Invacare®HomeFill®System
HF2ECE9,HF2ECE6
ENOxygenConservingCylinders
UserManual................................3
FRBouteilledeconservationd’oxygène
Manueld'utilisation..........................23
DESauerstoffAufbewahrungs-Zylinder
Gebrauchsanweisung.........................45
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.

©2013Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout
priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare
ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
WD-40and3–IN-ONEareregisteredtrademarksoftheWD-40Company.

Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
1General......................................4
1.1IntendedUse..............................4
1.2Features.................................4
2Safety........................................5
2.1Symbols..................................5
2.2GeneralGuidelines..........................5
2.3Disposal.................................7
3Setup........................................8
3.1ReceivingInspection.........................8
3.2CylinderPrefillInspection.....................8
3.3TheHydrostaticTestingDate..................9
3.4ExternalExamination........................9
4Usage........................................11
4.1Connecting/DisconnectingtheCylinderto/from
theCompressor...........................11
ConnectingtheCylindertothe
Compressor............................11
DisconnectingtheCylinderfromthe
Compressor............................13
4.2CylinderOperation.........................14
5Maintenance..................................17
5.1Maintenance..............................17
5.2ReplacementParts—Accessories..............17
6Troubleshooting...............................18
6.1Troubleshooting...........................18
7TechnicalData................................19
7.1TypicalProductParameters...................19

HomeFill®System
1General
1.1IntendedUse
Thiscylinderisintendedtostore,regulate,anddeliversupplemental
oxygentopatientswhomayhavedifficultyextractingoxygenfrom
theairtheybreathe.Itisforpatientswhowouldnormallyreceivethe
oxygenviaanasalcannula.ThedevicedeliversUSP93±3%oxygenat
anumberofdifferentflowsettings.Itisintendedtobeusedasan
oxygendeliverydeviceandmayincreasethedryingoutofairways.
1.2Features
4
1176148-A

Safety
2Safety
2.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheEC
declarationofconformity.
2.2GeneralGuidelines
WARNING!
–Thissectioncontainsimportantinformationforthe
safeoperationanduseofthisproduct.
–DONOTUSETHISPRODUCTORANYAVAILABLE
OPTIONALEQUIPMENTWITHOUTFIRST
COMPLETELYREADINGANDUNDERSTANDING
THESEINSTRUCTIONSANDANYADDITIONAL
INSTRUCTIONALMATERIALSUCHASUSER
MANUALS,SERVICEMANUALSORINSTRUCTION
SHEETSSUPPLIEDWITHTHISPRODUCTOR
OPTIONALEQUIPMENT.IFYOUAREUNABLE
TOUNDERSTANDTHEWARNINGS,CAUTIONS
ORINSTRUCTIONS,CONTACTAHEALTHCARE
PROFESSIONAL,DEALERORTECHNICAL
PERSONNELBEFOREATTEMPTINGTOUSETHIS
EQUIPMENT-OTHERWISE,INJURYORDAMAGE
MAYOCCUR.
–GENERALGUIDELINEScontainsimportant
informationforthesafeoperationanduseofthis
product.
ACCESSORIESWARNING
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyother
manufacturershavenotbeentestedbyInvacareand
arenotrecommendedforusewithInvacareproducts.
1176148-A5

