Invacare®ISA™
ISA™COMPACT,ISA™STANDARD,ISA™PLUS,ISA™XPLUS
enStandassistpatientlift
UserManual...................................3
deAufstehlifter
Gebrauchsanweisung...........................33
esGrúadebipedestaciónparapacientes
Manualdelusuario.............................65
frLève-personneverticalisateur
Manueld'utilisation............................97
itSollevatoreperpazientiStandAssist
Manualed’uso.................................129
nlActievepatiëntentillift
Gebruiksaanwijzing.............................161
ptElevadordetransferênciaposiçãovertical
Manualdeutilização...........................193
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsaved
forfuturereference.

©2019InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibitedwithoutpriorwritten
permissionfromInvacare.Trademarksareidentiedby™and®.AlltrademarksareownedbyorlicensedtoInvacare
Corporationoritssubsidiariesunlessotherwisenoted.
–
AlleRechtevorbehalten.DieWeiterveröffentlichung,VervielfältigungoderÄnderungimGanzenoderinT eilenohne
vorherigeschriftlicheGenehmigungvonInvacareistuntersagt.Markensinddurch™und®gekennzeichnet.Soweit
nichtandersangegeben,sindalleMarkenEigentumderInvacareCorporationbzw.derenTochtergesellschaften
oderwerdenvondieseninLizenzgenutzt.
–
Todoslosderechosreservados.Seprohíbevolverapublicar ,copiaromodicarelpresentedocumento,enparteopor
completo,sinelprevioconsentimientoporescritodeInvacare.Lasmarcascomercialesseidenticancon™y®.T odas
lasmarcascomercialessonpropiedaddeInvacareCorporationodesusliales,obienInvacareCorporationosus
lialescuentanconlascorrespondienteslicencias,salvoqueseindiquelocontrario.
–
Tousdroitsréservés.Larepublication,laduplicationoulamodicationdetoutoupartieduprésentdocumentest
interditesansl'accordécritpréalabled'Invacare.Lesmarquescommercialessontidentiéespar™et®.Toutes
lesmarquescommercialessontdétenuesparoucédéessouslicenceàInvacareCorporationousesliales,sauf
stipulationcontraire.
–
Tuttiidirittiriservati.Èproibitalariproduzione,laduplicazioneolamodicaparzialeocompleta,salvoprevia
autorizzazionescrittadapartediInvacare.Imarchisonocontrassegnatida™e®.Tuttiimarchisonodiproprietào
licenzadiInvacareCorporationodisueafliate,salvoindicazionecontraria.
–
Allerechtenvoorbehouden.Herpublicatie,duplicatieofgeheleofgedeeltelijkewijzigingisverbodenzonder
voorafgaandeschriftelijketoestemmingvanInvacare.Handelsmerkenzijnteherkennenaan™en®.Allehandelsmerken
zijneigendomvanofgelicentieerdaanInvacareCorporationofhaardochterondernemingen,tenzijandersaangegeven.
–
Todososdireitosreservados.Arepublicação,duplicaçãooumodicaçãototalouparcialestáinterditasema
autorizaçãopréviaporescritodaInvacare.Asmarcascomerciaissãoidenticadaspelossímbolos™e®.Todasas
marcascomerciaissãopropriedadedaouestãolicenciadasàInvacareCorporationouàssuassubsidiárias,exceto
quandoapresentadainformaçãoemcontrário.

Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsaved
forfuturereference.
1General.........................................4
1.1Introduction..................................4
1.1.1Symbolsinthisdocument.....................4
1.2Servicelife...................................4
1.2.1Additionalinformation........................4
1.3Limitationofliability............................4
1.4Warrantyinformation...........................4
1.5Compliance...................................4
1.5.1Product-specicstandards.....................4
2Safety..........................................5
2.1Generalsafetyinformation.......................5
2.1.1Pinchpoints................................6
2.2Safetyinformationonaccessories..................6
2.3Safetyinformationonelectromagneticinterference.....6
2.3.1Batterychargingmode........................6
2.4Labelsandsymbolsontheproduct.................7
2.4.1Labellocation..............................7
2.4.2Identicationlabel...........................7
2.4.3Othersymbols..............................7
3ProductOverview.................................8
3.1Intendeduse..................................8
3.2Mainpartsofthelift............................8
3.3Accessories...................................9
4Setup...........................................10
4.1Generalsafetyinformation.......................10
4.2Scopeofdelivery..............................10
4.3Installingthemasttothebase....................10
4.4Installingthefootplate..........................11
4.5Installingthelegband...........................11
4.6Installingthelegsupport.........................11
4.7Installingtheleverformanuallegspreader...........11
4.8Resettingtheserviceindicator.....................12
5Usage..........................................13
5.1Generalsafetyinformation.......................13
5.2Locking/Unlockingtherearcastors.................13
5.3Raising/LoweringanElectricLift...................13
5.4Closing/OpeningLegs...........................13
5.4.1Closing/OpeningElectricLegs...................13
5.4.2Closing/Openinglegsmanually..................13
5.5Adjustingtheliftingarmtopatientbodyheight........13
5.6Adjustingtheheightofthelegsupport..............14
5.7Usingthelegband.............................14
5.8Emergencyfunctions............................15
5.8.1Performinganemergencystop..................15
5.8.2Activatinganemergencylowering(CBJHome
controlunit)...............................15
5.8.3Activatinganemergencylowering(CBJCare,CBJ1,
CBJ2controlunit)...........................15
5.8.4Activatinganemergencylifting(CBJCare,CBJ1
controlunit)...............................15
5.8.5Activatingamechanicalemergencylowering.......15
5.9Chargingthebattery............................16
5.9.1CBJHomecontrolunit........................16
5.9.2CBJCare,CBJ1,CBJ2controlunit................16
5.9.3Optionalbatterycharger......................17
6PatientTransfer...................................18
6.1Generalsafetyinformation.......................18
6.2Attachingtheslingtothelift......................19
6.3Transferringapatient...........................20
7TransportationandStorage..........................21
7.1Generalinformation............................21
7.2Disassemblingthelift...........................21
8Maintenance.....................................22
8.1Generalmaintenanceinformation..................22
8.2Dailyinspections...............................22
8.3CleaningandDisinfection........................22
8.3.1Generalsafetyinformation.....................22
8.3.2CleaningIntervals............................22
8.3.3Cleaninginstructions.........................22
8.3.4Disinfectioninstructions.......................23
8.4Serviceinterval................................23
8.4.1LOLERStatement............................23
9AfterUse........................................24
9.1Disposal.....................................24
9.2Reconditioning................................24
10Troubleshooting..................................25
10.1Identifyingfaultsandpossiblesolutions.............25
11T echnicalData...................................27
11.1Maximumsafeworkingload.....................27
11.2Dimensionsandweights........................27
11.3Electricalsystem..............................28
11.4Environmentalconditions........................28
11.5Materials....................................29
11.6Operatingforcesofcontrols......................29
12Electromagneticcompatibility(EMC)..................30
12.1GeneralEMCinformation.......................30
12.2Electromagneticemission........................30
12.3ElectromagneticImmunity.......................30

