Installation & User Manual

Ed 19UK

June 2013

MAS2600

Tank Contents Transmitter

Installation & User Manual

Ed 19UK June 2013

Table of contents

 

Description and Operation .............................................................................................

3

MAS2600 Tank Contents Transmitter ..............................................................................................

3

Application.........................................................................................................................................

3

Intrinsically Safe in Hazardous Areas ...............................................................................................

3

Type Approvals .................................................................................................................................

3

MAS2600 Ordering Information

 

Technical Specificatio s ................................................................................................

4

.................................................................................

5

Type ...................................................................................................................................................

6

Measuring Ranges .............................................................................................................................

6

Selecting the Measuring Range.....................................................................................................

7

Selecting the Transducer Type ......................................................................................................

7

Absolute Transducers ....................................................................................................................

8

Basic Rules ....................................................................................................................................

8

Forepeak Tank ...............................................................................................................................

8

Draft ...............................................................................................................................................

9

Temperature Sensor...........................................................................................................................

9

Cable ..................................................................................................................................................

9

Brackets for Internal Mounting - drawing no. G022P010 .........................................................

10

Pole Mounting - drawing no. G022P011 ...................................................................................

10

1” Pipe End Mounting - drawing no. G022P015.......................................................................

11

Flange Mounting - drawing no. G022P013................................................................................

11

Flexible rubber tube mounting 80 ..............................................................................................

12

Flexible PTFE tube mounting ....................................................................................................

12

DN25 flange with 1” ball valve – drawing no. G022P019........................................................

13

Amplifier Box.................................................................................................................................

14

Deck Box 165B9035....................................................................................................

15

Handling .......................................................................................................................

16

Installation....................................................................................................................

17

Intrinsically Safe Installation .......................................................................................

19

Testing and Recalibrating ............................................................................................

22

Testing.............................................................................................................................................

22

Recalibrating Gauge Transmitter using Test Cup type G022S100................................................

22

Recalibrating Gauge Transmitter using Vacuum (for gauge transmitters only)............................

23

Recalibrating the Absolute Transmitter using Test Cup type G022S100......................................

24

Maintenance and Trouble Shooting.............................................................................

26

Maintenance....................................................................................................................................

26

Trouble Shooting ............................................................................................................................

26

MAS2600 Test sheet ......................................................................................................................

27

Page 2 of 28

 

Installation & User Manual

June 2013

Ed 19UK

 

 

Description and Operation

MAS2600 Tank Contents Transmitter

The MAS2600 is a 2-wire4-20mA level transmitter consisting of a transducer and an amplifier interconnected by a6-corevented cable.

Transducer:

The transducer is a pressure sensitive silicon micro strain gauge sensor mounted in a glass to metal seal. The sensor is protected by an isolation diaphragm, electron beam welded to the transducer housing, with an oil filling between the sensor and the diaphragm. Pressure changes on the front of the diaphragm will bring a resistance change in the Wheatstone bridge of the transducer. This change in the Wheatstone bridge will be transmitted to the amplifier as a change in the electrical signal.

The transducer is fully welded, housed in titanium with a titanium diaphragm.

All transducer types are submersible (IP68).

The transducer is available in three versions:

Gauge version.

Absolute version.

High temperature gauge version.

As optional the transducer is available with a built-inPt100 temperature sensor.

Amplifier:

The amplifier is available in the following options:

Programmable version: Calibration is made by dip switches and potentiometers.

Differential version: For measuring of both pressure and vacuum measuring.

The amplifier is housed in a sea water resistant polyester casing (IP56).

Application

The MAS2600 has been developed for level measuring in ballast, oil, service and fresh water tanks as well as tanks containing media which are not hostile to titanium.

Intrinsically Safe in Hazardous Areas

The MAS2600 is DEMKO approved for use with standard transmitter zener barriers or Ex isolation amplifiers and is Ex ia IIC T4 compliant in accordance with IEC 60079- 0:2007 and IEC 60079-11:2006.

Type Approvals

DNV, GL, LRS, MRS, BV, RINA, KRS, ABS.

CE marked in accordance with EU directive.

DMS

MAS 2600

PDCR-941-11.5 bar g

S/N 46374

Transducer

Amplifier box

Page 3 of28

Installation & User Manual

Ed 19UK

Technical Specifications

Transducer Ranges:

Gauge:0-3.5 /0-7 /0-16 /0-35m H2O. Absolute:0-20.394 /0-35.690m H2O.

