GSI TympStar User manual Versions 1

Loading...

Grason-Stadler

GSI® TympStar Versions 1 and 2

Middle-Ear Analyzer

Service Manual

Reference Instruction Manual 2000-0110, Rev3.0

Printed October, 2002

GSI

 

 

 

 

 

Grason-Stadler - A Division of VIASYS Healthcare

 

 

Grason-Stadler

 

 

 

 

5225 Verona Road, Building 2, Madison, WI 53711

 

 

 

 

TEL:608-441-2323,800-700-2282,FAX:608-441-2234

 

 

 

 

 

 

 

 

Title: GSI TympStar Versions 1 and 2

Middle-Ear Analyzer

Service Manual

Date:

October, 2002

Copyright © 2001 Grason-Stadler. All rights reserved. No part of this publication may be reproduced or transmitted in any form or by any means without the prior written permission of Grason-Stadler. The information in this publication is proprietary to Grason-Stadler.

Part number:

2000-0110

 

 

Printing history:

November, 2001

First printing

Rev 1.0

 

June, 2002

Second printing

Rev 2.0

 

October, 2002

Thirdprinting

Rev 3.0

Warranty for the GSI TympStar Versions 1 & 2

Grason-Stadler warrants the GSI TympStar Versions 1 and 2 to be free of original defects in material and workmanship and to perform in accordance with manufacturer’s specifications for a period of one year from the date of purchase. If this instrument or any component thereof is found to be defective or at variance from the manufacturer’s specifications during the warranty period, Grason-Stadler will repair, replace, or recalibrate the instrument or component at no cost to the purchaser.

This warranty only applies to instruments purchased new from Grason-Stadler or its authorized distributors or representatives. The purchaser must return the instrument directly to Grason-Stadler or an authorized GSI TympStar distributor or representative and bear the costs of shipping.

This warranty does not cover breakage or failure due to tampering, misuse, neglect, accidents, modification, or shipping, and is void if the instrument is not used in accordance with manufacturer’s recommendations or if repaired or serviced by other than Grason-Stadler or a Grason-Stadler authorized representative.

WARNING

The GSI TympStar Versions 1 and 2 are designed to be used with a hospital grade outlets. Injury to personnel or damage to equipment can result when a three-prong adapter is connected between the instruments’s power plug and an A/C outlet or extension cord.

Accessory Hazard Warning

This IEC 601-l/CSA C22.2 No. 601.1M90 listed medical instrument should be interconnected with accessories that have proper electrical compatibility and which are listed as conforming to Part 1: General Requirements for Safety of the UL Medical Electrical Equipment Standard UL2601-1. Connection of accessories not meeting these requirements may result in electrical leakage currents in excess of those allowed by the standard and present a potential electrical shock hazard to the person being tested.

User manuals

Installation, setup and

operating information can be found in the Reference

 

Instruction Manuals:

 

 

TympStar Version 1:

2000-0100

 

TympStar Version 2:

2000-0120

Service personnel

Repair and/or bench testing of the GSI TympStar Version 1 and Version 2

 

instruments should be performed only by trained personnel. The following

 

instructions are provided primarily for use by persons who are skilled in the

 

repair of electronic equipment.

Electrical safety

CAUTION

 

The GSI TympStar Version 1 and Version 2 instruments are

 

IEC 601-1/CSA approved; consequently if any parts are re-

 

placed during the repair of these units, only exact replace-

 

ments should be made. Any alterations of the present elec-

 

trical or mechanical construction or components will void

 

these safety approvals.

CMOS handling

Many of the integrated circuits on the PC Boards are constructed of CMOS

 

and NMOS materials. Please observe the following precautions:

CAUTION

Failure to observe the following precautions whenever a circuit board or an integrated circuit package is handled can result in damage to the GSI TympStar Version 1 or Version 2 instrument. Please observe these precautions:

1) Place the instrument and parts on a grounded, conductive work surface.

2) Ground yourself (with a strap having about 1 Meg-Ohm resistance) to discharge or prevent static charges.

3) Ground the frame of any test instrument or soldering iron to be used. 4) If any circuit boards are to be stored or transported, enclose them in

conductive (antistatic) envelopes.

Contents

Introduction ..................................

1-1

General .....................................................................................................................

1-1

Version 1 ...................................................................................................................

1-2

Version 1 equipment connections and options .....................................................

1-2

Version 2 ...................................................................................................................

1-3

Version 2 equipment connections and options .....................................................

1-3

The probe .................................................................................................................

1-4

Instrument controls .................................................................................................

1-5

Hardkeys and softkeys .........................................................................................

1-5

Rotary pressure control .......................................................................................

1-5

Test modes and menu navigation ..........................................................................

1-5

Changing parameter settings..................................................................................

1-7

Menu diagrams ........................................................................................................

1-7

GSI default parameter settings ...............................................................................

1-7

Specifications ...............................

2-1

Introduction ..............................................................................................................

2-1

Standards .................................................................................................................

2-1

Version 1 Specifications ...................................................

2-1

Sensitivity ranges ....................................................................................................

2-2

Probe signal .............................................................................................................

2-2

Pneumatic System ...................................................................................................

2-3

Acoustic Reflex Activating (Stimulus) Signal .........................................................

2-3

Stimulus Presentation Control ................................................................................

2-5

Temporal Specifications of Stimulus Presentation ...............................................

2-5

Environmental ..........................................................................................................

2-6

Warm-Up Time..........................................................................................................

2-6

Calibration Stability ..................................................................................................

2-7

Connectors ...............................................................................................................

2-7

Electrical ...................................................................................................................

2-8

Supplied Accessories ..............................................................................................

2-9

Mechanical ...............................................................................................................

2-9

Materials of manufacture ........................................................................................

2-9

GSI TympStar Version 1 and Version 2 Service Manual

Contents - 1

Table of Contents

 

Calibration requirements .........................................................................................

2-9

Version 2 Specifications .................................................

2-10

Standards ...............................................................................................................

2-10

Sensitivity ranges ..................................................................................................

2-11

Temporal Latency in ARLT Mode ..........................................................................

2-12

Multi Frequency .....................................................................................................

2-12

Probe Signal (Sinusoidal) ......................................................................................

2-12

Pneumatic System .................................................................................................

2-13

Acoustic Reflex Activating (Stimulus) Signal .......................................................

2-13

Click Stimulus ........................................................................................................

2-16

Stimulus ..................................................................................................................

2-16

Presentation Control..............................................................................................

2-16

Temporal Specifications Of Stimulus Presentation ............................................

2-17

Environmental ........................................................................................................

2-17

Warm-Up Time........................................................................................................

2-17

Calibration Stability................................................................................................

2-18

Connectors .............................................................................................................

2-18

Electrical .................................................................................................................

2-19

Supplied Accessories ............................................................................................

2-20

Mechanical .............................................................................................................

2-20

Materials of manufacture ......................................................................................

2-20

Calibration requirements .......................................................................................

2-20

Operation summary......................

3-1

Front panel controls .........................................................

3-1

Hardkeys...................................................................................................................

3-2

Tymp ....................................................................................................................

3-2

Reflex ...................................................................................................................

3-2

Etf .........................................................................................................................

3-2

Special .................................................................................................................

3-2

Page .....................................................................................................................

3-2

Print ......................................................................................................................

3-2

Remote .................................................................................................................

3-2

Data Transfer .......................................................................................................

3-2

Pressure control ...................................................................................................

3-2

Manual .................................................................................................................

3-2

Hold ......................................................................................................................

3-2

Stop ......................................................................................................................

3-2

Start ......................................................................................................................

3-3

Stimulus ...............................................................................................................

3-3

Intensity ................................................................................................................

3-3

Contents - 2

Grason-Stadler

Table of Contents

 

Present .................................................................................................................

3-3

LCD hardkeys...........................................................................................................

3-3

Clear ....................................................................................................................

3-3

Erase ....................................................................................................................

3-3

Return...................................................................................................................

3-3

Softkeys ....................................................................................................................

3-3

Menus and LCD screens .................................................

3-4

Menus .......................................................................................................................

3-4

Program modes........................................................................................................

3-5

Version 1 probe tone frequencies ..........................................................................

3-5

Version 2 probe tone frequencies ..........................................................................

3-5

Version 1 LCD graphic traces .................................................................................

3-5

Version 2 LCD graphic traces .................................................................................

3-5

Probe lights ..............................................................................................................

3-6

LCD screen ...............................................................................................................

3-6

Erasing and clearing test data ................................................................................

3-7

Version 2 External reflex test stimuli .....................................................................

3-7

Paging test data .......................................................................................................

3-7

Version 2 cursor.......................................................................................................

3-8

Printing tests ............................................................................................................

3-8

Preparing an external printer ..................................................................................

3-8

Setting the printer parameters ..............................................................................

3-9

Left column printout data ....................................................................................

3-10

Color or black and white ....................................................................................

3-10

Changing Instrument Versions .......................................

3-11

Changing from a Version 2 to a Version 1 ...........................................................

3-11

Switching to V1 temporarily................................................................................

3-12

Switching back to V2 ..........................................................................................

3-12

Switching to V1 permanently ..............................................................................

3-12

Changing from a Version 1 to a Version 2 ...........................................................

3-13

Obtaining a License Code ..................................................................................

3-14

Replacing the V1 Probe with a V2 Probe ...........................................................

3-15

Switching to Version 2 software .........................................................................

3-16

Version 1 test procedures ..............................................

3-17

Tymp diagnostic.....................................................................................................

3-17

Manual Tymp ..........................................................................................................

3-17

Tymp screening .....................................................................................................

