Invacare®Kirilos/Atlante
Kirilos,AtlanteElectric,AtlanteHydraulic
enMobilePatientLift
UserManual...............................3
frLève-personnemobile
Manueld'utilisation.........................52
itSollevatoremobile
Manualed’uso..............................108
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.

©2017InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited
withoutpriorwrittenpermissionfromInvacare.Trademarksareidentiedby™and®.Alltrademarks
areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
–
Tousdroitsréservés.Larepublication,laduplicationoulamodicationdetoutoupartieduprésent
documentestinterditesansl'accordécritpréalabled'Invacare.Lesmarquescommercialessont
identiéespar™et®.T outeslesmarquescommercialessontdétenuesparoucédéessouslicence
àInvacareCorporationousesliales,saufstipulationcontraire.
–
Tuttiidirittiriservati.Èproibitalariproduzione,laduplicazioneolamodicaparzialeocompleta,
salvopreviaautorizzazionescrittadapartediInvacare.Imarchisonocontrassegnatida™e®.T utti
imarchisonodiproprietàolicenzadiInvacareCorporationodisueafliate,salvoindicazione
contraria.

Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
1General.........................................4
1.1Introduction..................................4
1.2Symbolsinthismanual..........................4
1.3Intendeduse..................................4
1.4Scopeofdelivery..............................5
1.5ServiceLife...................................6
1.6WarrantyInformation...........................6
1.7Limitationofliability............................6
2Safety..........................................7
2.1Safetyinformation..............................7
2.2OperatingInformation...........................7
2.2.1General...................................7
2.2.2PinchPointsandPositioning....................8
2.3RadioFrequencyInterference.....................8
2.4Productlabeling...............................8
3Components......................................10
3.1Mainpartsofthelift............................10
3.2Accessories...................................11
4Setup...........................................12
4.1SafeAssembly.................................12
4.2AssemblingtheLift.............................12
4.3DisassemblingtheLift...........................13
5Usage..........................................14
5.1Introduction..................................14
5.2Locking/UnlockingtheRearCasters.................14
5.3Raising/LoweringtheLift.........................14
5.3.1Raising/LoweringanElectricLift.................14
5.3.2Raising/LoweringaHydraulicLift................15
5.4Closing/OpeningLegs...........................16
5.4.1Closing/Openinglegsmanually..................16
5.5PerforminganEmergencyStop....................16
5.6ActivatinganEmergencyRelease...................16
5.7Chargingthebattery............................17
6LiftingthePatient.................................19
6.1SafeLifting...................................19
6.2PreparingtoLift...............................21
6.3AttachingtheSlingtotheLift.....................23
6.4LiftingandTransferringthePatientFromaBed........24
6.4.1FloorTransfers(LiftingfromtheFloor)............28
6.4.2CommodeTransferGuidelines..................30
6.4.3WheelchairTransfer..........................32
7Maintenance.....................................34
7.1MaintenanceandSafetyInspection.................34
7.1.1Safetyinspectionchecklist.....................36
7.2LubricatingtheLift.............................37
7.3CleaningtheSlingandLift........................37
7.4Checkingandtighteningbolts.....................38
7.5MaintainingtheHydraulicPump...................38
8AfterUse........................................39
8.1TransportationandStorage.......................39
8.2Disposal.....................................39
8.3Reuse.......................................39
9Troubleshooting...................................40
9.1Identifyingandrepairingfaults....................40
10T echnicaldata...................................42
10.1Dimensionsandweight.........................42
10.2Electricalsystem..............................43
10.3Hydraulicsystem..............................44
10.4Environmentalconditions........................44
10.5Materials....................................44
10.6Electromagneticcompliance(EMC)information.......44
10.7Electromagneticcompliance(EMC).................46

