iM3 GS Deluxe START Operating Manual

GS Deluxe START
Manual
iM3 - The Veterinary Dental Company
iM3 Inc iM3 Pty Ltd 12414 95th Street 21 Chaplin Drive Vancouver WA 98682 USA Lane Cove NSW 2066 Australia Tel (360) 254 2981 Fax (360) 254 2940 Tel (02) 9420 5766 Fax: (02) 9420 5677
iM3 Dental Limited Unit 9, Block 4, City North Business Park Stamullen, Co Meath
K32 V634
Ireland Tel Main office: +353 16911277 Tel UK Direct : 01423 224297
www.im3vet.com
2015
2
TABLE OF CONTENTS
Table of contents
2
iM3 Unique Design Features.........................................................................................................
3
iM3 GS DeluxeControl Panel Guide............................................................................................
4
Rear of iM3 GS Deluxe & Ultrasonic Scaler coolant outlet…………………….…………….
5
Getting Started …………………………………………………………………………………..
6-7
iM3 Distilled Water and CLS Enviro Solution Supply..............................................................
8
iM3 Filtration System....................................................................................................................
9
3 Way Air Water Syringe..............................................................................................................
10
iM3 Advantage Low Speed Handpiece and attachments ……..................................................
11-13
High Speed Handpiece .............................................................................
14
iM3 Dental System Bur Kits and Rodent Kits………................................................................
15
Handpiece Pressure Adjustment..................................................................................................
16
Handpiece Lubrication .................................................................................................................
17
Handpiece Sterilization ………………………………………………........................................
18
Handpiece Maintenance................................................................................................................
19
Silent Hurricane Oil Free Compressor ……………...................................................................
Operating Instructions..............................................................................................................
Identifying Air Leaks.................................................................................................................
Compressor Component Details……………………………………………………………...
Adjusting Regulator Pressure...................................................................................................
20
20
21 22
Silent Hurricane Compressor Maintenance................................................................................
23
Trouble Shooting; Cartridge Toggle switches ……………………………………………………………………
High Speed Handpiece ……….................................................................................................
Low Speed Handpiece ..............................................................................................................
24 24 25
iM3 Operator Safety
Im3 Dental equipment should be operated by a qualified person—for animal use only
For operator safety, it is recommended that safety goggles
or visor, face mask and surgical gloves be worn
during all procedures.
Visit Our Web Site: www.im3vet.com
3
Unique Design Features
iM3 Dental Systems are designed with unique and practical features to make your work even more satisfying and productive. Please note; some features listed below, may be optional depending on the system you purchase.
Control Top
The Corian® top is a non-porous work surface which is easy to sanitize. It incorporates a groove around the edge to prevent burs falling on the floor during procedures. The control box is made from powder coated aluminum that wont warp or deteriorate with age.
Color Coded System
All supply lines on iM3 dental systems are color coded to simplify service and maintenance requirements.
Maintenance Instructions
All maintenance instructions are conveniently printed on the card hanging from the unit, or printed
on the back of the dental machine ensuring the system is kept in perfect working order.
Power Board
A four socket electric power board is conveniently located on the side of the control panel, featuring an IEC female socket. (Optional on the GS, standard on the Deluxe)
Adjustable Stainless Steel Stand
The iM3 stand is made from high quality stainless steel and is height adjustable for work and ease of storage. (Optional on the GS, standard on the Deluxe)
Wall Arm
An air operated approx 54”/1.50m long arm is available
Ultrasonic Scaler Coolant Outlet
The ultrasonic scaler coolant outlet is mounted conveniently on the rear of the control box, to allow the use of the water or system to supply the coolant to your ultrasonic dental scaler.
CLS Enviro System
CLS Enviro System is an ingenious occupational health and patient safety system. CLS is designed to help reduce the risk of atmospheric pollution and the transfer of viruses, such as FIV from patient to patient and reduce unpleasant mouth odors.
Oil Free Silent Hurricane Compressor Mounts underneath the unit on the stainless steel stand—supplying air wherever the unit is to be
used, and is considered more convenient than a noisy compressor which is usually situated in an­other room and requires a length of air line. iM3s has produced an Oil free compressor - low maintenance, oil free air.
