No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
N5242-90023
Edition
Edition 1, September 2019
Supersedes: August 9, 2019
Printed in USA/Malaysia
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
This upgrade converts your N52xxA microwave network analyzer or N5264A
microwave receiver to an equivalent N52xxB or N5264B. This is done by
removing the existing front panel assembly, outer covers, and installing a new
front panel assembly, SSDA, and outer covers.
As part of this upgrade, there are other assemblies that need to be verified,
because the boards must be of a certain type to be compatible with the B
model upgrade. Refer to “Initial Instrument Inspection” on page 18.
After installation of this upgrade, your analyzer will be a “B” model rather than
an “A” model.
STOP!This repair must be done at a service center! Refer to “Getting
Assistance from Keysight” on page 4.
IMPORTANT! As part of the upgrade, the service center needs a copy of
the customer’s Keysight License files that were sent to them from
Keysight. Refer to “License Key File Redemption” on page 5.
IMPORTANT! N5244/5A PNAs with H29 cannot upgrade to N5244/5B.
Included in the contents of this kit is a Software Entitlement Certificate
instructions on how to redeem it. Do NOT attempt to redeem this
until you have read “License Key File Redemption” on page 5.
For additional information on using the Keysight Software Manager and
redeeming a Software Entitlement Certificate, refer to
Description of this Upgrade
Getting Assistance from Keysight
Getting Assistance from Keysight
Installing this upgrade kit requires special skills and experience. If you think
you may not be qualified to do the work, or need advice, contact Keysight.
If You Have Problems With the Upgrade Kit Contents
Keysight stands behind the quality of the upgrade kit contents. If you have
problems with any item in the kit, refer to www.keysight.com and the Contact
Keysight () link.
4 Installation Note N5242-90023
Description of this Upgrade
Getting Prepared
Getting Prepared
License Key File Redemption
IMPORTANT!As part of the upgrade, the service center needs to
request a the customer’s Keysight License files from Keysight. Refer to
“License Key File Redemption” on page 5.
IMPORTANT! Also, you will need to contact the Keysight factory support
node to request multiple license key files. Refer to “Obtain Multiple
License Key Files” on page 7.
Ensure that you are connected to an external server, before attempting to
download your email and license key file.
The enclosed Software Entitlement Certificate is a receipt, verifying that
you have purchased a licensed option for the analyzer of your choice. You
must now use a Keysight Web page to request a license key file for the
instrument that will receive the option.
To upgrade model(s), you must request an Option 690 license key file from the
Keysight Software Manager:
http://www.keysight.com/find/softwaremanager:
To complete the request, you will need to gather the following information:
— From the certificate
— Order number
— Certificate number
— From your instrument
(Instrument information is available in the network analyzer - on the
toolbar, click Help, then click About Network Analyzer.)
— Model number
— Serial number
Using the information just gathered, you must request license key file from the
Keysight Software Manager:
http://www.keysight.com/find/softwaremanager.
The following Keysight Software Manager (KSM) procedure has been added—
but is abbreviated—for users that are familiar with the KSM process and has
steps that are specific to the A model to B model upgrade (Option 690).
1. Enter your login and password or setup a login as required.
Installation Note N5242-90023 5
-
Description of this Upgrade
Getting Prepared
2. In the Keysight License Manager welcome window, click You can add a
new certificate.
3. In the "Keysight Software Manager: Add Certificate" window that opens, in
the "Enter your certificate information" section, refer to your Software
License Certificate that you received with your upgrade kit and enter the
following:
— Order Number
— Certificate Number
4. Click Continue (to return to the Keysight Software Manager welcome
screen).
5. If there are multiple "A" model PNAs that are being upgraded to a "B"
model, repeat steps 3 and 4.
Else, continue to next step.
6. In the Keysight Software Manager welcome screen, click "You can request
new licenses".
7. In the "Keysight Software Manager: Request Licenses" window that opens,
in the "Select the host to assign licenses to" section, click the Click to add
or select a host drop-down.
8. Click Add a new License Notifier Host ID host...
9. In the License Notifier Host ID: box that is displayed, enter your
instrument’s model and serial number, separated by only a comma—no
spaces). Example:
— New License Host Notifier Host ID host: N5242A,MY12345678
10. In the Select the licenses to assign section, scroll down until you see a
Option 690 upgrade with your instrument’s "A" and "B" model values—
similar to this example—and then click the box to select:
— N5247AU-690 Upgrade N5247A to N5247B
IMPORTANT: To avoid the loss of an option license, ensure that you only
have a single Option 690 upgrade checked.
11. Click Assign Licenses
12. In the "Assigned licenses:" section that opens, listing all of the License
Notifier Host IDs and associated instrument information, verify the
following:
— License Notifier Host ID: contains the correct "B" model and serial
number
— the associated N52xxAU-690Upgrade lists the correct "A" model to
"B" model
6 Installation Note N5242-90023
Description of this Upgrade
Getting Prepared
If any of the information listed in the "Assigned licenses:" section is
incorrect, click Remove next to the incorrect License Notifier Host ID: or
N52xxAU-690 Upgrade to remove them from the "Assigned Licenses" list.
13. If there are multiple "A" model PNAs that are being upgraded to a "B"
model, repeat steps 7 through 12.
Else, continue to next step.
14. Once you have verified the License Notifier Host ID & N52xxAU-690
Option models are correct, click Continue > (an email is sent to you with
an Option 690 license file for each specific License Notifier Host ID’d PNA).
15. Keysight will promptly email your license key file(s) to your profile’s email
address.
Obtain Multiple License Key Files
All of the licenses in your instrument need to be converted to a new format for
your "B" model instrument. Contact Keysight Technologies factory support
early in your upgrade process to receive your licenses in a timely manner. See
“Getting Assistance from Keysight” on page 4.
You will need to provide Keysight with the foll
owing information:
—model number
— serial number
— email address of the technician performing this upgrade (i.e., to send the
license key files)
Installation Note N5242-90023 7
-
Description of this Upgrade
Protecting Your Workspace from Electrostatic Discharge
Downloading the Online PNA Service Guide
To view the Service Guide for your PNA model number, use the following steps:
1. Go to www.keysight.com.
2. In the Search box, enter the model number of the analyzer (Ex: N5242B)
and click Search.
3. Click Technical Support > Document Library > Manuals or under Refine
Your Results:
4. Click Manuals > Service Manual.
5. Click the service guide title to download the PDF file.
6. When the PDF of the Service Guide is displayed, scroll through the
Contents section bookmarks to locate the information needed.