Invacare®Kirilos/Atlante
1General
1.1Introduction
ThankyouforchoosinganInvacareproduct.
Thisusermanualcontainsimportantinformationaboutthe
handlingoftheproduct.Inordertoensuresafetywhen
usingtheproduct,readtheusermanualcarefullyandfollow
thesafetyinstructions.
Pleasenotethattheremaybesectionsinthisusermanual,
whicharenotrelevanttoyourproduct,sincethismanual
appliestoallavailablemodels(onthedateofprinting).If
nototherwisestated,eachsectioninthismanualrefersto
allmodelsoftheproduct.
1.2Symbolsinthismanual
Symbolsandsignalwordsareusedinthismanualandapply
tohazardsorunsafepracticeswhichcouldresultinpersonal
injuryorpropertydamage.Seetheinformationbelowfor
denitionsofthesignalwords.
WARNING
Indicatesahazardoussituationthatcouldresult
inseriousinjuryordeathifitisnotavoided.
CAUTION
Indicatesahazardoussituationthatcouldresult
inminororslightinjuryifitisnotavoided.
IMPORTANT
Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
4
TipsandRecommendations
Givesusefultips,recommendationsand
informationforefcient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.Thelaunchdate
ofthisproductisstatedintheCEdeclaration
ofconformity.
Manufactureroftheproduct.
1.3Intendeduse
Mobilepatientliftsareindoortransferdevices,designed
tobeusedinmostofthecommonliftingsituationsin
hospitals,nursingfacilitiesanddomesticareas,forexample:
•Betweenthebedandawheelchair
•Toandfromthetoilet
•Loweringandraisingpatientsto/fromtheoor
Themobilepatientliftcanbeusedtotransferandposition
completelyorpartiallyimmobilepatients,whocannotbe
transferredwithothertypesofliftsortransferaids.All
positionchangesarepossiblewithoutassistanceofthe
patient.Themobilepatientliftisonlyintendedtolift
patientsuptothemaximumweightlimitstatedintechnical
data.Therearenoknowncontraindicationsforthisproduct.
Selectingtheappropriateslingsandaccessoriesforeach
individualisimportanttoassuresafetywhenusingapatient
lift.SeeInvacare’sslingandaccessoryusermanualsfor
furtherinformationonthosedevices.
1551341-E

General
Invacarerecommendsthatthepatientbetransferredtoa
showerchairorothermeansforbathing.
Themobilepatientliftcanbeturned(rotated)inplacefor
transferswithlimitedoorspace.
1.4Scopeofdelivery
A
B
C
D
E
F-notshown
G-notshown
H-notshown
Mastassemblyincludingboom,hanger
barandactuatororhydraulicpump(1
piece)
Battery(1piece,KirilosandAtlante
Electriconly)
Handcontrol(1piece,Kirilosand
AtlanteElectriconly)
Legspreaderhandle(1piece)
Basewithlegs(1piece)
Cableforcharging(1piece,Kirilosand
AtlanteElectriconly)
Sling(1piece)
UserManual(1piece)
Theitemslistedinthetableareincludedinyourpackage.
1551341-E5

Invacare®Kirilos/Atlante
1.5ServiceLife
WARNING!
RiskofInjuryorDamage
Theproducthasbeentestedfortheservicelife
statedinthismanual.Useoftheproductbeyond
thistimeperiodmaycauseproductdamageand
injury.
–Onlyusetheproductfortheservicelifestated
inthismanual.DONOTexceedtheservicelife
oftheproduct.
–Removetheproductfromservicewhenthe
servicelifehasbeenmet.DONOTcontinueto
usetheproduct.
–Performallmaintenanceaccordingtothe
recommendedscheduleinthismanual.
Theproducthasanexpectedlifetimeof8yearswhenused
inaccordancewithsafetyinstructions,maintenanceintervals
andcorrectuse,statedinthismanual.Theeffectiveproduct
lifetimecanvaryaccordingtofrequencyandintensityofuse.
ActuatorServiceLife
Numberofliftsperday
1–210
39
46
55
64
Servicelifeofactuator
(inyears)
Numberofliftsperday
7–93
10–132
14–271
Servicelifeofactuator
(inyears)
1.6WarrantyInformation
Termsandconditionsofthewarrantyarepartofthegeneral
termsandconditionsparticulartotheindividualcountriesin
whichthisproductissold.
ContactinformationforyourlocalInvacareofceislocated
insidethebackcoverofthismanual.
1.7Limitationofliability
Invacareacceptsnoliabilityfordamagearisingfrom:
•Non-compliancewiththeusermanual
•Incorrectuse
•Naturalwearandtear
•Incorrectassemblyorset-upbythepurchaserora
thirdparty
•Technicalmodications
•Unauthorizedmodicationsand/oruseofunsuitable
spareparts
61551341-E

