Keysight Technologies N5280A User's And Service Manual

K
eysight Technologies
N5280A
Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business Technologies. For more information, go to www.keysight.com.
User’s and Service Guide
Notices
© Keysight Technologies, Inc. 2008-2017
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
N5280-90001
Edition
June 2017
Supersedes: August 2015
Printed in USA
Published by: Keysight Technologies
1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hard ware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional
government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Table of Conte nts
N5280A
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Network Analyzer Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Verifying the Shipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Available Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Test Set Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Equipment Heating and Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Required Conditions for Accuracy Enhanced Measurement. . . . . . . . . . . . . . . . . . . . . 13
Dimensions and Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Maximum Power Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Contents
Front and Rear Panel Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
RF INPUTS (A, B, C and D) (all options) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
IF OUTPUTS (A, B, C and D) (Option 002) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
LO INPUT (Option 002). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Standby Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
IF OUT (A, B, C/R1 and D/R2) (Option 700 and 001) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
LO IN (Option 700 and 001) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
LO AUX (All Options) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Attenuators (A, B, C and D) (Option 001) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Line Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Available Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Controlling the Test Set with N5242A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Controlling the Test Set with N5264A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Controlling the N5280A Option 002 using an Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Controlling the RF Receiver Attenuators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Operational Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Equipment Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Verification Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Typical Plots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Operational Check Procedure (Option 700 and 001) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Preparing the N5242A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Attenuator Element Verification Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Operational Check Procedure (Option 002). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Requirements for Operational Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Keysight N5280A User’s and Service Guide 3
Contents
Required Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Instrument Setup (Source Calibration). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Check IF Output A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Check LO AUX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Safety and Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Safety Earth Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Statement of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Connector Care and Cleaning Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Instrument Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
South Korean Class A EMC Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Acoustic Statement (European Machinery Directive): . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
Electrostatic Discharge Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Keysight Support, Services, and Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
Service and Support Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Contacting Keysight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Shipping Your Product to Keysight for Service or Repair. . . . . . . . . . . . . . . . . . . . . . . . . .54
4 Keysight N5280A User’s and Service Guide
Keysight N5280A
User’s and Service Guide
N5280A
Introduction
The N5280A replaces the 8511A Four Channel Frequency Converter Test Set. This document describes how to use the N5280A Test Set with the Keysight N5242A 2-Port, or 4-Port PNA-X Network Analyzer and N5264A Measurement Receiver.
Figure 1 N5242A 2-Port PNA-X with N5280A
5
-
N5280A Introduction
Figure 2 N5242A 4-Port PNA-X with N5280A
6 Keysight N5280A User’s and Service Guide
N5280A Introduction
Figure 3 N5264A Measurement Receiver with N5280A
Keysight N5280A User’s and Service Guide 7
-
N5280A Description
Description
The Keysight N5280A is a four channel frequency converter Test Set. This Test Set is used with the Keysight N5242A 2-Port or 4-Port PNA-X Network Analyzer, and a N5264A Measurement Receiver. It can be operated with other microwave accessories (couplers, power splitters). The N5280A provides a convenient means of customizing a test configuration for a variety of applications within a frequency range of 10 MHz to 26.5 GHz. In addition to configurations for measuring reflection and transmission parameters of one-port or two-port devices, you can build configurations to characterize antenna parameters, radar cross sections and frequency translation devices.
Figure 4 on page 9 illustrates one possible measurement set-up configuration.
Network Analyzer Requirements
— The N5242A 2-Port or 4-Port PNA-X Network Analyzer (10 MHz to 26.5
GHz) requires Option 020, which adds IF inputs.
— The N5264A Measurement Receiver requires Option 108, which adds LO
Source 26.5 GHz.
More PNA information is available on the following websites:
—Documentation - http://www.keysight.com/find/pna
— Network Analyzer Firmware -
http://na.support.keysight.com/pna/firmware
Verifying the Shipment
To verify the contents shipped with your product, refer to Tab le 1.
Inspect the shipping container. If the container or packing material is damaged, it should be kept until the contents of the shipment have been checked mechanically and electrically. If there is physical damage refer to
“Contacting Keysight” on page 54. Keep the damaged shipping materials (if
any) for inspection by the carrier and an Keysight Technologies representative.
Table 1 Content List
Keysight Part Number
Description Qty
9320-6636 Functional Test Certificate 1
9320-0333 Envelope-Cal Certificate 1
5063-9232 Rack Mount Kit with Handles 1
5063-9226 Front Handle Kit 1
5061-9038 Cable Assembly 5
N5280-90001 User’s and Service Guide 1
8 Keysight N5280A User’s and Service Guide
RF
RF
C
A
LO
Source Antenna
Reference Antenna
DUT
N5264A
Measurement Receiver
N5280A
Frequency Converter
A
C
RF Source
RF Source
Positioner
Positioner
Controller
Amplifier
Control Lines
RF Cable
RF Cable
RF Cable
RF Cable
IF Cables
10 MHz REF
LAN/GPIB
10 MHz REF
LAN/GPIB
N5280A Description
Figure 4 Measurement Setup
Keysight N5280A User’s and Service Guide 9
-
N5280A Description
Available Options
Test Set Options
The N5280A has two available Test Set options: Refer to Figure 29 on page 43, Figure 30 on page 44 and Figure 31 on
page 45.
