Invacare HomeFill Oxygen System User Manual

Invacare®HomeFill®Oxygen System
ModelIOH200
enCompressor
UserManual
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
Contents
1General.........................................4
1.1Symbols.....................................4
1.2IntendedUse.................................5
2Safety..........................................7
2.1GeneralGuidelines.............................7
2.2LabelLocations................................10
2.3DealerInformation.............................11
2.4Double-InsulatedProducts........................11
2.5EMCInformation...............................11
3Setup...........................................13
3.1Unpacking....................................13
3.2CompressorInitialSetupChecklist..................13
3.3ChoosingaLocation............................14
3.4TransportingtheCompressor......................14
3.5InstallingthePowerCord........................15
3.6ConnectingtheCompressortotheConcentrator.......16
4Usage..........................................17
4.1CompressorOperationChecklist...................17
4.2CylinderPrellInspection........................17
4.3ConnectingtheCylindertotheCompressor...........19
4.4DisconnectingtheCylinderfromtheCompressor.......20
4.5TurningtheCompressorOnandOff................21
5Maintenance.....................................23
5.1MaintenanceInfo..............................23
5.2WearandTear................................23
5.3ServiceLife...................................23
5.4CleaningtheCompressorFilter....................24
5.5CleaningtheCabinet............................24
5.6ReplacingtheFuse.............................25
6AfterUse........................................26
6.1Disposal.....................................26
7Troubleshooting...................................27
7.1IndicatorLights................................27
7.2Troubleshooting...............................29
8TechnicalData....................................32
8.1TechnicalDescription............................32
8.2Specications.................................32
9Warranty........................................34
9.1LimitedWarranty..............................34
Invacare®HomeFill®OxygenSystem
1General
1.1Symbols
SymbolsintheManual
Signalwordsareusedinthismanualandapplytohazards orunsafepracticeswhichcouldresultinpersonalinjuryor propertydamage.Seetheinformationbelowfordenitions ofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin deathorseriousinjury.
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin deathorseriousinjury.
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin propertydamageorminorinjuryorboth.
IMPORTANT
–Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformation forefcient,trouble-freeuse.
SymbolsontheProduct
Thefollowingsymbolsarefoundonthecompressor , compressorpackagingortheoxygencylinder .
ReadtheManual(Compressor)
ReadtheManual(Cylinder)
ElectricalHazard
AlternatingCurrent
TypeBFequipment—AppliedParts
Powerswitch“ON”
Powerswitch“OFF”
DONOTsmoke
4
1100873-M
General
NoOpenFlame
ClassII,DoubleInsulation.Referto
2.4Double-InsulatedProducts,page
11.
Keepdryintransportorstorageand duringuse
Fuse
Oxidizer
TransportandStorageT emperature
TransportandStorageHumidity
Recyclethisproduct.Referto6.1 Disposal,page26.
1.2IntendedUse
WARNING!
–Theuseofthisdeviceislimitedtorellinga
specicpatient’sownoxygencylinders.The designatedcylindersMUSTbeusedONL Yby thespecicoxygenpatientandarenotto bedistributedtoanyotherindividualforany purpose.
–Nootheruseofthisdeviceisindicated.
WARNING! RiskofDeath,InjuryorDamage
TheInvacareHomeFillcompressorwas designedtobeusedonlywithInvacareoxygen concentrators.Usewithothertypesofdevices maycauseinjury,deathorpropertydamage.
–DONOTusetheInvacareHomeFillcompressor
withanythingotherthananInvacareoxygen concentrator .
–UseonlywithspeciallydesignedInvacare
oxygenconcentratorsthatareHomeFill compatible.
TodetermineifyouroxygenconcentratorisHomeFill compatible,examinethelowerrightcorneroftherearof theoxygenconcentratorAforanoutletttingB.
1100873-M5
Invacare®HomeFill®OxygenSystem
Platinum®—TopView
Perfecto2™—RearView
TheintendeduseofanInvacareHomeFillcompressoris anaccessorydevicetoanInvacareoxygenconcentrator thatenablesthepatienttorellcustomportableoxygen cylindersforpersonaluse.
Thisproductisintendedforsinglepatientuseintheindoor homecareenvironmentonlyandisnottobeusedforbulk cylinderrelling.