3-18

Reflex threshold .....................................................................................................

3-18

Automatic Auto Zero ...........................................................................................

3-18

Mark Threshold ..................................................................................................

3-19

Reflex Threshold Seeking ..................................................................................

3-19

Reflex decay ...........................................................................................................

3-20

Eustachian tube function ......................................................................................

3-20

Pressure Swallow Test (intact TM) .....................................................................

3-20

GSI TympStar Version 1 and Version 2 Service Manual

Contents - 3

Table of Contents

 

Perforated Ear Test (Perforated TM) ..................................................................

3-20

Auto sequence testing ..........................................................................................

3-21

Programming auto sequence tests ......................................................................

3-21

Version 2 test procedures ..............................................

3-22

Tymp diagnostic.....................................................................................................

3-22

Manual tymp ...........................................................................................................

3-23

Tymp screening .....................................................................................................

3-23

Reflex threshold .....................................................................................................

3-24

Automatic auto zero ...........................................................................................

3-25

Mark Threshold ..................................................................................................

3-25

Reflex Threshold Seeking ..................................................................................

3-25

Reflex decay ...........................................................................................................

3-26

Acoustic reflex latency test (ARLT) ......................................................................

3-26

Acoustic reflex sensitization (A.R. SENSI) ...........................................................

3-27

Eustachian tube function ......................................................................................

3-28

Pressure Swallow Test (intact TM) .....................................................................

3-28

Perforated Ear Test (Perforated TM) ..................................................................

3-28

Multiple frequency tympanometry (MULTIPLE Hz) .............................................

3-29

Auto sequence testing ..........................................................................................

3-30

Programming auto sequence tests ......................................................................

3-31

Calibration ....................................

4-1

Introduction ..............................................................................................................

4-1

When to calibrate the TympStar .............................................................................

4-2

Calibration requirements .........................................................................................

4-2

Complete calibration sequence ............................................................................

4-2

Routine calibration sequence ...............................................................................

4-3

Preparing for calibration ...................................................

4-4

Tools and equipment required ................................................................................

4-4

Cleaning the probe and Contra phone ...................................................................

4-4

Probe....................................................................................................................

4-4

Contra phone........................................................................................................

4-5

Cal/Normal switch ............................................................

4-6

Cal Option DIP switch ......................................................

4-6

Switch 1: Factory use only ...................................................................................

4-7

Switch 2: GSI/custom RTL transducer calibration..............................................

4-7

Switch 3: Unused ...................................................................................................

4-7

Switch 4: Unused ...................................................................................................

4-7

Switch 5: Unused ...................................................................................................

4-7

Switch 6: Diagnostic mode ...................................................................................

4-8

Contents - 4

Grason-Stadler

Table of Contents

 

Hardware Diagnostics .............................................................................................

4-8

DAC LEVELS .......................................................................................................

4-9

IPSI SOURCE.....................................................................................................

4-11

IPSI ON/OFF ......................................................................................................

4-11

CONTRA SOURCE ............................................................................................

4-11

CONTRA ON/OFF.............................................................................................

4-12

MUX INPUTS .....................................................................................................

4-12

PROBE/FILER ....................................................................................................

4-13

ROUTING ...........................................................................................................

4-14

PROBE ON/OFF ................................................................................................

4-15

PORTS DIAG. ..........................................................................................................

4-15

RS-232 INT. ON/OFF .........................................................................................

4-15

RS-232 INT. ON/OFF .........................................................................................

4-16

PARALLEL ON/OFF ...........................................................................................

4-16

MISC. DIAG .............................................................................................................

4-17

CALENDAR CLOCK ...........................................................................................

4-18

DISPLAY & LEDS ...............................................................................................

4-20

KEYS ..................................................................................................................

4-21

DIP SWITCHES ..................................................................................................

4-22

EEPROM ............................................................................................................

4-22

SOFTWARE TIMER ...........................................................................................

4-23

WATCHDOG TIMER ..........................................................................................

4-24

ERROR LOG ...........................................................................................................

4-25

Switch 7: Self-Cal..................................................................................................

4-26

Switch 8: Load default calibration data...............................................................

4-26

Probe serial number.......................................................

4-26

Resolving problems .......................................................

4-27

Calibration procedures...................................................

4-28

Hardkeys ............................................................................................................

4-28

Softkeys .............................................................................................................

4-29

Printing calibration sheets ..................................................................................

4-30

Adjusting Cal HL and Target SPL levels.............................................................

4-31

Custom transducers ..............................................................................................

4-32

Transducerconnections ........................................................................................

4-32

Calibration requirements .......................................................................................

4-33

Complete calibration sequence ..........................................................................

4-33

Routine calibration sequence .............................................................................

4-33

TympStar Version 1 Calibration......................................

4-33

Version 1 Calibration Procedures...................................

4-34

Self-Cal ...................................................................................................................

4-34

Loading default data ..............................................................................................

4-35

Contra, Ipsi and Probe tone calibration ...............................................................

4-36

Contra SPL Cal ..................................................................................................

4-36

GSI TympStar Version 1 and Version 2 Service Manual

Contents - 5

Table of Contents

 

Ipsi SPL Cal .......................................................................................................

4-37

Probe Tone SPL Cal...............................................................................................

4-38

Y Cal ........................................................................................................................

4-39

Altitude Cal .........................................................................................................

4-39

Pressure Cal ...........................................................................................................

4-41

Set ambient ........................................................................................................

4-42

Set -600/400 ............................................................................................................

4-44

Leak rate ............................................................................................................

4-45

Verify Cal ................................................................................................................

4-47

Diagnostic...............................................................................................................

4-49

Returning to the Normal Test Mode .....................................................................

4-50

TympStar Version 2 Calibration .....................................

4-51

Calibration requirements ......................................................................................

4-51

Complete calibration sequence ..........................................................................

4-51

Routine calibration sequence.............................................................................

4-51

Version 2 Calibration Procedures ..................................

4-52

Self-Cal ...................................................................................................................

4-52

Loading default data...........................................................................................

4-53

Contra, Ipsi and Probe tone calibration ...............................................................

4-54

Contra and Ipsi Cal ............................................................................................

4-54

External Input Calibration ...................................................................................

4-55

Click Calibration .................................................................................................

4-55

Contra SPL Cal .......................................................................................................

4-56

Ipsi Channel .......................................................................................................

4-56

Contra Channel ..................................................................................................

4-57

Ipsi SPL Cal ............................................................................................................

4-59

Ipsi Channel .......................................................................................................

4-59

Contra Channel ..................................................................................................

4-60

Probe Tone SPL Cal...............................................................................................

4-62

Y/B/G Cal ................................................................................................................

4-64

Altitude Cal .........................................................................................................

4-64

Pressure Cal and diagnostics ...............................................................................

4-66

Returning to Normal Test Mode ...........................................................................

4-66

Disassembly .................................

5-1

Introduction ..............................................................................................................

5-1

Disassembly Procedure ...................................................

5-3

Tools Required .........................................................................................................

5-3

Opening the cover ...................................................................................................

5-3

Removing the analog board ...................................................................................

5-4

Removing the Pump ................................................................................................

5-6

Contents - 6

Grason-Stadler

Table of Contents

 

Removing the power supply...................................................................................

5-8

Removing the power module .................................................................................

5-9

Fuse holder ..........................................................................................................

5-9

Removing the LCD contrast control ....................................................................

5-10

Removing the printer board ..................................................................................

5-11

Removing the printer .............................................................................................

5-12

Removing the PC104 Board ..................................................................................

5-15

Removing the digital board ...................................................................................

5-17

Removing the LCD or Softkeypad cables ............................................................

5-20

Removing the LCD .................................................................................................

5-25

Removing the LCD front panel label (softkey panel) ...........................................

5-27

Removing the instrument front panel label..........................................................

5-29

Removing panel keypads.......................................................................................

5-29

System Level Parts ......................

6-1

General .....................................................................................................................

6-1

Assembly drawings ..................................................................................................

6-1

Case/chassis ............................................................................................................

6-2

Overall assembly .....................................................................................................

6-4

LCD assembly ..........................................................................................................

6-5

Top case assembly ..................................................................................................

6-6

Bottom case assembly ............................................................................................

6-7

Instrument assembly ...............................................................................................

6-8

Labels .......................................................................................................................

6-9

Printer Assembly ....................................................................................................

6-11

Hardware description ...................

7-1

General .....................................................................................................................

7-1

System boot sequence ....................................................

7-1

Pre-CP Control .........................................................................................................

7-1

Main() ........................................................................................................................

7-1

Set_op_mode() ........................................................................................................

7-3

Tymp_sk() .................................................................................................................

7-4

Interconnection diagram ..................................................

7-5

Block diagram description of assemblies .............................................................

7-6

LCD assembly .....................................................................................................

7-6

Printer...................................................................................................................

7-6

Probe ...................................................................................................................

7-6

GSI TympStar Version 1 and Version 2 Service Manual

Contents - 7

Table of Contents

 

Pump ...................................................................................................................

7-6

Digital board .........................................................................................................

7-7

PC104 board ........................................................................................................

7-7

Analog board........................................................................................................

7-7

Power supply board ..............................................................................................

7-8

Printer board ........................................................................................................

7-8

System assembly diagram ..............................................

7-9

Troubleshooting ...........................

8-1

System power supply measurements ..............................

8-1

Power Supply (J2) ....................................................................................................

8-1

Power Supply (set +5VDC) ......................................................................................

8-2

Analog Board (J7) ....................................................................................................

8-2

Analog Board (J6) ....................................................................................................

8-2

Analog Board (J10) ..................................................................................................

