Hanna Instruments PH 210 User Manual

4 (1)

HANNA LITERATURE

Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as:

Water Treatment

Process

Swimming Pools

Agriculture

Food

Laboratory

Thermometry

and many others. New reference material is constantly being added to the library.

For these and other catalogs, handbooks and leaflets, contact your dealer or the Hanna Customer Service Center nearest to you. To find the Hanna Office in your vicinity, check our home page at www.hannainst.com.

Instruction Manual

pH 210

pH 211

pH 212

pH 213

Microprocessor-based

Bench pH/mV/°C Meters

02/00 MANPH213R2

 

 

 

 

 

 

These Instruments are in

 

 

 

 

 

 

 

 

 

 

 

 

h t t p : / / w w w . h a n n a i n s t . c o m

Manufacturers since 1978

Compliance with the CE Directives

Dear Customer,

Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument.

This manual will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility.

If you need more technical information, do not hesitate to e-mail us at tech@hannainst.com.

These instruments are in compliance with directives.

TABLE OF CONTENTS

 

PRELIMINARY EXAMINATION ......................................................

3

GENERAL DESCRIPTION .............................................................

3

FUNCTIONAL DESCRIPTION pH 210 AND pH 211 .......................

4

SPECIFICATIONS pH 210 AND pH 211 ......................................

5

FUNCTIONAL DESCRIPTION pH 212 AND pH 213 .......................

6

SPECIFICATIONS pH 212 AND pH 213 ......................................

7

OPERATIONAL GUIDE ................................................................

8

pH CALIBRATION ...................................................................

11

TEMPERATURE CALIBRATION (pH 212 AND pH 213 ONLY) ........

14

mV CALIBRATION (pH 213 ONLY) ...........................................

15

INTERFACE WITH PC (pH 212 AND pH 213 ONLY) ....................

16

pH VALUES AT VARIOUS TEMPERATURES .................................

18

ELECTRODE CONDITIONING & MAINTENANCE .............................

19

TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS ..........

22

TROUBLESHOOTING GUIDE .....................................................

23

ACCESSORIES .........................................................................

24

WARRANTY ............................................................................

30

CE DECLARATION OF CONFORMITY ...........................................

31

CE DECLARATION OF CONFORMITY

Recommendations for Users

Before using this product, make sure that it is entirely suitable for the environment in which it is used.

Operation of this instrument in residential areas could cause unacceptable interferences to radio and TV equipment, requiring the operator to take all necessary steps to correct interferences.

The glass bulb at the end of the pH electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.

During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges.

Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance.

To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24VAC or 60VDC.

To avoid damages or burns, do not perform any measurement in microwave ovens.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

31

WARRANTY

All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are guaranteed for a period of six months. This warranty is limited to repair or replacement free of charge.

Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered.

If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.

To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., 584 Park East Drive, Woonsocket, Rhode Island, 02895 , USA.

Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Custumer Service Center.

Each meter comes supplied complete with:

HI 1131B glass-body combination pH Electrode with 1 m (3.3') cable

HI 7669/2W Temperature Probe

HI 76404 Electrode Holder

pH 4.01 & 7.01 Buffer Solutions (20mL each)

HI 7071S Electrolyte Solution

12VDC Power Adapter

Instruction Manual

Note: Save all packing material until you are sure that the instrument functions correctly. All defective item must be returned in the original packing with the supplied accessories.

GENERAL DESCRIPTION

The Hanna pH 210, pH 211, pH 212 and pH 213 are micropro- cessor-based bench meters for pH and temperature measurements.

pH 211 and pH 213 can also measure ion concentration (ISE) and ORP (Oxidation Reduction Potential) in the mV range.

The pH measurements are compensated for temperature effect manually or automatically with the HI 7669/2W temperature probe.

The meter comes equipped with a large easy-to-read LCD which shows the pH (or mV) and temperature simultaneously together with graphic symbols.

The calibration process is guided step by step through clear indications on the LCD. A stability indicator make the calibration procedure errorfree.

Through the RS 232C serial port you can transfer the data to a PC (pH 212 and pH 213 only).