— Standard (700) - There are no attenuators in the RF Input paths.
— Option 001 - There are four attenuators (0 to 35 dB) in the RF Input paths to
reduce the power levels.
— Option 002 - There are four IF Channels (A, B, C and D) and LO Input on the
front panel.
10 Keysight N5280A User’s and Service Guide
N5280A General Specifications
General Specifications
Specifications for the N5280A Frequency Converter Test Set (10 MHz to 26.5 GHz) are characteristic for the System performance of the PNA and Test Set. Actual performance of the system is based on the customers PNA that is used with the Test Set.
A functional certificate is supplied for the N5280A. The N5280A performance is based on external components such as the calibration kit, network analyzer, external sources, and measurement receiver. There are no internal adjustments in the N5280A, therefore an annual calibration is not required.
Frequency range and connectors are listed in Table 2 for specific ports.
Table 2 Frequency Range and Connectors
Port Frequency Range Connectors
RF Port 0.01 to 26.5 GHz 3.5 mm female
LO Port 0.01 to 26.5 GHz 3.5 mm female
IF Port 7 to 20 MHz (with jumper, LP OUT)
0.007 to 1.5 GHz (without jumper, Max BW)
SMA female
Keysight N5280A User’s and Service Guide 11
-
N5280A General Specifications
Power Requirements
Verify that the required ac power is available before installing the Test Set to the PNA.
— 100/120/220/240 VAC (50/60Hz)
— The instruments can operate with mains supply voltage fluctuations up to ±
10% of the nominal voltage.
— Air conditioning equipment (or other motor–operated equipment) should
not be placed on the same ac line that powers the Test Set and PNA.
Table 3 contains the maximum wattage for all instruments. This table can
be use to determine the electrical and cooling requirements.
Table 3 Maximum Power Requirements
Standard Equipment
Instrument Max Wattage
N5242A 450
N5264A 450
N5280A 350
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall be inserted only into a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited.
12 Keysight N5280A User’s and Service Guide
N5280A General Specifications
Environmental Requirements
The environmental requirements of the Test Set are listed in Table 4. Refer to the N5242A PNA-X and N5264A Measurement Receiver standard documentation for environmental requirements.
Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used.
Table 4 PNA Operating Environment
Operating Environmental Requirements:
Temp erature 0 °C to 40 °C ( 32 ° F to 1 04 ° F)
M e a s u re m e n t C a l i b r a t i o n 2 0 ° C to 2 6 ° C ( 6 8 ° F t o 7 9 ° F )
P er f o r ma n c e V e r i fi c a ti o n T e mp e ra t u re m us t be w i th i n 1 ° C ( 1 .8 ° F) o f t h e t e m pe r a tu r e
at which the measurement calibration was performed.
Pressure Altitude (Operating)
3000 meters (9,842 feet)
Environmental
The N5280A complies with all applicable safety and regulatory requirements for the intended location of use.
— The instrument can safely operate in a relative humidity of 80% for
temperatures to 31 °C, decreasing linearly to 50% relative humidity at 40 °C.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature within the appropriate range. Air conditioning capacity must be consistent with the BTU ratings given in Table 3.
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient temperature of the PNA and Test Set to be maintained within ± 1 °C of the ambient temperature at calibration.
Keysight N5280A User’s and Service Guide 13
-
N5280A General Specifications
Dimensions and Space Requirements
Standard installation of the N5280A and PNA includes configuration and installation on a customer provided lab bench, or table top of adequate size and strength.
Table 5 Instrument Dimensions
Model Weight Height Wid th Depth
N 5 2 4 2 A 37 kg
82 lb (±0.5 lb)
N 5 2 6 4 A 2 2 k g
48 lb (±0.5 lb)
N5280A 9.1 kg
(20 lb)
Maximum Power Levels
Tab le 6 Power Level s
RF Input Power Damage Levels
RF Port +18 dBm
LO Port +5 dBm
Optimum LO Power 0 dBm (±1 dB)
RF Input @ 0.1 dB Typical Compression
A- D Receiver 10 dBm
26.67 cm (10.5 in)
26.67 cm (10.5 in)
8.9 cm (3.5 in)
42.5 cm (16.75 in)
42.5 cm (16.75 in)
42.5 cm (16.75 in)
55.8 cm (21.97 in)
55.8 cm (21.97 in)
48.3 cm (19 in)
Refer to your PNA specifications to determine the maximum input power levels for the PNA access and test ports, or to optimize the power levels in the receivers.
Damage and maximum levels are not necessarily the optimum level.
14 Keysight N5280A User’s and Service Guide
IF Output Level1
Ports A- D 10 dBm
1.
IF Output level is based on the RF Input @
0.1 dB typical compression.