Protecting Your Workspace from Electrostatic Discharge
For information, click on the Chapter 1 bookmark, "Electrostatic Discharge
Protection" in the PDF of your instrument’s Service Guide on
1
www.keysight.com
.
ESD Equipment Required for the Installation
DescriptionKeysight Part Number
ESD grounding wrist strap9300-1367
5-ft grounding cord for wrist strap9300-0980
2 x 4 ft conductive table mat and 15-ft grounding wire9300-0797
ESD heel strap (for use with conductive floors)9300-1308
1. See “Downloading the Online PNA Service Guide” on page 8.
8 Installation Note N5242-90023
Description of this Upgrade
Tools Required for the Installation
Tools Required for the Installation
DescriptionQty
T-10 TORX driver - set to 9 in-lbs (1.02 N.m)1
T-20 TORX driver - set to 21 in-lbs (2.38 N.m)1
5/16-in (8 mm) nutsetter or open end torque wrench - set to 10 in-lbs (1.13 N.m)1
5/16-in (8 mm) nutsetter or open end torque wrench - set to 21 in-lbs (2.38 N.m)1
9 mm nutsetter or open end torque wrench - set to 21 in-lbs) (2.38 N.m)1
1/4-in (6 mm) open end wrench1
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front and rear panel bulkhead connectors. On these, use a
9 mm nutsetter or open end torque wrench set to 21 in-lb.
About Installing the Upgrade
Products affectedAll N524xA microwave network analyzers and all N5264A
Installation to be performed byKeysight service center, personnel qualified by Keysight, or
Estimated installation time4 hours
Estimated adjustment time1.5 hours
Estimated full instrument calibration time 4.5 to 8.0 hours
microwave receivers
customer
Installation Note N5242-90023 9
-
Description of this Upgrade
Items Included in the Upgrade Kit
Items Included in the Upgrade Kit
Check the contents of your kit against the following list. If any part is missing or
damaged, contact Keysight Technologies. Refer to “Description of this
Upgrade” on page 3.
This section applies to multiple models and upgrade kits. Please refer to
the model and upgrade kit for your instrument’s upgrade.
Extra quantities of items such as protective plastic caps, screws, cable ties,
and cable clamps may be included in this upgrade kit. It is normal for some
of these items to remain unused after the upgrade is completed.
Table 1 Contents of Upgrade Kit N5242-60117 for N5241AU-690 (N5241B),
N5242AU-690 (N5242B), and N5249AU-690 (N5249B)
QtyDescription Part Number
1Installation Note (this document)N5242-90023
1Software Entitlement Certificate5964-5145
2Front handle5023-3074
4Rear foot5041-7903
4Bottom foot5041-7906
2Trim strip, side, phantom gray5041-7908
1Caution label5190-1799
1Win10 IoT Ent LTSB 2016 Multi-Lang OEI Value9010-0335
1China RoHS Addendum9320-6722
1Front frame – (assembly)N5240-60071
4Machine screw, M4.0 x 25 pan head (To attach rear feet to chassis.)0515-1619
14
10 Installation Note N5242-90023
Machine screw, M3.0 x 10 mm flat head (To attach lower dress panel
chassis).
5Machine screw, M4.0 x 10 flat head (To attach front frame to chassis.)0515-2044
1Cable Assembly–Flat Ribbon 230 mm LG - Handler I/O cable (LFE)8121-2919
1Solid-state drive assembly (SSDA), imaged, for 2.2 GHz CPU and Win 10N5242-60135
1Lower front panel overlay, 2-port (N5241/2/9B Options 201/219)
Lower front panel overlay, 2-port (N5241/2/9B Option 224)
Lower front panel overlay, 4-port (N5241/2/9B Option 401/419 /423)
Lower front panel overlay, 2-port (N5241/2/9B Option 219 with Option 029) N5242-80026
overlay frame to
0515-1946
N5222-80010
N5242-80024
N5222-80013
Description of this Upgrade
Items Included in the Upgrade Kit
Table 1 Contents of Upgrade Kit N5242-60117 for N5241AU-690 (N5241B),
N5242AU-690 (N5242B), and N5249AU-690 (N5249B)
QtyDescription Part Number
Lower front panel overlay, 2-port (N5241/2/9B Option 224 with Option 029) N5242-80027
Lower front panel overlay, 4-port (N5241/2/9B Option 419/423 with Option 029) N5242-80028
Lower front panel overlay, 2-port (N5241/2/9B Option 217) N5242-80029
Lower front panel overlay, 2-port (N5241/2/9B Option 222) N5242-80030
Lower front panel overlay, 4-port (N5241/2/9B Option 417 or 422) N5242-80031
Lower front panel overlay, 2-port (N5241/2/9B Option 217 with Option 029)N5242-80032
Lower front panel overlay, 2-port (N5241/2/9B Option 222 with Option 029)N5242-80033
Lower front panel overlay, 4-port (N5241/2/9B Option 417/422 with Option 029)N5242-80034
1Nameplate, N5241BN5241-80002
Nameplate, N5242BN5242-80022
Nameplate, N5249BN5249-80003
Lower Dress Panel, 2-Port (ALL PNA models). Goes on, before lower front
1
overlay. Attach with 0515-1946 subpanel screws.
Lower Dress Panel, 4-Port (ALL PNA models). Goes on, before lower front
overlay. Attach with 0515-1946 subpanel screws.
Not Included - (For Reference Only) - External Hard ware and Miscellaneous Parts, All Options
Machine screw M5.0 x 18, flat head (To attach strap handle.)0515-0710
Hole plug (When analyzer is rack mounted.)N9020-40007
Installation Note N5242-90023 17
-
Description of this Upgrade
Initial Instrument Inspection
Initial Instrument Inspection
STOP! This repair must be done at a service center! Refer to “Getting
Assistance from Keysight” on page 4.
This section and all steps must be completed before proceeding to the
following sections or customer data may be lost or instrument repair
delayed!
The purpose of this section is to verify where the instrument should begin the
model conversion process. This section verifies that the instrument has all of
the required board assemblies for a B model upgrade.
If your instrument does not have all of the required board assemblies, contact
the customer, to provide a quote on parts and labor. Upon customer approval,
order up all of the new upgrade kits and boards, before the "B" model upgrade
can be completed.
