Instruction Manual
HI 92701 - HI 92704
HI 92801 - HI 92804
Portable Microprocessor
Printing and Logging
K-type Thermocouple
Thermometers
|
|
|
|
These Instruments are in |
|
|
|
|
|
||
|
|
|
|
||
http://www.hannainst.com |
|||||
Compliance with the CE Directives |
2
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications.
These instruments are in compliance with |
directives (EN 50081- |
|
1 and EN50082-1). |
|
|
TABLE OF CONTENTS |
|
|
PRELIMINARY EXAMINATION ........................................................ |
|
3 |
GENERAL DESCRIPTION ................................................................ |
|
3 |
LCD DISPLAY FUNCTIONAL DESCRIPTION ...................................... |
|
4 |
FUNCTIONAL DESCRIPTION HI 92701 AND HI 92704 .................. |
5 |
|
FUNCTIONAL DESCRIPTION HI 92801 AND HI 92804 .................. |
6 |
|
SPECIFICATIONS HI 92701 AND HI 92704 .................................. |
|
7 |
SPECIFICATIONS HI 92801 AND HI 92804 .................................. |
|
7 |
OPERATIONAL GUIDE ................................................................... |
|
8 |
PRINTING / RECORDING WITH HI 92701 AND HI 92704 ........... |
13 |
|
PRINTING / LOGGING WITH HI 92801 AND HI 92804 .............. |
15 |
|
DATA TRANSFER TO PC .............................................................. |
|
24 |
FAULT FUNCTIONS ..................................................................... |
|
25 |
MEMORY ORGANIZATION ........................................................... |
|
25 |
PRINTER MAINTENANCE ............................................................. |
|
26 |
BATTERY REPLACEMENT .............................................................. |
|
28 |
CALIBRATION ............................................................................. |
|
29 |
ACCESSORIES ............................................................................. |
|
30 |
WARRANTY ................................................................................ |
|
34 |
CE DECLARATION OF CONFORMITY .............................................. |
|
35 |
ISO 9000 Certified
Company since 1992
2
Untitled-27
3
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.
Each printing/logging thermometer is supplied complete with:
•Paper Rolls (5 pcs)
•1.5V AA size Alkaline Batteries (4 pcs)
•Rugged Carrying Case
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in its original packaging together with the supplied accessories.
GENERAL DESCRIPTION
The Hanna Instruments portable K-type thermocouple thermometers with built-in printers are microprocessor-based and enable you to accurately measure temperature and record data.
The instrument housing is made of rugged and lightweight material. Equipped with a large, easy-to-read LCD, they feature an extended battery life and a special design that enables results to be obtained even in humid, wet or dusty conditions.
Measurements can be performed with lab-grade precision in the field as well as in the laboratory without compromising accuracy. A 12VDC car battery or a battery charger can also be used to power the unit for extended use.
HI 92801 and HI 92804 have a memory capacity of 16,000 individual temperature readings with recording intervals selected by the user.
Using the HI 9200 infrared transmitter, all the recorded data can be transferred to a PC for easy reproduction, storage or elaboration without the interference of cables and cords between the meter and the receiver.
Other features include: on-board clock and date (HI 92801 and HI 92804 only), user-selectable printing/logging intervals, automatic memorization of measured values (HI 92801 and HI 92804 only), use of pre-calibrated completely interchangeable K-type thermocouple probes, printing on plain paper, automatic shut-off capability, low
3
4
battery warning and 2 year warranty. Two versions are available:
HI 92701C, HI 92704C, HI92801C and HI92804C perform temperature readings in the Centigrade scale;
HI 92701F, HI 92704F, HI92801F and HI92804F perform temperature readings in the Fahrenheit scale.
