Invacare HomeFill HF2ECE9, HomeFill HF2ECE6 User Manual

Invacare®HomeFill®System
HF2ECE9,HF2ECE6
ENOxygenConservingCylinders
UserManual................................3
FRBouteilledeconservationd’oxygène
Manueld'utilisation..........................23
DESauerstoffAufbewahrungs-Zylinder
Gebrauchsanweisung.........................45
©2013Invacare®Corporation Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. WD-40and3–IN-ONEareregisteredtrademarksoftheWD-40Company.
Contents
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuture reference.
1General......................................4
1.1IntendedUse..............................4
1.2Features.................................4
2Safety........................................5
2.1Symbols..................................5
2.2GeneralGuidelines..........................5
2.3Disposal.................................7
3Setup........................................8
3.1ReceivingInspection.........................8
3.2CylinderPrefillInspection.....................8
3.3TheHydrostaticTestingDate..................9
3.4ExternalExamination........................9
4Usage........................................11
4.1Connecting/DisconnectingtheCylinderto/from
theCompressor...........................11
ConnectingtheCylindertothe
Compressor............................11
DisconnectingtheCylinderfromthe
Compressor............................13
4.2CylinderOperation.........................14
5Maintenance..................................17
5.1Maintenance..............................17
5.2ReplacementParts—Accessories..............17
6Troubleshooting...............................18
6.1Troubleshooting...........................18
7TechnicalData................................19
7.1TypicalProductParameters...................19
HomeFill®System

1General

1.1IntendedUse

Thiscylinderisintendedtostore,regulate,anddeliversupplemental oxygentopatientswhomayhavedifficultyextractingoxygenfrom theairtheybreathe.Itisforpatientswhowouldnormallyreceivethe oxygenviaanasalcannula.ThedevicedeliversUSP93±3%oxygenat anumberofdifferentflowsettings.Itisintendedtobeusedasan oxygendeliverydeviceandmayincreasethedryingoutofairways.

1.2Features

4
1176148-A
Safety

2Safety

2.1Symbols

Signalwordsareusedinthismanualandapplytohazardsorunsafe practiceswhichcouldresultinpersonalinjuryorpropertydamage. Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor efficient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC concerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheEC declarationofconformity.

