Keysight Technologies N524 A Series, N5264B, N524 B Series, N5264A, N5244B Installation Note

...
Keysight 2-Port/4-Port
N524xA to N524xB and N5264A to N5264B Model
PNA-X Upgrade Kit
This manual provides documentation for the following models: N524xA and N5264A
To Upgrade 2-Port/4-Port "A" models to 2-Port/4-Port "B" models
N5241AU-690 (For N5241B), N5242AU-690 (For N5242B), N5244AU-690 (For
N5244B), N5245AU-690 (For N5245B), N5247AU-690 (For N5247B), N5249AU-690
(For N5249B), and N5264AU-690 (For N5264B)
Upgrade Kit Numbers: N5242-60117, N5245-60115, N5247-60114, N5264-60102
Installation Note

Notices

© Keysight Technologies, Inc. 2007-2019
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
N5242-90023
Edition
Edition 1, September 2019
Supersedes: August 9, 2019
Printed in USA/Malaysia
Published by: Keysight Technologies
1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or
commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Keysight Licensed Functionality Upgrade Kit Upgrade Kit Numbers: N5242-60117, N5245-60116, N5247-60114,
N5264-60102 Installation Note

Description of this Upgrade

This upgrade converts your N52xxA microwave network analyzer or N5264A microwave receiver to an equivalent N52xxB or N5264B. This is done by removing the existing front panel assembly, outer covers, and installing a new front panel assembly, SSDA, and outer covers.
As part of this upgrade, there are other assemblies that need to be verified, because the boards must be of a certain type to be compatible with the B model upgrade. Refer to “Initial Instrument Inspection” on page 18.
After installation of this upgrade, your analyzer will be a “B” model rather than an “A” model.
STOP! This repair must be done at a service center! Refer to “Getting
Assistance from Keysight” on page 4.
IMPORTANT! As part of the upgrade, the service center needs a copy of
the customer’s Keysight License files that were sent to them from Keysight. Refer to “License Key File Redemption” on page 5.
IMPORTANT! N5244/5A PNAs with H29 cannot upgrade to N5244/5B.
Included in the contents of this kit is a Software Entitlement Certificate instructions on how to redeem it. Do NOT attempt to redeem this until you have read “License Key File Redemption” on page 5.
For additional information on using the Keysight Software Manager and redeeming a Software Entitlement Certificate, refer to
http://literature.cdn.keysight.com/litweb/pdf/N5242-90024.pdf.
with
certificate
3
-
Description of this Upgrade Getting Assistance from Keysight

Getting Assistance from Keysight

Installing this upgrade kit requires special skills and experience. If you think you may not be qualified to do the work, or need advice, contact Keysight.

If You Have Problems With the Upgrade Kit Contents

Keysight stands behind the quality of the upgrade kit contents. If you have problems with any item in the kit, refer to www.keysight.com and the Contact
Keysight () link.
4 Installation Note N5242-90023
Description of this Upgrade Getting Prepared

Getting Prepared

License Key File Redemption

IMPORTANT! As part of the upgrade, the service center needs to
request a the customer’s Keysight License files from Keysight. Refer to
“License Key File Redemption” on page 5.
IMPORTANT! Also, you will need to contact the Keysight factory support node to request multiple license key files. Refer to “Obtain Multiple
License Key Files” on page 7.
Ensure that you are connected to an external server, before attempting to download your email and license key file.
The enclosed Software Entitlement Certificate is a receipt, verifying that you have purchased a licensed option for the analyzer of your choice. You must now use a Keysight Web page to request a license key file for the instrument that will receive the option.
To upgrade model(s), you must request an Option 690 license key file from the Keysight Software Manager:
http://www.keysight.com/find/softwaremanager:
To complete the request, you will need to gather the following information:
— From the certificate
— Order number
— Certificate number
— From your instrument
(Instrument information is available in the network analyzer - on the toolbar, click Help, then click About Network Analyzer.)
— Model number
— Serial number
Using the information just gathered, you must request license key file from the Keysight Software Manager:
http://www.keysight.com/find/softwaremanager.
The following Keysight Software Manager (KSM) procedure has been added— but is abbreviated—for users that are familiar with the KSM process and has steps that are specific to the A model to B model upgrade (Option 690).
1. Enter your login and password or setup a login as required.
Installation Note N5242-90023 5
-
Description of this Upgrade Getting Prepared
2. In the Keysight License Manager welcome window, click You can add a new certificate.
3. In the "Keysight Software Manager: Add Certificate" window that opens, in
the "Enter your certificate information" section, refer to your Software License Certificate that you received with your upgrade kit and enter the following:
Order Number
Certificate Number
4. Click Continue (to return to the Keysight Software Manager welcome
screen).
5. If there are multiple "A" model PNAs that are being upgraded to a "B" model, repeat steps 3 and 4.
Else, continue to next step.
6. In the Keysight Software Manager welcome screen, click "You can request new licenses".
7. In the "Keysight Software Manager: Request Licenses" window that opens, in the "Select the host to assign licenses to" section, click the Click to add
or select a host drop-down.
8. Click Add a new License Notifier Host ID host...
9. In the License Notifier Host ID: box that is displayed, enter your
instrument’s model and serial number, separated by only a comma—no spaces). Example:
— New License Host Notifier Host ID host: N5242A,MY12345678
10. In the Select the licenses to assign section, scroll down until you see a
Option 690 upgrade with your instrument’s "A" and "B" model values— similar to this example—and then click the box to select:
N5247AU-690 Upgrade N5247A to N5247B
IMPORTANT: To avoid the loss of an option license, ensure that you only
have a single Option 690 upgrade checked.
11. Click Assign Licenses
12. In the "Assigned licenses:" section that opens, listing all of the License
Notifier Host IDs and associated instrument information, verify the following:
License Notifier Host ID: contains the correct "B" model and serial
number
the associated N52xxAU-690 Upgrade lists the correct "A" model to
"B" model
6 Installation Note N5242-90023
Description of this Upgrade Getting Prepared
If any of the information listed in the "Assigned licenses:" section is incorrect, click Remove next to the incorrect License Notifier Host ID: or N52xxAU-690 Upgrade to remove them from the "Assigned Licenses" list.
13. If there are multiple "A" model PNAs that are being upgraded to a "B" model, repeat steps 7 through 12.
Else, continue to next step.
14. Once you have verified the License Notifier Host ID & N52xxAU-690 Option models are correct, click Continue > (an email is sent to you with an Option 690 license file for each specific License Notifier Host ID’d PNA).
15. Keysight will promptly email your license key file(s) to your profile’s email address.
Obtain Multiple License Key Files
All of the licenses in your instrument need to be converted to a new format for your "B" model instrument. Contact Keysight Technologies factory support early in your upgrade process to receive your licenses in a timely manner. See
“Getting Assistance from Keysight” on page 4.
You will need to provide Keysight with the foll
owing information:
—model number
— serial number
— email address of the technician performing this upgrade (i.e., to send the
license key files)
Installation Note N5242-90023 7
-
Description of this Upgrade Protecting Your Workspace from Electrostatic Discharge

Downloading the Online PNA Service Guide

To view the Service Guide for your PNA model number, use the following steps:
1. Go to www.keysight.com.
2. In the Search box, enter the model number of the analyzer (Ex: N5242B)
and click Search.
3. Click Technical Support > Document Library > Manuals or under Refine Your Results:
4. Click Manuals > Service Manual.
5. Click the service guide title to download the PDF file.
6. When the PDF of the Service Guide is displayed, scroll through the
Contents section bookmarks to locate the information needed.

Protecting Your Workspace from Electrostatic Discharge

For information, click on the Chapter 1 bookmark, "Electrostatic Discharge Protection" in the PDF of your instrument’s Service Guide on
1
www.keysight.com
.