LCD DISPLAY
FUNCTIONAL DESCRIPTION
1
°C
DATE
TIME 2
LOG |
INTV |
LO BAT |
1.Primary Display
2.Secondary Display
4
Untitled-27
5
FUNCTIONAL DESCRIPTION
HI 92701 AND HI 92704
1.Temperature Probe Connector/s
2.Power adapter plug
3.LCD Display
4.PAPER key to pull the paper
5.ON/OFF key, to turn the meter on or off
6.INTV key to select the printing interval
7.0.1/1 key, to select the resolution
8.UP key to set up
9.DOWN key to set up
10.PRINT key to obtain a printout (printing present time and temperature/s)
11.LOG key to enter the recording mode
5
6
FUNCTIONAL DESCRIPTION
HI 92801 AND HI 92804
1.Temperature Probe Connector/s
2.Power adapter plug
3.LCD Display
4.PAPER key to pull the paper
5.PRINT key to obtain a printout (printing present time and temperature/s)
6.ALT key, alternate function key
7.TEMP key, to read temperature and to reactivate the meter when is "sleeping"
8.UP key to scan data or to set up
9.DOWN key to scan data or to set up
10.TIME key to display present time and printing interval
11.LOG key to enter and exit the logging mode
6
Untitled-27
7
SPECIFICATIONS
HI 92701 AND HI 92704
|
HI 92701C |
HI 92701F |
|
HI 92704C |
HI 92704F |
|
|
|
Range |
-200.0 to 999.9°C |
-300.0 to 999.9°F |
|
1000 to 1370°C |
1000 to 2500°F |
Resolution |
0.1/1°C |
0.2/1°F |
Accuracy |
±0.3% of reading |
±0.3% of reading |
(@25°C/77°F) |
or ±0.5°C |
or ±1°F |
|
for one year, excluding probe error |
|
|
|
|
Typical EMC Deviation |
±5°C |
±10°F |
|
|
|
Power Supply |
4x1.5V AA size, 500 |
hours with 60' printing |
|
interval. Socket for 12VDC adapter |
|
Auto Shut-off |
After 5 minutes of non-use |
|
Channels |
1 for HI 92701; 4 for HI 92704 |
|
Printer |
Low-power impact belt type, 14 characters per line |
|
|
using 38 mm plain paper |
|
Printing Intervals |
1, 2, 5, 10, 15, 30, 60, 120 and 180 minutes |
|
Environment |
0 to 50 °C (32 to 122°F) Max. 95% RH |
|
Dimensions |
220 x 82 x 66 mm (8.7 x 3.2 x 2.6") |
|
Shipping |
Instrument: 500g (18 oz.) |
|
Weight |
Kit: 1.4 kg (3.1 lb.) |
SPECIFICATIONS
HI 92801 AND HI 92804
|
HI 92801C |
HI 92801F |
|
HI 92804C |
HI 92804F |
|
|
|
Range |
-200.0 to 999.9°C |
-300.0 to 999.9°F |
|
1000 to 1370°C |
1000 to 2500°F |
Resolution |
0.1/1°C |
0.2/1°F |
Accuracy |
±0.3% of reading |
±0.3% of reading |
(@25°C/77°F) |
or ±0.5°C |
or ±1°F |
|
for one year, excluding probe error |
|
|
|
|
Typical EMC Deviation |
±5°C |
±10°F |
Power Supply |
4x1.5V AA size, 500 |
hours with 60' printing |
|
interval. Socket for 12VDC adapter |
|
Auto Shut-off |
After 5 minutes of non-use |
|
|
|
|
Channels |
1 for HI 92801; 4 for HI 92804 |
|
Printer |
Low-power impact belt type, 14 characters per line |
|
|
using 38 mm plain paper |
|
Printing/Logging Intervals |
1, 2, 5, 10, 15, 30, 60, 120 and 180 minutes |
|
Environment |
0 to 50 °C (32 to 122°F) Max. 95% RH |
|
Dimensions |
220 x 82 x 66 mm (8.7 x 3.2 x 2.6") |
|
Shipping |
Instrument: 500g (18 oz.) |
|
Weight |
Kit: 1.4 kg (3.1 lb.) |
|
|
7 |
|
8
OPERATIONAL GUIDE
INITIAL PREPARATION
Each meter is supplied complete with batteries. Remove the back cover, unwrap the batteries and install them while paying attention to the polarity.
To prepare the instrument for use, connect a K-type thermocouple temperature probe to the connector located on the top of the instrument.
To switch the HI92701 and the HI92704 on, |
|
press the ON/OFF key. |
ON/OFF |
To switch the HI92801 and the HI92804 on, press the TEMP key.
TEMP
Choose the most appropriate temperature probe(s) for your application (see page 30) and connect it (them) to the meter.