Invacare®ISA™
1General
1.1Introduction
Thisusermanualcontainsimportantinformationaboutthe
handlingoftheproduct.T oensuresafetywhenusingthe
product,readtheusermanualcarefullyandfollowthe
safetyinstructions.
Notethattheremaybesectionsinthisdocument,whichare
notrelevanttoyourproduct,sincethisdocumentapplies
toallavailablemodels(onthedateofprinting).Ifnot
otherwisestated,eachsectioninthisdocumentreferstoall
modelsoftheproduct.
Themodelsandcongurationsavailableinyourcountrycan
befoundinthecountry-specicpricelists.
Invacarereservestherighttoalterproductspecications
withoutfurthernotice.
Beforereadingthisdocument,makesureyouhavethe
latestversion.YoundthelatestversionasaPDFonthe
Invacarewebsite.
Ifyoundthatthefontsizeintheprinteddocumentis
difculttoread,youcandownloadthePDFversionfromthe
website.ThePDFcanthenbescaledonscreentoafont
sizethatismorecomfortableforyou.
Formoreinformationabouttheproduct,forexampleproduct
safetynoticesandproductrecalls,contactyourInvacare
representative.Seeaddressesattheendofthisdocument.
Incaseofaseriousincidentwiththeproduct,youshould
informthemanufacturerandthecompetentauthorityin
yourcountry.
maintenanceintervalsandcorrectuse,statedinthismanual.
Theeffectiveservicelifecanvaryaccordingtofrequencyand
intensityofuse.
1.2.1Additionalinformation
Theexpectedservicelifeisbasedonanestimatedaverage
of4liftingcyclesperday.
1.3Limitationofliability
Invacareacceptsnoliabilityfordamagearisingfrom:
•Non-compliancewiththeusermanual
•Incorrectuse
•Naturalwearandtear
•Incorrectassemblyorset-upbythepurchaserorathird
party
•Technicalmodications
•Unauthorisedmodicationsand/oruseofunsuitable
spareparts
1.4Warrantyinformation
Weprovideamanufacturer’swarrantyfortheproduct
inaccordancewithourGeneralTermsandConditionsof
Businessintherespectivecountries.
Warrantyclaimscanonlybemadethroughtheprovider
fromwhomtheproductwasobtained.
1.5Compliance
Qualityisfundamentaltothecompany’soperation,working
withinthedisciplinesofISO13485.
1.1.1Symbolsinthisdocument
Symbolsandsignalwordsareusedinthisdocumentand
applytohazardsorunsafepracticeswhichcouldresultin
personalinjuryorpropertydamage.Seetheinformation
belowfordenitionsofthesignalwords.
WARNING
Indicatesahazardoussituationthatcouldresult
inseriousinjuryordeathifitisnotavoided.
CAUTION
Indicatesahazardoussituationthatcouldresult
inminororslightinjuryifitisnotavoided.
IMPORTANT
Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
TipsandRecommendations
Givesusefultips,recommendationsand
informationforefcient,trouble-freeuse.
1.2Servicelife
Theexpectedservicelifeofthisproductiseightyearswhen
useddailyandinaccordancewiththesafetyinstructions,
ThisproductfeaturestheCEmark,incompliancewiththe
MedicalDeviceRegulation2017/745Class1.Thelaunchdate
ofthisproductisstatedintheCEdeclarationofconformity.
Wearecontinuouslyworkingtowardsensuringthatthe
company’simpactontheenvironment,locallyandglobally ,
isreducedtoaminimum.
WeonlyuseREACHcompliantmaterialsandcomponents.
WecomplywiththecurrentenvironmentallegislationsWEEE
andRoHS.
1.5.1Product-specicstandards
TheproducthasbeentestedandconformstoISO10535
(Hoistsforthetransferofdisabledpersons)andallrelated
standards.
Forfurtherinformationaboutlocalstandardsandregulations,
contactyourlocalInvacarerepresentative.Seeaddresses
attheendofthisdocument.
4
1654371-A

Safety
2Safety
2.1Generalsafetyinformation
Thissectionofthemanualcontainsgeneralsafety
informationaboutyourproduct.Forspecicsafety
information,refertotheappropriatesectionofthemanual
andprocedureswithinthatsection.
WARNING!
Riskofinjuryordamage
–Donotusethisproductoranyavailableoptional
equipmentwithoutrstcompletelyreading
andunderstandingtheseinstructionsandany
additionalinstructionalmaterialsuchasuser
manualsorinstructionsheetssuppliedwith
thisproductoroptionalequipment.Ifyouare
unabletounderstandthewarnings,cautionsor
instructions,contactahealthcareprofessional,
Invacareproviderorqualiedtechnicianbefore
attemptingtousethisproduct.
–Donotmakeanyunauthorizedalterationsor
modicationstotheproduct.
WARNING!
Themaximumsafeworkingloadmustnotbe
exceeded
–Donotexceedthemaximumsafeworkingload
ofthisproductorusedaccessorieslikeslings
etc.Seedocumentationorthelabelingforthe
statedmaximumsafeworkingload.
–Thecomponentwiththelowestloadlimit
determinesthemaximumsafeworkingloadof
theentiresystem.
WARNING!
Riskofinjuryordamage
Improperuseofthisproductmaycauseinjury
ordamage.
–Donotattemptanytransferwithoutapproval
ofthepatient’shealthcareprofessional.
–Readtheinstructionsinthisusermanualand
observetrainedpersonnelperformingtransfer
procedures.Thenpracticetransfersunder
supervisionandwithacapablepersonacting
asapatient.
–Patientswhoaretransferredandpositioned
usingthestandassistliftMUSTbecooperative,
coherentandhaveheadandneckcontrol.
Otherwise,injurymayoccur .
–Patientswhoaretransferredandpositioned
usingaslingwithbacksupportonly,MUSTbe
abletosupportamajorityoftheirownweight.
Otherwise,injuryordamagemayoccur .Refer
totheslingusermanualformoreinformation.
–Donotusetheliftasatransportdevice.It
isintendedtotransferanindividualfromone
seatingsurfacetoanother.
WARNING!
Riskofinjuryordamage
Inappropriatehandlingofcables,incorrect
connectionsanduseofunauthorizedequipment
cancauseelectricalshockandproductfailure.
–Donotkink,shearorotherwisedamagethe
cablesoftheproduct.
–Makesurethatnocablesarejammedor
damaged,whentheproductisused.
–Makesurethewiringiscorrectandconnections
areproper.
–Donotuseunauthorizedequipment.
WARNING!
Riskofinjuryordamage
Excessivemoisturewilldamagetheproductand
maycauseelectricalshock.
–Thepatientliftcanbeusedinabathorshower
areabutmustNOTbeusedundertheshower.
Thepatientmustbetransferredtoashower
chairoruseothermeansforshowering.
–Ifthepatientliftisusedinamoisture
environment,ensurethatthepatientliftis
wipedcleanofanymoistureafteruse.
–Donotplugorunplugthepowercableina
moistureenvironmentorwithwethands.
–Donotstoretheproductinadampareaor
dampcondition.
–Periodicallyinspectallcomponentsofthe
productforsignsofcorrosionordamage.
Replacepartsthatarecorrodedordamaged.
–See11.4Environmentalconditions,page28
WARNING!
Riskofinjuryordamage
Ignitionsourcescancauseburnsorre.
–Patienttransfermustbeperformedwitha
safetyspacebetweentheliftandpossible
ignitionsources(heater,stove,replace,etc.)
–Thepatientandassistantsmustnotsmoke
duringtransfer.
–Theslingmustnotbeplacedoverheatsources
(heater,stove,replace,etc.)
WARNING!
Riskofinjuryordamage
Toavoidinjuryordamagewhenoperatingthe
product:
–Closesupervisionisnecessarywhentheproduct
isusednearchildrenorpets.
–Donotletchildrenplaywiththeproduct.
CAUTION!
Riskofinjuryordamage
Theproductcangethotwhenexposedtosunlight
orotherheatsources.
–Donotexposetheproducttodirectsunlightfor
prolongedperiods.
–Keeptheproductawayfromheatsources.
.
IMPORTANT!
Accumulationoflint,dustandotherdirtcan
impairtheproduct.
–Keeptheproductclean.
1654371-A5

Invacare®ISA™
2.1.1Pinchpoints
WARNING!
RiskofInjury
Pinchpointsarepresentinseverallocationson
theliftandbodypartscouldbepinched.
–Alwayskeephands,ngersandfeetclearof
movingparts.
2.2Safetyinformationonaccessories
CAUTION!
Riskofinjury
Non-originalorincorrectaccessoriesmayaffect
thefunctionandsafetyofthisproduct.
–Duetoregionaldifferences,refertoyour
localInvacarewebsiteorcatalogforavailable
accessoriesorcontactyourInvacareprovider .
–Seethemanualdeliveredwiththeaccessory
forfurtherinformationandinstructions.
–Useonlyoriginalaccessoriesfortheproduct
inuse.Undercertaincircumstances,theuse
slingsfromothermanufacturersispossible.See
additionalinformationinthissection.
–Ariskassessmentisalwaystobecarriedoutby
aprofessionalpriortoissuingliftingequipment.
ItisimportantthattheT ask,Individual,Load,
EnvironmentandEquipmentareconsideredin
theriskassessment.
–Alwayschoosetheslingdesignandsize
accordingtothepatient’sweight,sizeand
physicalabilitywhilstconsideringthetypeof
transfertobecarriedout.
CAUTION!
Compatibilityofslingswithattachmentsystem
Invacareusesacommonattachmentsystem
basedonhooksandloops.Loopsontheslings
areattachedtohooksonthelift.Therefore,
suitableslingsofothermanufacturerscanbe
usedonthisliftaswell.
–Useonlyslingswhichareintendedforstand
assistlifts.
–Useonlyslingswithloopattachmentsthatare
suitableforhooksasattachmentpoints.
–Donotuseslingsdesignedfor"Keyholeor
ClipAttachmentSystems"or"TiltingFrame
Systems".
2.3Safetyinformationon
electromagneticinterference
WARNING!
Riskofmalfunctionduetoelectromagnetic
interference
Electromagneticinterferencebetweenthis
productandotherelectricalequipmentcanoccur
anddisturbtheelectricaladjustmentfunctions
ofthisproduct.Toprevent,reduceoreliminate
suchelectromagneticinterference:
–Onlyuseoriginalcables,accessoriesandspare
parts,tonotincreaseelectromagneticemission
orreduceelectromagneticimmunityofthis
product.
–Donotuseportableradiofrequency(RF)
communicationsequipmentcloserthan30cm
toanypartofthisproduct(includingcables).
–Donotusethisproductnearactive
high-frequencysurgicalequipmentandthe
RFshieldedroomofasystemformagnetic
resonanceimaging,wheretheintensityof
electromagneticdisturbancesishigh.
–Ifdisturbancesoccur,increasethedistance
betweenthisproductandtheotherequipment
orswitchitoff.
–Refertothedetailedinformationandfollow
theguidanceinchapter12Electromagnetic
compatibility(EMC),page30
WARNING!
Riskofmalfunction
Electromagneticinterferencemaycauseimproper
operation.
–Donotusethisproductadjacenttoorstacked
withotherelectricalequipment.Ifsuchuseis
necessarythisproductandtheotherequipment
mustbecloselyobservedtoverifythatthey
areoperatingnormally.
2.3.1Batterychargingmode
IMPORTANT!
Duringbatterychargingthecontrolunitis
moresusceptiblefordamagebyanElectrostatic
Discharge(ESD).TheoccurrenceofanESDduring
batterychargingmodemayleadtocontrolunit
malfunctionorlossoffunction.
–AvoidanypossiblesourcesofESDduring
batterycharging.
.
61654371-A

