Defibrillator Analyzer Specifications
Energy Output Measurement
Compatible Defibrillator Waveshapes ................ Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Note
AC Pulsed Biphasic waveform has not been approved in the United States.
Autoranged Measurement.................................... |
0.1 to 600 J |
Accuracy |
|
0.1 to 360 J ..................................................... |
±(1 % of reading + 0.1 J) |
360 to 600 J .................................................... |
±(1 % of reading + 0.1 J), typical |
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges.
Load resistance |
|
Resistance .......................................................... |
50 Ω |
Accuracy ......................................................... |
±1 %, non-inductive (<2 μH) |
Pulse trigger level ................................................. |
20 V |
Pulse width |
|
Range ................................................................. |
1.0 to 50.0 ms |
4/08 |
1 |
Manual Supplement |
Impulse 6000D/7000DP Users |
Accuracy............................................................. |
±0.1 ms |
Voltage |
|
Range ................................................................. |
20 to 5000 V |
Accuracy............................................................. |
±(1 % of reading + 2 V) |
Current |
|
Range ................................................................. |
0.4 to 100.0 A |
Accuracy............................................................. |
±(1 % of reading + 0.1 A) |
Tilt (biphasic and pulsed biphasic) |
|
Range ................................................................. |
1 % to 99 % |
Accuracy............................................................. |
±1 digit |
Interphase delay (biphasic and pulsed biphasic) |
|
Range ................................................................. |
0.1 ms to 9.9 ms |
Accuracy............................................................. |
±0.1 ms |
Frequency (pulsed biphasic only) |
|
Range ................................................................. |
2000 Hz to 8000 Hz |
Accuracy............................................................. |
±1 % of reading |
Duty cycle (pulse biphasic only) |
|
Range ................................................................. |
1 % to 99 % |
Accuracy............................................................. |
±1 digit |
Sample rate ........................................................... |
250 kHz (4 μs sample) |
Maximum Average Power .................................... |
12 W, equivalent to 10 defib pulses of 360 J every 5 minutes |
2 |
8/08 |
Impulse 6000D/7000DP Users |
Manual Supplement |
Scope Output |
|
Autorange ........................................................... |
2000:1, 400:1 and 80:1: dependant on the range |
Waveform Playback |
|
Output ............................................................. |
BNC |
Output impedance........................................... |
50 Ω |
Amplitude Accuracy ........................................ |
±5 % |
Charge Time Measurement |
|
Range ................................................................. |
0.1 to 100.0 s |
Accuracy ............................................................. |
±0.05 s, typical |
Synchronization Test (Elective Cardioversion) |
|
Delay Time Measurement |
|
Timing window ................................................ |
ECG R-wave peak to the defib pulse peak |
Range ............................................................. |
-120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 |
|
ms following the R-wave peak. |
Resolution....................................................... |
1 ms |
Accuracy ......................................................... |
±1 ms |
ECG waves |
|
Normal Sinus Rhythm (NSR).......................... |
10 to 180 (by 1) BPM |
Atrial fibrillation................................................ |
Coarse and fine |
Monomorphic Ventricular Tachycardia ........... |
120 to 240 (by 5) BPM |
Asystole .......................................................... |
Flat line |
Automated Defibrillator Test ECG Waves |
|
Normal Sinus ...................................................... |
10 to 300 (by 1) BPM |