8-3

Digital Board (J28) ...................................................................................................

8-3

Digital Board (J34) ...................................................................................................

8-3

Printer Board (Full View) .........................................................................................

8-4

Printer Board (J6) ....................................................................................................

8-4

Printer Board (J4) ....................................................................................................

8-4

Printer Board (J8) ....................................................................................................

8-5

Pressure Potentiometer (Digital Board) .................................................................

8-5

Error Codes .....................................................................

8-6

CP Error codes ........................................................................................................

8-6

SP Error codes .......................................................................................................

8-16

Error Messages..............................................................

8-19

LCD problem symptoms and probable causes ..............

8-22

Probe problem symptoms and probable causes ............

8-22

Pump problem symptoms and probable causes ............

8-23

Printer problem symptoms and probable causes ...........

8-23

Hardware Diagnostic mode............................................

8-24

Functional description ...........................................................................................

8-24

Pressure A/D MUX .................................................................................................

8-24

Mic. MUX.................................................................................................................

8-25

Probe tone and filter mode ...................................................................................

8-25

DAC levels ..............................................................................................................

8-25

Routing ...................................................................................................................

8-25

IPSI/CONTRA and OSC l/OSC 2 ............................................................................

8-26

Contents - 8

Grason-Stadler

Introduction

1

General

The GSI TympStar Version 1and Version 2 Middle-Ear Analyzers are techni-

 

cally advanced, computer-based admittance instruments designed to be used

 

in clinical or research settings. The TympStar builds on the sophistication,

 

functionality and flexibility of the GSI 33, offering unparalleled testing capa-

 

bilities. It contains total capabilities for complete, manual or automatic diag-

 

nostic testing for analysis of middle ear function.

 

Operators have a choice of using GSI

 

preprogrammed test parameters, or programming

 

their own test criteria. A large liquid crystal

 

display (LCD) clearly displays test param-

 

eter choices and the possible alterna-

 

tives. Admittance and pressure

 

indications are shown on the

 

LCD along with a

 

continuous digital

 

readout.

Test sta-

 

tus and

invalid

 

choices are also shown

 

on the LCD.

Test results are displayed in real-time. The user can view the results as they are being measured and then has the choice of printing the display or retesting the patient. The high-speed printer generates reports in concise graphical formats that are easy to read.

This manual addresses the service requirements of the Version 1 and the Version 2 instruments, and calls attention to the differences when necessary. Please refer to the appropriate Reference Instruction Manual for more detailed information regarding instrument installation and operation.

GSI TympStar Version 1 and Version 2 Service Manual

1 - 1

 

 

 

 

Chapter 1

Version 1

Admittance (Y) may be measured with a probe tone frequency of 226 Hz. The

 

extensive battery of test mode choices include:

 

• Diagnostic Tympanometry

 

• Acoustic Reflex Threshold and Decay Measurements

 

• Eustachian-Tube Function Testing (Both intact and perforated eardrums)

 

• Screening Tympanometry/Reflex (Automatic Only)

 

The tympanometric measurement results are automatically

 

scaled and presented in equivalent ml of compliance at “Y”,

 

226 Hz. Sensitivity scales for the display of reflex mea-

 

surement results can be selected manually. A cur-

 

sor is available in all test modes for calling out

 

numeric positions on the X and Y

 

axes.

Version 1 equipment

TympStar Version 1 options that can be

connections and

connected to the rear panel include RS232

options

serial communication, a keyboard for entering patient information and a VGA

 

monitor for displaying test results. Other options offered for the TympStar

 

Version 1 for managing and archiving data, include:

 

• Internal memory for storing up to 26 test results

 

• Data export to an external PC via RS232 serial interface

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Not used in TympStar

Version 1

1 - 2

Grason-Stadler

 

 

 

 

Introduction

Version 2

Admittance (Y), and its components Susceptance (B) and Conductance (G), may

 

be measured with probe tone frequencies of 226, 678, and 1000 Hz. The extensive

 

battery of test mode choices include:

 

• Diagnostic Tympanometry

 

• Acoustic Reflex Threshold and Decay Measurements

 

• Eustachian-Tube Function Testing (Both intact and perforated eardrums)

 

• Screening Tympanometry/Reflex (Automatic Only)

 

• Acoustic Reflex Latency Testing

 

• Acoustic Reflex Sensitization

 

• Multiple Frequency Tympanometry (250 Hz to 2000 Hz)

 

The tympanometric measurement results are automatically scaled and presented

 

in equivalent ml of compliance at “Y”, 226 Hz. All “B” and “G” measure-

 

ments and measurements performed at probe tone frequencies of 678 Hz and

 

1000 Hz are expressed in mmhos. Sensitivity scales for the display of reflex

 

measurement results can be selected manually. Reflex test stimuli can be in-

 

put from an external source and presented via external control. A cursor is

 

available in all test modes for calling out numeric positions on the X and Y

 

axes.

Version 2 equipment

TympStar options that can be connected to the rear panel include an external

connections and

stimulus source, an external Present control, RS232 serial communication, a

options

keyboard for entering patient information and a VGA monitor for displaying

 

test results. Other options offered for the TympStar for managing and archiving

 

data, include:

 

• Internal memory for storing up to 26 test results

 

• Data export to an external PC via RS232 serial interface

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Used only in TympStar

Version 2

GSI TympStar Version 1 and Version 2 Service Manual

1 - 3

 

Chapter 1

The probe

The innovative lightweight probe is designed for patient comfort, ease-of-seal, and

 

accurate test results. A wide variety of both standard and special sized eartips are

 

supplied with the GSI TympStar to hermetically seal the ear canal. In addition, a set

 

of screening eartips is provided for screening tymp and reflex tests.

 

The operator has a choice of three mountings to support the probe box; the

 

standard lightweight shoulder mounting, standard clothes clip, or an optional

 

operator wrist attachment. The probe box has 2 LED’s to indicate test status.

 

Within the probe box there are two small loudspeakers, a microphone and a

 

pressure transducer. One loudspeaker delivers the probe tone to the ear canal,

 

while the microphone monitors the intensity of the probe tone within the ear canal.

 

The second loudspeaker delivers the ipsilateral (ipsi) stimuli to the ear canal.

 

The contralateral (contra) insert phone contains its own loudspeaker. The

 

pressure within the ear canal can be varied within the range of +400 daPa

 

to -600 daPa. At no time is it possible to exceed specified maximum

 

limits. Pressure can be varied automatically or manually and from nega-

 

tive to positive or from positive to negative values. Pressure within

 

the ear canal is monitored continuously in order to maintain pressure

 

accuracy throughout each test sequence.

1 - 4

Grason-Stadler

 

Introduction

 

 

Instrument controls

A combination of hardkeys and softkeys are used to select the test modes and

 

parameters and to conduct tests.

 

 

Hardkeys and softkeys

Hardkeys are located on the front panel and the sides of the LCD panel and provide

 

fixed functions that do not change. Softkeys are located directly under the LCD and

 

change to support the requirements of a test session.

 

 

 

LCD panel

 

 

 

Softkey

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hardkeys

 

 

 

 

 

 

Front panel

 

 

 

 

 

 

 

 

 

 

 

Pressure control

Rotary pressure

A rotary pressure control is also provided to change or fine-tune pressure within the

control

ear canal.

 

 

Test modes and menu

Selecting a test mode by pressing a hardkey causes the required test screen to be

navigation

displayed on the LCD with the appropriate menu of test parameters shown across

 

the bottom. The softkeys are then used in conjunction with a few hardkeys to

 

navigate through the menus and set parameter values for the selected tests. For

 

example, pressing the TYMP hardkey causes the Tymp Diagnostic screen to be

Tymp hardkey

displayed.

 

 

Tymp Diagnostic screen

Tymp parameter menu

GSI TympStar Version 1 and Version 2 Service Manual

1 - 5

First level

Second level

First level

Second level

Chapter 1

Pressing the softkeys displayed across the bottom of the LCD allows the user to navigate through the Tymp Diagnostic parameter menus diagrammed below. These menus are from the Version 1, however, the process is the same in both versions. GSI default softkey selections are circled on menu diagrams.

The first level of parameter menu selections includes AUTO SEQUENCE, P-RANGE, START, GRADIENT, EAR and MORE.

AUTO

P-RANGE

START

 

GRADIENT

EAR

MORE

SEQUENCE

daPa

daPa

 

 

 

 

 

 

NORMAL WIDE -600 -400

-200

+200

+400

MORE TYMP WIDTH

RATIO OFF

RIGHT LEFT

 

 

 

 

 

 

daPa

ml

 

 

 

-500

-300

0

+300

MORE

 

 

 

 

 

 

 

 

P-RATE

BASELINE

MORE

 

 

 

 

 

daPa/s

 

 

 

 

 

 

 

12.5 50

200 600/200

ON OFF

 

Default selection

A second level of menus can be displayed by pressing the MORE softkey and includes P-RATE, BASELINE and MORE.

The MORE softkey is used to toggle between the top and lower levels of menus while the test screen portion above the softkeys remains unchanged.

1 - 6

Grason-Stadler

Introduction

Pressing a parameter menu softkey causes a sub-menu of parameter settings to be displayed. Often submenus will also contain MORE softkey selections that provide access to additional setting alternatives.

Sub-menu of

START

 

 

daPa

 

 

START daPa

 

 

 

 

-600 -400

-200

+200

+400

MORE

-500

-300

0

+300

MORE

Sub-menu of

START daPa

Changing parameter

Settings can be changed for a selected parameter by pressing the desired softkey as

settings

shown in this example of changing the Probe Hz from 226 to 1000.

