Invacare HomeFill II IOH200AW User Manual

Invacare®HomeFill®II
ModelIOH200AW
enCompressor
UserManual................................3
csKompresor
Návodkobsluze............................29
frCompresseur
Manueld'utilisation..........................57
Gebrauchsanweisung.........................87
itCompressore
Manualed’uso...............................117
plCompresor
Instrukcjaobsługi............................147
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
©2014Invacare®Corporation Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. WD–40and3–IN-ONEareregisteredtrademarksofWD-40Company. UpdatedversionsofthismanualareavailableonyourregionalInvacarewebsite.Foralistofregional Invacarewebsites,refertotheaddresseslistedattheendofthismanual.
Contents
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuture reference.
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................5
2Safety..........................................6
2.1GeneralGuidelines.............................6
2.2LabelLocation.................................9
2.3DealerInformation.............................11
2.4Disposal.....................................11
2.5EMCInformation...............................12
3Set-Up.........................................13
3.1Unpacking....................................13
3.2CompressorInitialSetupChecklist..................13
3.3ChoosingaLocation............................14
3.4TransportingtheCompressor.....................14
3.5InstallingthePowerCord.........................15
3.6ConnectingtheCompressortotheConcentrator.......15
4Usage..........................................16
4.1CompressorOperationChecklist...................16
4.2CylinderPrefillInspection.........................16
4.3TheHydrostaticTestingDate......................17
4.4ExternalExamination............................17
4.5Connecting/DisconnectingtheCylinderto/fromthe
Compressor.................................17
4.5.1ConnectingtheCylindertotheCompressor........17
4.6TurningtheCompressorOnandOff................19
5Maintenance....................................21
5.1MaintenanceInfo...............................21
5.2CleaningtheCompressorFilter....................21
5.3CleaningtheCabinet............................21
5.4ReplacingtheFuse..............................22
6Troubleshooting.................................23
6.1IndicatorLights................................23
6.2Troubleshooting...............................24
7TechnicalData..................................27
7.1TechnicalDescription...........................27
7.2CylinderFillTimes..............................27
7.3TypicalProductParameters.......................28
Invacare®HomeFill®II
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe practiceswhichcouldresultinpersonalinjuryorpropertydamage. Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage orminorinjuryorboth.
IMPORTANT
–Indicatesahazardoussituationthatcouldresultin
damagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformationfor efficient,trouble-freeuse.
ReadtheManual
AlternatingCurrent TypeBFequipment—AppliedParts
Powerswitch“ON”
Powerswitch“OFF”
DONOTsmoke
ClassII,DoubleInsulation
Keepdryintransportorstorageandduringuse
Fuse
Oxidizer
4
1145804-E
General
Non-flammablegas,class2type(oxygen)
EuropeanRepresentative
Manufacturer
TransportandStorageTemperature
TransportandStorageHumidity
ThisproductcomplieswithDirective93/42/EEC concerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheCE declarationofconformity.
1.2IntendedUse
TheintendeduseofanInvacareHomeFillcompressorisanaccessory devicetoanInvacareoxygenconcentratorthatenablesthepatient torefillcustomportableoxygencylindersforpersonaluse.
Thisproductisintendedforsinglepatientuseintheindoorhomecare environmentonlyandisnottobeusedforbulkcylinderrefilling.
Nospecificproductknowledgeortrainingisrequired,otherthan whatiscontainedinthismanual,tooperatetheproduct.
Itisnotintendedforsupplyingoxygentherapyduringrefilling.
Thisproductisintendedtobeusedbyorunderadultsupervision onlyafterreadingandunderstandingtheinstructionsandwarnings ofthisusermanual.
NOTICE
TheHomeFillIIcompressorisforusewithspecially designedInvacareoxygenconcentratorsthatare HomeFillcompatible.
TodetermineifyouroxygenconcentratorisHomeFillcompatible, examinethelowerrightcorneroftherearoftheoxygen concentratorAforanoutletfittingB.
Platinum—TopView
Perfecto
2
—RearView
1145804-E5
Invacare®HomeFill®II
2Safety
2.1GeneralGuidelines
WARNING!
Thissectioncontainsimportantinformationforthesafe operationanduseofthisproduct.
–Readandunderstandtheseinstructions.DONOTuse
thisproductwithoutreadingandunderstandingthese instructions.Ifyouneedhelpwiththeseinstructions, contactahealthcareprofessionaloranInvacare dealer.Otherwiseinjuryordamagemayoccur.
ACCESSORIESWARNING
–Invacareproductscanonlybeusedwithaccessories
madebyInvacare.AccessoriesnotmadebyInvacare havenotbeentestedbyInvacare.Accessoriesnot madebyInvacareshouldnotbeused.
DANGER! RiskofFire
–UsersMUSTNOTSMOKEwhileusingthisdevice.
Keepallmatches,lightedcigarettesorothersources ofignitionoutoftheroominwhichthisproductis located.Textilesandothermaterialsthatnormally wouldnotburnareeasilyignitedandburnwithgreat intensityinoxygenenrichedair.Failuretoobserve thiswarningcanresultinseverefire,propertydamage andcausephysicalinjuryorDEATH.
