Hanna Instruments BL 931700 User Manual

4.5 (2)
Hanna Instruments BL 931700 User Manual

Instruction Manual

BL 931700-0

BL 931700-1

Panel-Mounted

pH Indicators &

Controllers

WARRANTY

These instruments are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Probes are warranted for six months. This warranty is limited to repair or replacement free of charge.

Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment. The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc. To avoid damages or burns, do not perform any measurement in microwave ovens.

Dear Customer,

Thank you for choosing a Hanna product.

This manual will provide you with the necessary information for the correct operation of the meter. Please read it carefully before using the instrument.

If you need additional technical information, do not hesitate

to e-mail us at

tech@hannainst.com.

 

These instruments

are in compliance with the

directives.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.

Each meter is supplied with:

Mounting brackets

Instruction manual

Note: Conserve all packing material until the instrument has been observed to function correctly. Any defective item must be returned in its original packing.

SPECIFICATIONS

Range

0.00 to 14.00 pH

Resolution

0.01 pH

Accuracy (@ 20°C/68°F)

±0.02 pH

Typical EMC Deviation

±0.10 pH / ±0.10 mA

 

 

Calibration

Manual, through Offset & Slope trimmers

 

 

Dosing Contact

Maximum 2A (fuse protected),

 

250 Vac, 30 Vdc

 

 

Dosing Selection

Acid or Alk, selectable on the back panel

Contact Open=Acid dosage=Relay ON if Measure > Setpoint Contact Close=Alk dosage=Relay ON if Measure < Setpoint

Setpoint

 

Adjustable, from 0 to 14 pH

Overtime

Adjustable, typically from 5 to approx. 30 minutes

 

 

 

Recorder Output

 

4 to 20 mA, accuracy ±0.20 mA,

 

 

500 Ω maximum load

 

 

Power Consumption

10 VA

 

 

Installation Category

II

 

 

 

Power supply:

 

External (fuse protected)

BL931700-0

 

12 Vdc

BL931700-1

 

115/230 Vac ; 50/60Hz

 

 

 

Dimensions

 

83 x 53 x 99 mm (3.3x2.1x3.9”)

 

 

 

GENERAL DESCRIPTION

ASSEMBLING VIEW

BL931700-0 and BL931700-1 are panel-mounted pH indi-

 

cators and controllers designed for simplicity of use in a wide

 

range of industrial applications.

 

Connections and wiring to electrode, power supply, contacts

 

and external recorders are made via the terminal blocks on

 

the rear panel. The meters are equipped with a BNC socket

 

and accept input from conventional pH electrodes.

 

Other features include: 4-20 mA output, overtime control

 

system, selection of dosing direction (Acid/Alk), contact for

 

external control and disabling of dosing action, multi-colour

 

LED for indicating if the meter is in measurement/dosing/

 

alarm condition, possibility to set (Off-Auto-On switch) dosing

 

action mode.

 

Two models are available:

 

• BL 931700-0 powered at 12 Vdc

 

• BL 931700-1 powered at 115 or 230 Vac

 

ACCESSORIES

pH CALIBRATION SOLUTIONS

HI 7004M/L pH 4.01 buffer solution, 230/500 mL bottle HI 7006M/L pH 6.86 buffer solution, 230/500 mL bottle HI 7007M/L pH 7.01 buffer solution, 230/500 mL bottle HI 7009M/L pH 9.18 buffer solution, 230/500 mL bottle

HI 7010M/L pH 10.01 buffer solution, 230/500 mL bottle

OTHER SOLUTIONS

HI 70300M/L Electrode storage solution, 230/500 mL bottle

HI 7061M/L Electrode general cleaning solution, 230/500 mL HI 7073M/L Protein cleaning solution, 230/500 mL bottle HI 7074M/L Inorganic cleaning solution, 230/500 mL bottle HI 7077M/L Oil & Fat cleaning solution, 230/500 mL

REFILLING ELECTROLYTE SOLUTIONS (50 mL, 4 pcs.) HI 7071 3.5M KCl+AgCl, for single junction electrodes HI 7072 1M KNO3 solution

HI 7082 3.5M KCl, for double junction electrodes

pH ELECTRODES

 

HI 1002/5

BNC connector, double junction, plastic-body

 

w/external thread & 5 m (16.5’) cable

HI 1090T

Screw connector, double junction, glass-body

 

with external thread

 

HI 1110S

Screw connector, single junction, glass-body

HI 1210T

Screw connector, double junction, plastic-body

 

with external thread

 

HI 2114P/2

BNC connector, double junction, plastic-body

 

with 2 m (6.6’) cable

 

HI 1210B/5

BNC connector, double

junction, plastic-body,

 

with 5 m (16.5') cable

 

HI 2910B/5

BNC connector, double

junction, plastic-body,

 

with built-in amplifier and 5 m (16.5') cable

EXTENSION CABLES, SCREW TO BNC CONNECTOR

HI 7855/5

Extension cable 5 m (16.5') long

HI 7855/10

Extension cable 10 m (33') long

OTHER ACCESSORIES

 

BL PUMPS

Dosing Pumps (flow rate from 1.5 to 20LPH)

HI 6050

Submersible electrode holder, 60 cm (24”)

HI 6051

Submersible electrode holder, 110 cm (43”)

HI 6054B

Electrode holder for in-line applications

HI 6054T

Electrode holder for in-line applications

HI 710005/6

12 Vdc power adapter, US/European plug

HI 731326

Calibration screwdriver (20 pcs)

HI 740146

Mounting brackets

 

HI 7871

Level Controller (min and max)

HI 7873

Level Controller (min, max and overflow)