HomeFill®System
DANGER!
–Thisproductisnotintendedasalife-sustainingor
life-supportingdevice.
–DONOTSMOKEwhileusingthisdevice.Keepall
matches,lightedcigarettesorothersourcesofignition
outoftheroominwhichthisproductislocatedand
awayfromwhereoxygenisbeingdelivered.
–NOSMOKINGsignsshouldbeprominentlydisplayed.
Textilesandothermaterialsthatnormallywouldnot
burnareeasilyignitedandburnwithgreatintensityin
oxygenenrichedair.Failuretoobservethiswarning
canresultinseverefire,propertydamageandcause
physicalinjuryordeath.
–Theinformationcontainedinthisdocumentissubject
tochangewithoutnotice.
OperatingInformation
CylinderMUSTbeemptiedofgaspriortoshippingbycourieror
postalservice.
OnlypersonnelinstructedandtrainedintheuseoftheHomeFill
ConservingCylindershouldoperatethisproduct.
Productcontainsmagnetic,ferrousmaterialthatmayaffectthe
resultsofanMRI.
Besureallconnectionsaretightandleakfree.
DONOTusealiquidleakdetectortotestforleaks.
DONOTautoclave.
DONOTgassterilizewithethyleneoxide.
DONOTcleanwitharomatichydrocarbons,including,butnot
limitedtobenzeneandtoluene.
DONOTattempttorepairproduct.
DONOTimmerseproductinanykindofliquid.
Storeproductinacleanareawhennotinuse.
OnlyuseInvacareCorporationcarryingbag.
NEVERsmokeinanareawhereoxygenisbeingadministered.
ALWAYSconfirmprescribeddosebeforeadministeringtopatient
andmonitoronafrequentbasis.
DONOTuseifdirtorcontaminantsarepresentonoraround
cylinder,valve,productorconnectingdevices.
NEVERusenearanytypeofflameorflammable/explosivesubstances,
vapoursoratmosphere.
NOOXYGENisdeliveredwhenthepointerisalignedwith“O”.
DONOTblocktheoutletfittingorkinkthecannulawhenthe
productisinuse,asthismaydamagetheunit.
Theuseofthisdeviceislimitedtotheoxygenpatient.Cylinders
MUSTbeusedonlybytheoxygenpatientandarenottobe
distributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathethrough
theirmouths.
DONOTallowoilfromyourhandsorothersourcestocome
intocontactwiththecompressorcouplingorbottlefittings.These
substancesmaybecomeflammableinthepresenceofoxygenand
causeinjury.Avoidtouchingthebottlefittingwheneverpossible.
Exampleofitemstoavoidarehandcreamsorlotions,cookingoils,
suntanoilsorsunblocksandsimilaroilyproducts.Ifyouusethese
61176148-A

Safety
typesofproductsInvacaresuggestsyouwashyourhandspriorto
usingtheHomeFillunit.Ifthebottlefittingshouldcomeincontact
withsuchsubstancesitshouldbecleanedoffwithadryclothprior
toconnectingittothecompressor.
NEVERoilorlubricatethecompressorcouplerorbottlefittings.
NEVERuseanypenetratingoilorlubricantsuchasWD-40®or
3-IN-ONE®.Ifconnectiondifficultyisexperienced,verifythatthe
bottlefittingisbeinginsertedstraightdownintothecompressor
coupler.Ifcontinueddifficultyisexperienced,contactyoursupplier
forassistance.
DONOTuseahumidifierwiththisproduct.
DONOTallowcylinderstotiporfall.Securegascylinderssothey
cannotfall.Ensurecylinderisalwaysinuprightposition.
DONOTstorecylindersnearsourcesofheatorflame.
Thisproductisequippedwithareliefvalve.Ifyouhearaloudhissing
orpoppingsoundcomingfromtheproduct,discontinueuseof
product,turntheflowselectortothe“O”position,andcontactyour
dealer/HomeCareequipmentprovider.
DONOTusewhilesleepingwithoutconsultingyourhealthcare
provider.
Thisproductisdesignedtooperatewithastandardorhighflow
duallumen,adultoxygennasalcannulawithamaximumlengthof
2.1m(7ft).
Thecannulaisforsinglepatientuseonly.
CannulaMUSTbeinstalledonoutletconnectionpriortouse.
DONOTplacetheunitunderclothingwhileinuse.Whenthis
productisinuseasmallamountofoxygenisvented.Wearingthe
unitunderclothingmaysaturatefabricswithoxygenandcausethem
toburnrapidlyifexposedtosparksorflame.Itmaytakeseveral
hoursforoxygenlevelsinfabricstoreturntonormal.
DONOTuseHomeFillcompressorunit,tofillcylinderoutdoors
oranywherebelow13°C(55°F).
NEVERattempttofillcylinderwithanythingotherthantheHomeFill
compressor.
2.3Disposal
Thisproducthasbeensuppliedfromanenvironmentallyaware
manufacturerthatcomplieswiththeWasteElectricalandElectronic
Equipment(WEEE)Directive2002/96/CE.Thisproductmaycontain
substancesthatcouldbeharmfultotheenvironmentifdisposedofin
places(landfills)thatarenotappropriateaccordingtolegislation.
Followlocalgoverningordinancesandrecyclingplansregarding
disposalofthedeviceorcomponentsnormallyusedinoperation.
Thedevicedoesnotgeneratewasteorresidueinoperation.DO
NOTdisposeofthedeviceinthenormalwastestream.Any
accessoriesnotpartofthedeviceMUSTbehandledinaccordance
withtheindividualproductmarkingfordisposal.
1176148-A7

