Invacare CHECK O2 PLUS Assembly, Installation And Operating Instructions

INVACARE CHECK O2 PLUS HAND-HELD ANALYZER
ANALYSEUR PORTABLE INVACARE CHECK O
INVACARE CHECK O
INVACARE CHECK O
KOMBI-PRÜFGERÄT
2
2
PLUS
2
Assembly, Installation and Operating Instructions - ENGLISH
Installation, Montage et Instructions de Fonctionnement - FRENCH
Zusammenbau, Installation und Gebrauchsanweisung - GERMAN
Montaje, Instalación e Instrucciones de Funcionamiento - SPANISH
USER: BEFORE USING THIS ANALYZER, READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE.
NOTE: Check all parts for shipping damage. If dam­age is found, DO NOT use. Contact Carrier/Invacare for further instructions.
SAFETY SUMMARY
The following recommendations are made for the safe and proper use of the Invacare Check O
Plus Hand-Held Analyzer:
2
WARNING
DO NOT install this equipment without first read­ing and understanding this instruction sheet. If you are unable to understand these instruc­tions, contact a healthcare professional, dealer or technical personnel if applicable before attempting to install this equipment ­otherwise, injury or damage may occur.
Oxygen rapidly accelerates combustion:
DO NOT smoke while using oxygen. DO NOT use oxygen in the presence of
flames, sparks, or glowing objects. Ensure that all items are properly and se-
curely installed.
INSTRUCTIONS FOR USE (FIGURE 1)
1. Press the ON ( I ) button. The LCD display should
come on.
NOTE:The analyzer automatically shuts off after ap­proximately two (2) minutes of operation. This helps save battery life.
2. The display will show “FLO” for approximately two (2) seconds and then switch to readings in L/min.
3. If no reading is displayed or if display is erratic or frozen, refer to the troubleshooting section of this instruction sheet.
4. In order to measure oxygen concentration or flow rate, connect the analyzer’s FLOW/CON­CENTRATION inlet with a piece of ¼-inch or crushproof tubing to the outlet fitting of the gas source to be measured.
5. In order to measure pressure, connect the analyzer’s PRESSURE inlet with a piece of ¼­inch or crushproof tubing to the outlet fitting of the gas source to be measured.
CAUTION
DO NOT expose any port of the analyzer to any pressure greater than 68.95 kPa or dam­age may result to internal components.
DO NOT use the analyzer on pure oxygen from high pressure cylinders or liquid systems.
DO NOT use the Invacare Check O
Plus
2
Hand-Held Analyzer on any gas stream that is humidified. Use of the analyzer with water vapor can cause erroneous readings and in­ternal damage.
INTENDED USE
NOTE: The Invacare Check O2 Plus Hand-Held Analyzer is designed and calibrated for use in mea­suring unhumidified output gas of an oxygen con­centrator ONLY.
NOTE: The analyzer is not for continuous monitoring of patient gases.
NOTE: There are no user serviceable parts inside the analyzer. Field calibration of the unit is not re­quired.
NOTE: When using the analyzer after storage in ei­ther extremely hot or cold conditions, allow the gas being measured to flow through the analyzer for one (1) minute to bring the internal sensor up to the gas stream temperature.
DISPLAY MEASURES VALUE
FLO Flow rate Liters per minute O2 Oxygen concentration Percentage oxygen PSI Pressure Pounds per square inch KPA Pressure KiloPascals
NOTE: The Invacare Check O2 Plus Hand-Held Ana­lyzer can be used to measure the performance of an oxygen concentrator from its outlet fitting, or any con-
centrator accessory such as the Invacare Remote Flowmeter or Pediatric Flowmeter. However, Invac­are does NOT recommend using the Invacare Check O
Plus Hand-Held Analyzer to measure flow rates
2
below 1 L/min. See product specifications. NOTE: The concentrator output gas, measured from
the concentrator outlet fitting or from a concentrator accessory outlet fitting,must NOT contain humidified gas. Disconnect any humidifier bottle before taking any readings to prevent erroneous readings and dam­age to the internal flow sensor. Be sure to reconnect any required humidifier bottle before returning any equipment to patient use.
Pressure
Inlet
Flow/
Concentration
Inlet
LCD
Display
Function
On Button
Button
FIGURE 1-INSTRUCTIONS FOR USE
2
6. Pressing the FUNCTION ( ) button will toggle the analyzer through its various detecting functions.
7. With "O2" readings, values below 70% will be displayed as "LO".
8. When the battery is completely discharged, BAt will FLASH once on the LCD and the analyzer will turn itself off. Refer to section on changing the battery in this Instruction sheet.
