Invacare MicroAir MA90Z, MicroAir MA95Z, MicroAir MA90ZB42, MicroAir MA90ZB48, MicroAir MA95ZB42 User Manual

...
User Manual
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
EN
Invacare® MA90Z Series Mattresses
MA90Z Lateral Rotation with Alternating Pressure and On-Demand Low Air Loss System
Invacare® MA95Z Series Mattresses
MA95Z Lateral Rotation with Alternating Pressure and True Low Air Loss System
MA95ZB42 Lateral Rotation with Alternating Pressure and True Low Air Loss Bariatric 42”
MA95ZB48 Lateral Rotation with Alternating Pressure and True Low Air Loss Bariatric 48”
Invacare® MA90Z Series Mattresses 2 Part No 1148139
© 2011 Invacare Corporation. All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
CONTENTS
Part No 1148139
3 Invacare
®
MA90Z Series Mattresses
1 GENERAL 4
Warnings.................................................................................................................................................................................................................................................. 4
Limited Warranty .................................................................................................................................................................................................................................. 5
2OVERVIEW 6
Label Location......................................................................................................................................................................................................................................... 6
Typical Product Parameters ................................................................................................................................................................................................................ 7
3SAFETY 10
General Guidelines ..............................................................................................................................................................................................................................10
4OPERATION 14
Installing the Mattress Replacement System.................................................................................................................................................................................. 14
Installing the Side Rails........................................................................................................................................................................................................................16
Installing the Power Unit....................................................................................................................................................................................................................16
Connecting the Hose.......................................................................................................................................................................................................................... 16
Connecting the Power Cord.............................................................................................................................................................................................................18
Using the Front Panel..........................................................................................................................................................................................................................19
Powering Up the System....................................................................................................................................................................................................................26
Placing the Patient on the Mattress.................................................................................................................................................................................................27
Transferring Patient From/To a Gurney......................................................................................................................................................................................... 28
Transferring Patient From/To a Wheelchair.................................................................................................................................................................................28
Preparing for CPR Procedure........................................................................................................................................................................................................... 29
5 MAINTENANCE AND TROUBLESHOOTING 31
Cleaning the System............................................................................................................................................................................................................................ 31
Changing the Filter...............................................................................................................................................................................................................................33
Troubleshooting...................................................................................................................................................................................................................................35
1 GENERAL
Invacare® MA90Z Series Mattresses 4 Part No 1148139
1 General
1.1 Warnings
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. See the information below for definitions of the signal words.
DANGER
Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury or both.
!
IMPORTANT
Indicates a hazardous situation that could result in damage to property if it is not avoided.
Gives useful tips, recommendations and information for efficient, trouble-free use.
1 GENERAL
Part No 1148139 5 Invacare® MA90Z Series Mattresses
1.2 Limited Warranty
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS
MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare or a dealer. This warranty is not extended to any other person or entity and is not transferable or assignable to any subsequent purchaser or owner. Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person. This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state. Invacare warrants the mattress and cover when purchased new and unused to be free from defects in materials and workmanship for a period of one year from the date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty. Invacare warrants the control unit when purchased new and unused to be free from defects in materials and workmanship for a period of two years from the date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty.If within such warranty period any such product shall be proven to be defective, such product shall be repaired or replaced, at Invacare's option. This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product. Invacare's sole obligation and your exclusive remedy under this warranty shall be limited to such repair and/or replacement. For warranty service, please contact the dealer from whom you purchased your Invacare product. In the event you do not receive satisfactory warranty service, please write directly to Invacare at the address on the back cover. Provide dealer's name, address, model number, and the date of purchase, indicate nature of the defect and, if the product is serialized, indicate the serial number. Invacare Corporation will issue a return authorization. The defective unit or parts must be returned for warranty inspection using the serial number, when applicable, as identification within thirty days of return authorization date. DO NOT return products to our factory without our prior consent. C.O.D. shipments will be refused; please prepay shipping charges. LIMITATIONS AND EXCLUSIONS: THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS. IN ADDITION, THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE; OR PRODUCTS MODIFIED WITHOUT INVACARE'S EXPRESS WRITTEN CONSENT INCLUDING, BUT NOT LIMITED TO: MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS: PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE; PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE'S CONTROL; PRODUCTS REPAIRED BY ANYONE OTHER THAN AN INVACARE DEALER, SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE. THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER EXPRESS WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.
2 OVERVIEW
Invacare® MA90Z Series Mattresses 6 Part No 1148139
2 Overview
2.1 Label Location
DANGER-EXPLOSION HAZARD: DO NOT use in the presence of flammable anesthetics. CAUTION: Equipment should be connected to a properly grounded receptacle (3-prong). Risk of
Electrical shock. DO NOT remove back. Disconnect air hose before administering CPR.
WARNING
Patient entrapment with bed side rails may cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment. Follow the manufacturer’s instructions. Monitor patient frequently. Read and understand the Owner’s/Operator’s Manual prior to using this equipment. Invacare product manuals are available at www.invacare.com or your dealer. p/n 1150708 Rev A
ATTENTION
Clean Filter Every 5 months and Whenever Dirty. Remove Filter, Wash, Dry & Replace.
