Equipment described in these instructions is covered by the following Patents:
FNF UK Patent No: GB2306009B, FNF US Patent No: 5886522
This product meets the Electromagnetic Compatibility Directive.
The product is Class B, Group 1 ISM equipment according to CISPR 11
Group 1 ISM product: A product in which there is intentionally generated and/or used conductively coupled
radio-frequency energy which is necessary for the internal functioning of the equipment itself.
Class B product are suitable for use in domestic establishments and in establishments directly connected
to a low voltage power supply network which supplies buildings used for domestic purposes.
system or otherwise) or translated into any language, in any form or by any means (electronic, mechanical,
magnetic, optical, manual or otherwise) without the prior written permission of Elcometer Limited.
A copy of this Instruction Manual is available for download on our Website via www.elcometer.com.
Thank you for your purchase of this Elcometer 415 Coating Thickness Gauge. Welcome to Elcometer.
R
415
Elcometer are world leaders in the design, manufacture and supply of coatings inspection equipment. Our
products cover all aspects of coating inspection, from development through application to post application
inspection.
The Elcometer 415 Coating Thickness Gauge is a world beating product. With the purchase of this gauge
you now have access to the worldwide service and support network of Elcometer. For more information visit
our website at www.elcometer.com
1 ABOUT YOUR GAUGE
The Elcometer 415 Coating Thickness Gauge is a handheld gauge for fast and
accurate measurement of the thickness of coatings on smooth, flat, unblasted
metal substrates.
Your gauge features an easy to use menu-driven graphical interface which guides
the user through tasks such as gauge configuration and calibration adjustment.
The probe in your gauge is suitable for measurement on ferrous steel and on nonferrous aluminium substrates.
These instructions include images of Elcometer 415 screens with units set to
microns (µm). Similar screens will be seen when the gauge is set to mils.
a. The Elcometer 415 is not suitable for use on blasted substrates. For this application, contact Elcometer for informa-
tion on the Elcometer 456 Coating Thickness Gauge.
2
a
R
1.1 STANDARDS
The Elcometer 415 can be used in accordance with the following National and International Standards:
AS 2331.1.4, AS/NZS 1580.108.1, ASTM B 499, ASTM D 7091 supersedes ASTM D 1186-B & ASTM D
1400, ASTM E376, BS 5599, EN 13523-1 supersedes ECCA T1, IS0 2360 supersedes BS 5411-3,
ISO 2808-7C supersedes ISO 2808-6A, BS 3900-C5-6A, BS 5411-11 & DIN 50981, ISO 2808-7D
supersedes BS 3900-C5-6B, BS 5411-3, DIN 50984 & ISO 2808-6B, ISO 2808-12, NF T30-124.
2 WHAT THE BOX CONTAINS
•Elcometer 415 Gauge
•4 x Check Foils: 50, 125, 500 and 1000 µm / 2, 5, 20, and 40 mils (Paint and Powder Gauge only)
•4 x Check Foils: 25, 75, 125 and 250 µm / 1, 3, 5, and 10 mils (Automotive Gauge only)
•Ferrous Checkpiece (Automotive Gauge only)
•Non-Ferrous Checkpiece (Automotive Gauge only)
•Carry Case
•Batteries
•Operating Instructions
The gauge is packed in cardboard and plastic packaging. Please ensure that this packaging is disposed of
in an environmentally sensitive manner. Consult your Local Environmental Authority for further guidance.
To maximise the benefits of your new Elcometer 415, please take some time to read these Operating
Instructions. Do not hesitate to contact Elcometer or your Elcometer supplier if you have any
questions.
3
R
3 GETTING STARTED
Before you use your gauge for the first time, remove the paper strip from the battery compartment.
3.1 FITTING BATTERIES
Batteries are located under the cover at the rear of the gauge. Use two LR03 (AAA),
alkaline dry batteries, or rechargeable equivalent; take care to ensure correct battery
polarity.
3.2 SWITCHING THE GAUGE ON AND OFF
To switch your gauge on, press , or place the probe on any surface.
The gauge switches off automatically after 60 seconds; to switch off manually, press and
hold .
3.3 THE KEYPAD
The gauge is controlled by four ‘softkeys’.
The function of these softkeys varies and is described by symbols and words on the
bottom line of the screen.
3.4 INTERFACE
An interface is located on the side of your gauge. This interface is used to program the
gauge at the factory and has no function for users of the gauge.
4
R
3.5 SELECTING A LANGUAGE
The Elcometer 415 has over 20 languages built-in.
When the gauge is switched on for the first time after batteries have been fitted, the display will show the
language selection screen.
Press or to locate your language, then press SEL. The gauge operates in the new language until
changed.
TO CHANGE LANGUAGE AT ANY TIME
Either:
1. Switch gauge off.
2. Press and hold left hand softkey.
3. Press to switch on gauge.
4. Release left hand softkey.
5. Press
or to locate your language, then press SEL. The gauge operates in the new language until
changed.
