GE OEC 9800, OEC 2800, OEC 6800, OEC 8800 User manual

5 (1)

GE Medical Systems

Technical

Publications

00-879878-01 Revision C

GE OEC 9800, 2800, 6800, 8800

1K x 1K Workstation X-ray Systems CONFORMANCE STATEMENT for DICOM

Copyright© 2002 By General Electric Co.

Do not duplicate

 

GE

OEC 9800, 2800, 6800 8800 X-ray Systems

GE MEDICAL

SYSTEMS

DICOM CONFORMANCE STATEMENT

00-879878

Rev C

 

 

 

 

 

 

Table

of Contents

 

 

0.

INTRODUCTION

....scsssccssssssscccsssssssecssecssscsssccscesssssnscsscsessesscscsssscsscssccessscossesscsosseseess

4

I.

IMPLEMENTATION MODE LL.,.....ssscsssssesssssssssssoccssssccsssscesessssscescsessscsesssessssesseesees

89

1.1

Application

data

flow

diagram

 

 

9

1.2

Functional

Definition

of AE’s

 

 

10

 

1.2.1

Storage

Application

Entity

ss

 

10

 

1.2.2

Query Application Entity sister

10

 

1.2.3

Printer

Application

Entity

eee

eeseereseanen

11

1.3

Sequencing

of Real World Activities

 

 

11

 

AE

Specification

....sccssscssssccssssssscsssssccsveccsccsssscssscsscscccessccessssssessevescssessescssescaseseseees

11

2.1

Storage AE

Specification

 

 

11

2.1.1Association Establishment Policies...eee nenene nene tn 11

 

2.1.2

Association

Initiation by Real-World Activity...

ee

12

2.2

Query

AE

Specification

 

20

 

2.2.1

Association

Establishment Policies

 

nee nenene 20

 

2.2.2

“Association Initiation by Real-World Activity .Ne

21

2.3

Printer

AE

Specification

 

24

2.3.1AssociationEstablishmentPolci«............................................ 24

2.3.2Association Initiation by Real-World Activity...eee een eee eee ne nene 25

24

Verification

AE

Specification

 

 

 

28

 

24.1

Association

Establishment

Policies

 

ss

 

28

 

2.4.2

Association

Initiation by Real-World Activity .be

 

29

 

COMMUNICATION PROFILES..........serrrrrrererrrrrcrrererrererereeresenrescore oneereseseneneciosene

29

3.1

Supported

Communications

stacks

 

 

 

29

3.2

TCP/IP

Stack

 

 

 

 

 

29

 

3.2.1

Physical Media Supported ie

 

 

29

4.

Extensions

/ Specializations

/ 77 7 075.

 

 

29

5.Configuration ..........00001s0000s0030s0003000000550005000000000000000000005000000000000000000000000090000n000002008 3029

5.1

AE Title/presentation address

mapping

30

5.2

Configurable parameters

 

30

6.

Support of extended character

sels..............

0xs00500s000000000000000000000000000000000030000n00000000009 30

 

GE

OEC 9800, 2800, 6800 8800 X-ray Systems

GE MEDICAL

SYSTEMS

DICOM CONFORMANCE STATEMENT

00-879878

Rev C

 

0.INTRODUCTION

0.1OVERVIEW

This DICOM Conformance Statement is divided into Sections with Section 0 (Introduction) describing the overall structure, intent, and references for this Conformance Statement

0.2 OVERALL DICOM CONFORMANCE STATEMENT DOCUMENT STRUCTURE

The Documentation Structure of the GEMS Conformance Statements and their relationship with the DICOM Conformance Statements is shown in the Illustration

below.

 

GE

OEC 9800, 2800, 6800 8800 X-ray Systems

GE MEDICAL

SYSTEMS

DICOM CONFORMANCE STATEMENT

00-879878

Rev C

 

ID/Net v3.0

Integrated

DICOM/Network

(ID/Net v3.0)

Conformance

Statement

Direction: 211878

APPLICATION ENTITY SPECIFICATION

(SERVICE CLASSES, INFORMATION OBJECTS, MESSAGE EXCHANGES, ETC.)

 

OEC 9800

 

P rod u ct

Conformance

Conformance

Implementation:

Statement

Statement

Direction:

Direction:

 

00-879878

 

DICOM STANDARD

Standard

Specification:

DICOM

Part 18

 

GE

OEC 9800, 2800, 6800 8800 X-ray Systems

GE MEDICAL

SYSTEMS

DICOM CONFORMANCE STATEMENT

00-879878

Rev C

 

This

document

specifies

the

DICOM

 

implementation.

