Defibrillator Analyzer Specifications

Energy Output Measurement

Compatible Defibrillator Waveshapes ................ Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic

Note

AC Pulsed Biphasic waveform has not been approved in the United States.

Autoranged Measurement....................................

0.1 to 600 J

Accuracy

 

0.1 to 360 J .....................................................

±(1 % of reading + 0.1 J)

360 to 600 J ....................................................

±(1 % of reading + 0.1 J), typical

Note

For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges.

Load resistance

 

Resistance ..........................................................

50 Ω

Accuracy .........................................................

±1 %, non-inductive (<2 μH)

Pulse trigger level .................................................

20 V

Pulse width

 

Range .................................................................

1.0 to 50.0 ms

4/08

1

Manual Supplement

Impulse 6000D/7000DP Users

Accuracy.............................................................

±0.1 ms

Voltage

 

Range .................................................................

20 to 5000 V

Accuracy.............................................................

±(1 % of reading + 2 V)

Current

 

Range .................................................................

0.4 to 100.0 A

Accuracy.............................................................

±(1 % of reading + 0.1 A)

Tilt (biphasic and pulsed biphasic)

 

Range .................................................................

1 % to 99 %

Accuracy.............................................................

±1 digit

Interphase delay (biphasic and pulsed biphasic)

Range .................................................................

0.1 ms to 9.9 ms

Accuracy.............................................................

±0.1 ms

Frequency (pulsed biphasic only)

 

Range .................................................................

2000 Hz to 8000 Hz

Accuracy.............................................................

±1 % of reading

Duty cycle (pulse biphasic only)

 

Range .................................................................

1 % to 99 %

Accuracy.............................................................

±1 digit

Sample rate ...........................................................

250 kHz (4 μs sample)

Maximum Average Power ....................................

12 W, equivalent to 10 defib pulses of 360 J every 5 minutes

2

8/08

Impulse 6000D/7000DP Users

Manual Supplement

Scope Output

 

Autorange ...........................................................

2000:1, 400:1 and 80:1: dependant on the range

Waveform Playback

 

Output .............................................................

BNC

Output impedance...........................................

50 Ω

Amplitude Accuracy ........................................

±5 %

Charge Time Measurement

 

Range .................................................................

0.1 to 100.0 s

Accuracy .............................................................

±0.05 s, typical

Synchronization Test (Elective Cardioversion)

 

Delay Time Measurement

 

Timing window ................................................

ECG R-wave peak to the defib pulse peak

Range .............................................................

-120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380

 

ms following the R-wave peak.

Resolution.......................................................

1 ms

Accuracy .........................................................

±1 ms

ECG waves

 

Normal Sinus Rhythm (NSR)..........................

10 to 180 (by 1) BPM

Atrial fibrillation................................................

Coarse and fine

Monomorphic Ventricular Tachycardia ...........

120 to 240 (by 5) BPM

Asystole ..........................................................

Flat line

Automated Defibrillator Test ECG Waves

 

Normal Sinus ......................................................

10 to 300 (by 1) BPM

4/08