Setting USB ..........................................................................................................................................7
Connecting the USB Cable ................................................................................................................... 8
This instruction manual provides details on the communication interfaces of the IM7580A series, IM7581
series, IM7583 series, IM7585 series and IM7587 series Impedance Analyzer.
In this document, the “instrument” means the IM7580A series, IM7581 series, IM7583 series , IM7585
series and IM7587 series.
1
Safety Information
This manual contains information and warnings essential for safe operation of the instrument and for
maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety
precautions.
Notation
In this manual
WARNING
CAUTION
IMPORTANT
the risk seriousness and the hazard levels are classied as follows.
,
Indicates a potentially hazardous situation that may result in death or serious
injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate
injury to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance
tasks with which the operators must be fully familiar.
Indicates prohibited actions.
Indicates the action which must be performed.
*
[ ]
Indicates the action which must be performed.
Menus, commands, dialogs, buttons in a dialog, and other names on the screen
and the keys are indicated in brackets.
2
3
4
5
6
7
CURSOR
( Bold character)
Windows
DialogueDialogue box represents a Windows dialog box.
Mouse Operation
Click:Press and quickly release the left button of the mouse.
Right-click:Press and quickly release the right button of the mouse.
Double click:Quickly click the left button of the mouse twice.
Drag:
Bold characters within the text indicate operating key labels.
Unless otherwise specied
Windows 8 or Windows 10.
While holding down the left button of the mouse, move the mouse and then release the
left button to deposit the chosen item in the desired position.
“Windows” represents Windows Vista, Windows 7,
,
8
9
10
付録索引
1
Safety Information
2
1
Specications
1.1 RS-232C Specications
Transmission MethodCommunication method: Full duplex
• When the data in the receive buffer exceeds 75% of the
buffer, XOFF(13H) is sent and the controller is notied
that there is not much space remaining in the buffer.
• Processing of the data in the buffer continues, and then
XON(11H) is sent and the controller is notied that there
is sufcient remaining space in the buffer when the
amount of data becomes less than 25%.
Control during Sending
XON/XOFF control:
• When XOFF is received, the sending of data is halted. When XON is received, the sending of data is
resumed.
Input Buffer
Usage
Amount
85%
75%
25%
Buffer Space
Send XOFF
Send XON
3
GP-IB Specications
1.2 GP-IB Specications
SH1Supports all source handshake functions.
AH1Supports all acceptor handshake functions.
T6Supports standard talker functions.
Supports serial poll functions.
Talk only mode is not supported.
Supports the talker cancel function by MLA (My Listen Address).
L4Supports standard listener functions.
Listener only mode is not supported.
Supports the listener cancel function by MTA (My Talk Address).
SR1Supports all service request functions.
RL1Supports all remote/local functions.
PP0Parallel poll functions are not supported.
DC1Supports all device clear functions.
DT1Supports all device trigger functions.
C0Controller functions are not supported.
Code used: ASCII code
1.3 USB Specications
ConnectorSeries B receptacle
Compliance tandardUSB2.0 (Full Speed/High Speed)
No. of ports1
ClassCommunication class
Supported OSWindows Vista, Windows 7, Windows 8, Windows 10
1.4 LAN Specications
ConnectorRJ-45 connector × 1
Compliance
standard
IEEE 802.3-compliant Ethernet
Transfer system10BASE-T/ 100BASE-TX/ 1000BASE-T Auto detected
ProtocolTCP/IP
FunctionCommand control
4
2
Connection and Setting
2.1 Overview of Communication
You can control the instrument with communication commands from a computer via the USB, GP-IB, RS232C and LAN interfaces.
There are the following four communication methods. To enable communication, the communication
conditions need to be set on the instrument.
USB communication (p. 7)
The instrument is communication class compatible.
LAN communication (p. 9)
2
Command control using the TCP/IP protocol is possible.
GP-IB communication (when the Z3000 is connected) (p. 14)
• Commands common to IEEE-488-2 1987 (requirement) can be used.
• The instrument has been designed with reference to the following standard.
(Reference standard: IEEE-488.2 1987)
RS-232C communication (when the Z3001 is connected) (p. 16)
WARNING
• Always turn both devices OFF when connecting and disconnecting an interface
connector. Otherwise, an electric shock accident may occur.
• To avoid damage to the instrument, do not short-circuit the terminal and do not
input voltage to the terminal.
• Failure to fasten the connectors properly may result is sub-specication
performance or damage to the equipment.
5
Overview of Communication
Screen Displayed while Setting Interfaces
When you set an interface, the icon for the set interface is displayed on the right side of the screen.
When USB is set
When GP-IB is set
When RS-232C is set
When LAN is set
6
2.2 USB Settings and Connection
USB Settings and Connection
To connect the instrument to a computer the rst time
Before connecting the instrument to the computer, install the USB driver.