2.2GeneralGuidelines

WARNING!
–Thissectioncontainsimportantinformationforthe
safeoperationanduseofthisproduct.
–DONOTUSETHISPRODUCTORANYAVAILABLE
OPTIONALEQUIPMENTWITHOUTFIRST COMPLETELYREADINGANDUNDERSTANDING THESEINSTRUCTIONSANDANYADDITIONAL INSTRUCTIONALMATERIALSUCHASUSER MANUALS,SERVICEMANUALSORINSTRUCTION SHEETSSUPPLIEDWITHTHISPRODUCTOR OPTIONALEQUIPMENT.IFYOUAREUNABLE TOUNDERSTANDTHEWARNINGS,CAUTIONS ORINSTRUCTIONS,CONTACTAHEALTHCARE PROFESSIONAL,DEALERORTECHNICAL PERSONNELBEFOREATTEMPTINGTOUSETHIS EQUIPMENT-OTHERWISE,INJURYORDAMAGE MAYOCCUR.
–GENERALGUIDELINEScontainsimportant
informationforthesafeoperationanduseofthis product.
ACCESSORIESWARNING
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare accessories.Accessoriesdesignedbyother manufacturershavenotbeentestedbyInvacareand arenotrecommendedforusewithInvacareproducts.
1176148-A5
HomeFill®System
DANGER!
–Thisproductisnotintendedasalife-sustainingor
life-supportingdevice.
–DONOTSMOKEwhileusingthisdevice.Keepall
matches,lightedcigarettesorothersourcesofignition outoftheroominwhichthisproductislocatedand awayfromwhereoxygenisbeingdelivered.
–NOSMOKINGsignsshouldbeprominentlydisplayed.
Textilesandothermaterialsthatnormallywouldnot burnareeasilyignitedandburnwithgreatintensityin oxygenenrichedair.Failuretoobservethiswarning canresultinseverefire,propertydamageandcause physicalinjuryordeath.
–Theinformationcontainedinthisdocumentissubject
tochangewithoutnotice.
OperatingInformation
CylinderMUSTbeemptiedofgaspriortoshippingbycourieror postalservice.
OnlypersonnelinstructedandtrainedintheuseoftheHomeFill ConservingCylindershouldoperatethisproduct.
Productcontainsmagnetic,ferrousmaterialthatmayaffectthe resultsofanMRI.
Besureallconnectionsaretightandleakfree. DONOTusealiquidleakdetectortotestforleaks. DONOTautoclave. DONOTgassterilizewithethyleneoxide.
DONOTcleanwitharomatichydrocarbons,including,butnot limitedtobenzeneandtoluene.
DONOTattempttorepairproduct. DONOTimmerseproductinanykindofliquid. Storeproductinacleanareawhennotinuse. OnlyuseInvacareCorporationcarryingbag. NEVERsmokeinanareawhereoxygenisbeingadministered. ALWAYSconfirmprescribeddosebeforeadministeringtopatient
andmonitoronafrequentbasis. DONOTuseifdirtorcontaminantsarepresentonoraround
cylinder,valve,productorconnectingdevices. NEVERusenearanytypeofflameorflammable/explosivesubstances,
vapoursoratmosphere. NOOXYGENisdeliveredwhenthepointerisalignedwith“O”. DONOTblocktheoutletfittingorkinkthecannulawhenthe
productisinuse,asthismaydamagetheunit. Theuseofthisdeviceislimitedtotheoxygenpatient.Cylinders
MUSTbeusedonlybytheoxygenpatientandarenottobe distributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathethrough theirmouths.
DONOTallowoilfromyourhandsorothersourcestocome intocontactwiththecompressorcouplingorbottlefittings.These substancesmaybecomeflammableinthepresenceofoxygenand causeinjury.Avoidtouchingthebottlefittingwheneverpossible. Exampleofitemstoavoidarehandcreamsorlotions,cookingoils, suntanoilsorsunblocksandsimilaroilyproducts.Ifyouusethese
61176148-A
Safety
typesofproductsInvacaresuggestsyouwashyourhandspriorto usingtheHomeFillunit.Ifthebottlefittingshouldcomeincontact withsuchsubstancesitshouldbecleanedoffwithadryclothprior toconnectingittothecompressor.
NEVERoilorlubricatethecompressorcouplerorbottlefittings. NEVERuseanypenetratingoilorlubricantsuchasWD-40®or 3-IN-ONE®.Ifconnectiondifficultyisexperienced,verifythatthe bottlefittingisbeinginsertedstraightdownintothecompressor coupler.Ifcontinueddifficultyisexperienced,contactyoursupplier forassistance.
DONOTuseahumidifierwiththisproduct. DONOTallowcylinderstotiporfall.Securegascylinderssothey
cannotfall.Ensurecylinderisalwaysinuprightposition. DONOTstorecylindersnearsourcesofheatorflame. Thisproductisequippedwithareliefvalve.Ifyouhearaloudhissing
orpoppingsoundcomingfromtheproduct,discontinueuseof product,turntheflowselectortothe“O”position,andcontactyour dealer/HomeCareequipmentprovider.
DONOTusewhilesleepingwithoutconsultingyourhealthcare provider.
Thisproductisdesignedtooperatewithastandardorhighflow duallumen,adultoxygennasalcannulawithamaximumlengthof
2.1m(7ft). Thecannulaisforsinglepatientuseonly. CannulaMUSTbeinstalledonoutletconnectionpriortouse. DONOTplacetheunitunderclothingwhileinuse.Whenthis
productisinuseasmallamountofoxygenisvented.Wearingthe unitunderclothingmaysaturatefabricswithoxygenandcausethem
toburnrapidlyifexposedtosparksorflame.Itmaytakeseveral hoursforoxygenlevelsinfabricstoreturntonormal.
DONOTuseHomeFillcompressorunit,tofillcylinderoutdoors oranywherebelow13°C(55°F).
NEVERattempttofillcylinderwithanythingotherthantheHomeFill compressor.