Nospecicproductknowledgeortrainingisrequired, otherthanwhatiscontainedinthismanual,tooperate theproduct.
Itisnotintendedforsupplyingoxygentherapyduring relling.
Thisproductisintendedtobeusedbyorunderadult supervisiononlyafterreadingandunderstandingthe instructionsandwarningsofthisusermanual.
61100873-M
Safety
2Safety
2.1GeneralGuidelines
DANGER! RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath, injuryordamage.Thissectioncontainsimportant informationforthesafeoperationanduseofthis product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanuals,servicemanualsorinstruction sheetssuppliedwiththisproductoroptional equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professional,dealerortechnicalpersonnel beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct isnotworkingcorrectly,contactatechnician orInvacareforrepair .
–Inordertoensurethesafeinstallation,
assemblyandoperationoftheHomeFill CompressortheseinstructionsMUSTbe followed.
–THEINFORMA TIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–DONOTSMOKEwhileusingthisdevice. –DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonthecompressor
unlessrecommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Keepallmatches,lightedcigarettesorother
sourcesofignitionoutoftheroominwhich thiscompressorislocatedandawayfrom whereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andcompressor
outfromundersuchitemsasblankets, bedcoverings,chaircushions,clothing,and awayfromheatedorhotsurfacesincluding spaceheaters,stoves,andsimilarelectrical appliances.
1100873-M
7
Invacare®HomeFill®OxygenSystem
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–DONOTallowoilfromyourhandsorother
sourcestocomeintocontactwiththe compressorcouplingorcylinderttings.These substancesmaybecomeammableinthe presenceofoxygenandcauseinjury.Avoid touchingthecylinderttingwheneverpossible. Exampleofitemstoavoidarehandcreamsor lotions,cookingoils,suntanoilsorsunblocks andsimilaroilyproducts.Ifyouusethese typesofproductsInvacaresuggestsyouwash yourhandspriortousingtheHomeFillunit. Ifthecylinderttingshouldcomeincontact withsuchsubstancesitshouldbecleanedoff withadampclothpriortoconnectingittothe compressor .
–NEVERoilorlubricatethecompressorcoupler
orcylinderttings.NEVERuseanypenetrating oilorlubricantsuchasWD-40®or3-in-1 Oil®.Ifconnectiondifcultyisexperienced verifythatthecylinderttingisbeinginserted straightdownintothecompressorcoupler.If continueddifcultyisexperiencedcontactyour supplierforassistance.
DANGER! RiskofElectricShock
Thisunitcontainsmovingandhighpressure components.Therearenouserserviceableparts insidethecover .T oavoidelectricshock,injury ordeath:
–DONOTremovethecover. –Referservicingtoqualiedpersonnel.
WARNING! RiskofSeriousInjuryorDamage
Useofnon-Invacareaccessoriesmayresultin seriousinjuryordamage.
–Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunctionwith Invacareaccessories.Accessoriesdesignedby othermanufacturershavenotbeentestedby Invacareandarenotrecommendedforuse
withInvacareproducts. –DONOTusenon-Invacareaccessories. –ToobtainInvacareaccessories,contactInvacare
byphoneoratwww .invacare.com
WARNING! Riskofinjuryordamage
Blockingtheairowintoandoutofthe compressormaycauseinjuryordamage.
–NEVERblocktheairopeningsoftheproduct
orplaceitonasoftsurface,suchasabedor
couch,wheretheairopeningmaybeblocked. –Keeptheopeningsfreefromlint,hairand
similarforeignitems.
81100873-M
Safety
WARNING! RiskofInjuryorDamage
Adamagedproductmaymalfunctionandcause injuryordamage.T oavoidinjuryordamage:
–DONOTgetthecompressorwet. –Ifthecompressorhasadamagedcordorplug,
ifitisnotworkingproperly,ifithasbeen droppedordamaged,orsubmersedinwater , callaqualiedtechnicianforexaminationand repair .
WARNING!
–DONOTusethecompressoroutdoors.Indoor
useonly.
–Ifanyleakageoftheoxygencylinderisdetected
DONOTattempttousethecylinder.Turnthe owselectorknobontheregulatorto“Off”.If leakagepersists,placethecylinderoutdoors andnotifyyourHomeHealthCareProvideror ServiceRepresentativeofthiscondition.
