Activates the calibration process for the tester. A
calibration process performs open, short and load (OSL)
calibrations. You can either follow the instructions from
the tester to run a calibration or customize a calibration.
You can select either an electronic calibrator or a
T-combo mechanical calibrator to perform an
open-short-load calibration.
• T-combo calibrator
• Electronic calibrator
When a calibration is completed, a “Calibrated”
indicator displays on the left top corner the screen.
Refer to "Calibrate the N9330B before making
measurements" on page 25, for more information.
Is a T-shaped mechanical calibrator. The three parts
of the T-combo are the open, short and 50 W load. The
T-combo calibrator is a more accurate calibration
than the electronic calibrator and will improve the
measurement range of the tester.
The electronic calibrator enables you to perform a
one-connection calibration process. It performs an
open, short, and 50 W load calibration consecutively.
The electronic calibrator is faster but less accurate
than the mechanical calibrator.
Bypass Electronic Calibrator
The electronic calibrator is the default setting for the
tester. However, a user may choose to use either the to
electronic calibrator, or T-combo mechanical calibrator.
When using the T-combo mechanical calibrator, the
user will bypass the default electronic calibrator process
by pressing [CAL] > {Bypass E Calib}. The tester will then
access the manual open-short-load calibration.
Key access: [CAL] > {Bypass E Calib}
64
Page 71
Key Reference
NOTE
User Cal Mode
The user may define a custom calibration process
instead of using the tester default, open-short-load
combination. Pressing [Cal] > {User Cal Mode} to select
the calibration item for your customized calibration.
User defined calibration accepts only the T-combo calibrator. The
electronic calibrator is not compatible with any user defined
calibration mode. A User Cal requires the user to perform an
open-short-load calibration process.
Key access: [CAL] > {User Cal Mode}
CAL
65
Page 72
Key Reference
Distance
Distance
When you are using the DTF (Distance to Fault) mode,
pressing [FREQ/DIST] brings up a submenu of setting
DTF parameters, instead of frequency range. The tester
saves the latest defined DTF parameters and recalls
them when you power on the tester next time.
Start
Sets the displayed start distance. You can pre-define the
default unit to be either meters or feet by pressing [SYS] > {Units}. The minimum distance increment is 0.1 m or
0.1 foot.
Key access: {Start}
Stop
Sets the displayed stop distance. Stop distance is
always greater than the start distance. The maximum
stop distance is calculated from the following equation:
(Resolution * 150 * V
In this equation,
• Resolution means the current resolution of the trace.
•V
•V
•F
Key access: {Stop}
) / (F2–F1)
t/Vc
Pressing [MEAS/VIEW] > {Resolution} enables you to
select resolution setups: 131, 261 or 521 points.
means the velocity factor of a cable.
t /Vc
means velocity of light.
c
means the stop frequency and F1 refers to the start
2
frequency. The offset between F2 and F1 should
always be greater than a tenth of current resolution
(Resolution/10). For example, if the current resolution
is 521, then F
should be greater than 52.1 MHz.
2–F1
66
Page 73
Key Reference
DTF Config
Lists and edits the DTF parameters.
Key access: {DTF Config}
Pressing {DTF Config} brings up a check list, showing the
parameters in a DTF measurement:
• Start Distance
When you rotate knob to highlight this parameter and
press [ENTER], the displayed start distance becomes
editable. This is equivalent to pressing [FREQ/DIST] > {Start} in DTF mode.
• Stop Distance
When you rotate knob to highlight this parameter and
press [ENTER], the displayed stop distance becomes
editable. This is equivalent to pressing [FREQ/DIST] > {Stop} in DTF mode.
• Start Frequency
When you rotate knob to highlight this parameter and
press [ENTER], the start frequency becomes editable.
This is equivalent to pressing [FREQ/DIST] > {Start}
when in non-DTF mode. Changing frequency range
changes the maximum measured and displayed
distance.
• Stop Frequency
When you rotate knob to highlight this parameter and
press [ENTER], the stop frequency becomes editable.
This is equivalent to pressing [FREQ/DIST] > {Stop}
when in non-DTF mode. Changing frequency range
changes the maximum measured and displayed
distance. Refer to
more information.