4
iM3 GS Deluxe START Control Panel
(A) Corian ® work surface with recessed magnetized bur holder, to prevent burs falling on the floor. (B) Handpiece Selector switch for the selection of low speed or High Speed Handpiece. (C) Low Speed Handpiece – used for polishing, and low speed cutting – 20,000 RPM. (D) 3 way syringe – air, water, or mist when both buttons are pressed together. (E) Coolant bottle pressure switch. (F) Pressure gauge for setting handpiece pressure. (G) Coolant Flow Control – regulates the amount of liquid to the high speed bur. (H) High Speed Handpiece – for high speed cutting (I) The amber bottle allows the use of CLS, which is light sensitive (contains chlorhexidine) (J) iM3 R&R oscillating prophy angle L7465 Maintenance Instructions are printed on the back of the machine. Choice of Stand – Wall mounted air activated arm, Stainless Steel Height adjustable stand and Stainless Steel Height adjustable Stand with iM3 silent hurricane oil free compressor.
iM3 is a licensed trademark of iM3 Inc. Corian ® is a registered trademark of the DUPONT Company
iM3 Dental Systems
B
E
C
D
F
G
H
A
I
J
5
REAR OF iM3 GS Deluxe START
With the coolant bottle pressure switch (F - Page 4) in the ONposition, the bottle will be pressurized to 35PSI (approx. 2.5 bar) to supply water/CLS to the 3 way syringe and the High Speed Handpiece. It must be switched to the off position to release the pressure before re filling the bottle.
(1) Ultra Sonic Coolant outlet – This will supply pressurized water/CLS to your ultrasonic dental scaler – A male fitting can be supplied by iM3 for your ultrasonic scaler water line. DO NOT use any coolant other than distilled water or CLS – any other solution may cause a blockage and will void your warranty.
(2) Amber coolant bottle for the use of CLS or distilled water in the GS Deluxe.
(1) Ultra Sonic Coolant Outlet
(2) PET Pressure Coolant Bottle amber - for CLS solution
iM3 Dental Systems
Machine maintenance instructions printed on back of machine
Main power supply. IEC Male power cord is required. Power board overload protector located inside unit.
6
Getting Started
1. Attach all four castor wheels to the stainless steel base.
2. Loosen the stand height adjustment knob and lift the dental control box to working height. Tighten the stand height adjustment knob to secure.
4. Fill the amber plastic bottle with the CLS Concentrate Solution (8.5oz or 250 ml) and
dilute with distilled water (25.4oz or 750ml) to make a 33.82 oz or 1 litre solution. Screw the bottle into the holder with the green tubing. Do not over tighten.
5. Place the 3 way air/water syringe in its handpiece holder—center position.
6. Connect the handpieces by screwing them to the grey handpiece line (see picture below) and
place in correct handpiece holder labeled on the dental system. Page 4 (D) and (I)
7. Check the main dental machine air line
(grey in colour) is connected to the quick disconnect fitting on the compressor outlet tap (on/off) (See page 23)
8. Plug the unit into an electrical outlet with power cord supplied and turn power board ON.
(see page 5)
iM3 Dental Systems
7
iM3 Dental Systems
Getting Started
9. Turn on the compressor and allow to pressurize, (see page 21.)
10. Make sure your compressor is set to deliver the correct working pressure. The air pressure
regulator gauge should read 80 psi (5.5Bar). If adjustment is required, please see page 27 for compressor pressure adjustment instructions. The pressure is set at the iM3 factory.
11. Place the foot control pedal on the floor in front of the unit. The foot pedal operates the High
and Low speed handpieces when depressed.
12. Check handpiece operation:
a) Select HS on the Toggle, Remove High Speed Handpiece from holder and depress handpiece foot control pedal—handpiece should now operate. Page 4 (C) b) Select LS and Remove Low Speed Handpiece from holder and depress handpiece foot control pedal—handpiece should now operate. Page 4 (C)
13. Adjust individual handpiece pressure if required, (see page 16). Each handpiece—Low
speed and high speed has a specific pressure requirement.