Refer to Table 6 on page 19 and then to “Verifying the Board Assemblies” on
page 20.
18 Installation Note N5242-90023
Description of this Upgrade
Initial Instrument Inspection
Table 6 Overview of Assembly Requirements for B Model Upgrade
AssemblyPart NumberSerial Number for PNAs that start using
these boards (PNAs with these serial
prefixes do not need to be upgraded)
CPU, version 7W1312-60211MY/US5540 and aboveN8989A-PC7
SPAM 5N5240-60078
For 2-port PNAs, also order x1
N5247-60023 and x1
N5247-60024
IF Multiplexer,
version B
Synthesizers,
versions E, F, or G
N5240-60062MY/US5540 and aboveN889A-MX2
N5240-60076 (single board) or
N5242-60168 (3 board kit)
MY/US5200 and aboveN8989A-SP5
MY/US5120 and aboveN8989A-SR6
Can be
ordered
using the
part number,
or:
2-port
PNAs
without
option 224
need two
synthesizer
boards.
4-port
PNAs or
2-port
option 224
PNAs need
three
synthesizer
boards.
Test Set
Motherboard –
N5245-60163MY/US4940 and above––––
N5241/2A only
Installation Note N5242-90023 19
-
Description of this Upgrade
Initial Instrument Inspection
Verifying the Board Assemblies
The following board assemblies need to be verified and installed during the
upgrade:
1. Press Help > About NA...
2. In the About Network Analyzer window that opens, make a note of the
following:
— System CPU Version
— The system CPU must be a version 7 or greater. This CPU
is not included in this kit. Please contact Keysight if you
need to replace the CPU and order CPU board number:
W1312-60211.
— DSP Version
— The DSP (SPAM) board must be version 5 or greater. The
SPAM V boards are not included in this kit. Please contact
Keysight if you need to replace the SPAM boards and order
N5240-60077 SPAM V. If you have a 2-Port instrument,
you will also need to order one each of cable numbers:
N5247-60023 and N5247-60024.
3. Press Utility > System > Service > Diagnostics > EEPROM Headers
— Verify the LO Synth, Src1 Synth, and Src2 Synth boards part number
is either "N524063074" or "N524263150". The synthesizer boards
are not included in this kit. Please contact Keysight if you need to
replace the synthesizer boards and order N5240-60076.
4. Remove the instrument outer covers and if necessary the power supply, to
verify the Midplane Board’s part number reads: "W1312-63376". Refer to
the Service Guide on www.keysight.com for removing the power supply.
If the midplane board’s part number does not read "W1312-63376", the
Midplane Board needs to be modified. Refer to “Step 13. Inspect or Modify
Midplane Board (W1312-63095 to a W1312-63376)” on page 35.
5. N5244/5/A models (ONLY): If the subpanels need to be changed, the LED
boards need to be rotated 180 degrees and the LED cables replaced (refer
to “Step 11. Inspect and If Required, Replace Front Subpanel” on page 33).
Regardless, ensure that LED boards are installed correctly and the LED
cables (N5225-60001 (x2)) are the correct part numbers. Refer to “Step
12. If Required, Rotate the LED Boards and Replace the LED Bias Cables
(N5244/5A Only)” on page 34 and to the Service Guide’s "Removing and
Replacing the Front Panel LED Boards" on www.keysight.com. If not, these
parts should be ordered and replaced at no cost to the customer. Refer to
Table 2 on page 13.
20 Installation Note N5242-90023
Description of this Upgrade
Initial Instrument Inspection
6. If your instrument has the minimum criteria mentioned above or you have
received the required parts proceed to “Step 1. Copy All Unique Files to an
External Disk Drive” on page 24.
Overview of the Installation Procedure
Windows XP OS only: Before installing the new SSDA with Windows 7 OS—
that is part of this upgrade
≥A.10.40.00 must be loaded
initializes the instrument to a known
This section assumes you have completed the “Initial Instrument
Inspection” section and verified, that all of the board assemblies meet the
requirements for this A to B model upgrade. If not, refer to the “Initial
Instrument Inspection” on page 18.
kit—the service center’s SSDA with firmware
and initialized in the instrument. This
stable state.
Installation Note N5242-90023 21
-
Description of this Upgrade
Initial Instrument Inspection
All instruments:
“Step 1. Copy All Unique Files to an External Disk Drive” on page 24
Windows 7 Operating System (Windows 7 OS) only:
“Step 2. Update Windows 7 OS Instruments with A.10.40 or above” on
page 25
“Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or Greater” on
page 26
All instruments:
“Step 4. Remove the outer covers” on page 27
“Step 5. Remove the Power Supply Cover” on page 28
“Step 6. Install the *New* Version 7 CPU Assembly” on page 28
Windows XP Operating System (Windows XP OS) only:
“Step 7. Install the Service Center’s N5242-60133 Solid-State Disk Drive
Assembly (SSDA) with A.10.40 or above—(Windows XP OS Only)” on page 29
All instruments:
“Step 8. Install the *New* Power Supply Cover” on page 30
“Step 9. Power Up Your Instrument with the *New* Version 7 CPU and Allow
Windows 7 to Stabilize” on page 31
“Step 10. Remove the Front Panel Assembly” on page 32
“Step 11. Inspect and If Required, Replace Front Subpanel” on page 33
N5241/2/4/5/9A Only:
“Step 12. If Required, Rotate the LED Boards and Replace the LED Bias
Cables (N5244/5A Only)” on page 34
All instruments:
“Step 13. Inspect or Modify Midplane Board (W1312-63095 to a
W1312-63376)” on page 35
“Step 14. Install the Dress Panel on the New Front Panel Assembly” on
page 37
“Step 15. Install the New Overlays on the New Front Panel Assembly” on
page 38
“Step 16. Install the New Front Panel Assembly and the Lower Overlay” on
page 39
“Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test Set
Mother boards (If Necessary)” on page 43
22 Installation Note N5242-90023
Description of this Upgrade
Initial Instrument Inspection
“Step 18. Install the New Handler I/O Cable” on page 44
“Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly
(SSDA) from the Upgrade Kit” on page 45
“Step 20. Reboot the Instrument” on page 46
“Step 21. Reinstall the Outer Covers” on page 47
“Step 22. Install Option N52xxAU-690” on page 48
“Step 23. Changing your PNA’s Computer Name from an A model to a B
model” on page 50
“Step 24. Install and Verify Multiple License Files” on page 51
“Step 25. Verify the PNA Analyzer Program is Running with the Correct
Options” on page 52
“Step 26. Copy the Unique Files Back to the Analyzer’s SSD” on page 53
“Step 27. Perform Post-Upgrade Adjustments and Calibration” on page 54
“Step 28. Prepare the PNA for the User” on page 55
Installation Note N5242-90023 23
-
Description of this Upgrade
Step 1. Copy All Unique Files to an External Disk Drive
Step 1. Copy All Unique Files to an External Disk Drive
The hard disk drive (HDD/SSD) in your analyzer will be replaced as part of
this upgrade. Therefore,
stored on the HDD/SSD be saved to an external
later on the new SSD.