The USB driver can be downloaded from the bundled CD, or our web site. (
The USB driver is compatible with the Windows Vista (32-bit, 64-bit version), Windows 7 (32-bit,64-bit
version), Windows 8 (32-bit, 64-bit version) and Windows 10 (32-bit, 64-bit version) operating systems.
Additionally, do not put the computer into the sleep state while the instrument is connected to the
computer.
a dedicated USB driver must be installed.
,
http://www.hioki.com
Setting USB
Procedure
You can congure the setting from any of mode and mode.
1
LCR Measurement Screen
Interface Settings
)
2
2
USB Setting
Press .
7
USB Settings and Connection
3
USB Setting
Select the terminator setting.
CR+LF
CR
4
Press to conrm the setting.
Connecting the USB Cable
Connect a USB cable (commercially available USB cable) to the USB port of the instrument.
Type B
USB cable (commercially
available product)
USB interface port of
computer
CAUTION
• To avoid faults, do not disconnect or reconnect the USB cable during instrument
operation.
• Connect the instrument and the computer to a common earth ground. Using different
grounds could result in potential difference between the instrument and the computer.
Potential difference on the USB cable can result in malfunctions and faults.
• Do not disconnect or connect other USB devices from or to, respectively, PC during
USB communications between instrument and computer. The USB communications
between instrument and PC may stop.
8
LAN Settings and Connection
2.3 LAN Settings and Connection
LAN Settings
You can perform command control using the TCP/IP protocol.
Set the instrument to match your network environment in advance.
• Make these settings before connecting to a network. Changing settings while connected can duplicate
IP addresses of other network devices, and incorrect address information may otherwise be presented
to the network.
• The instrument does not support DHCP (automatic IP address assignment) on a network.
Setting Items
Identies each device connected on a network.
IP address
Subnet mask
Each network device must be set to a unique address.
The instrument supports IP version 4, with IP addresses indicated as four decimal octets,
e.g.,“192.168.0.1”.
This setting is for separating the IP address into the network address that indicates the
network and the host address that indicates the instrument. On this instrument, the subnet
mask is represented as four decimal numbers separated by “. “ such as “255.255.255.0.”
2
When the computer and instrument are on different but overlapping networks (subnets), this
Default Gateway
Network Environment Conguration
Example 1. Connecting the instrument to an existing network
Example 1. Connecting the instrument to an existing network
When connecting the instrument to an existing network
in advance.
An IP address which is not the same as that of another network device needs to be assigned.
Conrm the following items with the network administrator
IP Address _________._________._________._________
Subnet Mask _________._________._________._________
Default Gateway _________._________._________._________
Example 2. Connecting multiple instruments to a single computer using a hub
When building a local network with no outside connection, the following private IP addresses are
recommended.
Example of private IP address:
IP Address.....................Computer: 192.168.0.100
Subnet Mask..................255.255.255.0
Default Gateway.............OFF(0.0.0.0)...
Example 3. Connecting one instrument to a single computer using the 9642 LAN Cable
The 9642 LAN Cable can be used with its supplied connection adapter to connect one instrument
to one computer, in which case the IP address is freely settable. Use the recommended private IP
addresses.
IP Address.....................Computer: 192.168.0.100
Subnet Mask..................255.255.255.0
Default Gateway....OFF(0.0.0.0)
IP address species the device to serve as the gateway between the networks.
If the computer and instrument are connected one-to-one, no gateway is used, and the
instrument’s default setting “0.0.0.0” can be kept as is.
the network settings need to be conrmed
,
and write them down.
,
Instrument: 192.168.0.1, 192.168.0.2, 192.168.0.3...
(Set an IP address that differs from that of other network devices.)
Instrument: 192.168.0.1 (Set to a different IP address than the computer.)
9
LAN Settings and Connection
Procedure
You can congure the setting from any of mode and mode.
1
2
LCR Measurement Screen
LAN Settings
Interface Settings
Press .
3
4
LAN Settings
Select the IP address.
IP address Settings
or to set the IP address.
Press
to conrm the setting.
10
5
LAN Settings and Connection
LAN Settings
Subnet mask Settings
6
Any of the following 30 subnet masks can be set for the instrument.
If the default gateway does not need to
be set, for example, when connecting the
instrument and computer on a one-to-one
basis using a cross cable, leave this set
to OFF.
11
LAN Settings and Connection
Default gateway Settings
8
9
LAN Settings
or to set the default gateway.
Press
to conrm the setting.
10
11
Select the port number.
Port number Settings
or to set the port number to
use for communication commands.
Settable range : 1024 to 65535
Press to conrm the setting.
LAN Settings
12
12
Press to conrm the setting.
Select the terminator setting.
