Hanna Instruments HI 92801 User Manual

5 (1)

Untitled-27

1

Instruction Manual

HI 92701 - HI 92704

HI 92801 - HI 92804

Portable Microprocessor

Printing and Logging

K-type Thermocouple

Thermometers

 

 

 

 

These Instruments are in

 

 

 

 

 

 

 

 

http://www.hannainst.com

Compliance with the CE Directives

2

Dear Customer,

Thank you for choosing a Hanna Instruments Product.

Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications.

These instruments are in compliance with

directives (EN 50081-

1 and EN50082-1).

 

 

TABLE OF CONTENTS

 

PRELIMINARY EXAMINATION ........................................................

 

3

GENERAL DESCRIPTION ................................................................

 

3

LCD DISPLAY FUNCTIONAL DESCRIPTION ......................................

 

4

FUNCTIONAL DESCRIPTION HI 92701 AND HI 92704 ..................

5

FUNCTIONAL DESCRIPTION HI 92801 AND HI 92804 ..................

6

SPECIFICATIONS HI 92701 AND HI 92704 ..................................

 

7

SPECIFICATIONS HI 92801 AND HI 92804 ..................................

 

7

OPERATIONAL GUIDE ...................................................................

 

8

PRINTING / RECORDING WITH HI 92701 AND HI 92704 ...........

13

PRINTING / LOGGING WITH HI 92801 AND HI 92804 ..............

15

DATA TRANSFER TO PC ..............................................................

 

24

FAULT FUNCTIONS .....................................................................

 

25

MEMORY ORGANIZATION ...........................................................

 

25

PRINTER MAINTENANCE .............................................................

 

26

BATTERY REPLACEMENT ..............................................................

 

28

CALIBRATION .............................................................................

 

29

ACCESSORIES .............................................................................

 

30

WARRANTY ................................................................................

 

34

CE DECLARATION OF CONFORMITY ..............................................

 

35

ISO 9000 Certified

Company since 1992

2

Untitled-27

3

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.

Each printing/logging thermometer is supplied complete with:

Paper Rolls (5 pcs)

1.5V AA size Alkaline Batteries (4 pcs)

Rugged Carrying Case

Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in its original packaging together with the supplied accessories.

GENERAL DESCRIPTION

The Hanna Instruments portable K-type thermocouple thermometers with built-in printers are microprocessor-based and enable you to accurately measure temperature and record data.

The instrument housing is made of rugged and lightweight material. Equipped with a large, easy-to-read LCD, they feature an extended battery life and a special design that enables results to be obtained even in humid, wet or dusty conditions.

Measurements can be performed with lab-grade precision in the field as well as in the laboratory without compromising accuracy. A 12VDC car battery or a battery charger can also be used to power the unit for extended use.

HI 92801 and HI 92804 have a memory capacity of 16,000 individual temperature readings with recording intervals selected by the user.

Using the HI 9200 infrared transmitter, all the recorded data can be transferred to a PC for easy reproduction, storage or elaboration without the interference of cables and cords between the meter and the receiver.

Other features include: on-board clock and date (HI 92801 and HI 92804 only), user-selectable printing/logging intervals, automatic memorization of measured values (HI 92801 and HI 92804 only), use of pre-calibrated completely interchangeable K-type thermocouple probes, printing on plain paper, automatic shut-off capability, low

3

4

battery warning and 2 year warranty. Two versions are available:

HI 92701C, HI 92704C, HI92801C and HI92804C perform temperature readings in the Centigrade scale;

HI 92701F, HI 92704F, HI92801F and HI92804F perform temperature readings in the Fahrenheit scale.