This procedure is written assuming a USB flash memory drive is used,
although any external disk
You may skip this step, if you plan to retain the installed SSD for future
access. (i.e., To connect your old SSD as a secondary drive, a USB to data
or USB to SATA converter cable is required and these are available
commercially).
1. Connect the power cord to the analyzer and turn ON the power.
it is necessary that all unique and personal files
disk drive for re-installation
drive will suffice.
2. Insert a USB flash memory drive into a USB port.
3. Open Windows Explorer.
4. Copy all of your files that you want to transfer to the USB drive:
— For a list of possible files (and their directory paths) you might want
to copy, refer to: http://na.support.keysight.com/pna/win7.html
5. Remove the USB drive from the USB port.
6. Exit Windows Explorer.
24 Installation Note N5242-90023
Description of this Upgrade
Step 2. Update Windows 7 OS Instruments with A.10.40 or above
Step 2. Update Windows 7 OS Instruments with A.10.40 or above
Skip this step, if your PNA is already running firmware A.10.40 or above.
1. If the PNA is running, click File > Exit
2. Go to http://na.support.keysight.com/pna/firmware/firmware.htm and
download the latest PNA firmware.
3. Click on the firmware to be downloaded.
4. Save the program to disk (hard drive of your PC or to your PNA).
5. If necessary, transfer the file from your PC to your PNA using LAN, CD, or
USB Pen drive.
6. Double-click the file on the PNA. Follow the prompts to complete the
firmware installation.
7. If the PNA application does not start up automatically, double-click the
network analyzer icon on the desktop.
If you have problems with updating the PNA’s firmware, refer to “Getting
Assistance from Keysight” on page 4.
Installation Note N5242-90023 25
-
Description of this Upgrade
Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or Greater
Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or
Greater
Skip this step, if your PNA is already running firmware A.10.40 or above.
1. After the new firmware is installed, verify the firmware operates without
errors.
2. After the instrument powers up, on the toolbar click or press Help > About
NA to verify the firmware has been updated to A.10.40.00 or greater.
3. Power down your instrument.
If you have problems with updating the PNA’s firmware, refer to “Getting
Assistance from Keysight” on page 4.
26 Installation Note N5242-90023
Description of this Upgrade
Step 4. Remove the outer covers
Step 4. Remove the outer covers
Remove the Outer Cover
1. Turn OFF the analyzer and disconnect the power cord.
2. With a T-20 TORX driver, remove the strap handles (item
the screw (item
3. With a T-20 TORX driver, remove the four rear panel feet (item
removing the center
Description of this Upgrade
Step 5. Remove the Power Supply Cover
Step 5. Remove the Power Supply Cover
1. On the rear panel, remove the screws (x7) and the power supply cover.
Save the 0515-0372 (x7) screws for reuse.
2. If you do not have a new Hi-Pot label, then you need to peel off the old
Hi-Pot sticker and save for reuse.
Figure 2 Hi-Pot Sticker on the Power Supply Cover
Step 6. Install the *New* Version 7 CPU Assembly
IMPORTANT! If your instrument already has a version 7 CPU, skip to “Step
10. Remove the Front Panel Assembly” on page 32.
Install the new version 7 CPU into the analyzer by referring to Figure 3 on
page 29 and to your instrument’s Service Guide on www.keysight.com.
28 Installation Note N5242-90023
Description of this Upgrade
N5242-60133
Step 7. Install the Service Center’s N5242-60133 Solid-State Disk Drive Assembly
(SSDA) with A.10.40 or above—(Windows XP OS Only)
Step 7. Install the Service Center’s N5242-60133 Solid-State Disk
Drive Assembly (SSDA) with A.10.40 or above—(Windows XP OS
Only)
IMPORTANT! If your instrument already had a Windows 7 operating system
and already had a Version 7 CPU prior to this upgrade, skip to “Step 10.
Remove the Front Panel Assembly” on page 32.
Install the service center’s SSDA with A.10.40 or above, into the analyzer by
referring Figure 3 and to your instrument’s Service Guide on
www.keysight.com.
If the PNA’s serial number is 4742 or below, the display and touchscreen
may not initially work. This is normal. This issue will be fixed when the new
front panel is installed.
Figure 3 Service Center’s Solid-State Disk Drive Assembly (SSDA) with A.10.40 or
above
Installation Note N5242-90023 29
-
Description of this Upgrade
Step 8. Install the *New* Power Supply Cover
Step 8. Install the *New* Power Supply Cover
Refer to Figure 4 and Figure 5 for this step.
1. Install the new power supply cover by reusing the 0515-0372 (x7) screws.
2. Install the KCC label (N5240-80041) as shown.
Figure 4 Install the Power Supply Cover (N5240-00012) with reused screws (x7)
Figure 5 Install the KCC label (N5240-80041)
30 Installation Note N5242-90023
Description of this Upgrade
Step 9. Power Up Your Instrument with the *New* Version 7 CPU and Allow Windows 7
to Stabilize
Step 9. Power Up Your Instrument with the *New* Version 7 CPU
and Allow Windows 7 to Stabilize
IMPORTANT! If your instrument already had a Windows 7 operating system
and already had a Version 7 CPU prior to this upgrade, skip to “Step 10.
Remove the Front Panel Assembly” on page 32.
If your PNA’s serial number prefix is 4742 or below, the display and
touchscreen may not initially work. This is normal. This issue will be fixed
when the new front panel is installed.
If the front panel display blanking occurs, to verify your PNA boots up and
stabilizes, connect a PC monitor to the rear video output connector and
verify the PNA’s behavior.
1. Power up your instrument, allow to a couple of minutes for Windows 7 to
stabilize and verify that there are no errors.