HomeFill®System
3Setup
3.1ReceivingInspection
Removeproductfrompackageandinspectfordamage.Ifthereisany
damage,DONOTuse.Contactyourdealer/HomeCareequipment
provider.
3.2CylinderPrefillInspection
RefertotheHomeFillowner’smanualforcompressor
operatinginstructions.
DANGER!
–Improperfillingorabuseofthiscylinder,orfailureto
followthesesafetyinstructionsmaycauseserious
bodilyinjuryordeath.Highpressureoxidizinggas
vigorouslyacceleratescombustion.
–DONOTalterormodifycylinderorrelated
components.
–DONOTuseoilorgrease.Keepcylinder,cylinder
valve,allequipmentandconnectionsfreeofoiland
greasetoavoidviolentignitionresultinginserious
bodilyinjuryordeath.
WARNING!
–Formedicalapplications,useonlyasdirectedbya
licensedpractitioner.Uninterrupteduseofhigh
concentrationsofoxygenoveralongduration,
withoutmonitoringitseffectonoxygencontentof
arterialblood,maybeharmful.
WARNING!
–OxygencylindersfilledbytheHomeFillshallbeused
forpersonaluseonly.“Nottobefilledforresaleor
usebyprofessionalusers”.
–ThiscylinderistobeusedwiththeHomeFillsystem.
Thiscylindercontains93±3%oxygenproducedby
anoxygenconcentratorusingthepressureswing
adsorptionprocess.
–DONOThandlecylinderorusecontentsuntil
youareprofessionallytrained,includingemergency
procedures.
–UsethiscylinderinaccordancewiththeHomeFill
owner’smanual.Openvalveslowly,Closevalveafter
eachuseandwhenempty.Securecylinderduring
storageanduse.Nosmokingincylinderarea.Keep
awayfromheat,flameorspark.Keepoutofthereach
ofchildren.DONOTdrop.
–FillingofcylinderisperformedbyInvacareHomeFill
oxygensystemonly.
81176148-A

Setup
WARNING!
–Cylindersthathavebeenrefinishedand/orexposed
toelevatedtemperaturesMUSTbehydrostatically
testedbeforefilling.Cylindertemperatureshouldnot
exceed52°C(125°F).Aluminumcylinderssubjectto
theactionoffireorheatedtemperaturesinexcessof
177°C(350°F)MUSTbewithdrawnfromservice
andcondemnedbytrainedpersonnel.DONOT
usecausticpaintstrippers.Valveandsafetyrelief
devicesMUSTberemovedandreplacedbytrained
andauthorizedpersonnel.DONOTalterorchange
valveassembly.
–AllcylindersMUSTbeinspectedbeforeattemptingto
fill;otherwise,injuryordamagemayoccur.
–NEVERsmokeinanareawhereoxygenisbeing
administered.
–NEVERusenearanytypeofflameor
flammable/explosivesubstances,vapoursor
atmosphere.
3.3TheHydrostaticTestingDate
AluminumcylindersMUSTundergotestingeveryfiveyears.
WARNING!
–DONOTfillcylindersthathavenotbeentested
inthepastfiveyears.Otherwise,seriouspersonal
injurymayresult.Contactyourdealer/HomeCare
equipmentproviderforreplacement.
3.4ExternalExamination
1.Examinetheoutsideofthecylinderforthefollowingconditions,
andreplacethecylinderiftheyexist:
•Dentsordings
•Arcburns
•Oilorgrease
•Anyothersignsofdamagethatmightcauseacylindertobe
unacceptableorunsafeforuse.
2.Examinethecylinderforevidenceoffireorthermaldamage.
evidenceincludescharringorblisteringofthepaint,orother
protectivecoatingorheatsensitiveindicator.Iffireorthermal
damageisfound,replacethecylinder.
3.Inspectthecylinderfillportforthefollowing:
•Debris,oilorgrease
•Noticeablesignsofdamage
•Signsofcorrosioninsidethevalve
•Signsofexcessiveheatorfiredamage
•Signsofpressureleaks
DONOTusethecylinderifanyoftheseitemsare
found.Contactyourdealer/HomeCareequipment
providerforreplacement.
1176148-A9

