Invacare RPS 440ee Instruction Manual

Invacare® RPS 440ee
Instruction Manual
Version 07/08
DEALER: This manual MUST be given to the user of the patient lift. USER: BEFORE using this patient lift, read this manual and save for future reference. SERVICE AGENT: Only Invacare authorised service agents are permitted to install,
service or repair Invacare product.
!
Safety Precautions
Please read and understand the following before using a Invacare patient lifter:
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THE WARNINGS, CAUTIONS AND INSTRUCTIONS IN THIS MANUAL AND THE OWNER'S MANUAL PROVIDED WITH THE SLINGS. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR TECHNICAL PERSONNEL IF APPLICABLE BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE INJURY OR DAMAGE MAY RESULT.
RADIO FREQUENCY INTERFERENCE
Most electronic equipment is influenced by Radio Frequency Interference (RFI). CAUTION should be exercised with regard to the use of portable communication equipment in the area around such equipment. If RFI causes erratic behaviour, PUSH the Red Emergency Stop Button IMMEDIATELY. DO NOT turn the Red Emergency Stop Button OFF while transmission is in progress. This product meets EN60601-1-2 Collateral Standard: Electromagnetic Compatibility.
WARNING notices as used in this manual apply to hazards or unsafe practices which could result in serious bodily harm.
CAUTION notices as used in this manual apply to hazards or unsafe practices which could result in minor personal injury or property damage.
NOTES highlight procedures and contain information which assist the operator in understanding the information contained in this manual.
MAINTENANCE
Maintenance MUST be performed ONLY by qualified personnel.
SAVE THESE INSTRUCTIONS AND KEEP WITH PATIENT LIFT AT ALL TIMES.
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE.
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
2
3
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
RPS 440ee User's Manual
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Special Notes
!
!
ALWAYS be aware of the Lift Arms. Injury to the patient and/or assistant may occur.
ALWAYS be aware of the Footrest, especially the patient’s position on the footrest. Injury to the patient and/or assistant may occur.
Device contains Lead Acid Gel Cell batteries. DO NOT dispose of batteries in normal household waste. They MUST be taken to a proper disposal site. Contact your local waste management company for information.
3
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Contents
Special Notes ..........................................................2
Specifications ..........................................................4
Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Key to Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Operating the Stand Up Lift .............................................12
Locking/Unlocking the Rear Swivel Casters ....................................12
Raising/Lowering Stand Up Lift ..............................................12
Closing/Opening the Legs of the Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Mounting Battery Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Charging the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Transferring From .....................................................15
Lifting Preparation ........................................................15
Lifting the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Moving the Patient .......................................................16
Transferring To ........................................................17
Transferring the Patient to a Wheelchair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Transferring the Patient to a Bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Transferring the Patient to a Commode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Lifter Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Troubleshooting .......................................................22
Care and Maintenance of your Stand Up Lifter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Lubrication ..............................................................23
Wear And Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Electric Actuator Replacement ..............................................24
Base Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Log Book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4
5
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
RPS 440ee User's Manual
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Specifications
Hoisting Range/Reach
1. Hoisting reach at max. height of C.S.P ....................300
2. Flexible device
3. C.S.P (Central Suspension Point)
4. Spreader bar
5. Lifting machinery
6. Hoisting range ............................................................... 630
7. Max height of C.S.P ....................................................1700
8. Min height of C.S.P .....................................................1070
9. Height of C.S.P at max. height hoisting reach ........NA
10. Hoisting reach at min. height of C.S.P ..................... NA
11. Max hoisting reach .......................................................NA
12. Base
Base (Radial type)
1. Max external length ...................................................1140
2. Max internal length ......................................................820
3. Min internal width ........................................................ 575
4. Min external width .......................................................675
5. Max internal width .....................................................1180
6. Max external width ....................................................1270
7. Base
Base height/clearance
1. Base
2. Min clearance................................................................... 70
3. Max height ......................................................................120
Model RPS440
Product Code LPS-RPS440-AUS
Capacity 200 kg (440 lb)
Leg Spread Adjustment Electric
Configuration
Frame Material Epoxy Coated Steel
Castors Single Wheel
Size 100mm (4")
Brakes Yes - Rear
Product Weight 52 kg (115 lb)
Battery 24VDC
Charger Input 100 - 240 VAC
Charger Output/Charging Time 29.5 VDC - Max. 6 hours
Approx. Lifts per Charge* 100 - 200 cycles
Safety Devices
Motor Anti-Entrapment Yes
Emergence Stop Yes
Emergence Lowering Yes
Hoisting Range/Reach Base Base height/clearance
5
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Safety Summary
NOTE: Check all parts for shipping damage before using. In case of damage, DO NOT use the equipment. Contact the Carrier/Dealer for further instructions.
NOTE: The stand up lift may be operated by one (1) healthcare professional for ALL lifting preparation, transferring from and transferring to procedures with a cooperative, weight-bearing individual able to support the majority of his/her own weight. However, since medical conditions vary, Invacare recommends that the healthcare professional evaluate the need for
assistance and determine whether more than one (1) assistant is appropriate in each case to safely perform the transfer.
DO NOT attempt any transfer without approval of the patient’s physician, nurse or medical assistant. Thoroughly read the instructions in this Owner’s Manual, observe a trained team of experts perform the lifting procedures and then perform the entire lift procedure several times with proper supervision and a capable individual acting as a patient.
Individuals that use the Standing Sling MUST be able to support the majority of their own weight, otherwise injury can occur.
DO NOT raise the patient to a full standing position while using the transport sling otherwise injury can occur.