• Resolution
When you rotate knob to highlight this
parameter and press [ENTER], the trace resolution
becomes editable. This is equivalent to pressing
[Meas/View] > {Resolution}. Changing the
resolution changes the maximum measured and
displayed distance.
"Distance" on page 66, for
Distance
67
Page 74
Key Reference
Distance
• Windowing
• Cable
• Veloci ty Factor
• Cable Loss
Cable Loss
Sets the average loss of a transmission cable, with
minimum increment is 0.001. Use either dB/m or
dB/foot as its unit.
Key access: {Cable Loss}
When you rotate knob to highlight this
parameter and press [ENTER], the windowing menu is
presented. This is equivalent to pressing {Window} in DTF mode.
When you rotate knob to highlight this
parameter and press [ENTER], a cable library is
brought up. This is equivalent to pressing {Cable} in
DTF mode.
When you rotate knob to highlight this
parameter and press [ENTER], the velocity factor
becomes editable. This is equivalent to pressing
{Velocity Factor} in DTF mode.
When you rotate knob to highlight this
parameter and press [ENTER], the cable loss becomes
editable. This is equivalent to pressing {Cable Loss} in
DTF mode.
Velocity Factor
Sets the velocity factor of a transmission cable. Velocity
factor has no unit. It refers to the rate of a signal’s
transmission speed in a particular cable versus its
transmission speed in vacuum. Velocity factor ranges
from 0.01 to 1.
Key access: {Velocity Factor}
68
Page 75
Key Reference
Cable
The cable library consists of four categories of cables:
• Standard cables below 1000 MHz
• Standard cables below 2000 MHz
• Standard cables below 2500 MHz
• Customized cables
Key access: {Cable}
In each category, commonly used cables are listed with
cable type, velocity factor, and cable loss. If you need to
customize cables, use the N9330B Post Analysis Tool
(PC software) to send customized cable list to the tester.
Window
• Adds window to traces. Five options of
windowing include:
• Rectangular windowing
• Hamming windowing
• 3-sample Blackman windowing
• 4-sample Blackman windowing
• None
Key access: {Window}
Distance
69
Page 76
Key Reference
Enter and ESC/CLR
Enter and ESC/CLR
[ENTER] and [ESC/CLR] are the frequently used keys on
the tester.
Enter
• Terminates date entry and enters into the analyzer a
numerical value keyed from numeric keypad on the
front panel.
• Terminates filename entries.
Key access: [ENTER]
ESC/CLR
• Clears data entries.
• Exits from a functional block.
Key access: [ESC/CLR]
70
Page 77
Frequency
NOTE
Sets the measurement frequency range. The default unit
is MHz and the tester provides this unit only. The tester
saves the latest defined frequency range and recalls the
frequency setup when you power on the tester next time.
Start
Sets the start frequency. The minimum increment is 0.1
MHz. Use numerical keypad to input the frequency and
press [ENTER] to confirm your input. Or rotate the knob
to increase or decrease the frequency. You can also
press {Left} or {Right} to move the cursor to the digit you
want to edit on, then rotate the knob to modify the
value.
Key access: [FREQ/DIST] > {Start}
Stop
Sets the stop frequency. The minimum increment is 0.1
MHz.
Key access: [FREQ/DIST] > {Stop}
Signal Standard
Includes a frequency range list of several signal
standards. When you select a signal standard from the
list, the tester automatically sets the start and stop
frequency according to the frequency of the signal
standard you just selected. Refer to
standards" on page 54
Key access: [FREQ/DIST] > {Signal Standard}
When in DTF mode, pressing [FREQ/DIST] brings up a submenu of
setting the distance range, instead of frequency range.
Key Reference
Frequency
"Using signal
for more information.
71
Page 78
Key Reference
Limit
Limit
Sets the pass/fail bar for traces. Two individual single
limit lines and single limit line with five multiple limit
segments are selectable. The tester also provides a limit
beep function to alert a failure of limit judgement. If
traces are below a limit or a limit segments, the tester
judges the measurement as “passed”, if the traces are
above the limit bars, the tester judges the measurement
as “failed”.
Limit 1
Pressing [LIMIT] > {Limit 1} brings up a sub menu for
editing a single limit applied horizontally across the
displayed bandwid th.
The amplitude range of a single limit varies according to
the current measurement mode. Refer to
"AMPTD" on page 62 for more information.