HomeFill®System
101176148-A

Usage
4Usage
4.1Connecting/DisconnectingtheCylinder
to/fromtheCompressor
ConnectingtheCylindertotheCompressor
WARNING!
–NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe
injuryand/ordamagemayoccur.
–DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise,
injuryordamagemayoccur.
1.Examinethecylindergauge.Ifthecylinderpressureislessthan
1500psig(10342kPa),proceedtoSTEP2.
Ifthecylinderpressureisgreaterthan1500psig(10342
kPa),DONOTattempttotopoffthiscylinder.Itmay
notfill.
2.Settheconserverflowdialonthecylinderto“O”.
3.Slowlyopentheshutoffvalvefully(counterclockwise).
4.Removethecylinderandconnectorfillportcovers.
1176148-A
CAUTION!
–DONOTconnecttheoutlet(cannula)connection
totheHomeFillunit,otherwisetheconserving
cylinderwillnotworkproperly.
5.MomentarilypushDOWNontheouterring(sleeve)ofthe
connectorfillportuntilGREENdotsarevisibletoresetthe
connector.
Iftheouterring(sleeve)isintheUPposition(GREEN
dotsnotvisible),theconnectorfillportwillnotbe
abletoacceptthecylinderfillport.PushingDOWN
momentarilywillresettheconnectorfillport(GREEN
dotsvisible)toacceptthecylinderfillport.
11

HomeFill®System
12
1176148-A

Fillportcoversnotshownforclarity.
6.Graspthecylinder/regulatorassemblyintheareabehindthe
cylindergauge.
7.Positionthecylinderinthecompressorcradle.
8.Alignthecylinderfillportwiththeconnectorfillport.
9.PullUPontheouterring(sleeve)oftheconnectorfillportwhile
pushingDOWNonthecylinder/regulatorassemblytocouple
thecylinderfillportintotheconnectorfillport.
Thecylinderisproperlyconnectedwhenanaudible
“click”isheard.
DisconnectingtheCylinderfromtheCompressor
WARNING!
–NEVERusetoolsofanykindtoconnect/disconnect
thecylinderandthecompressor.Otherwise,severe
injuryand/ordamagemayoccur.
Usage
1.PressthecompressorpowerswitchtotheOff(O)position.
2.Graspthecylinder/regulatorassemblyintheareabehindthe
cylindergauge.
1176148-A13

HomeFill®System
3.Withtheotherhand,grasptheouterring(sleeve)ofthe
connectorfillportandpushDOWN.
4.Liftuponthecylinder/regulatorassemblytoremovefromtheconnectorfillport.
5.Whenthecylinderfillportisdisconnectedfromtheconnector
fillport,releasetheouterring(sleeve)oftheconnectorfillport
andusetwohandstoremovethecylinderfromthecompressor
cradle.
WARNING!
–Thefillportcoverontheconnectorfillportand
thecylinderfillportMUSTbereplacedafter
fillingandwhenevernotinuse.Ifeitherofthe
fillportcoversaremissingcontactyourhealthcare
providerbeforeusing.
6.Placethefillportcoversontotheconnectorfillportandcylinder
fillport.
7.Examinethecylindergaugetoensurecylinderisfull(needle
pointingintoGREENarea).Ifthecylinderisnotfull,perform
thefollowing:
•RepeatSTEPS1–9ofConnectingtheCylindertothe
Compressor,page11
.
•Ifcylinderisnotfullaftersecondattempttofill,contact
theprovider.
8.Slowlyclosetheshutoffvalvefully(clockwise).
4.2CylinderOperation
WARNING!
–NEVERsmokeinanareawhereoxygenisbeing
administered.
–NEVERusenearanytypeofflameor
flammable/explosivesubstances,vapoursor
atmosphere.
1.Priortoeachuse,inspectproductforvisibledamage.Referto
3.4ExternalExamination,page9
isfound.
.DONOTuseifanydamage
14
1176148-A