Safety
2Safety
2.1Safetyinformation
WARNING!
–Donotusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.Ifyouareunabletounderstand
thewarnings,cautionsorinstructions,contact
ahealthcareprofessional,dealerortechnical
personnelbeforeattemptingtousethis
equipment-otherwise,injuryordamagemay
occur.
Theinformationcontainedinthisdocumentis
subjecttochangewithoutnotice.
Checkallpartsforshippingdamagebeforeusing.Incaseof
damage,Donotusetheequipment.Contactthedealeror
Invacarerepresentativeforfurtherinstructions.
2.2OperatingInformation
Thissectionofthemanualcontainsgeneralsafety
informationaboutyourproduct.Forspecicsafety
information,refertotheappropriatesectionofthemanual
andprocedureswithinthatsection.Forinstance,forsafety
informationrelatedtoassemblingthelift,refertotheSetup
sectionofthemanual.
1551341-E
2.2.1General
WARNING!
RiskofFalling
TheInvacaremobilepatientliftisNOTatransport
device.Itisintendedtotransferanindividual
fromonerestingsurfacetoanother(suchasa
bedtoawheelchair).
Invacareslingsandpatientliftaccessoriesare
specicallydesignedtobeusedinconjunction
withInvacarepatientlifts.
WARNING!
RiskofFalling
Donotattemptanytransferwithoutapproval
ofthepatient’sphysician,nurseormedical
assistant.Thoroughlyreadtheinstructionsin
thisUserManual,observeatrainedteamof
expertsperformtheliftingproceduresandthen
performtheentireliftprocedureseveraltimes
withpropersupervisionandacapableindividual
actingasapatient.
–Usecommonsenseinalllifts.Specialcare
mustbetakenwithpeoplewithdisabilities
whocannotcooperatewhilebeinglifted.
–Usesteeringhandleonthemastatalltimesto
pushorpullthepatientlift.
–Besuretochecktheslingattachmentseach
timetheslingisremovedandreplaced,to
ensurethatitisproperlyattachedbeforethe
patientisremovedfromastationaryobject
(bed,chairorcommode).
7

Invacare®Kirilos/Atlante
Product name
YYYY-MM
= xxx kg
Invacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio, Portugal
XXXXXX XXXXX_XX
XXXXXXXXXXXXX
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
ISO 10535
1590833-A
WARNING!
–Thepatientliftcanbeusedindoorsonalevel
surface.Ifthepatientliftisusedintheareaof
ashowerorbath,ensurethatthepatientliftis
wipedcleanofanymoistureafteruse.
–Periodicallyinspectallcomponentsofthe
patientliftforsignsofcorrosion.Replaceall
partsthatarecorrodedordamaged.
2.2.2PinchPointsandPositioning
2.3RadioFrequencyInterference
WARNING!
RiskofInjuryorDamage
MostelectronicequipmentisinuencedbyRadio
FrequencyInterference(RFI).
CAUTIONshouldbeexercisedwithregardtothe
useofportablecommunicationequipmentinthe
areaaroundsuchequipment,otherwiseinjury
ordamagemayoccur.
IfRFIcauseserraticbehavior:
–PUSHtheRedPowerSwitchOFFIMMEDIATELY.
–DONOTturnthePowerSwitchONwhile
transmissionisinprogress.
2.4Productlabeling
Labellocation
Theproductlabelisplacedonthemastofthepatient
liftandcontainsthemainproductinformation,including
WARNING!
RiskofInjury
Pinchpointsarepresentinseverallocationson
theliftandngerscouldbepinched.
Thehangerbarcanmovesuddenlyandcause
injury.
–Alwayskeephandsandngersclearofmoving
parts.
–Whenpositioninglift,beawareoftheposition
ofthehangerbarandthepatient.
81551341-E
technicaldata.

Safety
Symbolsontheproductlabel
Manufactureraddress
Dateofmanufacture
TypeReferenceNumber
SerialNumber
SafeWorkingLoad
ClassIIequipment
(KirilosandAtlanteElectriconly)
TypeBappliedpart
(KirilosandAtlanteElectriconly)
ThisproductcomplieswithDirective
93/42/EECconcerningmedicaldevices.The
launchdateofthisproductisstatedinthe
ECdeclarationofconformity.
WEEEconform.
(KirilosandAtlanteElectriconly)
Otherlabelsandsymbols
ReadManual
(KirilosandAtlanteElectriconly)
ReadManual
(AtlanteHydrauliconly)
1551341-E9

Invacare®Kirilos/Atlante
3Components
3.1Mainpartsofthelift
Base
A
Leg
B
Caster
C
Rearcasterwithbrake
D
Boom
E
Mast
F
Hangerbar
G
Hookforsling
H
Actuator(KirilosandAtlanteElectriconly)
I
Controlunitwithbattery(KirilosandAtlanteElectric
J
only)
EmergencyStop(KirilosandAtlanteElectriconly)
K
Steeringhandle
L
Handcontrol(KirilosandAtlanteElectriconly)
M
Legspreaderhandle
N
OHydraulicpump–replacesactuatorIforAtlante
Hydraulic
Manualemergencyloweringdevice(Kirilosand
P
AtlanteElectriconly)
101551341-E

