Note: All the above is the general dissembling course of 40-6065, and the assembling course of 40-6065 is
just the contrary. While repairing the instrument, it should take the disassembling level, procedure and
method according to the practical situation.
2. Accuracy Check and Calibration
This function works only when the angle is in “UNLCOK” state.
3
Page 4
2.1 Accuracy check of horizontal angle
Reference platform setting
As shown in figure 6, set a reference platform that has four flat and smooth sides and is 10m away from the wall.
Side A is horizontal, and its horizontal angle is 0°±10", Side C is parallel with and below side A, with the horizontal
angle is 0°±1°. Side B and side D are both perpendicular to side A, and the included angles between side A and side
B or side A and side D are both 90°±1°.
Wall
Side B Side D
Side A
Side C
Figure 6
0 degree check
(1) Put the instrument against the side A, as shown in figure 7. Install into the battery, and press the first key (from
left to right), i.e. ON/OFF key. Then, notice the LCD display (left), wait 10 seconds till the displayed data stable and
record the angle value.
Figure 7
(2) As shown in figure 8, rotate the instrument by 180
degree in A plane, wait 10 seconds again till the displayed data
stable then record the second angle value.
Figure 8
180 degree check
(1) As shown in figure 9, place the instrument against C plane and power on, notice the LCD display (left), wait 10
seconds till the displayed data stable and record the angle value.
Figure 9
(2) As shown in figure 10, rotate the instrument by 180
degree in C plane, wait 10 seconds again till the displayed
data stable then record the second angle value.
Figure 10
90 degree check
(1) As shown in figure 11, place the instrument against B plane and power on, notice the LCD display (left), wait 10
seconds till the displayed data stable and record the angle value.
(2) As shown in figure 12, rotate the instrument by 180 degree in B plane, wait 10 seconds again till the displayed
data stable then record the second angle value.
4
Page 5
Figure 11 Figure 12 Figure 13 Figure 14
270 degree check
(1) (1) As shown in figure 13, place the instrument against D plane and power on, notice the LCD display (left), wait
10 seconds till the displayed data stable and record the angle value.
(2) As shown in figure 14, rotate the instrument by 180 degree in D plane, wait 10 seconds again till the displayed
data stable then record the second angle value.
In the above four characteristic angle checks, if the difference between two angle values is over 0.1o, it is
necessary to recalibrate the instrument. After calibration, please check the four reference angles again till
they completely complies with the accuracy requirements
2.2 Accuracy calibration of horizontal angle
Please make calibration for the four angles 0°,90°,180°,270°simultaneously.
0 degree calibration
(1) Place the instrument according to the step 1 of 0 degree check (as shown in figure 7). Power on and press “Zero
calibration” key once, i.e. the third key (from left to right) with a “CAL “ sign. Then the LCD will display “-0-“ (figure 15)
Figure 15
(2) Wait ten seconds and then press again the “Zero calibration” key, the LCD will display “-1-“ (figure 16);
Figure 16
(3) Place the instrument according to the step 2 of 0 degree check (as shown in figure 8), wait 10 seconds and press
the “Zero calibration” key again, the LCD will display “-2-” (as shown in figure 17). After 10 seconds, the calibration is
completed and the LCD will display the current angle value. If it is failure to perform calibration, it means the
horizontal-angle sensor is defective. Please replace it.
Figure 17
The calibration methods of other reference angles are the same as the above. While calibrating, the placement of
instrument should refer to the placement as shown in their checks.
5
Page 6
2.3 Accuracy check and calibration of laser spot
(1) Mark a reference point on the wall, and its horizontal accuracy should not be less than 5″.
(2) Power on the instrument, and observe the distance from the laser spot to the reference point. If the vertical
distance error from the laser spot to the reference point is out of ±3mm range, or the horizontal distance error from
the laser spot to the reference point is out of ±20mm range, it means that it is necessary to recalibrate the accuracy
of the laser spot.
(3) While calibrating, unclench the laser cover (3-3#), and adjust the 3pcs crosshead plate bolts M2.5×8(3-6#) until
it comply with the accuracy requirement, as shows in figure 18.
Bolt
Figure 18
2.4 Accuracy check of included angle
Check of 180°included angle
Setting a reference platform of 180 degree
A plane
Figure 19
As it shows in figure 19, set a reference platform with a flat and smooth A plane. A plane should be long enough to
ensure that the angle square board could be unfolded entirely but not exceed the range of A plane.
Unfold the angle square board entirely and place the instrument against A plane, as shows in figure 20. Then, power
on the instrument, notice the LCD display (right), and wait 10 seconds till the displayed data stable.
Check of 0°included angle
Fold the angle square board, as shown in figure 21. Then, power on the instrument, notice the LCD display (right),
and wait 10 seconds till the displayed data stable.