N5280A
RF INPUT Ports
A B C D
A B C D
A B C D LO INPUT
IF OUTPUT Ports
RF INPUT Ports
Front and Rear Panel Features
Front and Rear Panel Features
Figure 5 Front Panel N5280A (Option 700 and 001)
Figure 6 Front Panel N8280A (Option 002)
RF INPUTS (A, B, C and D) (all options)
These input signals (10 MHz to 26.5 GHz) will be downconverted and directed to the IF Outputs.
IF OUTPUTS (A, B, C and D) (Option 002)
These outputs transmit IF signals (10 MHz to 1.5 GHz) derived from the RF Inputs.
LO INPUT (Option 002)
This input signal (10 MHz to 26.5 GHz) is used as the LO to downconvert the RF Inputs and to provide a buffered LO Output.
Standby Switch
The switch is only a Standby switch, not a AC line power switch.The main power cord can be used as the system disconnecting device. It disconnects the mains circuits from the mains supply.
Keysight N5280A User’s and Service Guide 15
-
RF Jumper LO AUX (load)
Line Module
LP OUT
LP IN
Max BW
LO AUX (load)
N5280A Front and Rear Panel Features
Figure 7 Rear Panel (Option 700 and 001)
LO IN
Attenuators (Option 001)
Figure 8 Rear Panel (Option 002)
IF OUT (A, B, C/R1 and D/R2) (Option 700 and 001)
LP OUT from the IF Outputs are connected to the IF Inputs on the PNA-X. The RF Jumper (N5222-20091 was E8356-20072) is connected between LP IN and Max BW. These connectors transmit the IF signal from the Test Set to the N5242A 2-Port or 4-Port PNA-X and N5264A Measurement Receiver.
16 Keysight N5280A User’s and Service Guide
N5280A Front and Rear Panel Features
LO IN (Option 700 and 001)
LO IN is connected to the LO OUT from the EXT TSET DRIVE on the PNA-X.
LO AUX (All Options)
The load (1810-0118) is connected to LO AUX.
Attenuators (A, B, C and D) (Option 001)
These connectors are used only in Test Sets with Option 001. The four Viking attenuator connectors are controlled by the 11713C Attenuator Switch Driver.
Line Module
The line fuse, as well as a spare, reside within the line module. Figure 9 illustrates where the fuses are located and how to access them.
Available Fuses
— Fuse (F 5 A/250V, 2110-0709) UL listed and CSA certified
For continued protection against fire hazard replace line fuse only with same type and rating: Fuse 5A/250V, Part Number 2110-0709. The use of other fuses or material is prohibited.
Figure 9 Line Fuse
Verify that the premise electrical voltage supply is within the range specified on the instrument.
Keysight N5280A User’s and Service Guide 17
-
RF INPUTS
A
BC
D
N5280A Frequency Converter
N5242A 4-Port PNA-X
N5280A Controlling the Test Set with N5242A
Controlling the Test Set with N5242A
This section will describe how to setup and operate the N5280A Frequency Converter Test Set with the N5242A 2-Port or 4-Port PNA-X Network Analyzer.
A PNA-X must be used to control the N5280A. The internal LO Source from the PNA-X can be used for testing. Refer to Figure 10 and Figure 11 for cable connections from the N5280A to the N5242A PNA-X.
Figure 10 Front Panel N5280A and N5242A PNA-X
18 Keysight N5280A User’s and Service Guide
N5280A
LO IN
LO AUX
D/R2 C/R1
A
B
IF OUTPUTS
A
B
CD
ATTENUATOR
IF OUTPUTS
N5242A 2-Port PNA-X
N5280A Frequency Converter
A
B
C/R1
D/R2
LO
Controlling the Test Set with N5242A
Figure 11 Rear Panel N5280A and N5242A PNA-X
Keysight N5280A User’s and Service Guide 19
-
N5264A Measurement Receiver
N5280A Frequency Converter
RF Source
N5280A Controlling the Test Set with N5264A
Controlling the Test Set with N5264A
This section will describe how to setup and operate the N5280A Frequency Converter Test Set with the N5264A Measurement Receiver.
A N5264A must be used to control the N5280A. The N5264A Option 108 adds an internal LO Source 26.5 GHz for testing. Refer to Figure 12 and Figure 16 on
page 24 for cable connections.
You may also use an external LO Source for testing. Refer to Figure 14 on
page 22 and Figure 15 on page 23 for cable connections.