 

Making the new selection returns the display to the previous menu level with the new

 

setting shown above the selected parameter.

 

If no change is desired, the display can be returned to the previous level by

 

pressing the RETURN hardkey.

 

In the manner described above, menus can be navigated and settings can be

 

changed for any of the test modes.

Menu diagrams

Menu structure diagrams like the diagrams shown on these pages will be used

 

throughout the remainder of this manual as a convenience to the user.

GSI default param- GSI default parameters are circled on eter settings menu structure diagrams shown throughout the remainder of this

manual.

START daPa

-600 -400 -200 +200 +400 MORE

GSI default value

-500 -300 0 +300 MORE

GSI TympStar Version 1 and Version 2 Service Manual

1 - 7

Chapter 1

1 - 8

Grason-Stadler

Specifications 2

Introduction

Detailed specifications are provided in this chapter for the TympStar Version 1

 

and TympStar Version 2 instruments. The specifications for each instrument

 

can also be found in its corresponding Reference Instruction Manual.

Version 1 Specifications

Standards

The GSI TympStar Version 1 meets or exceeds the following standards and

 

specifications for aural acoustic admittance instruments:

 

IEC 1027 1991-03

 

ANSI S3.39-1987

 

ANSI S3.6-1996

 

ANSI S3.7-1995

 

IEC 645-1 1992

 

IEC 126-1973 (also BS 6111-1981)

 

BS ISO 389-2 1994

 

Y2K Compliant

 

UL 2601-1 Part 1: General Requirements for Safety

 

CSA C22.2 No. 601.1-M90 (Canada)

CE Mark per Medical Device Directive (93/42/EEC) European Representative:

Mr. Leo Hoogendoorn Nicolet Biomedical GMBH Saalackerstrasse 8

63801 Kleinostheim, Germany Tel: 011-49-6027-46980

Fax: 011-49-6027-469815

EN60601-1:1990 Safety Requirements for Medical Electrical Equipment

EN60601-1-2 Medical Electrical Equipment Emissions and Immunity

Requirements

This equipment has been tested for radio frequency emissions and has been verified to meet Radiated and Conducted Emissions per EN 55011-1998, Group 1, Class A and per CISPR, Class A.

GSI TympStar Version 1 and Version 2 Service Manual

2 - 1

 

 

 

Chapter 2

 

Sensitivity ranges

The following admittance measurements give maximum range at 226 Hz, Y

 

in ml.

 

 

 

 

 

Table 1: Tymp Mode

 

 

 

 

 

 

Frequency

 

Digital Read-Out

Graphical

 

 

 

Including Cursor

Display

 

 

 

 

 

 

226 Hz

 

-7.0 to +7.0

-1.0 to +7.0

 

 

 

 

 

 

Accuracy at 226Hz:

0.1 ml or 5%, whichever is greater.

 

 

 

 

Table 2: Reflex Mode

 

 

 

 

 

 

 

Frequency

 

Digital Read-Out

 

Graphical

 

 

 

Including Cursor

 

Display

 

 

 

 

 

 

 

226 Hz

 

-7.0 to +7.0

 

-0.16 to +0.80

 

 

 

 

 

 

+0.16 to -0.80

 

 

 

 

 

Accuracy: 226 Hz is 0.02 ml or 5%,whichever is greater.

Probe signal

Sinusoidal signal with the following characteristics:

 

Frequency:

 

226 Hz

 

 

Frequency accuracy:

 

±1%

 

 

 

Harmonic distortion:

<2%

 

 

 

(Measured in an HA-1 2cc coupler)

 

 

Signal level:

 

85 dB SPL

 

 

(In Real Ear and in Normal Test Mode)

 

 

NOTE

 

 

 

 

 

The probe tone level is set to be nominally 70 dB HL.

 

Signal level accuracy:

±1.5 dB SPL

 

2 - 2

Grason-Stadler

 

 

Specifications

Pneumatic System

Pressure Maximum Limits: -800 daPa to +600 daPa

 

Programmed Pressure Ranges:

 

Normal:

+ 200 to - 400 daPa

 

Wide:

 

+ 400 to - 600 daPa

 

Pressure Accuracy:

±10% or ±10 daPa, Whichever is greater in cavities

 

 

 

from 0.5 cc to 5.0 cc.

 

Pressure Sweep Rate:

12.5 daPa/sec

 

 

 

50 daPa/sec

 

 

 

600/200 daPa/sec

 

Manual Sweep Rate Limit: 600 daPa/sec

 

Sweep Rate Accuracy:

±10%

 

Pressure System Leak Rate: < 1. 0 daPa/sec

 

(Measured at -600 and +400 daPa, while pressure servo is disabled.)

Acoustic Reflex

Pure Tone Stimulus

 

Activating (Stimulus)

Frequencies for Contra phone and for Ipsi phone with time multiplexed

Signal

stimulus. See Table 3.

 

 

FrequencyAccuracy: ±1%

 

 

Total Harmonic Distortion (Acoustically): <2%

 

(Measured 5 dB HL below guaranteed maximum HL)

 

Noise Stimulus

 

 

 

The uniformity of the spectrum level of acoustic pressure for the noise signal over 20

 

averages measured acoustically within their respective band limits will be:

 

±10 dB for insert or probe type earphones

 

±5 dB for supra-aural type earphones

 

Noise Band Widths:

 

 

Low Band: 125 -1600 Hz

 

High Band: 1600 - 4000 Hz

 

Broad Band: 125 - 4000 Hz (Relative to level at 1 kHz)

 

Band-edges accurate to within ±15%

 

Roll off rate:

>12 dB/Octave

Stimulus Level Control

Tone Stimulus: The transfer of reference equivalent threshold values are based on the article; “Reference Threshold Levels For The ER-3A Insert Phone”, by Laura Ann Wilber, Barbara A. Krueger and Mead C. Killion, J. Acoustic Soc. Am. Suppl. 1, Vol. 81 Spring 1987. GSI determined the transfer data from an IEC 711 coupler to an ANSI HA-1 coupler. Using this data, the reference threshold values were determined for both the Ipsi and Contra insert earphones for calibration in an ANSI HA-1 2cc coupler.

Noise Stimulus: The transfer of reference threshold values was done by GSI using the “Threshold Determination Method”. The transfer data from an IEC 711 to an ANSI HA- 1 coupler was determined by GSI.

GSI TympStar Version 1 and Version 2 Service Manual

2 - 3

Chapter 2

Intensity levels are reduced as a function of volume at a rate of 1 dB SPL for each .1 ml. Intensity reduction begins at 1.2 ml.

Table 3: Upper limit of HL range in Reflex Threshold Mode

110 105

Lower limit of HL range for all stimulus signals (35 dB HL).

Hearing Level Increment:

5.0 dB

Hearing Level Increment Accuracy:

±0.5 dB

Hearing Level Control Linearity:

±1.0 dB

2 - 4

Grason-Stadler

 

Specifications

 

Stimulus

Signal ON/OFF Ratio:

>70 dB

Presentation Control

OFF mode signal level:

<20 dB SPL

 

Signal To Noise Ratio:

>70 dBA

 

Measured with disabled probe signal and “A” weighting for noise

 

measurement.

 

 

Residual noise:

<25 dBA SPL

 

(Stimulus Present switch in OFF position)

 

Unwanted Acoustic Probe Signals:

<60 dBA

Measured while pump is operating and probe tone is disabled. Measure it with “A” weighted filter in “SLOW” Time mode. The noise in normal operating mode will not effect the immittance measurement accuracy.

Signal separation between Ipsi and Contra channels; all frequencies: >70 dB

 

(Measured at the probes with the “ON” channel set to 90 dB)

 

Leaked signal: <20 dB SPL

 

The radiated acoustic noise from the instrument (with reflex stim off) when

 

measured at 1 meter from the instrument, shall be: <50 dBA

 

(“A” weighting and “SLOW” averaging)

Temporal Specifica-

Steady State Stimulus

tions of Stimulus

Initial Delay (elapsed time from present bar activation to 10% stimulus

Presentation

amplitude): < 100 msec

 

Terminal Delay (elapsed time from present bar deactivation to 90% stimulus

 

amplitude): <100 msec

 

Rise time: 7.5 ± 2.5 msec

 

Fall time: 7.5 ± 2.5 msec

 

Multiplexed Stimulus (Used in Reflex Threshold test mode)

Period data for frequencies: 250 and 500 Hz:

Period

124

msec

Stimulus on time

44

msec

Stimulus off time

62

msec

Rise and fall time

18

msec

Period data for all other frequencies:

Period

115 msec

Stimulus on time

44

msec

Stimulus off time

53

msec

Rise and fall time

18

msec

Temporal Spec. Accuracy:

±10% or 5 msec, whichever is greater

GSI TympStar Version 1 and Version 2 Service Manual

2 - 5

 

Chapter 2

 

Environmental

The GSI TympStar Version 1 meets ANSI S3.6-1996 Standards for temperature

 

and humidity specifications, and it meets the UL 2601-1, CSA 22.2 and IEC60601-

 

1 standards for safety.

 

 

Temperature:

 

 

Storage: -40 degrees C to +75 degrees C

 

 

Operating: +15 degrees C to +35 degrees C

 

Humidity: 90% at 35 degrees C (non-condensing)

Warm-UpTime

At room temperature; +15°C to +35°C:

10 Minutes

 

At room temperature; below +15°C:

1 Hour

2 - 6

Grason-Stadler

 

Specifications

 

Calibration Stability

All GSITympStar Version 1 specifications are met over the range of specified power

 

line, temperature and humidity variations.