Components
3.2Accessories
CAUTION!
Compatibilityofslingsandhangerbars/strap
hooks
Invacare®usesa"LoopandCoatHanger
Bar/StrapHookSystem"asdomanyother
manufacturers.Thereforeothersuitablepatient
AvailableAccessories
Slingmodelsfor"LoopandCoatHangerBarSystem":
•Fullbodysupportslings–withoutheadsupport
•Fullbodysupportslings–withheadsupport
•Slingsfordress/toileting–withorwithoutheadsupport
•Slingsforamputee
transfersystems(slings),manufacturedbyother
companies,canbeusedontheInvacarepatient
liftrangeaswell.
Howeverwedorecommend:
–Ariskassessmentisalwaystobecarriedoutby
aprofessionalpriortoissuingliftingequipment.
ItisimportantthattheTask,Individual,Load,
EnvironmentandEquipmentareconsideredin
theriskassessment.
–Alwayschoosetheslingdesignandsize
accordingtothepatient’sweight,sizeand
physicalabilitywhilstconsideringthetypeof
transfertobecarriedout.
–Doonlyuseslingsthataresuitablefora"Loop
andCoatHangerBar/StrapHookSystem".
–Donotuseslingsfor"KeyholeHangerBar"or
for"TiltingFrameHangerBar"designs.
1551341-E
11

Invacare®Kirilos/Atlante
4Setup
4.1SafeAssembly
WARNING!
RiskofInjury
Improperassemblymaycauseinjuryordamage.
–AssemblyMUSTbeperformedonlybyqualied
personnel.
–UseonlyInvacarepartsintheassemblyof
thispatientlift.Thebaselegs,themast,
boom,pumpassemblyandthehangerbar
aremanufacturedtospecicationsthatassure
correctalignmentofallpartsforsafefunctional
operation.
–DONOTovertightenthemountinghardware.
Thiswilldamagethemountingbracket.
Ifthereareanyissuesorquestionsduringassembly,
contactalocalInvacarerepresentative.Refertothe
contactinformationonthelastpageinthismanual.
4.2AssemblingtheLift
WARNING!
–Checkallpartsforvisibledefectsordamage
beforeassembly.Incaseofanydamage,do
notusetheproductandcontactInvacare.
–Makesuretheemergencystopisactivated
beforeassemblyordisassembly.
–T akecarewhenliftingcomponentsduring
assembly.Somepartsareheavy.Always
remembertoadoptthecorrectliftingposition.
Assemblingthemasttothebase
12
1551341-E

Setup
1.PutthebaseAonalevelsurfaceandlocktherear
castorsB.
2.LoosenthemastknobC,butleaveitscrewedintothe
baseA.
3.InsertthetubeofthemastDintotheholeofthebase
A.
4.WhilesupportingthemastassemblyD,tightenthe
mastknobC.
5.PlacethelegspreaderhandleEontoitssupportF
andpushthespringpin,tolockthehandleEonits
supportF.
Installingtheactuatororhydraulicpumptotheboom
2x17mmwrench
Wiring
(KirilosandAtlanteElectriconly)
1.PlugthehandcontrolcableonsocketAandmotor
cableonsocketBcontrolunit.
2.PlacethebatteryConthecontrolunit.Pushthetop
ofthebatteryagainstthemountingbracketuntilthere
isanaudibleclick.
4.3DisassemblingtheLift
1.Putthebaseonalevelsurfacesothatallcastorsmake
contactwiththeoor.
2.Locktherearcastors.
3.Reversetherelevantstepsoftheassemblingprocedure
foryourmodel.
1.LoosenthenutAandremoveboltBfromtheboom
mountingbracketC.
2.PlacetheactuatorDorhydraulicpumpEintheboom
mountingbracketCandaligntheholes.
3.ReinstalltheboltBandtightenthenutA.
1551341-E13

Invacare®Kirilos/Atlante
5Usage
5.1Introduction
Theoperationofthepatientliftisaneasyandsafe
procedure.
Beforeusingtheliftwithapatient,refertothe
followingproceduresforsafetyinformationand
instruction:
•OperatingInformationintheSafetysection
•LiftingandTransferringthePatientintheLifting
thePatientsection
5.2Locking/UnlockingtheRearCasters
1.ToLocktheCasterA:PresstabBdown.
2.ToUnlocktheCasterA:PresstabBup.
5.3Raising/LoweringtheLift
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
–Invacaredoesrecommendthattherearcasters
beleftunlockedduringliftingproceduresto
allowthepatientlifttostabilizeitselfwhenthe
patientisinitiallyliftedfromachair,bedor
anystationaryobject.
5.3.1Raising/LoweringanElectricLift
Thehandcontrolisusedtoraiseorlowerthelift.
1.Toraisethelift—Press
andholdtheUPAbutton
toraisetheboomandthe
patient.
2.Tolowerthelift—Press
andholdtheDOWNB
buttontolowertheboom
andthepatient.
Releasethebuttontostopraisingorloweringthelift.
14
1551341-E

