Excell ELW Max, ESW Max User Manual

User Manual
ELW Max / ESW Max
© Excell Precision Limited 2017. All rights reserved Worldwide. The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorized by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.
02026010 ZSME300000369
1
EXCELL PRECISION CO., LTD.
Table of Content
INSTRUCTION FOR USE ................................................................................................... 2
PREPARING TO USE THE SCALE ..................................................................................... 2
CHAPTER 1 INTRODUCTION ............................................................................................ 3
1-1 FEATURES AND SPECIFICATIONS ..................................................................................... 3
1-2 POWER SUPPLY ............................................................................................................ 3
1-3 PANEL AND KEYBOARD INTRODUCTION............................................................................ 4
1-4 OPERATING THE SCALE ................................................................................................. 5
1-5 SELF-TEST MODE ......................................................................................................... 6
1-6 ERROR MESSAGES ....................................................................................................... 7
1-7 WEIGHT UNIT ............................................................................................................... 7
CHAPTER 2 ADVANCED FUNCTIONS .............................................................................. 8
2-1 ADVANCED FUNCTION SETTING TABLE ............................................................................ 8
2-2 ADVANCED FUNCTION SETTING WORKFLOW ................................................................... 9
2-3 GENERAL FUNCTION SETTING .............................................................. 10
2-3-1 Automatic Backlight Function Setting ....................................... 11
2-3-2 Automatic Power-off Timer Setting ........................................... 12
2-3-3 Hi/Lo/OK Function Setting ........................................................ 13
2-3-4 Restore to the Default Setting ................................................... 14
2-3-5 Noise Filter Setting ................................................................... 15
2-3-6 Hold Function Setting .............................................................. 16
2-3-7 Auto Unit Weight Averaging Setting ........................................ 17
2-3-8 Two Weighing Units Setting ..................................................... 18
2-3-9 Unstable Tare ........................................................................... 19
2-4 WEIGHT CALIBRATION .............................................................................. 20
APPENDIX 1 COMMAND MODE &OUTPUT DATA FORMAT .......................................... 21
APPENDIX 2 7-SEGMENT DISPLAY CHARACTERS....................................................... 23
02026010 ZSME300000369
2
EXCELL PRECISION CO., LTD.
Instruction for Use
1. Please keep the scale in a cool dry place. Do not store it at high temperature.
2. Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the weigh pan to any strong shock loads.
3. The load placed on the weigh pan must not exceed the maximum weighing capacity of the scale.
4. If the scale is not going to be used for some time, please clean it and store it in a plastic bag in dry conditions. A desiccant sachet may be included to prevent any moisture build up.
Preparing To Use the Scale
1. Adjust the four levelling feet (if fitted) to set the scale pan level using the spirit level
bubble located at the front of the scale.
2. Avoid operating the scale in direct sunlight or drafts of any kind.
3. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to
other appliances to minimise the possibility of interference affecting the performance of the scale.
4. Remove any weight that might be on the weigh pan
before the scale is switched on and avoid leaving weight on the pan for long periods of time
5. All goods weighed should be placed in the centre of
the weigh pan for accurate weighing. The overall dimensions of the goods being weighed should not exceed the dimension of the weigh pan.
6. We suggest to warm up the scale for 15~20 minutes
before operation to ensure best accuracy.
7. Please note when the symbol keeps flashing on
the screen, the batteries need to be replaced. Please ensure batteries are facing to the front before placing them back.
Thank you for purchasing EXCELL waterproof weighing scale. In order to use the scale properly, please read this instruction carefully before use. If you have a problem concerning the scale, please contact your supplier.
02026010 ZSME300000369
3
EXCELL PRECISION CO., LTD.
Chapter 1 Introduction
1-1 Features and Specifications
Features:
Sealed waterproof silica gel strip blocks water from infiltrating into the scale. Surrounded by waterproof grade sheeting to ensure the water free. 1/3,000~1/6,000 display resolution available. ESW Max adopts stainless steel housing while ELW Max and plastic ABS housing. High speed of 24bits AD fast reacts and shortens the weighing operation duration. Selectable units: Kilogram (kg), gram (g), ounce (oz), and pound (lb) weighing units
available.
Well-designed protection point for transportation. Low power indication and auto power off.