14. Pressurize water system by selecting the coolant bottle pressure switch to ON, located on
the front of the control panel. (see page 4 F)
Important. When filling distilled water and/or CLS Enviro bottles de-pressurize the water system, by switching the coolant bottle pressure switch to off. Check water and air
supply to the 3 way air water syringe. Depress the right hand button for air and left hand button for water or both together for a mist.
15. Water volume. Remove high speed handpiece, from holder and depress foot control pedal.
Adjust water spray volume to a fine mist by rotating the flow control knob. (see page 4 H”)
Please observe manufacturers handpiece maintenance instructions for
handpiece longevity.
8
iM3 Distilled Water and CLS Enviro Solution Supply
Each iM3 dental system is supplied with an amber water bottle which should be filled only with distilled water to assist in preventing mineral or bacteria build up in the handpieces. This water bottle supplies the high speed handpieces, 3 way air water syringe and ultra sonic scaler coolant outlet . The CLS solution contains chlorhexidine which is light sensitive and should be kept in the amber bottle provided. The coolant system for your iM3 dental system, has been preset at 25 psi.
Refilling Coolant Bottles
To refill the distilled water or CLS solution, flip the coolant pressure switch (p. 4 F) on your dental system to OFF - depressurizing the system. After the system has been depressurized, unscrew the bottle and refill with either distilled water or CLS Solution. After filling the coolant bottle screw it back into the holder and flip the coolant switch to ON, repressurizing the system.
DO NOT OVERTIGHTEN THE BOTTLES.
Please Note: When the days dental procedures are complete, flip the ON/OFF coolant switch to the OFF position, depressurizing the coolant bottles for storage.
Heres what Dr Jeanie Hawkins, DVM, Diplomat AVDC had to say about CLS
Solution and the Auto Flush System…..
Having CLS Solution at my fingertips to flush periodontal pockets is a real plus. iM3® is the only Company to address the problem of cleaning the airlines and high speed handpieces.
iM3 Dental Systems
iM3 CLS solution 250ml - Ordering Code C3130
9
iM3 Dental Systems
iM3 Filtration System CHECK
Water and CLS Enviro Solution Supply
Inline filters in the supply bottles help prevent particles from entering the system and blocking handpieces. These filters are located in the distilled water and CLS Enviro bottles.
The inline filters will need to be checked and replaced if they appear dirty.
The use of distilled water is mandatory in the water bottles, as local water may clog the system and can corrode the brass valves in the system, apart from the risk of a bacterial
build up.
Note: Failure to use distilled water will void the iM3 machine warranty.
Note: The use of other chlorhexidine solutions may cause crystal deposits
in handpieces and valves. Use of such solutions will void the warranty.
Do not dilute CLS beyond the recommended label instructions, as it may
result in crystal deposits forming and block handpieces and valves. Chlorine
in water will cause chlorhexidine to flocculate—use only distilled water.
10
3 Way Air/Water Syringe
Right-hand button for air (a) and the left-hand button for water (b) and both for a fine mist
(a & b).
The 3 Way Syringe is ideal for flushing away debris during prophys or drying areas during endodontic procedures.
Tip Removal
The tip of the 3 Way Syringe may be removed for cleaning and autoclaving.
Simply depress the outer ring housing (c) and remove the tip. (See diagram below).
iM3 Dental Systems
Ultrasonic Scaler Coolant Outlet
The Ultra Sonic coolant outlet is located on the back of the control box. Page 5
This outlet provides coolant under pressure for an ultrasonic scaler.
(a)
(b)
(c)
(a) Air (b) Water (c) Outer Ring Housing
11
iM3 Dental Systems
Operation Recommendations
Air pressure The dental unit air pressure
should be adjusted to be 40 psi See page 16
Lubrication To assure optimal performance and life the handpiece should be lubricated daily with MD-30 oil lubricant. Instructions for lubrication are found on page 20. Air supply Moist or dirty air from the compressor can cause damage.
Check the regulator filter regularly. If dirt or moisture is present, replace the regulator filter and bleed the system, to remove water or oil.
