Invacare HF2ECE6, HF2ECE9 User Manual

Invacare®1.0L/1.7LCylinderwith DualLumenConserver
ForHomeFill®OxygenSystem
enHF2ECE6,HF2ECE9
UserManual
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
93/42/EEC
©2017InvacareCorporation Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited withoutpriorwrittenpermissionfromInvacare.Trademarksareidentiedby™and®.Alltrademarks areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A. WD-40and3–IN-ONEareregisteredtrademarksoftheWD-40Company.
Contents
1General.........................................4
1.1Symbols.....................................4
1.2IntendedUse.................................5
1.3UserProle...................................5
1.4ServiceLife...................................5
1.5Warranty.....................................5
1.6ComponentIdentication........................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................8
3Setup...........................................14
3.1DealerSetup..................................14
3.2UserSetup...................................14
3.3ReceivingInspection............................14
3.4CylinderPrellInspection........................14
3.4.1TheHydrostaticTestingDate...................15
3.4.2ExternalExamination.........................16
4Usage..........................................17
4.1BeforeUse...................................17
4.2Connecting/DisconnectingtheCylinderto/fromthe
Compressor.................................17
4.3CylinderAssemblyOperation......................17
5Maintenance.....................................21
5.1Maintenance..................................21
5.2EmptyingtheCylinderAssembly...................22
5.3Cleaning.....................................23
5.4Disinfecting...................................23
5.5Reuse.......................................23
5.6Storage......................................24
5.7OptionalAccessories............................24
6AfterUse........................................25
6.1Disposal.....................................25
6.2WearandTear................................25
7Troubleshooting...................................26
7.1Troubleshooting...............................26
8T echnicaldata....................................27
8.1TechnicalDescription............................27
8.2Specications.................................28
Invacare®1.0L/1.7LCylinderwithDualLumenConserver
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazards orunsafepracticeswhichcouldresultinpersonalinjuryor propertydamage.Seetheinformationbelowfordenitions ofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin deathorseriousinjury.
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin deathorseriousinjury.
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin propertydamageorminorinjuryorboth.
–Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformation forefcient,trouble-freeuse.
Readthemanual
Noopename
DONOTSmoke
Oxidizer
Recycle
DONOTdisposeofinhouseholdwaste
Manufacturer
EURepresentative
ThisproductcomplieswithDirective93/42/EEC concerningmedicaldevices.
Thelaunchdateofthisproductisstatedinthe CEdeclarationofconformity.
4
1176148-D
General
1.2IntendedUse
Thiscylinderisintendedtostore,regulate,anddeliver supplementaloxygentopatientswhomayhavedifculty extractingoxygenfromtheairtheybreathe.Itisforpatients whowouldnormallyreceivetheoxygenviaanasalcannula. ThedevicedeliversUSP93±3%oxygenatanumberof differentowsettingsandmayincreasethedryingoutof airways.Thisproductistobeusedasanoxygensupplement andisnotintendedtobelifesupportingorlifesustaining.
DANGER! RiskofInjuryorDeath
Thisproductistobeusedasanoxygen supplementandisnotintendedtobe life-supportingorlife-sustaining.ONLYusethis productifthepatientiscapableofspontaneous breath,abletoinhaleandexhalewithouttheuse ofamachine.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse andspecicationshasnotbeentestedandmay leadtoproductdamage,lossofproductfunction, orpersonalinjury .
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended usesectionsofthismanual.
1.3UserProle
Nospecicproductknowledgeortrainingisrequired,other thanwhatiscontainedinthismanual,tooperatethe
product.Itisnotintendedforsupplyingoxygentherapy duringrelling.Thisproductisintendedtobeusedbyor underadultsupervisiononlyafterreadingandunderstanding theinstructionsandwarningsofthisusermanual.
Onlypersonnelinstructedandtrainedintheuseofthe HomeFill®ConservingCylindershouldoperatethisproduct.
Theuseofthisproductislimitedtotheoxygenpatient. CylindersMUSTbeusedonlybytheoxygenpatientandare nottobedistributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathe throughtheirmouths.
1.4ServiceLife
Theexpectedservicelifeoftheconserverisveyearswhen usedinaccordancewiththesafetyinstructions,maintenance intervalsandcorrectusestatedinthismanual.Refertothe Maintenancesection.
Theexpectedservicelifeofthecylinderisbasedontesting thecylindereverytenyearsandusingitinaccordancewith thesafetyinstructions,maintenanceintervalsandcorrect usestatedinthismanual.Refertothefollowingprocedures:
TheHydrostaticTestingDate,ExternalExamination, Maintenance
1.5Warranty
Termsandconditionsofthewarrantyarepartofthegeneral termsandconditionsparticulartotheindividualcountriesin whichthisproductissold.
1176148-D5
Invacare®1.0L/1.7LCylinderwithDualLumenConserver
1.6ComponentIdentication
Item
Description
A
CylinderAssembly
B
Cylinder
C
ContinuousFlowConserver
D
FlowDial
E
IndicatorWindow
F
Switch
G
OutletConnection
H
CylinderFillport
I
ShutoffValve
J
CylinderGauge
K
CylinderGaugeNeedle
61176148-D
Safety
2Safety
2.1LabelLocations
Serialnumberlocatedaboveproductlabelshown here.