ESD Equipment Required for the Installation

Description Keysight Part Number
ESD grounding wrist strap 9300-1367
5-ft grounding cord for wrist strap 9300-0980
2 x 4 ft conductive table mat and 15-ft grounding wire 9300-0797
ESD heel strap (for use with conductive floors) 9300-1308
1. See “Downloading the Online PNA Service Guide” on page 8.
8 Installation Note N5242-90023
Description of this Upgrade Tools Required for the Installation

Tools Required for the Installation

Description Qty
T-10 TORX driver - set to 9 in-lbs (1.02 N.m) 1
T-20 TORX driver - set to 21 in-lbs (2.38 N.m) 1
5/16-in (8 mm) nutsetter or open end torque wrench - set to 10 in-lbs (1.13 N.m) 1
5/16-in (8 mm) nutsetter or open end torque wrench - set to 21 in-lbs (2.38 N.m) 1
9 mm nutsetter or open end torque wrench - set to 21 in-lbs) (2.38 N.m) 1
1/4-in (6 mm) open end wrench 1
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections except the front and rear panel bulkhead connectors. On these, use a 9 mm nutsetter or open end torque wrench set to 21 in-lb.

About Installing the Upgrade

Products affected All N524xA microwave network analyzers and all N5264A
Installation to be performed by Keysight service center, personnel qualified by Keysight, or
Estimated installation time 4 hours
Estimated adjustment time 1.5 hours
Estimated full instrument calibration time 4.5 to 8.0 hours
microwave receivers
customer
Installation Note N5242-90023 9
-
Description of this Upgrade Items Included in the Upgrade Kit