2. Power down your instrument.
Installation Note N5242-90023 31
-
Description of this Upgrade
Step 10. Remove the Front Panel Assembly
Step 10. Remove the Front Panel Assembly
1. With a 5/16-inch wrench, remove all semi-rigid jumpers (item ➀) from the
front panel. s
2. With a T-10 TORX driver, remove the eight 0515-0244 screws (item
from the sides of the frame. Retain these screws for later use.
Before removing the front panel from the analyzer, lift and support the
front of the analyzer chassis.
3. Slide the front panel over the test ports and other front panel connectors.
4. Disconnect the front panel interface ribbon cable (item
is now free from the analyzer.
Figure 6 Front Panel Assembly Removal (Instruments Vary—Example Only— Shown)
③). The front panel
➁)
32 Installation Note N5242-90023
Description of this Upgrade
Step 11. Inspect and If Required, Replace Front Subpanel
Step 11. Inspect and If Required, Replace Front Subpanel
IMPORTANT! If your front panel has the screw holes shown in Figure 7 on
page 33, skip this step.
STOP! If you have a N5244/5A and it is necessary to replace the front
subpanel, you will also need to rotate the LED boards and replace the bias
cables. Refer to “Step 12. If Required, Rotate the LED Boards and Replace
the LED Bias Cables (N5244/5A Only)” on page 34.
This section is only required, if the front subpanel is missing the four screw
holes as shown in Figure 7. If you need to order a subpanel, refer to “Getting
Assistance from Keysight” on page 4.
If your subpanel is missing the four screw holes indicated in Figure 7, then you
need to replace the subpanel. Contact Keysight to order a new subpanel and
refer to the Service Guide for your instrument that can be downloaded from
www.keysight.com.
Figure 7 Verifying the front subpanel screw holes
Installation Note N5242-90023 33
-
Description of this Upgrade
Step 12. If Required, Rotate the LED Boards and Replace the LED Bias Cables
(N5244/5A Only)
Step 12. If Required, Rotate the LED Boards and Replace the LED
Bias Cables (N5244/5A Only)
IMPORTANT! If the PNA is a N5244A/45A and it was necessary to replace
the front subpanel, you will also need to rotate the LED boards 180
degrees and replace the bias cables. This rework is done at no cost to the
customer. Refer to Table 2 on page 13 and to “Getting Assistance from
Keysight” on page 4.
Refer to the Service Guide’s "Removing and Replacing the Front Panel LED
1
Boards" on www.keysight.com
replaced at no cost to the customer. Refer to Table 2 on page 13.
1. When installing the new subpanels, remove and reinstall the LED boards
after rotating them 180 degrees.
. If not, these boards should be ordered and
2. Replace the old LED board bias cables with the new cables (N5225-60001
(x2)).
1. See “Downloading the Online PNA Service Guide” on page 8.
34 Installation Note N5242-90023
Description of this Upgrade
Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376)
Step 13. Inspect or Modify Midplane Board (W1312-63095 to a
W1312-63376)
If your instrument already has the correct W1312-63376 midplane board,
skip to “Step 14. Install the Dress Panel on the New Front Panel Assembly”
on page 37.
STOP!!! Please read all content before proceeding with this upgrade
procedure. This is a complex procedure that requires Keysight support
training, before beginning any repairs. If you have not been properly
trained by Keysight support personnel, attempting to do this procedure
could result in damage to the midplane board and or the instrument! See
also Appendix A: “Midplane Board Upgrade (W1312-63095 to a
W1312-63376),” on page 57.
If you do not have the following equipment, do not attempt to do the
process in the Appendix, because you may damage the board or your
instrument. Call the factory.
This process requires the following:
— Training by Keysight support personnel to perform this upgrade
— RoHS compliant soldering materials and components
— Variable power soldering iron for surface-mount components
— 10x magnifier or greater
— Else, you may damage your midplane board and or your instrument
This section is only required if the Midplane board is not a W1312-63376 part
number. If your midplane board does not require this modification, skip this
process and continue to the next step. After this section is completed, your
midplane board will be a W1312-63376 midplane board. Refer to Figure 8 on
page 36.
Installation Note N5242-90023 35
-
Description of this Upgrade
Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376)
Verify that the sticker has the "W1312-63376" part number. Refer to Figure 8.
If not, and you have read and understand the Caution on page 35, then refer to
Appendix A: “Midplane Board Upgrade (W1312-63095 to a W1312-63376),”
on page 57.
If you need to remove your midplane board to verify the part number, refer to
the Service Guide for your instrument that can be downloaded from
www.keysight.com.
Figure 8 Verifying the midplane board number part number is: W1312-63376—Top
Side View
36 Installation Note N5242-90023
Description of this Upgrade
Step 14. Install the Dress Panel on the New Front Panel Assembly
Step 14. Install the Dress Panel on the New Front Panel Assembly
Install the new lower dress panel onto the new front panel using the 10 flat
head screws (0515-1946) provided.
Figure 9 Front Dress Panel Installation
Installation Note N5242-90023 37
-
Description of this Upgrade
Step 15. Install the New Overlays on the New Front Panel Assembly
Step 15. Install the New Overlays on the New Front Panel Assembly
IMPORTANT! To avoid possible damage to the lower front panel overlay
(label), do not attempt to attach the lower front panel label until “Step 16.
Install the New Front Panel Assembly and the Lower Overlay” on page 39,
step 7.
There should also be a nameplate overlay included with the kit that matches
your new analyzer; the same number as the old one except it will be a “B”
model. There are multiple overlays provided in the kit; be sure that you use the
proper one for your analyzer. If there is not an overlay that matches your
analyzer, contact Keysight. Refer to “Getting Assistance from Keysight” on
page 4 of this installation note. Refer to Figure 12 on page 41.
38 Installation Note N5242-90023
Description of this Upgrade
Step 16. Install the New Front Panel Assembly and the Lower Overlay
Step 16. Install the New Front Panel Assembly and the Lower
Overlay
Before installing the front panel assembly onto the analyzer, place the
analyzer right side up and lift and support the front of the analyzer chassis.
1. Reconnect the ribbon cable (item ➁) to the A1 front panel interface board.
Refer to Figure 10.
2. Slide the front panel over the test port connectors. Ensure that the ribbon
cable (item
Figure 10 New Front Panel Assembly Installation
➁) connectors are seated.
Installation Note N5242-90023 39
-
Description of this Upgrade
Step 16. Install the New Front Panel Assembly and the Lower Overlay
3. Hand start the twelve screws (item 0515-2044) in the sides of the frame.
Use the new screws from the kit and those retained earlier. Do not torque.