Figure 12 Front Panel N5280A and N5264A with Option 108
20 Keysight N5280A User’s and Service Guide
N5280A
LO IN
LO AUX
D/R2 C/R1
A
B
IF OUTPUTS
A
B
CD
ATTENUATOR
IF OUTPUTS
GPIB
10 MHZ REF
N5264A Measurement Receiver
N5280A Frequency Converter
RF Source
D/R2
C/R1
B
A
LO
Controlling the Test Set with N5264A
Figure 13 Rear Panel N5280A and N5264A with Option 108
Keysight N5280A User’s and Service Guide 21
-
RF INPUTS
A
BC
D
N5264A Measurement Receiver
N5280A Frequency Converter
LO Source
RF Source
N5280A Controlling the Test Set with N5264A
Figure 14 Front Panel N5280A and N5264A without Option 108
22 Keysight N5280A User’s and Service Guide
N5280A
LO IN
LO AUX
D/R2 C/R1
A
B
IF OUTPUTS
A
B
CD
ATTENUATOR
IF OUTPUTS
GPIB
10 MHZ REF
N5264A Measurement Receiver
N5280A Frequency Converter
LO Source
RF Source
D/R2
C/R1
LO
B
A
Controlling the Test Set with N5264A
Figure 15 Rear Panel N5280A and N5264A without Option 108
Keysight N5280A User’s and Service Guide 23
-
RF INPUTS
AB C
D
B
C
D
LO INPUT
A
Vertical
Trigger
Horizontal
4
3
1
2
LO Source
RF Source
N5280A
Oscilloscope
AUX Trig IN
10 MHz IN
10 MHz OUT
IF OUTPUTS
N5280A Controlling the N5280A Option 002 using an Oscilloscope
Controlling the N5280A Option 002 using an Oscilloscope
Figure 16 Front Panel N5280A and Oscilloscope
24 Keysight N5280A User’s and Service Guide
N5280A
ATTEN X
ATTEN X
ATTEN Y
ATTEN Y
GPIB
LAN
USB
BANK 1
BANK 2
11713C Attenuator Switch Driver
LO IN
LO AUX
D/R2 C/R1
A
B
IF OUTPUTS
A
B
CD
ATTENUATOR
IF OUTPUTS
A
B
C
D
N5280A Frequency Converter
Controlling the RF Receiver Attenuators
Controlling the RF Receiver Attenuators
The 11713C attenuator switch driver controls the Test Set through the Viking connector on the rear panel.
Figure 17 Rear Panel N5280A Option 001 and 11713C
Keysight N5280A User’s and Service Guide 25
-
N5280A Controlling the RF Receiver Attenuators
Table 7 illustrates the key combinations that are required to set the
attenuation.
Example: In the Attenuator X panel, press 1 & 3 to set A, RF INPUTS to 15 dB.
Table 7 Attenuators
Bank 1 1
Bank 2 1
Figure 18 Front Panel
Attenuator X (A, RF INPUTS)
2 3
Attenuator X (B, RF INPUTS)
2 3
Attenuator Y (C, RF INPUTS)
5 6 7
Attenuator Y (D, RF INPUTS)
5 6 7
Attenuation (dB)
10 20 5
Attenuation (dB)
10 20 5
26 Keysight N5280A User’s and Service Guide
N5280A Operational Check
Operational Check
This section provides operational check to confirm the N5280A and PNA-X operational performance. The operation verification limits provided ensure that your N5280A and PNA-X are operating properly.
Equipment Required
The Keysight N5280A requires that the user be familiar with the equipment and components listed in Tab le 8.
This section provides an equipment list and setup of the PNA-X and Test Set.
Table 8 Equipment List
Description Qty
N5242A 2-Port or 4-Port PNA-X Network Analyzer (Option 020) or 1
N5264A Measurement Receiver (Option 108) 1
Keysight N5280A User’s and Service Guide 27
-
N5280A Operational Check
Verification Limits
Specifications for the N5280A Multiport Test Set are typical. System performance for the PNA-X and Test Set are only characteristic and intended as non warranted information. Only a functional certificate is provided for the N5280A.
It is recommended that you return your instrument to Keysight Technologies for servicing or repair if the Test Set and PNA-X performance exceed the operational verification limits.
Typical specifications are based on 1 to 2 units performance.
Table 9 RF Receiver Tracking
RF Port Magnitude Tracking
Frequency Value
10 MHz to 20 GHz ±2.0 dB
20 GHz to 22 GHz ±3.0 dB
22 GHz to 26.5 GHz ±4.0 dB
Table 10 Noise Floor
Direct Receiver Access Input Noise Floor1 IF Band wid th equal to 10 Hz
Frequency Receiver Access Input
10 MHz to 100 MHz 128 dBm
100 MHz to 500 MHz 132 dBm
500 MHz to 2 GHz 133 dBm
2 GHz to 20 GHz 129 dBm
20 GHz to 24 GHz 122 dBm
24 GHz to 26.5 GHz 119 dBm
1. Noise floor measured with a 50 Ohm load at port, and measured as the mean value of a 801 point trace.
28 Keysight N5280A User’s and Service Guide
N5280A Operational Check
Table 11 Trace Noise
Trace Noise Magnitude (dB, rms)
1,2,3
1 kHz IF Band wid th
Frequency Typical
10 MHz to 100 MHz 
100 MHz to 13.5 GHz 
13.5 GHz to 16 GHz 
16 GHz to 22.5 GHz 
22.5 GHz to 24 GHz 
24 GHz to 26.5 GHz 
,4
Trace Noise Phase (deg, rms)
1
1 kHz IF Bandwid th
10 MHz to 100 MHz 
100 MHz to 13.5 GHz 
13.5 GHz to 16 GHz 
16 GHz to 22.5 GHz 
22.5 GHz to 26.5 GHz 
Trace Noise, sweep to sweep variation.
