Invacare BAR600IVC, IVC BAR600IVC User Manual

User Manual
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
EN
Invacare®IVC™Model BAR600IVC Bariatric Bed
For more information regarding bed rail safety, refer to the Bed Rail Entrapment Risk Notification Guide at www.invacare.com.
Invacare®IVC™Model BAR600IVC Bariatric Bed 2 Part No 1123842
© 2011 Invacare Corporation. All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written
permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries
unless otherwise noted.
CONTENTS
Part No 1123842
3 Invacare
®
IVC™Model BAR600IVC Bariatric Bed
1 GENERAL 5
Symbols .................................................................................................................................................................................................................................................... 5
Limited Warranty .................................................................................................................................................................................................................................. 6
2OVERVIEW 7
Label locations ........................................................................................................................................................................................................................................ 7
Typical Product Parameters ................................................................................................................................................................................................................ 9
Bed Components: .................................................................................................................................................................................................................................. 9
3SAFETY 10
General Guidelines ..............................................................................................................................................................................................................................10
4 OPERATION AND ADJUSTMENT 18
Operating Bariatric Bed...................................................................................................................................................................................................................... 18
Emergency Crank................................................................................................................................................................................................................................. 19
Optional Bed Rails ...............................................................................................................................................................................................................................20
5 MAINTENANCE 21
Washing Bed Ends ...............................................................................................................................................................................................................................21
Cleaning Head and Foot Sections....................................................................................................................................................................................................22
Maintenance and Cleaning Checklist............................................................................................................................................................................................... 22
6 TROUBLESHOOTING 24
Troubleshooting: Electrical................................................................................................................................................................................................................24
Troubleshooting: Mechanical ............................................................................................................................................................................................................25
7 ASSEMBLY, MAINTENANCE AND ADJUSTMENT SECTIONS - FOR DEALERS ONLY 26
Overview of Procedures.................................................................................................................................................................................................................... 26
8 PACKAGING, HANDLING AND SET-UP 27
CONTENTS
Invacare
®
IVC™Model BAR600IVC Bariatric Bed 4 Part No 1123842
Receiving ................................................................................................................................................................................................................................................27
Unpacking ..............................................................................................................................................................................................................................................27
Inspecting ...............................................................................................................................................................................................................................................29
Set-Up.....................................................................................................................................................................................................................................................30
Storing the Beds................................................................................................................................................................................................................................... 45
9 MAINTENANCE 46
Washing Bed Ends ...............................................................................................................................................................................................................................46
Cleaning Head and Foot Sections....................................................................................................................................................................................................47
Maintenance Checklist........................................................................................................................................................................................................................47
10 TROUBLESHOOTING 50
Troubleshooting: Electrical................................................................................................................................................................................................................50
Troubleshooting: Mechanical ............................................................................................................................................................................................................51
1 GENERAL
Part No 1123842 5 Invacare®IVC™Model BAR600IVC Bariatric Bed
1 General
1.1 Symbols
Warnings
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. See the information below for definitions of the signal words.
DANGER
Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury or both.
!
IMPORTANT
Indicates a hazardous situation that could result in damage to property if it is not avoided.
Gives useful tips, recommendations and information for efficient, trouble-free use.
1 GENERAL
Invacare®IVC™Model BAR600IVC Bariatric Bed 6 Part No 1123842
1.2 Limited Warranty
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare or a dealer. This warranty is not extended to any other person or entity and is not transferable or assignable to any subsequent purchaser or owner. Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person. This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state. Invacare warrants the mechanical and electrical components of this product when purchased new and unused to be free from defects in materials and workmanship for a period of one year from date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty. Invacare warrants all welds of this product when purchased new and unused to be free from defects in materials and workmanship for a period of five years from date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty. If within such warranty periods any such product shall be proven to be defective, such product shall be repaired or replaced, at Invacare's option. This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product. Invacare's sole obligation and your exclusive remedy under this warranty shall be limited to such repair and/or replacement. For warranty service, please contact the dealer from whom you purchased your Invacare product. In the event you do not receive satisfactory warranty service, please write directly to Invacare at the address on the back cover, provide dealer's name, address, and the date of purchase, indicate nature of the defect and, if the product is serialized, indicate the serial number. Do not return products to our factory without our prior consent. LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED, PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, PRODUCTS MODIFIED WITHOUT INVACARE'S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS; PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE, OR TO A PRODUCT DAMAGED BY CIRCUMSTANCES BEYOND INVACARE'S CONTROL, AND SUCH EVALUATION WILL BE SOLELY DETERMINED BY INVACARE. THE WARRANTY SHALL NOT APPLY TO NORMAL WEAR AND TEAR OR FAILURE TO ADHERE TO THE PRODUCT INSTRUCTIONS. THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT BE APPLICABLE. THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.
