Invacare IOH200 Owners Operating & Maintenance Manual

Owner's Operator And Maintenance Manual
Venture® HomeFill
Compressor
Model IOH200
II
THIS MANUAL MUST BE GIVEN TO THE USER OF THE COMPRESSOR.
REFER TO THE INFORMATION ON PAGE 3 OF THIS MANUAL.
USER:
BEFORE USING THE COMPRESSOR, READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE.
SPECIAL NOTES
SPECIAL NOTES
WARNING
DO NOT INSTALL, ASSEMBLE OR OPERATE THIS
EQUIPMENT WITHOUT FIRST READING AND
UNDERSTANDING THIS INSTRUCTION MANUAL,
THE CONCENTRATOR OWNERS MANUAL, AND
THE OXYGEN CYLINDER WARNING LABEL. IF
YOU ARE UNABLE TO UNDERSTAND THE
WARNINGS, CAUTIONS AND INSTRUCTIONS,
CONTACT A DEALER OR TECHNICAL PERSONNEL
IF APPLICABLE BEFORE ATTEMPTING TO INSTALL
THIS EQUIPMENT - OTHERWISE, INJURY OR
DAMAGE MAY OCCUR.
SPECIAL NOTES
Check all parts for shipping damage. In case of damage, DO NOT use. Contact Invacare for further instruction.
WARNING/CAUTION notices used in this manual apply to hazards or unsafe practices which could result in personal injury or property damage.
NOTICE
The Venture HomeFill II Compressor is for use with specially designed Invac­are 5 liter HomeFill II compatible concentrators set at 3 L/min or less ONLY. To determine if your Invacare 5 liter concentrator is HomeFill II compatible, examine the lower right corner of the rear of the concentrator for an outlet fitting, as shown below.
Concentrator
VENTURE
®
HOMEFILL II COMPRESSOR 2
Outlet Fitting
CONTENTS/DEALER INFORMATION
TABLE OF CONTENTS
SPECIAL NOTES .............................................................................. 2
SPECIFICATIONS ............................................................................. 4
LABEL LOCATIONS
PROCEDURE 1 - GENERAL GUIDELINES ............................................ 5
GENERAL WARNINGS ....................................................................................................................... 5
HANDLING WARNINGS ................................................................................................................... 5
PROCEDURE 2 - INITIAL SETUP ....................................................... 6
COMPRESSOR INITIAL SETUP CHECKLIST ................................................................................6
CHOOSING A LOCATION ............................................................................................................... 6
TRANSPORTING THE COMPRESSOR .......................................................................................... 6
INSTALLING THE POWER CORD ................................................................................................. 7
CONNECTING THE COMPRESSOR TO THE CONCENTRATOR .....................................7
PROCEDURE 3 - COMPRESSOR OPERATION ..................................... 8
COMPRESSOR OPERATION CHECKLIST .................................................................................... 8
CYLINDER PREFILL INSPECTION ...................................................................................................8
CONNECTING THE CYLINDER TO THE COMPRESSOR ...................................................10
TURNING THE COMPRESSOR ON .......................................................................................... 11
REMOVING THE CYLINDER ........................................................................................................ 12
SETTING THE FLOW KNOB TO THE PRESCRIBED SETTING....................................... 13
PROCEDURE 4 - INDICATOR LIGHTS ............................................. 14
PROCEDURE 5 - TROUBLESHOOTING ............................................ 15
COMPRESSOR .................................................................................................................................... 15
CYLINDER............................................................................................................................................ 16
PROCEDURE 6 - MAINTENANCE .................................................. 17
CLEANING THE COMPRESSOR FILTERS ................................................................................ 17
CABINET .............................................................................................................................................. 17
CYLINDER FILL TIMES .................................................................. 18
WARRANTY ................................................................................. 19
CONTENTS
DEALER INFORMATION
NOTE: Invacare recommends leaving a full cylinder of oxygen with the patient AFTER setting up the HomeFill II compressor.
CHECKLIST
Before leaving a Venture HomeFill II Compressor with a user, the following checklist MUST be completed:
Make sure the pressure test on cylinder has not expired.Instruct the user on the safe operation of the concentrator and review ALL warnings.Instruct the user on the safe operation of the Venture HomeFill II compressor and review
ALL warnings.
Leave a full cylinder of oxygen with the user.Leave a copy of ALL warnings and filling instructions with the user.
3 VENTURE
®
HOMEFILL II COMPRESSOR
DEALER INFORMATION
SPECIFICATIONS/LABEL LOCATIONS
SPECIFICATIONS
Regulatory Listing:
Operating temperature:
Storage temperature:
SPECIFICATIONS
Input Pressure Required:
Input FLOW Required:
Input O2% Required:
Shipping Weight:
Double Insulated Product ETL Certified Complying with UL1097 (US) ETL Certified Complying with CAN/CSA C22.2 No. 68 (Canada)
COMPRESSOR SPECIFICATIONS
50 to 95° F (10 to 35° C) @ 20 to 60% non-condensing humidity
-10 to 150° F (-23 to 65° C) @ 15 to 95% humidity 14 - 21 psi 2 L/min
Width:
Height:
Depth:
Weight:
>90% O 20-1/4-inches 15-inches 16-inches 33 lbs.