Installation Instructions
1. Preset the air pressure to 40 psi (see page 16)
1. Connect the low speed handpiece to the grey handpiece line
2. Fit the 4:1 Advantage straight nose cone. Attach the Blue R&R oscillating prophy head
3. Depress the foot control fully to operate the low speed handpiece
Note: Polishing should be carried out at 5,000 rpm or less. This can be
accomplished by running the motor at full revs (20,000 Rpm) and use a 4:1
reduction head, to achieve 5,000 revs, maintaining torque.
iM3 Advantage Low Speed Handpiece L6200
The low speed handpiece features include a light weight balanced profile and smooth swivel action. Its vane type power system is quiet and vibration free. The motor accommodates both doriot or ISO heads and its powerful torque is suitable for all low speed procedures. The Low speed Motor operates at 20,000 RPM with delivery pressure of 40 psi.
12
iM3 Dental Systems
iM3 Low Speed Handpiece Attachments
(1) 4:1 Advantage Straight Nose Cone Attachment L6250
The reducing head is designed for polishing (prophy). The speed of the polishing cup should be maintained at less than 5,000 r.p.m. The reducing attachment therefore allows the motor speed and torque to be maintained while reducing the polishing speed. iM3 disposable oscillating prophy angles are fitted to the SNC. They wont trap muzzle hair, or generate excessive heat.
(iM3 R&R prophy cup Code L7465)
(2) 1:1 Straight Nose Cone Attachment Blue Band
For (HP burs) and the use with the iM3 rodent kit.
This includes a soft tissue protector with diamond bur and diamond disc.
(3) Diamond Saw Attachment
This saw blade may be single or double sided and operates in the 1:1 attachment blue band straight nose cone on the low speed handpiece. It should always be used with a guard.
(4) Prophy Paste
iM3 recommends the use of iM3 R&R individual tub prophy paste. The paste contains no fluoride or flavor oils. It is a medium grit suitable for use with the R&R prophy cups.
13
iM3 Dental Systems
Advantage Low Speed Motor and 4:1 straight Nose Cone
See operating instructions supplied with iM3 low speed handpiece
Fitting R&R Prophy Cups (L7465)
Place the R&R prophy angle over the chuck housing on the 4:1 SNC mak­ing sure the slot in the R&R lines up with the small screw in the side of the chuck housing.
See picture to right.
14
High Speed Handpiece
The High Speed Handpiece is designed to operate at a pressure of 40 psi at the handpiece and at speeds of up to 350,000 rpm. A very light touch is necessary as handpieces operates with very low torque and high speed. Pressure on the handpiece of 30gms (1ounce) or more, will cause the bur to stall.
Note: All handpieces should be lubricated prior to use with iM3 MD-30
iM3 Dental Systems
15
iM3 Autoclavable Bur Kits
There are three Bur Kits in the iM3 range, the F.G. Dog (Friction Grip), F.G. Cat and H.P. (Handpiece Burs). Each kit contains 10 burs of various sizes and applications. Each bur has been selected by a specialist Veterinary Dentist. iM3 Bur Kits come complete with a laminated suggested application guide. Each bur holder carries individual re-order codes and can be ordered in packs of five. The blue holder is autoclavable and prevents the burs from falling out.
iM3 Dental Systems
Feline FG Bur kit (Order code) D7530
Canine FG Bur kit (Order code) D7510
HP Bur kit (Order code) D7540
44.5mm long
The iM3 Rabbit/Rodent Soft tissue kit comes complete with Long diamond bur and soft tissue protector and diamond disc with safety shield. Complete Kit (optional) Code: D2245
The soft tissue kit should be used on an iM3 1:1 straight nose cone - Blue
16
iM3 Handpiece Pressure Adjustment
Handpiece Pressure Adjustment
Both the High Speed and Low speed handpiece should be set to a specific pressure. The Handpiece pressure adjustment Knobs for the High Speed and Low Speed handpieces are located underneath the dental machine behind the grey tubing for each handpiece. The adjustment knobs are a small silver knurled knob (see picture below).
Turn your unit on and select the Handpiece (high or low speed) to be adjusted, remove the handpiece and depress the foot pedal to operate the handpiece. Turn the adjustment knob until the desired pressure is displayed on the handpiece pressure gauge picture 2 below.
