Invacare MA55, MA51, MA50 User Manual

Owner’s Operator and Maintenance Manual
MA50 Series
MA50 Alternating Pressure
Overlay System
MA51 Alternating Pressure
Mattress System
MA55 Alternating Pressure
Low Air Loss Mattress System
DEALER: This manual MUST be given to the user of the product.
USER:
manual and save for future reference.
BEFORE using this product, read this
For more information regarding
Invacare products, parts, and services,
SPECIAL NOTES
WARNING
DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment
- otherwise, injury or damage may occur.
Procedures other than those described in this manual must be performed by a qualified technician.
ACCESSORY WARNING
Only use Invacare accessories on Invacare products.
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the table below for definitions of the signal words.
SIGNAL WORD MEANING
DANGER
WARNING
CAUTION
Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury.
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE.
Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact Invacare Customer Service Department for further instruction.
NOTE: Updated versions of this manual are available on www.invacare.com.
MA50 Series 2 Part No 1148136
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES ................................................................................ 2
REGISTER YOUR PRODUCT ............................................................... 4
TYPICAL PRODUCT PARAMETERS .................................................... 5
Electrical Parameters .................................................................................................................................5
Performance Parameters ..........................................................................................................................5
Mechanical Parameters..............................................................................................................................5
Environmental Parameters .......................................................................................................................6
LABEL LOCATION ............................................................................ 7
SECTION 1—GENERAL GUIDELINES ................................................... 8
Contraindications .......................................................................................................................................8
Installation ....................................................................................................................................................8
Entrapment May Occur.............................................................................................................................8
Fire Hazard...................................................................................................................................................9
Smoking ....................................................................................................................................................9
Anesthesia Equipment...........................................................................................................................9
Oxygen .....................................................................................................................................................9
Electrical .......................................................................................................................................................9
SECTION 2—OPERATION ................................................................ 10
Installing the Mattress..............................................................................................................................10
Overlay System (MA50) .....................................................................................................................10
Mattress Replacement System (MA51 and MA55) ......................................................................11
Installing the Side Rails ............................................................................................................................12
Installing the Power Unit ........................................................................................................................12
Connecting the Hose...............................................................................................................................12
Connecting the Power Cord .................................................................................................................13
Using the Front Panel ..............................................................................................................................14
Power Button........................................................................................................................................14
Static/Alternating Pressure Button ..................................................................................................15
Adjusting the Minimum Pressure Setting (A/P Low) ..............................................................15
Firm/Soft Buttons.................................................................................................................................16
Fowler Button.......................................................................................................................................16
Max Inflate Button ...............................................................................................................................16
Mode Button (Max Inflate/Low Air Loss).......................................................................................17
Lock Button...........................................................................................................................................17
Alarm Silence Button ..........................................................................................................................17
Power Fail LED.....................................................................................................................................17
Low Pressure LED...............................................................................................................................18
Powering Up the System ........................................................................................................................18
Placing the Patient on the Mattress......................................................................................................18
Transferring Patient From/To a Gurney .............................................................................................19
Transferring Patient From/To a Wheelchair......................................................................................20
Preparing for CPR Procedure................................................................................................................20
About Power Outage and Transportation .........................................................................................20
Part No 1148136 3 MA50 Series
TABLE OF CONTENTS
TABLE OF CONTENTS
SECTION 3—MAINTENANCE AND TROUBLESHOOTING ................... 21
Cleaning the System.................................................................................................................................21
Storing the System...............................................................................................................................22
Troubleshooting........................................................................................................................................23
LIMITED WARRANTY ..................................................................... 24
REGISTER YOUR PRODUCT
The benefits of registering include:
1. Safeguarding your investment.
2. Ensuring long-term maintenance and servicing of your product.
3. Receiving updates with product information, maintenance tips and industry news.
Register ONLINE at warranty.invacare.com
Please have your model number and purchase date available to complete your registration.
Any registration information you submit will only be used by Invacare Corporation and protected as required by applicable laws and regulations.