Invacare standing and transport slings are specifically designed to be used in conjunction with the Invacare Stand Up Lift. Slings and accessories designed by other manufacturers or other Invacare slings are not to be utilized as a component of Invacare’s stand up lift system. Use of these products is prohibited and will void Invacare’s stand up lift warranty. Use only genuine Invacare standing and transport slings and lift accessories to maintain patient safety and product utility. Use the sling that is recommended by the individual’s doctor, nurse or medical assistant for the comfort and safety of the individual being lifted.
Before positioning the legs of the stand up lift around the patient, make sure the patient's feet are out of the way of the footplate, otherwise injury can occur.
Invacare recommends locking the rear swivel casters ONLY when positioning or removing the sling (standing or transport) from around the patient.
Invacare does NOT recommend locking the rear swivel casters of the stand up lift when lifting and transferring an individual. Doing so could cause the lift to tip and endanger the patient and assistants. Invacare DOES recommend that the rear swivel casters be left UNLOCKED during lifting and transferring procedures to allow the stand up lift to stabilize itself when the patient is initially lifted from and transferred to a chair, bed or any stationary object.
STANDING SLINGS - DO NOT use the standing sling in combination with the stand up lift as a transport device. It is intended to transfer an individual from one resting surface to another (such as a bed to a wheelchair). Moving a person using the standing sling in combination with the stand up lift over ANY distance is NOT recommended.
STANDING SLINGS - Before lifting the patient, make sure the bottom edge of the standing sling is positioned on the lower back of the patient and the patient's arms are outside the standing sling.
6
7
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
RPS 440ee User's Manual
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Safety Summary (continued)
STANDING SLINGS - The belt MUST be snug, but comfortable on the patient, otherwise the patient can slide out of the sling during transfer, possibly causing injury.
TRANSPORT SLINGS - Before lifting the patient, make sure the bottom edge of the transport sling is at the base of the spine and the patient's arms are outside the transport sling.
DO NOT use any kind of material (such as a plastic back incontinence pad or seating cushion) between the patient and sling material that may cause the patient to slide out of the sling during transferring.
ALWAYS use the colour coded strap on the standing sling closest to the patient while still maintaining patient stability and comfort. Bleached, torn, cut, frayed, or broken slings are unsafe and could result in injury. Discard Immediately.
DO NOT alter slings.
When elevated a few inches off the surface being transferred from and before moving the patient, check again to make sure that the sling is properly connected to the attachment points of the stand up lift. If any attachments are NOT properly in place, lower the patient back onto the surface and correct this problem.
During transfer, with patient suspended in a sling attached to the lift, DO NOT roll stand up lift over objects such as carpet, raised carpet bindings, door frames, or any uneven surfaces or obstacles that would create an imbalance of the stand up lift and could cause the stand up lift to tip over.
Use push handle on the mast at ALL times to push or pull the stand up lift.
Casters and axle bolts require inspection every six (6) months to check for tightness and wear.
After the first 12 months of operation, inspect all pivot points and fasteners for wear. If the metal is worn, the parts MUST be replaced. Make this inspection every six (6) months thereafter.
The electric motor is sealed at the factory and if service is required, the motor unit MUST BE returned to the factory for repair. DO NOT attempt to open the motor or obtain local service as this will VOID the warranty and may result in damage and a costly repair. Consult your Dealer or Invacare for further information.
Regular maintenance of Stand Up Lift and accessories is necessary to assure proper operation.
WEIGHT LIMITATION DO NOT exceed maximum weight limitation of the Stand Up Lift. The maximum weight limitation is 440 lbs. regardless of any additional weight limitations on accessories.
7
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Key to Symbols
The following symbols are used on lifter and attachment labels:
The symbol required to be displayed on regulated products for sale on the European Market. It indicates that the product complies with applicable European Directives related to health, safety, environment and consumer protection.
Safe Working Load. Represents the maximum load the lifter is rated at for safe operation.
The following symbols are used on sling labels and relate to washing instructions:
Wash at maximum temperature 60
o
C (140oF) permanent press cycle.
Wash at maximum temperature 85oC (198oF) permanent press cycle.
Do not use bleach.
Please refer to individual sling labels for complete instructions regarding washing and laundering.
Do not dry clean.
Dry cleanable.
Do not tumble dry.
Do not iron.
8
9
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
RPS 440ee User's Manual
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Documentation
With each lifter you should receive the following documents:
• thisInstructionManual;
• aTestCerticate;and
• aWarranty/RegistrationCard.
Warranty/Registration Card
The tear-off Registration Card must be completed and returned to Invacare within fourteen days of purchase.
9
RPS 440ee User's Manual
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2008 Invacare Asia Pacific
Assembly
The RPS400ee is packed partly assembled for shipping purposes. To assemble your RPS400ee, remove all contents from box and place unit on workbench or table. Using 16 mm and 17 mm spanners remove the two bolts connecting the mast to the chassis.
Remove mast and boom assembly from base (1). Position upright and attach mast to base using the previously removed bolts.
Note: Ensure these bolts are firmly tightened.
Attach the rod ends as shown in (2).
Cut cable tie securing actuator (3), and swing actuator forward so that it is out of the way.
Swing boom up and over to the front (4) .
Attaching the kneepad receiver
Note: The kneepad is an integral part of the lift and must be attached so that the lifter can function correctly and safely.
Remove the two hex screws on front of mast behind the actuator. Align the kneepad receiver assembly over the screw holes (5). Ensure the assembly is correctly orientated - the kneepad release lever should be on top. Replace the screws. Tighten firmly.
!
When assembling the patient lift, it is important to tighten bolts according to instruction. Failure to do so may result in damage to the lift or injury to the care­giver or client.
Loading...