Checkallpartsforshippingdamage.Incaseofdamage,DO NOTuse.ContactInvacareforfurtherinstruction.
Inordertoensurethesafeinstallation,assemblyandoperationofthe HomeFillIICompressortheseinstructionsMUSTbefollowed.
CAUTION!
–Statutorylawcanrestrictthisdevicetosaleorrental
byoronorderofaphysician,oranyotherpractitioner licensedbythelawofthegovernmentagencyinwhich he/shepracticestouseorordertheuseofthisdevice.
WARNING!
–Theuseofthisdeviceislimitedtorefillingaspecific
patient’sownoxygencylinders.Thedesignated cylindersMUSTbeusedONLYbythespecificoxygen patientandarenottobedistributedtoanyother individualforanypurpose.
–Nootheruseofthisdeviceisindicated.
Themaximumoxygenconcentratorflowrateforpatientuse islistedinthetablebelow.Otherwise,theoxygenlevelto thecompressorwillbe<90%andthecompressorwillnot fillthecylinder.
InvacareOxygen Concentrator
MaximumOutputFlowfor PatientUse
Platinum5
2.5L/min Platinum95L/min Perfecto
2
3L/min
Patientswithahigherprescribedflowshouldnotrefilla cylinderwhilereceivingsupplementaloxygentreatment fromtheoxygenconcentrator.
61145804-E
Safety
WARNING!
–NEVERblocktheairopeningsoftheproductorplace
itonasoftsurface,suchasabedorcouch,wherethe airopeningmaybeblocked.Keeptheopeningsfree fromlint,hairandthelike.
DONOTuseoutdoors.Indooruseonly.NEVERusenearanytypeofflameor
flammable/explosivesubstances,vaporsor atmosphere.
–DONOTdisassemble.Nomodificationofthis
equipmentisallowed.Theunitisprovidedwith doubleinsulationandhasnointernaluserserviceable parts.Ifanyserviceisrequired,contactInvacare TechnicalService.
–Ifthecompressorhasadamagedcordorplug,ifit
isnotworkingproperly,ifithasbeendroppedor damaged,orsubmersedinwater,callaqualified technicianforexaminationifrepairisneeded.
–IfanyleakageoftheoxygencylinderisdetectedDO
NOTattempttousethecylinder.Turntheflow selectorknobontheregulatorto“Off”.Ifleakage persists,placethecylinderoutdoorsandnotifyyour HomeHealthCareProviderorServiceRepresentative ofthiscondition.
–ChildrenshouldALWAYSbesupervisedaroundthe
HomeFillIIunit.WhenusedwiththeHomeFilltable orsimilarelevatedsupportsurfaces,achildcouldpull overanysuchtablecausingtheunittofall.Failure toobservethiswarningmayresultinpersonalinjury orproductdamage.
WARNING!
–DONOTallowoilfromyourhandsorothersources
tocomeintocontactwiththecompressorcoupling orcylinderfittings.Thesesubstancesmaybecome flammableinthepresenceofoxygenandcauseinjury. Avoidtouchingthecylinderfittingwheneverpossible. Exampleofitemstoavoidarehandcreamsorlotions, cookingoils,suntanoilsorsunblocksandsimilaroily products.IfyouusethesetypesofproductsInvacare suggestsyouwashyourhandspriortousingthe HomeFillunit.Ifthecylinderfittingshouldcomein contactwithsuchsubstancesitshouldbecleaned offwithadampclothpriortoconnectingittothe compressor.
–NEVERoilorlubricatethecompressorcoupleror
cylinderfittings.NEVERuseanypenetratingoil orlubricantsuchasWD-40®or3-in-1Oil®.If connectiondifficultyisexperiencedverifythatthe cylinderfittingisbeinginsertedstraightdowninto thecompressorcoupler.Ifcontinueddifficultyis experiencedcontactyoursupplierforassistance.
–Keeptheoxygentubing,cord,andunitawayfrom
heatedorhotsurfaces,includingspaceheaters, blankets,stovesandsimilarelectricalappliances.
1145804-E7
Invacare®HomeFill®II
WARNING!
–TheHomeFillIIisequippedwithahighpressurerelief
valvetoensuretheuser’ssafety.Whenactivated,this safetyfeaturewillmakeanextremelyloudnoise.Ifthis noiseoccurs,turntheunitOffandcontactyourHome HealthCareProviderorServiceRepresentative.
–Beforemovingorrepositioningeitherthecompressor
orconcentrator,ALWAYSpushthepowerswitchto theOff(O)position,disconnecttheACpowercords andtheinterconnecthosebetweenthecompressor andconcentrator.Failuretodosomayresultin damagetotheunitorpersonalinjury.
81145804-E
Safety
2.2LabelLocation
1145804-E9
Invacare®HomeFill®II
A
SerialNumberLabel
B
SpecificationLabel
101145804-E
Safety
2.3DealerInformation
Invacarerecommendsleavingafullcylinderofoxygenwith thepatientaftersettinguptheHomeFillIIcompressor.