Usage
5.3.2Raising/LoweringaHydraulicLift
Thepumphandleandcontrolvalveareusedtoraiseor
lowerthehydrauliclift.
RefertothesafetyinformationintheIntroductionof
theUsagesectionbeforeperformingthisprocedure.
Therearetwocontrolsonthepumpassembly:
•ThecontrolvalveA
•ThepumphandleB
Toraisethelift:
1.TurnthecontrolvalveAclockwisetotheCLOSED
position.
2.MovethepumphandleBupanddowntoelevatethe
boomC.
Tolowerthelift:
1.TurnthecontrolvalveAcounterclockwisetotheOPEN
position.
2.TheweightofthepatientintheslingDwillcausethe
boomtodescend.
3.Ifthereisnopatientinthesling,pulldownonthe
boomC.
Therateofdescentoftheboomiscontrolledbythe
amountthatthecontrolvalveisopen.
Asafetygateispartofthehydraulicsystem
thatcontrolsthemaximumdescentoftheboom
regardlessofhowfarthecontrolvalveisopened.
1551341-E15

Invacare®Kirilos/Atlante
5.4Closing/OpeningLegs
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
–Thelegsoftheliftmustbeinthemaximum
openpositionforoptimumstabilityandsafety.
Ifitisnecessarytoclosethelegsoftheliftto
maneuvertheliftunderabed,closethelegs
oftheliftonlyaslongasittakestoposition
theliftoverthepatientandliftthepatientoff
thesurfaceofthebed.Whenthelegsofthe
TheshifterhandleMUSTlockintoitsmountingslot
tolockthelegsinthefullclosedposition.
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
–IfthelegspreaderhandleisNOTpositioned
completelyintoitsmountingslot,DONOTuse
thepatientliftuntilthelegspreaderhandle
isproperlyseatedandthelegsofthepatient
liftarelockedinplace.
liftarenolongerunderthebed,returnthe
legsofthelifttothemaximumopenposition.
5.4.1Closing/Openinglegsmanually
5.5PerforminganEmergencyStop
(KirilosandAtlanteElectriconly)
1.PresstheredemergencybuttonAonthecontrolunit
tostoptheboomandpatientfromraisingorlowering.
2.Toreset,rotatetheemergencybuttonclockwise.
1.Toopenthelegs,pullthelegspreaderhandleAoffand
thentotheleftuntilitlocksinthenotchofthebracket.
2.Toclosethelegs,pushthelegspreaderhandleAoff
andthentotherightuntilitlocksinthenotchofthe
bracket.
161551341-E
5.6ActivatinganEmergencyRelease
Electricalemergencyreleasewithcontrolunit
(KirilosandAtlanteElectriconly)

1.InsertapenintotheholelabeledEmergencyDownA
onthecontrolunitB
Secondary(manual)emergencyreleasewithactuator
(KirilosandAtlanteElectriconly)
Usage
Itisrecommendedthattheprimaryemergency
releasebeused.Thesecondary(manual)emergency
releaseisonlyaback-uptotheprimaryemergency
release.
1.RotatetheemergencyreleaseknobAcounterclockwise.
Theboomwilllowerslowlywithoutadditional
downwardforceontheboom.
Emergencyreleasewithhydraulicpump
(AtlanteHydrauliconly)
Incaseofpartialortotalpowerfailure,orifthebattery
runsdownwhileusingthelift,itisequippedwithamanual
emergencyreleasesystemlocatedatthebottomofthe
actuator.
1551341-E
1.TurnthecontrolvalveAcounterclockwisetotheOPEN
position.
5.7Chargingthebattery
(KirilosandAtlanteElectriconly)
17

Invacare®Kirilos/Atlante
IMPORTANT!
–Makesuretheemergencystopisnotactivated
whilechargingthebattery.
–Donotusetheliftwhilechargingthebattery.
–Makesurethatchargingtakesplaceinaroom
withgoodairventilation.
–Donotuseormovetheliftwithoutunplugging
fromthewalloutletaftercharging.
–Donotattempttousetheliftifthebattery
housingisdamaged.
–Replaceadamagedbatteryhousingbefore
furtheruse.
Itisrecommendedtochargethebatterydailytoensure
optimaluseoftheliftandprolongthelifeofthebattery.
Furthermore,itisrecommendedtochargethebattery
beforerstuse.
Controlunit
Thecontrolunitisequippedwithasoundsignal.Abeep
indicatesthatthebatteryhaslowcapacity,butloweringthe
patientisstillpossible.Itisrecommendedtochargethe
batteriesassoonasthesoundsignalisheard.
1.PlugthecableforchargingCintoapoweroutlet.
Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhen
thebatteriesarefullycharged.
TherightyellowchargediodeAwillcontinuous
lightduringchargingandswitchoffwhenfully
charged.
TheleftgreendiodeBwilllightcontinuously
whilethecontrolunitisconnectedtothemains,
andlightsupwhenanybuttononthehand
controlispressed.
2.Disconnectthepowercordfromthepoweroutletafter
thebatteryhasbeenfullycharged.
181551341-E