 

 

Power Line:

 

 

 

Voltage Variation:

±10%

 

 

Frequency Variation:

± 5%

 

 

Power line short term variation which affects the performance of the instru-

 

ment will turn off all probe and stimulus signals.

 

Power Rating:

120

Watts maximum

 

Line Voltage Range:

100

VAC to 240 VAC

 

Power Line Frequency Range: 50 - 60 Hz

 

Temperature Operating Range: +15°C to +35°C

 

Relative Humidity Operating Limit: 90%

 

Guaranteed Operating Elevation: 6000 Ft. (1800m)

Connectors

STIMULUS: External Stimulus Input

 

 

(Phone Jack) Peak Voltage: 3 VAC

 

 

Input Impedance: 15,000 Ohms

 

PRESENT: External Present Control Input that turns the stimulus signal ON and OFF (Phone Jack).

Voltage Range: STIM OFF: +5.0 VDC STIM ON: 0.0 v

Input Impedance: 11,000 Ohms

Contra PHONE: Output Voltage: 7 VAC Output Impedance: 2.5 Ohms

Standard 9-pin RS232C serial port for interfacing with outside computer.

RXD

CTS

TXD

RTS

Standard VGAport for external monitor.

GSI TympStar Version 1 and Version 2 Service Manual

2 - 7

GSI TympStar User manual Versions 1

Chapter 2

Standard PS-2 keyboard port for external Keyboard.

Standard parallel printer port for external printer.

Electrical

The following apply to the TympStar system:

 

1)

Class 1 Medical Equipment

 

2)

Type B Medical Equipment

 

3)

IPXO ingress of water (ordinary equipment)

 

4)

Equipment not suitable for use in the presence of flammable anesthetic

 

 

mixture with air or with oxygen or nitrous oxide.

 

5)

Mode of operation - continuous

Input Voltage:

100 – 240 VAC

Input Frequency:

50 – 60 Hz

Input Current:

3.2 A maximum

Power Consumption:

120W maximum

2 - 8

Grason-Stadler

 

Specifications

 

SuppliedAccessories

Contra insert phone

GSI Part # 8000-0078

 

Calibration cavity (V1)

GSI Part # 2000-1036

 

Probe cleaning kit (2 boxes)

GSI Part # 2000-9610

 

Eartips:

 

 

1 pkg. 8 standard sizes, 4 ea. (Color coded)

GSI Part # 1700-9660

 

1 pkg. 6 special sizes, 2 ea.

GSI Part # 1700-9670

 

1 pkg. 6 screening sizes, 2 ea.

GSI Part # 1700-9622

 

Reference Instruction Manual

GSI Part # 2000-0100

 

Quick User Guide

GSI Part # 2000-0108

 

Printer paper (2 rolls) for orders including printer GSI Part # 1700-9619

 

Spare set of probe tubing

GSI Part # 2000-9617

 

Probe mount - shoulder

GSI Part # 1700-9646

 

Probe mount - wrist

GSI Part # 1700-9642

 

Probe mount - clothes

GSI Part # 1700-9608

 

Printer paper, adhesive-backed

 

 

for orders including a printer

GSI Part # 1770-9643

 

GSI TympStar dustcover

GSI Part # 1700-9618

 

Power cord with hospital-grade plug

GSI Part # 4204-0251

 

 

(USA)

 

Power cord part # varies depending on location

Mechanical

DIMENSIONS AND WEIGHT

 

 

W x D x H: 20.38 inches x 15 inches x 12.6 inches (LCD raised)

 

52cm x 38cm x 32cm

 

 

Height with LCD lowered - 6 inches (15cm)

 

 

Weight: 16.58 pounds, 7.53 kg

 

 

Shipping weight: 29.50 pounds, 13.38 kg

 

Materials of manufacture

Calibration requirements

Top Case, LCD Housing

 

Front & Rear & Hinges:

Lexan 500 w/10% Glass & 2% BlowingAgent,

 

UL 94V0 Rated

LCD Lense:

GE HP40S -OR- Duralan II

Switches:

Shincor Shin-Etsu/Novacor KE-951 U

Labels:

Lexan & Polycarbonate

Softkey Panel:

Mylar

Probe Cord:

Polyvinyl Chloride (PVC)

Probe Top & Bottom Housings:

Cycolac KJW, UL 94V0 Rated

Probe Tip:

Polypropolene

Eartips:

Kraton 3226

Tubing:

Vinyl and Polyurethane

GSI recommends quarterly calibration checks for the GSI TympStar along with annual certification. ASHArequires quarterly electro-acoustic calibration checks and annual electro-acoustic calibration. It is good practice to perform daily biologic checks. See Chapter 4: Calibration for complete calibration instructions.

GSI TympStar Version 1 and Version 2 Service Manual

2 - 9

Chapter 2

Version 2 Specifications

Standards

The GSI TympStar Version 2 meets or exceeds the following standards and speci-

 

fications for aural acoustic admittance instruments:

 

IEC 1027 1991-03

 

ANSI S3.39-1987

 

ANSI S3.6-1996

 

ANSI S3.7-1995

 

IEC 645-1 1992

 

IEC 126-1973 (also BS 6111-1981)

 

BS ISO 389-2- 1994

 

Y2K Compliant

 

UL 2601-1 Part 1: General Requirements for Safety

 

CSA C22.2 No. 601.1-M90 (Canada)

CE Mark per Medical Device Directive (93/42/EEC) European Representative:

Mr. Leo Hoogendoorn Nicolet Biomedical GMBH Saalackerstrasse 8

63801 Kleinostheim, Germany Tel: 011-49-6027-46980

Fax: 011-49-6027-469815

EN60601-1:1990 Safety Requirements for Medical Electrical Equipment

EN60601-1-2 Medical Electrical Equipment Emissions and Immunity

Requirements

This equipment has been tested for radio frequency emissions and has been verified to meet Radiated and Conducted Emissions per EN 55011-1998, Group 1, Class A and per CISPR, Class A.

2 - 10

Grason-Stadler

 

 

Specifications

 

 

 

Sensitivity ranges

The following admittance measurements give maximum range at standard probe

 

tone frequencies. Compliance “Y”, 226 Hz is measured in ml. All other units are in

 

mmhos. (1 acoustic mmho = 10-8 m3) /Pa.s)

 

 

 

Table 1: Tymp Mode (Y, B, G)

 

 

 

 

 

Frequency

Digital Read-Out

Graphical

 

 

Including Cursor

Display

 

 

 

 

 

 

226 Hz

-7.0 to +7.0

 

-1.0 to +7.0

 

 

 

 

 

 

678 Hz

-21 to + 21

 

-5.0 to +25

 

 

 

 

 

 

1000 Hz

-30 to +30

 

-5.0 to +30

 

 

 

 

 

 

 

Accuracy:

 

 

 

 

 

226Hz:

0.1 ml or 5%, whichever is greater.

 

Above 226Hz:

(F/226) x 0.1 mmho or K% whichever is greater.

 

K FACTOR:

From 250 to 1500 Hz = 5%

 

 

Above 1500 Hz =10%

 

 

 

Table 2: Reflex Mode (Y, B, G)

 

 

 

 

 

Frequency

Digital Read-Out

Graphical

 

 

Including Cursor

Display

 

 

 

 

 

226 Hz

-7.0 to +7.0

-0.16 to +0.80

 

 

 

+0.16 to -0.80

 

 

 

 

 

678 Hz

-21 to + 21

-0.48 to +0.80

 

 

 

+0.48 to -0.80

 

 

 

 

 

1000 Hz

-30 to +30

-0.64 to +0.80

 

 

 

+ 0.64 to -0.80

 

 

 

Accuracy: 226 Hz is 0.02 ml or 5%,whichever is greater.

GSI TympStar Version 1 and Version 2 Service Manual

2 - 11

 

 

Chapter 2

Temporal Latency in

Initial Latency Li:

< 5 ± 5 msec.

ARLT Mode

(From signal onset to 10% ofAmplitude)

 

Terminal Latency Lt:

< 5 ± 5 msec

 

(From signal offset to 90% of Amplitude)

 

Rise Time Tr:

 

<30 ±5 msec.

 

(From 10% to 90% of Amplitude)

 

Fall Time Tt:

 

<25 ±5 msec.

 

(From 90% to 10% of Amplitude)

Multi Frequency

Resonant Frequency Measurement Accuracy: 50 Hz or 5%,

 

 

 

Whichever Is Greater

 

NOTE

 

 

 

Resonant frequency is defined as the frequency at which “Delta B “ is

 

zero (B is measured at + 200 daPa and at “Peak Pressure”).

Probe Signal

Frequencies:

 

 

(Sinusoidal)

Discrete:

226 Hz, 678 Hz, 1000 Hz

 

Multi Freq: From 250 Hz to 2000 Hz

 

Multi Frequency Increment:50 Hz

 

Frequency Accuracy:

 

 

Discrete and Sweep ±1%

 

From 250 to 1000 Hz ±1%

 

Above 1000 Hz

±2%

 

Harmonic Distortion:

<2%

 

(Measured in an HA-1 2cc Coupler)

 

Signal Level (In Real Ear and In Normal Test mode):

 

226 Hz:

85 dB SPL

 

678 Hz:

80 dB SPL

1000 Hz: 75 dB SPL

NOTE

All Probe Tone levels at all frequencies are set to be nominally 70 dB HL.

Signal Level Accuracy:

226 Hz: ±1.5 dB SPL

Other Frequencies: ±3.0 dB SPL

2 - 12

Grason-Stadler

 

 

Specifications

Pneumatic System

PressureMaximumLimits: -800 daPa to +600 daPa

 

Programmed Pressure Ranges:

 

Normal:

+ 200 to - 400 daPa

 

Wide:

 

+ 400 to - 600 daPa

 

Pressure Accuracy:

±10% or ±10 daPa, Whichever is greater in cavities

 

 

 

from 0.5 cc to 5.0 cc.