BeforeleavingaHomeFillIICompressorwithauser,thefollowing checklistMUSTbecompleted:
qMakesurethepressuretestoncylinderhasnotexpired. qChecktheEMCsectionforoperationalinformationandpossible
RFissues.Referto2.5EMCInformation,page12
.
qInstructtheuseronthesafeoperationoftheconcentratorand
reviewallwarnings.
qInstructtheuseronthesafeoperationoftheHomeFillII
compressorandreviewallwarnings.
qLeaveafullcylinderofoxygenwiththeuser. qLeaveacopyofallwarningsandfillinginstructionswiththeuser.
2.4Disposal
Recycle
Thisproducthasbeensuppliedfromanenvironmentallyaware manufacturerthatcomplieswiththeWasteElectricalandElectronic Equipment(WEEE)Directive2002/96/CE.
Thisproductmaycontainsubstancesthatcouldbeharmfultothe environmentifdisposedofinplaces(landfills)thatarenotappropriate accordingtolegislation.
The'crossedoutwheeliebin'symbolisplacedonthisproductto encourageyoutorecyclewhereverpossible.
Followlocalgoverningordinancesandrecyclingplansregarding disposalofthedeviceorcomponentsnormallyusedinoperation. Thedevicedoesnotgeneratewasteorresidueinoperation.DO NOTdisposeofthedeviceinthenormalwastestream.Any accessoriesnotpartofthedeviceMUSTbehandledinaccordance withtheindividualproductmarkingfordisposal.
Pleasebeenvironmentallyresponsibleandrecyclethisproductatits endoflife.ContactInvacareoryourserviceproviderforproduct recyclinginformation.RefertotheregionalInvacare.
1145804-E
11
Invacare®HomeFill®II
2.5EMCInformation
MedicalElectricalEquipmentneedstobeinstalledandusedaccording totheEMCinformationinthismanual.
ThisequipmenthasbeentestedandfoundtocomplywithEMC limitsspecifiedbyIEC/EN60601-1-2forTypeBFequipment. Theselimitsaredeterminedtoprovidereasonableprotection againstelectromagneticinterferenceinatypicalhomehealthcare environment.Thisequipmentisnotintendedforuseinaprofessional healthcareenvironment.
PortableandmobileRFcommunicationsequipmentcanaffectthe operationofthisequipment.Thecompressormaybestackedontop onanInvacareHomeFillcompatibleoxygenconcentratorduring setup.Thecompressoroperationshouldbeobservedtoverify normalcylinderfillingineithersituation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels ofelectromagneticemissionspermittedbytheabovestandard.To determineiftheemissionsfromthecompressorarecausingthe interference,turnthecompressorOff.Iftheinterferencewiththe otherdeviceoperationstops,thenthecompressoriscausingthe interference.Insuchrarecases,interferencemaybereducedor correctedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetweenthe
devices.
Connecteitherlinepowereddevicetoadifferentelectrical
powercircuit.
ContactInvacareforanyadditionalEMCinformationorEMCtables forthedeviceenvironment.
12
1145804-E
Set-Up
3Set-Up
3.1Unpacking
PARTNUMBER
DESCRIPTION
1116440CompressorAssembly 1102315PackagingAssembly 1105177SupplyLineAssembly 1131308PowerCord 1145804UserManual(notshown)
1.Checkforanyobviousdamagetothecartonoritscontents.If damageisevident,notifythecarrier,oryourlocaldealer.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallofthecomponentsfromthecarton.
UnlessthecompressoristobeusedIMMEDIATELY,leave thecompressorinitspackagingforstorageuntiluseofthe compressorisrequired.
3.2CompressorInitialSetupChecklist
DealersONLY—Referto2.3DealerInformation,page11 beforeperformingthisprocedure.
UsersONLY—ContactyourInvacaredealerorInvacare forassistancewiththesetupprocedureandtoreport unexpectedoperationorevents.
BeforeoperatingtheHomeFillIICompressorforthefirsttime, completethefollowingchecklist:
qChoosealocationforthecompressor. qTransportthecompressortothedesiredlocation. qConnectthepowercordtothecompressor. qConnectthecompressortotheconcentrator.
1145804-E13
Invacare®HomeFill®II
3.3ChoosingaLocation
WARNING!
–NEVERusenearanytypeofflameor
flammable/explosivesubstances,vaporsor atmosphere.
–Keeptheoxygentubing,cord,andunitawayfrom
heatedorhotsurfaces,includingspaceheaters, blankets,stovesandsimilarelectricalappliances.
Thecompressorshouldbeatleastthreeinchesawayfromwalls, draperiesorfurnituretoensuresufficientairflowandclearancefor thepowercord.
DONOTuseinacloset. Thecompressorshouldsitonahardflatstablesurface,suchasa
tableortheReadyRack.
TheHomeFillReadyRackissoldseparatelyandusedto positiontheHomeFillcompressorontopofaPerfecto
2
oxygenconcentrator.RefertotheHomeFillReady Rackinstructionsheetpartnumber1148363,formore information.