2.4Labelsandsymbolsontheproduct
Product Name
YYYY-MM
= xxx kg
Invacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio, Portugal
XXXXXX XXXXX_XX
XXXXXXXXXXXXX
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
ISO 10535
2.4.1Labellocation
Max.SafeWorkingLoad
CLASSIIequipment
TypeBAppliedPart
WEEEconform
Europeanconformity
Abbreviationsfortechnicaldata:
•Iin=IncomingCurrent
•Uin=IncomingVoltage
•Int.=Intermittence
•AC=AlternatingCurrent
•Max=maximum
•min=minute
Formoreinformationabouttechnicaldata,see11T echnical
Data,page27.
2.4.3Othersymbols
Safety
Identicationlabel
A
Readusermanuallabel
B
Nameandsafeworkingloadoflift–depending
C
onmodel
Environmentalconditionsandtotalmaximum
D
weightlabel
Safeworkingloadoflift–dependingonmodel
E
Warrantyvoidlabel–notpresentonallmodels
F
Warninglabel
G
Liftingarmadjustmentlabel
H
2.4.2Identicationlabel
Theidenticationlabelcontainsthemainproduct
information,includingtechnicaldata.
RefertoUserManual
Totalweightoftheproductwiththe
maximumsafeworkingloadapplied
Donotusetheactuatorasapushbar
Donotremove
Warrantyvoidifremovedorbroken
Temperaturelimit
Humiditylimitation
Atmosphericpressurelimitation
Transportandstorageconditions
Symbols
1654371-A
SerialNumber
ReferenceNumber
ManufacturerAddress
ManufacturingDate
Operationconditions
Liftingarmadjustmentincorrelationto
thebodyheightofthepatient
7

Invacare®ISA™
3ProductOverview
3.1Intendeduse
Thestandassistpatientliftisabattery-poweredtransfer
deviceandisintendedtotransferandpositionanindividual
fromoneseatingsurfacetoanother .Forexample:
•Betweenthebedandawheelchair
•Toandfromthetoilet
•Rehabilitation,assistingapatientfromaseatedposition
toastandingposition
Themaximumsafeworkingloadisstatedin11Technical
Data,page27
Thestandassistpatientliftisdesignedtobeusedindoorson
alevelsurface,inhospitals,nursingfacilitiesanddomestic
areas.Thestandassistpatientliftcanbeturned(rotated)in
placefortransferswithlimitedoorspace.
Ahealthcareprofessionalorprivatepersonwhohasreceived
propertrainingistheintendedoperatorofthisproduct.
Indications
Thestandassistliftisintendedtotransferandposition
partiallyimmobileweight-bearingpatients.
3.2Mainpartsofthelift
Contraindications
Thestandassistliftiscontraindicatedforpatientswho:
•arenotcooperative
•arenotabletounderstandandfollowinstructions
•donothaveheadandneckcontrol
•donothavereasonabletrunkstability
Someslingsdesignedforthestandassistliftare
contraindicatedforpatientswhocannotsupportthemajority
oftheirweight.Refertotheslingusermanualformore
information.
Boom
A
Hookforsling
B
Handgrip
C
Liftingarm
D
Pushbar
E
Liftingactuator
F
Mast
G
Legsupport(xedorswiveling)
H
Basewithfootpedalformanuallegspreader
I
Footplate(high,loworinclined)
J
Frontcastors
K
Leg
L
Basewithactuatorsforelectricallegspreader—
M
withorwithoutactuatorcovers
Rearcastorswithbrake
N
Handcontrol
O
Emergencystop
P
CBJHomecontrolunitwithintegratedbattery
Q
81654371-A

ProductOverview
CBJCare,CBJ1orCBJ2controlunitwithdetachable
R
battery
Mechanicalemergencylowering
S
Leverformanuallegspreader
T
3.3Accessories
Duetoregionaldifferences,refertoyourlocal
Invacarewebsiteorcatalogfordetailsaboutavailable
accessoriesorcontactyourInvacareprovider.
•Leverformanuallegspreader
•Protectivecoversforlegs
•Legband
•Wallchargerforadetachablebattery
•Extrabattery
•Extrafootplate(high,loworinclined)
Slingmodelswithloopattachments,suitableforhooksas
attachmentpoints:
•Invacare®StandAssistsling(Europe)/Invacare®Standup
sling(APAC):2–pointattachmentwithbacksupportonly.
•Invacare®TransferStandAssistsling:4–pointattachment
withbackandlegsupport.
1654371-A9

Invacare®ISA™
4Setup
4.1Generalsafetyinformation
WARNING!
–Checkallpartsforshippingdamagebeforeuse.
–Incaseofdamage,donotusetheequipment.
ContactyourInvacareproviderforfurther
instructions.
WARNING!
RiskofInjury
Improperassemblymaycauseinjuryordamage.
–UseonlyInvacarepartsintheassemblyofthis
patientlift.
–Aftereachassembly ,checkthatallttings
areproperlytightenedandallpartshavethe
correctfunction
–Donotovertightenthemountinghardware.
Thismaydamagethemountingbrackets.
Therearenotoolsrequiredtoassemblethepatient
lift.
Ifthereareanyissuesorquestionsduringassembly,
contactyourInvacareprovider.
UserManual(1piece)
Battery(1or2pieces)*–CBJCare,CBJ1and
CBJ2only
Leverformanuallegspreader(1piece)*
Legband(1piece)*
Sling(1piece)*
*Dependsonmodeland/orconguration
Iftheliftisdeliveredwithasling,refertotheuser
manualoftheslingforuse,application,maintenance
andwashinginstructions.
4.3Installingthemasttothebase
WARNING!
RiskofinjuryorDamage
Themaximumsafeworkingloadstatedonthe
boomandonthebasemustbethesame.
–Alwayscomparethemaximumsafeworking
loadvaluesstatedonthebaseandonthe
boom.
1.
4.2Scopeofdelivery
Theitemsincludedinyourpackagedependonthemodels
andcongurationsavailableinyourcountry.See1.1
Introduction,page4
LockbothrearcastorsBandrotatethehandscrews
anti-clockwiseAtoremovethemfromthebase
2.
A
B
C
D
E
101654371-A
Mast/boomassembly(1piece)—incl.
controlunitandactuator
Basewithorwithoutactuators(1piece)
Legsupport
Footplate
Handcontrol(1piece)
Mainscable(1piece)
Withtheboompointinginforwarddirection,lowerthe
mastCintothesupportofbaseDandreinstallthe
handscrewsAtoxthemastCtothebase.