 

Pressure Sweep Rate:

12.5 daPa/sec

 

 

 

50 daPa/sec

 

 

 

600/200 daPa/sec

 

 

 

200 daPa/sec

 

Manual Sweep Rate Limit: 600 daPa/sec

 

Sweep Rate Accuracy:

±10%

 

Pressure System Leak Rate: < 1. 0 daPa/sec

 

(Measured at -600 and +400 daPa, while pressure servo is disabled.)

Acoustic Reflex

Pure Tone Stimulus

 

Activating (Stimulus)

Frequencies for Contra phone and for Ipsi phone with time multiplexed

Signal

stimulus. See Table 3.

 

 

FrequencyAccuracy: ±1%

 

 

Total Harmonic Distortion (Acoustically): <2%

 

(Measured 5 dB HL below guaranteed maximum HL)

 

Noise Stimulus

 

 

 

The uniformity of the spectrum level of acoustic pressure for the noise signal over 20

 

averages measured acoustically within their respective band limits will be:

 

±10 dB for insert or probe type earphones

 

±5 dB for supra-aural type earphones

 

Noise Band Widths:

 

 

Low Band: 125 -1600 Hz

 

High Band: 1600 - 4000 Hz

 

Broad Band: 125 - 4000 Hz (Relative to level at 1 kHz)

 

Band-edges accurate to within ±15%

 

Roll off rate:

>12 dB/Octave

Stimulus Level Control

Tone Stimulus: The transfer of reference equivalent threshold values are based on the article; “Reference Threshold Levels For The ER-3A Insert Phone”, by Laura Ann Wilber, Barbara A. Krueger and Mead C. Killion, J. Acoustic Soc. Am. Suppl. 1, Vol. 81 Spring 1987. GSI determined the transfer data from an IEC 711 coupler to an ANSI HA-1 coupler. Using this data, the reference threshold values were determined for both the Ipsi and Contra insert earphones for calibration in an ANSI HA-1 2cc coupler.

Noise Stimulus: The transfer of reference threshold values was done by GSI using the “Threshold Determination Method”. The transfer data from an IEC 711 to an ANSI HA- 1 coupler was determined by GSI.

GSI TympStar Version 1 and Version 2 Service Manual

2 - 13

Chapter 2

Intensity levels are reduced as a function of volume at a rate of 1 dB SPL for each .1 ml. Intensity reduction begins at 1.2 ml.

Table 3A: Upper limit of HL range in Reflex Threshold Mode

 

 

 

 

 

 

Pure Tone Stimulus (Hz)

 

 

 

Noise Stimulus

Other Stimulus

 

Probe

 

250

 

500

 

1000

 

2000

4000

 

LBN

 

HBN

BBN

Click

EXT

 

Tone

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(SPL)

(SPL)

Ipsi

226 Hz

 

95

 

110

 

110

 

105

100

 

95

 

95

95

110

110

678 Hz

 

95

 

110

 

110

 

105

100

 

95

 

95

95

110

110

Pulsed

 

 

 

 

 

 

1000 Hz

 

95

 

110

 

110

 

105

100

 

95

 

95

95

110

110

 

 

 

 

 

 

 

Contra

226 Hz

 

110

 

120

 

120

 

120

115

 

115

 

115

115

120

120

678 Hz

 

110

 

120

 

120

 

120

115

 

115

 

115

115

120

120

Pulsed

 

 

 

 

 

 

1000 Hz

 

110

 

120

 

120

 

120

115

 

115

 

115

115

120

120

 

 

 

 

 

 

 

Contra

226 Hz

 

n/a

 

120

 

120

 

120

115

 

115

 

115

115

120

120

678 Hz

 

110

 

n/a

 

120

 

120

115

 

115

 

115

115

120

120

Steady

 

 

 

 

 

 

1000 Hz

 

110

 

120

 

n/a

 

120

115

 

115

 

115

115

120

120

 

 

 

 

 

 

 

 

 

 

 

Lower limit of HL range for all stimulus signals (35 dB HL).

 

 

 

 

 

 

Hearing Level Increment:

 

 

1.0, 2.0, 5.0 dB

 

 

 

 

 

 

 

Hearing Level IncrementAccuracy:

±0.5 dB

 

 

 

 

 

 

 

 

 

Hearing Level Control Linearity:

 

±1.0 dB

 

 

 

 

 

 

 

Table 3B: Upper limit of HL range in ARLT and Reflex Decay Modes

 

 

 

 

 

 

Stimulus Selections (Hz)

 

 

 

Noise Stimulus

Other Stimulus

 

Probe

 

250

 

500

 

1000

 

2000

4000

 

LBN

 

HBN

BBN

Click

EXT

 

Tone

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(SPL)

(SPL)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Contra

226Hz

1

n/a

 

120

 

120

 

120

115

 

115

 

115

115

120

120

678Hz

2

110

 

n/a

 

120

 

120

115

 

115

 

115

115

120

120

 

 

 

 

 

 

Ipsi

226Hz

1

n/a

 

105

 

110

3

105

100

 

n/a

 

n/a

n/a

n/a

110

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

678Hz

2

90

 

n/a

 

110

 

100

100

 

n/a

 

n/a

n/a

n/a

110

 

 

 

 

 

 

1Limits apply for Reflex Decay and ARLT modes.

2Limits apply for Reflex Decay only. ARLT not available at 678 Hz.

3Limits apply for Reflex Decay only. ARLT limit is 105 dB.

2 - 14

Grason-Stadler

Specifications

TABLE 4: UPPER HL LIMIT IN ACOUSTIC REFLEX SENSITIZATION MODE

 

 

 

 

Cf Versus la

 

 

 

Ipsi Actuator

 

Cf (Contra facilitator)

 

 

(Hz)

(HL)

500Hz

1000

2000

4000

6000

BBN

EXT(SPL)

 

 

(dbHL)

(dbHL) (dbHL) (dbHL) (dbHL) (dbHL) (dbSPL)

500

105

---

120

120

115

115

115

120

1000

110

120

---

120

115

115

115

120

2000

105

120

120

---

115

115

115

120

4000

100

120

120

120

---

115

115

120

EXT

101

120

120

120

115

115

115

120

 

 

 

 

Cf Versus Ca

 

 

 

Contra Actuator

 

Ct (Contra facilitator)

 

 

(Hz)

(HL)

500Hz

1000

2000

4000

6000

BBN

EXT(SPL)

 

 

(dbHL)

(dbHL) (dbHL) (dbHL) (dbHL) (dbHL) (dbSPL)

500

120

---

120

120

115

115

115

120

1000

120

120

---

120

115

115

115

120

2000

120

120

120

---

115

115

115

120

4000

115

120

120

120

---

115

115

120

BBN

115

120

120

120

115

115

115

120

EXT

120

120

120

120

115

115

115

120

 

 

 

 

Cf Versus Ca

 

 

 

Contra Actuator

 

If (Ipsi facilitator)

 

 

 

(Hz)

(HL)

500Hz

1000

2000

4000

6000

EXT(SPL)

 

 

(dbHL)

(dbHL) (dbHL) (dbHL) (dbHL) (dbSPL)

 

500

120

---

110

105

100

90

110

 

1000

120

105

---

105

100

90

110

 

2000

120

105

110

---

100

90

110

 

4000

115

105

110

105

---

90

110

 

BBN

115

105

110

105

100

90

110

 

EXT

120

105

110

105

100

90

110

 

 

 

 

 

If Versus la

 

 

 

Ipsi Actuator

 

If (Ipsi facilitator)

 

 

 

(Hz)

(HL)

500Hz

1000

2000

4000

6000

EXT(SPL)

 

 

(dbHL)

(dbHL) (dbHL) (dbHL) (dbHL) (dbSPL)

 

500

105

---

110

105

100

90

110

 

1000

110

105

---

105

100

90

110

 

2000

105

105

110

---

100

90

110

 

4000

100

105

110

105

---

90

110

 

EXT

110

105

110

105

100

90

110

 

GSI TympStar Version 1 and Version 2 Service Manual

2 - 15

 

 

Chapter 2

 

Click Stimulus

Guaranteed Peak Equivalent SPL Levels:

 

 

Ipsi:

110 dB SPL

 

 

Contra:

120 dB SPL

 

 

Peak hold SPL to peak equivalent SPL transfer data.

 

(Peak hold SPL = peak equivalent SPL + transfer data)

 

Ipsi:

8.5 dB

 

 

Contra:

5.5 dB

 

 

ClickRate Range: 50-300 Pulse/sec.

 

Click rate accuracy: ±1pulse/sec.

 

 

Pulse Width (Electrically Measured): 100 msec.

 

Pulse Rise/Fall Time (Electrically):

5.0 msec.

 

Frequency Spectrum:

 

 

Ipsi:

50 - 4000 Hz

 

 

Contra:

50 - 3600 Hz

 

NOTE

Frequency spectrum uniformity better than: 10 dB

 

External Input: At 0.5 VRMS:

 

 

1 KHz Upper Limits:

 

 

Ipsi:

110 dB SPL

 

 

Contra:

120 dB SPL

 

Stimulus

Signal ON/OFF Ratio:

>70 dB

Presentation Control

OFF mode signal level:

<20 dB SPL

 

Signal To Noise Ratio:

>70 dBA

 

Measured with disabled probe signal and “A” weighting for noise

 

measurement.