3.4TransportingtheCompressor
WARNING!
–NEVERtransportacompressorwithacylinder
connectedtoorstoredonthecompressor,otherwise injuryordamagecanoccur.
CAUTION!
–Whentransportingthecompressor,becarefulnot
tobumpthecompressorortheconnectorsinto obstacles.Otherwise,damagetothecompressormay occur.
Therearetwomethodstosafelytransportthecompressor:
1.Onehand-GraspthehandgripAatthebottomrearofthe compressor.
2.Twohands-GraspthehandgripsBonthesidesofcompressor.
14
1145804-E
Set-Up
3.5InstallingthePowerCord
1.EnsurethecompressorAisonahardflatstablesurface,suchas atableorReadyRack.
2.MakesurethepowerswitchisintheOff(O)position.
3.Plugthepowercord(Invacaresupplied)intotheelectrical connectorBinthebackofthecompressor.
4.PlugtheoppositeendoftheInvacaresuppliedpowercordD intothewalloutletC.
5.Leaveaminimumof3”(7.6cm)clearancetothepowercord endsfordisconnectionaccess.
Thepowerinletactsasthemainsdisconnectiondevice toremoveallpowertotheunitwhenthelinecordis disconnected.
3.6ConnectingtheCompressortothe Concentrator
1.InsertoneendBoftheinterconnecthoseA(InvacareSupplied) intotheoutletfittingBonthebackoftheconcentratorC.
2.Inserttheoppositeendoftheinterconnecthoseintotheinlet fittingDonthesideofthecompressorE.
ThemetaltabsGontheconcentratoroutletfitting andthecompressorinletfittingFwillpopoutwithan audible“click”whentheendoftheinterconnecthose isproperlyinstalled.
3.Loopanyexcessinterconnecthoseandsecuretothebackof theconcentratorwiththefasteningstraps.
1145804-E15
Invacare®HomeFill®II
4Usage
4.1CompressorOperationChecklist
EachtimetheHomeFillIICompressorisusedtofillacylinder, completethefollowingchecklist:
qEnsuretheconcentratorhasbeenOnforatleast30minutes.
RefertotheconcentratorUserManual.
qPerformtheprefillinspectiononthecylinder. qConnectthecylindertothecompressor. qPushthecompressorpowerswitchtotheOn(
)position.
qExaminetheindicatorlightsonthecontrolpanel. qDisconnectandremovethefullcylinder. qPushthecompressorpowerswitchtotheOff(O)position. qIffillinganothercylinder,repeatthischecklist.
4.2CylinderPrefillInspection
WARNING!
ONLYusecylindersthathavetheInvacareHomeFillIIconnectionandthecylinderlabel(shownforreference).AllcylindersMUSTbeinspectedbeforeattemptingtofill.Otherwise,injuryordamagemayoccur.
161145804-E
Usage
4.3TheHydrostaticTestingDate
AluminumcylindersMUSTundergohydrostatictestingeveryfive years.Contactyourdealerformoreinformation.
WARNING!
–DONOTfillcylindersthathavenotbeentestedin
thepastfiveyears.Otherwise,seriouspersonalinjury mayresult.Contactyourdealerforreplacement.
4.4ExternalExamination
1.Examinetheoutsideofthecylinderforthefollowingconditions, andreplacethecylinderiftheyexist:
Dentsordings
Arcburns
Oilorgrease
Anyothersignsofdamagethatmightcauseacylindertobe
unacceptableorunsafeforuse.
2.Examinethecylinderforevidenceoffireorthermaldamage. Evidenceincludescharringorblisteringofthepaint,orother protectivecoatingorheatsensitiveindicator.Iffireorthermal damageisfound,replacethecylinder.
3.Inspectthecylinder/regulatorassemblyforthefollowing,and replaceiffound:
Debris,oilorgrease
Noticeablesignsofdamage
Signsofcorrosioninsidethevalve
Signsofexcessiveheatorfiredamage
4.5Connecting/DisconnectingtheCylinder to/fromtheCompressor
4.5.1ConnectingtheCylindertotheCompressor
WARNING! RiskofInjuryorDamage
–NEVERusetoolsofanykindtoconnect/disconnect
thecylinderAandthecompressorB.Otherwise, severeinjuryand/ordamagemayoccur.
–DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise, injuryordamagemayoccur.
–DONOTusealiquidleakdetectortotestforleaks.
1.ExaminethecylindergaugeC.Ifthecylinderpressureisless than1500psig(10342kPa),proceedtostep2.
Ifthecylinderpressureisgreaterthan1500psig(10342 kPa),DoNOTattempttotopoffthiscylinder.Itmay notfill.
2.SetregulatorflowdialDonthecylinderAtoOff.
1145804-E17
Invacare®HomeFill®II
3.Removethecylinderfillportcover EandconnectorfillportcoverF.
CAUTION!
–DONOTconnecttheoutlet(cannula)G
connectiontothecompressor,otherwisethe conservingcylinderwillnotworkproperly.
4.MomentarilypushDOWNontheouterring(sleeve)Hofthe connectorfillportIuntilGREENdotsJarevisibletoreset theconnector.