30

3

Hanna Instruments PH 210 User Manual

FUNCTIONAL DESCRIPTION pH 210 AND pH 211

1)Liquid Crystal Display (LCD)

2)CFM key, to confirm calibration values

3)CAL key, to enter or exit calibration mode

4)W C key, to manually decrease temperature or select pH buffer

5)V C key, to manually increase temperature or select pH buffer

6)MR key, to recall the stored value

7)MEM key, to store a value in memory

8)RANGE key, to select measurement range (pH 211 only)

9)ON/OFF switch

10)Power adapter socket

11)BNC electrode connector

12)Electrode reference socket

13)Temperature probe socket

14)RESET button

Extension cables for screw-type electrodes (screw to BNC adapter)

HI 7855/1 Extension cable 1m (3.3') long HI 7855/3 Extension cable 3m (9.9') long

HI 7855 SERIES CABLE CONNECTORS

CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC

CONNECT TO

CONNECT TO THE

SCREW TYPE

BNC SOCKET

ELECTRODES

OF THE METER

OTHER ACCESSORIES

HI 710005 Voltage adapter from 115 VAC to 12 VDC (USA plug)

HI 710006 Voltage adapter from 230 VAC to 12 VDC (European plug)

HI 710012 Voltage adapter from 240 VAC to 12 VDC (UK plug)

HI 710013 Voltage adapter from 230 VAC to 12 VDC (South Africa plug)

HI 710014 Voltage adapter from 230 VAC to 12 VDC (Australia plug)

ChecktempC Pocket-size thermometer (range -50.0 to 150.0 C) HI 76405 Electrode holder

HI 8427 pH and ORP electrode simulator with 1 m (3.3') coaxial cable ending in female BNC connectors

HI 931001 pH and ORP electrode simulator with LCD and 1 m (3.3') coaxial cable ending in female BNC connectors

HI 7669/2W Temp. probe with 1 m (3.3') cable HI 92000 Windows® compatible software

HI 920010 9 to 9-pin RS232 cable

Windows® is registered Trademark of "Microsoft Co."

4

29

HI 1413B

Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement.

12 mm

0.5"

HI 1413

110 mm

4.3"

ORP ELECTRODES

HI 3131B

Glass-body, refillable, combination platinum ORP electrode. Use: titration.

 

12 mm

 

0.5"

HI 3131

 

 

150 mm

"S" VERSION

5.9"

HI 3230B

Plastic-body (Ultem®), gel-filled, combination platinum ORP electrode. Use: general purpose.

 

12 mm

 

0.5"

HI 3230

 

 

120 mm

"S" VERSION

4.7"

 

HI 4430B

Plastic-body (Ultem®), gel-filled, combination gold ORP electrode. Use: general purpose.

 

12 mm

 

0.5"

HI 4430

 

 

120 mm

"S" VERSION

4.7"

 

Consult the Hanna General Catalog for more electrodes with screwtype or BNC connectors.

Ultem® is registered Trademark of "General Electrics Co."

Kynar® is registered Trademark of "Pennwalt Corp."

SPECIFICATIONS pH 210 AND pH 211

Range

pH

0.00 to 14.00

 

mV

±399.9 (ISE, pH 211 only)

 

mV

±1999 (ORP, pH 211 only)

 

°C

0.0 to 100.0

 

 

 

Resolution

pH

0.01

 

mV

0.1 (ISE, pH 211 only)

 

mV

1 (ORP, pH 211 only)

 

°C

0.1

Accuracy

pH

±0.01

(@20°C/68°F)

mV

±0.2 mV (ISE, pH 211 only)

 

mV

±1mV (ORP, pH 211 only)

 

°C

±0.5

Typical EMC

pH

±0.03

Deviation

mV

±2 mV (pH 211 only)

 

°C

±0.3

pH Calibration

 

Automatic 1 or 2 points with 5 memorized buffers

 

 

(4.01, 6.86, 7.01, 9.18, 10.01)

 

 

 

Temperature

 

Automatic or manual

Compensation

 

from0to100 C(32to212 F)

Electrode

 

HI 1131B glass-body combination

 

 

refillable pH electrode (included)

Temperature probe

 

HI 7669/2W (included)

 

 

 

Input impedance

 

1012

Power supply

 

12 VDC adapter (included)

 

 

 

Environment

 

0to50 C(32to122 F)

 

 

max. 95% RH non-condensing

 

 

 

Dimensions

 

240x182x74 mm (9.4x7.1x2.9")

Weight

 

1.1 kg (2.5 lb.); kit with holder: 2.5 kg (5.5 lb.)

28

5