 

 

Residual noise:

<25 dBA SPL

(Stimulus Present switch in OFF position) Unwanted Acoustic Probe Signals: <60 dBA

Measured while pump is operating and probe tone is disabled. Measure it with “A” weighted filter in “SLOW” Time mode. The noise in normal operating mode will not effect the immittance measurement accuracy.

Signal separation between Ipsi and Contra channels; all frequencies: >70 dB (Measured at the probes with the “ON” channel set to 90 dB)

Leaked signal: <20 dB SPL

The radiated acoustic noise from the instrument (with reflex stim off) when measured at 1 meter from the instrument, shall be: <50 dBA

(“A” weighting and “SLOW” averaging)

2 - 16

Grason-Stadler

 

Specifications

Temporal Specifica-

Steady State Stimulus

tions Of Stimulus

Initial Delay (elapsed time from present bar activation to 10% stimulus

Presentation

amplitude): < 100 msec

 

Terminal Delay (elapsed time from present bar deactivation to 90% stimulus

 

amplitude): <100 msec

 

Rise time: 7.5 ± 2.5 msec

 

Fall time: 7.5 ± 2.5 msec

 

NOTE

 

Initial and terminal delays do not effect temporal measurement of ARLT

 

test (software compensates for them).

 

Multiplexed Stimulus (Used in Reflex Threshold test mode)

 

Period data for frequencies: 250 and 500 Hz:

Period

124

msec

Stimulus on time

44

msec

Stimulus off time

62

msec

Rise and fall time

18

msec

Period data for all other frequencies:

Period

115 msec

Stimulus on time

44

msec

Stimulus off time

53

msec

Rise and fall time

18

msec

 

Temporal Spec. Accuracy:

±10% or 5 msec, whichever is greater

Environmental

The GSI TympStar Version 2 meets ANSI S3.6-1996 (Rl986) Standards for tem-

 

perature and humidity specifications, and it meets the UL 2601 Standards

 

in terms of shock hazards and leakage.

 

 

Temperature:

 

 

 

Storage: -40 degrees C to +75 degrees C

 

 

Operating: +15 degrees C to +35 degrees C

 

Humidity: 90% at 35 degrees C (non-condensing)

Warm-UpTime

At room temperature; +15°C to +35°C:

10 Minutes

 

At room temperature; below +15°C:

1 Hour

GSI TympStar Version 1 and Version 2 Service Manual

2 - 17

 

Chapter 2

 

 

Calibration Stability

All GSITympStar Version 2 specifications are met over the range of specified power

 

line, temperature and humidity variations.

 

 

Power Line:

 

 

 

Voltage Variation:

±10%

 

 

Frequency Variation:

± 5%

 

 

Power line short term variation which affects the performance of the instru-

 

ment will turn off all probe and stimulus signals.

 

Power Rating:

120

Watts maximum

 

Line Voltage Range:

100

VAC to 240 VAC

 

Power Line Frequency Range: 50 - 60 Hz

 

Temperature Operating Range: +15°C to +35°C

 

Relative Humidity Operating Limit: 90%

 

Guaranteed Operating Elevation: 6000 Ft. (1800m)

Connectors

STIMULUS: External Stimulus Input

 

 

(Phone Jack) Peak Voltage: 3 VAC

 

 

Input Impedance: 15,000 Ohms

 

PRESENT: External Present Control Input that turns the stimulus signal ON and OFF (Phone Jack).

Voltage Range: STIM OFF: +5.0 VDC STIM ON: 0.0 v

Input Impedance: 11,000 Ohms Contra PHONE: Output Voltage: 7 VAC

Output Impedance:2.5 Ohms

Standard 9-pin RS232C serial port for interfacing with outside computer.

RXD

CTS

TXD

RTS

Standard VGAport for external monitor.

2 - 18

Grason-Stadler

Specifications

Standard PS-2 keyboard port for external Keyboard.

Standard parallel printer port for external printer.

Electrical

The following apply to theTympStar system:

 

1)

Class 1 Medical Equipment

 

2)

Type B Medical Equipment

 

3)

IPXO ingress of water (ordinary equipment)

 

4)

Equipment not suitable for use in the presence of flammable anesthetic

 

 

mixture with air or with oxygen or nitrous oxide.

 

5)

Mode of operation - continuous

Input Voltage:

100 – 240 VAC

Input Frequency:

50 – 60 Hz

Input Current:

3.2 A maximum

Power Consumption:

120W maximum

GSI TympStar Version 1 and Version 2 Service Manual

2 - 19

 

Chapter 2

 

SuppliedAccessories

Contra insert phone

GSI Part # 8000-0078

 

Calibration cavity (V2)

GSI Part # 2000-1035

 

Probe cleaning wire (2 boxes)

GSI Part # 2000-9610

 

Eartips:

 

 

1 pkg. 8 standard sizes, 4 ea. (Color coded)

GSI Part # 1700-9660

 

1 pkg. 6 special sizes, 2 ea.

GSI Part # 1700-9670

 

1 pkg. 6 screening sizes, 2 ea.

GSI Part # 1700-9622

 

Reference Instruction Manual

GSI Part # 2000-0120

 

Quick User Guide

GSI Part # 2000-0126

 

Printer paper (2 rolls) for orders including printer

GSI Part # 1738-9619

 

Spare set of probe tubing

GSI Part # 2000-9617

 

Probe mount - shoulder

GSI Part # 1700-9646

 

Probe mount - wrist

GSI Part # 1700-9642

 

Probe mount - clothes

GSI Part # 1700-9608

 

Printer paper, adhesive-backed

 

 

for orders including a printer

GSI Part # 1770-9643

 

GSI TympStar dustcover

GSI Part # 1700-9618

 

Power cord with hospital-grade plug

GSI Part # 4204-0251

 

 

(USA)

 

Power cord part # varies depending on location

Mechanical

DIMENSIONS AND WEIGHT

 

 

W x D x H: 20.38 inches x 15 inches x 12.6 inches (LCD raised)

 

52cm x 38cm x 32cm

 

 

Height with LCD lowered - 6 inches (15cm)

 

 

Weight: 16.58 pounds, 7.53 kg

 

 

Shipping weight: 29.50 pounds, 13.38 kg

 

Materials of manufacture

Calibration requirements

Top Case, LCD Housing

 

Front & Rear & Hinges:

Lexan 500 w/10% Glass & 2% BlowingAgent,

 

UL 94V0 Rated

LCD Lense:

GE HP40S -OR- Duralan II

Switches:

Shincor Shin-Etsu/Novacor KE-951 U

Labels:

Lexan & Polycarbonate

Softkey Panel:

Mylar

Probe Cord:

Polyvinyl Chloride (PVC)

Probe Top & Bottom Housings:

Cycolac KJW, UL 94V0 Rated

Probe Tip:

Polypropolene

Eartips:

Kraton 3226

Tubing:

Vinyl and Polyurethane

GSI recommends quarterly calibration checks for the GSI TympStar along with annual certification. ASHArequires quarterly electro-acoustic calibration checks and annual electro-acoustic calibration. It is good practice to perform daily biologic checks. See Chapter 4: Calibration for complete calibration instructions.

2 - 20

Grason-Stadler

Operation summary

3

This chapter summarizes the operation of the TympStar Verrsion 1 and TympStar Version 2 Middle-ear Analyzers. A brief introduction to the TympStar controls, menus and LCD screens will be followed by summarized test procedures for each instrument. Please refer to the Reference Instruction Manuals for more detailed operating instructions.

NOTE

Most generic LCD screen and menu examples use TympStar Version 2 images. However, the concepts and processes described apply equally well to both instruments.

Front panel controls

Front panel controls consist of a combination of “hardkeys” with fixed functions that do not change, and “softkeys” located under the LCD that change to support the changing requirements of a test session. A rotary pressure control is provided to change or fine-tune pressure within the ear canal.

LCD housing

Softkey

Front panel

Pressure control

Hardkey

GSI TympStar Version 1 and Version 2 Service Manual

3 - 1

 

Chapter 3

Hardkeys

All of the push-button controls on the

 

front panel are hardkeys. Three hard-

 

keys are also located on the left and

 

right sides of the LCD housing. Their

 

functions are printed directly on the

 

push-button or nearby on the panel

 

and do not change. The front panel

 

hardkey functions include:

Tymp

Causes GSI (or operator pro-

 

grammed) default criteria for diagnos-

 

tic tymp test to be automatically set

 

up.

Reflex

Causes GSI (or operator pro-

 

grammed) default criteria for reflex

 

threshold testing to be automatically

 

set up.

Etf

Causes GSI (or operator pro-

 

grammed) default criteria for eusta-

 

chian tube function (intact eardrum) test to be automatically set up.

Special

Causes GSI TympStar to initialize to GSI (or operator programmed) default

 

criteria for reflex decay test.

Page

Allows operator to recall and display tests in memory or in progress.

Print

Allows operator to print selected test results.

Remote

Used when GSI TympStar is interfaced with external computer. Implements a

 

common handshake routine to ensure that the RS232 link is in place.

Data Transfer

Used to send test data to an external computer.

Pressure control

Rotary knob used to manually change or fine-tune the pressure within the ear

 

canal.

Manual

Allows the user to run each applicable test procedure manually. The pressure

 

control is used to change pressure manually within the ear canal.

Hold

Allows the current test sequence to be temporarily halted without venting the

 

ear canal pressure.

Stop

Allows the current test sequence to be terminated. The ear canal is vented.

 

Data on the current page is stored in memory.

3 - 2

Grason-Stadler

 

Operation Summary

Start

Causes the selected test sequence to begin in the direction indicated by the associ-

 

ated arrow.

