Hanna Instruments BL 983314 User Manual

4 (2)
Hanna Instruments BL 983314 User Manual

Instruction Manual

BL 983314-0

BL 983314-1

Panel-Mounted

Resistivity

Indicators & Controllers

WARRANTY

These instruments are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Probes are warranted for six months. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment. The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching this metal band at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the probe by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc. To avoid damages or burns, do not perform any measurement in microwave ovens.

Dear Customer,

Thank you for choosing a Hanna product.

This manual will provide you with the necessary information for the correct operation of the meter. Please read it carefully before using the instrument.

If you need additional technical information, do not hesitate

to e-mail us at

tech@hannainst.com.

 

These instruments

are in compliance with the

directives.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.

Each meter is supplied with:

HI 3314 resistivity probe

Mounting brackets

Instruction manual

Note: Conserve all packing material until the instrument has been observed to function correctly. Any defective item must be returned in its original packing.

GENERAL DESCRIPTION

BL983314-0 and BL983314-1 resistivity indicators and controllers have been designed for continuous monitoring of process solutions.

The setpoint value can be fixed by acting on the trimmer on the front panel, and after selection of the required limit, the output relay allows to control an external dosing device.

Connections and wiring to probe, power supply and contacts are made via the terminal blocks on the rear panel.

The probe is easy to clean and requires little maintenance. Other features include: user-selectable temperature coefficient for readings compensation, overtime control system, multicolour LED for indicating if the meter is in measurement/ dosing/alarm condition, possibility to set (Off-Auto-On switch) dosing action mode.

Two models are available:

BL 983314-0 powered at 12 Vdc

BL 983314-1 powered at 115 or 230 Vac

SPECIFICATIONS

Range

 

 

0.00 to 19.90 MΩ/cm

 

 

 

 

Resolution

 

 

0.10 MΩ/cm

 

 

Accuracy (@ 20°C/68°F)

±2% f.s.

 

 

 

Typical EMC Deviation

 

±2% f.s.

 

 

 

Probe

 

HI 3314 resistivity probe (included)

Temp. Compensation

 

Automatic and Linear

 

 

 

from 5 to 50°C (41 to 122°F)

 

 

Temperature Coefficient

β=2.4 ; 3.5 ; 4.5 %/°C

 

 

 

User-selectable through rear jumper

Calibration

 

 

Factory calibrated

 

 

Dosing Contact

Maximum 2A (fuse protected), 250 Vac, 30 Vdc

 

Contact close when measure < setpoint

Setpoint

 

Adjustable, from 0 to 19.90 MΩ/cm

 

 

Overtime

Adjustable, typically from 5 to approx. 30 minutes

 

 

 

Power Consumption

 

10 VA

 

 

 

Installation Category

 

II

Power supply:

 

 

External (fuse protected)

BL983314-0

 

 

12 Vdc

BL983314-1

 

 

115/230 Vac ; 50/60Hz

 

 

 

 

Dimensions

 

 

83 x 53 x 99 mm (3.3x2.1x3.9”)

ASSEMBLING VIEW

CE DECLARATION OF CONFORMITY

ACCESSORIES

HI 3314

Resistivity probe

HI 710005

12

Vdc power adapter, US plug

HI 710006

12

Vdc power adapter, European plug

HI 710012

12

Vdc power adapter, Australian plug

HI 710013

12 Vdc power adapter, South African plug

HI 710014

12

Vdc power adapter, UK plug

HI 731326

Calibration screwdriver (20 pcs)

HI 740146

Mounting brackets