Invacare Platinum IRC9LXO2AWQ, Platinum IRC5LXO2AW, Platinum Series, Platinum IRC5LXO2AWQ, Platinum IRC5LXAW User Manual

Invacare® Platinum™ Series
IRC5LXAW, IRC5LXO2AW, IRC5LXO2AWQ, and IRC9LXO2AWQ
en HomeFill® System Compatible Oxygen
Concentrators with SensO
User Manual ................................ 3
da HomeFill® System Kompatibel
Brugsanvisning .............................. 35
sv HomeFill® systemet kompatibelt
syrgaskoncentrerare med SensO
Bruksanvisning .............................. 67
no HomeFill®System Kompatibla
konsentratoren med SensO
Bruksanvisning .............................. 99
HomeFill® System yhteensopiva Happi
keskittimet kanssa SensO
Käyttöohje .................................133
2
2
2
2
2
This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
© 2017Invacare Corporation Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™ and®.Alltrademarksare ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. Pine-solandLestoilaretrademarksofTheCloroxCompany. DawnisatrademarkofTheProctorandGambleCompany. Making Life's Experiences Possible is a registered trademark in the U.S.A.
Contents
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuture reference.
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................6
1.3Description...................................7
1.4Contraindications..............................7
1.5OptionalAccessories............................7
2Safety..........................................8
2.1LabelLocations................................8
2.2GeneralGuidelines.............................10
2.3RadioFrequencyInterference......................13
3Components....................................14
3.1ComponentIdentification.........................14
4Setup..........................................15
4.1ShippingandHandling............................15
4.2Unpacking....................................15
4.3Inspection....................................15
4.4Storage......................................15
5Usage..........................................16
5.1Introduction..................................16
5.2SelectaLocation...............................16
5.3SettingUptheConcentrator......................18
5.3.1SettingUptheHumidifier......................18
5.3.2TurningtheConcentratorOn...................20
5.3.3Flowrate..................................20
5.3.4SensO
OxygenPurityIndicator.................21
2
5.3.5UnitswithoutSensO
........................22
2
5.4MainPowerLossAlarmSystem....................22
5.5ElapsedTimeMeter(HourMeter)..................22
6Maintenance....................................23
6.1CleaningtheCabinet............................23
6.2CleaningtheCabinetFilter........................24
6.3HumidifierCleaningandThermicDisinfection..........25
6.4CleaningandDisinfectionBetweenPatients............25
6.5PreventiveMaintenanceChecklist...................26
7AfterUse.......................................27
7.1RecyclingInformation............................27
7.2WearandTear................................27
8Troubleshooting.................................28
8.1Troubleshooting...............................28
9Technicaldata...................................30
9.1TypicalProductParameters.......................30
Invacare®Platinum™Series
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe practiceswhichcouldresultinpersonalinjuryorpropertydamage. Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor efficient,trouble-freeuse.
SymbolsinDocumentation
Readmanual
NoSmoking
NoOpenFlame
ClassII,DoubleInsulated
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
ElectricalHazard
Keepdry
ECRepresentative
Manufacturer
4 1195661-A
General
Callsupplier
**ThisproductcomplieswithDirective 93/42/EECconcerningmedicaldevices.
Thelaunchdateofthisproductisstatedinthe CEdeclarationofconformity.
FormoredetailsaboutCEmarks,refertoTypicalProduct Parameters.
SymbolsonProduct
Unitrunning
Unitnotrunning
O
Indicators
2
SymbolO
O
SYSTEMOKAY
2
O O
85%
SYSTEMFAILURE O
PurityIndicatorLights(LED)
2
over85%
2
between73%to
2
Below73%
2
GREENIndicatorLight
YELLOWIndicatorLight A.YELLOWSolid B.YELLOWFlashingSensor Failure.Callaqualified
technician REDIndicatorLight RefertoTroubleshooting.
LXIndicators
SymbolO
I/O
PurityIndicatorLights(LED)
2
SYSTEMOKAYGREENIndicatorLight
SYSTEMFAILURE
REDIndicatorLight ContinuousAudibleAlarm
Sieve—GARD™ CompressorShutdown Callaqualifiedtechnician.
1195661-A 5
Invacare®Platinum™Series
1.2IntendedUse
Youroxygenconcentratorisintendedforindividualusebypatients withrespiratorydisorderswhorequiresupplementaloxygen.The deviceisnotintendedtosustainorsupportlife.Theconcentratoris intendedforusewithinahomeorinstitutionalenvironment.
DANGER! RiskofInjuryorDeath
Thisproductistobeusedasanoxygensupplementand isnotintendedtobelife-supportingorlife-sustaining. ONLYusethisproductifthepatientiscapableof spontaneousbreath,abletoinhaleandexhalewithout theuseofamachine.