Low
Speed
Handpiece
40 p.s.i. - Pre set by iM3
High Speed Handpiece
40 p.s.i. Pre set by iM3
iM3 Dental Systems
3 way syringe hold­er
Low speed hand­piece holder
Handpiece pressure adjustment knob
PRESSURE IS READ ON THE HANDPIECE
PRESSURE GAUGE
17
iM3 Dental Systems
Lubrication of HS & LS Handpieces L5200 & L6200
Refer to separate handpiece manual.
All handpieces should be lubricated regularly. We recommend the use of MD-30
handpiece oil before use and after 30 minutes of running time.
The foot pedal should be depressed for 30 seconds after oiling to ensure
the lubricant will reach the turbines and working portions of the handpiece.
The lubricant is placed in the smaller of the 2 large holes.
Air entry holeat the base of the handpiece.
NOTE:
DONT FORGET TO DEPRESS THE FOOT CONTROL AFTER OILING TO ENSURE LUBRICANT REACHES THE TURBINES.
Maintenance instructions are printed on
the back of the dental machine
18
iM3 Dental Systems
Sterilization
Refer to handpiece operating manual.
HIGH SPEED HANDPIECE
Handpieces should be thoroughly cleaned to remove any foreign particles, eg; blood, saliva etc. Make sure that the turbine cartridge and head cavity are clean before autoclaving. If autoclaved with bloody particles in the turbine housing, autoclaving will cause particles to be scorched and damage to the turbine may result. The exterior of the handpiece may be cleaned with any good surgical disinfectant. Be sure to rinse and dry thoroughly. Lubricate handpiece turbine with MD-30 Lubricant. Refer to handpiece operating manual.
LOW SPEED HANDPIECE
Prepare handpiece and attachments for sterilization by carefully cleaning the exterior surface. Thoroughly scrub, rinse and clean away any residual solution and particles. Remove moisture with a towel or use air from the 3 way air water syringe.
1. Prior to sterilization, lubricate the handpiece using MD-30 Lubricant.
HANDPIECES: Apply oil through the handpiece air inlet tube. (see page 17) ATTACHMENT: Remove attachment from handpiece and oil through both end openings.
Note: After lubricating items, assemble and operate for a few seconds to expel excess lubricant.
2. Seal items in sterilization tubing bags. Make sure to remove bur from handpiece prior to sealing.
3. Place handpiece or attachments in sterilization trays of dry heat sterilizer or steam autoclave. Do not exceed temperatures of 275F or 135C. Sterilize according to manufacturers instructions.
4. After sterilization cycle is completed, allow handpieces and attachments to cool down then lubricate handpiece and attachments thoroughly. Make sure to use a different container after sterilization cycle, to prevent recontaminating.
Note: Lubricate handpieces using MD-30 Lubricant after cooling down, do not allow hand-
piece to sit in sealing bag overnight or for a long period. Immediately operate handpiece as
described in the following step.
5. After lubricating items, insert bur into handpieces and operate for a few seconds to
remove excess lubricant.
Do not autoclave the high speed handpiece with a bur in the chuck.
19
Handpiece Maintenance
All Handpieces have a working life which can be extended by maintaining the correct operating pressures and speeds as well as adequate lubrication.
The working life of most Handpieces turbines is 12-24 months, however, this will depend upon the degree of care taken and the amount of use.
Preventative maintenance will ensure a working Handpiece at all times.
iM3 Dental Systems
Handpiece Daily Weekly Monthly Yearly
iM3 Low Speed Handpiece
Oil in smaller of 2 big holes. Depress the foot pedal for 10 seconds to ensure the oil reaches the Turbine (page 17)
Oil AM or PM before or after use.
2-3 drops
Use iM3 MD-30 Lubricant.
Replace turbine and ‘o’ rings and sealing gasket. If worn
iM3 High Speed Handpiece as
per the instruc­tions for the Low speed above. Page 17.
Oil AM or PM before or after use. 2-3 drops
Oil with MD-30 Lubricant.