Programmable Measuring Ranges:

Each transducer range is programmable in 8 steps.

Accuracy:

± 0.25% F.S. at 20°C.

Stability:

Max. ± 0.1% / Year.

TEB (Total Error Band):

± 2.0% F.S. at -20to +80°C.

EMC immunity influence < ±0.5%FS

Overload Capability:

Min. 4 x transducer range with no changes in calibration.

6 x transducer range burst pressure.

Built-inTemperature Sensor Pt100:

Optional.

Power Supply:

The Power Supply can vary from 17-33VDC. Permissible load resistance is shown graphically below. Calculation formula:

ULoop-17V

Rloop max. (Kohm) =25mA

Max. load resistance at various power supply

(ohm)

700

 

 

 

 

 

600

 

 

 

 

 

500

 

 

 

 

 

resistance

400

 

 

 

 

 

300

 

 

 

 

 

200

 

 

 

 

 

100

 

 

 

 

 

0

 

 

 

 

 

17

20

23

26

29

32

Load

 

Power supply (V)

 

 

 

 

Page 4 of 28

June 2013

Output Current:

4-20mA DC, loop powered2-wiresystem.

Current Limiting:

Max:

25 mA

Sensor Materials:

Housing: Titanium Grade 2

Diaphragm: Titanium Grade 4

Operating Temperature Ranges:

 

Transducer:

-20to +125°C

Standard cable:

-20to +

80°C

High temperature cable:

-20to +125°C

Amplifier:

-40to +

85°C

Protection Class:

 

Transducer:

IP 68

Amplifier:

IP 56

Intrinsic Safety:

Ex ia IIC T4 compliant.

Max. 75 m cable between transducer and transmitter amplifier.

Transmitter dimensions

Amplifier box:

 

110.0

55.0

 

98.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2600 SERIES

 

 

 

 

4-20mA TWO WIRE TRANSMITTER

75.0

4 pcs. ø4.5

45.0

x EEX ia

C T5

S/N: 2600

 

CERTIFICATE REF.: 90-C.95948Tamb.:-40to +85 °C

 

Lint = 20 uH

Cint = 10 nF

Umax = 33 V

IK=< 45 mA

 

 

 

 

 

MARINE AUTOMATION DENMARK

 

 

 

 

 

 

 

 

 

PG11

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1/8" NPT

cable gland

 

Use 4x20mm bolts and nuts

 

 

 

 

 

for suspension on a 4mm

 

 

vent fitting

 

thick surface

Transducer:

3.0

6-corevented cable, ø8.0

 

 

 

 

 

 

 

30.0

 

 

 

25.0

 

 

 

 

 

 

 

 

 

 

 

 

 

120.0

Front view

 

 

Installation & User Manual

June 2013

Ed 19UK

 

 

MAS2600 Ordering Information

Type

A absolute transducer

Ggauge transducer

Hhigh temperature gauge transducer

Transducer Ranges

13,5 mH2O gauge or high temp.

27,0 mH2O gauge or high temp.

316,0 mH2O gauge or high temp.

435,0 mH2O gauge or 0,8 – 3,5 bar absolute

70,8 – 2,0 bar absolute

Temperature Sensor

0without

1Built-inPt100 (Not available for IS installations )

Cable Max. 300 meters – max 75 meters for IS installations

XX XLPE standard cable length in meters (up to 80°C).

XX FEP high temperature cable length in meters (up to 125°C) (Not IS inst).

Mounting

0without fittings

1brackets for internal mounting

2pole mounting

31" pipe end mounting

4flange mounting DN 25

5flange mounting DN 40

6flange mounting DN 50

9 flexible rubber tube mounting – 80 ºC (add length in meters)

P flexible PTFE tube mounting – 125 ºC (add length in meters) (Not IS inst) V DN25 flange / 1” ball valve

A-Mreplacement

Amplifier Box

0not supplied

1standard box with PG 11 / PG 11

2standard box with PG 11 / PG 13,5

3standard box with PG 11 / PG 16

5standard box with PG 13,5 / PG 13,5

6standard box with PG 11 / PG 11 – filling hole

7standard box with PG 11 / PG 13,5 – filling hole

8standard box with PG 11 / PG 16 – filling hole

Amplifier PCB

0not supplied

without terminals for temperature sensors:

P programmable output range gauge

Ddifferential output range gauge

with terminals for temperature sensors:

S programmable output range gauge

M differential output range gauge

X X X - XX - X / X X

Page 5 of28

Installation & User Manual

Ed 19UK

 

June 2013

 

 

 

Type

MAS2600-

XXX-XX-X/XX

A:Absolute Transducer. The absolute transducer has no breather pipe, because it refers to absolute vacuum. An atmospheric pressure sensor is used to compensate for the atmospheric pressure. The absolute transducer is especially suitable for open deck applications, where the amplifier box may be exposed to flooding.