Setup
4.4Installingthefootplate
1.LowerthefootplateontothebaseuntilthebracketsA
arefullyseatedontubeB.
4.5Installingthelegband
(optional)
Thetwopartsofthelegbandareattachedontheback
sideofthelegsupport.
1.
4.6Installingthelegsupport
1.PushandholdupthereleasehandleAontheleg
supportB.
2.SlidetheguidebaronthebackofthelegsupportB
ontoitsmountingplateConthemast.
3.Whenthedesiredheightisreached,releasethehandle
AandslightlymovethelegsupportBupordownuntil
anaudibleclickisheardandthelegsupportBislocked
intooneofthe6availableheightpositions.
ThreadthetwohooksAintolugsBonthebackside
ofthelegsupport.
2.
4.7Installingtheleverformanualleg
spreader
2x13mmwrench
1.InserttheboltAfrombelowintothelowerholeofthe
bracketBandxitwiththenutC.
2.InsertthethreadedendofleverDintotheupperhole
ofbracketBontothexedboltA.
3.TurntheleverDclockwisetoscrewitontothebolt.
ThreadtheloopsCofthetwolegbandpartsintothe
hooksA.
1654371-A
11

Invacare®ISA™
4.8Resettingtheserviceindicator
(CBJCarecontrolunitonly)
IMPORTANT!
Iftheservicelightisashingyellowduringdaily
usageorafterareassembly,theliftrequires
service.
–ContactyourInvacareproviderforservice.
Aftertheinitialassemblyofthelift,theserviceindicator
needstoberesetbeforeusingthelift.
Toperformaninitialresetoftheserviceindicator:
1.Locatethehandcontrol.
2.PressandholdtheUPbuttonandtheDOWNbuttonat
thesametimeforveseconds.
3.Youwillhearasoundwhentheserviceindicatorhas
beenreset.
12
1654371-A

Usage
5Usage
5.1Generalsafetyinformation
WARNING!
Riskofinjuryordamage
Beforeusingtheliftwithapatient,refertothe
followingsafetyinformationandinstructions:
–2Safety,page5
–6PatientTransfer,page18
5.2Locking/Unlockingtherearcastors
WARNING!
RiskofInjuryorDamage
Theliftcouldtipandendangerthepatientand
assistants.
–Invacarerecommendsthattherearcastors
beleftunlockedduringliftingproceduresto
allowthepatientlifttostabilizeitselfwhenthe
patientisinitiallyliftedfromachair,bedorany
stationaryobject.
–Invacarerecommendslockingtherearcastors
oftheliftonlywhenpositioningorremoving
theslingfromaroundthepatient.
5.4Closing/OpeningLegs
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
–Thelegsoftheliftmustbeinthemaximum
openpositionforoptimumstabilityandsafety.
Ifitisnecessarytoclosethelegsoftheliftto
maneuvertheliftunderabed,closethelegsof
theliftonlyaslongasittakestopositionthe
liftoverthepatientandliftthepatientoffthe
surfaceofthebed.Whenthelegsofthelift
arenolongerunderthebed,returnthelegsof
thelifttothemaximumopenposition.
5.4.1Closing/OpeningElectricLegs
Thehandcontrolisusedtoopenorclosethelegsofthe
base.
1.Toclosethelegs,press
andholdthelegsclosed
buttonA.
2.Toopenthelegs,pressand
holdthelegsopenbutton
B.
•Tolockthecastor,pushdownpedalAwiththefoot.
•Tounlockthecastor,pushuppedalAwiththefoot.
5.3Raising/LoweringanElectricLift
Thehandcontrolisusedtoraiseorlowerthelift.
1.Toraisethelift—Press
andholdtheUPAbutton
toraisetheboomandthe
patient.
2.Tolowerthelift—Press
andholdtheDOWNB
buttontolowertheboom
andthepatient.
Releasethebuttontostopraisingorloweringthelift.
Thelegswillstopmovingwhenthebuttonis
released.
5.4.2Closing/Openinglegsmanually
Themanuallegspreaderisoperatedbytwopedals(Aand
B)orbytheleverC.
1.Toopenthelegs,presstherightpedalBwithafoot.
2.Toclosethelegs,presstheleftpedalAwithafoot.
Withthelever:
1.Toopenthelegs,pulltheleverCtotheright.
2.Toclosethelegs,pushtheleverCtotheleft.
5.5Adjustingtheliftingarmtopatient
bodyheight
CAUTION!
Riskofinjuryordamage
Awrongboompositionmaycausetheliftingarm
tosuddenlyslideinoroutwhenreleased.
–Alwaysadjusttheliftingarmwhenunloaded
andwiththeboominhorizontalposition.
Themainadjustmentcriteriaisthepatientsbodyheight,but
thecorrectsettingalsodependsonotherfactorssuchas
lengthoftheslingloops,patientconditionandpreference.
1654371-A13

Invacare®ISA™
Thefollowingguidelinesarebasedonaveragevaluesand
deviationscananddooccur:
•Position1(fullyshortened):patientheightfromabout
140cmandtaller.
•Position5:patientheightfromabout170cmandtaller.
•Position9(fullyextended):patientheightuptoabout
200cm.
IMPORTANT
Anindividualtrialttingisstronglyrecommended
todetermineacorrectandsafesettingforeach
patient.
1.PushandholdthereleasehandleAonthelegsupport
Bupwards.
2.SlidethelegsupportBupordown.
3.Whenthedesiredheightisreached,releasethehandle
AandslightlymovethelegsupportBupordownuntil
anaudibleclickisheardandthelegsupportBislocked
intooneofthe6availableheightpositions.
5.7Usingthelegband
1.
1.PushandholdtheredleverAontheboomupwards.
2.AdjustthelengthoftheliftingarmBtooneofthenine
availablepositionsindicatedonthescaleC,byplacing
thedesiredpositioneldbeforetheedgeoftheboom.
3.ReleasetheleverAandslightlymovetheliftingarmB
backandforthuntilanaudibleclickisheardanditis
lockedintoposition.
5.6Adjustingtheheightoftheleg
support
Toapplythelegband,wrapthelongpaddedpart
aroundthelegsupportandthepatient´slowerlegsand
pushthecatchAintothebuckleclaspBontheback
sideofthelegsupport.
2.
Toopen,pressthereleasemechanismCandpullthe
catchAoutofthebuckleclaspB.
3.
Toadjustthelengthofthelegband,eitherpullonthe
lugAtoshortenorpushthelugAtowardsthebuckle
tolengthen.
14
1654371-A

Usage
Whenthereisnopatientinthelift,wraptheleg
bandaroundthelegsupport,closeandshortenit
totighten.
5.8Emergencyfunctions
5.8.1Performinganemergencystop
1.PresstheredemergencybuttonAonthecontrolunit
tostoptheboomandpatientfromraisingorlowering.
2.Toreset,rotatetheemergencybuttonclockwise.
5.8.2Activatinganemergencylowering(CBJ
Homecontrolunit)
Ifthehandcontrolfails,theboomcanbeloweredbyusing
thecircularswitchforemergencyrelease.
5.8.4Activatinganemergencylifting(CBJCare,
CBJ1controlunit)
Ifthehandcontrolfails,theboomcanberaisedbyusing
thecircularswitchforemergencylifting.Useasharpobject,
suchasapenciltopressthebutton.
1.RaisetheboombypressingandholdingbuttonBatthe
frontofthecontrolunit.
2.Stopraisingtheboombyreleasingthebutton.
5.8.5Activatingamechanicalemergency
lowering
Iftheemergencyloweringonthecontrolunitisnotworking,
amechanicalemergencyloweringisavailableasaback-up.
Thismayhappenincaseofpartialortotalpowerfailure,or
ifthebatteryrunsdownduringuse.
1.LowertheboombypressingandholdingbuttonAat
thefrontofthecontrolunit.
2.Stoploweringtheboombyreleasingthebutton.
5.8.3Activatinganemergencylowering(CBJ
Care,CBJ1,CBJ2controlunit)
Ifthehandcontrolfails,theboomcanbeloweredbyusing
thecircularswitchforemergencyrelease.Useasharp
object,suchasapenciltopressthebutton.
1.LowertheboombypressingandholdingbuttonAat
thefrontofthecontrolunit.
2.Stoploweringtheboombyreleasingthebutton.
WARNING!
RiskofInjuryorDamage
Repeatedmechanicalemergencyloweringleads
toahighincreaseoftheloweringspeed.
–Onlyusethemechanicalemergencylowering
asaback-upwhentheemergencyloweringon
thecontrolunitisnotworking.
–Donotusethemechanicalemergencylowering
repeatedlyinclosesuccession.
–Alwayslowerthepatientontoasupporting
surfacelikeabedorachair .
–Afteramechanicalemergencyloweringwas
required,theliftmustbecheckedtoresolve
themalfunctionofthecontrolunit.
1.Locatethered
emergencygripA
atthebottomofthe
actuatorpiston.
2.Slowlypullupthered
emergencygripAand
holditinpositionwhen
asafeloweringspeed
isreached.
3.Ifnoloweringoccurs
withafullypulled
emergencygripA,
pushdownonthe
boomatthesame
time.
Theeffectiveloweringspeedisdependingonthepatient
weight.Ifthespeedistoolowortoohigh,itcanbeadapted
tothepatientweight.
Adjustingtheloweringspeed:
1.Locatethescrewinthebottomholeofthered
emergencygripA.
2.Turnthescrewcounter-clockwisetoincreasethespeed.
3.Turnthescrewclockwisetodecreasethespeed.
1654371-A15

