Invacare 3600XL, 3600 User Manual

0 (0)

Owner’s Operator and Maintenance Manual

Side Entry Whirlpool Tub

Model 3600 & Model 3600XL

Model 3650

AFTER INSTALLATION, PLEASE CALL CUSTOMER SERVICE AT 1-800-668-2337 TO ARRANGE FOR AN IN-SERVICE ON THE TUB

DEALER: This manual MUST be given to the user of the tub.

USER: BEFORE using this tub, read this manual and save for future reference.

For more information regarding Invacare Continuing Care Group products, parts, and services, please visit www.invacare-ccg.com

WARNING

DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.

The initial set up of this tub MUST be performed by a qualified technician.

Procedures other than those described in this manual MUST be performed by a qualified technician.

ACCESSORIES WARNING

Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.

For further information on this product, please call the following:

Customer Service 1 800 668 2337

Technical Support 1 800 668 2337

Side Entry Whirlpool Tub

2

Part No 1118387

TABLE OF CONTENTS

TABLE OF CONTENTS

 

SPECIAL NOTES ................................................................................

5

LABEL LOCATION ............................................................................

6

TYPICAL PRODUCT PARAMETERS ....................................................

7

FEATURES ........................................................................................

8

Models 3600 and 3600XL.........................................................................................................................

8

Model 3650 ..................................................................................................................................................

9

SECTION 1—GENERAL GUIDELINES .................................................

10

Important Safety Instructions ................................................................................................................

10

Instructions Pertaining to a Risk of Fire, Electric Shock, or Injury to Persons .....................

10

Installation ..................................................................................................................................................

10

Electrical.................................................................................................................................................

11

Plumbing.................................................................................................................................................

11

Maintenance and Inspection ..............................................................................................................

12

SECTION 2—PARTS AND INSTALLATION ........................................

14

Parts Information......................................................................................................................................

14

About Installation .....................................................................................................................................

14

Preparing the Work Area.......................................................................................................................

14

Securing Tub ..............................................................................................................................................

16

Electrical and Plumbing Hook-up Locations on the Tub.................................................................

17

Electrical Installation ................................................................................................................................

17

Plumbing Installation ................................................................................................................................

20

Post Installation Checklist ......................................................................................................................

21

SECTION 3—OPERATING INSTRUCTIONS ........................................

22

Overview....................................................................................................................................................

22

Disinfecting the Tub Using the Closed Loop System.......................................................................

23

Opening/Closing the Tub Door ............................................................................................................

24

Opening the Tub Door.......................................................................................................................

24

Closing the Tub Door.........................................................................................................................

24

Tub Operation ..........................................................................................................................................

25

Water Temperature Setting..............................................................................................................

25

Operating Shower/Massage Wand ..................................................................................................

26

Filling Tub...............................................................................................................................................

26

Hydromassage.......................................................................................................................................

27

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING ....................

28

Part No 1118387

3

Side Entry Whirlpool Tub

TABLE OF CONTENTS

TABLE OF CONTENTS

 

Safety Inspection Checklists...................................................................................................................

28

Initially.....................................................................................................................................................

28

Before Each Use...................................................................................................................................

28

Every 80-150 Baths..............................................................................................................................

28

Every Six Months..................................................................................................................................

28

Cleaning ......................................................................................................................................................

29

Cleaning the Door Seal Gasket ........................................................................................................

29

Removing Soap Film and Hard Water Deposits...........................................................................

29

Removing Calcium Deposits, Scale and Lime Build-up ...............................................................

29

Cleaning Your Tub...............................................................................................................................

29

Heavy Duty Cleaning Procedure ..........................................................................................................

30

Calibrating the Disinfectant Flow Meter.............................................................................................

31

Troubleshooting........................................................................................................................................

33

Maintenance Record ................................................................................................................................

34

LIMITED WARRANTY .....................................................................

36

Side Entry Whirlpool Tub

4

Part No 1118387

SPECIAL NOTES

SPECIAL NOTES

Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.

 

SIGNAL WORD

MEANING

 

 

 

 

DANGER

Danger indicates an imminently hazardous situation which, if not avoided, will result in

 

 

death or serious injury.

 

 

 

 

WARNING

Warning indicates a potentially hazardous situation which, if not avoided, could result in

 

 

death or serious injury.

 

 

 

 

CAUTION

Caution indicates a potentially hazardous situation which, if not avoided, may result in

 

 

property damage or minor injury or both.

 

 

 

 

 

 

NOTICE

THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE.