Specifications:
Model
Capacity
Division
Resolution
ELW Max-3 ESW Max-3
3kg
0.5g 1g
1/6,000 1/3,000
ELW Max-6
ESW Max-6
6kg
1g 2g
1/6,000 1/3,000
ELW Max-15 ESW Max-15
15kg
2g 5g
1/7,500 1/3,000
ELW Max-30 ESW Max-30
30Kg
5g
10g
1/6,000 1/3,000
Operating Temperature -10°C ~ 40°C (14°F ~ 104°F)
Dimensions : ELW Max: 270 x 124 x 310mm (W x H x D)
ESW Max: 240 x 120 x 280mm (W x H x D)
Weight of the scale: ELW Max: 3.3 kg approximately;
ESW Max: 3.5 kg approximately;
1-2 Power Supply
Power Supply and Power Consumption
4 D size dry batteries (Alkaline)
Load Cell
Power Consumption
(single display without backlight)
Battery Life
(single display without backlight)
35
10.5 mA (max.)
About 900 hours
100
7 mA (max.)
About 2000 hours
02026010 ZSME300000369
4
EXCELL PRECISION CO., LTD.
.
. .
. . .
. . . .
Pcs
Low Battery Warning
When shows up, it will have approximately 10 hours of operation left. Then symbol will flash on the display; please replace the dry batteries soon when one of those two symbols appears.
When the battery voltage is less than 3.8 V, the scale would power off automatically.
1-3 Panel and Keyboard Introduction
Icon Introduction
STABLE
: The weight is stable.
GROSS
: The scale is in the gross mode. The display shows the goods and any
container weight. This Gross status indication is on when the TARE function is used.
: The unit weight is not sufficient. When the icon is on, the counting
function is operational but the count may contain errors.
: The sampling size is not sufficient When the icon is on, the counting
function is operational but the count may contain errors.
RANGE 1
: Only available for multi-range models
RANGE 2
: Only available for multi-range models
Keyboard Function
ON/ZERO KEY
This key possesses two functions: Power On and Zero function.
OFF KEY
When the scale is switched on, press the OFF key, the scale will switch off.
TARE KEY
The tare function will not operate during the following conditions:
1. When the scale powers on if the weight is negative and after a container is placed on the weigh pan if the weight is still below zero.
2. The tare value is over the full scale capacity.
RANGE 1 RANGE 2
02026010 ZSME300000369
5
EXCELL PRECISION CO., LTD.
UNITS KEY
Press the UNITS key to switch weight units; the icons will indicate the active units.
NET/GROSS KEY
In the Tare mode, the screen displays the “TARE” icon; press the NET/GROSS key to switch between the “Net value” and the “Gross value”.
1-4 Operating the Scale
POWER ON
When the scale is off, press the ON/ZERO key, the scale will switch on.
POWER OFF
When the scale is on, press the OFF key, the scale will switch off.
ZERO
When the weigh pan is empty (free of load) and the display is not showing zero, press the ON/ZERO key to zero the scale. At zero, the “0” indication is on.
When the weight value is within the zero range, the zero function operates to zero the
scale or cancel the tare function.
Zero range: The OIML & NTEP models have a zero range of 2% of Full Scale. The
Sri Lanka model has a zero range of 4% of Full Scale.
SWITCHING UNITS
Press the UNITS key to switch weight units, the icons or arrows will indicate the active units as appropriate. The units available are dependent on the exact scale model.
After power off, the scale will memorize the active units. When the scale is powered on
again, it displays the previously active units.
TARE FUNCTION
(1) Put a container on the weigh pan and after the weight is stable, press the TARE
key to zero the weight of the container. The screen displays the “Net” icon.
(2) Put the goods in the container, the screen displays the net weight value of the
goods.
(3) Remove the full container; the screen displays the negative weight value of the
container. At this time pressing the TARE key again will cancel the tare and the scale reverts back to zero. The “Net” icon is switched off.
The tare function can be operated continually to the full weighing capacity of the scale. Continual tare operation is adding or removing tare objects on weigh pan and pressing
the TARE key each time.
NET/GROSS FUNCTION
In the Tare mode, the screen displays the “Net” icon, press the NET/GROSS key to switch between the “Net value” and the “Gross value”.