HI 92701 and HI 92801 have a single probe connection. HI 92704 and HI 92804 have four probe connectors.
With the meter facing you, channel #1 is the first connector on the top left hand side.
If the temperature probe is not connected to the instrument, the meter will display and print "----" to alert the user.
This also indicates the possibility of a broken probe cable. The channel is displayed with "CH" and the number of channel (four channels versions only).
°C
CH
Note: To ensure accurate readings it is recommended that the temperature probes be connected to consecutive channels beginning with channel #1. For example, if only one probe is to be used, it must be connected to channel #1. If 2, 3 or 4 probes are to be used, they must be connected to channels 1, 2, 3, and 4 in ascending order.
8
Untitled-27
9
To maximize battery life, the display is automatically switched off after 5 minutes of non-use. However, the meter will continue to monitor (if in the logging/recording mode) temperature.
To revive the display, press the ON/OFF key (HI92701 and HI92704 only) or the TEMP key (HI92801 and HI92804 only).
ON/OFF |
TEMP |
SETTING THE PRINTING INTERVAL (FOR HI92701 AND HI92704 ONLY)
Turn the |
instrument on by pressing the ON/ |
OFF key. |
ON/OFF |
Press the INTV and the UP or DOWN keys simultaneously.
UP
OR |
|
|
INTV |
|
|
|
DOWN |
|
The display will show the log number. At |
|
|
the bottom of the display the printing |
|
|
interval will be flashing. |
|
|
Keep the INTV key pressed and set the |
INTV |
|
printing interval by pressing the UP or |
||
|
||
DOWN key. |
|
SETTING THE MEASUREMENT RESOLUTION (FOR HI 92701 AND HI 92704 ONLY)
Press the 0.1/1 key to select the 0.1° or 1° resolution.
°C
0.1/1 CH
With four channels versions, all four channels readings automatically change by pressing this key.
To return the reading to its original resolution, simply press the 0.1/ 1 key.
9
10
SETTING DATE / TIME / PRINTING INTERVAL (FOR HI92801 AND HI92804 ONLY)
Turn the instrument on by pressing the TEMP |
|
key. |
TEMP |
Press the ALT and the TIME keys simultaneously. The display will show the date setting. At the bottom of the display the year will be blinking.
ALT |
UP |
DATE |
TEMP TIME
Use the UP or the DOWN key to select the year.
UP |
DOWN |
When the correct year is selected, press the TIME key once. The left digit of the display blinks, showing the month.
DATE
TIME
Select the month by using the UP or the DOWN key.
UP |
DOWN |
Press the TIME key, the right hand side of the display shows the day.
DATE
TIME
Use the UP or the DOWN key to select the correct day.
UP |
DOWN |
10
Untitled-27
11
Press the ALT and the TIME keys simultaneously and the display will now show the clock time with the printing interval blinking.
ALT |
UP |
TIME |
|
|
INTV |
TEMP TIME
Any interval can be selected from 1, 2, 5, 10, 15, 30, 60, 120 or 180 minutes by using the UP and the DOWN key.
UP |
DOWN |
Once the desired interval is selected press the TIME key once to set it, the left digits of the display will start blinking.
TIME
INTV
TIME
To select the hour, press the UP or the DOWN key (24 hour clock).
UP |
DOWN |
To set the hour press the TIME key once, the minutes start blinking.
TIME
INTV
TIME
Use the UP or the DOWN key to select the minutes.
UP |
DOWN |
Press the ALT and the TIME keys to leave this mode.
ALT |
UP |
TEMP TIME
The time, date and printing interval are now set and stored in the
memory even when the display is switched off. 11
12
TO VIEW DATE / TIME / TEMPERATURE (FOR HI 92801 AND HI 92804 ONLY)
To view the time press the TIME key. This also displays the selected interval time on the secondary LCD.
TIME
INTV
TIME
To view the date, press the UP key when the LCD is displaying time.
DATE
UP
To view temperature press TEMP key.
°C
CHANNEL
CH
TEMP
Using UP or DOWN key select which channel to monitor (four channels versions only).
UP DOWN
If "----" is displayed, it indicates that the temperature probe is not connected.
°C
CH
This also indicates the possibility of a broken probe cable.
The channel is displayed with "CH" and the number of channel (four channels versions only).
12