WARNING

RADIO FREQUENCY INTERFERENCE

Most electronic equipment is influenced by Radio Frequency Interference (RFI). CAUTION should be exercised with regard to the use of portable communication equipment in the area around such equipment. If RFI causes erratic behavior, turn the power off IMMEDIATELY. DO NOT turn the power on while transmission is in progress.

MAINTENANCE

Maintenance MUST be performed only by qualified personnel.

Information on maintenance and adjustment of the thermoscopic mixing valve (RADA) is in the RADA Owner’s Manual which is included with this product.

Use this unit only for its intended use as described in this manual. DO NOT use attachments not recommended by the manufacturer.

Part No 1118387

5

Side Entry Whirlpool Tub

Invacare 3600XL, 3600 User Manual

LABEL LOCATION

LABEL LOCATION

Interior Hardware

Labels located inside the Disinfecting

Model 3650

 

 

 

Models 3600 and

 

 

Cabinet Door

 

 

 

 

 

 

3600XL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODEL NO.

U.S. AND

 

FOREIGN

 

PATENTS

SERIAL NO.

PENDING

 

 

% +

 

 

 

" #$% #&$" ' ' ()

 

' . /. 0 ' 1 2) . 3 ' 1"4. (

 

 

 

 

% #< ==9

#

 

#<

 

 

 

# %

% ,

 

 

 

%

 

!

 

> ? 0

!

 

 

# %

 

 

5 1

% #

 

 

 

'?

 

 

 

 

0 ()

#

# %-

 

 

%8

# ! # * & + # # ,

. 72 ;

 

CAUTION

Some state and local plumbing codes require the installation of a “reduced pressure zone” (RPZ) assembly on both the incoming hot and cold supply lines to prevent back flow contamination into the potable water system. Invacare recommends you check local plumbing code requirements to determine if RPZs are required on your tub installation. RPZs are NOT supplied by Invacare and are to be provided by the customer if required.

Disinfecting Cabinet Door

P/N 1150759 Rev A

Side Entry Whirlpool Tub

6

Part No 1118387

TYPICAL PRODUCT PARAMETERS

TYPICAL PRODUCT PARAMETERS

MODEL

3600

 

3600XL

 

3650

 

 

 

 

 

 

MINIMUM SPACE REQUIREMENTS:

 

 

7 feet x 6 feet

 

 

 

 

 

 

 

LENGTH

 

 

 

 

 

EXTERIOR:

63 inches

 

72 inches

 

67 inches

INTERIOR:

46.5 inches

 

56 inches

 

56 inches

 

 

 

 

 

 

WIDTH

 

 

 

 

 

EXTERIOR:

34 inches

 

32 inches

INTERIOR (SEAT):

32 inches

 

26 inches

 

 

 

 

 

 

HEIGHT

 

 

 

 

 

OVERALL:

44 inches

 

45 inches

RIM:

35.5 inches

 

35.5

 

 

 

 

SIDE ACCESS DOOR SWING:

40 inches

 

38.5 inches

 

 

 

 

 

VOLUME CAPACITY (WITHOUT USER):

44 gallons

 

64 gallons

 

 

 

 

 

LIFT BASE CLEARANCE:

6 inches

 

6 inches

 

6¼ inches

 

 

 

 

 

 

WHIRLPOOL PUMP

 

 

 

 

 

SANITATION:

 

Drain with back flush

 

MOTOR:

1 HP, 115 VAC, 60 HZ, 15

 

 

 

 

 

 

JET OUTLETS (ADJUSTABLE/DIRECTIONAL):

2 jets

 

4 jets

 

4 jets

 

 

 

 

TUB MATERIAL:

Double wall gel coated fiberglass

 

 

THERMOMETER:

2 inch dial, 40° F to 140° F range

 

 

ANTI-SCALD WATER REGULATOR:

RADA 222 thermoscopic mixing valve

 

 

 

 

WATER SUPPLY:

 

¾ inch hot and cold

 

 

 

 

 

 

DRAIN SIZE:

 

 

2 inches

 

 

 

 

 

FILL RATE:

 

24 gallons per minute

 

 

 

 

 

DRAIN RATE:

 

25 gallons per minute

 

 

 

 

 

 

COLOR:

 

 

White

 

 

 

 

 

 

 

WEIGHT

 

 

 

 

 

FILLED W/O USER:

796 pounds

 

1,012 pounds

 

903 pounds

EMPTY:

252 pounds

 

274 pounds

 

263 pounds

SHIPPING:

319 pounds

 

380 pounds

 

350 pounds

 

 

 

 

 

 

APPROVALS:

 

 

UL

 

 

 

 

 

 

 