Key access: [LIMIT] > {Limit 1}
Limit 2
Pressing [Limit] > {Limit 2} brings up a sub menu for
editing a single limit that applies a second horizontal
limit line across the displayed bandwidth.
The amplitude range of a single limit varies according to
the current measurement mode. Refer to
"AMPTD" on page 62 for more information.
Key access: [LIMIT] > {Limit 2}
72
Page 79
Key Reference
NOTE
Multiple Limit
Pressing {Multi Limit} brings up a submenu for editing
multiple limit segments. Up to five individual limit
segments may be applied. Each segment requires a
defined frequency and amplitude value of the start and
end point for the segment.
Limit segments may not overlap with each other and segments
must have a horizontal component, that is. they cannot be
vertical.
Key access: [LIMIT] > {Multi Limit}
Limit Beep
The tester emits a beeping sound when the trace is
above the limit bar to indicate the measurement fails as
defined in the limit setups.
Key access: [LIMIT] > {Limit Beep}
Limit
73
Page 80
Key Reference
Marker
Marker
Provides frequency and amplitude readout for a
particular point or multiple points on a trace. The tester
provides 6 markers, with 6 different colors. You can turn
on or off any one of them, or turn off all of them at them
same time.
Marker readout displays at the lower side of the screen,
in a format of (frequency, amplitude). The tester always
uses MHz as its default frequency unit and dB as its
default amplitude unit except for SWR measurement.
Marker 1...6
Provides editable markers, labeled from Marker1 to
Marker6. Two methods of placing a marker onto a
particular point on a trace:
• Turn on a marker and rotate the knob to move the
• Turn on a marker and press [Edit] to manually set the
Key access: [MARKER] > {Marker 1...6} > {Edit}
Delta
Calculates and reads out the frequency and amplitude
offsets of Markers, comparing to Marker1. When using
delta function, the tester keeps the marker on the trace
as it is, but the readout changes to the frequency and
amplitude offsets comparing to the Marker1.
Key access: [MARKER] > {Marker 2...6} > {Delta}
marker to the point.
marker to a specified point.
74
Marker to Peak
Places a marker to the point that has the highest
amplitude on the trace.
Key access: [MARKER] > {Marker 1...6} > {Marker to Peak}
Page 81
Marker to Minimum
Places a marker to the point that has the lowest
amplitude on the trace.
Key access:
[MARKER] > {Marker
1...6} > {Marker to Minimum}
Marker All off
Turn off all the active markers.
Key access:[MARKER] > {Marker All off}
Key Reference
Marker
75
Page 82
Key Reference
Meas/View
Meas/View
Five functional blocks are available under [MEAS/VIEW]:
• Setting the trace resolution
• Enabling the single sweep
• Performing mathematical operation on two traces
• Performing trace overlap comparison
• Enabling interference immunity
Resolution
Three types of trace resolution are selectable:
• 131 points
• 261 points
• 521 points
Key access: [MEAS/VIEW] > {Resolution}
Single
Toggles the single sweep between on and off. When
single sweep is off, the tester sweeps continuously.
Key access: [MEAS/VIEW] > {Single}
IIM (Interference Immunity)
Toggles the interference immunity between on and off.
When IIM is on, the sweep speed becomes slower than
normal.
Key access: [MEAS/VIEW] > {IIM}
76
Page 83
Key Reference
Trace Operation
Mathematically compares current trace with a
pre-stored trace. Trace operation requires both
measurements be made using the same parameters,
mode, resolution, and frequency/distance range.
Pressing {Trace Operation} brings up a check box for you
to choose a trace operation mode:
•Off
The tester turns off trace operation function.
•Trace – Memory
Current trace subtracts pre-stored trace.
• Trace + Memory
Current trace adds pre-stored trace.
Key access: [MEAS/VIEW] > {Trace Operation}
Trace Overlap
Overlaps currently displayed measurement trace with a
previously-stored trace, using a different color. Trace
Overlap requires both measurements be made using the
same parameters, mode, resolution, and
frequency/distance range.