Items Included in the Upgrade Kit

Check the contents of your kit against the following list. If any part is missing or damaged, contact Keysight Technologies. Refer to “Description of this
Upgrade” on page 3.
This section applies to multiple models and upgrade kits. Please refer to the model and upgrade kit for your instrument’s upgrade.
Extra quantities of items such as protective plastic caps, screws, cable ties, and cable clamps may be included in this upgrade kit. It is normal for some of these items to remain unused after the upgrade is completed.
Table 1 Contents of Upgrade Kit N5242-60117 for N5241AU-690 (N5241B),
N5242AU-690 (N5242B), and N5249AU-690 (N5249B)
Qty Description Part Number
1 Installation Note (this document) N5242-90023
1 Software Entitlement Certificate 5964-5145
2 Front handle 5023-3074
4 Rear foot 5041-7903
4 Bottom foot 5041-7906
2 Trim strip, side, phantom gray 5041-7908
1 Caution label 5190-1799
1 Win10 IoT Ent LTSB 2016 Multi-Lang OEI Value 9010-0335
1 China RoHS Addendum 9320-6722
1 Front frame – (assembly) N5240-60071
4 Machine screw, M4.0 x 25 pan head (To attach rear feet to chassis.) 0515-1619
14
10 Installation Note N5242-90023
Machine screw, M3.0 x 10 mm flat head (To attach lower dress panel chassis).
5 Machine screw, M4.0 x 10 flat head (To attach front frame to chassis.) 0515-2044
1 Cable Assembly–Flat Ribbon 230 mm LG - Handler I/O cable (LFE) 8121-2919
1 Solid-state drive assembly (SSDA), imaged, for 2.2 GHz CPU and Win 10 N5242-60135
1 Lower front panel overlay, 2-port (N5241/2/9B Options 201/219)
Lower front panel overlay, 2-port (N5241/2/9B Option 224)
Lower front panel overlay, 4-port (N5241/2/9B Option 401/419 /423)
Lower front panel overlay, 2-port (N5241/2/9B Option 219 with Option 029) N5242-80026
overlay frame to
0515-1946
N5222-80010
N5242-80024
N5222-80013
Description of this Upgrade Items Included in the Upgrade Kit
Table 1 Contents of Upgrade Kit N5242-60117 for N5241AU-690 (N5241B),
N5242AU-690 (N5242B), and N5249AU-690 (N5249B)
Qty Description Part Number
Lower front panel overlay, 2-port (N5241/2/9B Option 224 with Option 029) N5242-80027
Lower front panel overlay, 4-port (N5241/2/9B Option 419/423 with Option 029) N5242-80028
Lower front panel overlay, 2-port (N5241/2/9B Option 217) N5242-80029
Lower front panel overlay, 2-port (N5241/2/9B Option 222) N5242-80030
Lower front panel overlay, 4-port (N5241/2/9B Option 417 or 422) N5242-80031
Lower front panel overlay, 2-port (N5241/2/9B Option 217 with Option 029) N5242-80032
Lower front panel overlay, 2-port (N5241/2/9B Option 222 with Option 029) N5242-80033
Lower front panel overlay, 4-port (N5241/2/9B Option 417/422 with Option 029) N5242-80034
1 Nameplate, N5241B N5241-80002
Nameplate, N5242B N5242-80022
Nameplate, N5249B N5249-80003
Lower Dress Panel, 2-Port (ALL PNA models). Goes on, before lower front
1
overlay. Attach with 0515-1946 subpanel screws.
Lower Dress Panel, 4-Port (ALL PNA models). Goes on, before lower front overlay. Attach with 0515-1946 subpanel screws.
1 Rear panel, power supply N5240-00012
panel
panel
N5240-00010
N5240-00009
1 Cable, assembly 60C 30 AWG 60F 60F N5240-60090
1 Label, KCC N5240-80041
2 Guard, jumper cables, side—2-Port & 4-Port N5242-00048
1 Outer cover N5242-00047
1 Guard, jumper cables, center—4-Port N5242-00049
1 Front impact cover N9040-40007
1 Rear impact cover N9040-40008
2 Strap handle assembly N5242-60076
4 Key lock (for bottom foot) 5021-2840
Not Included - (For Reference Only) - External Hard ware and Miscellaneous Parts, All Options
Machine screw M5.0 x 18, flat head (To attach strap handle.) 0515-0710
Hole plug (When analyzer is rack mounted.) N9020-40007
Test set, subpanel (2-port)
1
- N5241/2/9B only
N5221-00006
Installation Note N5242-90023 11
-
Description of this Upgrade Items Included in the Upgrade Kit
Table 1 Contents of Upgrade Kit N5242-60117 for N5241AU-690 (N5241B),
N5242AU-690 (N5242B), and N5249AU-690 (N5249B)
Qty Description Part Number
Test set, subpanel (4-port)1 - N5241/2/9B only
Wire assy, LED PCA
1
- N5241/2/9B only
N5221-00007
N5242-60009
Synthesizer board (With tabs for s/n 5200 and above) N5240-60074
Synthesizer board (Without tabs for s/n’s below 5200) N5240-60076
Synthesizer board kit (Without tabs for s/n’s below 5200) – Qty is 3 boards N5242-60168
DSP (SPAM) version 5 board N5240-60077
1 DSP (SPAM) cable (2-port only) N5247-60023
1 N5247-60024
IF Multiplexer board N5240-60062
Version 7 CPU assembly W1312-60211
1. If it is necessary to replace these parts, Keysight will replace these parts at no cost to the customer.
12 Installation Note N5242-90023
Description of this Upgrade Items Included in the Upgrade Kit
Table 2 Contents of Upgrade Kit N5245-60116 for N5244AU-690 (N5244B) and
N5245AU-690 (N5245B)
Qty Description Part Number
1 Installation Note (this document) N5242-90023
1 Software Entitlement Certificate 5964-5145
2 Front handle 5023-3074
4 Rear foot 5041-7903
4 Bottom foot 5041-7906
2 Trim strip, side, phantom gray 5041-7908
1 Caution label 5190-1799
1 Win10 IoT Ent LTSB 2016 Multi-Lang OEI Value 9010-0335
1 China RoHS Addendum 9320-6722
1 Front frame – (assembly) N5240-60071
4 Machine screw, M4.0 x 25 pan head (To attach rear feet to chassis.) 0515-1619
14
Machine screw, M3.0 x 10 mm flat head (To attach lower dress panel chassis).
overlay frame to
0515-1946
5 Machine screw, M4.0 x 12 flat head (To attach front frame to chassis.) 0515-2044
1 Cable Assembly–Flat Ribbon 230 mm LG - Handler I/O cable (LFE) 8121-2919
1 Solid-state drive assembly (SSDA), imaged, for 2.2 GHz CPU and Win 10 N5242-60135
1 Lower front panel overlay, 2-port (N5244/5B Options 201/219) N5242-80023
Lower front panel overlay, 2-port (N5244/5B Option 224) N5245-80027
Lower front panel overlay, 4-port (N5244/5B Option 401/419 /423) N5242-80025
Lower front panel overlay, 2-port (N5244/5B Option 222) N5245-80028
Lower front panel overlay, 2-port (N5244/5B Option 224 with Option 029) N5245-80029
Lower front panel overlay, 4-port (N5244/5B Option 423 with Option 029) N5245-80030
Lower front panel overlay, 2-port (N5244/5B Option 217) N5242-80029
Lower front panel overlay, 4-port (N5244/5B Option 417/422) N5242-80031
Lower front panel overlay, 2-port (N5244/5B Option 222 with Options 029) N5245-80033
Lower front panel overlay, 4-port (N5244/5B Option 417/422 with Options
1 Nameplate, N5244B N5244-80002
Nameplate, N5245B N5245-80026
029) N5245-80031
Lower Dress Panel, 2-Port (ALL PNA models). Goes on, before lower front
1
overlay. Attach with 0515-1946 subpanel screws.
Installation Note N5242-90023 13
panel
N5240-00010
-
Description of this Upgrade Items Included in the Upgrade Kit
Table 2 Contents of Upgrade Kit N5245-60116 for N5244AU-690 (N5244B) and
N5245AU-690 (N5245B)
Qty Description Part Number
Lower Dress Panel, 4-Port (ALL PNA models). Goes on, before lower front panel overlay. Attach with 0515-1946 subpanel screws.
1 Rear panel, power supply N5240-00012
1 Cable, assembly 60C 30 AWG 60F 60F N5240-60090
1 Label, KCC n5240-80041
1 Outer cover N5245-00035
2 Guard, jumper cables, side—2-Port & 4-Port N5242-00048
1 Guard, jumper cables, center—4-Port N5242-00049
1 Front impact cover N9040-40007
1 Rear impact cover N9040-40008
2 Strap handle assembly N5247-60030
4 Key lock (for bottom foot) 5021-2840
N5240-00009
Not Included - (For Reference Only) - External Hard ware and Miscellaneous Parts, All Options
Machine screw M5.0 x 18, flat head (To attach strap handle.) 0515-0710
Hole plug (When analyzer is rack mounted.) N9020-40007
Synthesizer board (With tabs for s/n 5200 and above) N5240-60074
Synthesizer board (Without tabs for s/n’s below 5200) N5240-60076
Synthesizer board kit (Without tabs for s/n’s below 5200) – Qty is 3 boards N5242-60168
DSP (SPAM) version 5 board N5240-60077
1 DSP (SPAM) cable (2-port only) N5247-60023
1 N5247-60024
IF Multiplexer board N5240-60062
Version 7 CPU assembly W1312-60211
Test set, subpanel (2-port)
Test set, subpanel (4-port)
Wire assy, LED PCA
1. If it is necessary to replace these parts, Keysight will replace these parts at no cost to the customer.
1
- N5244/45B only
1
- N5244/45B only
1
- N5244/45B only
N5224-00004
N5224-00005
N5225-60001
14 Installation Note N5242-90023
Description of this Upgrade Items Included in the Upgrade Kit
Table 3 Contents of Upgrade Kit N5247-60114 for N5247AU-690 (N5247B))
Qty Description Part Number
1 Installation Note (this document) N5242-90023
1 Software Entitlement Certificate 5964-5145
1 Front frame – (assembly) N5240-60100
4 Machine screw, M4.0 x 25 pan head (To attach rear feet to chassis.) 0515-1619
14
12 Machine screw, M4.0 x 12 flat head (To attach front frame to chassis.) 0515-2044
Machine screw, M3.0 x 10 mm flat head (To attach lower dress panel chassis).
1 Win10 IoT Ent LTSB 2016 Multi-Lang OEI Value 9010-0335
1 China RoHS Addendum 9320-6722
1 Solid-state drive assembly (SSDA), imaged, for 2.2 GHz CPU and Win 10 N5242-60135
2 Front handle 5023-3074
overlay frame to
0515-1946
4 Rear foot 5041-7903
4 Bottom foot 5041-7906
2 Trim strip, filler (For analyzers without handles.) 5041-7908
1 Two person Caution label 5190-1799
1 Cable Assembly–Flat Ribbon 230 mm LG - Handler I/O cable (LFE) 8121-2919
1 Lower front panel overlay, 2-port (N5247B Options 201 or 219)
Lower front panel overlay, 2-port (N5247B Option 224)
Lower front panel overlay, 4-port (N5247B Option 401 or 419)
Lower front panel overlay, 2-port (N5247B Option 423)
Lower front panel overlay, 2-port (N5247B Option 224 with Option 029)
Lower front panel overlay, 4-port (N524xB Option 423 with Option 029) N5247-80023
Lower front panel overlay, 4-port (N524xB Option 425 with Option 029) N5247-80028
1 Nameplate, N5247B N5247-80017
Lower Dress Panel, 2-Port (ALL PNA models). Goes on, before lower front
1
overlay. Attach with 0515-1946 subpanel screws.
panel
N5247-80020
N5247-80018
N5247-80021
N5247-80019
N5247-80022
N5240-00010
Lower Dress Panel, 4-Port (ALL PNA models). Goes on, before lower front overlay. Attach with 0515-1946 subpanel screws.
1 Rear panel, power supply N5240-00012
1 Cable, assembly 60C 30 AWG 60F 60F N5240-60090
1 Label, KCC N5240-80041
Installation Note N5242-90023 15
panel
N5240-00009
-
Description of this Upgrade Items Included in the Upgrade Kit
Table 3 Contents of Upgrade Kit N5247-60114 for N5247AU-690 (N5247B))
Qty Description Part Number
1 Outer cover N5245-00035
2 Guard, jumper cables, side—2-Port & 4-Port N5242-00048
1 Guard, jumper cables, center—4-Port N5242-00049
1 Guard, jumper cables, center—4-Port N5247-00024
1 Front impact cover N9040-40007
1 Rear impact cover N9040-40008
2 Strap handle assembly N5247-60030
4 Key lock (for bottom foot) 5021-2840
Not Included - (For Reference Only) - External Hard ware and Miscellaneous Parts, All Options
Machine screw M5.0 x 18, flat head (To attach strap handle.) 0515-0710
Hole plug (When analyzer is rack mounted.) N9020-40007
Synthesizer board (With tabs for s/n 5200 and above) N5240-60074
Synthesizer board (Without tabs for s/n’s below 5200) N5240-60076
Synthesizer board kit (Without tabs for s/n’s below 5200) – Qty is 3 boards N5242-60168
DSP (SPAM) version 5 board N5240-60077
1 DSP (SPAM) cable (2-port only) N5247-60023
1 N5247-60024
IF Multiplexer board N5240-60062
Version 7 CPU assembly W1312-60211
16 Installation Note N5242-90023
Description of this Upgrade Items Included in the Upgrade Kit
Table 4 Contents of Upgrade Kit N5264-60102 for N5264AU-690 (N5264B)
Qty Description Part Number
1 Installation Note (this document) N5242-90023
1 Software Entitlement Certificate 5964-5145
1 Front frame – (assembly) N5240-60071
4 Machine screw, M4.0 x 25 pan head (To attach rear feet to chassis.) 0515-1619
12
12 Machine screw, M4.0 x 12 flat head (To attach front frame to chassis.) 0515-2044
Machine screw, M3.0 x 10 mm flat head (To attach lower dress panel chassis).
1 Win10 IoT Ent LTSB 2016 Multi-Lang OEI Value 9010-0335
1 China RoHS Addendum 9320-6722
1 Solid-state drive assembly (SSDA), imaged, for 2.2 GHz CPU and Win 10 N5242-60135
2 Front handle 5023-3074
overlay frame to
0515-1946
4 Rear foot 5041-7903
4 Bottom foot 5041-7906
2 Trim strip, filler (For analyzers without handles.) 5041-7908
1 Lower front panel overlay, N5264B N5264-80005
1 Nameplate, N5264B N5264-80004
1 Lower Dress Panel, attach with 0515-1946 subpanel screws. N5264-00005
1 Rear panel, power supply N5240-00012
1 Cable, assembly 60C 30 AWG 60F 60F N5240-60090
1 Label, KCC n5240-80041
2 Guard, jumper cables, side—2-Port & 4-Port N5242-00048
1 Outer cover N5242-00047
1 Guard, jumper cables, center—4-Port N5242-00049
1 Front impact cover N9040-40007
1 Rear impact cover N9040-40008
2 Strap handle assembly N5242-60076
4 Key lock (for bottom foot) 5021-2840
Not Included - (For Reference Only) - External Hard ware and Miscellaneous Parts, All Options
Machine screw M5.0 x 18, flat head (To attach strap handle.) 0515-0710
Hole plug (When analyzer is rack mounted.) N9020-40007
Installation Note N5242-90023 17
-
Description of this Upgrade Initial Instrument Inspection