Replace turbine
and ‘o’ rings and sealing gasket. If worn
Prophy Head R&R Oscillat­ing
Dispose of prophy head between pa­tients to prevent cross contamina­tion
20
Silent Hurricane Compressor Oil Free
The compressor is perhaps the most crucial part of any dental system as it supplies the valuable air needed to run handpieces and pressurize the water and cooling system.
The regulator/pressure gauge should be set to read 70 p.s.i./5 BAR. The tank pressure gauge should read 110 p.s.i./7.5 BAR. See page 21
Operating the Silent Hurricane Oil Free Compressor
1. Connect the unit to the appropriate voltage.
2. Check that the air tank on/off valve is closed. See page 23.
3. Switch unit on at the mains (power outlet).
4. Turn unit on at switch situated on top of the pressure switch. See page 21.
6. Motor should now run and will stop running when 110 p.s.i./7.5 BAR. has been reached.
7. Connect the grey air line from dental unit to the compressor and turn air tank valve to the ON position. See page 23
Identifying Air Leaks
To ensure the Silent Hurricane Compressor Motor is not subjected to higher than normal
operating temperatures, each iM3 Dental unit should be checked for air leaks at regular intervals, the easiest way to check for air leaks is to pump the compressor tank up to 110 p.s.i./7.5 BAR. and turn the motor off. Make sure the air line to the dental head is connected and the tank valve left in the open position with the pressure switch for the water bottles also turned on.
Once the tank has been pumped to 110 p.s.i./7.5 BAR., check the pressure in the tank 10 minutes later. If the pressure has dropped more than 10% during that time then the unit is leaking air which should be identified and remedied to ensure the compressor motor does not overheat due to excessive running caused by the air leak.
Model: Silent Hurricane Oil Free compressor
Voltage: 220-240 50Hz
900W Maximum operating Pressure 110psi (770KPa)
Manufactured By: iM3 Pty Ltd, Sydney Australia.
iM3 Dental Systems
21
iM3 Dental Systems
iM3 Silent Hurricane Oil Free Compressor
Circuit breaker located on the side of Black electrical box. Push to Reset
Compressor ON/OFF pressure switch I = On O = Off
Outlet pressure regulator
Air to dental machine ON/ OFF tap is on when in line with outlet Pictured in Off/closed posi­tion
Air tank pressure gauge
Regulator pressure gauge
Compressor air tank
Non return valve
Air inlet filter
Oil Free Compressor motor under cover
Tank drain under middle of tank see P.23
Regulator drain bowl
Maintenance Instructions
22
Silent Hurricane Compressor
Adjusting Regulator Air Pressure
With the Dental Unit pressurised and air lines turned ON, the black knob on top of the regulator is pulled up from the locked to the free position to adjust air pressure (see picture page 21). Depress the black knob after setting to lock pressure adjustment.
The regulator gauge registers working or air line pressure which may be increased by turning clockwise and lowered by turning anti clockwise.
The pressure air line gauge will register the working pressure, which should be
70 p.s.i./ 5BAR..
Warning
Unit starts automatically. Isolate power supply and relieve air pressure by opening tank ON/OFF valve prior to commencing non electrical service tasks.
This appliance is not intended for use by persons
(including children) with reduced physical, sensory or
mental capabilities, or lack of experience and knowledge, unless they have
been given supervision or instruction concerning use of the appliance by a
person responsible for their safety.
.
Danger
High voltage used on this equipment. Isolate from mains before removing any covers, which should only be carried out by a qualified electrician . Ensure power outlet is adequately rated for this unit.
iM3 Dental Systems
23
iM3 Dental Systems
Feature Daily Weekly Monthly Yearly
Silent Hurricane Compressor
Turn power off.
Drain moisture from main compressor tank by opening the valve on the top of the white tank marked with a yellow stick­er—Drain. Catch the fluid from the white drain tube into a con-
tainer. Keep the drain open
until all air is removed from tank. See image below
Change air inlet
filter on compres­sor motor.
Regulator ­Auto Drain Filter
Automatic, when tank drained of moisture
Replace regulator
filter.
Regulator ­Air Pressure
Check that correct air pressure is being delivered.
See Adjusting Regulator Air Pressure on Page 22.