1.
2. Measured linear magnitude CW Frequency at 201 points.
3. Trace Noise in dB = 20 * Log 10 (1+ standard deviation).
4.
Trace Noise in degrees = standard deviation.
Table 12 Port Match
RF/LO Port Match
Frequency Value
10 MHz to 10 GHz < 9 dB
10 GHz to 26.5 GHz < 4 dB
Keysight N5280A User’s and Service Guide 29
-
N5280A Operational Check
Table 13 Dynamic Range
Dynamic Range @ 10 Hz IF Band wid th
Frequency Typical (dB) Direct
Receiver Access Input
Typical (dBm) Low-Level Noise Floor
Typical (dBm) Max
1
Power
10 MHz to 50 MHz   13
50 MHz to 100 MHz   13
100 MHz to 500 MHz   13
500 MHz to 2 GHz   13
2 GHz to 20 GHz   13
20 GHz to 24 GHz   13
24 GHz to 26.5 GHz   13
1. RCVR IN @ 0.1 dB typical compression at max power.
The receiver access input dynamic range is calculated as the difference between the receiver access input low-level noise floor, and the source maximum power to the receiver that results with a typical 0.1dB compression level.
Low-level noise floor measured with 50 Ohm load at the port.
The effective dynamic range must take measurement uncertainties and interfering signals into account. This set-up should only be used when the receiver input will not exceed its compression or damage level.
30 Keysight N5280A User’s and Service Guide
N5280A Operational Check
Typical Plots
Figure 19 CE Response at 10 MHz IF Outputs
Figure 20 IF Band width at 10 MHz to 1.5 GHz
Keysight N5280A User’s and Service Guide 31
-
N5280A Operational Check
Figure 21 LO AUX Output
32 Keysight N5280A User’s and Service Guide
N5280A Operational Check
Operational Check Procedure (Option 700 and 001)
The sequence of this procedure is very important and must be followed or the performance accuracy and results may vary from the reference plots provided.
The following procedures are used with the N5242A 4-Port PNA-X.
If you are using a N5242A 2-Port PNA-X, channels C and D will be replaced with R1 and R2.
Preparing the N5242A
1. Connect the Test Set to the N5242A 4-Port PNA-X using the cables as shown in Figure 10 on page 18 and Figure 11 on page 19.
2. Connect the 10 dB attenuator to the PNA-X Reference port cable.
3. Turn On the Test Set.
4. Press [Preset].
5. Verify that the Stop Frequency is set to the maximum of the PNA-X and
Test Set. If not, press [Freq] > Stop.
6. Verify that the Start Frequency is set to 10 MHz. If not, press [Freq] > Start > [10 MHz].
7. Verify that the Power is to set to 10 dBm. If not, press [Power] > Power Level and enter 10 dBm].
8. Select [Avg] > IF Band wid th > enter 1 kHz.
9. Select [Sweep] > Number of Points > enter 401.
10.Allow the Test Set and the PNA-X to warm up for a minimum of 30
minutes.
11.Set the IF Config On. Trace/Chan > Channel > Hardware Setup > IF Config.
Keysight N5280A User’s and Service Guide 33
-
N5280A Operational Check
Figure 22 IF Config
34 Keysight N5280A User’s and Service Guide
N5280A Operational Check
12.Select all of the IF Inputs External A, B, C and D > OK.
Figure 23 IF Input
13.Connect the cable to A, RF INPUTS on the Test Set and select [Meas] >
More > Receivers, select A > Apply > OK.
Figure 24 Receiver A
Keysight N5280A User’s and Service Guide 35
-
N5280A Operational Check
14.Set the Reference level to 20 dBm. Press [Scale] > Reference level > [20 dBm].
All of the observed traces should decrease from 17 dB (± 5 dB) at 10 MHz to 34 dB (± 5 dB) at 26.5 GHz.
Figure 25 Typical Receiver A
15.Connect the cable to B, RF INPUTS on the Test Set. Select [Meas] > More >
Receivers, select B > Apply > OK to observe the power level trace.
16.Connect the cable to C, RF INPUTS on the Test Set. Select [Meas] > More > Receivers, select C > Apply > OK to observe the power level trace.
17.Connect the cable to D, RF INPUTS on the Test Set. Select [Meas] > More > Receivers, select D > Apply > OK to observe the power level trace.
36 Keysight N5280A User’s and Service Guide
N5280A Operational Check
Attenuator Element Verification Procedure
The Attenuator Element Verification Procedure is a functional test only. This test verifies the functionality of the variable attenuator in each channel RF path. The following procedures are used with the 11713C attenuator switch driver.
1. Connect the cables from the Test Set to the 11713C as shown in Figure 17
on page 25.
2. Press [Preset].
3. Set the IF Config On. Select Trace/Chan > Channel > Hardware Setup > IF
Config. Refer to Figure 22 on page 34.