2
39 lbs.
REGULATOR SPECIFICATIONS
Oxygen Output: 0, 0.25, 0.50, 0.75, 1.0, 1.5, 2.0, 2.5, 3.0, 4.0, 5.0, 6.0 L/min
LABEL LOCATIONS
WARNING
DO NOT REMOVE UNIT COVER.
This unit contains moving and high pressure components. Manufacturer service only.
!
LABEL LOCATIONS
FRONT LEFT SIDE
WARNING
!
PART NO.
1075221
Keep all connections free of oil and grease to avoid
violent ignition. See Owner's Manual before cleaning.
PART NO.
1075220
WARNING
!
Keep all connections free of oil and grease.
See owners manual before cleaning. Cylinders filled on this unit must be labeled "Contents of this cylinder is the product of an oxygen concentrator."
VENTURE
®
HOMEFILL II COMPRESSOR 4
PART NO.
1075238
GENERAL GUIDELINES PROCEDURE 1
This Procedure includes the following:
General Warnings
Handling Warnings
In order to ensure the safe installation, assembly and operation of the VENTURE HOMEFILL II COMPRESSOR these instructions MUST be followed:
GENERAL WARNINGS
Caution: Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.
The use of this device is limited to the oxygen patient. Cylinders MUST be used ONLY by the oxygen patient and are not to be distributed to any other indi­vidual for any purpose.
DO NOT use parts, accessories or adapters other than those authorized by Invacare or warranty will be voided.
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like.
DO NOT disassemble. There are minimal serviceable parts on the Venture HomeFill II Compressor. If any service is required, contact Invacare technical service.
If the compressor has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or submersed in water, call a qualified techni­cian for examination and repair.
If any leakage of the oxygen cylinder is detected DO NOT attempt to use the cylinder. Turn the flow selector knob on the regulator to 0. If leakage per­sists, place the cylinder outdoors and notify your Home Health Care Provider or Service Representative of this condition.
DO NOT allow oil from your hands or other source to come into contact with the regulator or cylinder valve connection. These solutions may be flammable and cause injury.
The HomeFill II is equipped with a High Pressure Relief Valve to ensure the users safety. When activated, this safety feature will make an extremely loud noise. If this noise occurs, turn the unit OFF and contact your Home Health Care Provider or Service Representative.
Before moving or repositioning either the compressor or concentrator, always disconnect the AC power cords and the interconnect hose between the com­pressor and concentrator. Failure to do so may result in damage to the unit or personal injury.
Children should always be supervised around the HomeFill unit. When used with the HomeFill table or similar elevated support surfaces, a child could pull over any such table causing the unit to fall. Failure to observe this warning may result in personal injury or product damage.
GENERAL GUIDELINES
Use extreme care when handling and filling an oxygen cylinder. Full oxygen cylin­ders are under pressure and can become a projectile if dropped or mishandled.
NEVER transport a compressor with a cylinder connected to or stored on the compressor, otherwise injury or damage can occur.
HANDLING WARNINGS
5 VENTURE
®
HOMEFILL II COMPRESSOR
INITIAL SETUPPROCEDURE 2
This Procedure includes the following:
Compressor Initial Setup Checklist
Choosing a Location
Transporting the Compressor
Installing the Power Cord
Connecting the Compressor to the Concentrator
NOTE: DEALERS ONLY - Refer to DEALER INFORMATION on PAGE 3 of this manual BEFORE performing this procedure.
INITIAL SETUP
COMPRESSOR INITIAL SETUP CHECKLIST
Before operating the HomeFill II Compressor for the first time, complete the following checklist:
Choose a location for the compressor.Transport the compressor to the desired location.Connect the power cord to the compressor.Connect the compressor to the concentrator.
CHOOSING A LOCATION
The compressor should be at least three (3) inches away from walls, draperies or furniture to ensure sufficient air flow.
Avoid placing the compressor next to heaters, radiators or hot air registers. DO NOT use in a closet. The compressor should sit on a hard flat stable surface, such as a table.
TRANSPORTING THE COMPRESSOR (FIGURE 1)
NEVER transport a compressor with a cylinder connected to or stored on the compressor, otherwise injury or damage can occur.
When transporting the compressor, be careful not to bump the compressor or the connectors into obstacles. Otherwise damage to the compressor may occur.
There are two (2) methods to safely transport the compressor:
A. ONE HAND - Grasp the handle at the
bottom rear of the compressor.
B. TWO HANDS - Grasp the hand grips on
the sides of the compressor.
VENTURE
®
HOMEFILL II COMPRESSOR 6
WARNING
CAUTION
Hand
Grip
FIGURE 1 - TRANSPORTING THE
Hand
Grip
Handle
COMPRESSOR
Loading...