Invacare®ISA™
5.9Chargingthebattery
IMPORTANT!
–Makesuretheemergencystopisnotactivated
whilechargingthebattery.
–Makesurethatchargingtakesplaceinaroom
withgoodairventilation.
–Electricalfunctionsarenotworkingwhenthe
liftisconnectedtopowersupply.
–Donotattempttousetheliftifthebattery
housingisdamaged.
–Replaceadamagedbatteryhousingbefore
furtheruse.
–Donotmovetheliftwithoutunpluggingfrom
thepoweroutlet.
Itisrecommendedtochargethebatterydailytoensure
optimaluseoftheliftandprolongthelifeofthebattery.
Furthermore,itisrecommendedtochargethebattery
beforerstuse.
5.9.1CBJHomecontrolunit
Thecontrolunitisequippedwithasoundsignal.Abeep
indicatesthatthebatteryhaslowcapacity,butloweringthe
patientisstillpossible.Itisrecommendedtochargethe
batteriesassoonasthesoundsignalisheard.
patientisstillpossible.Itisrecommendedtochargethe
batteriesassoonasthesoundsignalisheard.
1.PlugthepowercordAintoapoweroutlet.
Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhenthe
batteriesarefullycharged.
TherightyellowdiodeBwilllightcontinuously
duringcharging,andswitchoffwhenfully
charged.
TheleftgreendiodeCwilllightcontinuously
whilethecontrolunitisconnectedtothemains.
2.DisconnectthepowercordAfromthepoweroutlet
afterthebatteryhasbeenfullycharged.
BatteryIndicator
Thecontrolunitmaybeequippedwithabatteryindicator
D,indicatingtheremainingbatterycapacity.
CBJCare
Battery
Indicator
Type
Battery
State
Full
Charge
Description
ThebatteryisOK—noneed
forcharging(100–50%).The
topLEDisGREEN.
1.PlugthepowercordCintoapoweroutlet.
Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhenthe
batteriesarefullycharged.
TheupperyellowdiodeAwillashduring
charging,andswitchtocontinuouslightwhen
fullycharged.
ThelowergreendiodeBwilllightcontinuously
whilethecontrolunitisconnectedtothe
mains,andlightsupwhenanybuttononthe
handcontrolispressedorwhentheelectric
emergencyloweringisactivated.
2.Disconnectthepowercordfromthepoweroutletafter
thebatteryhasbeenfullycharged.
5.9.2CBJCare,CBJ1,CBJ2controlunit
CBJ1andCBJ2withLCD
Battery
Indicator
Type
Partial
Charge
Low
Charge
Low
Charge
(LED
ashing)
Battery
State
Full
Charge
(100%)
Thebatteryneedstobe
charged(50–25%).Themiddle
LEDisYELLOW.
Thebatteryneedstobe
charged(Lessthan25%).
Acousticsignalwhenabutton
ispressed.ThebottomLEDis
YELLOW .
Thebatteryneedstobe
charged.
Someofthefunctionalityof
theliftislostanditisonly
possibletolowertheboom.
Description
ThebatteryisOK—noneed
forcharging(100%).
Partial
Charge
(75%)
Partial
Charge
(50%)
Thecontrolunitisequippedwithasoundsignal.Abeep
indicatesthatthebatteryhaslowcapacity,butloweringthe
161654371-A
ThebatteryisOK—noneed
forcharging(75%).
Thebatteryneedstobe
charged(50%).

Usage
Battery
Indicator
Battery
State
Description
Type
Low
Charge
(25%)
Low
Charge
(0%)
Thebatteryneedstobe
charged(25%).Acousticsignal
whenabuttonispressed.
Thebatteryneedstobe
charged.
Someofthefunctionalityof
theliftislostanditisonly
possibletolowertheboom.
Optionalhandcontrol
ForCBJCareaoptionalhandcontrolwithbatteryindicator
isavailable.
Battery
Indicator
Battery
State
Description
Type
Full
Charge
ThebatteryisOK—noneed
forcharging(100–50%).The
rightLEDisGREEN.
Partial
Charge
Thebatteryneedstobe
charged(50–25%).Themiddle
LEDisYELLOW.
Low
Charge
Thebatteryneedstobe
charged(Lessthan25%).
Acousticsignalwhenabutton
ispressed.TheleftLEDis
YELLOW .
Low
Charge
(LED
ashing)
Thebatteryneedstobe
charged.
Someofthefunctionalityof
theliftislostanditisonly
possibletolowertheboom.
5.9.3Optionalbatterycharger
(forcontrolunitswithdetachablebatteryonly)
Removingthebattery
1.2.
1.LiftuponthehandleAonthebackofthebatteryB.
2.Liftthebatteryupandoutawayfromthecontrolunit
orcharger.
Installingthebattery
CAUTION!
Installingthebatteryimproperlymaycauseinjury
ordamage.
–Makesurethereisanaudibleclickwhen
installingthebatteryonthecontrolunitor
chargertoconrmproperinstallation.
1.PlacethebatteryB
onthecontrolunitor
chargerasshownand
makesurethereisan
audibleclick.
Whenthebatteryisplacedonthecharger,thecharge
LEDwillilluminate.Whenchargingiscomplete,
chargeLEDwillswitchoff .
Abatteryneedingtobefullyrechargedwilltake
approximatelyfourhours.
Theprocedureofremovingorinstallingthebattery
isthesameforthecontrolunitandthecharger.
1654371-A
17

Invacare®ISA™
6PatientTransfer
6.1Generalsafetyinformation
WARNING!
RiskofInjuryorDamage
Improperuseofthisproductmaycauseinjury
ordamage.
–Beforetransferringtoastationaryobject
(wheelchair,bed,commodeorothersurface)
checkthatitsweightcapacitycanwithstand
thepatient’sweight.
–Ifapplicablethebrakesofthestationaryobject
(wheelchair,bedetc.)mustbeappliedbefore
loweringthepatientontoorliftingthepatient
off.
–Whenusingtheliftinconjunctionwithbedsor
wheelchairs,beawareofthepositionofthelift
inrelationshiptothoseotherdevicessothat
theliftdoesnotbecomeentangled.
–Beforepositioningthelegsofthepatientlift
underabed,makesurethattheareaisclear
ofanyobstructions.
WARNING!
RiskofInjuryorDamage
Improperuseofthisproductmaycauseinjury
ordamage.
–Usethepushbarsonthemastatalltimesto
pushorpullthelift.
–Avoidusingtheliftonanincline.Invacare
recommendsthattheproductonlybeusedon
aatsurface.
–Duringtransfer,withthepatientsuspendedin
aslingattachedtothelift,donotrollliftover
unevensurfacesthatcouldcausethelifttotip
over.
–Duringtransfer,withthepatientsuspendedin
aslingkeepthelegsatthemaximumopen
positionforoptimumstability .Ifitbecomes
necessarytomovethroughanarrowpassage,
closethelegsonlyaslongasrequiredand
returntomaximumopenpositionassoonas
possible.
WARNING!
RiskofInjury
Injurytothepatientand/orassistantmayoccur
–Alwaysbeawareofthepositionofthelifting
armandthepatient.
–Alwaysbeawareofthepositionofthefoot
plate,especiallythepatient'spositiononthe
footplate.
–Beforepositioningthelegsoftheliftaround
thepatient,makesurethatthepatient'sfeet
areoutofthewayofthefootplate.
–Alwaysensurethatthefootplateand/or
patient'sfeetaredrybeforeperforminga
transfer .
WARNING!
RiskofEntrapmentorStrangulation
Thehandcontrolcordcancauseinjuryif
improperlypositionedandsecured.
–Alwaysbeawareofthelocationofthehand
controlcordinrelationtothepatientand
caregivers.
–Donotallowthehandcontrolcordtobecome
entangledaroundthepatientandcaregivers.
–Thehandcontrolmustbesecuredproperly.
ALWAYSstorethehandcontrolproperlywhen
notinuse.
WARNING!
RiskofEntrapmentorStrangulation
Itemsinthepatient’ssurroundingscancause
entrapmentstrangulationduringlifting.Toavoid
entrapmentorstrangulation:
–Beforelifting,checkthatthepatientis
completelyfreeofhis/hersurroundings.
WARNING!
RiskofEntrapment
Thereisariskofentrapmentbetweenthehooks
andthesling.
–Usecautionwhenlifting.
–Neverputhandsorngersonornearthe
hookswhenlifting.
–Ensurethepatient’shandsandngersareaway
fromthehooksbeforelifting.
WARNING!
RiskofInjuryorDamage
Damagetopartsofthelift(handcontrol,castors,
etc.)causedbyimpactwiththeoor,wallsor
otherstationaryobjectsmaycausedamageto
theproductandleadtoinjury.
–Donotallowpartsofthelifttoimpactthe
oor,wallsorotherstationaryobjects.
–Alwaysstorethehandcontrolproperlywhen
notinuse.
181654371-A
WARNING!
Riskofinjury
Injurytothepatientand/orassistantmayoccur:
–Ifliftingarmand/orlegsupportisnotadjusted
correctlytothepatientheight.
–Iftheliftingarmadjustmentisnotlocked,
damagedorwithwornparts.
IMPORTANT!
Alltransferproceduresdescribedinthefollowing
canbeperformedbyone(1)assistant.However,
Invacarerecommendsthattheproceduresare
performedbytwo(2)assistantswhenever
possible.
Onlyliftthepatientashighasrequired.Alower
centerofgravityprovidesbetterstability,makesthe
patientfeelmoresecureandthelifteasiertomove.