4. Select all of the IF Inputs External A, B, C and D > OK. Refer to Figure 23 on
page 35.
5. Select [Analysis] > Statistics On.
6. Select [Meas] > More > Receivers, select A > Apply > OK. Refer to Figure
24 on page 35.
7. Select [Memory] > Normalize.
8. Select [1] on the 11713C. The mean value should read 10 dB (± 1 dB).
Select [1] again, it should read 0 dB.
9. Select [2] on the 11713C. The mean value should read 20 dB (± 1 dB). Select [2] again, it should read 0 dB.
10.Select [3] on the 11713C. The mean value should read 5 dB (± 1 dB). Select [3] again, it should read 0 dB.
11.Repeat See “Select [Meas] > More > Receivers, select A > Apply > OK.
Refer to Figure 24 on page 35.” on page 37. through See “Select [3] on the 11713C. The mean value should read 5 dB (± 1 dB). Select [3] again, it should read 0 dB.” on page 37. for B, C and D RF INPUTS.
Keysight N5280A User’s and Service Guide 37
-
N5280A Operational Check
Operational Check Procedure (Option 002)
The N5280A operational check can be verified using two Sources with a Power Meter and Power Sensor.
Requirements for Operational Check
The N5280A operational check measures the signal paths from the LO INPUT to the LO OUTPUT (LO AUX) and each of the four internal mixers, using a power divider. Internally, the LO signal is mixed with RF input signals supplied through each of the RF INPUTS (A, B, C, and D) and measured at each of the IF OUTPUTS (A, B, C, and D). The following process demonstrates that all associated switches, connectors, cables and circuitry are operational.
Required Hardware
To demonstrate that the N5280A is operating properly requires external equipment - the same equipment that would be used to perform a field calibration of the N5280A in a system. This external equipment includes two precision signal generators and a power meter with a power sensor and power sensor cable. The module interconnect cables are not used when performing this operational check; instead, two high-quality flexible 3.5 mm cables are needed along with adapters.
Please refer to the following table for recommended equipment.
Table 14 Keysight Recommended Equipment
Equipment Description
E8257D-532 or E8267D-532 or equivalent
N1913A or N1914A or equivalent Single/Dual Channel Power Meter
N8485A Option 100 or equivalent 10 MHz to 26.5 GHz Power Sensor
11730A or equivalent Power Sensor Cable
PSG #1 and PSG #2 Precision Signal Generator (250 kHz to 31.8 GHz)
38 Keysight N5280A User’s and Service Guide
N5280A
Source
Power Meter
Power Sensor
Operational Check
Instrument Setup (Source Calibration)
Using the power meter and power sensor, measure the PSG #1 (RF Source) and PSG #2 (LO Source) power levels at the frequencies listed in the Table 15
on page 40 at the end of test port cables. Record The RF and LO values in the
table.
Figure 26 Source and Power Meter Setup
Keysight N5280A User’s and Service Guide 39
-
IF OUTPUTS
N5280A Frequency Converter
LO Source
RF Source
RF INPUTS
AB C
D
BC
D
LO INPUT
A
Power Meter
Power Sensor
N5280A Operational Check
Check IF Output A
1. On the PSG #1 connect RF OUT using a high-quality flexible cable to RF INPUT A on the N5280A.
2. On the PSG #2 connect RF OUT using a high-quality flexible cable to LO INPUT on the N5280A.
3. Connect the power meter to IF OUTPUT A on the N5280A.
4. Use the output frequencies and power levels from the Tab le 15, set PSG #1
and PSG #2. Record the result in table for the measure IF OUTPUT A.
5. Repeat See “On the PSG #1 connect RF OUT using a high-quality flexible
cable to RF INPUT A on the N5280A.” on page 40. to See “Use the output frequencies and power levels from the Table 15, set PSG #1 and PSG #2. Record the result in table for the measure IF OUTPUT A.” on page 40. for IF
Outputs B, C, and D.
Figure 27 IF Output Channel A Setup
Table 15 Measured IF OUT Power Levels
PSG #1 Output Freq to RF IN A (GHz)
RF IN Power Level (dBm)
PSG #2 Output Freq to RF IN A (GHz)
LO IN Power Level (dBm)
10 dBm 0 dBm A BCDLO AUX
3.5 4 5 to 11
Power Level Range Expected at IF OUT A (dBm)
Power Meter Reading Measured for IF OUtputs and LO AUX (dBm)
40 Keysight N5280A User’s and Service Guide
N5280A Operational Check
Table 15 Measured IF OUT Power Levels
PSG #1 Output Freq to RF IN A (GHz)
RF IN Power Level (dBm)
PSG #2 Output Freq to RF IN A (GHz)
LO IN Power Level (dBm)
Power Level Range Expected at IF OUT A (dBm)
10 dBm 0 dBm A BCDLO AUX
8.5 9 5 to 11
17.5 18 5 to 11
26.0 26.5 5 to 11
Power Meter Reading Measured for IF OUtputs and LO AUX (dBm)
Keysight N5280A User’s and Service Guide 41
-
IF OUTPUTS
N5280A Front Panel
LO Source
RF INPUTS
AB C
D
BC
D
LO INPUT
LO IN
LO AUX
D/R2 C/R1
A
B
IF OUTPUTS
A
B
CD
ATTENUATOR
IF OUTPUTS
N5280A Rear Panel
A
Power Sensor
Power Meter
N5280A Operational Check
Check LO AUX
1. Connect the power meter to the LO AUX on the N5280A.
2. Use the output frequencies and power levels from the Table 15 on
page 40. Record the result in table for the measure LO AUX. The LO AUX
specification is greater than 10 dBm.
