Invacare PLATINUM 10L, IRC10LX User Manual

Invacare®Platinum™10L OxygenConcentrator
IRC10LX
enHomeFill®SystemCompatible
UserManual
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
Contents
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................5
1.3IndicationsForUse.............................6
1.4Description...................................6
1.5Contraindications..............................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................9
2.3RadioFrequencyInterference......................12
3Components....................................13
3.1ComponentIdentification.........................13
3.2PneumaticDiagram.............................14
4Accessories.....................................15
4.1OptionalAccessories............................15
5Setup..........................................16
5.1Unpacking....................................16
5.2Inspection....................................16
5.3Storage......................................16
6Usage..........................................17
6.1Introduction..................................17
6.2SelectaLocation...............................17
6.3SettingUptheConcentrator......................18
6.3.1SettingUptheHumidifier......................18
6.3.2TurningtheConcentratorOn...................20
6.3.3Flowrate..................................21
6.3.4UnitswithoutSensO
2
........................21
7Maintenance....................................22
7.1ServiceLife...................................22
7.2CleaningtheCabinet............................22
7.3CleaningtheCabinetFilter........................23
7.4HumidifierCleaningandThermicDisinfection..........24
7.5CleaningandDisinfectionBetweenPatients............24
7.6PreventiveMaintenanceChecklist...................25
8AfterUse.......................................26
8.1RecyclingInformation............................26
8.2WearandTear................................26
9Troubleshooting.................................27
9.1Troubleshooting...............................27
10Technicaldata..................................28
10.1Specifications.................................28
11Warranty......................................31
11.1NorthAmericanLimitedWarranty.................31
Invacare®Platinum™10LOxygenConcentrator
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe practiceswhichcouldresultinpersonalinjuryorpropertydamage. Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor efficient,trouble-freeuse.
Refertothefollowingtableforthemeaningofsymbols markedontheequipmentand/orpackaging.
SymbolsinDocumentation
GeneralWarningSign
Thebackgroundcolorinsidethe triangleisyellowonproductlabels.
Readmanual
Thecolorofthesymbolbackground isblueonproductlabels.
NoSmoking
Thecolorofthecirclewithdiagonal barisredonproductlabels.
NoOpenFlame
Thecolorofthecirclewithdiagonal barisredonproductlabels.
ClassII,DoubleInsulated
Protectedagainstsolidforeignobjectsof12.5 mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
IndoorUseONLY
KeepDry
4
1118353-M-00
General
-20F MIN
150F MAX
TransportandStorageTemperature
15%
MIN
95% MAX
TransportandStorageHumidity
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
ElectricalHazard
Manufacturer
Callsupplier
SymbolsonProduct
Unitrunning
Unitnotrunning
LXIndicators
SymbolO
2
PurityIndicatorLights(LED)
I/O
SYSTEMOKAYGREENIndicatorLight
SYSTEMFAILURE
REDIndicatorLight
ContinuousAudibleAlarm Sieve—GARD™
CompressorShutdown
Callaqualifiedtechnician.
1.2IntendedUse
TheintendedfunctionanduseoftheInvacare®Platinum™Oxygen Concentratoristoprovidesupplementaloxygentopatientswith respiratorydisorders,byseparatingnitrogenfromroomair,byway ofamolecularsieve.Itisnotintendedtosustainorsupportlife.
DANGER! RiskofInjuryorDeath
Thisproductistobeusedasanoxygensupplementand isnotintendedtobelife-supportingorlife-sustaining. ONLYusethisproductifthepatientiscapableof spontaneousbreath,abletoinhaleandexhalewithout theuseofamachine.
–DONOTuseinparallelorserieswithotheroxygen
concentratorsoroxygentherapydevices.
1118353-M-005
Invacare®Platinum™10LOxygenConcentrator
DANGER! RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientson flowratesgreaterthan5L/minmaybeatanincreased riskforseriousinjuryordeathintheeventoffailure.
–ALWAYSdiscussthisincreasedriskwithyourhealth
caresupplierBEFOREusingthisproductifyouare prescribedaflowrategreaterthan5L/min.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand specificationshasnotbeentestedandmayleadto productdamage,lossofproductfunction,orpersonal injury.
–DONOTusethisproductinanywayotherthan
describedintheSpecificationsandIntendedUse sectionsofthismanual.
DANGER! RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen concentratorinthemarkettoday,thisoxygen concentratorcanfailtoproduceoxygenduetopower failureordevicemalfunction.
–ALWAYShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient toswitchtotheirbackupsourceofoxygen.Referto Troubleshootingformoredetail.