Key access: [MEAS/VIEW] > {Trace Overlap}
Meas/View
77
Page 84
Key Reference
Mode
Mode
Accesses the check box menu for selecting the
measurement mode from the following five options:
• Standing wave ratio (SWR)
• Return loss
•Cable loss
• Distance-to-Fault (DTF)
• Power Meter
Key access: [MODE]
Power Meter
Use the arrow keys to highlight the mode.
Select this mode, and press [ENTER] to make a
measurement using the power meter function. When
selected, the Power Meter softkey menu will be
displayed and many of the instrument hardkeys will be
disabled as they are not necessary for this mode.
Key access: [MODE] > {Power Meter}
• Standing wave ratio
• Return loss
78
Zero
Access the associated Zeroing submenu.
• Zero
Make a Zero operation of the power sensor.
The U2000 Series USB Power Sensors have their sensor calibration tables stored in non-volatile memory
which means that the reference calibration factor is
automatically loaded upon power-up.
• Zero Type
Page 85
Key Reference
Toggles the zeroing type of the Power Sensor
between internal and external. When this type is set
to EXT, pressing Zeroing will bring up a message
which indicates the external signal need be pulled
out. This setting is not affected by the preset in Power
Meter menu.
Key access: [MODE] > {Power Meter} > {Zero}
Mode
Meas Disp
Pressing this key accesses the Meas Disp submenu for
defining the display range and resolution while in Power
Meter mode.
• Disp Range
Access the associated Disp Range submenu.
• Auto Range
Toggles the auto range function between On and Off.
When the Auto Range is set to On, the Top and Bottom value of the power meter’s display range will be
set automatically according to the measurement
data:
Top = Max. data + 0.5
Bottom = Min. data - 0.5
The default setting is Off.
• Top
Set the top value of the power meter’ display range.
Range from (Bottom + 0.01) to 200 dBm.
The default value is 20 dBm.
• Bottom
Set the bottom value of the power meter’ display
range.
Range from -200 to (Top - 0.01) dBm.
The default value is -70 dBm.
•Resolution
Access the associated Resolution submenu.
79
Page 86
Key Reference
Mode
• Disp Mode
•Chart View
Key access: [MODE] > {Power Meter} > {Meas Disp}
When the measurement suffix is W or %, this parameter represents the number of significant digits. When
the measurement suffix is dB or dBm, 1 through 4
represents 1, 0.1, 0.01, and 0.001 dB respectively.
The default value is 3.
Change the display mode between Meter and Chart.
The default mode is Meter.
Access the Chart View submenu. This soft key is disabled when the measurement is in process or if the
display mode is set to Meter.
This submenu is used to view the measurement chart.
• Home
View the first 21 points of current measurement chart.
• End
View the lastest 21 points of current measurement
chart.
• Prev
View the previous points of the chart.
• Next
View the next points of the chart.
80
Meas Setup
Accesses the Meas Disp submenu which is used to
configure the mesurement settings.
•Freq
Sets a frequency between 1 kHz and 90.0 GHz using
the numeric keypad, arrow key or knob. If the frequency does not correspond directly to a frequency in
the sensor calibration table, the power sensor calculates the calibration factor using linear interpolation.
Page 87
Key Reference
•Power Range
Accesses the Power Range submenu.
• Auto Range
Enables and disable auto ranging. When auto ranging is ON, the power sensor selects the best measuring range for the measurement. When auto ranging is
OFF, the power sensor remains in the current range
and may be manually set by the user.
• Range
Selects the Lower or Upper of power ranges when the
Auto Range is Off.
• Limits
Accesses the Limits submenu.
• Limits
Enables or disables the limit function.
• Upper Limit
Sets a value of the upper test limit for the current
measurement. The units used are dependent on the
current setting of power unit. When the measured
power is greater than the value specified, a failure will
be reported.
• Lower Limit
Sets the lower test limit value for the current measurement. The units used are dependent on the current setting of power unit. When the measured power
is less than the value specified, a failure will be
reported.
• Limit Beep
Enables or disables the limit beep.
• Rel/Offset
Accesses the Rel/Offset submenu.
• Rel
Enables the comparison of a measurement result to a
reference value and sets a reference value on the current measured data.
Mode
81
Page 88
Key Reference
Mode
•Average
•Meas Interval
Key access: [MODE] > {Power Meter} > {Meas Setup}
• Offset
Sets a value of the display offset on the current measurement, and enable or disable the display offset.