LiftingthePatient
6LiftingthePatient
6.1SafeLifting
WARNING!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycausedeath,
injuryordamage.
Theliftcouldtipandendangerthepatientand
assistants.
TheInvacaremobilepatientliftisNOTatransport
device.Itisintendedtotransferanindividual
fromonerestingsurfacetoanother(suchasa
bedtoawheelchair).
–WheelchairandbedwheellocksMUSTbein
alockedpositionbeforeloweringthepatient
ontoorliftingthepatientoffofthewheelchair
orbedtopreventthewheelchairorbedfrom
movingduringtransfer .
–Beforetransferring,checkthatthewheelchair
(bed,commodeorothersurface)weight
capacitycanwithstandthepatient’sweight.
WARNING!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycausedeath,
injuryordamage.
Theliftcouldtipandendangerthepatientand
assistants.
TheInvacaremobilepatientliftisNOTatransport
device.Itisintendedtotransferanindividual
fromonerestingsurfacetoanother(suchasa
bedtoawheelchair).
–Thelegsoftheliftmustbeinthemaximum
openpositionforoptimumstabilityandsafety.
Ifitisnecessarytoclosethelegsoftheliftto
maneuvertheliftunderabed,closethelegs
oftheliftonlyaslongasittakestoposition
theliftoverthepatientandliftthepatientoff
thesurfaceofthebed.Whenthelegsofthe
liftarenolongerunderthebed,returnthe
legsofthelifttothemaximumopenposition.
–Invacarerecommendslockingtherearcasters
oftheliftONL Ywhenpositioningorremoving
theslingfromaroundthepatient.
–Invacarerecommendsthattherearcasters
beleftunlockedduringliftingproceduresto
allowthepatientlifttostabilizeitselfwhenthe
patientisinitiallyliftedfromachair,bedor
anystationaryobject.
1551341-E19

Invacare®Kirilos/Atlante
WARNING!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycausedeath,
injuryordamage.
–Usethesteeringhandleonthemastatall
timestopushorpulltheproduct.
–Avoidusingtheproductonanincline.Invacare
recommendsthattheproductonlybeusedon
aatsurface.
–Duringtransfer,withthepatientsuspended
inaslingattachedtothelift,DONOTrolllift
overunevensurfacesthatcouldcausethelift
totipover .
WARNING!
RiskofInjuryorDamage
Damagetopartsofthelift(pendant,casters,
etc.)causedbyimpactwiththeoor,wallsor
otherstationaryobjectsmaycausedamageto
theproductandleadtoinjury.
–DONOTallowpartsofthelifttoimpactthe
oor,wallsorotherstationaryobjects.
–ALWAYSstorethependantproperlywhennot
inuse.
WARNING!
RiskofDeath
Thependantcordcancauseinjuryifimproperly
positionedandsecured.
–ALWAYSbeawareofthelocationofthe
pendantcordinrelationtothepatientand
caregivers.
–DONOTallowthependantcordtobecome
entangledaroundthepatientandcaregivers.
–Thependantmustbesecuredproperly.ALWAYS
storethependantproperlywhennotinuse.
WARNING!
RiskofEntrapmentorStrangulation
Itemsinthepatient’ssurroundingscancause
entrapmentstrangulationduringlifting.Toavoid
entrapmentorstrangulation:
–Beforelifting,checkthatthepatientis
completelyfreeofhis/hersurroundings.
WARNING!
RiskofEntrapment
Thereisariskofentrapmentbetweenthehanger
barhooksandthesling.
–Usecautionwhenlifting.
–NEVERputhandsorngersonornearthe
hookswhenlifting.
–Ensurethepatient’shandsandngersareaway
fromthehooksbeforelifting.
201551341-E

LiftingthePatient
6.2PreparingtoLift
WARNING!
RiskofInjury
Duringtransfersandliftoperation,theboomcan
impactthepatientorcaregiversandcauseinjury.
–ALWAYSbeawareofthepositionoftheboom
duringtransfers.
–Ensuretheboomispositionedinawaythatit
cannotimpactthepatientorbystanders.
–ALWAYSbeawareofyourbodypositionin
relationtotheboomduringtransfers.
1.RefertotheSafetysectionofthemanualandreview
theinformationinSafeLiftingintheLiftingthePatient
sectionofthemanualbeforeproceedingfurther.
Observeallwarningsindicated.
2.PositionthepatientontotheslingA.Refertoyour
slingusermanual.
4.Openthelegsofthelift.RefertoClosing/OpeningLegs
oftheLiftintheUsagesectionofthemanual.
3.Unlocktherearcasters.RefertoLocking/Unlockingthe
RearCastersintheUsagesectionofthemanual.
1551341-E
21