Iftheouterring(sleeve)isintheUPposition(GREEN dotsnotvisible),theconnectorfillportwillnotbe abletoacceptthecylinderfillport.PushingDOWN momentarilywillresettheconnectorfillport(GREEN dotsvisible)toacceptthecylinderfillport.
5.Graspthecylinder/regulatorKassemblyintheareabehindthe cylindergaugeL.
6.PositionthecylinderinthecompressorcradleM.
7.AlignthecylinderfillportNwiththeconnectorfillportI.
8.PullUPontheouterring(sleeve)oftheconnectorfillportwhile pushingDOWNonthecylinder/regulatorassemblytocouple thecylinderfillportintotheconnectorfillport.
Thecylinderisproperlyconnectedwhenanaudible “click”isheard.
DisconnectingtheCylinderfromtheCompressor
WARNING!
–NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe injuryand/ordamagemayoccur.
1.Whilestandinginfrontoftheunit,pressthecompressorpower switchtotheOff(O)Oposition.
2.Graspthecylinder/regulatorassemblyKintheareabehindthe cylindergaugeL.
3.Withtheotherhand,grasptheouterring(sleeve)Hofthe connectorfillportIandpushDOWN.
181145804-E
Usage
4.Liftuponthecylinder/regulatorassemblytoremovefromthe connectorfillport.
5.Whenthecylinderfillportisdisconnectedfromtheconnector fillport,releasetheouterring(sleeve)oftheconnectorfillport andusetwohandstoremovethecylinderfromthecompressor cradle.
WARNING!
–Thefillportcoverontheconnectorfillportand
thecylinderfillportMUSTbereplacedafter fillingandwhenevernotinuse.Ifeitherofthe fillportcoversaremissingcontactyourhealthcare providerbeforeusing.
6.Placethefillportcoversontothecon­nectorfillportFandcylinderfillportE.
7.Examinethecylindergaugetoensurecylinderisfull(needle pointingintoGREENarea).Ifthecylinderisnotfull,perform thefollowing:
Repeat4.5.1ConnectingtheCylindertotheCompressor,
page17
.
Ifcylinderisnotfullaftersecondattempttofill,contact
theprovider.
4.6TurningtheCompressorOnandOff
1145804-E19
Invacare®HomeFill®II
1.MakesuretheconcentratorisOn.Refertotheconcentrator UserManual.
Themaximumoxygenconcentratorflowratefor patientuseislistedinthetablebelow.Otherwise,the oxygenleveltothecompressorwillbe<90%andthe compressorwillnotfillthecylinder.
InvacareOxygen Concentrator
MaximumOutputFlow forPatientUse
Platinum5
2.5L/min Platinum95L/min Perfecto
2
3L/min
Patientswithahigherprescribedflowshouldnotrefilla cylinderwhilereceivingsupplementaloxygentreatment fromtheoxygenconcentrator.
2.Settheconcentratorflowratetothepatientto2.5L/min.orless forPlatinum5and5L/min.forPlatinum9oxygenconcentrator. RefertotheconcentratorUserManual.
3.Makesurethecylinderisconnectedtothecompressor.Refer to4.5.1ConnectingtheCylindertotheCompressor,page17.
4.PushthepowerswitchBonthecontrolpanelAtotheOn ()position.
5.Examinethecontrolpanel.Thefollowingsequenceofevents shouldoccur:
a.0-3minutes-TheO
2
belownormal(YELLOW)lightCwill
belitwhilethecompressorwarmsup.
b.Afterthreeminutes-Thefilling(green)lightDwillbeon
whilethecylinderisfilling.
c.Thefull(green)lightEwillbeonwhenthecylinderis
finishedfilling.Proceedtostep6.
TheO
2
belownormal(YELLOW)lightmaycomeon iftheoxygenlevelfromtheconcentratorhasnotyet reachedordropsbelow90%orhigher.Thecompressor willbeginorresumefillingwhentheoxygenlevel reaches90%.Referto7.2CylinderFillTimes,page27
if
lightdoesnotgoOffwithin10minutes. Referto7.2CylinderFillTimes,page27
iftheAttention
(red)lightFandaudiblealarmareon.
6.Removethefullcylinder.RefertoDisconnectingtheCylinder fromtheCompressor,page18.
7.PushthepowerswitchtotheOff(O)position.
Forcylinderoperationanduse,refertothespecificuser manualforthecylinder/regulatorassembly.
201145804-E
Maintenance
5Maintenance
5.1MaintenanceInfo
WARNING!
–Onlythepartsandproceduresidentifiedinthis
sectionaretobeperformedbytheoperator.All otherpartsandproceduresaretobeperformedbya qualifiedtechnician.
5.2CleaningtheCompressorFilter
WARNING!
–PushthepowerswitchtotheOff(O)positionand
unplugthecompressorbeforecleaning.
–DONOToperatethecompressorwithoutthefilter
installed.
CAUTION!
–DONOTunplugbypullingoncord,otherwise
damagemayoccur.