Ifforanyreason,anylabelbecomesillegibleorlost,
contactyourdealer/HomeCareequipmentprovider.
2.Inspectthecylindercontentsgaugetoensurethepressureisat
least500psig(3447kPa).Ifbelow500psig(3447kPa),thenrefill
thecylinderusingtheHomeFillsystem.
WARNING!
–Whenpointeroncontentspressuregaugeenters
redzone,itisrecommendedtorefillthecylinder
ontheInvacareHomeFillCompressor.
3.Positionthecylindersotheoxygencylindervalveoutletis
pointingawayfromtheuserandanyotherperson(s).
WARNING!
–DONOTdirectflowofoxygenatanyperson,or
flammablematerialwhenadjustingtheflowdial.
4.Besuretheflowdialisinthe“O”position.Align“O”withthe
indicatorwindow.
5.Attachaduallumenadultoxygennasalcannulawithamaximum
lengthof2.1m(7feet)totheproduct’stwooutletconnections
accordingtothecannulamanufacturer’sinstructions.
Itdoesnotmatterwhichtubeisinstalledonwhich
outletconnection.
DONOTusepediatric,lowflownasalcannulasor
oxygenmaskswiththisproduct.
6.Placethecannulaoveryourearsandpositiontheprongsinyour
noseasinstructedbyyourHealthCareproviderorcannula
manufacturer.
Usage
DANGER!
RiskofDeath
Thecannulacancauseinjuryifimproperlypositioned
andsecured.
–ThecannulaMUSTberoutedandsecured
properly.
–DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhilewearing
thecannula.
–Avoidpositioningcannulasacrossareasofhigh
traffic.
7.Ensuretherearenokinksortwistsinthecannulatubing.
8.SlowlyturntheshutoffvalveknobintheONdirection
(counterclockwise).
Thiswillallowcylinderpressuretotravelthroughthe
regulator.
9.Usetheswitchtoselectoneofthefollowingmodes:
•ContinuousFlow
•ConservingPulse
1176148-A15

HomeFill®System
Theregulatorislabeled“CONTINUOUS”and“PULSE”.
Thefirstthreeflows(0.25,0.50and0.75L/min)arein
continuousmodeatalltimes,regardlessoftheswitch
position.
10.Turnflowdialuntiltheindicatorwindowisalignedwiththe
prescribedsetting.
Whentherearetwonumbersdisplayedintheindicator
window,thelargernumberisthePULSEflowsettingand
thesmallernumberistheCONTINUOUSflowsetting.
WARNING!
–NOOXYGENisdeliveredbetweensettings.
–Toavoidinjurytopatient,ALWAYSconfirm
prescribedsettingbeforeadministeringtopatient
andmonitorflowonafrequentbasis.
–OnlyusetheInvacareCorporationcarryingbag.
11.Breathethroughthenoseandfeelapulseofoxygenasyouinhale.
Whenonapulsesetting,thereisflow,orapulse,onlyat
thebeginningofeachbreath.Ifyoudonotfeelthepulse
atthebeginningofeachbreath,checkthesetting.
Becauseeachpatient’sbreathingpatternisdifferentand
theenvironmentvaries,itmaybedifficulttofeelsome
lowsettingpulses.
Ifoperatingthecylinderbelow1.7ºC(35ºF)usethe2
L/Mincontinuoussettingonly.
Whentheunitisfirstswitchedfrompulsetocontinuous
flow,asingleinhalationisrequiredtoinitiatethe
continuousflow.
12.Whenfinishedusingthecylinder,performthefollowingsteps:
a.Turntheflowdialtothe“O”position.
b.Turntheshutoffvalveclockwisetoturnofftheregulator.
161176148-A

5Maintenance
5.1Maintenance
CAUTION!
–DONOTusecleaningsolutionsonthecoupleror
fillportfitting.
–DONOTimmerseproductinanykindofliquid.
–AllrepairsMUSTbedonebyInvacareCorporationor
inanInvacareapprovedrepaircenter.
1.Asrequiredwhendirty,cleanexterioroftheproductwitha
dry,lintfreeclothonly.
2.Storeproductinacleanareafreefromgrease,oil,andother
sourcesofcontamination.
5.2ReplacementParts—Accessories
HF2POST6BAG
HF2POST9BAG
Maintenance
1176148-A17