Invacare®Kirilos/Atlante
5.UsethesteeringhandleBtopushthepatientliftinto
position.
WARNING!
–Whenusingtheliftinconjunctionwith
bedsorwheelchairs,beawareofthe
positionoftheliftinrelationshiptothose
otherdevicessothattheliftdoesnot
becomeentangled.
Beforepositioningthelegsofthepatientlift
underabed,makesurethattheareaisclearof
anyobstructions.
6.Lowerthepatientliftforeasyattachmentofthesling.
7.Locktherearcasters.RefertoLocking/Unlockingthe
RearCastersintheUsagesectionofthemanual.
8.Attachtheslingtothelift.RefertoAttachingthe
SlingtotheLiftintheLiftingthePatientsectionof
themanual.
22
1551341-E

LiftingthePatient
6.3AttachingtheSlingtotheLift
WARNING!
RiskofInjuryorDeath
Improperlyattached,improperlyadjusted,or
damagedslingscancausethepatienttofallor
causeinjurytoassistants.
–UseanInvacareapprovedslingthatis
recommendedbytheindividual’sdoctor,nurse
ormedicalassistantforthecomfortandsafety
oftheindividualbeinglifted.
–Invacareslingsandpatientliftaccessoriesare
specicallydesignedtobeusedinconjunction
withInvacarepatientlifts.
–Aftereachlaundering(inaccordancewith
instructionsonthesling),inspectsling(s)for
wear,tear,andloosestitching.
–Bleached,torn,cut,frayed,orbrokenslings
areunsafeandcouldresultininjury.Discard
immediately.
–DONOTalterslings.
–Besuretochecktheslingattachmentseach
timetheslingisremovedandreplaced,to
ensurethatitisproperlyattachedbeforethe
patientisremovedfromastationaryobject
(bed,chairorcommode).
–Positionthepatientintheslingasdirectedby
theinstructionsprovidedwiththesling.
–Adjustmentsforpatientsafetyandcomfort
shouldbemadebeforemovingthepatient.
WARNING!
RiskofInjuryorDeath
Improperlyattached,improperlyadjusted,or
damagedslingscancausethepatienttofallor
causeinjurytoassistants.
–DONOTuseanykindofplasticback
incontinencepadorseatingcushionbetween
patientandslingmaterialthatmaycausethe
patienttoslideoutoftheslingduringtransfer.
–Whenconnectingslingsequippedwithcolor
codedstrapstothepatientlift,useshorter
strapsatthebackofpatientforupright
support.Usinglongerstrapswillleaveless
supportforthepatient'sbackandthepatient
willbeinamorereclinedposition.Theloops
oftheslingarecolorcodedandcanbeused
toplacepatientinvariouspositions.Thecolors
makeiteasytoconnectbothsidesofthesling
equally.Makesurethatthereissufcienthead
supportwhenliftingapatient.
–ThehangerbarMUSTbeattachedtothelift
BEFOREattachingthesling.
1551341-E23

Invacare®Kirilos/Atlante
1.PlacethestrapsAoftheslingBoverhooksCofthe
hangerbarD.Theslingsmaybeequippedwithcolor
codedstrapstoassistwithproperattachment.Match
thecorrespondingstrapsoneachsideoftheslingfor
anevenliftofthepatient.
2.Liftandtransferthepatient.RefertoLiftingand
TransferringthePatientintheLiftingthePatientsection
ofthemanual.
24
6.4LiftingandTransferringthePatientFrom
aBed
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
–Refertothesafetyinformationandinstructions
inthefollowingproceduresBEFOREperforming
thisprocedure:
SafeLiftingintheLiftingthePatientsection
ofthemanual
PreparingtoLiftintheLiftingthePatient
sectionofthemanual
AttachingtheSlingtotheLiftintheLiftingthe
Patientsectionofthemanual
Raising/LoweringtheLiftintheUsagesection
ofthemanual
Invacarerecommendsatleasttwo(2)individuals
assistintransferringthepatientintoandoutofa
bedusingthisproduct.
1.Movethelifttothepatientarea,openlegsandprepare
tolift.RefertoPreparingtoLiftintheLiftingthePatient
sectionofthemanual.
2.Attachtheslingtothelift.RefertoAttachingthe
SlingstotheLiftintheLiftingthePatientsectionof
themanual.
3.Unlocktherearcasters.
1551341-E

LiftingthePatient
4.Lowerthebedtothelowestposition.
5.Liftthepatienthighenoughtoclearthestationary
objectwiththeirweightfullysupportedbythelift.
RefertoRaising/LoweringtheLiftintheUsagesection
ofthemanual.
TheboomwillstayinpositionuntiltheDOWN(
6.Beforemovingthepatient,checkagaintomakesure
thattheslingisproperlyconnectedtothehooksofthe
hangerbar.RefertoAttachingtheSlingtotheLiftin
theLiftingthePatientsectionofthemanual.Ifany
attachmentsarenotproperlyinplace,lowerthepatient
backontothestationaryobjectandcorrecttheproblem.
7.Usingthesteeringhandle,movetheliftawayfromthe
stationaryobject.
)buttonispressed.
1551341-E25