LCD DISPLAY

FUNCTIONAL DESCRIPTION

1

°C

DATE

TIME 2

LOG

INTV

LO BAT

1.Primary Display

2.Secondary Display

4

Untitled-27

5

FUNCTIONAL DESCRIPTION

HI 92701 AND HI 92704

1.Temperature Probe Connector/s

2.Power adapter plug

3.LCD Display

4.PAPER key to pull the paper

5.ON/OFF key, to turn the meter on or off

6.INTV key to select the printing interval

7.0.1/1 key, to select the resolution

8.UP key to set up

9.DOWN key to set up

10.PRINT key to obtain a printout (printing present time and temperature/s)

11.LOG key to enter the recording mode

5

6

FUNCTIONAL DESCRIPTION

HI 92801 AND HI 92804

1.Temperature Probe Connector/s

2.Power adapter plug

3.LCD Display

4.PAPER key to pull the paper

5.PRINT key to obtain a printout (printing present time and temperature/s)

6.ALT key, alternate function key

7.TEMP key, to read temperature and to reactivate the meter when is "sleeping"

8.UP key to scan data or to set up

9.DOWN key to scan data or to set up

10.TIME key to display present time and printing interval

11.LOG key to enter and exit the logging mode

6

Untitled-27

7

SPECIFICATIONS

HI 92701 AND HI 92704

 

HI 92701C

HI 92701F

 

HI 92704C

HI 92704F

 

 

 

Range

-200.0 to 999.9°C

-300.0 to 999.9°F

 

1000 to 1370°C

1000 to 2500°F

Resolution

0.1/1°C

0.2/1°F

Accuracy

±0.3% of reading

±0.3% of reading

(@25°C/77°F)

or ±0.5°C

or ±1°F

 

for one year, excluding probe error

 

 

 

Typical EMC Deviation

±5°C

±10°F

 

 

 

Power Supply

4x1.5V AA size, 500

hours with 60' printing

 

interval. Socket for 12VDC adapter

Auto Shut-off

After 5 minutes of non-use

Channels

1 for HI 92701; 4 for HI 92704

Printer

Low-power impact belt type, 14 characters per line

 

using 38 mm plain paper

Printing Intervals

1, 2, 5, 10, 15, 30, 60, 120 and 180 minutes

Environment

0 to 50 °C (32 to 122°F) Max. 95% RH

Dimensions

220 x 82 x 66 mm (8.7 x 3.2 x 2.6")

Shipping

Instrument: 500g (18 oz.)

Weight

Kit: 1.4 kg (3.1 lb.)

SPECIFICATIONS

HI 92801 AND HI 92804

 

HI 92801C

HI 92801F

 

HI 92804C

HI 92804F

 

 

 

Range

-200.0 to 999.9°C

-300.0 to 999.9°F

 

1000 to 1370°C

1000 to 2500°F

Resolution

0.1/1°C

0.2/1°F

Accuracy

±0.3% of reading

±0.3% of reading

(@25°C/77°F)

or ±0.5°C

or ±1°F

 

for one year, excluding probe error

 

 

 

Typical EMC Deviation

±5°C

±10°F

Power Supply

4x1.5V AA size, 500

hours with 60' printing

 

interval. Socket for 12VDC adapter

Auto Shut-off

After 5 minutes of non-use

 

 

 

Channels

1 for HI 92801; 4 for HI 92804

Printer

Low-power impact belt type, 14 characters per line

 

using 38 mm plain paper

Printing/Logging Intervals

1, 2, 5, 10, 15, 30, 60, 120 and 180 minutes

Environment

0 to 50 °C (32 to 122°F) Max. 95% RH

Dimensions

220 x 82 x 66 mm (8.7 x 3.2 x 2.6")

Shipping

Instrument: 500g (18 oz.)

Weight

Kit: 1.4 kg (3.1 lb.)

 

7

 

8

OPERATIONAL GUIDE

INITIAL PREPARATION

Each meter is supplied complete with batteries. Remove the back cover, unwrap the batteries and install them while paying attention to the polarity.

To prepare the instrument for use, connect a K-type thermocouple temperature probe to the connector located on the top of the instrument.

To switch the HI92701 and the HI92704 on,

 

press the ON/OFF key.

ON/OFF

To switch the HI92801 and the HI92804 on, press the TEMP key.

TEMP

Choose the most appropriate temperature probe(s) for your application (see page 30) and connect it (them) to the meter.

HI 92701 and HI 92801 have a single probe connection. HI 92704 and HI 92804 have four probe connectors.