 

It is entitled:

 

 

 

 

 

 

 

 

 

GE

OEC

1K x

1K

Workstation

X-ray

Systems

 

 

 

 

 

 

 

 

 

 

 

 

 

Conformance

Statement

for

DICOM

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

00-879878

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This

DICOM

Conformance

Statement

documents

the

 

DICOM

Conformance

Statement

and

Technical

Specification required to interoperate with the GEMS

network

interface.

Introductory information, which is applicable to

all

GEMS

Conformance

Statements,

is

described

in

the

document:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Introduction to the Integrated DICOM/Network

 

v3.0 (ID/Net

v3.0)

 

 

 

 

 

 

 

Conformance

Statement

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Direction:

2118780

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This

Introduction

familiarizes

the

reader with

DICOM

terminology

and

general

concepts.

It should be

read

prior

to

reading

the

 

individual

products'

GEMS

Conformance

 

 

Statements.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The GEMS Conformance Statement, contained

in

this

document,

also

specifies

the

 

Lower

Layer

communications

which

 

it supports

(e.g.,

TCP/IP).

 

However,

the

Technical

Specifications

are

defined

in

the

DICOM

Part

8

standard.

 

 

 

 

 

 

 

 

 

For more information including Network Architecture and

basic

 

DICOM

concepts,

 

 

please

refer

to

the

Introduction.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

For more information regarding DICOM, copies

of

the

 

Standard

may

be

obtained

on

the

Internet

at

http://medical.nema.org.

 

Comments

on

the

Standard

may

be

addressed

to:

 

 

DICOM

Secretariat

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NEMA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1300

N.

17"

Street,

Suite

1847

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rosslyn,

VA

22209

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

USA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Phone:

+1.703.841.3200

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.3

INTENDED

AUDIENCE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

reader

 

of

this

document

is concerned

with

software

design

and/or

system

 

 

integration

issues.

It

is

assumed

that the reader of this

document

is

familiar

with

the

 

DICOM

Standards

and

with

the

terminology

and

concepts which

are

used

in

those

 

Standards.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If readers

are

unfamiliar

with

DICOM

 

terminology

they

 

should

first

refer

to

the document

listed

below,

then

read

the

DICOM

Standard

itself,

prior

to reading

this

DICOM

 

 

Conformance

Statement

document.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GE

OEC 9800, 2800, 6800 8800 X-ray Systems

GE MEDICAL

SYSTEMS

DICOM CONFORMANCE STATEMENT

00-879878

RevC

 

Introduction to the Integrated DICOM/Network (ID/Net v3.0)

Conformance Statement

Direction: 2118780

0.4 SCOPE AND FIELD OF APPLICATION

It is the intent of this document, in conjunction with the Introduction to the Integrated DICOM/Network (ID/Net v3.0) Conformance Statement, Direction: 2118780, to provide an unambiguous specification for GEMS implementations. This specification, called a Conformance Statement, includes a DICOM Conformance Statement and is necessary to ensure proper processing and interpretation of GEMS medical data exchanged using

DICOM.

 

The

GEMS Conformance

Statements

are

available to

the

public.

 

 

The

reader of

this

DICOM

Conformance

Statement

should

be

aware that

different

GEMS

devices

are capable of using different Information Object

Definitions.

For

example,

a GEMS

CT

Scanner

may

send

images

using

the

CT

 

Information

Object, MR

Information Object, Secondary Capture Object, etc.

 

 

 

 

 

 

 

 

 

Included

in this DICOM

Conformance

Statement

are

the

Module

Definitions

which

define

all

data

elements

used

by this

GEMS

implementation.

If

the

user

encounters

unspecified private data elements while parsing a GEMS

Data

Set,

the user

is

well

advised to ignore those data

elements

(per

the

DICOM

standard).

Unspecified

private

data

element

information

is

subject

to

change

without notice.

If,

however,

the

device is

acting as a "full fidelity storage device", it should retain and re-transmit all of the private data elements which are sent by GEMS devices.

0.5

IMPORTANT

REMARKS

 

 

 

 

 

 

 

 

 

The

use

of

these

DICOM

Conformance

Statements,

in

conjunction

with

the DICOM

Standards,

is

intended

to facilitate

communication

with

GE

imaging

equipment.