When the GROSS icon is on, the weight value on the display is the total amount of
the tare value and net value.
At the Gross status, only OFF and NET/GROSS keys are functional. NET/GROSS key is only used in Tare mode.
02026010 ZSME300000369
6
EXCELL PRECISION CO., LTD.
Pcs
.
. .
. . .
. . . .
Pcs
.
. .
. . .
. . . .
SIMPLE COUNTING FUNCTION
(1) Use the UNITS key to enter into the “PCS” mode (2) Press the NET/GROSS key to select the counting sample size
(S = 10, S = 20, S = 50, S = 100, S=200). The LCD shows , , , , in order.
(3) Put the samples on the weigh pan and press the UNITS key, the screen displays
------”. After the sampling process is complete, put the goods on the weigh pan and the screen shows the quantity of the items.
The sample weight should be heavier than the minimum capacity of the scale (20d), If
not the arrow pointing to the icon will be activated.
The weight of a sample should be heavier than the 0.2d (d=division), or the arrow
pointing to the icon will be on.
When the or are indicated, the scale is still operational but the count may
contain errors.
To power off in this mode, the scale will memorize the “Pcs” unit. When the scale is
powered on again, it directly enters the simple counting mode.
While the Auto unit weight average function is available in the Advanced Function,
the goods on the weigh pan are 5pcs greater than the sample size and less than double the sample size, the scale will automatically re-sample the unit weight.
1-5 Self-Test Mode
Set the switch SWA1 on the bottom of machine to the LOCK position.When power is off,
hold
NET/GROSS
, and press
ON/ZERO
key, Wait till display shows to
enter “Self-Test Mode”.
Press
UNITS
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
UNITS
key
Press
UNITS
key
Press
UNITS
key
02026010 ZSME300000369
7
EXCELL PRECISION CO., LTD.
1-6 Error Messages
Initial zero is higher than the zero range when switching the indicator on.
Initial zero is lower than the zero range when switching the indicator on.
Internal value is below Zero.
The weight of the object is over 9 divisions of the maximum capacity.
----- For weight < -20d without tare or pretare device in operation.
1-7 Weight Unit
kg
1 g = 0.001 kg
g
1 g = 1 g
lb
1 g = 0.002204623 lb
oz
1 g = 0.03527396 oz
INTERNAL VALUE MODE (must hook up Load Cell to test)
Press
TARE
to enter, and the display shows internal value
Please check whether the internal value has changed obviously with weight
changing.
Please check the backlight. Press
ON/ZERO
key to back to the last screen , the display shows
KEYPAD TEST MODE
Press
TARE
to enter, display shows
Keypads internal code:
TARE
key= 04,
NET/GROSS
key=03,
UNITS
key= 02
Press
ON/ZERO
key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
Press
TARE
to enter , display shows the firmware version
Press
TARE
key again, the display shows maintenance number for 2 secs
Press
ON/ZERO
key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press
TARE
key to exit self-test mode, the scale will restart automatically.
02026010 ZSME300000369
8
EXCELL PRECISION CO., LTD.
Chapter 2 Advanced Functions
2-1 Advanced Function Setting Table
Below is an overview of the advanced functions. For detailed settings refer to the following sections:
DISPLAY
LEVEL 1 FUNCTION
DISPLAY
LEVEL 2 FUNCTIONS
Exit the ADVANCED
FUNCTION setting mode
---
---
General Function setting
mode
Return to ADVANCED
FUNCTION setting menu
Automatic backlight
function setting
Automatic power-off
timer setting
Hi/Lo/OK function setting
Restore the default
settings
Noise filter setting
Hold function setting
Auto unit weight
averaging setting
Two Weighing Units
Setting
Unstable Tare
External Weight
Calibration
---
---
02026010 ZSME300000369
9
EXCELL PRECISION CO., LTD.
2-2 Advanced Function Setting Workflow
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to enter the Advanced Function setting mode. The LCD shows .
Overall workflow of the Advanced Function setting mode:
General Function setting mode External Weight Calibration Exit the Advanced Function setting mode
Refer to the following sections for the detailed operation procedures of each function setting.
Press UNITS key
Press ON/ZERO
key
Press UNITS key
Press UNITS key
Press ON/ZERO key
Press ON/ZERO
key
02026010 ZSME300000369
10
EXCELL PRECISION CO., LTD.