Refer to Figure 10 on page 39 and Figure 11.
4. Install 0515-1946 (x4) screws into lower front panel. Do not torque. Refer
to Figure 11.
Figure 11 New Front Panel Assembly Torque Screws
5. Torque the twelve 0515-2044 screws in the order shown in Figure 10 on
page 39.
6. Torque the two 0515-1946 screws (2-port) or four 0515-1946 screws
(4-port) to 9 in-lbs. 4-port is shown, refer to Figure 11.
40 Installation Note N5242-90023
Description of this Upgrade
Step 16. Install the New Front Panel Assembly and the Lower Overlay
7. Install each of the overlays as follows (refer to Figure 12):
— Remove the protective backing to expose the adhesive.
— Carefully align the overlay to its location and gently position it.
— Once the overlay is properly positioned, press it firmly into place.
Figure 12 Front Panel Overlays Installation
Installation Note N5242-90023 41
-
Description of this Upgrade
Step 16. Install the New Front Panel Assembly and the Lower Overlay
8. If equipped with front panel jumpers, reinstall the semi-rigid jumpers
(Figure 13) on the front panel using a 5/16 inch torque wrench set to
8 in-lbs (N5247B only) and 10 in-lbs (N5241/2/4/5/9B).
9. If available, install the jumper covers (not shown).
Figure 13 Install front panel jumpers
42 Installation Note N5242-90023
Description of this Upgrade
Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test Set Mother boards
(If Necessary)
Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test
Set Mother boards (If Necessary)
A) Install the new SPAM V Boards
If your PNA already has the SPAM V boards, skip to next step.
Install the new SPAM V boards by referring to your Service Guide on
www.keysight.com.
B) Install the new Synthesizer Boards
If your PNA already has the correct synthesizer boards, skip to next step.
Install the new synthesizer boards by referring to your Service Guide on
www.keysight.com.
C) Install the new IF Multiplexer Board
If your PNA already has the correct IF multiplexer board, skip to next step.
Install the new IF multiplexer board by referring to your Service Guide on
www.keysight.com.
D) Install the new Test Set Motherboard
Whether you need to replace your TSMB("N524063043") or your PNA already
has the correct test set motherboard, skip to next step.
Installation Note N5242-90023 43
-
Description of this Upgrade
Step 18. Install the New Handler I/O Cable
Step 18. Install the New Handler I/O Cable
Refer to your instrument Service Guide’s section "Remove and Replace the A19
Test Set Motherboard" on www.keysight.com.
1. Remove the A19 TSMB and be careful to mark all of the cables before
removal (i.e., If your TSMB needs to be replaced ("N524063043"), discard
this older board.)
2. Remove the old handler I/O cable and discard, it will not be reused.
3. Install the new I/O cable (8121-2919) using (x2) 0515-0372 screws.
Torque to 9 l b-in s.
4. Reinstall the A19 TSMB by reversing the procedure in the Service Guide’s
"Remove and Replace the A19 Test Set Motherboard" section.
Figure 14 Install the Handler I/O Cable (8121-2919 and 0515-0372 (x2))
44 Installation Note N5242-90023
Description of this Upgrade
Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from
the Upgrade Kit
Figure 15 Fold the Handler I/O Cable (8121-2919 and N5242-60005)
Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive
Assembly (SSDA) from the Upgrade Kit
1. Install the new SSDA in the analyzer by reversing the Hard Disk Drive
Removal procedure in the Service Guide for your model. Refer to Figure 3
on page 29 and to your instrument’s Service Guide on www.keysight.com.
2. Install the Win 10 sticker (9010-0335) to the rear-panel side of the SSD
tray.
Installation Note N5242-90023 45
-
Description of this Upgrade
Step 20. Reboot the Instrument
Step 20. Reboot the Instrument
When you power up your instrument there will be many errors, because
your instrument does not have any licenses installed. This is as expected
and will be corrected in later steps.
IMPORTANT! T
drive that comes with the upgrade kit is first booted, it is as expected that the PNA
display is blurry for 10 to 15 seconds, then go blank and reset to a normal screen
.
view
he "B" firmware automatically updates the CPU BIOS. When the
If your display remains blurry beyond ~15 seconds, the PNA BIOS needs to
be manually set to 24-bit. (i.e., this display blurriness is indicated by a
lavender background and screen text that is difficult to read.) Refer to
Appendix B:, “Manually Setting the PNA Display to 24-Bit”.
1. Power up the instrument and allow it to run for a couple of minutes to
stabilize.
2. Verify the instrument has the correct "A" model number and serial number
using the embedded Keysight Licensing Manager (KLM) tool on your
analyzer: Press Start > All Programs > Keysight Licensing Manager > Keysight Licensing Manager.
This verifies the instrument is working correctly, so the licenses can be
loaded. Refer to “Obtain Multiple License Key Files” on page 7.
3. If the instrument stabilizes and has the correct "A" model number and
serial number, continue to the next step.
Else, if the unit still has initialization errors which look like problems that
cannot be resolved, contact the Keysight Technologies, because the model
and serial number need to be updated via the Keysight network. Refer to
“Getting Assistance from Keysight” on page 4.
If your display remains blurry, refer to Appe
ndix B:, “Manually Setting the PNA
Display to 24-Bit”.
If the PNA does not have the correct "A" model number or serial number, it
means that somewhere in the process the instrument was initialized with
the incorrect model number or serial number. For these cases, connect the
PNA to the LAN and contact to the support team for assistance. Refer to
“Getting Assistance from Keysight” on page 4.
46 Installation Note N5242-90023
Description of this Upgrade
Step 21. Reinstall the Outer Covers
Step 21. Reinstall the Outer Covers
Reinstall the Outer Cover
1. Slide the cover onto the frame.
2. Using a T-20 TORX driver, install the four rear panel feet (item
installing the center
3. Slide the four bottom feet (item
4. Using a T-20 TORX driver, install the strap handles (item
the screws (item
Figure 16 Outer Cover Reinstallation
screws (item ➃).
➁)
on both ends.
③) by
➄) onto the cover.
➀) by tightening
Installation Note N5242-90023 47
-
Description of this Upgrade
Step 22. Install Option N52xxAU-690
Step 22. Install Option N52xxAU-690
After completing this process your instrument will be a "B" model Network
Analyzer.