Initial Instrument Inspection

STOP! This repair must be done at a service center! Refer to “Getting
Assistance from Keysight” on page 4.
This section and all steps must be completed before proceeding to the following sections or customer data may be lost or instrument repair delayed!
The purpose of this section is to verify where the instrument should begin the model conversion process. This section verifies that the instrument has all of the required board assemblies for a B model upgrade.
If your instrument does not have all of the required board assemblies, contact the customer, to provide a quote on parts and labor. Upon customer approval, order up all of the new upgrade kits and boards, before the "B" model upgrade can be completed.
Refer to Table 6 on page 19 and then to “Verifying the Board Assemblies” on
page 20.
18 Installation Note N5242-90023
Description of this Upgrade Initial Instrument Inspection
Table 6 Overview of Assembly Requirements for B Model Upgrade
Assembly Part Number Serial Number for PNAs that start using
these boards (PNAs with these serial prefixes do not need to be upgraded)
CPU, version 7 W1312-60211 MY/US5540 and above N8989A-PC7
SPAM 5 N5240-60078
For 2-port PNAs, also order x1 N5247-60023 and x1 N5247-60024
IF Multiplexer, version B
Synthesizers, versions E, F, or G
N5240-60062 MY/US5540 and above N889A-MX2
N5240-60076 (single board) or N5242-60168 (3 board kit)
MY/US5200 and above N8989A-SP5
MY/US5120 and above N8989A-SR6
Can be ordered using the part number, or:
2-port PNAs without option 224 need two synthesizer boards. 4-port PNAs or 2-port option 224 PNAs need three synthesizer boards.
Test Set Motherboard –
N5245-60163 MY/US4940 and above ––––
N5241/2A only
Installation Note N5242-90023 19
-
Description of this Upgrade Initial Instrument Inspection

Verifying the Board Assemblies

The following board assemblies need to be verified and installed during the upgrade:
1. Press Help > About NA...
2. In the About Network Analyzer window that opens, make a note of the
following:
System CPU Version
The system CPU must be a version 7 or greater. This CPU
is not included in this kit. Please contact Keysight if you need to replace the CPU and order CPU board number: W1312-60211.
DSP Version
The DSP (SPAM) board must be version 5 or greater. The
SPAM V boards are not included in this kit. Please contact Keysight if you need to replace the SPAM boards and order N5240-60077 SPAM V. If you have a 2-Port instrument, you will also need to order one each of cable numbers: N5247-60023 and N5247-60024.
3. Press Utility > System > Service > Diagnostics > EEPROM Headers
Verify the LO Synth, Src1 Synth, and Src2 Synth boards part number
is either "N524063074" or "N524263150". The synthesizer boards are not included in this kit. Please contact Keysight if you need to replace the synthesizer boards and order N5240-60076.
4. Remove the instrument outer covers and if necessary the power supply, to verify the Midplane Board’s part number reads: "W1312-63376". Refer to the Service Guide on www.keysight.com for removing the power supply.
If the midplane board’s part number does not read "W1312-63376", the Midplane Board needs to be modified. Refer to “Step 13. Inspect or Modify
Midplane Board (W1312-63095 to a W1312-63376)” on page 35.
5. N5244/5/A models (ONLY): If the subpanels need to be changed, the LED boards need to be rotated 180 degrees and the LED cables replaced (refer to “Step 11. Inspect and If Required, Replace Front Subpanel” on page 33). Regardless, ensure that LED boards are installed correctly and the LED cables (N5225-60001 (x2)) are the correct part numbers. Refer to “Step
12. If Required, Rotate the LED Boards and Replace the LED Bias Cables (N5244/5A Only)” on page 34 and to the Service Guide’s "Removing and
Replacing the Front Panel LED Boards" on www.keysight.com. If not, these parts should be ordered and replaced at no cost to the customer. Refer to
Table 2 on page 13.
20 Installation Note N5242-90023
Description of this Upgrade Initial Instrument Inspection
6. If your instrument has the minimum criteria mentioned above or you have received the required parts proceed to “Step 1. Copy All Unique Files to an
External Disk Drive” on page 24.

Overview of the Installation Procedure

Windows XP OS only: Before installing the new SSDA with Windows 7 OS— that is part of this upgrade A.10.40.00 must be loaded initializes the instrument to a known
This section assumes you have completed the “Initial Instrument
Inspection” section and verified, that all of the board assemblies meet the
requirements for this A to B model upgrade. If not, refer to the “Initial
Instrument Inspection” on page 18.
kit—the service center’s SSDA with firmware
and initialized in the instrument. This
stable state.
Installation Note N5242-90023 21
-
Description of this Upgrade Initial Instrument Inspection
All instruments:
“Step 1. Copy All Unique Files to an External Disk Drive” on page 24
Windows 7 Operating System (Windows 7 OS) only:
“Step 2. Update Windows 7 OS Instruments with A.10.40 or above” on page 25
“Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or Greater” on page 26
All instruments:
“Step 4. Remove the outer covers” on page 27
“Step 5. Remove the Power Supply Cover” on page 28
“Step 6. Install the *New* Version 7 CPU Assembly” on page 28
Windows XP Operating System (Windows XP OS) only:
“Step 7. Install the Service Center’s N5242-60133 Solid-State Disk Drive Assembly (SSDA) with A.10.40 or above—(Windows XP OS Only)” on page 29
All instruments:
“Step 8. Install the *New* Power Supply Cover” on page 30
“Step 9. Power Up Your Instrument with the *New* Version 7 CPU and Allow Windows 7 to Stabilize” on page 31
“Step 10. Remove the Front Panel Assembly” on page 32
“Step 11. Inspect and If Required, Replace Front Subpanel” on page 33
N5241/2/4/5/9A Only:
“Step 12. If Required, Rotate the LED Boards and Replace the LED Bias Cables (N5244/5A Only)” on page 34
All instruments:
“Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376)” on page 35
“Step 14. Install the Dress Panel on the New Front Panel Assembly” on page 37
“Step 15. Install the New Overlays on the New Front Panel Assembly” on page 38
“Step 16. Install the New Front Panel Assembly and the Lower Overlay” on page 39
“Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test Set Mother boards (If Necessary)” on page 43
22 Installation Note N5242-90023
Description of this Upgrade Initial Instrument Inspection
“Step 18. Install the New Handler I/O Cable” on page 44
“Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit” on page 45
“Step 20. Reboot the Instrument” on page 46
“Step 21. Reinstall the Outer Covers” on page 47
“Step 22. Install Option N52xxAU-690” on page 48
“Step 23. Changing your PNA’s Computer Name from an A model to a B model” on page 50
“Step 24. Install and Verify Multiple License Files” on page 51
“Step 25. Verify the PNA Analyzer Program is Running with the Correct Options” on page 52
“Step 26. Copy the Unique Files Back to the Analyzer’s SSD” on page 53
“Step 27. Perform Post-Upgrade Adjustments and Calibration” on page 54
“Step 28. Prepare the PNA for the User” on page 55
Installation Note N5242-90023 23
-
Description of this Upgrade Step 1. Copy All Unique Files to an External Disk Drive