CHANGING THE BATTERY
NOTE: The internal battery needs to be replaced whenever the BAt display is seen on the LCD.
1. Turn the analyzer over.
2. Remove the six (6) screws holding the case bot­tom to the top and take the bottom cover off.
3. Remove the discharged battery from the bat­tery clips and dispose of or recycle properly.
CAUTION
Observe proper polarity for battery installation.
4. Install a new 9 volt battery on the battery clips. Make sure the battery snaps tightly on the clips.
NOTE: Use of a DURACELL® alkaline 9 volt battery will provide for maximum life of the unit.
5. Turn the unit over and press the ON button. The display should come on. If it does, replace the bottom cover and install the six cover screws.
6. If the unit does not come on, remove the battery, check it in another device or on a battery tester and reinstall it. Try turning the unit on again.
7. If unit still fails to operate with a known good bat­tery, contact Invacare Technical Service for repair.
TROUBLESHOOTING
NOTE: There are no user serviceable parts inside the analyzer.
1. If the unit fails to come on, replace the battery with a known good battery.
PRODUCT SPECIFICATIONS
OPERATIONAL
Pressure:
Maximum Input Pressure: 68.95 kPa, 0 to 10 PSIG Resolution: 0.7 kPa 0.1 PSI Accuracy:
Flow:
Maximum Flow Rate: 6.0 L/min. Input: Unhumidified oxygen
Resolution: 0.1 L/min. Accuracy: + 5% (+ 0.3 L/min.)
Gas Stream Temperature: +15 to +35°C
Concentration:
Range: 73% to 96% Input: Unhumidified oxygen
Resolution: 0.1 % Accuracy: + 2%
Gas Stream Temperature: +15 to +35°C
Alarm: Low battery
Power On Time: 1 minute, 45 seconds
Display: LCD
Battery: Single NEDA/ANSI 1604A
Battery life: 400+ hours (13,700+ read
Warranty: 1 year
PHYSICAL Weight: 0.30 kg.
ENVIRONMENTAL
Operating Temperature Range: +15 to +35°C
+ 2% F.S. + 2% F.S.
concentrator output only
concentrator output only
+ 5 seconds
alkaline 9 volt
cycles) with alkaline battery
Width: 8.380 cm Length: 18.54 cm Includung Inlet Fittings: 20.07 cm Thickness: 3.18 cm
2. If the unit fails to come on with known good bat­tery, contact Invacare Technical Service for repair.
3. If the unit comes on but the FUNCTION button does not select any functions, contact Invacare Technical Service for repair.
4. If the display comes on but the readings are er­ratic or frozen, contact Invacare Technical Ser­vice for repair.
5. If a segment of the display is missing or displayed intermittently, contact Invacare Technical Service for repair.
Storage Temperature Range: -20 to +70°C
Humidity range: 20 to 60% relative humidity
Altitude (maximum): Up to 1828 meters above
sea level
3
LIMITED WARRANTY
This warranty is extended only to the original purchaser of our products and does not affect statutory rights.
Invacare UK warrants the Invacare Check O2 Plus Hand-Held Analyzer to be free from defects in materials and workmanship for a period of one (1) year from date of pur­chase. If within such warranty period the product shall be proven to Invacare UK’s reasonable satisfaction to be defective, it shall be repaired or replaced, at Invacare UK’s option; Invacare UK’s sole obligation, and your exclusive remedy under this war­ranty being limited to such repair or replacement.
For warranty service, please contact Invacare UK Service Department during normal working hours. Upon receiving notice of an alleged defect in a product, Invacare UK will issue a serialised return authorisation. The product shall then be returned to either Invacare UK’s plant or service center as specified by Invacare UK. Defective compo­nent part(s) must be returned for warranty inspection using the serial number as iden­tification within thirty (30) days of return authorisation date. DO NOT return products to our factory without prior consent.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY DOES NOT COVER NOR­MAL WEAR AND TEAR AND SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, COMMERCIAL OR IN­STITUTIONAL USE, PRODUCTS MODIFIED WITHOUT INVACARE UK’S EXPRESS WRITTEN CON­SENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORISED PARTS OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE UK.
4
NOTE : Vérifiez qu’aucune pièce n’a été endommagée au cours du transport. En cas de dommage, NE PAS se servir de l'appareil. Contactez le Ttansporteur/In­vacare pour de plus amples informations.