–DONOTuseinparallelorserieswithotheroxygen
concentratorsoroxygentherapydevices.
DANGER! RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientsonflow ratesgreaterthan5l/mmaybeatanincreasedriskfor seriousinjuryordeathintheeventoffailure.
–ALWAYSdiscussthisincreasedriskwithyourhealth
careproviderBEFOREusingthisproductifyouare prescribedaflowrategreaterthan5l/m.
DANGER! RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen concentratorinthemarkettoday,thisoxygen concentratorcanfailtoproduceoxygenduetopower failureordevicemalfunction.
–ALWAYShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient toswitchtotheirbackupsourceofoxygen.Referto Troubleshootingformoredetail.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand productparametershasnotbeentestedandmayleadto productdamage,lossofproductfunction,orpersonal injury.
–DONOTusethisproductinanywayotherthan
describedintheproductparametersandintendeduse sectionsofthismanual.
6 1195661-A
General
1.3Description
TheInvacarePlatinumconcentratorisusedbypatientswith respiratorydisorderswhorequiresupplementaloxygen.Thedevice isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87% to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse ofanasalcannula.
TheInvacarePlatinumconcentratorusesamolecularsieveand pressureswingadsorptionmethodologytoproducetheoxygen gasoutput.Ambientairentersthedevice,isfilteredandthen compressed.Thiscompressedairisthendirectedtowardoneof twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe patient.
TheInvacarePlatinumconcentratoriscapableofoperationbythe patientinahomeenvironmentorinaninstitutionalenvironment. Deviceoperatesatanominal230VAC/50Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified technicalpersonnelonly.
1.4Contraindications
InvacareisnotawareofanycontraindicationsfortheInvacare PlatinumConcentratorSeries.
deviceswillbebestforyou.Theyshouldalsogiveyouadviceonthe properusage,maintenance,andcleaning.
Thesupplyaccessories(nasalcannula,mask,tubing,humidifier,etc.) usedtodeliveroxygentothepatientneedtoincludeameansto reducethepropagationoffireintheaccessoriesforthesafetyofthe patientandothers.Ifacommerciallyavailable,fire-activatedflow stopdeviceisusedintheaccessoriessetup,itshouldbeplacedas closetothepatientaspossible.
1.5OptionalAccessories
Therearemanydifferenttypesofhumidifiers,oxygentubing,cannulas andmasksthatcanbeusedwiththisdevice.Youshouldcontactyour localhomecareproviderforrecommendationsonwhichofthese
1195661-A 7
Invacare®Platinum™Series
2Safety
2.1LabelLocations
A
B
C
SerialNumberLabelislocatedontheresonator intakeassembly.
SpecificationLabelislocatedonthebackofthe concentratoratthebase.
DoubleInsulationLabelislocatedonthebackof theconcentratoratthebase.
8 1195661-A
IRC5LXO2AW,IRC5LXO2AWQ,IRC9LXO2AWQIRC5LXAW
Safety
D
E
1195661-A 9
Invacare®Platinum™Series
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyandoperationofthe concentratortheseinstructionsMUSTbefollowed.
DANGER! RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor damage.Thissectioncontainsimportantinformationfor thesafeoperationanduseofthisproduct.
–DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand understandingtheseinstructionsandanyadditional instructionalmaterialsuchasusermanuals,service manualsorinstructionsheetssuppliedwiththis productoroptionalequipment.
–Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional, dealerortechnicalpersonnelbeforeattemptingto usethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot workingcorrectly,contactatechnicianorInvacare forrepair.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy substancesandothercombustiblesareeasilyignited andburnwithgreatintensityinoxygenenrichedairand whenincontactwithoxygenunderpressure.Toavoid fire,death,injuryordamage:
DONOTSMOKEwhileusingthisdevice.DONOTnearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
NOSMOKINGsignsshouldbeprominentlydisplayed.Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby anytypeoffriction.
–Keepallmatches,lightedcigarettesorothersources
ofignitionoutoftheroominwhichthisconcentrator islocatedandawayfromwhereoxygenisbeing delivered.
–Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings, chaircushions,clothing,andawayfromheatedorhot surfacesincludingspaceheaters,stoves,andsimilar electricalappliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesaletoor ontheorderofamedicalpractitionerlicensedbya governmentalagencywherehe/shepractices.
–ONLYalicensedmedicalpractitionermayorderthe
purchaseoruseofthisdevice.
10 1195661-A
Safety
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor injurytopersons:
–DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceable parts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the concentratormustbelocatedinanotherroomat least7ft(2.5m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
–DONOTusefrayedordamagedACadaptercords.