Allcompressorshaveonefilteronthebackofthecabinet. Performthisprocedureatleastonceaweek.
1.Unplugthecompressor.Grasptheplugnotthecord.
2.RemovethefilterA.
3.Cleanthefilterwithavacuumcleanerorwashinwarmsoapy waterandrinsethoroughly.
4.Drythefilterthoroughlybeforereinstallation.
5.3CleaningtheCabinet
WARNING!
–PushthepowerswitchtotheOff(O)positionand
unplugthecompressorbeforecleaning.Toavoid electricalshock,DONOTremovecabinet.
Cleanthecabinetwithamildhouseholdcleanerandnon-abrasive clothorsponge.
1145804-E
21
Invacare®HomeFill®II
5.4ReplacingtheFuse1.Unplugthepowercordfromtheelectricaloutlet.
2.RemovethepowercordfromthepowerinletAonthe compressorbaseC.
3.PulloutthefusedrawerBatthebottomofthepowerinlet.
4.Usingafusepullerorasmallflatbladescrewdriver,removethe frontfuseinsidethefusedraweranddiscard.
Thefusedrawercontainsspacefortwofuses.The frontlocationisfortheoperationalfuse.Theunitis suppliedwiththefrontfuseonly.Therearlocationis foranoptionalsparefuse.Whenreplacingthefront fuse,ensurethatthereisasparefusewiththesame typeandratinginthefusedrawer.
5.InstallreplacementfuseinfrontfuseDslot.
Usea5mmx20mm,timelag,lowbreakingcapacity, 250V,2Amp,IEC60127styleonly(Invacarepart number1099820).
6.Pushfusedrawerbackintothepowerinlet.
7.Installpowercordintothepowerinletonthecompressorbase.
8.Plugpowercordintoelectricaloutlet.
9.TurnthepowerswitchtotheOn()positionandchecktosee thattheunitpowersupproperly.
10.Iftheunitdoesnotcomeon,turnthepowerswitchoff,unplug thepowercord,andcallyourdealer,providerorInvacarefor service.
22
1145804-E
Troubleshooting
6Troubleshooting
6.1IndicatorLights
Allindicatorsareforlowpriorityconditionsthatrequireoperatorawarenessonly.
INDICATORLIGHTCOLORCOMPRESSORSTATUSEFFECT
None
CompressorisOff.O
2
flowingtouserifconcentratorisOn.
Cylinderfillingnotstarted. CompressorisOnandwarmingup
(0-3minutesafterpowerisOn).
O
2
flowingtouserifconcentratorisOn.
Cylinderfillingstops,willrestartwhenO
2
>90%.
O
2
BelowNormal-
YELLOW
O
2
fromconcentratoris<90%(anytimeafter
threeminutewarmup).
O
2
flowingtouserifconcentratorisOn.
Cylinderisfilling Filling-GREENCompressorisOnandcylinderisfilling.O
2
fromconcentratoris>90%(anytimeafter threeminutewarmup).
O
2
flowingtouserifconcentratorisOn.
1145804-E23
Invacare®HomeFill®II
INDICATORLIGHTCOLORCOMPRESSORSTATUSEFFECT
Full-GREENCompressorisOn.Cylinderisfinishedfilling
(full)-removecylinder(FULLlightisOnin >10mins.)
or Cylinderisnotattachedproperly-check
connection(FULLlightisOnin<10mins.).
Cylinderfillingstops Cannotfillanothercylinderuntilcompressor
isturnedOffandOnagain O
2
flowingtouserifconcentratorisOn.
ATTENTION-RED(AudibleTone)CompressorisOn.CylinderNOTfilling
Causes: FullcylinderattachedwhenunitturnedOn-
replacewithemptycylinder or Internalfailure-contactdealer.
Compressordoesnotbegintofillorstops filling UnitMUSTbeturnedOffandOnbefore fillingcanoccur. O
2
flowingtouserifconcentratorisOn.
6.2Troubleshooting
PROBLEMSOLUTION
Noindicatorlightsonwhencompressor turnedon.
1.TurncompressorOff(O).
2.Checkthatpowercordisproperlypluggedintocompressorandwalloutlet.
3.TurnthecompressorOn(
).
4.Cleanairfilters.Referto5.2CleaningtheCompressorFilter,page21.
5.IfperformingSteps1-4doesnotcauseanyindicatorlightstocomeon,contactyourhome healthcareproviderorInvacareforservice.
REDlightOn1.Checkthecylindergaugetoseeifitisfull.Cylinderfull-Removethecylinderandturntheunit
Off(O).Performthe4.2CylinderPrefillInspection,page16withacylinderthatisnotfull. Cylindernotfull-Proceedtostep2.
24
1145804-E
Troubleshooting
PROBLEMSOLUTION
2.Checkthatthecylinderisconnectedproperly.TurnthecompressorOff(O).Removethe cylinderandreinstall.TurnthecompressorOn(
).
3.Ifcylinderisnotfullandisconnectedproperly,thecompressormayhaveinternalfailure.