Invacare®Kirilos/Atlante
8.Usingthehandlesonthesling,turnthepatientsothat
he/shefacestheassistantoperatingthepatientlift.
9.Lowerthepatientsothathis/herfeetrestonthebase
ofthelift,straddlingthemast.
Thelowercenterofgravityprovidesstability
makingthepatientfeelmoresecureandthelift
easiertomove.
10.Movethepatientliftwithbothhandsrmlyonthe
steeringhandle.
11.ReadandunderstandtheinformationintheLiftingthe
Patientsectionofthemanualthatpertainstotransfer
toorfromspecictypesofsurfacesBEFOREproceeding:
•BedTransfer
•FloorTransfers(LiftingfromtheFloor)
•CommodeTransferGuidelines
•WheelchairTransfer
261551341-E

LiftingthePatient
12.Raiseorlowerthelifttopositionthepatientoverthe
stationarysurface.Besuretoraiseorlowerthepatient
enoughtoclearthesidesofthestationaryobject.
13.Lowerthepatientontothestationarysurface.
14.Ensurethepatientisfullysupportedbythesurfaceto
whichyouaretransferring.
15.Locktherearcasters.
16.Raisethebedtoagoodworkingheight(usuallyhip
heightofcaregivers).
17.Detachtheslingfromthehangerbar.
18.Unlocktherearcasters.
19.Movetheliftawayfromthearea.
1551341-E
27

Invacare®Kirilos/Atlante
6.4.1FloorTransfers(LiftingfromtheFloor)
WARNING!
RiskofInjury
Theliftcaninjureapatientwhoisontheoor
oranassistantworkingontheoor .
–Alwaysusetwoassistantswhentransferringa
patientfromtheoor.
1.PerformthesestepsinadditiontothoseinLiftingand
TransferringthePatientFromaBedwhentransferring
fromtheoor .
2.Determineifthepatienthassufferedanyinjuriesfrom
afall.Ifnomedicalattentionisneeded,proceedwith
thetransfer.
3.PositiontheslingAunderthepatient.Refertothesling
usermanualformoreinformationaboutpositioning
slings.
4.Oneassistantshouldhavethepatientbendhis/her
kneesandraisehis/herheadoffoftheoor.This
assistantshouldsupportthepatient’sheadwithapillow
B.
5.Theotherassistantshouldopenthelegsofthelift.
RefertoOpening/ClosingLegsoftheLiftintheUsage
sectionofthemanual.
6.Unlocktherearcasters.
281551341-E

LiftingthePatient
7.Positiontheliftwithonelegunderthepatient’shead
andtheotherlegunderthepatient’sbentknees.Keep
theslingstrapsinsideofthelegsofthelift.
8.Lowertheboomsothehangerbarisdirectlyoverthe
patient’schest.
9.Attachthesling.RefertoAttachingtheSlingtotheLift
intheLiftingthePatientsectionofthemanual.
10.Proceedwiththetransfer .RefertoLiftingand
TransferringthePatientFromaBedintheLiftingthe
Patientsectionofthemanual.
1551341-E29

Invacare®Kirilos/Atlante
6.4.2CommodeTransferGuidelines
TheInvacarepatientliftisNOTintendedasa
transportdevice.Ifthebathroomfacilitiesare
NOTnearthebedorifthepatientliftcannotbe
easilymaneuveredtowardsthecommode,thenthe
patientMUSTbetransferredtoawheelchairand
transportedtothebathroomfacilitiesbeforeusing
thepatientliftagaintopositionthepatientona
standardcommode.
Theslingswithcommodeopeningsaredesignedto
beusedwitheitheracommodechairorstandard
commode.
Invacarerecommendsatleasttwo(2)individuals
assistintransferringthepatientontoandoffofa
commodeusingthisproduct.
1.Performthesestepswhentransferringtoacommodein
additiontothoseinLiftingandT ransferringthePatient
FromaBed.
2.Beforetransferringthepatient,thepatientliftshould
beguidedtothebathroomfacilitiestocheckthatitcan
beeasilymaneuveredtowardsthecommode.
3.Attachtheslingstothelift.RefertoAttachingthe
SlingstotheLiftintheLiftingthePatientsectionin
themanual.
4.Unlocktherearcasters.
5.Elevatethepatienthighenoughtoclearthecommode
chairarmsandhavetheirweightsupportedbythe
patientlift.RefertoRaising/LoweringtheLiftinthe
Usagesectioninthemanual.
301551341-E