2 OVERVIEW
Part No 1123842 7 Invacare®IVC™Model BAR600IVC Bariatric Bed
2 Overview
2.1 Label locations
! WARNING !
Grounding reliability can ONLY be achieved when the equipment is connected to a receptacle marked "Hospital Only" or "Hospital Grade".
P/N 034059
Rev D (1) 02/00
WARNING
For proper grounding, junction box MUST be securely connected to the bed with metal screws and spacer provided or electrical shock could occur.
1108103 Rev.A(1) - 08/01
All cable connections MUST be aligned properly. Align locking levers with locking tabs for proper connection. Mis-alignment of cable connections will cause permanent product damage.
Locking Lever
Crossover
Cable
Head Motor
Cable
Locking Tab
P/N 1171770 Rev A
CAUTION
Serial number
label located here.
Power Cord
2 OVERVIEW
Invacare®IVC™Model BAR600IVC Bariatric Bed 8 Part No 1123842
NOTE:Labelslocatedonbedframe.
2 OVERVIEW
Part No 1123842 9 Invacare®IVC™Model BAR600IVC Bariatric Bed
2.2 Typical Product Parameters
2.3 Bed Components:
Head spring with one motor (BAR5000IVC)
Foot spring with two motors (BAR5490IVC)
Bed Ends (BAR5301IVC)
IVC BARIATRIC BED
FOOT END
MINIMUM: MAXIMUM:
26.25 inches (66.7 cm)
33.75 inches (85.7 cm)
DECK HEIGHT*
MINIMUM (DOWN POSITION): MAXIMUM (UP POSITION):
14 inches (35 cm) ± 2 inches 22 inches (55 cm) ± 2 inches
BED
OVERALL LENGTH: WIDTH:
88 inches (223.5 cm) 42 inches (106.7 cm)
SPLIT SPRINGS TO ENDS
HEAD: FOOT:
43 inches (109.2 cm) 45 inches (114.3 cm)
SHIPPING WEIGHT 275 lbs
WEIGHT LIMITATION 700 lbs (Including accessories, matress, occupant
and any other objects positioned on the bed)
*NOTE:Heightfromfloortobeddeck.
3 SAFETY
Invacare®IVC™Model BAR600IVC Bariatric Bed 10 Part No 1123842
3 Safety
3.1 General Guidelines
WARNING
Do not use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
The initial set up of this bed must be performed by a qualified technician.
Procedures other than those described in this manual must be performed by a qualified technician.
For Dealers Only - Set-up and Assembly Instructions are in the rear of this manual. These procedures must be performed by a qualified technicians only.
ACCESSORIES WARNINGS
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
!
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE. Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact Invacare/Carrier for
further instruction.
3 SAFETY
Part No 1123842 11 Invacare®IVC™Model BAR600IVC Bariatric Bed
IMPORTANT SAFETY INSTRUCTIONS
When using the bed, basic precautions should always be followed, including the following:
Read all the instructions before using the bed.
DANGER - To reduce the risk of electric shock:
1. Always unplug the bed from the electrical outlet before cleaning.
WARNING - To reduce the risk of burns, fire, electric shock, or injury to persons:
1. Unplug from the outlet before servicing.
2. Close supervision is necessary when the bed is used by, or near children and/or disabled persons.
3. Use this bed only for its intended use as described in these instructions. DO NOT use attachments not recommended by the manufacturer.