6.2Attachingtheslingtothelift
WARNING!
RiskofInjury
Usageofwrongordamagedslingscancausethe
patienttofallorcauseinjurytoassistants.
–UseanInvacareapprovedslingthatis
recommendedbytheindividual’sdoctor ,nurse
ormedicalassistantforthecomfortandsafety
oftheindividualbeinglifted.
–Invacareslingsandpatientliftaccessoriesare
specicallydesignedtobeusedinconjunction
withInvacarepatientlifts.
–Aftereachlaundering(inaccordancewith
instructionsonthesling),inspectsling(s)for
wear,tear,andloosestitching.
–Bleached,torn,cut,frayed,ordamagedslings
areunsafeandcouldresultininjury.Discard
immediately.
–DONOTalterslings.
PatientT ransfer
1.PlacetheloopsAoftheslingBoverthehooksCon
theliftingarm.
WARNING!
RiskofInjury
Improperlyattachedoradjustedslingscancause
thepatienttofallorcauseinjurytoassistants.
–Besuretochecktheslingattachmentseach
timetheslingisremovedandreplaced,to
ensurethatitisproperlyattachedbeforethe
patientisremovedfromastationaryobject
(bed,chairorcommode).
–DONOTuseanykindofplasticback
incontinencepadorseatingcushionbetween
patientandslingmaterialthatmaycausethe
patienttoslideoutoftheslingduringtransfer.
–Positionthepatientintheslingasdirectedby
theinstructionsprovidedwiththesling.
–Adjustmentsforpatientsafetyandcomfort
shouldbemadebeforemovingthepatient.
Theloopsoftheslingmaybeequippedwithcolorcoded
strapsprovidingdifferentlength,toplacethepatientin
variouspositions.Matchthecorrespondingcolorsoneach
sideoftheslingforanevenliftofthepatient.
Invacare®StandAssistsling
WARNING!
RiskofInjury
–UsingaStandAssistslingthepatientmust
beabletosupportthemajorityoftheirown
weight.
–MakesurethebottomedgeoftheStandAssist
slingisplacedonthelowerbackofthepatient
andthepatient`sarmsareoutsideofthesling.
–Thebeltmustbesnugtopreventthepatient
fromslidingoutofthesling.
Invacare®TransferStandAssistsling
WARNING!
RiskofInjury
–MakesurethebottomedgeoftheTransfer
StandAssistslingisplacedatthebaseofthe
patient`sspineandthepatient`sarmsare
outsideofthesling.
–Donotraisethepatienttoafullstanding
positionusingtheTransferStandAssistsling.
1.PlacetheupperloopsAoftheslingBoverthehooks
Contheliftingarm.
2.PlacethelowerloopsDoftheslingBoverthehooks
Eontheboom.
1654371-A19

Invacare®ISA™
6.3Transferringapatient
Liftingfromaseatingsurface
1.Beforeproceeding,reviewtheinformationandobserve
allwarningsindicatedin2Safety,page5
Transfer ,page18.
2.Ifapplicable,ensurethebrakesofthestationaryobject
(e.g.bed,wheelchairetc.)areapplied.
3.Positiontheslingaroundthepatient.Refertoyoursling
usermanual.
4.Adjusttheliftingarmtothepatientheight,ifrequired.
See5Usage,page13.
5.Unlocktherearcastorsofthelift.See5Usage,page13.
6.Openthelegsofthelifttomaximum.See5Usage,
page13.
7.Usethepushbartomovetheliftintoposition.
•Fortransferringapatientfromabedthelegsof
theliftareplacedunderthebed
•Fortransferringapatientfromawheelchair ,
commode,chairetc.thelegsoftheliftareplaced
oneachsideofthestationaryobject.
8.Instructthepatienttoplacetheirfeetontothefoot
plateandtheirlowerlegsagainstthelegsupport.Assist
thepatientifnecessaryandensureproperplacementof
thefeetonthefootplate
and6Patient
9.Ensuretheupperedgeofthelegsupportcushionis
placedabout2ngersbelowthekneecaps(patella)
ofthepatient.Adjusttheheightofthelegsupportif
necessary.See5Usage,page13.
10.Ifapplicable,adjustandapplythelegbandaroundthe
patientslowerlegs.See5Usage,page13.
11.Lowerthepatientliftforeasyattachmentofthesling.
See5Usage,page13.
12.Locktherearcastorsofthelift.See5Usage,page13
13.Attachthesling.See6.2Attachingtheslingtothelift,
page19
14.Unlocktherearcastorsofthelift.See5Usage,page13.
15.Instructthepatienttoholdontothehandgripsonboth
sidesoftheliftingarm.
16.Instructthepatienttoleanbackintothesling.
17.PresstheUParrowbuttononthehandcontroltoraise
thepatientabovethestationaryobject.See5Usage,
page13
18.Beforemovingthepatient,checkagaintomakesure
thattheslingisproperlyconnectedtothehooksofthe
liftingarm.Ifanyattachmentsarenotproperlyinplace,
lowerthepatientbackontothestationaryobjectand
correcttheproblem.
19.Usingthepushbar,movetheliftawayfromthe
stationaryobject.
Loweringtoaseatingsurface
1.Beforeproceeding,reviewtheinformationandobserve
allwarningsindicatedin2Safety,page5and6Patient
Transfer ,page18.
2.Ensurethepatientisraisedhighenoughtoclearthe
surfacetobetransferredto.Raisethepatientand/or
lowerthestationaryobjectifapplicable.
3.Ifapplicable,ensurethebrakesofthestationaryobject
(e.g.bed,wheelchairetc.)areapplied.
4.Positionthepatientasfarovertheseatingsurfaceas
possible.
•Fortransferringapatienttoabedthelegsofthe
liftareplacedunderthebed
•Fortransferringapatienttoawheelchair,commode,
chairetc.thelegsoftheliftareplacedoneach
sideofthestationaryobject.
5.PresstheDOWNarrowbuttonandlowerthepatient
ontotheseatingsurface.See5Usage,page13
6.Locktherearcastorsofthelift.See5Usage,page13
7.Unhooktheslingfromallattachmentpointsonthe
standassistlift.
8.Ifapplicable,removethelegbandfromaroundthe
patientslowerlegs.See5Usage,page13
9.Instructthepatienttolifttheirfeetoffofthefootplate.
Assistthepatientifnecessary.
10.Unlocktherearcastorsofthelift.See5Usage,page13
11.Movetheliftawayfromthestationaryobject.
12.Removetheslingfromaroundthepatient.
ThepatientcanremainintheInvacare®Stand
Assistslingwhileusingthecommode.Inthis
casetheslingmustnotbeplacedaroundthe
patientagainbeforeliftingfromthecommode.
201654371-A

TransportationandStorage
7TransportationandStorage
7.1Generalinformation
Duringtransportation,orwhenthepatientliftisnottobe
usedforsometime,theemergencystopbuttonshouldbe
pushedin.
Adisassembledliftcanbelocatedinthepackagingbox,for
transportationandstorage.
Thepatientliftmustbestoredatnormalroomtemperature.
Ifitisstoredinadamp,coldorwetenvironmentthenthe
motorandothermountingpartsmaybepronetocorrosion.
See11.4Environmentalconditions,page28
.
7.2Disassemblingthelift
1.Removethelegspreaderleverifattached.
2.Lowertheboomandclosethelegscompletely.
3.Activatetheemergencystopbuttonandapplycastor
brakes.
4.Adjusttheliftingarmtominimumlength.See5.5
Adjustingtheliftingarmtopatientbodyheight,page13.
5.Reversetheproceduredescribedin4.6Installingthe
legsupport,page11.
6.Reversetheproceduredescribedin4.3Installingthe
masttothebase,page10.
7.Reversetheproceduredescribedin4.4Installingthe
footplate,page11.
1654371-A
21