Part No 1118387

7

Side Entry Whirlpool Tub

FEATURES

FEATURES

Models 3600 and 3600XL

Tub Fill Valve

Control Panel

Flow Water Thermometer

 

 

Shower ON/

 

OFF Valve

Tub Spout

Whirlpool

Aerator Valve

Water Temperature Control Valve

Whirlpool ON/OFF Button

 

 

 

Tub Interior

Control Panel (See Above)

 

Burp Fitting

 

Shower Massage

Wand Holder

Whirlpool Inlet

 

 

 

 

 

 

 

Drain

 

 

 

 

 

 

 

 

 

 

Whirlpool Jets

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Tub Exterior

 

 

 

Door Handle

 

 

Door

 

 

 

 

Motor Access

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hot and Cold Water

Disinfectant Cabinet

 

 

 

Side Entry Whirlpool Tub

8

Part No 1118387

FEATURES

Model 3650

Control Panel

 

Water Temperature Control Valve

Flow Water Thermometer

 

Tub Fill Valve

Shower ON/

 

OFF Valve

Whirlpool

Whirlpool

ON/OFF Button

Aerator Valve

Tub Interior

 

Burp Fitting

Control Panel (See Above)

 

Tub Spout

Drain Plug

Whirlpool Jets

Whirlpool Inlet

Door Handle

Safety Pin

Seat

Tub Exterior

Shower Massage Wand Holder

Door Handle

Door

Motor Access Hole

Hot Water Inlet

Disinfectant Cabinet

Cold Water Inlet

Part No 1118387

9

Side Entry Whirlpool Tub

SECTION 1—GENERAL GUIDELINES

SECTION 1—GENERAL GUIDELINES

WARNING

SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product.

Check all parts for shipping damage before using. In case of damage, DO NOT use the equipment. Contact the ICCG dealer for further instructions.

SAVE THESE INSTRUCTIONS

Important Safety Instructions

Instructions Pertaining to a Risk of Fire, Electric Shock, or Injury to Persons

DANGER

To reduce the risk of injury, DO NOT permit children to use this unit unless they are closely supervised at all times.

WARNING

When using this unit, basic precautions should be always be followed, including the following:

READ AND FOLLOW ALL INSTRUCTIONS.

Use this unit only for its intended use as described in this manual. DO NOT use attachments not recommended by the manufacturer.

NEVER drop or insert any object into any opening.

DO NOT operate this unit without the guard over the suction fitting.

Installation

ICCG assumes no responsibility for code infractions or damage to components due to improper installation.

Use extreme care during installation.

The installation area MUST be able to support the 65 pounds per square inch load capacity of the tub.

After installation and before use ensure all parts are properly and securely installed.

Ensure bathroom floor is dry and free of obstructions before transporting or assembling the unit.

Side Entry Whirlpool Tub

10

Part No 1118387

SECTION 1—GENERAL GUIDELINES

Electrical

Be certain ALL electrical work is performed by a qualified electrician and is in compliance with local electrical codes.

ALL electrical connections are to be water resistant.

Grounding

DANGER

Grounding is required. The unit should be installed by a qualified service representative and grounded.

WARNING

The unit MUST be connected only to a supply circuit that is protected by a ground fault circuit-interrupter (GFCI). Such a GFCI should be provided by the installer and should be tested on a routine basis. To test the GFCI, push the test button. The GCFI should interrupt power. Push the resest button. Power should be restored. If the GCFI fails to operate in this manner, GFCI is defective. If the GFCI interrupts power to the bathtub without the test button being pushed, a ground current is flowing, indicating the possibility of an electric shock. DO NOT use this hydro massage bathtub. Disconnect the hydro massage bathtub and have the problem corrected by a qualified service representative before using.

A GREEN-colored terminal (or a wire connector marked “G,” “GR,” “Ground,” or “Grounding”) is provided within the terminal compartment. To reduce the risk of electric shock, connect this terminal or connector to the grounding terminal of your electric service or supply panel with a conductor equivalent in size to the circuit conductors supplying this equipment.

Plumbing

Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes.

Some state and local plumbing codes require the installation of a “reduced pressure zone” (RPZ) assembly on both the incoming hot and cold supply lines to prevent back flow contamination into the potable water system. Invacare recommends you check local plumbing code requirements to determine if RPZs are required on your tub installation. RPZs are NOT supplied by Invacare and are to be provided by the customer if required.

When operating water pressure exceeds 55 PSI, pressure regulators (similar to the ¾ inch Watts U5) MUST be installed.

The hot water supply to the tub MUST not exceed 110° F (43° C) to protect the user from scalding.

Part No 1118387