Step 1. Copy All Unique Files to an External Disk Drive

The hard disk drive (HDD/SSD) in your analyzer will be replaced as part of this upgrade. Therefore, stored on the HDD/SSD be saved to an external later on the new SSD.
This procedure is written assuming a USB flash memory drive is used, although any external disk
You may skip this step, if you plan to retain the installed SSD for future access. (i.e., To connect your old SSD as a secondary drive, a USB to data or USB to SATA converter cable is required and these are available commercially).
1. Connect the power cord to the analyzer and turn ON the power.
it is necessary that all unique and personal files
disk drive for re-installation
drive will suffice.
2. Insert a USB flash memory drive into a USB port.
3. Open Windows Explorer.
4. Copy all of your files that you want to transfer to the USB drive:
For a list of possible files (and their directory paths) you might want
to copy, refer to: http://na.support.keysight.com/pna/win7.html
5. Remove the USB drive from the USB port.
6. Exit Windows Explorer.
24 Installation Note N5242-90023
Description of this Upgrade Step 2. Update Windows 7 OS Instruments with A.10.40 or above

Step 2. Update Windows 7 OS Instruments with A.10.40 or above

Skip this step, if your PNA is already running firmware A.10.40 or above.
1. If the PNA is running, click File > Exit
2. Go to http://na.support.keysight.com/pna/firmware/firmware.htm and
download the latest PNA firmware.
3. Click on the firmware to be downloaded.
4. Save the program to disk (hard drive of your PC or to your PNA).
5. If necessary, transfer the file from your PC to your PNA using LAN, CD, or
USB Pen drive.
6. Double-click the file on the PNA. Follow the prompts to complete the firmware installation.
7. If the PNA application does not start up automatically, double-click the network analyzer icon on the desktop.
If you have problems with updating the PNA’s firmware, refer to “Getting
Assistance from Keysight” on page 4.
Installation Note N5242-90023 25
-
Description of this Upgrade Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or Greater

Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or Greater

Skip this step, if your PNA is already running firmware A.10.40 or above.
1. After the new firmware is installed, verify the firmware operates without errors.
2. After the instrument powers up, on the toolbar click or press Help > About NA to verify the firmware has been updated to A.10.40.00 or greater.
3. Power down your instrument.
If you have problems with updating the PNA’s firmware, refer to “Getting
Assistance from Keysight” on page 4.
26 Installation Note N5242-90023
Description of this Upgrade Step 4. Remove the outer covers

Step 4. Remove the outer covers

Remove the Outer Cover

1. Turn OFF the analyzer and disconnect the power cord.
2. With a T-20 TORX driver, remove the strap handles (item
the screw (item
3. With a T-20 TORX driver, remove the four rear panel feet (item removing the center
4. Slide the four bottom feet (item
5. Slide the cover off of the frame.
Figure 1 Outer Cover Removal (Instruments Vary—Example Only)
) on
each end of each handle.
screws (item ).
) off the cover.
) by loosening
) by
Installation Note N5242-90023 27
-
Description of this Upgrade Step 5. Remove the Power Supply Cover

Step 5. Remove the Power Supply Cover

1. On the rear panel, remove the screws (x7) and the power supply cover. Save the 0515-0372 (x7) screws for reuse.
2. If you do not have a new Hi-Pot label, then you need to peel off the old Hi-Pot sticker and save for reuse.
Figure 2 Hi-Pot Sticker on the Power Supply Cover

Step 6. Install the *New* Version 7 CPU Assembly

IMPORTANT! If your instrument already has a version 7 CPU, skip to “Step
10. Remove the Front Panel Assembly” on page 32.
Install the new version 7 CPU into the analyzer by referring to Figure 3 on
page 29 and to your instrument’s Service Guide on www.keysight.com.
28 Installation Note N5242-90023
Description of this Upgrade
N5242-60133

Step 7. Install the Service Center’s N5242-60133 Solid-State Disk Drive Assembly (SSDA) with A.10.40 or above—(Windows XP OS Only)

Step 7. Install the Service Center’s N5242-60133 Solid-State Disk Drive Assembly (SSDA) with A.10.40 or above—(Windows XP OS Only)
IMPORTANT! If your instrument already had a Windows 7 operating system and already had a Version 7 CPU prior to this upgrade, skip to “Step 10.
Remove the Front Panel Assembly” on page 32.
Install the service center’s SSDA with A.10.40 or above, into the analyzer by referring Figure 3 and to your instrument’s Service Guide on
www.keysight.com.
If the PNA’s serial number is 4742 or below, the display and touchscreen may not initially work. This is normal. This issue will be fixed when the new front panel is installed.
Figure 3 Service Center’s Solid-State Disk Drive Assembly (SSDA) with A.10.40 or
above
Installation Note N5242-90023 29
-
Description of this Upgrade Step 8. Install the *New* Power Supply Cover

Step 8. Install the *New* Power Supply Cover

Refer to Figure 4 and Figure 5 for this step.
1. Install the new power supply cover by reusing the 0515-0372 (x7) screws.
2. Install the KCC label (N5240-80041) as shown.
Figure 4 Install the Power Supply Cover (N5240-00012) with reused screws (x7)
Figure 5 Install the KCC label (N5240-80041)
30 Installation Note N5242-90023
Description of this Upgrade Step 9. Power Up Your Instrument with the *New* Version 7 CPU and Allow Windows 7 to Stabilize

Step 9. Power Up Your Instrument with the *New* Version 7 CPU and Allow Windows 7 to Stabilize

IMPORTANT! If your instrument already had a Windows 7 operating system and already had a Version 7 CPU prior to this upgrade, skip to “Step 10.
Remove the Front Panel Assembly” on page 32.
If your PNA’s serial number prefix is 4742 or below, the display and touchscreen may not initially work. This is normal. This issue will be fixed when the new front panel is installed.
If the front panel display blanking occurs, to verify your PNA boots up and stabilizes, connect a PC monitor to the rear video output connector and verify the PNA’s behavior.
1. Power up your instrument, allow to a couple of minutes for Windows 7 to stabilize and verify that there are no errors.
2. Power down your instrument.
Installation Note N5242-90023 31
-
Description of this Upgrade Step 10. Remove the Front Panel Assembly

Step 10. Remove the Front Panel Assembly

1. With a 5/16-inch wrench, remove all semi-rigid jumpers (item ) from the front panel. s
2. With a T-10 TORX driver, remove the eight 0515-0244 screws (item
from the sides of the frame. Retain these screws for later use.
Before removing the front panel from the analyzer, lift and support the front of the analyzer chassis.
3. Slide the front panel over the test ports and other front panel connectors.
4. Disconnect the front panel interface ribbon cable (item
is now free from the analyzer.
Figure 6 Front Panel Assembly Removal (Instruments Vary—Example Only— Shown)
). The front panel
➁)
32 Installation Note N5242-90023
Description of this Upgrade Step 11. Inspect and If Required, Replace Front Subpanel