Invacare®ISA™
8Maintenance
8.1Generalmaintenanceinformation
CAUTION!
Riskofinjuryordamage
–Donotperformanymaintenanceorservice
procedureswhiletheproductisinuse.
Followthemaintenanceproceduresdescribedinthismanual
tokeepyourproductincontinuousservice.
Furthermaintenanceandinspectionprocedures,tobe
performedbyaqualiedtechnician,aredescribedinthe
servicemanualofthisproduct.Servicemanualscanbe
obtainedfromInvacare.
8.2Dailyinspections
WARNING!
Riskofinjuryordamage
Damagedorwornpartscanaffectthesafetyof
thelift.
–Thepatientliftshouldbecheckedeachtime
itisused.
–Donotusetheliftifdamageisfoundoryou
questionthesafetyofanypartofthelift.
ContactyourInvacareproviderimmediately
andensuretheliftisnotuseduntilrepairsare
made.
CAUTION!
Riskofelectricshockandproductdamage
–Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
–Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
–Makesurethatnowatersplashestotheplug
orthewalloutlet.
–Donottouchthepowersocketwithwethands.
IMPORTANT!
Wronguidsormethodscanharmordamage
theproduct.
–Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedto
clean.
–Neverusecorrosiveuids(alkalines,acidetc.)
orabrasivecleaningagents.Werecommend
anordinaryhouseholdcleaningagentsuchas
dishwashingliquid,ifnotspeciedotherwisein
thecleaninginstructions.
–Neveruseasolvent(cellulosethinner ,acetone
etc.)thatchangesthestructureoftheplastic
ordissolvestheattachedlabels.
–Alwaysmakesurethattheproductiscompletely
driedbeforetakingintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term
careenvironments,followyourin-houseprocedures.
DailyInspectionChecklist
qVisuallyinspectthepatientlift.Checkallpartsfor
externaldamageorwear.
qCheckallhardware,attachmentpointsandstressed
parts,suchasslings,hooksandanypivotpointsforsigns
ofwear,cracking,fraying,deformationordeterioration.
qVerifythatthehandcontrolisfunctional(liftingand
legmovements).
qChargethebatteryeverydaytheliftisused.
qChecktheemergencystopfunction.
qChecktheservicelight(CBJCarecontrolunitonly).
Theservicelightisindicatedbyawrench
symbolonthecontrolunitandifapplicable
onthehandcontrol.
–Iftheservicelightisashingyellowtheliftrequires
service.DonotusetheliftandcontactyourInvacare
providerforservice.
–Iftheservicelightisnotashingtheliftisreadyfor
use.
8.3CleaningandDisinfection
8.3.1Generalsafetyinformation
CAUTION!
Riskofcontamination
–T akeprecautionsforyourselfanduse
appropriateprotectiveequipment.
8.3.2CleaningIntervals
IMPORTANT!
Regularcleaninganddisinfectionenhances
smoothoperation,increasestheservicelifeand
preventscontamination.
Cleananddisinfecttheproduct
–regularlywhileitisinuse,
–beforeandafteranyserviceprocedure,
–whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser .
8.3.3Cleaninginstructions
IMPORTANT!
–Theproductdoesnottoleratecleaningin
automaticwashingsystems,withhigh-pressure
cleaningequipmentorsteam.
Cleaningthelift
Method:Wipeoffwithawetclothorsoftbrush.
Max.temperature:40°C
Solvent/chemicals:Ordinaryhouseholdcleaningagentand
water .
Drying:Wipedrywithasoftcloth.
Cleaningthesling
Refertothewashinginstructionsontheslingandtothe
slingmanualforcleaningdetails.
22
1654371-A

Maintenance
8.3.4Disinfectioninstructions
Informationonrecommendeddisinfectants
andmethodscanbefoundonhttps://vahonline.de/en/for-users.
Indomesticcare
Method:Followtheapplicationnotesfortheused
disinfectantandwipedisinfectallaccessiblesurfaces.
Disinfectant:Ordinaryhouseholddisinfectant.
Drying:Allowtheproducttoair-dry.
Inclinicalandlongtermcare
Followyourin-housedisinfectionproceduresandonlyuse
thedisinfectantsandmethodsspeciedtherein.
8.4Serviceinterval
WARNING!
Riskofinjuryordamage
Servicemustbeperformedonlybyaqualied
technician.
–ContactyourInvacareproviderforservice.
Servicemustbeperformedatleastevery12monthsunless
otherwisestatedinlocalrequirements.
8.4.1LOLERStatement
TheUKHealthandSafetyExecutive’sLiftingOperationsand
LiftingEquipmentRegulations1998,requireanyequipment
thatisusedintheworkplacetoliftaloadbesubjectto
safetyinspectiononasixmonthlybasis.Pleaserefertothe
HSEwebsiteforguidancewww.hse.gov.uk.
Thepersonresponsiblefortheequipmentmustensure
adherencetoLOLERregulations.
1654371-A23

Invacare®ISA™
9AfterUse
9.1Disposal
WARNING!
EnvironmentalHazard
Devicecontainsbatteries.
Thisproductmaycontainsubstancesthatcould
beharmfultotheenvironmentifdisposedof
inplaces(landlls)thatarenotappropriate
accordingtolegislation.
–DONOTdisposeofbatteriesinnormal
householdwaste.
–BatteriesMUSTbetakentoaproperdisposal
site.Thereturnisrequiredbylawandfreeof
charge.
–Doonlydisposedischargedbatteries.
–Coverterminalsoflithiumbatteriespriorto
disposal.
–Forinformationonthebatterytypeseebattery
labelorchapter11TechnicalData,page27
Beenvironmentallyresponsibleandrecyclethisproduct
throughyourrecyclingfacilityatitsendoflife.
Disassembletheproductanditscomponents,sothedifferent
materialscanbeseparatedandrecycledindividually.
Thedisposalandrecyclingofusedproductsandpackaging
mustcomplywiththelawsandregulationsforwaste
handlingineachcountry.Contactyourlocalwaste
managementcompanyforinformation.
9.2Reconditioning
Thisproductissuitableforreuse.T oreconditiontheproduct
foranewuser ,carryoutthefollowingactions:
•Cleaninganddisinfection
•Inspectionaccordingtoserviceplan
Fordetailedinformationaboutinspection,cleaningand
disinfection,consulttheusermanualandservicemanual
forthisproduct.
.
24
1654371-A

10Troubleshooting
10.1Identifyingfaultsandpossiblesolutions
SymptomsFaultsSolution
Patientliftfeelsloose
Castors/brakesnoisyorstiff
NoisyordrysoundfrompivotsPivotsarewornordamagedContactyourInvacareprovider
Electricactuatorfailstoraiseorlower
theboomorlegsfailtoopenorclose
whenbuttonispressed
UnusualnoisefromactuatorActuatorisdamagedContactyourInvacareprovider
Boomwillnotlowerwhenactivating
themanualemergencyrelease
Thecontrolunitemitsabeepingsound
duringlifting,andthemotorstops
BatterydoesnotchargePowercordisnotproperlyconnectedto
LiftingarmcannotbeadjustedTheredleverontheboomisnot
Legsupportcannotbeinstalledor
adjusted
Mast/BasejointlooseRetightentheMast/Baseconnection.
See4.3Installingthemasttothebase,
page10
ManuallegspreadermechanismlooseContactyourInvacareprovider
FluffordebrisinbearingsCleancastorsfromuffanddebris
Castors/brakesaredamaged
Handcontroloractuatorconnector
loose
BatterylowChargethebattery.See5.9Charging
EmergencystopbuttonispressedinRotatetheemergencystopbutton
Batterynotconnectedproperlyto
controlunit
TheconnectingterminalsaredamagedContactyourInvacareprovider
PowercordconnectedtopoweroutletDisconnectpowercordfrompower
Boomactuatororlegactuatoris
damaged
Max.loadisexceeded
Boomrequiresaminimumweightload
tolowerfromtheuppermostposition
Max.loadisexceeded
thecontrolunit/andorpoweroutlet
PowercordisdamagedContactyourInvacareprovider
pushedandheldupwards
Liftingarmisalreadyatthemaximum
orminimumposition
TheadjustmentmechanismisdamagedContactyourInvacareprovider
Boomisinwrongposition
Thereleasehandlenotpushedandheld
upwards
TheadjustmentmechanismisdamagedContactyourInvacareprovider
Legbandiscloggingthebackoftheleg
support
ContactyourInvacareprovider
Connecthandcontroloractuator
connector.Ensureconnectorsareseated
properlyandfullyconnected
thebattery,page16
clockwiseuntilitpopsout
Reconnectthebatterytothecontrol
unit.See5.9.3Optionalbatterycharger,
page17
outlet
ContactyourInvacareprovider
Reducetheload(andtheliftwill
functionnormally)
Pulldownslightlyontheboom
Reducetheload(andtheliftwill
functionnormally)
Ensureconnectorsareseatedproperly
andfullyconnected.See5.9Charging
thebattery,page16
See5.5Adjustingtheliftingarmto
patientbodyheight,page13
See5.5Adjustingtheliftingarmto
patientbodyheight,page13
Movetheboomtohorizontalposition.
See5.3Raising/LoweringanElectricLift,
page13
See4.6Installingthelegsupport,page
11
Clearthebackofthelegsupport
Troubleshooting
1654371-A25