The display offset is applied to the measurement signal after any math calculation.
Accesses the Average submenu.
• Average
Enables or disables the Average function.
• Avg Count
Sets a value of the filter length and to set the average
count value to be Auto or Manual.
The power sensor automatically sets the number of
readings averaged together to satisfy most power
measurements averaging requirements. The number
of average readings depends on the resolution and
the power level settings the power sensor is currently
using.
• Step Detect
Enables or disables step detection. In AUTO filter
mode, the average of the last four values entered into
the filter is compared to the entire filter average. If
the difference between the two averages is greater
than 12.5%, the digital filter is cleared. The filter will
then start storing new measurement data. This feature shortens the filter time required when the input
power changes substantially. Note that the result
appears to settle faster, although true settling to the
final value is unaffected.
Sets the interval time between two measurements.
The default value is 100 ms.
82
Page 89
Key Reference
Mode
Continue/Pause
Continues or pauses the current measurement. When
the measurement is in process, this softkey is displayed
as “Pause”; when the measurement is paused, the
softkey is displayed as “Continue”.
Sets the power meter softkey menu options to the preset
conditions. For example, the upper and lower limits are
set to their preset limit values.
Key access: [MODE] > {Power Meter} > {Preset}
83
Page 90
Key Reference
System
System
Personalizes the tester and manages system setups.
Cal Mode
Toggles calibration mode between full span calibration
and selected span calibration.
Key access: [SYS] > {Cal Mode}
Power Manager
Pressing [SYS] > {Power Manager (on) } enables the
power saving mode. The tester will turn off the LCD
screen and key light after non-use of from 1 minute to
60 minutes, user defined by pressing [SYS] > {Power Manager} > {Edit}. Pressing any key on the front panel,
except the power switch, brings the tester back to its
previous state immediately.
Key access: [SYS] > {Power Manager}
Units
Toggles the default setting unit of distance between
meter and foot.
Key access: [SYS] > {Units}
Language
Clock
84
Enables you to select a regional language displayed on
the screen. Choices are: English, French, German,
Italian, Spanish, Portuguese, Russian, Korean, Japanese,
Simplified Chinese and Traditional Chinese.
Key access: [SYS] > {Language}
Accesses the date and time settings.
Key access: [SYS] > {Clock}
Page 91
Key Reference
Upgrade
Enables the firmware upgrade process.
1 Copy the upgrade application onto the root level of a
USB memory stick.
2 Insert the USB memory stick into the USB
interface of the tester on its top panel.
3Press [SYS] > {Upgrade} to initiate firmware upgrade.
4 The tester will promotes a message to indicate
firmware upgrading is in process.
5 After upgrading completes, power off the tester then
reboot it.
Key access: [SYS] > {Upgrade}
Status
Shows the systematic information of the tester.
Key access: [SYS] > {Status}
Self Test
Runs a self test on temperature and hardware.
Key access: [SYS] > {Self Test}
System
E Calib Test
Verifies the communication between the electronic
calibrator and the tester.
Key access: [SYS] > {E Calib Test}
Intelligent
Toggles the ability to identify the calibration loads
(Open/Short/Load) from each other between on and off.
Key access: [SYS] > {Intelligent}
85
Page 92
Key Reference
System
Fast Meas
Toggles the fast sweep between on and off.
Key access: [SYS] > {Fast Meas}
License Option
Brings up a window for inputting a license code for an
option.
Key access: [SYS] > {License Opt}
Key Light
The key light of the tester can be turned on or be turned
off resting with using conditions.
Key access: [SYS] > {Key Light}
Key Light Edit
The brightness of the key light can be adjusted based on
personal preference.
Key access: [SYS] > {Key Light Edit}.
Light Set
The Keysight N9330B provides two brightness adjusting
modes. Addressing “Auto” for adjusting brightness
automatically by the light sensor, while “Man” for
manually operation.
Key access: [SYS] > {Light Set}
86
Page 93
Instrument Messages
6Instrument Messages
87
Page 94
Instrument Messages
Error Message List
Error Message List
Error
Code
-310System errorFailed to write to file.
-311System errorFailed to load system file.
-312System errorFailed to delete file.
-313System errorFailed to read file.
-314System errorFailed to save file.
525Read Ext. power voltage error Failed to read Ext. power