WARNING! RiskOfInjuryOrDamage
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare accessories.Accessoriesdesignedbyother manufacturershavenotbeentestedbyInvacareand arenotrecommendedforusewithInvacareproducts.
–Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis device.Youshouldcontactyourlocalhomecare providerforrecommendationsonwhichofthese deviceswillbebestforyou.Theyshouldalsogiveyou adviceontheproperusage,maintenance,andcleaning.
DANGER! RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway contamination:
–DONOTusetheconcentratorinthepresenceof
pollutants,smoke,fumes,flammableanesthetics, cleaningagents,orchemicalvapors.
WARNING! RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratorisusedby
ornearchildrenorphysically-challengedindividuals.
–Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING! RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing entanglement:
–Keepchildrenandpetsawayfromnasalcannulaand
tubing.
–Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/ordisabledpersons.
WARNING! RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.Check
withyourequipmentproviderorphysicianto determinehowoftenthecannulashouldbereplaced.
–DONOTsharecannulasbetweenpatients.
1195661-A 11
Invacare®Platinum™Series
WARNING! RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto you.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings shouldbechanged.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorbypulling
onthecord.
–DONOTuseextensioncordswithACpower
adaptersprovided.
–Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhenplugged
in.
–Makesureconcentratorisoffwhennotinuse.
WARNING! RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned tominimizeroutinepreventivemaintenance.Toprevent injuryordamage:
–Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas factorytrainedpersonnelshouldperformpreventive maintenanceorperformanceadjustmentsonthe oxygenconcentrator,exceptfortasksdescribedin thismanual.
–UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION! RiskofDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto water,callequipmentprovider/qualifiedtechnicianfor examinationandrepair.
–NEVERdroporinsertanyobjectorliquidintoany
opening.
–Invacarerecommendstheconcentratornotbeused
intherain.
12 1195661-A
2.3RadioFrequencyInterference
WARNING! RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamagefrom interferencewithwirelessequipment:
–Keepconcentratoratleast9.8ft(3.0m)awayfrom
wirelesscommunicationequipmentsuchaswireless homenetworkdevices,mobilephones,cordless phonesandbasestations,walkie-talkies,etc.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimits specifiedbyIEC/EN60601-1-2.Theselimitsaredesignedtoprovide areasonableprotectionagainstelectromagneticinterferenceina typicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels ofelectromagneticemissionspermittedbytheabovestandards. Todetermineiftheemissionsfromtheconcentratorarecausing theinterference,turntheconcentratorOff.Iftheinterference withtheotherdevice(s)stops,thentheconcentratoriscausingthe interference.Insuchrarecases,interferencemaybereducedor correctedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetweenthe
equipment.
Connecttheequipmentintoanoutletonacircuitdifferentfrom
thattowhichtheotherdevice(s)isconnected.
Safety
1195661-A 13
Invacare®Platinum™Series
3Components
3.1ComponentIdentification
D
E
F
G
PowerSwitch CircuitBreaker ElapsedTimeMeter CabinetFilter
RearView
H
I
Accessories(NotShown):HomeFIllhomeoxygen compressor—IOH200AW
TheHomeFilloutletfittingIistobeusedonlyfor fillingoxygencylinderswiththeHomeFillhomeoxygen compressor.Theoutletfittingdoesnotaffectconcentrator performance.RefertotheHomeFIllusermanualfor
A
B
C
OxygenOutlet Flowmeter OxygenPurityIndicatorLights/FaultandPower
IndicatorLights
connectionandoperatinginstructions.Whennotinuse,the plugprovidedwiththeconcentratorshouldbeinsertedinto theoutletfitting.FormoreinformationabouttheHomeFill, contactyourInvacaredealer.
PowerCord HomeFillOutletFitting
14 1195661-A
Setup
4Setup
4.1ShippingandHandling
Beforeyouinstallandoperatetheconcentrator,pleasereferto PlatinumSeriesUserManual.
ThePlatinumconcentratorshouldALWAYSbekeptintheupright positiontopreventcabinetdamagewhilebeingtransported.The shippingcontainerhasbeendesignedtoassuremaximumprotection oftheconcentrator.
Iftheconcentratoristobereshippedbycommoncarrier,itshould bepackedinanewcarton.Additionalcartonsareavailablefrom Invacare.
Theaircompressorsuspensionsystemhasbeenengineeredto withstandseveremotionandorientation.
4.2Unpacking
1.Checkforanyobviousdamagetothecartonoritscontents.If damageisevident,notifythecarrier,oryourlocaldealer.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallthecomponentsfromthecarton.
Unlesstheoxygenconcentratoristobeused IMMEDIATELY,leaveconcentratorinitspackagingfor storageuntiluseisrequired.