However,

by

itself, it

is

not

sufficient

to

ensure

that

inter-operation will be

successful.

 

The

user

(or

user's

agent)

needs

to

proceed

with caution

and address

at least four

issues:

 

 

 

 

 

 

 

 

 

 

 

 

©Integration - The integration of any device into an overall system of

 

interconnected

 

devices

goes

beyond

 

the

 

scope

of

standards

 

(DICOM),

and

of

this

introduction

 

and

associated

DICOM

 

Conformance

 

Statements

when_

interoperability

 

with

non-GE

 

equipment is desired. The responsibility

 

to

analyze

the

applications

 

requirements and to design a solution

that

integrates

 

GE

imaging

 

equipment

with

non-GE systems is

the

user's

responsibility

and

 

should

not

be

underestimated.

The

user

is

strongly

advised

to

 

ensure

that

such an

integration analysis

is

correctly performed.

 

©

Validation

-

Testing

the

complete

range

of

possible

interactions

 

between

any

 

GE

device

required

for

 

their

connection

with

GE

 

devices.

This

 

includes

the

accuracy

of

the image data once it

has

 

crossed

the

interface

between

the GE

imaging

equipment

and

the

 

non-GE device and non-GE devices,

 

before

the

connection

is

 

declared operational, should not be overlooked. Therefore,

the user

 

should

ensure

that

any

 

non—GE

provider

accepts

full

responsibility

for

 

all validation and the stability of the image

 

data

for

 

the

intended

 

applications.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GE

 

OEC

9800,

 

2800, 6800 8800 X-ray Systems

GE

MEDICAL

SYSTEMS

 

 

 

 

 

 

 

 

 

 

DICOM

CONFORMANCE

 

STATEMENT

00-879878

Rev

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Such a validation is required before any clinical use (diagnosis and/or

 

 

 

 

 

 

 

treatment)

is

performed.

It

applies

when

images

acquired

on

GE

 

 

 

 

 

 

 

 

 

imaging equipment

are

processed/displayed on a non-GE

device,

as

 

 

 

 

 

 

 

well as when images acquired on non-GE equipment is

 

 

 

 

 

 

 

 

 

 

 

processed/displayed

 

on

a

GE

console

 

or

workstation.

 

 

 

 

 

 

 

 

 

 

 

 

©

Future

Evolution

-

GE

understands

 

that

the

DICOM

Standard

will

 

 

 

 

 

 

 

evolve

to

meet

the user's growing requirements. GE is actively

 

 

 

 

 

 

 

involved

in

the

development

of

 

the

 

DICOM Standard.

DICOM

will

 

 

 

 

 

 

 

incorporate

new

features and technologies and GE may follow the

 

 

 

 

 

 

 

evolution

of

the

Standard.

The

GEMS

 

protocol

is

based

on

DICOM

 

 

 

 

 

 

 

as

specified

 

in

each

DICOM

Conformance

Statement.

Evolution

of

 

 

 

 

 

 

 

the

Standard

 

may require changes to devices,

 

 

which

have

 

 

 

 

 

 

 

implemented

 

DICOM.

In addition, GE reserves the

right

to

 

 

 

 

 

 

 

discontinue or make changes to the support of communications

 

 

 

 

 

 

 

features

(on

 

its products) reflected on by these DICOM

 

 

 

 

 

 

 

Conformance

Statements.

The

 

user

 

should

ensure

 

that any

non—

 

 

 

 

 

 

 

GE

provider,

 

which

 

connects with

GE

devices,

also

plans

for

the

 

 

 

 

 

 

 

future

evolution

of

the

DICOM

Standard.

Failure

to

do

so

will

likely

 

 

 

 

 

 

 

result in the loss of function

and/or

 

connectivity

as

the

DICOM

 

 

 

 

 

 

 

Standard changes and GE Products are enhanced to support

these

 

 

 

 

 

 

 

changes.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

e

Interaction

-

it

is the sole responsibility of the non-GE

provider

to

 

 

 

 

 

 

 

ensure

that

communication

with

 

the

interfaced

equipment

does

not

 

 

 

 

 

 

 

cause

degradation

 

of

GE

imaging

equipment

performance

and/or

 

 

 

 

 

 

 

function.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.6

REFERENCES

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A list of

references, which

is

applicable

to

all

GEMS

Conformance

Statements,

is

 