6
Figure 20
Figure 21
Page 7
Check of other included angles
Reference platform setting
Baseboard
Figure 22
As shown in figure 22, set a reference baseboard and several reference blocking boards, which should be flat and
smooth enough. The included angle between two adjacent blocking boards is 10°. The included angle between the
baseboard and the blocking board could reach a maximum of 170° and a minimum of 10°.
Check of other included angles (take 30°included angle as an example)
Figure 23
Make the mainframe of the instrument against the baseboard. Turn the angle square board to the blocking board
that forms an included angle of 30 degree with the baseboard, as shown in figure 23. Power on the instrument,
notice the LCD display (right), and wait 10 seconds till the displayed data stable.
As for the included angle of 0°, 90° or 180°, if the accuracy of the display angle is out of ±0.1° range by
comparing with the reference angle, or as for other included angles, if the accuracy of the display angle is
out of ±0.2° range by comparing with the reference angle, then it is necessary to recalibrate the instrument.
After calibration, please check again until it complies with the accuracy requirement.
The check of other included angles is similar to the above, and it just needs to turn the angle square board
to the required blocking board. Any angle that is integral multiple of 10 degree should be checked.
2.5 Accuracy calibration of the included angle
Calibrati on of 180° included angle
(1) In the state as shown in figure 20, if you find the angle is beyond tolerance, it is necessary to recalibrate the
instrument.
(2) Tear out the membrane (1-3#)
(3) Power on the instrument, and press the switch 1, as shown in figure 24.
(4) Press the switch 2 or 3 again, adjust the reading displaying on LCD (right) until it displays “-18-”.
(5) Press the switch 1 again, and wait for 10 seconds.
(6) Finally, after the buzzer gives a beep, press the switch 4 and observe whether the included-angle reading
displaying on LCD (right) is correct or not. If it is not correct, please recalibrate the instrument until the reading
becomes correct. If the reading is still wrong after recalibrating, it means the included angle sensor is defective.
Please replace it.
7
Page 8
Figure 24
Calibration of 0° included angle
In the state as shown in figure 21, if you find the angle is beyond tolerance, it is necessary to recalibrate the
instrument.
The detailed calibration procedures are the same as the step 2-6 described in the 180° included angle calibration.
The only difference is that in the calibration of 0°included angle,, the reading displaying on the LCD should be set
as “-0-”.
Calibration of other included angles (take the included angle of 30 degree as an example):
In the state as shown in figure 23, if you find the angle is beyond tolerance, please make recalibration for the
instrument. The detailed calibration procedures are the same as the step 2-6 described in the calibration for included
angle of 180 degree. The only difference is that in the calibration for 30°included angle, the reading displaying on
the LCD should be set as “-3-”.
The calibration of other included angles is similar to the above. Any angle that is integral multiple of 10°
should be calibrated.
In the calibration of 10°included angle, the reading displaying on the LCD will be set as “-1-”.
In the calibration of 20°included angle, the reading displaying on the LCD will be set as “-2-”.
And by analogy--------
3. Electric connection drawing
103
0. 1U
0. 1U
105
105
100P
103
103
184
103
104
103
153
4MHz
104
104
104
184
103
103
202
106
106
106
103
103
104
104
104
104
Included angle sensor
3
1
3
1
8
1
夹角
传感器
水平
角度
传感器
L
D
BAT
8
Figure 25
Horizontal-angle sensor
Page 9
4. Solutions for usual malfunctions
No. Malfunction phenomenon Cause Repair method
1 The battery is empty Replace with the new batteries
2 The switch is loose contact Replace the switch
Failure to power on
3 The power line is cut off Weld the leading wire
4 The connection line is loose. Reinsert the connection line tightly
5
Defective 1# main circuit board Replace the 1# main circuit board
6 The switch is loose contact Replace the switch
7 The connection line is loose Reinsert the connection line tightly
No laser is output
8 The connection line is cut off Re-weld the connection line
9 Defective 1# main circuit board Replace 1# main circuit board
10
11
No horizontal angle display
12
13
Defective light source Replace the light source
The connection line of the
horizontal-angle sensor is loose
Defective horizontal-angle
sensor
Reinsert the connection line tightly
Replace horizontal-sensor sensor
Defective 1# main circuit board Replace the 1# main circuit board
After alignment, the
accuracy of the horizontal
14
angle is still beyond
Defective horizontal-angle
sensor
Replace the horizontal-angle sensor
tolerance
15 The switch is loose contact Replace the switch
No sound indication
16 Defective buzzer Replace the buzzer
17
18
Defective 1# main circuit board Replace 1# main circuit board
The connection line of the
included angle sensor is loose
Reinsert the connection line tightly
19 Defective included angle sensor Replace the included angle sensor
No included angle display
20 Defective 1# main circuit board Replace 1# main circuit board
21 Defective electrical brush Replace the electrical brush
22
After alignment, the included
23 Defective electrical brush Replace the electrical brush
angle is still beyond
24
tolerance
The electrical brush failures to
slide freely on the potentiometer
Defective included angle sensor Replace the included angle sensor
Reassembly