4.3Inspection
1.Inspect/examineexterioroftheconcentratorfornicks,dents, cracks,scratchesorotherdamage.
2.Inspectallcomponents.
4.4Storage
1.Storetherepackagedoxygenconcentratorinadryarea.Referto TypicalProductParametersforstoragetemperatureparameters.
2.DONOTplaceobjectsontopofrepackagedconcentrator.
1195661-A 15
Invacare®Platinum™Series
5Usage
5.1Introduction
Youroxygenconcentratorisintendedforindividualuseindoors. Itisanelectronicallyoperateddevicethatseparatesoxygenfrom roomair.Itprovideshighconcentrationofoxygendirectlytoyou throughanasalcannula.Clinicalstudieshavedocumentedthat oxygenconcentratorsaretherapeuticallyequivalenttoothertypesof oxygendeliverysystems.
Yourproviderwillshowyouhowtouseyouroxygenconcentrator. He/Sheshouldbecontactedwithanyquestionsorproblems regardingyouroxygenconcentrator.Thisusermanualwilltellyou aboutyourconcentratorandwillserveasareferenceasyouuse yourconcentrator.
5.2SelectaLocation
WARNING! RiskofInjuryorDamage
Toavoidinjuryordamagefromairbornepollutants and/orfumesandforoptimalperformance:
–Locateandpositiontheconcentratorinawell
ventilatedspacesothattheairintakeandtheair exhaustsarenotobstructed.
–NEVERblocktheairopeningsoftheconcentrator
orplaceitonasoftsurface,suchasabedorcouch, wheretheairopeningmaybeblocked.
–Keeptheopeningsfreefromlint,hairandsimilar
foreignitems.
–Keepconcentratoratleast12inches(30,5cm)away
fromwalls,draperiesandfurniture.
–Avoiduseinpresenceofpollutants,smokeorfumes,
flammableanesthetics,cleaningagentsorchemical vapors.
–DONOTuseinacloset.
WARNING! RiskofInjury
Toavoidinjuryduringtherapy:
–TheoxygenconcentratorMUSTbeplacedonalevel
surfaceforuse.
–DONOTrelocatetheoxygenconcentratorwhilein
use.
Youmayselectaroominyourhousewhereusingyouroxygen concentratorwouldbemostconvenient.Yourconcentratorcanbe easilyrolledfromroomtoroomonitscasters.
16 1195661-A
Usage
Youroxygenconcentratorwillperformbestwhenoperatedunder theconditionsoutlinedinthetablebelow.
Usageinenvironmentsotherthanthosedescribedmayresultinthe needforincreasedequipmentmaintenance.Theairintakeofthe unitshouldbelocatedinawellventilatedareatoavoidairborne pollutantsand/orfumes.
RecommendedGuidelinesforOptimumPerformance
Temperature: Electrical:Noextensioncords. Placement:
TubingandCannula:IRC5LXAW,IRC5LXO2AW,
Environment:
RelativeHumidity:
10°C-35°C(50°F-95°F)
Nocloserthan30,5cm(12 in)fromthewall,furniture, draperies,orsimilarsurfaces.
IRC5LXO2AWQ 2,1m(7ft)cannulawitha
maximum15,2m(50ft)of Crush-ProofTubing(DONOT pinch.)
IRC9LXO2AWQ Recommendeduseupto15.2
m(50ft)highflowtubingwith highflowcannulaatallflow rates.
Smokeandsoot-free.No confinedspaces
(Example:Noclosets). 20to60%
TimeofOperation:Upto24hoursperday. FlowRate
IRC5LXAW,IRC5LXO2AW, IRC5LXO2AWQ
0.5L/min.to5L/min.Flow rateslessthan1L/min.,arenot recommended.
IRC9LXO2AWQ 1L/min.to9L/min.Flowrates
lessthan1L/min.arenot recommended.
MinimumOperatingTime:30Minutes
Ensurethatyourconcentratorisatleast30,5cm(12in) awayfromwalls,draperiesorfurnituretoassuresufficient airflow.Avoiddeeppilecarpetsandheaters,radiatorsor hotairregisters.
1195661-A 17
Invacare®Platinum™Series
5.3SettingUptheConcentrator
1.Pluginpowercordtoanelectricaloutlet.
2.ConnectHumidifier(ifsoprescribed).
5.3.1SettingUptheHumidifier
WARNING! RiskofInjuryorDamage
Toavoidburnsfromsteamorhotwater,inhalationof waterand/orwaterdamagetotheconcentrator:
–DONOTfillhumidifierbottlewithhotwater.Allow
boiledwatertocooltoroomtemperaturebefore filling.
DONOToverfillhumidifier.Replacethehumidifiercapandsecurelytighten.