 

 

included

in

the

/ntroduction

to the

Integrated

DICOM/Network

 

(ID/Net

v3.0)

 

 

 

 

 

 

Conformance

Statement,

Direction:

2118780.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The information object implementation refers to DICOM

PS

3.3

(Information

 

Object

 

 

Definition).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.7

DEFINITIONS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A set of definitions, which

is

applicable

to

 

all

GEMS

 

Conformance

Statements,

is

 

 

 

included in the Introduction to the Integrated DICOM/Network (ID/Net v3.0)

 

 

 

 

 

 

Conformance

Statement,

Direction:

2118780.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.8 SYMBOLS AND ABBREVIATIONS

A list of symbols and abbreviations, which is applicable to all GEMS Conformance Statements, is included in the Introduction to the Integrated DICOM/Network (ID/Net v3.0) Conformance Statement, Direction: 2118780.

1. IMPLEMENTATION MODEL

Patient demographics may be obtained prior to a procedure using DICOM Query. After producing and storing digital images locally, the GE OEC Systems may be commanded

 

 

GE OEC 9800, 2800, 6800 8800 X-ray Systems

GE MEDICAL

SYSTEMS

DICOM

CONFORMANCE

STATEMENT

00-879878

RevC

 

 

 

 

 

to store single images or cine data to

DICOM compatible

storage

devices,

or may

be

commanded

to print images to DICOM

compatible printers. This

section presents

the

Application data flow diagram and provides functional definitions of the Application Entities and their related Real-World Activities.

1.1

Application data

flow

diagram

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure

1.1

is the Application data flow

diagram for the implementation model.

This

 

diagram represents the Application Entities present

in

this

implementation

and

 

 

 

graphically

 

depicts

the

relationship

of

the AE’s to real world

 

activities.

 

 

 

 

 

 

When

the

user has selected a single image or a sequence

 

of images (CINE loop) for

 

storage,

the

storage

AE

initiates

an

association with

the

storage

device.

After

verifying

the

device

is on-line, the AE

will

send

the

image

or

CINE

loop

selected

to the

device.

 

The association is then closed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When

the

user

selects Query RIS, the

Query

AE

opens

an

 

association

with

a

specified

server

and

 

issues a C-FIND

Request.

The

request

contains

keys

which

may

be

set

by

the

user.

 

The C-FIND responses received are displayed by

patient name.

The

user

 

may

then

select

one

of

the

patients

to

be entered

into

the

Patient

Information

screen

of

the x-ray device.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When

the

user

 

has

selected

one

or

more

single

images

to

print,

the

Printer

AE

initiates

an association with the printing device.

After

verifying

that

the

device

is

on-line

and

 

ready to

print,

the

AE

will send the appropriate

number

 

of

images to the device for

 

printing

 

until

all

 

of

the

selected images

are

printed.

The

association

will

then

be

closed.

 

 

GE

OEC 9800, 2800, 6800 8800 X-ray Systems

GE MEDICAL

SYSTEMS

 

DICOM CONFORMANCE STATEMENT

00-879878

RevC

 

 

 

 

 

DICOM

Store

 

 

Standard

 

 

Interface

 

 

 

Image

 

 

|

 

 

Storage AE

|

Verify

| Verification AE

!

Image Archive

 

 

|

Query/WorkList

 

 

|

Grayscale Print

 

 

|

Verification

 

Query/WorkList

|

AE

 

AE

 

 

>

 

|

GrayScale

|

Print

AE

|

Figure 1.1 Application

Data

Flow Diagram

1.2Functional Definition of AE’s

1.2.1Storage Application Entity

The Storage AE acts as an SCU and provides the DICOM Storage Service Class

support

required to

transfer X-ray RadioFluoroscopic,

X-ray

Angiographic

Images

and

X-ray cine loops to

an archive SCP.

Secondary Capture is

also supported.

The

AE

includes

Association

Negotiation for

the X-ray Storage

SOP

Class.

 

 

1.2.2Query Application Entity

The Query AE acts as an

SCU

and

provides the DICOM Query/Retrieve Service

Class

support

required to

Query

the

RIS

Server or Worklist Broker.

The

AE issues C-FIND

requests

using the

STUDY

level of

the

Patient Root

Query/Retrieve

Information

Model.

If Modality is specified as

a search

key,

the system

performs

a secondary Series

level

10