Invacare 1556922, 1557021, I-Transia User Manual

ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
IENI
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe practiceswhichcouldresultinpersonalinjuryorpropertydamage. Seetheinformationbelowfordefinitionsofthesignalwords.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage orminorinjuryorboth.
IMPORTANT
–Indicatesahazardoussituationthatcouldresultin
damagetopropertyifitisnotavoided.
Givesusefultips,recommendationsandinformationfor efficient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC concerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheEC declarationofconformity.
1.2IntendedUse
I-Transiaisahoistwhichcoverstheneedforliftingormovinga personwithdisabilitiesinthehomecaresector,nursinghomes, rehabilitationcentres,andinstitutions.
Thefreestandingrailsystemisaportablerailsystemwhichcanbeset upinmanylocations*andrequiresnomountingonwalls,ceilingor floors.Thefreestandingrailsystemislengthandheightadjustable.
*Thefreestandingrailsystemandhoistarenotdesignedfor useinmoistorwetareas.
1.3ConditionsforUse
TheuseoftheI-Transiahoistissubjecttothefollowing:
TheI-Transiashouldonlybeusedbytrainedpersonnel.
Thesafeworkingload(SWL),205kg(450lbs)or250kg(550
lbs),mustnotbeexceeded.Referto2.3LoadLimits,page2.
TheinstructionofferedbyInvacaretoallcustomergroupsin
connectionwiththepurchaseofahoisthasbeenreceived.
Thecaregiverpaysattentiontothewell-beingoftheuserwhen
usingthehoist.
Thehoistisusedinrailsystemswhichareinstalled,testedand
approvedaccordingtoDS/EN10535andInvacare’sstipulations.
OnlytechnicianswhohavebeencertifiedbyInvacaremayinstall
andtesttherailsystems.
ThehoistisusedwithanInvacarehangerbar.
ThehoistisusedwithanInvacareliftingslingorwithother
suitableslings.RefertotheI-Transiahoistusermanualformore informationaboutslings..
1.4ServiceLife
TheI-Transiafreestandingrailsystemhasanexpectedlifetimeof15 years,ifusedaccordingtotheproceduresinthismanual.
Invacare®I-Transia™
ENAdjustableFreestandingRailSystem
UserManual
I-Transia™
2Safety
2.1GeneralGuidelines
WARNING!
–Readandunderstandtheseinstructions.DONOT
usethisproductwithoutreadingandunderstanding theseinstructionsandthehoistusermanual.Ifyou needhelpwiththeseinstructions,contactahealthcare professionaloranInvacaredealer.Otherwiseinjuryor damagemayoccur.
ACCESSORIESWARNING
–Invacareproductscanonlybeusedwithaccessories
madebyInvacare.AccessoriesnotmadebyInvacare havenotbeentestedbyInvacare.Accessoriesnot madebyInvacareshouldnotbeused.
Theinformationcontainedinthisdocumentissubjectto changewithoutnotice.
Checkallpartsforshippingdamagebeforeusing.Incaseofdamage, DONOTusetheequipment.ContactanInvacaredealerforfurther instructions.
2.2OperatingInformation
Thissectionofthemanualcontainsgeneralsafetyinformationabout yourproduct.Forspecificsafetyinformation,refertotheappropriate sectionofthemanualandprocedureswithinthatsection.
General
WARNING!
Thismanualcoverstheinstallationofthefreestanding railsystem.RefertotheI-Transiausermanualforsafety informationrelatedtousingandmaintainingthehoist. –Readtheinstructionscarefullybeforeusingthe
I-Transiaandinconnectionwithcleaningandservice ofthehoist.
–Thefreestandingsystemcanbesetupandtakendown
byoneperson.
–Thefreestandingrailsystemshouldbesetuponalevel
surface.
TheI-Transia’smaximumloadmustneverbeexceeded.TheI-Transiamayonlybeusedtoliftormoveaperson
withdisabilities.
–Thefreestandingrailsystemmustnotbeusedforany
otherpurposethanthatstatedinthisusermanual.The railsystemisnotaplaygroundoraswingdevice.
–IftheI-Transiafailswhileinuse,stopusingthehoist,
transferthepatientusingadifferentmethodand contactanInvacaredealerforrepairbyacertified technician.
–Thisproductismechanicallyprotectedagainstderailing
andjamming.
–Theendstopshavetobeproperlyinstalledandlocked
withspringpins,andthe2yellowthumbscrewson thetopbracketshavetobeproperlytightenedbefore usingthesystem.
–Checkthefreestandingsystemregularlytoensureit
isstable.
PinchPointsandPositioning
WARNING! RiskofInjury
PinchpointsarepresentonthestrapcarabinerAand fingerscouldbepinched. ThehangerbarBcanmovesuddenlyandcauseinjury. –ALWAYSkeephandsandfingersclearofmovingparts
toavoidinjury.
–Whenpositioningthehoist,beawareofthepositionof
thehangerbarBandthepatient.Injurycouldoccur.
WARNING!
Pinchpointsarepresentinthepivotpointsofthevertical polesandbetweenthetoprailandtheverticalpoles. –ALWAYSkeephandsandfingersclearofmovingparts
toavoidinjury.
–ALWAYSkeephandsandfingersclearofpinchpoints
duringsetupofthefreestandingrailsystem.
2.3LoadLimits
Readthelabeloneachcomponentwhichindicatesthemaximumload limitsforeachcomponent.
Thecomponent,e.g.hangerbar,liftingsling,etc.labelledwiththe lowestloadlimitdeterminesthemaximumloadlimitfortheentire system.
Thismaximumloadlimitmustnotbeexceeded.
Pleasenotethatthemaxloadmaychangewhendifferentcomponents areused,suchashangerbars,liftingslings,etc.
2
1183379A
2.4ProductLabeling
ItemLabelText
1
WARNING
Tightentopwingknobnutsfirmlybyhand
beforesystemusage.
2
3
3Setup
3.1IncludedItems
Eachkitcontainsthefollowingitems:
1.Topbracketandrailmodule
2.Verticalpolemodule(x2)
3.Usermanual,partnumber1183379(notshown)
4.Hoist*
5.Hangerbar
6.5mmAllenkey(notshown)
*TheHoistcomeswithaseparateHoistUserManual(part number1183377).Refertothismanualforinformationon thesafeoperationandmaintenanceofthehoist.
3.2Inspection
WARNING! RiskofInjury
Damagedcomponentsofthefreestandingrailsystem couldcauseinjury.
Inspectthepackaginguponreceipt.Ifthepackagingisdamaged,thecomponentsofthe
freestandingrailsystemshouldbecarefullycheckedfor noticeableimperfections,defectsorflaws.
–Ifdamageissuspected,DONOTuseorassemblethe
freestandingrailsystem.ContactanInvacaredealer.
3.3AssemblingtheFreestandingRail
Foreaseofassembly,ensureverticalpostsareintheirlowest position.Referto3.6AdjustingtheHeight,page5.
1.OpenthelegsAofoneverticalpolemoduleB.
2.RaisetheverticalpolemoduleB.
3.Liftoneendofthetopbracketandrailmodule\Candmountit ontopoftheverticalpolemoduleB.
Thetopbracketmustbeflushwiththemark\Gonthe verticalpole.
Itwillbenecessarytotiltthetopbracketandrailmodule andverticalpoletoassemblethem.
4.LockthetopbracketCtotheverticalpoleBwiththelocking armD.
1183379A3