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— Obtain a license key file for installation of this upgrade by following the
instructions on the supplied Software Entitlement Certificate. Refer to
“License Key File Redemption” on page 5.
— On your computer, copy the downloaded license key file to a USB memory
device.
Option Enable Procedure
1. Verify your PNA is powered up.
2. Insert the USB memory device into your PNA’s USB drive slot.
3. Within ~5 seconds the PNA should display a small "New Licenses
Installed" message.
Open a Windows Explorer window(s) and manually move your license file
to C:\Program Files\Agilent\licensin g. It may take Keysight
License Manager an extra ~30 seconds to enable the license and
automatically power down your PNA.
Successful License Instal lation is displayed on the bottom of the
window.
If your PNA does not automatically reboot, contact Keysight Technologies.
See “Getting Assistance from Keysight” on page 4.
4. If the license file installed and the PNA rebooted as expected, remove your
USB memory device.
48 Installation Note N5242-90023
Description of this Upgrade
Step 22. Install Option N52xxAU-690
Option Verification Procedure
When you power up your instrument there will be many errors, because
your instrument does not
and will be corrected in later in this procedure.
Once the analyzer has restarted and the Network Analyzer program is again
running:
1. On the analyzer, click or press to open the KLM software from your PNA’s
Windows taskbar by pressing Start > More Programs > Keysight License
Manager folder > Keysight License Manager.
2. If the instrument stabilizes and has the correct "B" model number and
serial number, continue to the next “Step 24. Install and Verify Multiple
License Files” on page 51.
3. Else, proceed to “Step 23. Changing your PNA’s Computer Name from an A
model to a B model” on page 50.
have any licenses installed. This is as expected
If the PNA does not have the correct "B" model number and or serial
number, it means that somewhere in the process the instrument was
initialized with the incorrect model number or serial number. For these
cases, connect the PNA to the LAN. Refer to “Getting Assistance from
Keysight” on page 4.
IMPORTANT! For PNAs that are still A models, proceed to “Step 23.
Changing your PNA’s Computer Name from an A model to a B model” on
page 50.
Installation Note N5242-90023 49
-
Description of this Upgrade
Step 23. Changing your PNA’s Computer Name from an A model to a B model
Step 23. Changing your PNA’s Computer Name from an A model to
a B model
If the PNA does not have a B model after the upgrade, it is necessary to run the
ChangeHostname program to change the PNA from an A model to a B model.
1. Open a Windows Explorer window and navigate to:
C:\Keysight\PNA\tools\ChangeHostName
2. Right click on the ChangeHostname file and click:
Run as administrator
The program automatically runs and terminates in ~two seconds (i.e.,
without any user interaction).
3. To veri fy:
On the toolbar click Help > About NA…
And, verify that the Hostname has the correct B model number.
50 Installation Note N5242-90023
Description of this Upgrade
Step 24. Install and Verify Multiple License Files
Step 24. Install and Verify Multiple License Files
After completing this process all of your license files should be loaded into your
instrument.
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— Obtain your multiple license key files by following contacting Keysight
Technologies. Refer to “Obtain Multiple License Key Files” on page 7 and
“Getting Assistance from Keysight” on page 4.
— Use a USB memory device to download your license key file(s) to the top
level on a USB memory device (i.e., do not move the files to a folder).
Option Enable Procedure
1. Verify your PNA is powered up.
2. Insert the USB memory device in your PNA.
3. Locate the email(s) from Keysight which contain license file attachments.
These emails are the result of “License Key File Redemption” on page 5.
4. Copy the license file(s) from the email(s) to a USB flash drive.
More than one license file may be copied to the USB flash drive.
A single license file may contain licenses for more than one instrument.
5. Insert the USB storage device into the PNA’s USB drive slot.
Within ~5 seconds the PNA should display a small "New Licenses
Installed" message.
If your license did not load automatically via USB memory device, open a
Windows Explorer window(s) and manually move your license file to
C:\Program Files\Agilent\ licensing. It may take Keysight License
Manager an extra ~5 seconds to enable the license.
Successful License Instal lation is displayed on the bottom of the
window.
6. Shut down the PNA application and restart.
Optional: Cycle power on the instrument to activate the newly installed
option upgrade(s). (Press the power button once to power down the
instrument.)
Attempting to re-install a license file that is already installed may generate
a “Corrupt Media” error message. Ignore this message.
Installation Note N5242-90023 51
-
Description of this Upgrade
Step 25. Verify the PNA Analyzer Program is Running with the Correct Options
7. Disconnect the USB flash drive from the PNA.
Step 25. Verify the PNA Analyzer Program is Running with the
Correct Options
If the options have been enabled, skip to “Step 26. Copy the Unique Files
Back to the Analyzer’s SSD” on page 53.
Verify the Option Numbers
Verify that the Options are Enabled
Once the Network Analyzer program is again running:
1. On the analyzer, click or press to open the KLM software from your PNA’s
Windows taskbar by pressing Start > More Programs > Keysight License
Manager folder > Keysight License Manager and verify the options are
correct.
2. Press OK.
If the options have not been enabled, refer to “Getting Assistance from
Keysight” on page 4.
52 Installation Note N5242-90023
Description of this Upgrade
Step 26. Copy the Unique Files Back to the Analyzer’s SSD
Step 26. Copy the Unique Files Back to the Analyzer’s SSD
The files that were previously saved onto a USB flash memory drive must
now be installed onto the
— For a list of possible files (and their directory paths), refer to:
http://na.support.keysight.com/pna/win7.html
1. Connect the power cord to the analyzer and turn ON the power.
2. Insert the USB flash memory drive into a USB port.
3. Open Windows Explorer.
4. Navigate to:
— Data and state files storage: D:\
— Support files: c:\programdata\Keysight\network
analyzer
5. After all files have been copied, remove the USB flash memory drive from
the USB port.
new SSD.
6. Exit Windows Explorer.
Installation Note N5242-90023 53
-
Description of this Upgrade
Step 27. Perform Post-Upgrade Adjustments and Calibration
Step 27. Perform Post-Upgrade Adjustments and Calibration
Adjustments
The following adjustments must be made due to the hardware changes of the
analyzer.