Confirmthatthecapisnotcross-threadedonthe humidifierbottle.
–DONOTreversetheoxygeninputandoutput
connections.Waterfromthehumidifierbottlewill travelthroughthecannulabacktothepatientifinput andoutputconnectionsarereversed.
–Whenusingtubingconnectionslongerthan2.1m
(7feet)inlength,positionthehumidifierasclose tothepatientaspossibletoallowformaximum humidificationoutput.
WARNING! RiskofInjuryorDamage
Failuretoproperlyinstallthehumidifierbottleorother accessoriestotheconcentratorwillimpacttheflowof oxygen.
–Tocheckforproperoperationoftheoxygen
concentratorandaccessories,placetheendofthe nasalcannulaunderthesurfaceofahalf-fullcupof waterandlookforthebubbles.
–Iftherearenobubbles,checkallconnections
(includinghumidifierbottleandotheraccessories,if applicable)andrepeat.Contactyourdealerorservice providerimmediatelyifbubblesstilldonotappear.
FillingtheHumidifierBottleA A
1.Removecapfrombottle.
2.FillhumidifierbottleAwithboiledtapwaterorbottledwater tothelevelindicatedbythemanufacturer.Boiltapwaterfor approximately10minutesandcooltoroomtemperatureprior touse.
A
18 1195661-A
CAUTION! RiskofDamage
–Confirmthehumidifiercapisnotcross-threaded
onthehumidifierbottle.
3.InsertaflatheadscrewdriverintheplategrooveBonthetop edgeofthefilteraccessdoorCandgentlyprythefilteraccess dooroff.
4.PullupandremovethehumidifierbottleadapterD(nextto inletfilter).
Usage
WARNING! RiskofInjury
Toavoidimpactingtheflowofoxygen: –Installthehumidifierbottleand/orother
accessoriesproperly.
5.Attachthehumidifierbottleadaptertothehumidifierbottleby turningthewingnutonthehumidifierbottlecounterclockwise untilitissecurelyfastened.
6.WhenplacingthehumidifierbottleinthecompartmentEon theconcentrator,attachthehumidifierbottle/adaptertubeFto theoxygenoutletconnectorGontheconcentrator.
Whenusingtubingconnectionslongerthan2.1m(7 feet)inlength,positionthehumidifierasclosetothe patientaspossibletoallowformaximumhumidification output.
7.Attachthepatients’nasalcannulasupplytubetothehumidifier bottleoutletH.
8.Afterassembly,ensurethatoxygenisflowingthroughthenasal cannula.
9.Replacethefilteraccessdooronthesideofconcentratorby snappingitbackintotheplategroove.
1195661-A 19
Invacare®Platinum™Series
5.3.2TurningtheConcentratorOn
1.PresspowerswitchAtoOnposition. Allthepanellightsandtheaudiblealarmwillcomeon
foronesecond,indicatingthattheunitisfunctioning properly.
5.3.3Flowrate
WARNING! RiskofInjury
Toavoidinjuryfromexcessoxygenoradeficitof oxygen:
–DONOTchangetheL/minsettingontheflowmeter
unlessachangehasbeenprescribedbyyourphysician ortherapist.
–DONOTsettheflowgreaterthan9L/min.
1.TurntheflowrateknobAtothesettingprescribedbyyour physicianortherapist.
ToproperlyreadtheflowmeterB,locatethe prescribedflowratelineontheflowmeter.Next,turn theflowknobuntiltheballCrisestotheline.Now, centertheballontheL/minlineprescribed.
2.Iftheflowrateontheflowmetereverfallsbelow0.5L/minfor morethanaboutoneminute,thePotentialObstructionAlert willbetriggered.Thisisarapidbeepingoftheaudiblealarm. Checkyourtubingoraccessoriesforblockedorkinkedtubing oradefectivehumidifierbottle.Afterratedflowisrestoredto between0.5L/minand0.75L/min.,thePotentialObstruction Alertwillturnoff.
APotentialObstructionAlertindicatesaconditionthat maybeassociatedwithapartialorcompleteobstruction ofoxygenoutput.
TheuseofsomeaccessoriessuchasthePreciseRx™pediatric flowmeterandtheHomeFillcompressorwilldeactivatethe PotentialObstructionAlert.
20 1195661-A
Usage
5.3.4SensO
OxygenPurityIndicator
2
Thisfeaturemonitorsthepurityofoxygengeneratedbytheoxygen concentrator.Ifpurityfallsbelowfactorypresetstandards,indicator lightsonthecontrolpanelwillilluminate.
InitialStartupoftheConcentrator
Concentratormaybeusedduringtheinitialwarm-uptime (approximately30min.)whilewaitingfortheO reachmaximum.