Figure 28 IF LO AUX Setup
42 Keysight N5280A User’s and Service Guide
N5280A
LO
IF
RF
LO
IF
RF
LO
IF
RF
LO
IF
RF
LO AUX
LO IN
A
B
C
D
Power Supply
5087-7727
0dB Gain
Nominal
+/- 9 Volts DC
+ 15 Volts DC
Max BW
LP OUT
LP IN
AIF
OUT
Max BW
LP OUT
LP IN
BIF
OUT
Max BW
LP OUT
LP IN
CIF
OUT
Max BW
LP OUT
LP IN
DIF
OUT
RF IN
0950-4729
N5280A
Operational Check
Figure 29 N5280A Block Diagram (Standard 700)
Keysight N5280A User’s and Service Guide 43
-
LO
IF
RF
LO
IF
RF
LO
IF
RF
LO
IF
RF
LO AUX
LO IN
A
B
C
D
A
BC
D
Power Supply
20 dB 10 dB
5 dB
20 dB 10 dB
5 dB
20 dB 10 dB
5 dB
20 dB 10 dB
5 dB
33325-6001
was 33325-60011
0dB Gain
Nominal
+/- 9 Volts DC
+ 15 Volts DC
11713C
ATTENUATORS
5 Volt
Max BW
LP OUT
LP IN
A IF
OUT
Max BW
LP OUT
LP IN
B IF
OUT
Max BW
LP OUT
LP IN
C IF
OUT
Max BW
LP OUT
LP IN
D IF
OUT
RF IN
0950-4729
N5280A Opt 001
5087-7727
N5280A Operational Check
Figure 30 N5280A Block Diagram (Option 001)
44 Keysight N5280A User’s and Service Guide
N5280A
LO
IF
RF
LO
IF
RF
LO
IF
RF
LO
IF
RF
LO AUX
LO INPUT
A
B
C
D
Power Supply
0dB Gain
Nominal
+/- 9 Volts DC
+15VoltsDC
RF INPUTS
0950-4729
N5280A Opt 002
5087-7727
IF OUTPUTS
A
B
C
D
Operational Check
Figure 31 N5280A Block Diagram (Option 002)
Keysight N5280A User’s and Service Guide 45
-
N5280A Safety and Information
Safety and Information
Introduction
Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument.
This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Safety Earth Ground
This is a Safety Class I Product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
Always use the three prong AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause product damage and the risk of electrical shock.
Statement of Compliance
This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
46 Keysight N5280A User’s and Service Guide
N5280A Safety and Information
Before Applying Power
Verify that the premises electrical supply is within the range of the instrument. The instrument has an autoranging power supply.
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only
The Mains wiring and connectors shall be compatible with the connector used in the premise electrical system. Failure, to ensure adequate earth grounding by not using the correct components may cause product damage, and serious injury.
Always use the three prong AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause product damage and the risk of electrical shock.
This product is designed for use in Installation Category II and Pollution Degree.
Before switching on this instrument, make sure the supply voltage is in the specified range.
Verify that the premise electrical voltage supply is within the range specified on the instrument.
Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used.
Keysight N5280A User’s and Service Guide 47
-
N5280A Safety and Information
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Discard used batteries according to manufacturer’s instructions.
For continued protection against fire hazard replace line fuse only with same type and rating. The use of other fuses or material is prohibited.
These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so
The opening of covers or removal of parts is likely to expose the user to dangerous voltages. Disconnect the instrument from all voltage sources before opening.
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers.
The detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument. The front panel switch is only a standby switch and is not a LINE switch (disconnecting device)
Connector Care and Cleaning Precautions
Remove the power cord to the instrument. To clean the connectors use alcohol in a well ventilated area. Allow all residual alcohol moisture to evaporate, and fumes to dissipate prior to energizing the instrument.
To prevent electrical shock, disconnect the Keysight N5280A from mains electrical supply before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
If flammable cleaning materials are used, the material shall not be stored, or left open in the area of the equipment. Adequate ventilation shall be assured to prevent the combustion of fumes, or vapors.
48 Keysight N5280A User’s and Service Guide
N5280A Regulatory Information
Regulatory Information
This section contains information that is required by various government regulatory agencies.
Instrument Markings
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
The AC symbol indicates the required nature of the line module input power.
This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005. All electric and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE Directive, 2002/96/EC).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is in the STANDBY position.