ContactyourhomehealthcareproviderorInvacare. CompressorcyclesOnandOff or O
2
BelowNormal(yellow)andFilling
(green)lightsarealternating or O
2
BelowNormal(yellow)lightgoesOn
andOff
1.Checktheconcentratorliterflowsettingisat2.5L/min.orless.Ifnecessary,adjusttheliterflow
setting(refertotheconcentratorusermanual).Waitatleast20minutesfortheGREENlightto
appear.Ifitdoesnotappear,proceedtostep2.
2.Theconcentratormayneedservice,contactyourhomehealthcareproviderorInvacare.
O
2
BelowNormal(YELLOW)lightstays
On.
1.Ensuretheconcentratorhaswarmedupforatleast20minutes.Ifithasnotbeen20minutes,
turnthecompressorOff(O)untiltheconcentratorwarmsup.IftheYELLOWlightstillremains
onproceedtostep2.
2.Inspecttheconnectionbetweenthecompressorandtheconcentratorfordamageorpinching.If
damaged,replace.Ifpinched,straightentubing.TurnthecompressorOff(O)foratleast10seconds.
TurnthecompressorOn().IftheYELLOWlightstillremainsonafter10minutes,proceedtostep3.
3.Theconcentratormayneedservice,contactyourhomehealthcareproviderorInvacare.
1145804-E25
Invacare®HomeFill®II
PROBLEMSOLUTION
Filltimesexceedmaximumrecommended filltimesorGREENfillinglightstaysOn.
1.EnsurethecylinderregulatorissettoOff.
2.Checktherecommendedfilltimesshownin7.2CylinderFillTimes,page27
.
3.Checkthatthecylinderisconnectedproperly.TurnthecompressorOff(O).Removethe
cylinderandreinstall.TurnthecompressorOn(
).
4.Checkthecompressor/concentratorlineforproperconnectionand/ordamage(i.e.-leaks
orkinks).
5.Checktheconcentratorliterflowsettingisat3L/min.orless.Ifnecessary,adjusttheliter
flowsetting(refertotheconcentratorusermanual).IftheGREENfillinglightdoesnotgoOff,
thecompressorand/orconcentratormayneedservice.Contactyourhomehealthcareprovider
orInvacare. ExcessivenoisewhencompressorisOn.1.CallhomehealthcareproviderorInvacareforservice. HissingsoundfromRegulatorwhenflow
selectorisOff.
1.TurnflowselectorOff.
2.Ifhissingsoundscontinue,callhomehealthcareproviderorInvacare.
RegulatorisOnbutO
2
isnotbeing
delivered.
1.Refillcylinder
2.Checkthatcannulaisstraightandnotpinched.Ifdamaged,replace.
3.CallhomehealthcareproviderorInvacareforservice.
261145804-E
TechnicalData
7TechnicalData
7.1TechnicalDescription
TheHomeFillIIOxygenSystem(IOH200AW)isan electro-mechanicalprescriptionsystemforuseinthehome.Its intendedfunctionanduseistosupplypressurizedoxygentofill Invacareproprietarygascylindersforambulatoryusewhilethe patientreceivestherapeuticoxygenfromanoxygenconcentrator. TheIOH200AWgassystemisdesignedforusewithanInvacare oxygenconcentratorwithanintegralHF2accessoryportthatcan supplyagasconcentrationgreaterthan90%O
2
around2L/min.
whilesupplyingapatienttherapyflowrate.
Themaximumoxygenconcentratorflowrateforpatientuse islistedinthetablebelow.Otherwise,theoxygenlevelto thecompressorwillbe<90%andthecompressorwillnot fillthecylinder.
InvacareOxygen Concentrator
MaximumOutputFlowfor PatientUse
Platinum5
2.5L/min Platinum95L/min Perfecto
2
3L/min
TheHomeFillIIOxygensystemconsistsofIOH200AWgas compressor,connectiontubing,andtheInvacaresuppliedportable gascylinderswiththeintegralflowregulatorandinterlockingfill fitting.
TheIOH200AWcompressorconsistsofamultistagecompressor, internalreservoir,controlelectronicswithanintegraloxygen concentrationsensor,interlockingfillfitting,inletfitting,anddrive
motor.Thecompressoroperationiscontrolledbytheelectronics assembly.Themotor/compressorisallowedtofillacylinderonly whentheoxygenconcentrationoftheinletgasstreamisgreater than90%andthecylinderpressureisnotalreadygreaterthan1500 PSI(10342kPa).Themotor/compressorfilloperationiscompleted whenthegasinacylinderreachesthemaximumfillpressureof2000 PSI(13786kPa).
TheIOH200AWcompressorcanfillcylindersforportableuseby ambulatorypatients.TheInvacaresuppliedportablegascylinders haveaninterlockingfillfittingthatmatesonlywiththeInvacare IOH200AWcompressorassembly.Thegascylinderalsocontainsan integralflowregulatorwithfixed,selectableflowsettings.