CR+LF
CR
LAN Settings and Connection
Connecting a LAN Cable
Use a LAN cable to connect the instrument and computer.
Required items:
When connecting the instrument to an existing network (prepare any of the following):
• Straight-through Cat 5, 1000BASE-T-compliant Ethernet cable (up to 100 m, commercially available).
For 100BASE/10BASE communication, a 100BASE-TX/10BASE-T-compliant cable may also be used.
• Hioki 9642 LAN Cable (option)
(A cross adapter cannot be used.)
When connecting one instrument to a single computer (prepare one of the following):
• 1000BASE-T-compliant cross-over cable (up to 100 m)
• 1000BASE-T-compliant straight-through cable with cross-over adapter (up to 100 m)
• Hioki 9642 LAN Cable (option)
2
LINK UP LED
SPEED LED
On (Green) :1000BASE
On (Orange) :100BASE
Off:10BASE
The MAC address of the LAN is displayed below the serial number.
You can also check it on the instrument screen.
See: “Checking the Version of the Instrument” in the instruction manual.
When connecting the instrument to a single computer (connect the instrument to the computer)
Connecting with the 9642 LAN Cable and crossover adapter (supplied with the 9642)
Crossover adapter
LAN interface
Connect the LAN Cable to the
1
crossover adapter.
Connect the crossover adapter
2
to the LAN interface on the
instrument.
On: Link up
Off: Link down
Blinking: Sending or
receiving data
Connect the LAN Cable to the
3
100BASE-TX connector on the PC.
13
GP-IB Connection and Settings (when the Z3000 is connected)
2.4 GP-IB Connection and Settings
(when the Z3000 is connected)
Connecting the GP-IB Cable
Connect the GP-IB cable to the GP-IB connector.
Recommended cable:
9150-02 GP-IB connection cable (2 m)
Setting GP-IB
Procedure
You can congure the setting from any of mode and mode.
1
2
LCR Measurement Screen
GPIB Setting
Interface Settings
14
Press .
3
GP-IB Connection and Settings (when the Z3000 is connected)
GPIB Setting
Use or to set the GP-IB address.
Select the terminator setting.
4
Press to conrm the setting.
LF with EOI
LF with CR+EOI
2
15
RS-232C Connection and Settings (when the Z3001 is connected)
2.5 RS-232C Connection and Settings
(when the Z3001 is connected)
Connecting the RS-232C Cable
Connect the RS-232C cable to the RS-232C connector.
(Recommended cable: 9637 RS-232C cable)
Connector (D-sub)
Pin No.
1Unused
2Received Data104 BB RD RxD
3Transmitted Data 103 BA SD TxD
4
5Signal Ground102 AB SG GND
6Unused
7Unused
8Unused
9Unused
1 2 3 4 5
To connect the instrument to a controller (DTE), use a crossover cable compatible
with the connectors on both the instrument and the controller.
The I/O connector is a DTE (Data Terminal Equipment) conguration.
Interchange Circuit
Nam
e
Data Terminal Ready
6 7 8 9
Male 9-pin D-sub
#4-40 attaching
screws
CCITT
Circuit No.
108/2 CD ER DTR
EIA
Abbreviation
JIS
Abbreviation
Common
Abbreviation
Example: Connecting to a DOS/V PC
Specication: D-sub 9-pin female and female connector
reverse connection
,
16
RS-232C Connection and Settings (when the Z3001 is connected)
Setting RS-232C
Procedure
You can congure the setting from any of mode and mode.
1
2
LCR Measurement Screen
Interface Settings
2
RS-232C Settings
Press .
3
4
Press to conrm the setting.
RS-232C Settings
Select the baud rate setting.
Select the handshake setting.
No ow control
Software
(XON/XOFF control)
Select the terminator setting.
CR+LF
CR
17
Remote Mode
2.6 Remote Mode
When you connect a device to an interface and start communication, the mode becomes remote mode
(remote operation state) and the keys on the LCD are disabled.
Remote Mode State
Canceling Remote Mode
Procedure
Local State
1
All of the keys except are disabled.
Press
(local state).
to return to the normal state
18
2
LCR Measurement Screen
The measurement screen is redisplayed.
Appendix
Appx. 1 Checking the USB Virtual COM Port
The instrument’s USB interface supports communications-class performance, allowing control operations
on par with RS-232C to be performed from a computer. When you connect the instrument to a computer
and set its interface to USB, it will be recognized as a virtual COM port on the computer.
Device Manager starts.
1
2
3
4
5
Check the COM number to the right of “HIOKI USB Device” in the
The procedure to start Device Manager differs depending on the version of the Windows
operating system.
For details, refer to Help of the operating system.
[Ports (COM & LPT)]
6
7
list.
8
9
10
付録索引
付
1
Checking the USB Virtual COM Port
付
2
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.