URGENT RECALL NOTICE

PLEASE TAKE ACTION TO INFORM ALL USERS OF THE RELEVENT SYSTEM(S) OF

THESE ISSUES AND HOW TO ADDRESS THEM

February 21 2007

 

To:

Hospital Administrator

 

Director/Manager of Radiology

Subject:

Product Safety Issues

Affected Products:

OEC UroView 2800, OEC 9800 Plus Digital, OEC MiniView 6800, OEC

 

FlexiView 8800 Mobile C-Arm

Our records indicate that your facility has one or more of the following GE Healthcare OEC (GEHC OEC) devices that has received a replacement hard disk drive:

OEC UroView 2800OEC 9800 Plus DigitalOEC MiniView 6800

OEC FlexiView 8800 Mobile C-Arm

GEHC OEC has discovered an issue with the hard disk drives that were installed on the above listed products during service calls. The manufacture dates of the affected hard disk drives are between October 15, 2004 and December 8, 2004

This issue could potentially result in the system failing to store images when the hard disk drive capacity exceeds 170 images instead of the expected 400 images. Systems may also fail to boot or lose patient data as this image capacity is approached.