Instruction Manual |
|
DISPLAY FUNCTIONAL DESCRIPTION |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
HI 99300 • HI 99301
Portable Instruments
for EC/TDS/°C/°F measurements
with advanced features
01/05 IST99300R1
w w w . h a n n a i n s t . c o m
HI 99300 and HI 99301 come supplied complete with HI 76306 probe, 4 alkaline batteries (1.5V AAA), instructions and rugged carrying case.
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The probes are warranted for six months.
This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization
Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
SPECIFICATIONS
|
|
HI 99300 |
HI 99301 |
Range |
EC |
0 to 3999 µS/cm |
0.00 to 20.00 mS/cm |
|
TDS |
0 to 2000 ppm (mg/l) |
0.00 to 10.00 ppt (g/l) |
Temperature |
0.0 to 60.0°C / 32.0 to 140.0°F |
||
|
|
|
|
Resolution |
EC |
1 µS/cm |
0.01 mS/cm |
|
TDS |
1 ppm (mg/l) |
0.01 ppt (g/l) |
Temperature |
0.1°C / 0.1°F |
Accuracy (@20°C/68°F) |
±2% F.S. (EC/TDS) |
|
±0.5°C / ±1°F |
Typical EMC Deviation ±2% F.S. (EC/TDS) / ±0.5°C / ±1°F
Temperature |
Automatic, from 0 to 60°C |
Compensation |
β adjustable from 0.0 to 2.4%/°C (BETA) |
|
|
Calibration |
Automatic, single point |
EC/TDS Factor |
Selectable from 0.45 to 1.00 (CONV) |
Probe (included) HI 76306 with built-in temperature sensor,
|
DIN connector and 1 m cable |
Environment |
0 to 50°C (32 to 122°F); RH 100% |
|
|
Power Supply |
4 alkaline batteries, 1.5V AAA with BEPS; |
|
500 hours of continuous use |
|
|
Dimensions |
143x80x38 mm (5.6x3.2x1.5’’) |
|
|
Weight |
320 g (11.3 oz.) |
HI 99300 and HI 99301 are in compliance with the directives.