PRECAUTIONS D'UTILISATION
Les recommandations suivantes sont destinées à garantir une utilisation sûre et appropriée de l’analyseur portable Invacare Check O
Plus :
2
AVERTISSEMENT
N’INSTALLEZ pas cet appareil sans avoir préalablement lu et compris cette fiche d’instructions. En cas d'incompréhension de ces instructions, contactez un professionnel de la santé, un distributeur ou un technicien, le cas échéant, avant de tenter d’installer cet appareil pour éviter toute lésion ou dommage matériel.
L’oxygène accélère rapidement la combustion:
NE fumez PAS quand vous utilisez de l’oxygène.
N’utilisez PAS d'oxygène en présence de flammes, d’étincelles ou d’objets incandescents.
Vérifiez que toutes les pièces sont installées de façon correcte et sûre.
ATTENTION
N’EXPOSEZ aucun port de l’analyseur à une pression supérieure à 68,95 kPa, vous risqueriez d’endommager les composants internes.
N'UTILISEZ PAS l'analyseur avec de l'oxygène pur provenant d'une bouteille à haute pression ou d'un système liquide.
N’UTILISEZ PAS l’analyseur portable Invacare Check O humidifiés. L’utilisation de l’analyseur avec de la vapeur d’eau peut provoquer des erreurs de lecture et des dégâts internes.
Plus avec des flux gazeux
2
CHAMPS D’UTILISATION RECOMMANDE
NOTE : L’analyseur portable Invacare Check O2 Plus est conçu et calibré uniquement pour mésurer les gaz non humidifiés en sortie d’un concentrateur d’oxygène.
NOTE : L’analyseur n’est pas destiné au contrôle continu des gaz du patient.
NOTE : L’analyseur ne comporte aucune pièce dont la maintenance peut être assurée par l’utilisateur. Un calibrage de l’appareil est inutile.
INSTRUCTIONS D'UTILISATION (SCHEMA 1)
1. Pressez le bouton ON (I). L'écran à cristaux liquides doit s'allumer.
NOTE : L’analyseur s’arrête automatiquement après environ deux (2) minutes de fonctionnement. Cela permet d’économiser la batterie.
2. L’écran affiche "FLO" pendant environ deux (2) secondes puis passe aux relevés en L/mim.
3. Si aucune lecture ne s’affiche ou si l’affichage est discontinu ou figé, reportez-vous à la section résolution des pannes de cette fiche d’instructions.
4. Pour mesurer la concentration en oxygène ou le débit, raccordez l’entrée DEBIT/CONCENTRA- TION de l’analyseur avec une section de tube de ¼ de pouce anti-écrasement au raccord de sor­tie de la source d’oxygène à mesurer.
5. Pour mesurer la pression, raccordez l’entrée PRESSION de l’analyseur avec une section de tube de ¼ de pouce anti-écrasement au raccord de sortie de la source d’oxygène à mesurer.
6. Pressez le bouton FONCTION ( ) pour faire défiler les différentes fonctions de détection de l’analyseur.
AFFICHAGE MESURES VALEUR
FLO Débit Litres par minute O2 Concentration en oxygène Pourcentage d'oxygène PSI Pression Livres par pouce carré KPA Pression KiloPascals
NOTE : L’analyseur Invacare Check O2 Plus est destiné à mesurer la performance à la sortie patient d’un concentrateur d’oxygène ou d’un accessoire tel que le débitmètre portable et le débitmètre pédiatrique. Néanmoins Invacare recommande de ne pas utiliser l’analyseur pour mesurer des débits inférieur à 1 l/min. Voir spécifications produit.
NOTE : L’oxygène en sortie d’un concentrateur d’oxygène, mesuré à la sortie patient d’un concentrateur ou à la sortie patient d’un accessoire ne doit pas être humidifié. Afin d’éviter des mesures incorrectes et l’endommagement du sensor interne, déconnectez l’humidificateur. Assurez-vous d’avoir reconnectez l’humidificateur, s’il y a lieu, avant de réinstaller l’appareil chez le patient.
Entrée
Pression
Entrée
Débit/
Concentration
Ecran
d’affichage
à cristaux
liquides
NOTE : Lorsque vous utilisez l’analyseur après stockage dans un environnement extrêmement chaud ou froid, faites passer les gaz mesurés dans l’analyseur pendant une (1) minute pour amener le capteur interne à la température du flux gazeux.
Bouton
Fonction
SCHEMA 1-INSTRUCTIONS D'UTILISATION
5
Bouton ON