This label is on the back of the unit and also contains the serial number.
MA95Z, MA95ZB42 and MA95ZB48 only.
This label is on the foot end of the mattress cover (not shown).
2 OVERVIEW
Part No 1148139 7 Invacare® MA90Z Series Mattresses
2.2 Typical Product Parameters
Electrical Parameters
Performance Parameters
MA90Z power unit is also used on MA90ZB42 and MA90ZB48 systems.
MA95Z power unit is also used on MA95ZB42 and MA95ZB48 systems.
MA90Z MA95Z
INPUT VOLTAGE AC:
110 V
INPUT FREQUENCY:
60 Hz
CURRENT:
1 A 2 A
MAXIMUM POWER CONSUMPTION:
30 ± 10 W 150 ± 10 W
CIRCUIT PROTECTION:
Dual fused,
250 V,
1 A fast blow fuses
Dual fused,
250 V,
5 A fast blow fuses
MODE OF OPERATION:
Continuous
MA90Z MA95Z
WEIGHT CAPACITY
STANDARD MATTRESS: BARIATRIC MATTRESS
42 inch: 48 inch:
350 lbs
650 lbs
1000 lbs
2 OVERVIEW
Invacare® MA90Z Series Mattresses 8 Part No 1148139
PRESSURE ZONE:
2
MAXIMUM FLOW:
50 ± 15 LPM 1275 ± 100 LPM
MAXIMUM PRESSURE:
35 ± 5 mmHg
MAXIMUM FLOW TIMER:
30 minutes 15 minutes
SUPPORT SURFACE INFLATION TIME:
5 - 10 minutes 25 - 60 seconds
PATIENT COMFORT CONTROL PRESSURES:
SOFT PRESSURE: FIRM PRESSURE:
8 ± 4 mmHg
32 ± 4 mmHg
ROTATION TIME:
10, 20, 30, 60 minutes 5 minutes - 4 hrs
ROTATION ANGLE:
0 - 30° ± 5° 0 - 40° ± 5°
PATIENT CONTACT:
Control unit and mattress have
Latex-Free components
MA90Z MA95Z
2 OVERVIEW
Part No 1148139 9 Invacare® MA90Z Series Mattresses
Mechanical Parameters
Environmental Parameters
MA90Z MA95Z
CONTROL UNIT DIMENSIONS (L X W X H):
15” x 6½” x 11”
CONTROL UNIT WEIGHT:
15 lbs
CONTROL UNIT POWER CORD:
10 - 14 Feet Long, Hospital Grade
CONTROL UNIT CONNECTION:
Quad Flow - Four ¼” Single Quick Coupling
Triple Flow - Two ½” and
One ¼” Single Magnetic
Quick Coupling
CONTROL UNIT PACKAGING:
1 piece/box
AIR FILTER:
None
Charcoal Air Filter with
Fire Retardant
MA90Z/MA95Z
OPERATING CONDITIONS
AMBIENT TEMPERATURE: RELATIVE HUMIDITY: ATMOSPHERIC PRESSURE:
50° - 95° F
30% - 75% Non-Condensing
70 - 106 kPa
STORAGE AND SHIPPING CONDITIONS
AMBIENT TEMPERATURE: RELATIVE HUMIDITY: ATMOSPHERIC PRESSURE:
-40° - 158° F 10% - 100%
50 - 106 kPa
3 SAFETY
Invacare® MA90Z Series Mattresses 10 Part No 1148139
3 Safety
3.1 General Guidelines
Contraindications
WARNING
DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
ACCESSORIES WARNINGS
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
!
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE. Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact Invacare/Carrier for
further instruction.
WARNING
ALWAYS consult the patient’s physician before using the MA90Z and MA95Z systems.
3 SAFETY
Part No 1148139 11 Invacare® MA90Z Series Mattresses
Installation
WARNING
The MA90Z and MA95Z series (including bariatric) is recommended to be installed on medical bed frames with bed side or assist rails. It is preferred that the rails be in the raised position whenever a patient is on the bed. Health care professionals assigned to each case should make the final determination whether side or assist rails are warranted after assessing patient risks of entrapment and falls in accordance with State patient restraint legislation or facility interpretation of such legislation.
Controls on the footboard may be obstructed by the power unit on a few bed frames. It may be necessary to relocate the power unit. Refer to Installing the Power Unit
on page 16.
Check that air hoses and power cord are clear of moving bed components before placing a patient on the bed. Operate all bed frame motorized functions through their full range of motion to be certain that there is no pulling, interference or pinching.
DO NOT strap the mattress to the bed frame at the head and foot ends. Secure ALL mattress straps. Secure the straps to the bed deck at the head and foot ends and to the frame at the center of the bed. Otherwise damage to the mattress will occur when the head and foot ends are raised. Refer to Installing the Mattress Replacement System
on page 14.