2.3Disposal

Thisproducthasbeensuppliedfromanenvironmentallyaware manufacturerthatcomplieswiththeWasteElectricalandElectronic Equipment(WEEE)Directive2002/96/CE.Thisproductmaycontain substancesthatcouldbeharmfultotheenvironmentifdisposedofin places(landfills)thatarenotappropriateaccordingtolegislation.
Followlocalgoverningordinancesandrecyclingplansregarding disposalofthedeviceorcomponentsnormallyusedinoperation. Thedevicedoesnotgeneratewasteorresidueinoperation.DO NOTdisposeofthedeviceinthenormalwastestream.Any accessoriesnotpartofthedeviceMUSTbehandledinaccordance withtheindividualproductmarkingfordisposal.
1176148-A7
HomeFill®System

3Setup

3.1ReceivingInspection

Removeproductfrompackageandinspectfordamage.Ifthereisany damage,DONOTuse.Contactyourdealer/HomeCareequipment provider.

3.2CylinderPrefillInspection

RefertotheHomeFillowner’smanualforcompressor operatinginstructions.
DANGER!
–Improperfillingorabuseofthiscylinder,orfailureto
followthesesafetyinstructionsmaycauseserious bodilyinjuryordeath.Highpressureoxidizinggas vigorouslyacceleratescombustion.
–DONOTalterormodifycylinderorrelated
components.
–DONOTuseoilorgrease.Keepcylinder,cylinder
valve,allequipmentandconnectionsfreeofoiland greasetoavoidviolentignitionresultinginserious bodilyinjuryordeath.
WARNING!
–Formedicalapplications,useonlyasdirectedbya
licensedpractitioner.Uninterrupteduseofhigh concentrationsofoxygenoveralongduration, withoutmonitoringitseffectonoxygencontentof arterialblood,maybeharmful.
WARNING!
–OxygencylindersfilledbytheHomeFillshallbeused
forpersonaluseonly.“Nottobefilledforresaleor usebyprofessionalusers”.
–ThiscylinderistobeusedwiththeHomeFillsystem.
Thiscylindercontains93±3%oxygenproducedby anoxygenconcentratorusingthepressureswing adsorptionprocess.
–DONOThandlecylinderorusecontentsuntil
youareprofessionallytrained,includingemergency procedures.
–UsethiscylinderinaccordancewiththeHomeFill
owner’smanual.Openvalveslowly,Closevalveafter eachuseandwhenempty.Securecylinderduring storageanduse.Nosmokingincylinderarea.Keep awayfromheat,flameorspark.Keepoutofthereach ofchildren.DONOTdrop.
–FillingofcylinderisperformedbyInvacareHomeFill
oxygensystemonly.
81176148-A
Setup
WARNING!
–Cylindersthathavebeenrefinishedand/orexposed
toelevatedtemperaturesMUSTbehydrostatically testedbeforefilling.Cylindertemperatureshouldnot exceed52°C(125°F).Aluminumcylinderssubjectto theactionoffireorheatedtemperaturesinexcessof 177°C(350°F)MUSTbewithdrawnfromservice andcondemnedbytrainedpersonnel.DONOT usecausticpaintstrippers.Valveandsafetyrelief devicesMUSTberemovedandreplacedbytrained andauthorizedpersonnel.DONOTalterorchange valveassembly.
–AllcylindersMUSTbeinspectedbeforeattemptingto
fill;otherwise,injuryordamagemayoccur.
–NEVERsmokeinanareawhereoxygenisbeing
administered.
–NEVERusenearanytypeofflameor
flammable/explosivesubstances,vapoursor atmosphere.

3.3TheHydrostaticTestingDate

AluminumcylindersMUSTundergotestingeveryfiveyears.
WARNING!
–DONOTfillcylindersthathavenotbeentested
inthepastfiveyears.Otherwise,seriouspersonal injurymayresult.Contactyourdealer/HomeCare equipmentproviderforreplacement.