This symbol indicates that the power line switch is in the OFF position.
This symbol is used to identify a terminal which is internally connected to the product frame or chassis.
The CSA mark is a registered trademark of the CSA International.
This mark designates the pr (reference CISPR 11, Clause 5)
ICES/NMB-001
IP 2 0 The instrument has been designed to meet the requirements of IP 2 0 for egress and operational
This is a marking to indicate product compliance with the Canadian Interference-Causing Equipment Standard (ICES-001).
Direct Current.
environment.
The RCM mark is a registered trademark of the Australian Communications and Media Authority
Indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
oduct is an Industrial Scientific and Medical Group 1 Class A product
This symbol on all primary and secondary packaging indicates compliance to China standard GB 18455-2001.
South Korean Certification (KC) mark; includes the marking's identifier code which follows the format: MSIP-REM-YYY-ZZZZZZZZZZZZZZ.
Keysight N5280A User’s and Service Guide 49
-
N5280A Regulatory Information
50 Keysight N5280A User’s and Service Guide
Battery
EMC
N5280A Regulatory Information
Do not throw batteries away but collect as small chemical waste, or in accordance with your country’s requirements. You may return the battery to Keysight Technologies for disposal. Refer to “Contacting Keysight” on page 54 for assistance.
Complies with the essential requirements of the European EMC Directive as well as current editions of the following standards (dates editions are cited in the Declaration of Conformity):
— IEC/EN 61326-1
— CISPR Pub 11 Group 1, class A
— AS/NZS CISPR 11
— ICES/NMB-001
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
South Korean Class A EMC Declaration
This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
A 급 기기 ( 업무용 방송통신기자재 ) 이 기기는 업무용 (A 급 ) 전자파적합 기기로서 판 매자 또는 사용자는 이 점을 주 의하시기 바라 며 , 가정외의 지 역에서 사용하는 것을 목적으 로 합니다 .
Safety
Complies with the essential requirements of the European Low Voltage Directive as well as current editions of the following standards (dates and editions are cited int he Declaration of Conformity):
— IEC/EN 61010-1
— Canada: CSA C22.2 No. 61010-1
— USA: UL std no. 61010-1
Acoustic Statement (European Machinery Directive):
— Accoustical noise emission
LpA<70 dB Operator position Normal operation mode Per ISO 7779
Keysight N5280A User’s and Service Guide 51
-
N5280A Regulatory Information
Declaration of Conformity
To find a current Declaration of Conformity for specific Keysight product, go to:
http://regulations.about.keysight.com/DoC/search.htm
52 Keysight N5280A User’s and Service Guide
N5280A Electrostatic Discharge Protection
Electrostatic Discharge Protection
Electrostatic discharge (ESD) can damage or destroy electronic components. The product is shipped in materials that prevent damage from static, and should only be removed from the packaging in an anti-static area ensuring that the correct anti-static precautions are taken.
Two types of ESD protection are listed below. Purchase acceptable ESD accessories from your local supplier.
—Conductive table-mat and wrist-strap combination
— Conductive floor-mat and heal-strap combination
Both types, when used together, provide a significant level of ESD protection. To ensure user safety, static-safe accessories must provide at least 1 MΩof isolation from ground.
These techniques for a static-safe work station should not be used when working on circuitry with a voltage potential greater than 500 volts.
Keysight N5280A User’s and Service Guide 53
-
N5280A Keysight Support, Services, and Assistance
Keysight Support, Services, and Assistance
Service and Support Options
There are many other repair and calibration options available from the Keysight Technologies support organization. These options cover a range of service agreements with varying response times. Contact Keysight for additional information on available service agreements for this product.
Contacting Keysight
Assistance with test and measurement needs, and information on finding a local Keysight office are available on the Internet at:
http://www.keysight.com/find/assist
If you do not have access to the Internet, contact your field engineer.
In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number. With this information, the Keysight representative can determine the warranty status of your unit.
Shipping Your Product to Keysight for Service or Repair
Keysight Technologies reserves the right to reformat or replace the internal hard disk drive in your analyzer as part of its repair. This will erase all user information stored on the hard disk. It is imperative, therefore, that you make a backup copy of your critical test data located on the analyzer’s hard disk before shipping it to Keysight for repair.
If you wish to send your instrument to Keysight Technologies for service or repair:
— Include a complete description of the service requested or of the failure and
a description of any failed test and any error message.
— Remove and retain the front handles and all rack mount hardware. The
analyzer should be sent to Keysight in the same configuration as it was originally shipped.
— Remove and retain the front handles and all rack mount hardware. The
analyzer should be sent to Keysight in the same configuration as it was originally shipped.
— Contact Keysight for instructions on where to ship your analyzer.
54 Keysight N5280A User’s and Service Guide
Installation Note Xxxxx-xxxxx 3
This information is subject to change without notice. © Keysight Technologies 2008 – 2017 Edition 1: June 2017 Supersedes: August 2015
*N5280-90001*
N5280-90001
www.keysight.com