Whentheunitisturnedon,theGREENlightwillcomeon(SYSTEM OK/O
greaterthan85%).Afterfiveminutes,theoxygensensorwill
2
beoperatingnormallyandwillcontroltheindicatorlightsdepending onoxygenconcentrationvalues.
ExplanationofOxygenPurityIndicatorLights
2
purityto
Symbol
O
Purity
SYSTEM
A
O
OKAYO
2
over85%
Indicator
2
Light
(LED)
GREEN
2
light
YELLOW
light(Solid)
O
2
B
between 73%and
85%
YELLOW
light
flashing
Explanation
Noactionis needed.Normal foroperation.
Immediatelycall supplier.Youmay continuetousethe concentratorunless instructedotherwise byyoursupplier.Be certainthatbackup oxygenisnearby.
ContinuousAudible Alarm
Sieve-GARD™
SYSTEM
C
FAILURE
O
below
2
REDlight
(Solid)
73%
Compressor Shutdown
SYSTEMFAILURE. Totalunitshutdown. Immediatelyswitch tobackupoxygen supplyandcall supplier.
D
CallSupplier
1195661-A 21
Invacare®Platinum™Series
5.3.5UnitswithoutSensO
Concentratormaybeusedduringtheinitialstartwarm-up time(approximately30min).O2puritywillbuildtoa maximumduringthisperiod.
Light
GREEN
A
B
C
light
RED
light
Symbol
2
NormalOperation
ContinuousAudibleAlarm Sieve-GARD™Compressor Totalunitshutdown.Immediately
switchtobackupoxygensupply andcallsupplier.
CallSupplier
chargedandreadytotriggerthealarmintheeventofmainpower loss.
5.5ElapsedTimeMeter(HourMeter)
Thehourmeterdisplaysthecumulativenumberofhourstheunithas operated.RefertoComponentIdentificationforspecificlocation.
Explanation
5.4MainPowerLossAlarmSystem
Thealarmsystemcontainsnobattery-andistherebymaintenance free.Thealarmsystemispoweredbyacapacitorthatiscontinuously
22 1195661-A
Maintenance
6Maintenance
6.1CleaningtheCabinet
DANGER! RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned tominimizeroutinepreventivemaintenance.Toprevent injuryordamage:
Onlyqualifiedpersonnelshouldperformpreventive
DONOTremovecabinet.
DANGER! RiskofInjuryorDamage
Liquidwilldamagetheinternalcomponentsofthe concentrator.Toavoiddamageorinjuryfromelectrical shock:
TurnOfftheconcentratorandunplugthepowercord
DONOTallowanycleaningagenttodripinsidethe
DONOTsprayorapplyanycleaningagentdirectly
DONOThosedowntheproduct.
maintenanceontheoxygenconcentrator.
beforecleaning.
airinletandoutletopenings.
tothecabinet.
CAUTION! RiskofDamage
Harshchemicalagentscandamagetheconcentrator. Toavoiddamage:
–DONOTcleanthecabinetorfilterwithalcohol
andalcoholbasedproducts(isopropylalcohol), concentratedchlorine-basedproducts(ethylene chloride),andoil-basedproducts(Pine-Sol
®
,Lestoil oranyotherharshchemicalagents.Onlyusemild liquiddishdetergent(suchasDawn
).
Ataminimum,preventivemaintenanceMUSTbeperformed accordingtothemaintenancerecordguidelines.Inplaces withhighdustorsootlevels,maintenancemayneedtobe performedmoreoften.RefertotheMaintenanceChecklist.
Periodicallycleantheconcentrator’scabinetasfollows:
1.Useadampcloth,orsponge,withamilddetergentsuchas
Dawn
dishwashingsoaptogentlycleantheexteriorcase.
2.Allowtheconcentratortoairdry,oruseadrytowel,before operatingtheconcentrator.
®
)
1195661-A 23
Invacare®Platinum™Series
6.2CleaningtheCabinetFilter
CAUTION! RiskofDamage
Toavoiddamagefromclogging:
–DONOToperatetheconcentratorwithoutthefilter
installed.
Therearetwocabinetfilterslocatedoneachsideofthe cabinet.
1.RemovethefilterAandcleanasneeded.
Environmentalconditionsthatmayrequiremore frequentinspectionandcleaningofthefilterinclude,but arenotlimitedto:highdust,airpollutants,etc.
2.Cleanthecabinetfilterwithavacuumcleanerorwashwitha mildliquiddishdetergent(suchasDawn™)andwater.Rinse thoroughly.
3.Thoroughlydrythefilterandinspectforfraying,crumbling,tears andholes.Replacefilterifanydamageisfound.
4.Reinstallthecabinetfilter.