Step 11. Inspect and If Required, Replace Front Subpanel

IMPORTANT! If your front panel has the screw holes shown in Figure 7 on
page 33, skip this step.
STOP! If you have a N5244/5A and it is necessary to replace the front subpanel, you will also need to rotate the LED boards and replace the bias cables. Refer to “Step 12. If Required, Rotate the LED Boards and Replace
the LED Bias Cables (N5244/5A Only)” on page 34.
This section is only required, if the front subpanel is missing the four screw holes as shown in Figure 7. If you need to order a subpanel, refer to “Getting
Assistance from Keysight” on page 4.
If your subpanel is missing the four screw holes indicated in Figure 7, then you need to replace the subpanel. Contact Keysight to order a new subpanel and refer to the Service Guide for your instrument that can be downloaded from
www.keysight.com.
Figure 7 Verifying the front subpanel screw holes
Installation Note N5242-90023 33
-
Description of this Upgrade Step 12. If Required, Rotate the LED Boards and Replace the LED Bias Cables (N5244/5A Only)

Step 12. If Required, Rotate the LED Boards and Replace the LED Bias Cables (N5244/5A Only)

IMPORTANT! If the PNA is a N5244A/45A and it was necessary to replace the front subpanel, you will also need to rotate the LED boards 180 degrees and replace the bias cables. This rework is done at no cost to the customer. Refer to Table 2 on page 13 and to “Getting Assistance from
Keysight” on page 4.
Refer to the Service Guide’s "Removing and Replacing the Front Panel LED
1
Boards" on www.keysight.com replaced at no cost to the customer. Refer to Table 2 on page 13.
1. When installing the new subpanels, remove and reinstall the LED boards after rotating them 180 degrees.
. If not, these boards should be ordered and
2. Replace the old LED board bias cables with the new cables (N5225-60001 (x2)).
1. See “Downloading the Online PNA Service Guide” on page 8.
34 Installation Note N5242-90023
Description of this Upgrade Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376)

Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376)

If your instrument already has the correct W1312-63376 midplane board, skip to “Step 14. Install the Dress Panel on the New Front Panel Assembly”
on page 37.
STOP!!! Please read all content before proceeding with this upgrade procedure. This is a complex procedure that requires Keysight support training, before beginning any repairs. If you have not been properly trained by Keysight support personnel, attempting to do this procedure could result in damage to the midplane board and or the instrument! See also Appendix A: “Midplane Board Upgrade (W1312-63095 to a
W1312-63376),” on page 57.
If you do not have the following equipment, do not attempt to do the process in the Appendix, because you may damage the board or your instrument. Call the factory.
This process requires the following:
— Training by Keysight support personnel to perform this upgrade
— RoHS compliant soldering materials and components
— Variable power soldering iron for surface-mount components
— 10x magnifier or greater
— Else, you may damage your midplane board and or your instrument
This section is only required if the Midplane board is not a W1312-63376 part number. If your midplane board does not require this modification, skip this process and continue to the next step. After this section is completed, your midplane board will be a W1312-63376 midplane board. Refer to Figure 8 on
page 36.
Installation Note N5242-90023 35
-
Description of this Upgrade Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376)
Verify that the sticker has the "W1312-63376" part number. Refer to Figure 8.
If not, and you have read and understand the Caution on page 35, then refer to
Appendix A: “Midplane Board Upgrade (W1312-63095 to a W1312-63376),” on page 57.
If you need to remove your midplane board to verify the part number, refer to the Service Guide for your instrument that can be downloaded from
www.keysight.com.
Figure 8 Verifying the midplane board number part number is: W1312-63376—Top
Side View
36 Installation Note N5242-90023
Description of this Upgrade Step 14. Install the Dress Panel on the New Front Panel Assembly

Step 14. Install the Dress Panel on the New Front Panel Assembly

Install the new lower dress panel onto the new front panel using the 10 flat head screws (0515-1946) provided.
Figure 9 Front Dress Panel Installation
Installation Note N5242-90023 37
-
Description of this Upgrade Step 15. Install the New Overlays on the New Front Panel Assembly

Step 15. Install the New Overlays on the New Front Panel Assembly

IMPORTANT! To avoid possible damage to the lower front panel overlay (label), do not attempt to attach the lower front panel label until “Step 16.
Install the New Front Panel Assembly and the Lower Overlay” on page 39,
step 7.
There should also be a nameplate overlay included with the kit that matches your new analyzer; the same number as the old one except it will be a “B” model. There are multiple overlays provided in the kit; be sure that you use the proper one for your analyzer. If there is not an overlay that matches your analyzer, contact Keysight. Refer to “Getting Assistance from Keysight” on
page 4 of this installation note. Refer to Figure 12 on page 41.
38 Installation Note N5242-90023
Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay

Step 16. Install the New Front Panel Assembly and the Lower Overlay

Before installing the front panel assembly onto the analyzer, place the analyzer right side up and lift and support the front of the analyzer chassis.
1. Reconnect the ribbon cable (item ) to the A1 front panel interface board. Refer to Figure 10.
2. Slide the front panel over the test port connectors. Ensure that the ribbon cable (item
Figure 10 New Front Panel Assembly Installation
) connectors are seated.
Installation Note N5242-90023 39
-
Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay
3. Hand start the twelve screws (item 0515-2044) in the sides of the frame. Use the new screws from the kit and those retained earlier. Do not torque. Refer to Figure 10 on page 39 and Figure 11.
4. Install 0515-1946 (x4) screws into lower front panel. Do not torque. Refer to Figure 11.
Figure 11 New Front Panel Assembly Torque Screws
5. Torque the twelve 0515-2044 screws in the order shown in Figure 10 on
page 39.
6. Torque the two 0515-1946 screws (2-port) or four 0515-1946 screws (4-port) to 9 in-lbs. 4-port is shown, refer to Figure 11.
40 Installation Note N5242-90023
Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay
7. Install each of the overlays as follows (refer to Figure 12):
Remove the protective backing to expose the adhesive.
Carefully align the overlay to its location and gently position it.
Once the overlay is properly positioned, press it firmly into place.
Figure 12 Front Panel Overlays Installation
Installation Note N5242-90023 41
-
Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay
8. If equipped with front panel jumpers, reinstall the semi-rigid jumpers (Figure 13) on the front panel using a 5/16 inch torque wrench set to 8 in-lbs (N5247B only) and 10 in-lbs (N5241/2/4/5/9B).
9. If available, install the jumper covers (not shown).
Figure 13 Install front panel jumpers
42 Installation Note N5242-90023
Description of this Upgrade Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test Set Mother boards (If Necessary)

Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test Set Mother boards (If Necessary)

A) Install the new SPAM V Boards

If your PNA already has the SPAM V boards, skip to next step.
Install the new SPAM V boards by referring to your Service Guide on
www.keysight.com.

B) Install the new Synthesizer Boards

If your PNA already has the correct synthesizer boards, skip to next step.
Install the new synthesizer boards by referring to your Service Guide on
www.keysight.com.

C) Install the new IF Multiplexer Board

If your PNA already has the correct IF multiplexer board, skip to next step.
Install the new IF multiplexer board by referring to your Service Guide on
www.keysight.com.

