Hanna Instruments C 9800 User Manual

4.3 (4)
Hanna Instruments C 9800 User Manual

Instruction Manual

C9800

REACTOR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This Instrument is in

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

w w w . h a n n a i n s t . c o m

Compliance with the CE Directives

WARRANTY

C9800 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions.

This warranty is limited to repair or replacement free of charge.

Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.

If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.

To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.

Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.

Dear Customer,

Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the reactor. This manual will provide you with the necessary information for the correct use of the heater. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with directives.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.

Note: Save all packing material until you are sure that the instrument functions correctly. Any defective item must be returned in their original packing with the supplied accessories.

GENERAL DESCRIPTION

C9800 Reactor is a Hanna Instruments thermo-reactor with a metal dry-block and a metal case construction.

It allows to warm up to 25 samples, contained into 16 mm-diameter tubes, at 105 or 150 ºC.

The reactor is equipped with led indicators, to inform when power is on and heating is operating, and a 120 minutes timer, very useful in timed digestion procedures.

It is possible to select time intervals between 0 and 120 minutes and, during these intervals, temperature in the heater is maintained constant. At the end of the selected interval an acoustic alarm will sound and the heater will automatically shut off.

The reactor is also provided with a thermal fuse that prevents from overheating, by shutting down the heater.

Two models are available:

C9800-01 runs with 115 VAC mains line

C9800-02 runs with 230 VAC mains line.

HEALTH & SAFETY

Before unpacking, setting up or start using this reactor it is important to read carefully this manual. Pay particular attention to all warnings, cautions and notes. Failure to do so could result in serious injury to the operator or damage to the equipment.

To ensure a safety use of this instrument, it must not be installed or used in any manner other than that specified in this manual.

Reactor Operation: The Reactor incubates glass vials containing hazardous chemicals at high temperature. Read the Health & Safety Data Sheet (HSDS) before handling reagent vials.

Handling Reagent Vials: Handle the vials very carefully and wear protective clothing. After use, dispose of the reagent vials according to local regulations.

Safety Equipment: When operating with Reactor use protective clothing, including goggles or face mask, and gloves.

Use of the optional safety shield is highly recommended.

Reagent Spills: Wipe up spilled reagents immediately. If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors. Read the Health & Safety Data Sheet (HSDS) of each reagent for complete chemical information.

Fire Hazard: Avoid the presence of flammable liquids or paper near the operating Reactor. A fire hazard could be created.

SPECIFICATIONS

Capacity

25 16x100 mm vials and

1 receptacle for a stainless steel probe thermometer

 

 

Temperature

selectable, 105 or 150 ºC

Accuracy

±2 ºC @ 25 ºC ambient

Temperature stability

±0.5 ºC

 

 

Warm-up time

30-40 minutes,

 

depending on set temperature

 

 

Timer

0-120 minutes with acoustic alarm

 

and automatic shut-off,

 

or infinity mode feature

 

 

Ambient operating temperature

+5 to +50 ºC

Storage temperature

-20 to +60 ºC

 

 

 

Power supply:

 

 

C9800-01

115 VAC; 60Hz; 250W; Fuse 4A

C9800-02

230 VAC; 50Hz; 250W; Fuse 2A

 

 

Dimensions

190 x 300 x 95 mm (7.5 x 11.8 x 3.9’’)

Weight

approx. 4.8 kg (10.6 lb.)