G:Gauge Transducer. The gauge transducer is fitted with a breather pipe at the back of the diaphragm to compensate for the atmospheric pressure. The breather pipe is connected to a vent plug in the amplifier box. The gauge transducer is suitable for most applications.

H:High Temperature Transducer. The high temperature transducer is available in gauge version only. The transducer is mounted with a special high temperature cable and designed for heated tanks e.g. HFO tanks, and other applications with high tank media temperatures. The transducer can be flange mounted or internally mounted.

Measuring Ranges MAS2600-XXX-XX-X/XX

Gauge Transmitter:

The MAS2600 Gauge Transmitter is supplied in 4 ranges. In the programmable amplifier the measuring range is divided into 8 sub-rangesand operated by means of the range select switch.

The transmitter output signal can be adjusted to 20 mA within the following sub-ranges.

 

 

 

Transducer type

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Type

1:

Type

2:

 

Type

3:

 

Type

4:

Range No.

0 – 3.5 mH2O

0 – 7.0 mH2O

0 – 16.0 mH2O

0 – 35.0 mH2O

1

0.20 – 0.37 mH2O

0.20 – 0.30 mH2O

0.50

– 0.75 mH2O

1.00

– 1.90 mH2O

 

 

 

 

 

 

 

2

0.37 – 0.54 mH2O

0.30 – 0.50 mH2O

0.75

– 1.00 mH2O

1.90

– 2.80 mH2O

 

 

 

 

 

 

 

3

0.54 – 0.73 mH2O

0.50 – 0.70 mH2O

1.00

– 1.50 mH2O

2.80

– 3.90 mH2O

 

 

 

 

 

 

 

4

0.73 – 1.10 mH2O

0.70 – 1.10 mH2O

1.50

– 2.50 mH2O

3.90

– 5.80 mH2O

 

 

 

 

 

 

 

5

1.10 – 1.92 mH2O

1.10 – 1.80 mH2O

2.50

– 4.00mH2O

5.80

– 9.90 mH2O

 

 

 

 

 

 

6

1.92 – 3.10 mH2O

1.80 – 3.00 mH2O

4.00

– 6.00 mH2O

9.90 – 15.30 mH2O

 

 

 

 

 

 

7

3.10 – 3.50 mH2O

3.00 – 4.80 mH2O

6.00 – 10.00 mH2O

15.30

– 25.10 mH2O

 

 

 

 

 

 

 

 

 

 

8

NOT

4.80 – 7.10 mH2O

10.00

– 16.00 mH2O

25.10

– 35.60 mH2O

APPLICABLE

 

 

 

 

 

 

 

 

 

Page 6 of 28

Installation & User Manual

June 2013

Ed 19UK

 

 

Absolute Transmitter:

The MAS2600 Absolute Transmitter is available in two ranges.

The output signals are calibrated as follows:

Transducer type

 

: 0.8 - 3.5 bar abs

4

Transducer type

7

: 0.8 - 2.0 bar abs

Selecting the Measuring Range:

When selecting the measuring range it is important to consider the tank height, vent pipe height, the tank type and the specific gravity (SG) of the tank contents.

The transducer measures the height of the fluid (water gauge) from the transducer diaphragm to the surface of the fluid, including the fluid in the vent pipe.

MAS 2600 Amplifier box

4-20mA TWO WIRE TRANSMITTER

EE

CT5

S/N:

Vent pipe height

 

Tankheight

 

Sensor pos. above

Flange mounted

lowest point

G022P013

 

To ensure measuring accuracy a transducer type within the range no. 4, 5, 6, 7 and 8 is recommended.

Guidelines for selection of transducer measuring ranges are set out in the following.

Selecting the Transducer Type:

Gauge and High Temperature Gauge Transducers:

The Gauge Transmitter is normally used on dry and pressure neutral locations. The Gauge Transmitter is available in 4 ranges:

0-3.5mH2O.

0-7.0mH2O.

0-16.0mH2O.

0-35.0mH2O.