24 1195661-A
Maintenance
6.3HumidifierCleaningandThermic Disinfection
Cleananddisinfecttheoxygenhumidifierdailytoreduce limestonedepositsandeliminatepossiblebacterial contamination.Followtheinstructionsprovidedbythe manufacturer.Ifnoneareprovided,followthesesteps:
1.Washhumidifierinsoapywaterandrinsewithasolutionoften partswaterandonepartvinegar.
2.Rinsethoroughlywithhotwater.
3.Airdrythoroughly.
Tolimitbacterialgrowth,airdrythehumidifierthoroughly aftercleaningwhennotinuse.RefertoSetUpforuse.
6.4CleaningandDisinfectionBetweenPatients
WARNING! RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto concentrator:
–Onlyqualifiedpersonnelshouldperformcleaning
anddisinfectionoftheoxygenconcentratorand accessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossiblepathogen exchangebetweenpatientsduetocontaminationof componentsoraccessories.PreventiveMaintenanceshould alsobeperformedatthistimeifnecessary.
1.Disposeofandreplaceallpatientsideaccessoriesnotsuitable formultiplepatientuse,includingbutnotlimitedto:
NasalCannulaandTubing
Mask
Humidifier
2.Performmaintenanceproceduresdescribedinthismanualand itemsonPreventiveMaintenanceChecklist
3.Checkconcentratorforpossibleexternaldamageorsignsthatit mayrequireserviceorrepair.
4.Ensureconcentratorfunctionsproperlyandallalarmsarein workingorder.
5.Beforerepackaginganddistributiontonewpatient,ensure packagingcontentscontaintheconcentrator,powercord,air inletscoop,assemblyinstructions,humidifier,cannula,labels andusermanual.
1195661-A 25
Invacare®Platinum™Series
6.5PreventiveMaintenanceChecklist
ModelNo:
ONEACHINSPECTION
RecordDateofService RecordElapsedHoursonHourMeter CleanCabinetFilter(s)(RefertoCleaningtheCabinetFilter.) CheckPrescribedL/min.FlowRate
DURINGPREVENTIVEMAINTENANCESCHEDULE,ORBETWEENPATIENTS UNITSWITHOUTSensO
—Every6monthsofcontinuoususe(Equivalentto4,380hours)
2
CheckOxygenConcentration(greenindicatorlight) Clean/ReplaceCabinetFilter(s)(RefertoCleaningtheCabinetFilter.) Check/ReplaceOutletHEPAFilter* Check/ReplaceCompressorInletFilter* CheckPowerLossAlarm*
UNITSWITHSensO
—Every3yearsofcontinuoususe(Equivalentto26,280hours)
2
CheckOxygenConcentration(greenindicatorlight) Clean/ReplaceCabinetFilter(s)(RefertoCleaningtheCabinetFilter.) Check/ReplaceOutletHEPAFilter* Check/ReplaceCompressorInletFilter* CheckPowerLossAlarm* *Tobeconductedbyproviderorqualifiedservicetechnician.Refertoservicemanual.
SerialNo:
26 1195661-A
AfterUse
7AfterUse
7.1RecyclingInformation
Thisproducthasbeensuppliedfromanenvironmentallyaware manufacturerwhocomplieswiththeWasteElectricalandElectronic Equipment(WEEE)Directive2012/19/EU.Thisproductmaycontain substancesthatcouldbeharmfultotheenvironmentifdisposedofin places(landfills)thatarenotappropriateaccordingtolegislation.
Followlocalgoverningordinancesandrecyclingplansregarding disposaloftheconcentratororcomponentsnormallyusedin operation.Theconcentratordoesnotgeneratewasteorresiduein operation.
DONOTdisposeoftheconcentratorinthenormalwaste
stream.
AnyaccessoriesnotpartoftheconcentratorMUSTbehandled
inaccordancewiththeindividualproductmarkingfordisposal.
DONOTdisposeoftheinternalorsupplementalbatterypacks.
Batterypacksshouldbereturnedtoyourdealer/provider.
Invacare®iscontinuouslyworkingtowardsensuringthatthe company’simpactontheenvironment,locallyandglobally,isreduced toaminimum.Wecomplywiththecurrentenvironmentallegislation (e.g.WEEEandRoHSdirectives).WeonlyuseREACHcompliant materialsandcomponents.
7.2WearandTear
Invacarereservestherighttoaskforanyitembackthathasan allegeddefectinworkmanship.SeeWarrantythatshippedwiththe productforspecificwarrantyinformation.
RefertotheUserManualforproperpreventativemaintenance scheduleanduseoftheproduct.
Thisisjustageneralguidelineanddoesnotincludeitemsdamaged duetoabuseandmisuse.