MA50 Series 4 Part No 1148136
TYPICAL PRODUCT PARAMETERS
TYPICAL PRODUCT PARAMETERS
Electrical Parameters
MA50/MA51/MA55
INPUT VOLTAGE AC: 90 V
INPUT FREQUENCY: 60 Hz
CURRENT: 1 A
MAXIMUM POWER CONSUMPTION: 5 ± 3 W
CIRCUIT PROTECTION: Dual fused, 250 V, 1 A fast blow fuses
MODE OF OPERATION: Continuous
Performance Parameters
MA50/MA51/MA55
WEIGHT CAPACITY
STANDARD MATTRESS: BARIATRIC MATTRESS:
PRESSURE ZONE: MAXIMUM FLOW: MAXIMUM FLOW PRESSURE: MAXIMUM FLOW TIMER: SUPPORT SURFACE INFLATION TIME:
PATIENT COMFORT CONTROL PRESSURES
SOFT PRESSURE: FIRM PRESSURE: CYCLE TIME:
PATIENT CONTACT: Control unit and mattress have Latex-Free
350 lbs
1000 lbs
2
8 ± 4 LPM
35 ± 5 mmHg
45 minutes
20 - 45 minutes
8 ± 4 mmHg
32 ± 4 mmHg
5, 10, 15, 20 minutes
components
Mechanical Parameters
CONTROL UNIT MA50 MA51 MA55
DIMENSIONS (L X W X H):
9” x 5” x 5”
WEIGHT: 5 lbs
POWER CORD: 10 - 14 Feet Long, Hospital Grade
CONNECTION: Two ¼” Flow Couplings Three ¼” Flow Couplings
PACKAGING: 1 piece/box
AIR FILTER: None
Part No 1148136 5 MA50 Series
TYPICAL PRODUCT PARAMETERS
Environmental Parameters
OPERATING CONDITIONS
AMBIENT TEMPERATURE: RELATIVE HUMIDITY: ATMOSPHERIC PRESSURE:
STORAGE AND SHIPPING CONDITIONS
AMBIENT TEMPERATURE: RELATIVE HUMIDITY: ATMOSPHERIC PRESSURE:
MA50/MA51/MA55
50° - 104° F
30% - 75% Non-Condensing
70 - 106 kPa
-40° - 158° F
10% - 100% Non-Condensing
50 - 106 kPa
MA50 Series 6 Part No 1148136
LABEL LOCATION
LABEL LOCATION
DANGER-EXPLOSION HAZARD: DO NOT use in the presence of flammable anesthetics.
CAUTION: Equipment should be connected to a properly grounded receptacle (3-prong). Risk of Electrical shock. DO NOT remove back. Disconnect air hose before administering CPR.
NOTE: This label is on the back of the unit and also contains the serial number.
WARNING
Patient entrapment with bed side rails may cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment. Follow the manufacturer’s instructions. Monitor patient frequently. Read and understand the Owner’s/Operator’s Manual prior to using this equipment. Invacare product manuals are available at www.invacare.com or your dealer.
Label part number 1150708 Rev A
Part No 1148136 7 MA50 Series
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product.
Contraindications
ALWAYS consult the patient’s physician before using the MA50, MA51 and MA55 systems.
Installation
The MA50, MA51 and MA55 systems MUST be installed on medical bed frames with side rails. The side rails MUST be in the raised position whenever a patient is on the bed.
Controls on the footboard may be obstructed by the power unit on a few bed frames. It may be necessary to relocate the power unit. Refer to Installing the Power Unit on page 12.
Check that air hoses and power cord are clear of moving bed components before placing a patient on the bed. Operate all bed frame motorized functions through their full range of motion to be certain that there is no pulling, interference or pinching.
MA51 AND MA55 ONLY - DO NOT strap the mattress to the bed frame at the head and foot ends. Secure ALL mattress straps. Secure the straps to the bed deck at the head and foot ends and to the frame at the center of the bed. Otherwise damage to the mattress will occur when the head and foot ends are raised. Refer to Installing the Mattress on page 10.
Entrapment May Occur
WARNING
Patient entrapment with bed side rails may cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment. Follow the manufacturer’s instructions. Monitor patient frequently. Read and understand the Owner’s/Operator’s Manual prior to using this equipment. Invacare product manuals are available at www.invacare.com or your dealer.
Proper patient assessment and monitoring, and proper maintenance and use of equipment is required to reduce the risk of entrapment. Variations in bed rail dimensions, and mattress thickness, size or density could increase the risk of entrapment. Visit the FDA website at http://www.fda.gov to learn about the risks of entrapment. Review “A Guide to Bed Safety”, published by the Hospital Bed Safety Workgroup, located at www.invacare.com. Use the link located under each bed rail product entry to access this bed safety guide.
Refer to the owner’s manuals for beds and rails for additional product and safety information.
After any adjustments, repair or service and before use, make sure all attaching hardware is tightened securely. Assist rails with dimensions different from the original equipment supplied or specified by the bed manufacturer may not be interchangeable and may result in entrapment or other injury.
MA50 Series 8 Part No 1148136