Page 7 of28

Installation & User Manual

Ed 19UK June 2013

Absolute Transducers:

The Absolute Transmitter is typically used if the amplifier is exposed to water, e.g. on open deck. The Absolute Transmitter is available in 2 ranges: 0-20and0-35mH2O. In practice this means 10 mH2O less than the nominal range because of the atmospheric pressure which must be taken into consideration when choosing the transducer.

A transducer with an effective range larger than the tank height should be selected.

Basic Rules:

1. The selected measuring range must be larger than the tank height.

2. The selected measuring range must be larger than tank height + vent pipe height

2

EXAMPLE: Tank height

=

6 m

Vent pipe height =

14 m

1. The selected measuring range must be larger than 6 m due to the height of the tank.

2. The selected measuring range must be larger than

6 + 14

= 10 m due to the total height of the

2

 

 

tank and the vent pipe combined.

Select a transducer with a measuring range of 16 mH2O as this is the next measuring range larger than 10 m.

Turn the range select switch on the amplifier to range No. 7.

Normally the selection of transducer, range select and calibration are done by DAMCOS.

Special consideration applies to e.g. forepeak tanks and draft measuring when selecting the measuring range of the transducer.

Forepeak Tank:

Transducer measuring range > tank height: Select the next measuring range.

EXAMPLE: Tank height 12 m.

Based on the tank height a transducer with a measuring range of 16 mH2O should be selected.

However, due to the dynamic conditions in the forepeak tank, a transducer with a measuring range of 35 mH2O, i.e. the next in line, should be selected.

Turn amplifier range select switch to range No. 6.

Page 8 of 28

 

Installation & User Manual

June 2013

Ed 19UK

 

 

Draft:

Vent pipe min. 1”

MAS2600 amplifier box

Weather deck

Min. draft

Class approved

Sea valve

4-20mA TWO WIRE TRANSMITTER

EE

CT5

S/N:

Sensor pos. above

Ships hull

Ships hull

Transducer measuring range > max. draft minus sensor position.

This rule applies provided the transducer is mounted in accordance with the installation drawing for draft transmitters (vent pipe min. 1"). The sensor may be mounted direct on the sea valve with an air release valve if the ship is sailing in calm waters.

If the transducer is installed in tanks which may have excessive dynamic stresses such as e.g. high pump pressure, long bottomed tanks with a risk of slashing etc., contact Emerson Process Management Marine Solutions, Damcos.

Temperature Sensor MAS2600-XXX-XX-X/XX

The transducer can be supplied with a built-inPt100 temperature sensor. The temperature sensor will only give a proper reading, if the transducer is an internally mounted type.

0:Without temperature sensor.

1:With temperature sensor

Cable MAS2600-XXX-XX-X/XX

Two types of transducer cable is available. The standard cable for applications with temperatures never exceeding 80°C, and high temperature cable for applications with temperatures up to 125°C. The transducer cable is delivered in whole meters.

For further information require data sheet.

To make the MAS2600 transmitter easy to install, and for later service purposes, the transducer cable should be as short as possible, though considering the position of the transducer and amplifier box.

Max. cable length: 300 m.

When installing in hazardous areas max. cable length: 75 m.

Page 9 of28

Installation & User Manual

Ed 19UK

June 2013

Brackets for Internal Mounting - drawing no. G022P010:

4-20mA TWO WIRE TRANSMITTER

EE

CT5

S/N:

Tank top

 

 

 

Bulkhead

38.0

55.0

40.0

 

 

125.0

45.0

 

 

 

Tank bottom

 

160.0

 

 

 

 

The brackets are U-shapedand manufactured of 2 mm SS 316L plated steel. Diameter of mounting holes: 11.5 mm.

The transducer is secured by means of a spring fitted into two holes in the “U” shape, and can be mounted/dismounted without the use of tools.

When installing in hazardous areas the cable must be protected by a steel pipe welded to the tank.

Pole Mounting - drawing no. G022P011:

4-20mA TWO WIRE TRANSMITTER

EE

CT5

S/N:

Tank top

 

 

 

 

 

24.0

 

 

 

40.0

 

Bulkhead

 

 

 

34.0

26.0

26.0

41.0

Tank bottom

 

1" Thread

 

The weldable collar and the nut are manufactured of SS 316L steel. The gaskets are made of nitrile rubber.

Note: Max torque 20 Nm (slightly tighten by the hand + 1/8-1/4turn). Remember the gasket.