D) Install the new Test Set Motherboard

Whether you need to replace your TSMB("N524063043") or your PNA already has the correct test set motherboard, skip to next step.
Installation Note N5242-90023 43
-
Description of this Upgrade Step 18. Install the New Handler I/O Cable

Step 18. Install the New Handler I/O Cable

Refer to your instrument Service Guide’s section "Remove and Replace the A19 Test Set Motherboard" on www.keysight.com.
1. Remove the A19 TSMB and be careful to mark all of the cables before removal (i.e., If your TSMB needs to be replaced ("N524063043"), discard this older board.)
2. Remove the old handler I/O cable and discard, it will not be reused.
3. Install the new I/O cable (8121-2919) using (x2) 0515-0372 screws.
Torque to 9 l b-in s.
4. Reinstall the A19 TSMB by reversing the procedure in the Service Guide’s "Remove and Replace the A19 Test Set Motherboard" section.
Figure 14 Install the Handler I/O Cable (8121-2919 and 0515-0372 (x2))
44 Installation Note N5242-90023
Description of this Upgrade Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit
Figure 15 Fold the Handler I/O Cable (8121-2919 and N5242-60005)

Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit

1. Install the new SSDA in the analyzer by reversing the Hard Disk Drive Removal procedure in the Service Guide for your model. Refer to Figure 3
on page 29 and to your instrument’s Service Guide on www.keysight.com.
2. Install the Win 10 sticker (9010-0335) to the rear-panel side of the SSD tray.
Installation Note N5242-90023 45
-
Description of this Upgrade Step 20. Reboot the Instrument

Step 20. Reboot the Instrument

When you power up your instrument there will be many errors, because your instrument does not have any licenses installed. This is as expected and will be corrected in later steps.
IMPORTANT! T
drive that comes with the upgrade kit is first booted, it is as expected that the PNA display is blurry for 10 to 15 seconds, then go blank and reset to a normal screen
.
view
he "B" firmware automatically updates the CPU BIOS. When the
If your display remains blurry beyond ~15 seconds, the PNA BIOS needs to be manually set to 24-bit. (i.e., this display blurriness is indicated by a lavender background and screen text that is difficult to read.) Refer to
Appendix B:, “Manually Setting the PNA Display to 24-Bit”.
1. Power up the instrument and allow it to run for a couple of minutes to stabilize.
2. Verify the instrument has the correct "A" model number and serial number using the embedded Keysight Licensing Manager (KLM) tool on your analyzer: Press Start > All Programs > Keysight Licensing Manager > Keysight Licensing Manager.
This verifies the instrument is working correctly, so the licenses can be loaded. Refer to “Obtain Multiple License Key Files” on page 7.
3. If the instrument stabilizes and has the correct "A" model number and serial number, continue to the next step.
Else, if the unit still has initialization errors which look like problems that cannot be resolved, contact the Keysight Technologies, because the model and serial number need to be updated via the Keysight network. Refer to
“Getting Assistance from Keysight” on page 4.
If your display remains blurry, refer to Appe
ndix B:, “Manually Setting the PNA
Display to 24-Bit”.
If the PNA does not have the correct "A" model number or serial number, it means that somewhere in the process the instrument was initialized with the incorrect model number or serial number. For these cases, connect the PNA to the LAN and contact to the support team for assistance. Refer to
“Getting Assistance from Keysight” on page 4.
46 Installation Note N5242-90023
Description of this Upgrade Step 21. Reinstall the Outer Covers

Step 21. Reinstall the Outer Covers

Reinstall the Outer Cover

1. Slide the cover onto the frame.
2. Using a T-20 TORX driver, install the four rear panel feet (item
installing the center
3. Slide the four bottom feet (item
4. Using a T-20 TORX driver, install the strap handles (item
the screws (item
Figure 16 Outer Cover Reinstallation
screws (item ).
➁)
on both ends.
) by
) onto the cover.
) by tightening
Installation Note N5242-90023 47
-
Description of this Upgrade Step 22. Install Option N52xxAU-690

Step 22. Install Option N52xxAU-690

After completing this process your instrument will be a "B" model Network Analyzer.

Procedure Requirements

— The analyzer must be powered up and operating to perform this procedure.
— Obtain a license key file for installation of this upgrade by following the
instructions on the supplied Software Entitlement Certificate. Refer to
“License Key File Redemption” on page 5.
— On your computer, copy the downloaded license key file to a USB memory
device.

Option Enable Procedure

1. Verify your PNA is powered up.
2. Insert the USB memory device into your PNA’s USB drive slot.
3. Within ~5 seconds the PNA should display a small "New Licenses
Installed" message.
Open a Windows Explorer window(s) and manually move your license file to C:\Program Files\Agilent\licensin g. It may take Keysight License Manager an extra ~30 seconds to enable the license and automatically power down your PNA.
Successful License Instal lation is displayed on the bottom of the window.
If your PNA does not automatically reboot, contact Keysight Technologies. See “Getting Assistance from Keysight” on page 4.
4. If the license file installed and the PNA rebooted as expected, remove your USB memory device.
48 Installation Note N5242-90023
Description of this Upgrade Step 22. Install Option N52xxAU-690

Option Verification Procedure

When you power up your instrument there will be many errors, because your instrument does not and will be corrected in later in this procedure.
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer, click or press to open the KLM software from your PNA’s Windows taskbar by pressing Start > More Programs > Keysight License
Manager folder > Keysight License Manager.
2. If the instrument stabilizes and has the correct "B" model number and
serial number, continue to the next “Step 24. Install and Verify Multiple
License Files” on page 51.
3. Else, proceed to “Step 23. Changing your PNA’s Computer Name from an A
model to a B model” on page 50.
have any licenses installed. This is as expected
If the PNA does not have the correct "B" model number and or serial number, it means that somewhere in the process the instrument was initialized with the incorrect model number or serial number. For these cases, connect the PNA to the LAN. Refer to “Getting Assistance from
Keysight” on page 4.
IMPORTANT! For PNAs that are still A models, proceed to “Step 23.
Changing your PNA’s Computer Name from an A model to a B model” on page 50.
Installation Note N5242-90023 49
-
Description of this Upgrade Step 23. Changing your PNA’s Computer Name from an A model to a B model

Step 23. Changing your PNA’s Computer Name from an A model to a B model

If the PNA does not have a B model after the upgrade, it is necessary to run the
ChangeHostname program to change the PNA from an A model to a B model.
1. Open a Windows Explorer window and navigate to:
C:\Keysight\PNA\tools\ChangeHostName
2. Right click on the ChangeHostname file and click: Run as administrator
The program automatically runs and terminates in ~two seconds (i.e., without any user interaction).
3. To veri fy:
On the toolbar click Help > About NA…
And, verify that the Hostname has the correct B model number.
50 Installation Note N5242-90023
Description of this Upgrade Step 24. Install and Verify Multiple License Files

Step 24. Install and Verify Multiple License Files

After completing this process all of your license files should be loaded into your instrument.

Procedure Requirements

— The analyzer must be powered up and operating to perform this procedure.
— Obtain your multiple license key files by following contacting Keysight
Technologies. Refer to “Obtain Multiple License Key Files” on page 7 and
“Getting Assistance from Keysight” on page 4.
— Use a USB memory device to download your license key file(s) to the top
level on a USB memory device (i.e., do not move the files to a folder).

Option Enable Procedure

1. Verify your PNA is powered up.
2. Insert the USB memory device in your PNA.
3. Locate the email(s) from Keysight which contain license file attachments.
These emails are the result of “License Key File Redemption” on page 5.
4. Copy the license file(s) from the email(s) to a USB flash drive. More than one license file may be copied to the USB flash drive.
A single license file may contain licenses for more than one instrument.
5. Insert the USB storage device into the PNA’s USB drive slot.
Within ~5 seconds the PNA should display a small "New Licenses Installed" message.
If your license did not load automatically via USB memory device, open a Windows Explorer window(s) and manually move your license file to C:\Program Files\Agilent\ licensing. It may take Keysight License Manager an extra ~5 seconds to enable the license.
Successful License Instal lation is displayed on the bottom of the window.
6. Shut down the PNA application and restart.
Optional: Cycle power on the instrument to activate the newly installed option upgrade(s). (Press the power button once to power down the instrument.)
Attempting to re-install a license file that is already installed may generate a “Corrupt Media” error message. Ignore this message.
Installation Note N5242-90023 51
-
Description of this Upgrade Step 25. Verify the PNA Analyzer Program is Running with the Correct Options
7. Disconnect the USB flash drive from the PNA.

Step 25. Verify the PNA Analyzer Program is Running with the Correct Options

If the options have been enabled, skip to “Step 26. Copy the Unique Files
Back to the Analyzer’s SSD” on page 53.