1176148-D
7
Invacare®1.0L/1.7LCylinderwithDualLumenConserver
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyand operationoftheproducttheseinstructionsMUSTbe followed.
DANGER! RiskofDeath,InjuryorDamage
–Improperuseofthisproductmaycause
death,injuryordamage.Thissectioncontains importantinformationforthesafeoperation anduseofthisproduct.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanuals,servicemanualsorinstruction sheetssuppliedwiththisproductoroptional equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professional,dealerortechnicalpersonnel beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct isnotworkingcorrectly,contactatechnician orInvacareforrepair .
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskofDeath,InjuryorDamage
–Usethiscylinderinaccordancewiththe
HomeFillcompressorusermanual.
–DONOThandlethecylinderassemblyor
usethecontentsuntilyouareprofessionally trained,includingemergencyprocedures. Trainingrequiredtooperatetheproductis containedintheusermanual.
–Thisproductisnotintendedforsupplying
oxygentherapyduringrelling.
–Thisproductisintendedtobeusedbyorunder
adultsupervision.
–OxygenCylinderslledbytheInvacareHomeFill
oxygensystemshallbeusedforpersonaluse only.“Nottobelledforresaleoruseby professionalusers.”
–ThiscylinderistobeusedwiththeInvacare
HomeFilloxygensystem.Thiscylinder contains93±3%oxygenproducedbyan oxygenconcentratorusingthepressureswing adsorptionprocess.
–ALWAYSopenthevalveslowlyandclosethe
valveaftereachuseandwhenempty.
81176148-D
Safety
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–DONOTSMOKEwhileusingthisproduct. –DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonthecylinder
unlessrecommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutof theroominwhichthiscylinderislocatedand awayfromwhereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andcylinder
outfromundersuchitemsasblankets, bedcoverings,chaircushions,clothing,and awayfromheatedorhotsurfacesincluding spaceheaters,stoves,andsimilarelectrical appliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesale toorontheorderofamedicalpractitioner licensedbyagovernmentalagencywherehe/she practices.
–ONL Yalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
1176148-D9
Invacare®1.0L/1.7LCylinderwithDualLumenConserver
OperatingInformation
DANGER! RiskofFire
Dirtorcontaminantsonthecylinderassemblyor conservercancausere,injuryordeath.
–DONOTuseifdirtorcontaminantsarepresent
onoraroundcylinderorconserver .
–DONOTuseorllcylinderassemblyifthe
cylinderllportisdirtyorcontaminated. Contactyourproviderforserviceorrepair.
–DONOTallowoilfromyourhandsorother
sourcestocomeintocontactwiththe compressorcouplingorcylinderassembly ttings.Thesesubstancesmaybecome ammableinthepresenceofoxygenandcause injury.Examplesofitemstoavoidarehand creamorlotion,cookingoil,suntanoilor sunblockandsimilaroilyproducts.Ifyouuse thesetypesofproductsInvacaresuggestsyou washyourhandspriortousingtheHomeFill system.Ifthecylinderllportshouldcome incontactwithsuchsubstances,contact yourproviderforservice.Avoidtouchingthe cylinderllport.
DANGER! RiskofFire
Dirtorcontaminantsonthecylinderassemblyor conservercancausere,injuryordeath.
–DONOTcleanwitharomatichydrocarbons,
including,butnotlimitedtobenzeneand toluene.
–NEVERoilorlubricatethecompressorcoupler
orcylinderllport.NEVERuseanypenetrating oilorlubricantsuchasWD-40®or3-IN-ONE®. Ifconnectiondifcultyisexperienced,verify thatthecylinderllportisbeinginserted straightdownintothecompressorcoupler. Ifcontinueddifcultyisexperienced,contact yourproviderforassistance.
101176148-D
Safety
DANGER! RiskofFire
Dirtorcontaminantsonthecylinderassemblyor conservercancausere,injuryordeath.
–DONOTusecleaning/disinfectingsolutionson
thellporttting. –DONOTimmerseproductinanykindofliquid. –DONOTallowanykindofliquidtogetinside
openingsoftheconserver . –DONOTspraycleaning/disinfectingsolutions
onproduct. –DONOTautoclave. –DONOTgassterilizewithethyleneoxide. –DONOTattempttorepaircylinderassembly.
AllrepairsMUSTbeperformedbyInvacare
CorporationorinanInvacareapprovedrepair
center . –Storecylinderassemblyinaclean,dryandcool
areawhennotinuse.
DANGER! RiskInjuryorDeath
Toavoidchokingoringestionofchemicalsfrom airwaycontamination:
–DONOTusetheproductinthepresence
ofpollutants,smoke,fumes,ammable
anesthetics,cleaningagents,orchemical
vapors.
WARNING! RiskofInjuryorDeath
Contaminantsintheoxygenmaycauseinjuryor death.
–DONOTuseiftheoxygenhasanunusualodor . –DONOTuseifparticulatesareinhaled. –ContactInvacareoradealerforservice.
WARNING! RiskofInjuryorDamage
Toavoidinjuryordamagefromproductmisuse:
–Theowdialshouldbeturnedwithnormal
pressurefromone’shand.Iftheowdialis stifftoturn,donotforce,otherwisedamage orinjurycouldoccur .
–DONOTuseaowoutletfordrivingany
medicalequipment.
1176148-D
11