Invacare®ISA™
SymptomsFaultsSolution
MastcannotbeassembledonthebaseBoompointinginthewrongdirectionBringtheboomincorrectposition.See
4.3Installingthemasttothebase,page
10
LiftdoesnotrollThecastorsarelockedUnlockthecastorsSee5.2
Locking/Unlockingtherearcastors,page
13
ContactyourInvacareprovideriftheabovedoesnotsolveyourproblems.
261654371-A

11TechnicalData
11.1Maximumsafeworkingload
TechnicalData
Max.SafeWorkingLoad(patient+sling)
11.2Dimensionsandweights
ISA™
COMPACT
STANDARD
ISA™
ISA™
PLUS
140kg160kg180kg200kg
ISA™
XPLUS
*Forwarddirection
Dimensions[mm]
ISA™
COMPACT
STANDARD
ISA™
ISA™
PLUS
Front/rearcastordiameter75/75100/100100/100100/100100/125
Max.reachat600mm(a)*
Max.reachfrombase(b)*
Baselength(c)
Totallengthwithoutlever(n)
Totallengthwithlever(l)*
Reachfrombasewithlegsspreadto700
mm(d)
Max.height(f)*
Liftingrange(g)*
Min.heightatSlingHook-up(h
Max.heightatSlingHook-up(h
)*
min
)*
max
Totalwidth(legsopen),externalmeasure(i)
Totalwidth(legsopen)internalmeasure(j)
110-270110-270110-270110-270110-270
340-500340-500340-500340-500340-500
900925108012201280
925925
n/an/a
980-1000980-100011451290
420420225225225
1635-17851650-18001650-18001650-18001650-1800
505-675505-675505-675505-675505-675
945-970960-985960-985960-985960-985
1475-16201490-16351490-16351490-16351490-1635
865885111011701170
755
73596010201020
ISA™
XPLUS
1280
n/a
1654371-A
27

Invacare®ISA™
Dimensions[mm]
Totalwidth(legsclosed),externalmeasure
(r)
Min.internalwidth(s)
Internalwidthatmaximumreach(k)*
Turningdiameter
Heighttoupperedgeoflegs(m)
Min.freeheight(p)
*Rangedependingonfullyshortenedtofullyextendedpositionofliftingarm
WeightsMainParts[kg]
Totalweight
Weight,mast,incl.battery,
Weightmanuallegsection
Weightelectricallegsection
Legsupport
Footplate(high,lowandinclined)
515535635640640
380380495495495
640-725640-725775-870775-870775-870
10601060126014001430
100115115115115
ISA™
COMPACT
2035353535
ISA™
COMPACT
43484952
152021
n/a
ISA™
STANDARD
ISA™
STANDARD
1920
202123
5
4
ISA™
PLUS
ISA™
PLUS
ISA™
XPLUS
ISA™
XPLUS
n/a
11.3Electricalsystem
ISA™
COMPACT
Voltageoutput
Voltagesupply
280mA(for
Maximumcurrentinput
Protectionclass(entiredevice)IPX4*
InsulationclassClassIIequipment
TypeBAppliedPart
Soundlevel
Workingability
Intermittent(periodicmotoroperation)
Batteryspecications
Batterytype
Manualemergencylowering
Electricemergencylowering/lifting
CBJHome)/
400mA(for
CBJ2)
Appliedpartcomplyingwiththespeciedrequirementsforprotection
againstelectricalshockaccordingtoIEC60601-1.
82fulllifts(withbatterybetween100%–50%offullcapacity)
Yes/No(forCBJ2
andCBJHome)
ISA™
STANDARD
24VDC
100–240VAC,50/60Hz
280mA(for
CBJHome)/
400mA(for
CBJ1andCBJ2)
45–50dB(A)
10%,max.2min/18min
2x12V/2.9Ah
Leadacid(servicefree,sealed)
Yes(atbottomofactuator)
Yes/Yes
(forCBJ1)
Yes/No(for
CBJ2andCBJ
Home)
CBJ1,CBJ2and
(forCBJ1and
Yes/No(for
CBJ2andCBJ
280mA(for
CBJHome)/
400mA(for
CBJCare)
Yes/Yes
CBJCare)
ISA™
PLUS
Home)
ISA™
XPLUS
400mA(forCBJ1
andCBJcare)
Yes/Yes(for
CBJ1andCBJ
care)
*IPx4-Thesystemisprotectedagainstwatersplashedfromanydirection.
11.4Environmentalconditions
Storageand
transportation
Temperature
281654371-A
-10°Cto+50°C+5°Cto+40°C
Operation
Relativehumidity20%to90%at30°C,notcondensing
Atmospheric
pressure
800hPato1060hPa

TechnicalData
Allowtheproducttoreachoperationtemperature
beforeusage:
•Warmingupfromminimumstoragetemperature
mighttakemorethan30minutes.
•Coolingdownfrommaximumstorage
temperaturemighttakemorethan5minutes.
11.5Materials
Component
Base,legs,mastandboomSteel,powder-coated
Boomconnector,boltsand
nuts
Handgrips
Slinghooks
LegsupportcushionSelfskinPU
Footplatecover
Actuatorhousing,hand
control,mastprotector,
castorsandotherplastic
parts
Legband
Material
Steel,zinc-plated
TPE
TPU
PVC,PU
Materialaccordingto
marking(PA,PP ,PE)
PA,PES,PVC,PU
Allcomponentsoftheproductareeithercorrosionresistant
orcorrosionprotected.
11.6Operatingforcesofcontrols
ControlOperatingforce
Buttonsoncontrolunit
Buttonsonhandcontrol
FootpedalonISA™COMPACTwith
100mmcastors*
LegspreaderleveronISA™
COMPACT
LegspreaderleveronISA™
STANDARD
LegspreaderleveronISA™PLUS
*Forallothermodelsandcastorsizes,aleverforthe
manuallegspreaderismandatory.
4N
4N
270N
64N
70N
85N
1654371-A29

Invacare®ISA™
12Electromagneticcompatibility(EMC)
12.1GeneralEMCinformation
MedicalElectricalEquipmentneedstobeinstalledandusedaccordingtotheEMCinformationinthismanual.
ThisproducthasbeentestedandfoundtocomplywithEMClimitsspeciedbyIEC/EN60601-1-2forClassBequipment.
PortableandmobileRFcommunicationsequipmentcanaffecttheoperationofthisproduct.
Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabove
standard.Todetermineiftheemissionfromthisproductiscausingtheinterference,runandstoprunningthisproduct.If
theinterferencewiththeotherdeviceoperationstops,thenthisproductiscausingtheinterference.Insuchrarecases,
interferencemaybereducedorcorrectedbythefollowing:
•Reposition,relocate,orincreasetheseparationbetweenthedevices.
12.2Electromagneticemission
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis
productshouldassurethatitisusedinsuchanenvironment.
Emissionstest
RFemissions
CISPR11
RFemissions
CISPR11
Harmonic
emissions
IEC61000-3-2
Voltageuctuations
/ickeremissions
IEC61000-3-3
ComplianceElectromagneticenvironment-guidance
ThisproductusesRFenergyonlyforitsinternalfunction.Therefore,itsRF
GroupI
ClassB
ClassA
Complies
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearby
electronicequipment.
Thisproductissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
12.3ElectromagneticImmunity
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis
productshouldassurethatitisusedinsuchanenvironment.
Immunitytest
Electrostatic
discharge(ESD)
IEC61000-4-2
Electrostatic
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Test/Compliancelevel
±8kVcontact
±2kV,±4kV,±8kV,±
15kVair
±2kVforpowersupply
lines;100kHzrepetition
frequency
±1kVforinput/output
lines;100kHzrepetition
frequency
±1kVlinetoline
±2kVlinetoearth
Electromagneticenvironment–guidance
Floorsshouldbewood,concreteorceramictile.Ifoorsarecoveredwith
syntheticmaterial,therelativehumidityshouldbeatleast30%.
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
301654371-A