1.3IndicationsForUse
TheintendedfunctionanduseoftheInvacare®Platinum™Oxygen Concentratoristoprovidesupplementaloxygentopatientswith respiratorydisorders,byseparatingnitrogenfromroomair,byway ofamolecularsieve.Itisnotintendedtosustainorsupportlife.
1.4Description
TheInvacarePlatinumconcentratorisusedbypatientswith respiratorydisorderswhorequiresupplementaloxygen.Thedevice isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87% to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse ofanasalcannula.
TheInvacarePlatinumconcentratorusesamolecularsieveand pressureswingadsorptionmethodologytoproducetheoxygen gasoutput.Ambientairentersthedevice,isfilteredandthen compressed.Thiscompressedairisthendirectedtowardoneof twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe patient.
TheInvacarePlatinumconcentratoriscapableofoperationbythe patientinahomeenvironmentorinaninstitutionalenvironment. Deviceoperatesatanominal120VAC/60Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified technicalpersonnelonly.
1.5Contraindications
Therearenoknowncontraindications.
61118353-M-00
Safety
2Safety
2.1LabelLocations
A
SerialNumberLabelislocatedontheresonator intakeassembly.
B
C
SpecificationandDoubleInsulationLabelislocated onthebackoftheconcentratoratthebase
1118353-M-007
Invacare®Platinum™10LOxygenConcentrator
IRC10LX
D
81118353-M-00
Safety
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyandoperationofthe concentratortheseinstructionsMUSTbefollowed.
DANGER! RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor damage.Thissectioncontainsimportantinformationfor thesafeoperationanduseofthisproduct.
–DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand understandingtheseinstructionsandanyadditional instructionalmaterialsuchasusermanuals,service manualsorinstructionsheetssuppliedwiththis productoroptionalequipment.
–Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional, dealerortechnicalpersonnelbeforeattemptingto usethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot workingcorrectly,contactatechnicianorInvacare forrepair.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy substancesandothercombustiblesareeasilyignited andburnwithgreatintensityinoxygenenrichedairand whenincontactwithoxygenunderpressure.Toavoid fire,death,injuryordamage:
DONOTSMOKEwhileusingthisdevice.DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
NOSMOKINGsignsshouldbeprominentlydisplayed.Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby anytypeoffriction.
–Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutofthe roominwhichthisconcentratorislocatedandaway fromwhereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings, chaircushions,clothing,andawayfromheatedorhot surfacesincludingspaceheaters,stoves,andsimilar electricalappliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesaletoor ontheorderofamedicalpractitionerlicensedbya governmentalagencywherehe/shepractices.
–ONLYalicensedmedicalpractitionermayorderthe
purchaseoruseofthisdevice.
1118353-M-009
Invacare®Platinum™10LOxygenConcentrator
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor injurytopersons:
–DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceable parts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the concentratormustbelocatedinanotherroomat least7ft(2.1m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
–DONOTusefrayedordamagedACpowercords.
WARNING! RiskOfInjuryOrDamage
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare accessories.Accessoriesdesignedbyother manufacturershavenotbeentestedbyInvacareand arenotrecommendedforusewithInvacareproducts.
–Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis device.Youshouldcontactyourlocalhomecare supplierforrecommendationsonwhichofthese deviceswillbebestforyou.Theyshouldalsogiveyou adviceontheproperusage,maintenance,andcleaning.
DANGER! RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway contamination:
–DONOTusetheconcentratorinthepresenceof
pollutants,smoke,fumes,flammableanesthetics, cleaningagents,orchemicalvapors.
WARNING! RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratorisusedby
ornearchildrenorimpairedindividuals.
–Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING! RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing entanglement:
–Keepchildrenandpetsawayfromnasalcannulaand
tubing.
–Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/orimpairedpersons.
WARNING! RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.Check
withyourequipmentsupplierorphysicianto determinehowoftenthecannulashouldbereplaced.
–DONOTsharecannulasbetweenpatients.
101118353-M-00
Safety
WARNING! RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto you.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings shouldbechanged.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorbypulling
onthecord.
–DONOTuseextensioncordswithACpowercord
provided.
–Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhenplugged
in.
–Makesureconcentratorisoffwhennotinuse.
WARNING! RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned tominimizeroutinepreventivemaintenance.Toprevent injuryordamage:
–Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas factorytrainedpersonnelshouldperformpreventive maintenanceorperformanceadjustmentsonthe oxygenconcentrator,exceptfortasksdescribedin thismanual.
–UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION! RiskofDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto water,callequipmentsupplier/qualifiedtechnicianfor examinationandrepair.
–NEVERdroporinsertanyobjectorliquidintoany
opening.
–ForIndoorUseONLY.
1118353-M-00
11