With the meter facing you, channel #1 is the first connector on the top left hand side.

If the temperature probe is not connected to the instrument, the meter will display and print "----" to alert the user.

This also indicates the possibility of a broken probe cable. The channel is displayed with "CH" and the number of channel (four channels versions only).

°C

CH

Note: To ensure accurate readings it is recommended that the temperature probes be connected to consecutive channels beginning with channel #1. For example, if only one probe is to be used, it must be connected to channel #1. If 2, 3 or 4 probes are to be used, they must be connected to channels 1, 2, 3, and 4 in ascending order.

8

Untitled-27

9

To maximize battery life, the display is automatically switched off after 5 minutes of non-use. However, the meter will continue to monitor (if in the logging/recording mode) temperature.

To revive the display, press the ON/OFF key (HI92701 and HI92704 only) or the TEMP key (HI92801 and HI92804 only).

ON/OFF

TEMP

SETTING THE PRINTING INTERVAL (FOR HI92701 AND HI92704 ONLY)

Turn the

instrument on by pressing the ON/

OFF key.

ON/OFF

Press the INTV and the UP or DOWN keys simultaneously.

UP

OR

 

INTV

 

 

DOWN

The display will show the log number. At

 

the bottom of the display the printing

 

interval will be flashing.

 

Keep the INTV key pressed and set the

INTV

printing interval by pressing the UP or

 

DOWN key.

 

SETTING THE MEASUREMENT RESOLUTION (FOR HI 92701 AND HI 92704 ONLY)

Press the 0.1/1 key to select the 0.1° or 1° resolution.

°C

0.1/1 CH

With four channels versions, all four channels readings automatically change by pressing this key.

To return the reading to its original resolution, simply press the 0.1/ 1 key.

9

10

SETTING DATE / TIME / PRINTING INTERVAL (FOR HI92801 AND HI92804 ONLY)

Turn the instrument on by pressing the TEMP

 

key.

TEMP

Press the ALT and the TIME keys simultaneously. The display will show the date setting. At the bottom of the display the year will be blinking.

ALT

UP

DATE

TEMP TIME

Use the UP or the DOWN key to select the year.

UP

DOWN

When the correct year is selected, press the TIME key once. The left digit of the display blinks, showing the month.

DATE

TIME

Select the month by using the UP or the DOWN key.

UP

DOWN

Press the TIME key, the right hand side of the display shows the day.

DATE

TIME

Use the UP or the DOWN key to select the correct day.

UP

DOWN

10

Hanna Instruments HI 92801 User Manual

Untitled-27

11

Press the ALT and the TIME keys simultaneously and the display will now show the clock time with the printing interval blinking.

ALT

UP

TIME

 

 

INTV

TEMP TIME

Any interval can be selected from 1, 2, 5, 10, 15, 30, 60, 120 or 180 minutes by using the UP and the DOWN key.

UP

DOWN

Once the desired interval is selected press the TIME key once to set it, the left digits of the display will start blinking.

TIME

INTV

TIME

To select the hour, press the UP or the DOWN key (24 hour clock).

UP

DOWN

To set the hour press the TIME key once, the minutes start blinking.

TIME

INTV

TIME

Use the UP or the DOWN key to select the minutes.

UP

DOWN

Press the ALT and the TIME keys to leave this mode.

ALT

UP

TEMP TIME

The time, date and printing interval are now set and stored in the

memory even when the display is switched off. 11

12

TO VIEW DATE / TIME / TEMPERATURE (FOR HI 92801 AND HI 92804 ONLY)

To view the time press the TIME key. This also displays the selected interval time on the secondary LCD.

TIME

INTV

TIME

To view the date, press the UP key when the LCD is displaying time.

DATE

UP

To view temperature press TEMP key.

°C

CHANNEL

CH

TEMP

Using UP or DOWN key select which channel to monitor (four channels versions only).

UP DOWN

If "----" is displayed, it indicates that the temperature probe is not connected.

°C

CH

This also indicates the possibility of a broken probe cable.

The channel is displayed with "CH" and the number of channel (four channels versions only).

12