Normalwearandtearitemsandcomponentsforthisproductare listedbelow.
Alltypesoffilters
Alltypesoftubing
Alltypesofbatteries,includingleadacid/lithium,coincell(watch
type)andgel(6months)
Sieveisaporousfilteringmaterialandisconsideredawear item.Somefactorsthatcouldaffectsievemateriallifeinclude humidity,temperature,particulates,aircontaminates,air intake,vibrationandotherenvironmentalconditions.
1195661-A 27
Invacare®Platinum™Series
8Troubleshooting
8.1Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:MainPowerLoss:
ShortBeeps,longpause Concentratornotoperating,
powerswitchOn. Beep...Beep...
Alarm:
Continuous Concentratornotoperating,
powerswitchOn. Beep...
1.Powercordnotpluggedin.
2.Nopoweratoutlet.
3.Trippedcircuitbreaker.
SystemFailure:
1.Unitoverheatingduetoblockedair
2.Insufficientpoweratoutlet.
3.Internalrepairsrequired.
1.Insertplugintooutlet.
2.Inspecthousecircuitbreakersorfuses.Ifproblemrecurs,use adifferentoutlet.
3.Push/resetcircuitbreaker.Ifproblemrecurs,callservice provider.
1.
intake.
a.Removeandcleancabinetfilters. b.Moveoxygenconcentratoratleast12in(30,5cm)away
fromwalls,draperiesorfurniture.
2.DONOTuseextensioncords.Movetoanotherelectrical outletorcircuit.
3.Callserviceprovider.
28 1195661-A
SYMPTOMPROBABLECAUSESOLUTION
Alarm:
YELLOWorREDlight Illuminated.
Alarm:ContinuousOnRED light
*Onlyappliesto SensO
models.
2
GREENlightwithYELLOW LightFlashing.
Alarm:
Rapid Beep...Beep... Beep...Beep
SystemFailure:
1.Lowoxygenpurity.*
2.Kinkedorblockedtubing,cannula orhumidifier.*
3.Flowmetersetat0.5L/min*.
4.Unitoverheatingduetoblockedair intake.
5.Insufficientpoweratoutlet.
6.Internalrepairsrequired.
1.CleanorReplacefilters.
2.Inspectforkinksorblockages.Correct,cleanorreplaceitem. Oncecorrected,turnpowerOfffor60secondsandthenturn powerbackOn.
3.Checkflowmeterissetto1.0L/minormore.RefertoTypical ProductParameters.
4.
a.Removeandcleancabinetfilters. b.Moveconcentratoratleast12in(30,5cm)fromwalls,
5.DONOTuseextensioncords.Movetoanotherelectrical outletorcircuit.
6.CallServiceProvider.
1.Internalrepairsrequired.1.CallServiceProvider.
PotentialObstructionAlert
1.Possibleinternalobstructioninthe oxygenpath.Kinkedorblocked tubing,cannulaorhumidifier.
2.Flowmetersetat0.5L/min.
1.Inspectforkinksorblockages.Correct,cleanorreplaceitem. Oncecorrected,turnpowerOfffor60secondsandthenturn powerbackON.
2.Flowrateslessthan1L/minarenotrecommend.
Troubleshooting
draperies,andfurniture.
Theuseofthepediatricflowmeterwilldeactivatethe PotentialObstructionAlert.
1195661-A 29
Invacare®Platinum™Series
9Technicaldata
9.1TypicalProductParameters
ElectricalRequirements: RatedCurrentInput:1.4A(IRC5LXO2AWQ)
SoundLevel:39.5dBAverage
Altitude:Upto1,828m(6,000ft)abovesealevelwithoutdegradationofconcentrationlevels.
*OxygenOutput ConcentrationLevels:
All5LXO2AWQ/5LXAW /5LXO2AW /9LXO2AWQmodels
230VAC+10,–15%(253VAC/195.5VAC),50Hz
2.0A(IRC5LXAW,IRC5LXO2AW)
2.3A(IRC9LXO2AWQ)
44dBAverage(IRC5LXAW,IRC5LXO2AW) 50dBAverage(IRC9LXO2AWQ)
NOTRECOMMENDEDFORUSEABOVE1828m(6,000ft). AtmosphericPressureRange:101.33kPa–81.22kPa ForIRC9LXO2AWQ-Upto4,000ft(1230m)abovesealevelwithoutdegradationofconcentrationlevels. AtmosphericPressureRange:101.33kPa–88.0kPa IRC5LXAW,IRC5LXO2AWandIRC5LXO2AWQ 93%minimumat1to3L/min. 91%minimumat4L/min. 87%minimumat5L/min.(maximumrecommendedflow)
30 1195661-A