GE OEC Elite, MiniView C-arm User manual

4.7 (19)
GE OEC Elite, MiniView C-arm User manual

GE Healthcare

OEC Elite

MiniViewC-arm

Digital Mobile C-arm

Technical Data

Changing the Mini C-arm Experience

Cannot be placed on the market or put into service until it has been made to comply with the Medical Device Directive requirements for CE marking.

Specifications

Cannot be placed on the market or put into service until it has been made to comply with the Medical Device Directive requirements for CE marking.

X-ray System

X-Ray Tube

-Compact Monoblock

Focal Spot

-0.033 mm*

Generator

-12.8 W

kVP range

-40 - 80 kVp

mA range

-0.02 - 0.16 mA

Maximum Output

-0.16 mA 80 kVp

Dose Rate Calculation

-AKR/DAP

Imaging System

Detector

-CMOS flat detector

FOV Size

-4” (10 cm) Mag

-5” (13cm) Normal

Detector Size (resolution)

-1.3 k x 1.3 k pixels

Pixel Size (spacing)

-100 microns

DQE

-(0) - 70%

Viewing System

Monitors

-Dual medical 19” (48 cm)

-Monochromatic LCD

-Anti-glare panels

Monitor Resolution

-1280 x 1024 pixels

Displayed image: primary live image

-11.4” (29 cm) diameter

Displayed image: reference image

-11.4” (29 cm) diameter

Image Size

-1.3 k x 1.3 k 16 bit image processing

-1 k x 1 k displayed image

Tilt Motion

-10° up / 10° down

Swing + Swivel Rotation Motion

-180° + 270°

Viewing Angle

-170° horizontal & vertival

Max Panel Brightness

-1400 Cd/m2

Max Contrast Ratio

-1000:1

Touch Screen

-Right monitor

Surgeon Tube Head Controls

X-ray exposure button

Dual tube head control panels

-Auto Technique

-Low Dose Mode (max .08 mA -

80 kVp)

-Manual kV/mA adjustment

-Save

-Swap

-Print

-Laser Aimer

-Magnification

-Image rotate

-Auto Brightness/Contrast

-Manual Brightness/Contrast adjustments

-Alarm Reset

-Smart Lock Button

Imaging Features

Annotation/Measurements

Smart Metal

-Detects metal in the field and optimizes image quality

AutoTrak

-Automatically seeks anatomy anywhere in the field and selects optimal technique

Zoom-Roam

Auto X-Ray Technique Control

Edge Enhancement

Last Image Hold

Auto-save/Auto swap

-Configurable

Noise Reduction

Motion Artifact Reduction

Flip/Invert

Negate

C-arm Physical Specifications

Balanced Pivot C /Fork

-Carbon fiber Composite material

Smart Lock Button

-Anti Drift Locking Mechanism

-Single button locks 4 joints

-Electro-mechanical

SID

-17.6’’ (45cm)

Free Space

-13.4” (34 cm)

C-arm Depth

-18” (46 cm)

Frictionless Orbital Rotational Sleeve

-120° (90° Underscan, 30° Over

Scan)

C-arm Horizontal Travel

-77.2” (196 cm)

Counterbalanced C-arm Vertical Travel

-33.4” (85cm)

C-arm Panning Motion

-366°

C-arm Lateral Rotation

-380° (+/- 190°)

Weight

-220 kg (485 lbs)

Foot Print

-0.64 m²

Size

-70.5” x 29.0” x 38.3” (179 x 74 x 97 cm)

C-arm Motion

-5 joints: Orbital, Lateral, Upper and Lower Horizontal, Vertical

Footswitch

-Wired

-Configurable dual pedal ° Save, Print, Save & Print

Skin Spacer mount

Electrical

Backup Battery

Input Power (60/50 Hz)

-100/110/115/120/127VAC @ 4.5A

-200/220/230/240VAC @ 2.3A

Auto Power Sensing

Data Management & Connectivity

Water Resistant Keyboard

Removable Storage

-(3) USB Ports

Hard Drive Storage

-100,000 images

Printer

-optional

DICOM (3.0)

-MPPS

-RDSR - Radiation Dose Structured

Report

-Query and Retrieve

-Modality Worklist

-Print

-Storage

-Storage Commitment

Wireless DICOM

-optional

Wireless Frequency

-2.4G, 5G

Digital Video Interface

Network Interface

Room Control Interface

Regulatory Compliance

IEC60601-1

U.S. 21 CFR Subchapter J