2-3 General Function Setting
Workflow of the General Function setting:
ACTIONS
DISPLAY
NOTE
Return to the Advanced Function Setting Mode Menu Automatic Backlight Function Settings Automatic Power-off Timer Settings Hi/Lo/OK Settings Restore the Default Settings Noise Filter Settings Hold Function Settings Auto Unit Weight Averaging Setting
Two Weighing Units Setting Unstable Tare
Refer to the following sections for detailed operation procedures of each setting.
Press
TARE
key
Use the keys to select the function
Key in 00 and press TARE key to return to the Advanced function setting menu
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left
02026010 ZSME300000369
11
EXCELL PRECISION CO., LTD.
2-3-1 Automatic Backlight Function Setting
Select in the General Function setting mode to change the backlight function setting.
Automatic backlight function
When the scale is idle near zero (< 10d) without any key being pressed for 10 seconds, it will automatically disable the backlight. To activate the backlight, place weight over 10d or press any key.
DISPLAY
ACTIONS
NOTE
Key in the parameter “00”
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to select function “on” or “off
Press
TARE
key
Press
TARE
key to complete the general setting procedures
LCD displays the last saved status
02026010 ZSME300000369
12
EXCELL PRECISION CO., LTD.
2-3-2 Automatic Power-off Timer Setting
Select in the General Function setting mode to change the automatic power-off timer setting.
Automatic power-off function
When the scale is idle near zero (< 10d) without any key being pressed, the scale will automatically switch off once it reaches the set time (1~9 minutes).
ACTIONS
DISPLAY
NOTE
LCD displays the last saved status
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Automatic power-off timer setting Use
ON/ZERO
or
UNITS
key to key in
parameter 0  No auto power-off
1  When the scale is idle for 1
minute, the scale automatically switches off
2 When the scale is idle for 2
minutes, the scale automatically switches off
9  When the scale is idle for 9
minutes, the scale automatically switches off
Press TARE key
Use
ON/ZERO
or UNITS
key
to key in parameter
Key in the parameter 00
Press
TARE
key to complete
the general setting procedures Press TARE key
02026010 ZSME300000369
13
EXCELL PRECISION CO., LTD.
2-3-3 Hi/Lo/OK Function Setting
Select in the General Function setting mode to set the Hi/Lo/OK function. This function is available in all unit modes. In one specific unit mode,
enter to set the Hi/Lo/OK values. When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled.
DISPLAY
ACTIONS
NOTE
LCD displays the last saved status
LCD displays the last saved status
.
lb
LCD displays the last saved status
.
lb
.
lb
.
lb
Alarm setting
(a) (b) (c) The definition of a, b, c positions: (a) 1 = beep sound on
0 = beep sound off (b) 1 = beep sound on when stable
0 = beep sound on when unstable (c) 1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is
not between high and low limits and higher than 10d
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Key in the desired high limit value
Key in the parameter setting for alarm
Press
TARE
key
Press
TARE
key
Key in the desired low limit value
Press
TARE
key
Press
TARE
key
Key in the parameter 00
Press
TARE
key to complete
the general setting procedures
02026010 ZSME300000369
14
EXCELL PRECISION CO., LTD.
2-3-4 Restore to the Default Setting
Select in the General Function setting mode to restore to the default setting.
The default setting includes the following:
1) External weight calibration
2) HI/LO/OK setting value
3) Noise filter setting (External)
4) Sampling setting for the counting function
If is set to , and the scale has not been restarted
automatically. Please ensure to restart the scale manually.
ACTIONS
DISPLAY
NOTE
 Return (Cancel the
restoration)
Restore to the default
setting
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Press
TARE
key
Use
ON/ZERO
or
UNITS
keys
to select “return” or “format”
Press
TARE
key to complete the
general setting procedures
Key in the parameter 00
Press
TARE
key
02026010 ZSME300000369
15
EXCELL PRECISION CO., LTD.
2-3-5 Noise Filter Setting
Select in the General Function setting mode to set the noise filter setting.
When modifying , the parameters of remain un-altered.