3.4ExternalExamination

1.Examinetheoutsideofthecylinderforthefollowingconditions, andreplacethecylinderiftheyexist:
Dentsordings
Arcburns
Oilorgrease
Anyothersignsofdamagethatmightcauseacylindertobe
unacceptableorunsafeforuse.
2.Examinethecylinderforevidenceoffireorthermaldamage. evidenceincludescharringorblisteringofthepaint,orother protectivecoatingorheatsensitiveindicator.Iffireorthermal damageisfound,replacethecylinder.
3.Inspectthecylinderfillportforthefollowing:
Debris,oilorgrease
Noticeablesignsofdamage
Signsofcorrosioninsidethevalve
Signsofexcessiveheatorfiredamage
Signsofpressureleaks
DONOTusethecylinderifanyoftheseitemsare found.Contactyourdealer/HomeCareequipment providerforreplacement.
1176148-A9
HomeFill®System
101176148-A
Usage

4Usage

4.1Connecting/DisconnectingtheCylinder to/fromtheCompressor

ConnectingtheCylindertotheCompressor
WARNING!
–NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe injuryand/ordamagemayoccur.
–DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise, injuryordamagemayoccur.
1.Examinethecylindergauge.Ifthecylinderpressureislessthan 1500psig(10342kPa),proceedtoSTEP2.
Ifthecylinderpressureisgreaterthan1500psig(10342 kPa),DONOTattempttotopoffthiscylinder.Itmay notfill.
2.Settheconserverflowdialonthecylinderto“O”.
3.Slowlyopentheshutoffvalvefully(counterclockwise).
4.Removethecylinderandconnectorfillportcovers.
1176148-A
CAUTION!
–DONOTconnecttheoutlet(cannula)connection
totheHomeFillunit,otherwisetheconserving cylinderwillnotworkproperly.
5.MomentarilypushDOWNontheouterring(sleeve)ofthe connectorfillportuntilGREENdotsarevisibletoresetthe connector.
Iftheouterring(sleeve)isintheUPposition(GREEN dotsnotvisible),theconnectorfillportwillnotbe abletoacceptthecylinderfillport.PushingDOWN momentarilywillresettheconnectorfillport(GREEN dotsvisible)toacceptthecylinderfillport.
11
HomeFill®System
12
1176148-A
Fillportcoversnotshownforclarity.
6.Graspthecylinder/regulatorassemblyintheareabehindthe cylindergauge.
7.Positionthecylinderinthecompressorcradle.
8.Alignthecylinderfillportwiththeconnectorfillport.
9.PullUPontheouterring(sleeve)oftheconnectorfillportwhile pushingDOWNonthecylinder/regulatorassemblytocouple thecylinderfillportintotheconnectorfillport.
Thecylinderisproperlyconnectedwhenanaudible “click”isheard.
DisconnectingtheCylinderfromtheCompressor
WARNING!
–NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe injuryand/ordamagemayoccur.
Usage
1.PressthecompressorpowerswitchtotheOff(O)position.
2.Graspthecylinder/regulatorassemblyintheareabehindthe cylindergauge.
1176148-A13
HomeFill®System
3.Withtheotherhand,grasptheouterring(sleeve)ofthe connectorfillportandpushDOWN.
4.Liftuponthecylinder/regulatorassem­blytoremovefromtheconnectorfillport.
5.Whenthecylinderfillportisdisconnectedfromtheconnector fillport,releasetheouterring(sleeve)oftheconnectorfillport andusetwohandstoremovethecylinderfromthecompressor cradle.
WARNING!
–Thefillportcoverontheconnectorfillportand
thecylinderfillportMUSTbereplacedafter fillingandwhenevernotinuse.Ifeitherofthe fillportcoversaremissingcontactyourhealthcare providerbeforeusing.
6.Placethefillportcoversontotheconnectorfillportandcylinder fillport.
7.Examinethecylindergaugetoensurecylinderisfull(needle pointingintoGREENarea).Ifthecylinderisnotfull,perform thefollowing:
RepeatSTEPS1–9ofConnectingtheCylindertothe
Compressor,page11
.
Ifcylinderisnotfullaftersecondattempttofill,contact
theprovider.
8.Slowlyclosetheshutoffvalvefully(clockwise).

