Invacare Platinum IRC5LXO2, Platinum IRC10LXO2, Platinum IRC5LX, Platinum IRC10LX Operator's Manual

Operator’s Manual
Platinum™Series
XL, 5, 10
Oxygen Concentrators Standard,
with
SensO
®
and HomeFill™II
2
Invacare 5
with
SensO
2
Model IRC5LXO2
Invacare 10
with
SensO
2
Model IRC10LXO2
Dealer: This manual MUST be given to the end user. User: and save for future reference.
Invacare products, parts, and services,
Before using this product, read this manual
For more information regarding
Invacare 5
Model IRC5LX
Invacare 10
Model IRC10LX
WARNING
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT ­OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT.
ACCESSORIES
There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care provider for recommendations on which of these devices will be best for you. They should also give you advice on the proper usage, maintenance, and cleaning.
WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
Platinum™Series 2 Part No 1118353
TABLE OF CONTENTS
ACCESSORIES .................................................. 2
SPECIAL NOTES ............................................... 4
SECTION 1—GENERAL GUIDELINES .................. 5
SECTION 2—FEATURES .................................... 9
SECTION 3—HANDLING ................................. 10
Unpacking............................................................................................10
Inspection............................................................................................ 10
Storage.................................................................................................10
SECTION 4—SPECIFICATIONS ......................... 11
SECTION 5—OPERATING INSTRUCTIONS ....... 21
Introduction........................................................................................21
Select a Location ...............................................................................21
Recommended Guidelines for Optimum Performance............22
Plug in Power Cord .......................................................................... 22
Connect Humidifier (If So Prescribed).........................................22
Power Switch .....................................................................................25
Flowrate............................................................................................... 26
SensO2 Oxygen Purity Indicator (If Your Unit Features the O2
Sensor Feature) ........................................................................................27
Initial Startup of the Concentrator ...............................................27
Explanation of Oxygen Purity Indicator Lights...........................27
Indicators (If Your Unit Does Not Feature the O
Sensor) ...28
2
SECTION 6—MAINTENANCE .......................... 29
Routine Maintenance........................................................................ 29
SECTION 7—TROUBLESHOOTING GUIDE ......... 32
LIMITED WARRANTY .................................... 35
Part No 1118353 3 Platinum™Series
SPECIAL NOTES
SPECIAL NOTES
SIGNAL WORD MEANING
Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious
DANGER
WARNING
CAUTION
injury.
Warning indicates a potentially hazardous situation which, if not avoided, could result in death or seri­ous injury.
Caution indicates a potentially hazardous situation which, if not avoided, may result in property dam­age.
NOTICE
The information contained in this document is subject to change without notice.
DANGER
Users MUST not smoke while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located. No smoking signs should be prominently displayed. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.
Platinum™Series 4 Part No 1118353
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed.
CAUTION
“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”
WARNING
SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner’s Manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel.
The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air, and some that will not, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, it is necessary that all sources of ignition be kept away from the product and preferably out of the room in which it is being used. NO SMOKING signs should be prominently displayed.
Part No 1118353 5 Platinum™Series
SECTION 1—GENERAL GUIDELINES
A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless recommended by Invacare.
For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.
If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water, call qualified technician for examination and repair.
Keep the cord away from heated or hot surfaces.
DO NOT move or relocate concentrator by pulling on the cord.
NEVER drop or insert any object into any opening.
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like. Fill humidifier with water to the level shown by the manufacturer. DO NOT overfill.
Keep unit at least three (3) inches away from walls, draperies, furniture, and the like.
Invacare recommends that Crush-Proof oxygen tubing be used with this product and not exceed 50 ft. (15.2m) in length. Recommended use for Platinum 10 is up to 25 ft. in length of high flow tubing.
Platinum™Series 6 Part No 1118353
SECTION 1—GENERAL GUIDELINES
MAINTENANCE
The Invacare Oxygen Concentrator was specifically designed to minimize routine preventive maintenance. Only professionals of the healthcare field or persons fully conversant with this process such as factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator.
RADIO FREQUENCY INTERFERENCE
Most electronic equipment is influenced by Radio Frequency Interference (RFI). Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment.
TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS.
Avoid using while bathing. If continuous usage is required by the physician’s prescription:
The concentrator MUST be located in another room at least seven (7) feet from the bath.
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into water or other liquid.
DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.
A product should NEVER be left unattended when plugged in.
This device is to be used only in accordance with the prescription of a physician and this Owner’s Manual. If at any time the patient or attendant conclude that the patient is receiving an insufficient amount of oxygen, the supplier and/or physician should be contacted immediately.
DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.
Part No 1118353 7 Platinum™Series
SECTION 1—GENERAL GUIDELINES
Close supervision is necessary when this product is used near children or physically-challenged individuals.
Use this product for only intended use as described in this manual.
DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.
Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance.
In certain circumstances oxygen therapy can be hazardous. Invacare recommends that you seek medical advice before using this machine.
This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining.
Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.
POLARIZED PLUG INSTRUCTION
As a safety feature, this appliance has a polarized plug (one blade is wider than the other). This plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. DO NOT attempt to defeat this safety feature.
Platinum™Series 8 Part No 1118353
SECTION 2—FEATURES
SECTION 2—FEATURES
Oxygen Purity Indicator Lights
(SensO
Oxygen Outlet
Flowmeter
and Power Indicator Lights
model only)/ Fault
2
(LX model only)
Power Switch
Circuit Breaker
Elapsed Time Meter
Power
Cord
REAR VIEW
Outlet
Fitting
Cabinet Filter
(one of two shown)
ACCESSORIES (NOT SHOWN):
Remote Humidifier/Flowmeter Accessory -
IRCRF5
PreciseRX™ Pediatric Humidifier/Flowmeter
Accessory - IRCPF16
Venture® HomeFill II home oxygen
compressor - IOH200
NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the Venture HomeFill II home oxygen compressor. The outlet fitting does not affect concentrator performance. Refer to the HomeFill II owner's manual, part number 1100873, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet fitting. For more information about the HomeFill II, contact your Invacare dealer.
Part No 1118353 9 Platinum™Series
SECTION 3—HANDLING
SECTION 3—HANDLING
The concentrator should always be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator.
If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare.
Unpacking
NOTE: For this procedure, refer to FIGURE 3.1.
1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
NOTE: Unless the oxygen concentrator is to be used immediately, retain containers and packing materials for storage until use of the concentrator is required.
FIGURE 3.1 Unpacking
Inspection
1. Inspect/examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components.
Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place objects on top of repackaged concentrator.
Platinum™Series 10 Part No 1118353
SECTION 4—SPECIFICATIONS
SECTION 4—SPECIFICATIONS
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Rated Current Input: Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Sound Level: Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Altitude: Platinum 5 and XL - Up to 6,000 ft. (1828
meters) above sea level without degradation of concentration levels. From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below 90% efficiency
Platinum 10 - Up to 4,000 ft. (1230 meters) above sea level without degradation of con­centration levels. From 4,000 ft. (1230 meters) to 13,129 ft. (4000 meters) below 90% efficiency
*Oxygen Output Concentration Levels:
Maximum Outlet Pressure: 5 ± 0.5 psi. (34.5 ± 3.45 kPa.)
All 5LXO2/5LX models 95.6% to 87% at 0.5 to 5 L/min. and all 10LXO2/10LX models
95.6% to 87% at 2 L to 10 L/min.
*NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).
Part No 1118353 11 Platinum™Series