DISPLAY
ACTIONS
NOTE
Returning to the zero point setting
Use ON/ZERO or UNITS key to key in the parameters or zero point Default setting = 0 0  No skip 5 skip 5d 1  skip 1d 6  skip 6d 2  skip 2d 7 skip 7d 3  skip 3d 8  skip 8d 4  skip 4d 9 skip 9d When the weight on the scale is over
1/3 full capacity, the function is on.
Digital switch & Stabilization range setting LCD displays the last saved parameter setting
Filter parameter setting
LCD displays the last saved parameter setting
Filter parameter setting Use
ON/ZERO
or
UNITS
keys to key
in the parameters. Default setting = 5 Parameter 0 ~ 9, the larger the number, the faster the filter response. Fast response can lead to weight instability.
Digital switch & Stabilization range setting
Use ON/ZERO or UNITS keys to key in the parameters. Default setting = 0 Parameter 0 ~ 9, the larger the number the more stable the weight.
Returning to zero point setting
LCD displays the last saved status
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key to complete
general setting procedures
Press
TARE
key
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Key in the parameter 00
02026010 ZSME300000369
16
EXCELL PRECISION CO., LTD.
2-3-6 Hold Function Setting
Select in the General Function setting mode to set the hold function.
Setting is unavailable for Brazil models.
= Hold function disabled = “Peak hold” mode
Keep displaying the maximum weight when the weight is continually changing To exit this mode, press any key.
= “Stable hold 1” mode
When the weight is stable, the LCD shows the current weight value. To exit this mode, press any key.
= “Stable hold 2” mode
When the weight is stable, the LCD shows the current weight value. When the weight returns to zero (<10d), the hold mode is cancelled automatically.
ACTIONS
DISPLAY
NOTE
Hold function setting
Use
ON/ZERO
or
UNITS
keys to key in the parameters Default setting = 0 0  Hold function disabled 1 “Peak hold” mode 2 “Stable hold 1 mode 3 “Stable hold 2 mode
Hold function setting
LCD displays the last saved parameter setting
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Use
ON/ZERO
or
UNITS
key
to key in the parameters
Press
TARE
key to complete
the general setting procedures
Press
TARE
key
Key in the parameter 00
Press
TARE
key
02026010 ZSME300000369
17
EXCELL PRECISION CO., LTD.
2-3-7 Auto Unit Weight Averaging Setting
Select in the General Function setting mode to set Auto Unit Weight Averaging.
Setting is unavailable for Brazil models. When is set as in , setting is unavailable.
(See 3-4-5)
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Auto unit weight averaging setting Use
ON/ZERO
or
UNITS
keys to key in the parameters Default setting = 0 0  Auto unit weight averaging 1  Non auto unit weight averaging
Auto unit weight averaging setting
LCD displays the last saved parameter setting
DISPLAY
ACTIONS
NOTE
Use
ON/ZERO
or
UNITS
key
to key in the parameters
Press
TARE
key to complete
the general setting procedures
Press
TARE
key
Key in the parameter 00
Press
TARE
key
02026010 ZSME300000369
18
EXCELL PRECISION CO., LTD.
2-3-8 Two Weighing Units Setting
Select in the General Function setting mode to set the two weighing units Setting.
Two weighing units setting
Use
ON|ZERO
or
UNITS
key to key in the parameters for two weighing units Default setting = 0
0 two weighing units function is not
activated To activate two weighing units, please set the 2nd weighing unit to be: 1 2nd weighing unit in CSP 01 2  3rd weighing unit in CSP 01 3  4th weighing unit in CSP 01
If it set to 1~3, it only displays the 1st weighing unit and the selected 2nd weighing unit
Two weighing units setting
LCD displays the last saved parameter setting
ON|ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET|GROSS
key Move the cursor to left.
Press
TARE
key
Use
ON|ZERO
or
UNITS
key to key in the parameters Press
TARE
key
02026010 ZSME300000369
19
EXCELL PRECISION CO., LTD.
2-3-9 Unstable Tare
Unstable Tare setting
Use
ON|ZERO
or
UNITS
key to key in the parameters for second weighing unit Default setting = 0 0  Disable tare operation while unstable 1  Enable tare operation while unstable
Unstable Tare setting
LCD displays the last saved parameter setting
ON|ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET|GROSS
key Move the cursor to left.