–ChildrenshouldALWA YSbesupervisedaround
theHomeFillunit.Whenusedwiththe HomeFilltableorsimilarelevatedsupport surfaces,achildcouldpulloveranysuchtable causingtheunittofall.Failuretoobserve thiswarningmayresultinpersonalinjuryor productdamage.
WARNING! Riskofinjuryordamage
TheHomeFillisequippedwithahighpressure reliefvalvetoensuretheuser’ssafety .When activated,thissafetyfeaturewillmakean extremelyloudnoise.Toavoidinjuryordamage:
–Ifthisnoiseoccurs,turntheunitOffand
contactyourHomeHealthCareProvideror
ServiceRepresentative.
WARNING! Riskofinjuryordamage
Ifimproperlypositionedandsecured,thepower cordandinterconnecthosecancauseinjurydue totrippingandfalling.Productdamagemayalso occur.
–ThepowercordandinterconnecthoseMUST
beroutedandsecuredproperly. –Avoidpositioningpowercordsacrossareasof
hightrafc. –Beforemovingorrepositioningeitherthe
compressororconcentrator ,ALWA YSdisconnect
theACpowercordsandtheinterconnecthose
betweenthecompressorandconcentrator .
CAUTION!
Federal(statutory)lawrestrictsthisdevicesale toorontheorderofamedicalpractitioner licensedbyagovernmentalagencywherehe/she practices.
–ONLYalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
1100873-M9
Invacare®HomeFill®OxygenSystem
2.2LabelLocations
A
B
C
D
E
101100873-M
Safety
2.3DealerInformation
Invacarerecommendsleavingafullcylinderof oxygenwiththepatientaftersettinguptheHomeFill compressor .
BeforeleavingaHomeFillCompressorwithauser ,the followingchecklistMUSTbecompleted:
qMakesurethepressuretestoncylinderhasnotexpired. qInstructtheuseronthesafeoperationofthe
concentratorandreviewallwarnings.
qInstructtheuseronthesafeoperationoftheHomeFill
compressorandreviewallwarnings.
qLeaveafullcylinderofoxygenwiththeuser. qLeaveacopyofallwarningsandllinginstructionswith
theuser.
2.4Double-InsulatedProducts
CAUTION!
–DONOTunplugthepowercordbypullingon
thecord.T ounplug,grasptheplugnotthe cord,otherwisedamagemayoccur .
Inadouble-insulatedproduct,twosystemsofinsulation areprovidedinsteadofgrounding.Nogroundingmeansis requiredonadouble-insulatedproduct,norisameansfor groundingtobeaddedtotheproduct.Adouble-insulated productismarkedwiththewords“DOUBLEINSULATION”or “DOUBLEINSULATED”orthesymbol:
2.5EMCInformation
WARNING! RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamage frominterferencewithwirelessequipment:
–Keepthecompressoratleast9.8ft(3.0m)
awayfromwirelesscommunicationequipment
suchaswirelesshomenetworkdevices,mobile
phones,cordlessphonesandbasestations,
walkie-talkies,etc.
MedicalElectricalEquipmentneedstobeinstalledandused accordingtotheEMCinformationinthismanual.
Thisequipmenthasbeentestedandfoundtocomply withEMClimitsspeciedbyIEC/EN60601-1-2forType BFequipment.Theselimitsaredeterminedtoprovide reasonableprotectionagainstelectromagneticinterference inatypicalhomehealthcareenvironment.Thisequipment isnotintendedforuseinaprofessionalhealthcare environment.
PortableandmobileRFcommunicationsequipmentcan affecttheoperationofthisequipment.Thecompressor maybestackedontoponanInvacareHomeFillcompatible oxygenconcentratorduringsetup.Thecompressoroperation shouldbeobservedtoverifynormalcylinderllingineither situation.
Otherdevicesmayexperienceinterferencefromeven thelowlevelsofelectromagneticemissionspermitted bytheabovestandard.T odetermineiftheemissions fromthecompressorarecausingtheinterference,turn thecompressorOff.Iftheinterferencewiththeother
1100873-M
11