Stimulus

Allows available stimuli to be scrolled up or down and selected. This feature

 

is available only for Reflex type tests.

Intensity

Allows operator to select and set desired intensity (dB HL). This feature is

 

available only for Reflex type tests.

Present

Allows selected stimulus to be presented either manually or according to pre-

 

set automatic timing. This feature is available only for Reflex type tests.

LCD hardkeys

Three hardkeys are also located on the left and right sides of the LCD housing.

 

Their functions are printed directly on the push-button and do not change. The

 

LCD housing hardkey functions include:

Clear

Deletes selected tests stored in memory.

Erase

Erases current display of test results prior to placing test results in memory.

Return

Allows the operator to go back to the next higher level in the softkey menu.

 

Depressing Return while in Clear, Page, or Print mode restores the instrument

 

to the mode previously selected.

Softkeys

The push-buttons located directly under the LCD are the only softkeys. Their

 

functions are displayed on the LCD and change in support of the current test

 

activities.

Test parameters

Test parameter softkeys

GSI TympStar Version 1 and Version 2 Service Manual

3 - 3

Chapter 3

Menus and LCD screens

Hardkeys are located on the front panel and the sides of the LCD panel and provide fixed functions that do not change, such as selection of test type. Softkeys located directly under the LCD change to support the requirements of each test session.

Test parameters

 

Test parameter softkeys

Menus

Selecting a test mode by pressing a hardkey causes the required test screen to be

 

displayed with the appropriate menu of test parameters shown across the bottom of

 

the LCD. The softkeys are then used in conjunction with a few hardkeys to navigate

 

through menus and set parameter values for the selected tests. For example, press-

 

ing the TYMPhardkey causes the Tymp Diagnostic screen to be displayed.

 

GSI-developed default parameters for each selected test are displayed on the

 

lower portion of the LCD above the six softkeys. Alternative choices may be

 

made by pressing the appropriate test parameter softkey.

 

Pressing a test parameter softkey causes a new menu of choices to be pre-

 

sented across the bottom of the screen for that function. The menu structure

 

for the TympStar Version 2 Tymp diagnostic functions is diagrammed on the

 

next page as an example.

3 - 4

Grason-Stadler

Operation Summary

AUTO

PROBE

 

ADMITTANCE

 

P-RANGE

 

EAR

 

 

MORE

SEQUENCE

Hz

 

 

 

 

 

 

 

 

daPa

 

 

 

 

 

226 678

1000

Y

 

B

G

B/G

NORMAL WIDE

RIGHT LEFT

 

 

 

 

 

 

 

ml mmho

mmho mmho

 

 

 

 

 

 

 

P-RATE

 

 

START

 

 

 

BASELINE

GRADIENT

 

CURSOR

MORE

daPa/s

 

 

daPa

 

 

 

 

 

 

 

 

 

 

 

12.5 50 200 600/200

-600

-400

-200

+200

+400

MORE

ON

OFF

TYMP WIDTH

RATIO

OFF

OFF TYMP 1

TYMP 2

TYMP 3

 

 

 

 

 

 

 

 

 

 

daPa

ml

 

SET PEAK

SET PEAK

SET PEAK

GSI default

 

-500

-300

0

+300

MORE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

parameter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The MORE softkey is used to toggle between the top and lower levels of menus

 

 

 

 

while the test screen

portion above the softkeys remains unchanged.

 

 

 

 

 

NOTE

 

 

 

 

 

 

 

 

 

 

 

 

 

The GSI default parameters are circled on the menu structure diagrams.

Program modes

 

 

All test modes allow the user to individually program a unique set of default

 

 

 

 

test parameters. User-programmed parameters override GSI-developed de-

 

 

 

 

fault parameters.

 

 

 

 

 

 

 

Version 1 probe tone

 

The TympStar Version 1 measures compliance using a 226 Hz probe tone.

frequencies

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Version 2 probe tone

 

The following probe tone frequencies are available:

 

 

frequencies

 

 

 

• 226 Hz - All Tests

 

 

 

 

 

678 Hz - Tymp Diagnostic, Reflex Threshold, Reflex Decay, ETF

1000 Hz - Tymp Diagnostic, Reflex Threshold, ETF

250 Hz - 2000 Hz Sweep - Multiple Frequency Tympanometry

Version 1 LCD

The TympStar Version 1 measures compliance, and shows the tracing on the

graphic traces

graphic display as a solid line.

Version 2 LCD

The TympStar Version 2 measures admittance (Y), and its components:

graphic traces

Susceptance (B) and Conductance (G). Admittance (Y) and Susceptance (B)

 

tracings are solid lines, while tracings of conductance (G) values are dotted

 

lines.

GSI TympStar Version 1 and Version 2 Service Manual

3 - 5

Probe lights

Lights are provided on the probe to indicate test status in the following

 

manner:

 

 

• Solid Green

= Test In Progress

 

• Solid Amber

= Occlusion

 

• Blinking Amber

= Leak

 

• Blinking Green

= OK to begin a test

 

Probe lights

LCD screen

The liquid crystal display (LCD) shows axes for the graphic display of the

 

selected test, and pressure and compliance or admittance meters.

THE TEST STATUS line to the right of the graphics area of screen displays alert messages, invalid selections, and other messages to user.

Test Status Line

Test tracings, meter displays with digital readouts, and ear canal volume are recorded in real time.

3 - 6

Grason-Stadler

Operation Summary

Erasing and clearing The display of current test data may be erased prior to stopping a test or continuing test data to the next test by pressing the ERASE hardkey. Previous test data can be selectively cleared from the instrument’s storage by selecting the desired test with the

PAGE hardkey, then pressing the CLEAR hardkey

PAGE hardkey

CLEAR hardkey

ERASE hardkey PRINT hardkey

 

It is suggested that all testing be completed on one ear before proceeding to the

 

other ear. Changing the test ear causes a new page of test data to be generated.

Version 2 External

Reflex test stimuli may be input from an external source and presented via

reflex test stimuli

external control. The connections for these external inputs are provided on the

 

rear panel of the Version 2 instrument. These jacks are unused in the TympStar

 

Version 1.

External reflex test stimuli

External present control

input

input

Paging test data

Up to 26 test screens can be stored in memory as “pages”. These pages of data can

 

be recalled with appropriate titles and labels for viewing by pressing the PAGE

 

hardkey, then scrolling through pages using up/down arrows. Each screen of data

 

equals a page.

GSI TympStar Version 1 and Version 2 Service Manual

3 - 7

 

Chapter 3

Version 2 cursor

Acursor is available in the TympStar Version 2 on all test data pages for scrolling left

 

or right on the graphics area and can be used to set or reset tymp peak values for

 

tests. Acursor box describes the numeric position of the cursor on both the X and

 

Y axes. Cursors are not available in the Version 1.

Cursor Cursor box

 

NOTE

 

The cursor must be used to set tymp values when testing at anything other

 

than 226Hz, Y in the Version 2.

Printing tests

Up to 26 test screens stored in memory can be printed by selecting PAGE-ALL

 

followed by PRINT. Individual test results can be printed by selecting PAGE

 

and then using the arrow key to display the test. To begin printing, press the

 

PRINT hardkey.

Preparing an external The TympStar supports the full line of Hewlett Packard DeskJet® and LaserJet® printer printers. These printers provide full sheet printouts of test data. A test printout

of results containing:

One left and right tympanogram graph

A reflex threshold table for both left and right (Ipsi or Contra only)

A reflex decay table for both left and right (Ipsi or Contra only) will print on one 8.5 x 11 inch sheet of paper.

1)Turn the TympStar and printer off, then attach the printer data cable to the printer connector on the rear of the TympStar. Connect the printer power cable to AC power.

WARNING

ANY EQUIPMENT CONNECTED TO THE GSI TYMPSTAR AND USED IN THE PATIENT VICINITY MUST BE POWERED BYAN ISOLATED POWER SOURCE TO MAINTAIN THE ELECTRICAL SAFETY OF THE OVERALL SYSTEM. The isolated power source can be purchased directly from GSI, or elsewhere when approved for use by GSI.

3 - 8

Grason-Stadler

 

 

 

Operation Summary

 

2)

Turn the TympStar and printer on. Anytime the external printer is connected

 

 

and turned on, the TympStar will automatically direct printing to the external

 

 

printer. To redirect printing to the internal printer, turn the external printer off or

 

 

disconnect it.

Setting the printer

To set the external printer parameters:

parameters

 

 

 

 

 

 

 

1)

Press the ETF hardkey to select the

 

 

ETF test mode, then press the RETURN

 

 

hardkey to display the ETF sub-menu.

 

2)

Press the INSTRUMENT OPTIONS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INTACT

PERF

INSTRUMENT

PROGRAM

TM

TM

OPTIONS

MODE

softkey, then the MORE softkey to display the PRINT FORMAT softkey selection.

3)Press the PRINT FORMAT softkey, and then the MORE softkey to display the EXTERNAL PRINTER SETUP softkey selection.

GSI TympStar Version 1 and Version 2 Service Manual

3 - 9

Chapter 3

4)Press the EXTERNAL PRINTER SETUP softkey to display the External Printer Setup Menu.

Left column printout

5) Press the LEFT COLUMN SETTING softkey of the External Printer Setup

data

Menu to set the left column of the printed data to the left or right ear.

Color or black and

6) Press the PRINT TYPE softkey of the External Printer Setup Menu to set the

white

printout to Black and White or Color.

NOTE

The setting must be Black and White for grayscale printers.

3 - 10

Grason-Stadler

+ 174 hidden pages