4. NEVER operate this bed if it has a damaged cord or plug, if it is not working properly, if it has been dropped, damaged, or dropped into water. Return the bed to a service center for examination and repair.
5. Keep the cord away from heated surfaces.
6. DO NOT use outdoors.
7. Use masked or nasal type oxygen administering equipment only in conjunction with the manual/electric bed. The use of ANY other type of oxygen administering equipment can result in a fire hazard.
8. To disconnect, DO NOT depress pendant buttons.
9. Risk of electric shock - Connect this bed to a properly grounded outlet only. Refer to Grounding Instructions
on
page 16.
10. Risk of injury to persons - DO NOT place video equipment such as televisions or computer monitors on bed.
SAVE THESE INSTRUCTIONS
3 SAFETY
Invacare®IVC™Model BAR600IVC Bariatric Bed 12 Part No 1123842
Operating Information
WARNING
Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact a qualified technician for further instruction.
After any adjustments, repair or service and before use, make sure all attaching hardware is tightened securely.
Close supervision is necessary when the manual/electric bed is used by or near children and physically challenged individuals.
DO NOT let any individual underneath the bed or in between the raised bed frame components at anytime.
When bed is not to be used for an extended period, unplug electric bed from the wall outlet.
Keep the product a minimum of 12 inches away from any direct heat source.
DO NOT unplug power cord from junction box.
If a liquid is spilled in or around the electric bed, unplug the electric bed before cleaning. Clean up the spill and allow the electric bed or the area around the electric bed to dry thoroughly before using the electric controls again.
The manual/electric bed is not designed to be used as a patient transport. When transporting a patient, use an approved patient transport. Otherwise, injury or damage may result.
When operating/moving electric beds, ALWAYS ensure that the individual utilizing the bed is positioned properly within the confines of the bed. DO NOT let any extremities protrude over the side or between the bed rails when performing these functions.
After the electric bed has been assembled, ALWAYS test to make sure that all sections of the electric bed are properly and securely in place before using.
Make sure head and foot springs/sections are connected securely to the bed ends before use.
The pendant and power cords MUST be routed and secured properly to ensure that the cords DO NOT become entangled, pinched and/or severed during operation of the electric bed.
3 SAFETY
Part No 1123842 13 Invacare®IVC™Model BAR600IVC Bariatric Bed
Entrapment Warning
WARNING
NEVER operate the unit if the cords are damaged.
If the unit is not working properly, call a qualified technician to examine the unit and repair it.
Keep all moving parts, including the main frame, mattress deck (head and foot springs/sections) and all drive shafts, free of obstruction (i.e. blankets/sheets, heating blankets/pads, tubing, wiring, etc. and other types of products using electric cords which may get tangled around the bed, side rails or legs) during operation of the electric bed.
The electric bed is equipped with locking casters. When transferring into or out of the electric bed, ALWAYS lock the locking caster(s). Inspect the wheel locks for correct locking action before actual use. Even with casters properly locked, some flooring surfaces such as tile or wood, will allow the bed to move under some conditions. Use on surfaces such as these must be evaluated by the care provider.
WARNING
Refer to the Bed Rail Entrapment Risk Notification Guide at www.invacare.com for additional safety information. Mattress MUST fit bed frame and bed rails snugly to reduce the risk of entrapment. After any adjustments, repair or service and before use, make sure all attaching hardware is tightened securely.
3 SAFETY
Invacare®IVC™Model BAR600IVC Bariatric Bed 14 Part No 1123842
Replacement Parts/Accessories
WARNING
Use masked or nasal type oxygen administering equipment only in conjunction with the manual/electric bed. The use of any other type of oxygen administering equipment can result in a fire hazard.
When using nasal or masked type oxygen administering equipment, the oxygen or air tubing MUST be routed and secured properly to ensure that the tubing does not become entangled and/or severed during normal operation of manual/electric bed.
ALWAYS test to make sure that the side rails are properly and securely in place before using the manual/electric bed.