HomeFill®System
6Troubleshooting
6.1Troubleshooting
IfConservingCylinderfailstofunction,consulttheTroubleshootingguide.Ifproblemcannotbecorrected,consultyourDealer/HomeCare
equipmentprovider.
Problem
Noflow.1.Regulatorin“O”position.
PossibleCauseSolution
2.Cylinderempty.
3.Dialnotsetatincrement.
4.Conservernotsensingbreath.
1.Settoprescribedsetting.
2.Replace/fillcylinder.
3.Setdialsoindicatorpointstoasetting.
4.Checkpositionofcannulainnose.Donot
breathethroughmouth.
5.Checktoseeifthecannulaiskinked.
181176148-A

7TechnicalData
7.1TypicalProductParameters
Flows
TechnicalData
ContinuousFlow
Setting
PulseVolumes(L)
EquivalentFlowRateMinimumNominalMaximum
1/4L/minsetting:0.2130.2500.288
1/2L/minsetting:0.4250.5000.575
3/4L/minsetting:0.6380.7500.863
0.4L/minsetting:0.3400.4000.460
0.5L/minsetting:0.4250.5000.575
1L/minsetting:0.8501.0001.150
1.5L/minsetting:1.2751.5001.725
2L/minsetting:1.7002.0002.300
2.5L/minsetting:2.1252.5002.875
3.5L/minsetting:2.9753.5004.025
4.5L/minsetting:3.8254.5005.175
1176148-A19

HomeFill®System
PulseSetting
1,1.5,2,2.5,3,4,5,6
Conserving(Savings)Ratio,
BolusSizeandAccuracy
@20BPM
EquivalentFlowRateMinimumNominalMaximum
1L/minsetting:0.0140.01650.019
1.5L/minsetting:0.0210.02480.029
2L/minsetting:0.0290.03300.038
2.5L/minsetting:0.0360.04130.047
3L/minsetting:0.0430.04950.057
4L/minsetting:0.0560.06600.076
5L/minsetting:0.0710.08250.095
6L/minsetting:0.0840.09900.114
Sensitivity0.130to0.150in(0.330to0.381cm)H2O
Delivery
Pressures
InletPressureRange
MaxFillPressure14479kPa(2100psig)
InternalRegulatedPressure
SafetyBurstDisk
2068-14479kPa(300-2100psig)
345kPa(50psig)
Burstsat20.8—23.2Mpaat22°C(3025–3360psigat72°F)
PulseVolumes(L)
“ON”time<125ms
“OFF”time<300ms
201176148-A

EnvironmentalOperatingTemperatureRange
TechnicalData
StorageTemperatureRange
OperatingTemperatureRange
MaxOperatingHumidity
-20°Cto60°C(-4°Fto140°F)
1.7°Cto40.6°C(35°Fto105°F)
95%non-condensing
AssemblyDimensions
Model
Diameter11.1cm(4.4in)11.1cm(4.4in)
Weight2.07kg(4.57lb)2.75kg(6.07lb)
OverallLength
HF2ECE6HF2ECE9
43.3cm(17.03in)40.7cm(16.03in)
Cylinders
CylinderSize
Volume(H
OxygenCapacity
O)
2
M6M9
1.0L(0.035ft3)1.7L(0.061ft3)
170L(6ft3)255L(9ft3)
1176148-A
21

HomeFill®System
FillTimes
Model
HF2ECE6HF2ECE9
1hrs25min*2hrs20min*
*Concentratorflowratestopatient:
Platinum5,PlatinumSSeriesandPerfecto2
Upto2.5L/min
Series
Platinum9Upto5L/min
Allfillingtimesareapproximateandmayvaryonenvironmentalconditions.
Accessories
Model
DualLumenCannula5’(1.524m)or7’(2.1336m)tubing
CarryBagHF2POST6BAGHF2POST9BAG
HF2ECE6HF2ECE9
length(HUD1850)
RegulatoryCompliance
93/42/EEC-MedicalDeviceDirective(Class2b)
97/23/EC-PressureEquipmentDirective
5’(1.524m)or7’(2.1336m)tubing
length(HUD1850)
22
1176148-A