4.2CylinderOperation

WARNING!
–NEVERsmokeinanareawhereoxygenisbeing
administered.
–NEVERusenearanytypeofflameor
flammable/explosivesubstances,vapoursor atmosphere.
1.Priortoeachuse,inspectproductforvisibledamage.Referto
3.4ExternalExamination,page9 isfound.
.DONOTuseifanydamage
14
1176148-A
Ifforanyreason,anylabelbecomesillegibleorlost, contactyourdealer/HomeCareequipmentprovider.
2.Inspectthecylindercontentsgaugetoensurethepressureisat least500psig(3447kPa).Ifbelow500psig(3447kPa),thenrefill thecylinderusingtheHomeFillsystem.
WARNING!
–Whenpointeroncontentspressuregaugeenters
redzone,itisrecommendedtorefillthecylinder ontheInvacareHomeFillCompressor.
3.Positionthecylindersotheoxygencylindervalveoutletis pointingawayfromtheuserandanyotherperson(s).
WARNING!
–DONOTdirectflowofoxygenatanyperson,or
flammablematerialwhenadjustingtheflowdial.
4.Besuretheflowdialisinthe“O”position.Align“O”withthe indicatorwindow.
5.Attachaduallumenadultoxygennasalcannulawithamaximum lengthof2.1m(7feet)totheproduct’stwooutletconnections accordingtothecannulamanufacturer’sinstructions.
Itdoesnotmatterwhichtubeisinstalledonwhich outletconnection.
DONOTusepediatric,lowflownasalcannulasor oxygenmaskswiththisproduct.
6.Placethecannulaoveryourearsandpositiontheprongsinyour noseasinstructedbyyourHealthCareproviderorcannula manufacturer.
Usage
DANGER! RiskofDeath
Thecannulacancauseinjuryifimproperlypositioned andsecured. –ThecannulaMUSTberoutedandsecured
properly.
–DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhilewearing thecannula.
–Avoidpositioningcannulasacrossareasofhigh
traffic.
7.Ensuretherearenokinksortwistsinthecannulatubing.
8.SlowlyturntheshutoffvalveknobintheONdirection (counterclockwise).
Thiswillallowcylinderpressuretotravelthroughthe regulator.
9.Usetheswitchtoselectoneofthefollowingmodes:
ContinuousFlow
ConservingPulse
1176148-A15
HomeFill®System
Theregulatorislabeled“CONTINUOUS”and“PULSE”. Thefirstthreeflows(0.25,0.50and0.75L/min)arein
continuousmodeatalltimes,regardlessoftheswitch position.
10.Turnflowdialuntiltheindicatorwindowisalignedwiththe prescribedsetting.
Whentherearetwonumbersdisplayedintheindicator window,thelargernumberisthePULSEflowsettingand thesmallernumberistheCONTINUOUSflowsetting.
WARNING!
NOOXYGENisdeliveredbetweensettings.Toavoidinjurytopatient,ALWAYSconfirm
prescribedsettingbeforeadministeringtopatient andmonitorflowonafrequentbasis.
–OnlyusetheInvacareCorporationcarryingbag.
11.Breathethroughthenoseandfeelapulseofoxygenasyouinhale.
Whenonapulsesetting,thereisflow,orapulse,onlyat thebeginningofeachbreath.Ifyoudonotfeelthepulse atthebeginningofeachbreath,checkthesetting.
Becauseeachpatient’sbreathingpatternisdifferentand theenvironmentvaries,itmaybedifficulttofeelsome lowsettingpulses.
Ifoperatingthecylinderbelow1.7ºC(35ºF)usethe2 L/Mincontinuoussettingonly.
Whentheunitisfirstswitchedfrompulsetocontinuous flow,asingleinhalationisrequiredtoinitiatethe continuousflow.
12.Whenfinishedusingthecylinder,performthefollowingsteps:
a.Turntheflowdialtothe“O”position. b.Turntheshutoffvalveclockwisetoturnofftheregulator.
161176148-A

5Maintenance

5.1Maintenance

CAUTION!
–DONOTusecleaningsolutionsonthecoupleror
fillportfitting.
DONOTimmerseproductinanykindofliquid.AllrepairsMUSTbedonebyInvacareCorporationor
inanInvacareapprovedrepaircenter.
1.Asrequiredwhendirty,cleanexterioroftheproductwitha dry,lintfreeclothonly.
2.Storeproductinacleanareafreefromgrease,oil,andother sourcesofcontamination.