7.2CylinderFillTimes
CYLINDER FILLING TIMES
ConcentratorFlowRatetoPatient:
(Platinum5upto2.5L/min.and Platinum9upto5L/min.) *HF2PC9*HF2P-
CL4
*HF2P­CL6
HF2PCE8
NORMAL
2hr20 min
60min1hr25
min
2hr5min
MAXIMUM2hr45
min
1hr20 min
1hr50 min
N/A
Allfillingtimesareapproximateandmayvaryon environmentalconditions.
*NotforEUcountries.
1145804-E27
Invacare®HomeFill®II
7.3TypicalProductParameters
RegulatoryListing:EN/IEC60601–1
EN/IEC60601–1–2
OperatingEnvironmental Conditions:
41to104°F(5to40°C)at 15to60%non-condensing humidity
AtmosphericPressure:700to1060hPa StorageandTransport:
-25to158°F(-23to70°C)
at15to95%non-condensing humidity
InputPressureRequired:14-21PSI(96.5-144.7kPa) InputFlowRequired:2L/min InputO2%Required:
>90%O
2
Width:20¼inches(51.44cm) Height:15inches(38.1cm) Depth:
16inches(40.64cm)
Weight:33lbs(14.97kg) ShippingWeight:
39lbs(17.69kg)
MaxAppliedPart Temperature:
122°F(50°C)
ExpectedServiceLife:
Theexpectedservicelifeof thisproductis5yearsor 5000hoursofoperationwhen usedinaccordancewiththe safetyinstructions,maintenance intervalsandcorrectuse,stated inthismanual.Theeffective servicelifecanvaryaccording tofrequencyandintensityof use.Maintenanceshouldbe doneatleasteveryweek.Refer toMaintenancesection.
ElectricalRatings:230V~,1.0A,50Hz
281145804-E
Obsah
TentonávodkpoužitíMUSÍbýtposkytnutkaždémuuživatelitohoto výrobku.PEDpouitmvrobkusinvodprostudujteapothouschovejte probudoucpouit.
1Všeobecně......................................30
1.1Vysvětlenísymbolů.............................30
1.2Účelpoužití...................................31
2Bezpečnost.....................................32
2.1Všeobecnépokyny..............................32
2.2Umístěníštítku................................35
2.3Informaceproprodejce..........................37
2.4Likvidace.....................................37
2.5Informaceoelektromagnetickékompatibilitě...........38
3Nastavení.......................................39
3.1Vybalení.....................................39
3.2Kontrolníseznampředuvedenímkompresorudo
provozu....................................39
3.3Výběrumístění................................40
3.4Přenášeníkompresoru...........................40
3.5Připojenínapájecíšňůry..........................41
3.6Připojeníkompresorukekoncentrátoru..............41
4Použití.........................................42
4.1Kontrolníseznamprovozukompresoru..............42
4.2Prohlídkaválcepřednaplněním.....................42
4.3Datumhydrostatickéhotestu......................43
4.4Vnějšíprohlídka................................43
4.5Připojeníválcekekompresoru/odpojeníválceod
kompresoru.................................43
4.5.1Připojeníválcekekompresoru...................43
4.6Zapnutíavypnutíkompresoru.....................46
5Údržba.........................................48
5.1Informaceoúdržbě.............................48
5.2Čištěnífiltrukompresoru.........................48
5.3Čištěníkrytu..................................48
5.4Výměnapojistky................................49
6Řešeníproblémů.................................50
6.1Kontrolky....................................50
6.2Odstraňovánípotíží.............................51
7Technickéúdaje.................................55
7.1Technickýpopis................................55
7.2Dobynaplňováníválců...........................55
7.3Obvykléparametryvýrobku.......................56
Invacare®HomeFill®II
1Všeobecně
1.1Vysvětlenísymbolů
Vtomtonávodujsoupoužitasignálníslovavztahujícíse knebezpečnýmsituacímnebopostupům,kterébymohlyzapříčinit úrazosobnebovznikškodynamajetku.Definicetěchtosignálních slovjsouuvedenyníže.
NEBEZPEČÍ!
–Nebezpečíoznačujebezprostředněhrozící
nebezpečnousituaci,kterávpřípaděvznikupovede kvážnémunebosmrtelnémuúrazu.
VAROVÁNÍ!
–Varováníoznačujepotenciálněnebezpečnousituaci,
kterábyvpřípaděvznikumohlazapříčinitvážnýnebo smrtelnýúraz.
UPOZORNĚNÍ!
–Upozorněníoznačujepotenciálněnebezpečnousituaci,
kterábyvpřípaděvznikumohlazapříčinitlehčíúraz a/neboškodynamajetku.
DŮLEŽITÉ
–Označujenebezpečnousituaci,vjejímždůsledkuby
mohlodojítkpoškozenímajetku.
Poskytujeužitečnétipyadoporučenízaručujícíúčelné abezproblémovépoužívánívýrobku.
Přečtětesipříručku.
Střídavýproud VybavenítypuBF–použitédíly
VypínačvpolozeZapnuto
VypínačvpolozeVypnuto
Nekouřit
TřídaII,dvojitáizolace
Připřepravě,skladováníapoužitíudržujtevsuchu
Pojistka
Okysličovadlo
301145804-E