— EE default adjustment (Only, if the synthesizer board is changed)
— synthesizer Bandwidth adjustment (Only, if the synthesizer board is
changed)
—source adjustment
— IF gain adjustment
—receiver adjustment
— receiver characterization
— IF response adjustment (Option S93090xA, S93092A, S93093A, or S93094A
Only)
— Noise Adjustment (Option 029 with S93029A Only)
These adjustments are described in the PNA Service Guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide
After the specified adjustments have been performed, the analyzer should
operate and phase lock over its entire frequency range.
Else, all adjustments must be run per the instrument’s Service Guide on
www.keysight.com.
Operator’s Check
Perform the Operator’s Check to check the basic functionality of the analyzer.
For instructions, refer to the “Tests & Adjustments” chapter of the Service
Guide.
If you experience difficulty with the basic functioning of the analyzer, contact
Keysight. Refer to “Getting Assistance from Keysight” on page 4.
Calibration
1
.
Although the analyzer functions, its performance relative to its specifications
has not been verified. It is recommended that a full instrument calibration be
performed using the N7840A performance test software or the instrument’s
embedded performance tests. Refer to the analyzer’s service guide for
information on the performance test software.
1. See “Downloading the Online PNA Service Guide” on page 8.
54 Installation Note N5242-90023
Description of this Upgrade
Step 28. Prepare the PNA for the User
Step 28. Prepare the PNA for the User
1. If necessary, reinstall front jumper cables.
2. Install the cable guards, pushing them over the front jumper cables until
the cushioning material touches the front panel of the PNA.
3. Install the dust caps on the test ports.
4. Install labels included in kit as required (e.g., KCC, two person Caution).
5. Clean the analyzer, as needed, using a damp cloth.
Installation Note N5242-90023 55
-
Description of this Upgrade
Step 28. Prepare the PNA for the User
56 Installation Note N5242-90023
Keysight Licensed Funcitionality Upgrade Kit
N52xxA Microwave Network Analyzers
Installation Note
A:Midplane Board Upgrade (W1312-63095 to a
W1312-63376)
STOP!!! Please read all content before proceeding with this upgrade
procedure. This is a complex procedure that requires Keysight support
training, before beginning any repairs! If you have not been properly
trained by Keysight support personnel, attempting to do this procedure
could result in damage to the midplane board and or the instrument! See
also “Step 13. Inspect or Modify Midplane Board (W1312-63095 to a
W1312-63376)” on page 35.
If you do not have the following equipment, do not attempt to the process
in the Appendix, because you may damage the board and or your
instrument. Stop and return the instrument to Keysight for repair. Refer to
“Getting Assistance from Keysight”.
This process requires the following:
— Training by Keysight support personnel to perform this upgrade
— RoHS compliant soldering materials and components
— Variable power soldering iron for surface-mount components
— 10x magnifier or greater
— Else, you may damage your midplane board and or your instrument
This section is only required if the Midplane board is not a W1312-63376 part
number. If your midplane board does not require this modification, skip this
process and continue to the “Step 14. Install the Dress Panel on the New Front
Panel Assembly” on page 37. After this section is completed, your midplane
board will be a W1312-63376 midplane board. Refer to Figure 22 on page 64.
57
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Table 1 Parts List for Midplane Board Upgrade Kit Modification
Part numberDescription
0699-1429Resistor U403—383 ohm (attach to pin 5)
8150-343130 gauge wire (0.5 inch)
5184-4489Label-Brady blank label
0470-0954Loctite 495
a. The factory will provide these items upon request.
a
58 N5242-90023 Installation Guide
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Procedure
Procedure
1. To remove your midplane board, refer to the Service Guide for your
instrument that can be downloaded from www.keysight.com.
2. Cut pin 5 on U403, so that half of the lead is left near on the IC. Refer to
Figure 17.
Figure 17 Step 1 Cut pin 5 half way up as shown—Top Side View
N5242-90023 Installation Guide 59
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Procedure
3. Clean pad and remove cut off portion of U403–pin 5 from the pad. Refer to
Figure 18.
Figure 18 Cleaning the pad for pin 5—Top Side View
60 N5242-90023 Installation Guide
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Procedure
4. Add 383 ohm resistor as shown. Refer to Figure 19.
Figure 19 Add 383 ohm resistor
N5242-90023 Installation Guide 61
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Procedure
5. Add a 1/2 inch 30 gauge wire. Refer to Figure 20.
Figure 20 Add a 1/2 inch 30 gauge wire—Top Side View
62 N5242-90023 Installation Guide
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Procedure
6. Add glue as shown to hold the wire in place. Refer to Figure 21 on
page 63.
Figure 21 Add glue as shown to hold the wire in place—Top Side View
N5242-90023 Installation Guide 63
Midplane Board Upgrade (W1312-63095 to a W1312-63376)
Procedure
7. Add the W1312-63376 sticker. Verify that the sticker does not interfere
with the PCA screw holes. Refer to Figure 22.
Figure 22 Adding sticker to new board number W1312-63376—Top Side View
8. Re-install the midplane board per the Service Guide for your instrument
that can be downloaded from www.keysight.com.
64 N5242-90023 Installation Guide
Keysight Licensed Funcitionality Upgrade Kit
N52xxA Microwave Network Analyzers
Installation Note
B:Manually Setting the PNA Display to 24-Bit
This section is only required if, after installing the new SSD that comes with
this upgrade kit (refer to “Step 19. Install the *NEW* N5242-60135
Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit” on page 45),
your PNA reboots and does not automatically adjust the display to 24-bit. If
your PNA does not have a blurry display, skip this process and continue to the
“Step 21. Reinstall the Outer Covers” on page 47.
Procedure
In this procedure, you will need to navigate through the PNA’s BIOS menu,
even though you are unable to clearly read the menus.
1. During the bootup of your PNA: Press F7 on bootup until the BIOS screen
opens.
2. In the window that opens, the top row of the screen will have various menu
selections: Main, Advanced,Chipset,Boot,Security, and Save&Exit.
Use the arrow keys to move the selection from Main over to Chipset, then
press Enter.
3. In the window that opens, the first of three rows should be have white text
which says Graphics Configuration. With this row highlighted, press Enter.
4. In the window that opens, use the down arrow to move the selected text to
the bottom row: LCD Control. Press Enter.
5. In the window that opens, use the down arrow to move the selected text to
the bottom row: Panel Color Depth. Press Enter.
6. At this point, a small blue box with two items pops up. The top one is "18
bit", the bottom one is "24 bit". Select this bottom one and press Enter.
7. Now press F10 (to Save & Exit).
65
Manually Setting the PNA Display to 24-Bit
Procedure