5.2ReplacementParts—Accessories

HF2POST6BAG HF2POST9BAG
Maintenance
1176148-A17
HomeFill®System

6Troubleshooting

6.1Troubleshooting

IfConservingCylinderfailstofunction,consulttheTroubleshootingguide.Ifproblemcannotbecorrected,consultyourDealer/HomeCare equipmentprovider.
Problem
Noflow.1.Regulatorin“O”position.
PossibleCauseSolution
2.Cylinderempty.
3.Dialnotsetatincrement.
4.Conservernotsensingbreath.
1.Settoprescribedsetting.
2.Replace/fillcylinder.
3.Setdialsoindicatorpointstoasetting.
4.Checkpositionofcannulainnose.Donot breathethroughmouth.
5.Checktoseeifthecannulaiskinked.
181176148-A

7TechnicalData

7.1TypicalProductParameters

Flows
TechnicalData
ContinuousFlow Setting
PulseVolumes(L)
EquivalentFlowRateMinimumNominalMaximum
1/4L/minsetting:0.2130.2500.288
1/2L/minsetting:0.4250.5000.575
3/4L/minsetting:0.6380.7500.863
0.4L/minsetting:0.3400.4000.460
0.5L/minsetting:0.4250.5000.575
1L/minsetting:0.8501.0001.150
1.5L/minsetting:1.2751.5001.725
2L/minsetting:1.7002.0002.300
2.5L/minsetting:2.1252.5002.875
3.5L/minsetting:2.9753.5004.025
4.5L/minsetting:3.8254.5005.175
1176148-A19
HomeFill®System
PulseSetting
1,1.5,2,2.5,3,4,5,6
Conserving(Savings)Ratio, BolusSizeandAccuracy @20BPM
EquivalentFlowRateMinimumNominalMaximum
1L/minsetting:0.0140.01650.019
1.5L/minsetting:0.0210.02480.029
2L/minsetting:0.0290.03300.038
2.5L/minsetting:0.0360.04130.047
3L/minsetting:0.0430.04950.057
4L/minsetting:0.0560.06600.076
5L/minsetting:0.0710.08250.095
6L/minsetting:0.0840.09900.114
Sensitivity0.130to0.150in(0.330to0.381cm)H2O
Delivery
Pressures
InletPressureRange
MaxFillPressure14479kPa(2100psig)
InternalRegulatedPressure
SafetyBurstDisk
2068-14479kPa(300-2100psig)
345kPa(50psig)
Burstsat20.8—23.2Mpaat22°C(3025–3360psigat72°F)
PulseVolumes(L)
“ON”time<125ms
“OFF”time<300ms
201176148-A
EnvironmentalOperatingTemperatureRange
TechnicalData
StorageTemperatureRange
OperatingTemperatureRange
MaxOperatingHumidity
-20°Cto60°C(-4°Fto140°F)
1.7°Cto40.6°C(35°Fto105°F)
95%non-condensing
AssemblyDimensions
Model
Diameter11.1cm(4.4in)11.1cm(4.4in)
Weight2.07kg(4.57lb)2.75kg(6.07lb)
OverallLength
HF2ECE6HF2ECE9
43.3cm(17.03in)40.7cm(16.03in)
Cylinders
CylinderSize
Volume(H
OxygenCapacity
O)
2
M6M9
1.0L(0.035ft3)1.7L(0.061ft3)
170L(6ft3)255L(9ft3)
1176148-A
21
HomeFill®System
FillTimes
Model
HF2ECE6HF2ECE9
1hrs25min*2hrs20min*
*Concentratorflowratestopatient:
Platinum5,PlatinumSSeriesandPerfecto2
Upto2.5L/min
Series
Platinum9Upto5L/min
Allfillingtimesareapproximateandmayvaryonenvironmentalconditions.
Accessories
Model
DualLumenCannula5’(1.524m)or7’(2.1336m)tubing
CarryBagHF2POST6BAGHF2POST9BAG
HF2ECE6HF2ECE9
length(HUD1850)
RegulatoryCompliance
93/42/EEC-MedicalDeviceDirective(Class2b)
97/23/EC-PressureEquipmentDirective
5’(1.524m)or7’(2.1336m)tubing
length(HUD1850)
22
1176148-A