Press
TARE
key
Press
TARE
key
Use
ON|ZERO
or
UNITS
key to
key in the parameters
02026010 ZSME300000369
20
EXCELL PRECISION CO., LTD.
kg
kg
kg
2-4 Weight Calibration
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to enter the Advanced Function setting mode. The LCD shows and use the NET/GROSS or UNITS key to select to enter the weight calibration mode.
Weight calibration conditions:
The calibration weight value placed on the weight pan must be over 100e, and must be within the 90% and 110% of the full weight
The LCD displays the calibration weight value (the maximum capacity of 1st unit) and the unit. The digit on the right side keeps flashing.
The scale is calculating the internal value of the zero point. After stabilization, the LCD shows the calibration weight value.
Press
TARE
key
When the digit on the right flashes, ensure that the weigh pan is empty. Then Press the
TARE
key. After stabilization
Place the displayed weight on the weigh pan. Press the
TARE
key. The scale calculates the full capacity internal value and after stabilization, the buzzer beeps 3 times. Remove the weight from the weigh pan and the procedure is completed.
.
.
ACTIONS
DISPLAY
NOTE
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
02026010 ZSME300000369
21
EXCELL PRECISION CO., LTD.
Appendix 1 Command Mode &Output data format
only work with models have WIFI card or BLE card installed inside
Command Mode
Command Format A
Host
Command
Slave
Command
MZ
Zero
UA
Switch to the first weighing unit
MT
Tare
UB
Switch to the second weighing unit
MG
Gross weight
MN
Net weight
CT
Clear TARE value
Note: UB depends on the setting in FnC08
Command Format B
Host
Command
Slave
Data
RG
Read Gross weight
RN
Read Net weight
RT
Read TARE
Note: add % before the command to read continuously
Read HIGH/LOW values in FnC 03 RS
: Weighing unit (00 ~ 09) : Setting Items
HI
HIGH value
LO
LOW value
Note : ( weighing unit) is various depended on models
00  The first weighing unit
EX:
RS02LOCR><LF Read LOW values
ANS:
RS02LOXXXXXXCR><LF
Command Format C
Host
Command+ Data
Slave
Command+ Data
Write HIGH/LOW values in FnC 03 WSXXXXXX
: Weighing unit (00 ~ 09) : Setting Items XXXXXX: Setting Value
HI
HIGH value
LO
LOW value
02026010 ZSME300000369
22
EXCELL PRECISION CO., LTD.
˙
Note :  (Weighing unit) is various depended on models
00  The first weighing unit
EX:
WS00HI001000CR><LF Write HIGH values
ANS:
WS00HI001000CR><LF
Command Format D
Host
Data
Slave
Value (e.g. Price)
Position of decimal point
CR
LF
1 2 3 4 5
6
1
When the Slave receives this data format, it will transfer the data and display it on its LCD.
Only effective when the weight value is over 10d.
The above 4 (ABCD) command formats are RS232 bi-directional. The following error
messages might be received by Slave terminal (scale).
Error messages: E1: Wrong command E2: Command format error (Wrong parameters) E3: Do not match with the executing conditions for Command
Output Data Format
6 places ( first decimal place not included)
Weight format
Gross
S T , G S , + 1 2 3 4 5 6 7 SP
SP o z
CR
LF
Net S T , N T , + . 2 3 . 4 5 6 t l . g
Tare
S T , T R , + 1 2 . 3 4 5 6 SP
SP k g
Plus OL
O L , G S , +
SP
SP
SP
SP
SP
SP
SP
SP
SP
SP
SP
Minus OL
O L , G S , -
SP
SP
SP
SP
SP
SP
SP
SP
SP
SP
SP
Unstable
U S , G S , + 1 2 3 4 . 5 6 SP
SP l b
Serial Data Transfer/Receive Format
LSB MSB
1 0
n , 8 , 1 S 8-bit data
STOP
Note: S : Start bit STOP : Stop bit P : Parity bit
02026010 ZSME300000369
23
EXCELL PRECISION CO., LTD.
Appendix 2 7-Segment Display Characters
Number
Display
Letter
Display
Letter
Display
0 A N
1 B O
2 C P
3 D Q 4 E R 5 F S 6 G T 7 H U 8 I V 9 J W
K X
L Y
M Z