Verify the Option Numbers

Verify that the Options are Enabled
Once the Network Analyzer program is again running:
1. On the analyzer, click or press to open the KLM software from your PNA’s Windows taskbar by pressing Start > More Programs > Keysight License Manager folder > Keysight License Manager and verify the options are
correct.
2. Press OK.
If the options have not been enabled, refer to “Getting Assistance from
Keysight” on page 4.
52 Installation Note N5242-90023
Description of this Upgrade Step 26. Copy the Unique Files Back to the Analyzer’s SSD

Step 26. Copy the Unique Files Back to the Analyzer’s SSD

The files that were previously saved onto a USB flash memory drive must now be installed onto the
— For a list of possible files (and their directory paths), refer to:
http://na.support.keysight.com/pna/win7.html
1. Connect the power cord to the analyzer and turn ON the power.
2. Insert the USB flash memory drive into a USB port.
3. Open Windows Explorer.
4. Navigate to:
Data and state files storage: D:\Support files: c:\programdata\Keysight\network
analyzer
5. After all files have been copied, remove the USB flash memory drive from the USB port.
new SSD.
6. Exit Windows Explorer.
Installation Note N5242-90023 53
-
Description of this Upgrade Step 27. Perform Post-Upgrade Adjustments and Calibration

Step 27. Perform Post-Upgrade Adjustments and Calibration

Adjustments

The following adjustments must be made due to the hardware changes of the analyzer.
— EE default adjustment (Only, if the synthesizer board is changed)
— synthesizer Bandwidth adjustment (Only, if the synthesizer board is
changed)
—source adjustment
— IF gain adjustment
—receiver adjustment
— receiver characterization
— IF response adjustment (Option S93090xA, S93092A, S93093A, or S93094A
Only)
— Noise Adjustment (Option 029 with S93029A Only)
These adjustments are described in the PNA Service Guide and in the PNA on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide
After the specified adjustments have been performed, the analyzer should operate and phase lock over its entire frequency range.
Else, all adjustments must be run per the instrument’s Service Guide on
www.keysight.com.

Operator’s Check

Perform the Operator’s Check to check the basic functionality of the analyzer. For instructions, refer to the “Tests & Adjustments” chapter of the Service Guide.
If you experience difficulty with the basic functioning of the analyzer, contact Keysight. Refer to “Getting Assistance from Keysight” on page 4.

Calibration

1
.
Although the analyzer functions, its performance relative to its specifications has not been verified. It is recommended that a full instrument calibration be performed using the N7840A performance test software or the instrument’s embedded performance tests. Refer to the analyzer’s service guide for information on the performance test software.
1. See “Downloading the Online PNA Service Guide” on page 8.
54 Installation Note N5242-90023
Description of this Upgrade Step 28. Prepare the PNA for the User

Step 28. Prepare the PNA for the User

1. If necessary, reinstall front jumper cables.
2. Install the cable guards, pushing them over the front jumper cables until
the cushioning material touches the front panel of the PNA.
3. Install the dust caps on the test ports.
4. Install labels included in kit as required (e.g., KCC, two person Caution).
5. Clean the analyzer, as needed, using a damp cloth.
Installation Note N5242-90023 55
-
Description of this Upgrade Step 28. Prepare the PNA for the User
56 Installation Note N5242-90023
Keysight Licensed Funcitionality Upgrade Kit
N52xxA Microwave Network Analyzers
Installation Note
A: Midplane Board Upgrade (W1312-63095 to a
W1312-63376)
STOP!!! Please read all content before proceeding with this upgrade procedure. This is a complex procedure that requires Keysight support training, before beginning any repairs! If you have not been properly trained by Keysight support personnel, attempting to do this procedure could result in damage to the midplane board and or the instrument! See also “Step 13. Inspect or Modify Midplane Board (W1312-63095 to a
W1312-63376)” on page 35.
If you do not have the following equipment, do not attempt to the process in the Appendix, because you may damage the board and or your instrument. Stop and return the instrument to Keysight for repair. Refer to
“Getting Assistance from Keysight”.
This process requires the following:
— Training by Keysight support personnel to perform this upgrade
— RoHS compliant soldering materials and components
Variable power soldering iron for surface-mount components
— 10x magnifier or greater
— Else, you may damage your midplane board and or your instrument
This section is only required if the Midplane board is not a W1312-63376 part number. If your midplane board does not require this modification, skip this process and continue to the “Step 14. Install the Dress Panel on the New Front
Panel Assembly” on page 37. After this section is completed, your midplane
board will be a W1312-63376 midplane board. Refer to Figure 22 on page 64.
57
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Table 1 Parts List for Midplane Board Upgrade Kit Modification
Part number Description
0699-1429 Resistor U403—383 ohm (attach to pin 5)
8150-3431 30 gauge wire (0.5 inch)
5184-4489 Label-Brady blank label
0470-0954 Loctite 495
a. The factory will provide these items upon request.
a
58 N5242-90023 Installation Guide
Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure

Procedure

1. To remove your midplane board, refer to the Service Guide for your instrument that can be downloaded from www.keysight.com.
2. Cut pin 5 on U403, so that half of the lead is left near on the IC. Refer to
Figure 17.
Figure 17 Step 1 Cut pin 5 half way up as shown—Top Side View
N5242-90023 Installation Guide 59
Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure
3. Clean pad and remove cut off portion of U403–pin 5 from the pad. Refer to
Figure 18.
Figure 18 Cleaning the pad for pin 5—Top Side View
60 N5242-90023 Installation Guide
Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure
4. Add 383 ohm resistor as shown. Refer to Figure 19.
Figure 19 Add 383 ohm resistor
N5242-90023 Installation Guide 61
Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure
5. Add a 1/2 inch 30 gauge wire. Refer to Figure 20.
Figure 20 Add a 1/2 inch 30 gauge wire—Top Side View
62 N5242-90023 Installation Guide
Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure
6. Add glue as shown to hold the wire in place. Refer to Figure 21 on
page 63.
Figure 21 Add glue as shown to hold the wire in place—Top Side View
N5242-90023 Installation Guide 63
Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure
7. Add the W1312-63376 sticker. Verify that the sticker does not interfere with the PCA screw holes. Refer to Figure 22.
Figure 22 Adding sticker to new board number W1312-63376—Top Side View
8. Re-install the midplane board per the Service Guide for your instrument that can be downloaded from www.keysight.com.
64 N5242-90023 Installation Guide
Keysight Licensed Funcitionality Upgrade Kit
N52xxA Microwave Network Analyzers
Installation Note
B: Manually Setting the PNA Display to 24-Bit
This section is only required if, after installing the new SSD that comes with this upgrade kit (refer to “Step 19. Install the *NEW* N5242-60135
Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit” on page 45),
your PNA reboots and does not automatically adjust the display to 24-bit. If your PNA does not have a blurry display, skip this process and continue to the
“Step 21. Reinstall the Outer Covers” on page 47.

Procedure

In this procedure, you will need to navigate through the PNA’s BIOS menu, even though you are unable to clearly read the menus.
1. During the bootup of your PNA: Press F7 on bootup until the BIOS screen opens.
2. In the window that opens, the top row of the screen will have various menu selections: Main, Advanced, Chipset, Boot, Security, and Save&Exit.
Use the arrow keys to move the selection from Main over to Chipset, then press Enter.
3. In the window that opens, the first of three rows should be have white text which says Graphics Configuration. With this row highlighted, press Enter.
4. In the window that opens, use the down arrow to move the selected text to the bottom row: LCD Control. Press Enter.
5. In the window that opens, use the down arrow to move the selected text to the bottom row: Panel Color Depth. Press Enter.
6. At this point, a small blue box with two items pops up. The top one is "18 bit", the bottom one is "24 bit". Select this bottom one and press Enter.
7. Now press F10 (to Save & Exit).
65
Manually Setting the PNA Display to 24-Bit Procedure
66 N5242-90023 Installation Guide
Installation Note Xxxxx-xxxxx 3
This information is subject to change without notice. © Keysight Technologies 2007-2019
Edition 1, September 2019