Contact iM3 for your yearly compressor service kit.
Silent Hurricane Oil Free Compressor Maintenance
Air ON / OFF (valve pictured in ON or open position)
Compressor tank drain, located on­top of white air tank—open slowly Place container under white tube to
collect water.
Compressor air regulator adjustment knob
Air inlet quick discon­nect line, supplying air to dental machine
24
iM3 Dental Systems
PROBLEM CAUSE SOLUTION
No Water Flow in Handpiece
1. Compressor not ON.
2. Water Flow Control not ON (P4- E).
3. Water/CLS Bottle Empty.
4. Check for a ‘kink’ in the water line.
5. Check if 3 way syringe has Water.
6. Use of tap water, blocking water valves.
7. Low regulator pressure.
1. Turn compressor ON and check that ON/ OFF tap is in the ON position.
2. Turn water control to ON.
3. Fill water/CLS Bottle(s).
4. Check lines and pressure.
5. If yes, lack of water must be in the valves or the handpiece. The 3 way syringe is the quickest way to check for water.
6. Valves will need replacing.
7. Check regulator is set at 70-80 PSI.
Mineral Accumulation
1. Use of tap water - mineral or chemical action blockage will usually take 3-4 years to occur.
2. Verdigrease (Green Slime) build up in brass water valves due to chemical reaction with tap water.
1. Pass a fine wire through the water intake hole. Pass another fine wire through the hole in the handpiece head. Pass the wire as far as possible each way. Blow outthe loosened mineral deposits.
2. Replace valves.
Use of Chlorhexidine 1. Most brands of Chlorhex will de-
posit out of solution and block valves, etc. CLS is designed to be used in its diluted strength without depositing.
NOTE: Do not dilute CLS beyond
its label recommendations or it could result in deposits.
2. Chlorine in tap water will cause a flocculation and block filters and valves
1. Contact your local iM3 Official Service Center.
2. Contact your local iM3 Official Service Center. Use only distilled water
Handpiece Trouble Shooting
High Speed Handpiece
Cartridge Toggle
Toggle switches become floppy
All iM3 dental units use a unique cartridge system for the toggle switches. The toggles are used to pressurize the water/CLS bottles and select the low or high speed handpiece. When the o-rings be­come worn you simply push the cartridge in and turn a 1/4 turn anti clockwise. The cartridge can then be removed from the valve casing. To install a new toggle, locate the desired position, push and rotate a 1/4 turn clockwise. New toggles can be ordered from iM3 and installed by the user. Toggles can be installed in the 12, 3, 6 and 9 o'clock positions. If the toggle is in the wrong position simply remove and reposition. View this procedure on the iM3 website http://www.compad.com.au/cms/im3vet/pages/ iM_elite_film/222
25
Handpiece Trouble Shooting
Low Speed Handpiece
PROBLEM CAUSE SOLUTION
Handpiece wont run
1. No or low air pressure.
2. Air line improperly sealed.
3. Inferior oil; too much or excess more commonly too little oil or water in the turbine.
4. Broken drive head.
Note new handpieces require a Run in timethis can be up to 20 hours of use
1. Check the air line to be sure the air is turned on and the line is unobstructed.
2. Reconnect air line to
handpiece making sure the
gasket is fully contracted.
3. See lubrication instructions
on page 17.
4. Contact your local iM3 official Service Centre.
If the motor stops spinning place a fine screw driver down the nose and rotate shaft.
Low Speed Handpiece - 4:1 Straight Nose Cone
PROBLEM CAUSE SOLUTION
Cant fit R&R prophy angle
1. Plastic drive shaft broken off internally in Chuck housing.
2. Improper lubrication.
Only fit plastic iM3 R&R prophy
angles Code L7465.
1. Contact your local official iM3 Service Centre.
2. See lubrication instruc­tions, on page 17.
iM3 Inc iM3 Pty Ltd 12414 NE 95th Street 21 Chaplin Drive Vancouver WA 98682 USA Lane Cove NSW 2066 Australia
Tel (360) 254 2981 Fax (360) 254 2940 Tel (02) 9420 5766 Fax (02) 9420 5677
www.im3vet.com