Invacare recommends that the mattress be centered on the bed frame. Otherwise, individuals may become entangled between the bed rail and the bed frame.
Physically challenged individuals who cannot prevent themselves from rolling/climbing out of the manual/electric bed may require alternative safe means of restraint.
When used with a manual/electric bed, the bed rails DO NOT fall under any weight limitations. Bed rails can be deformed or broken if excessive side pressure is exerted on the bed rails. These bed rails are used for the purpose of preventing an individual from inadvertently rolling out of bed. The bed rails are not intended nor may be used for restraint purposes. If an individual is capable of injuring himself/herself, a physician or a healthcare professional should be consulted for alternative means of safe restraint.
After raising/lowering the head/foot end of the electric bed, check the distance between the bottom of the bed rail and the mattress. If there is excessive distance between the bottom of the bed rail and the mattress in which individuals may become entangled, adjust the height of the bed rail (if applicable), or provide alternative means of patient protection.
DO NOT use the side rails as push handles for moving the electric bed.
Replacement mattresses and bed side rails with dimensions different from the original equipment supplied or specified by the bed frame manufacturer are not interchangeable. Variations in bed side rail design, width and thickness or firmness of the mattress could cause/contribute to entrapment.
3 SAFETY
Part No 1123842 15 Invacare®IVC™Model BAR600IVC Bariatric Bed
Electrical
Repair or Service Information
DANGER
When using an extension cord, use only a three wire extension cord having at least 16 AWG (American Wire Gauge) wire and the same or higher electrical rating as the device being connected. Use of improper extension cord could result in a risk of fire and electric shock. Three prong to two prong adapters should not be used. Use of three prong adapters can result in improper grounding and present a shock hazard to the user.
WARNING
NEVER operate if the unit has a damaged cord or plug. If it is not working properly, call a qualified technician for examination and repair.
Keep all electrical cords away from heated or hot surfaces.
Ensure all cables and cords are routed such that they will not become entangled or pinched. Otherwise damage or injury may result.
WARNING
DO NOT open assemblies such as the motors, pendant, junction boxes or gear boxes. No user serviceable parts are inside. Only qualified technicians are permitted to repair these parts. If unqualified individuals perform any work on these beds, the warranty is void.
Unplug the power cord from its power source before performing any maintenance on the electric bed. DO NOT unplug the power cord from the junction box. Damage to cord will result.
CAUTION
All cable connections MUST be aligned properly. Align locking levers with locking tabs for proper connection. Mis-alignment of cable connections will cause permanent product damage.
3 SAFETY
Invacare®IVC™Model BAR600IVC Bariatric Bed 16 Part No 1123842
Grounding Instructions
Radio Frequency Interference
Weight Limitations
WARNING
Electric beds are equipped with three-prong (grounding) plugs for protection against possible shock hazards. Where a two-prong wall receptacle is encountered, it is the personal responsibility and obligation of the customer to contact a qualified electrician and have the two-prong receptacle replaced with a properly grounded three-prong wall receptacle in accordance with the National Electrical Code. If you must use an extension cord, use only a three-wire extension cord having the same or higher electrical rating as the device being connected. DO NOT, under any circumstances, cut or remove the round grounding prong from any plug used with or for Invacare products. In addition, Invacare has placed RED warning tags on some equipment. DO NOT remove these tags.
For proper grounding, junction box MUST be securely connected to the bed with metal screws and the spacer provided or electrical shock could occur.
WARNING
Electronic equipment may be influenced by Radio Frequency Interference (RFI). Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment. If RFI causes erratic behavior, unplug the electric bed immediately. Leave unplugged while transmission is in progress.
WARNING
The total weight limit of the Bariatric Bed (including accessories, mattress, occupant and any other objects positioned on the bed) is 700 pounds (318 kg); 600 pounds (272 kg) patient weight.
NEVER permit more than one person on the manual/electric bed at any time.
Body weight should be evenly distributed over the surface of the manual/electric bed. DO NOT lay, sit or lean in such a way that your entire body weight is placed only on raised head or foot sections of the bed. This includes when repositioning or transferring in or out of bed.