Or:
Select MENU/SETUP/LANGUAGES
5
R
3.6 THE SYMBOLS ON THE DISPLAY
Battery condition indicator.
When the battery condition indicator flashes and the gauge beeps every 10 seconds,
change the batteries.
When the batteries are exhausted, the gauge emits five loud beeps and then switches off.
Fe/NFSubstrate type.
Fe = Ferrous, NF = Non-Ferrous metal
µm/milMeasurement units.
MENUPress to access user-selectable features and gauge information
6
R
4 TAKING A READING
1. Hold the gauge with the probe pressed firmlyb against
the surface you are measuring.
If the gauge is off, it will switch on automatically.
2. Read the coating thickness shown on the display.
The display also shows the substrate type (Ferrous or
Non-ferrous) and the measurement units (µm or mil).
b. The reading may be inaccurate if the probe is not held flat against the surface.
7
R
5 THE MENU
SETUP
Units
Beep volume
Languages
ABOUT
Gauge information
Probe information
Contact
RESET
Factory cal
MENU
Zero
Setup
About
Reset
UNITS
µm
mil
BEEP VOLUME
5
4
3
2
1
0
LANGUAGES
English
Deutsch
Française
etc.
Press the MENU button to access all features of your gauge:
8
R
6 ADJUSTING GAUGE CALIBRATION
Before taking measurements, always adjust the calibration of your gauge by zeroing your gauge on
uncoated steel or aluminium sample having the same substrate thickness and a similar smooth surface
finish as the item to be measured.
1. Select MENU/ZERO
2. When indicated by the display, place the probe on the uncoated substrate.
The display will show 0 µm (or 0.0 mil) and is immediately ready to use.
You may wish to test the calibration using the foils supplied with your gauge.
If a sample of uncoated substrate is not available, use the factory calibration zero:
Select MENU/RESET/FACTORY CAL/YES.
7 PROBES
The probe is a consumable item and will eventually wear. Probe life will depend on the number of
measurements taken and how abrasive the coating is. Probe life can be prolonged by careful positioning of
the probe on the surface. If the probe becomes worn or damaged, return your gauge to Elcometer for probe
replacement and re-programming.
9
R
8 STORAGE
This gauge incorporates a Liquid Crystal Display (LCD). If the display is heated above 50°C (120°F)
it may be damaged. This can happen if the gauge is left in a car parked in strong sunlight.
Always store the gauge in its carrying pouch when it is not being used.
Remove the batteries from the gauge and store them separately if the gauge is to remain unused for a long
period of time. This will prevent damage to the gauge in the event of malfunction of the batteries.
9 MAINTENANCE
You own one of the finest hand-held coating thickness gauges in the world. If looked after, it will last a
lifetime.
The gauge does not contain any user-serviceable components. In the unlikely event of a fault, the gauge
should be returned to your local Elcometer supplier or directly to Elcometer. The warranty will be invalidated
if the gauge has been opened.
Contact details can be found:
•Stored in the gauge (Select MENU/ABOUT/CONTACT).
•On the outside cover of these operating instructions.
•At www.elcometer.com
10
R
10 TECHNICAL DATA
Range:0 µm to 1000 µm (0 mils to 40 mils)
Accuracy:±3% or ±3 µmc (±0.12 mil)
Resolution:1 µm (0.1 mils)
Measurement speed:>60 readings per minute
Weight:130g (4.1oz)
Dimensions:110 mm x 75 mm x 35 mm (4.3" x 3" x 1.38")
Operating temperature (ambient
Operating temperature, maximum (probe): 80°C (176°F )
Storage temperature:-10°C to 55°C (14°F to 130°F)
Case:High impact ABS
Batteries:2 x LR03 (AAA), alkalinee dry batteries
c. Whichever is the greater.
d. Operation outside these limits depends upon climatic conditions. If required, contact Elcometer for further information.
e. Alkaline batteries must be disposed of carefully to avoid environmental contamination. Please consult your local envi-
ronmental authority for information on disposal in your region.
Do not dispose of any batteries in fire.
f.Rechargeable batteries can be used if they are charged outside the gauge.
d
):0°C to 50°C (32°F to 120°F)
or rechargeable
f
equivalents.
11
R
11 ACCESSORIES
The following accessories and spares are available from Elcometer, or your local Elcometer supplier.
Elcometer produces a wide range of coating thickness gauges and associated paint inspection equipment.
Users of the Elcometer 415 may also benefit from the following Elcometer products:
•Uncured powder thickness gauges
•Inspection management software
•Mechanical coating thickness gauges
•Elcometer 456 Coating Thickness Gauge
•Appearance testers
•Adhesion testers
For further information contact Elcometer, your local Elcometer supplier or visit www.elcometer.com
13
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.