This document is the property of Druck Limited and may not, either in part or whole, be
copied or otherwise reproduced, communicated in any way to third parties nor stored in any
Data Processing system, without the express written authority of Druck Limited.
i
K239 Issue No. 1
INTRINSIC SAFETY CONDITIONS OF USE
This instrument is designed to be intrinsically safe when operated
in accordance with the BASEEFA Cer tification document and
schedule (see Appendix 2).
BASEEFA Certificate of Conformity No. Ex 99E2002X
BASEEF A being an Approv ed Certification Body , in accordance with
Article 14 of the Council Directive of the European Communities of
18th December , 1975 (76/117/EEC) certifies that the apparatus has
been found to comply with harmonised European Standards:
EN 50014: 1992
EEx ia IIC T4
EN 50020: 1994
and has successfully met the examination and test requirements
recorded in confidential report Number,
NOTE:Attention is drawn to Pages 4/5 of the Certificate
of conformity for electrical connection parameters
SPECIAL CONDITION OF USE
The DPI 610 IS Series Pressure calibrator is NOT capab le of
withstanding the 500V r.m.s. electric strength test between
the external connectors and the frame of the apparatus as
required by Clause 6.4.12 of EN50020 and this must be tak en
into account when using the apparatus for input measurements
in a system.
K239 Issue No. 1
ii
SAFETY
The Manufacturer has designed this product to be entirely safe when operated correctly.
lPlease pay close attention to the Safety Instructions outlined on this page and
elsewhere in this manual. They have been designed to protect the user from
personal injury and the equipment from damage.
Potentially hazardous operations are indicated in the text by means of a
hazard warning triangle. Specific warnings relating to each section of the
!
lPlease observe the installation advice and any operational limits given in this
lThis equipment must only be used for the purpose for which it was designed
Electrical Safety
The instrument is designed to be completely safe when used with options and
accessories supplied by the manufacturer for use with the instrument.
Test leads
Only use the test leads supplied with this instrument; the test leads must not be used
with any other test equipment.
T o xic Materials
No toxic materials are employed in this equipment
manual are given at the beginning of that section. On the instrument, this
symbol indicates that the user should refer to the User Manual.
manual.
Repair and Maintenance
The instrument must be maintained, either by the manufacturer or a competent person.
Please refer to supplier for details of approved ser vice agents. A list of Druck
Subsidiaries who will be able to assist and advise is given on Page 38.
Software Issue
This guide contains operating instructions for instruments with software Version 1.XX.
This product meets the essential protection requirements of the relevant
C E
WARNINGThis is a Class A product. In a domestic environment this product may
EEC directives. Fur ther details of applied standards may be found in the
product specification.
cause radio interference in which case the user may be required to take
adequate measures.
iii
K239 Issue No. 1
Specification
Pressure Ranges (Internal Transducer s)
Accuracy
Combined non-linearity, Hysteresis and repeatability
70 mbar to 20 bar (Calibrator):0.025% F.S.
35 bar to 700 bar (Indicator):0.025% F.S.
70 bar to 400 bar (Hydraulic)0.025% F.S.
T emperature Effects
±0.004% of reading/°C (averaged over -10° to -40°C w.r.t. 20°C)
Electrical Parameters
V oltage Inputs
Range:±50V
Accuracy±0.05% rdg, ±0.004% F.S.
Resolution100mV max
Current Inputs
Range:±55mA
Accuracy±0.05% rdg, ±0.004% F.S.
Resolution1mA max
Current sink
Range:24 mA
Accuracy±0.05% rdg, ±0.01% F.S.
Display
Size:60x60 mm LCD Graphics
Reading±99999, update rate 2 readings/sec
Environmental
Operating Temp.:-10°C to 50°C (Calibrated -10 °C to 40°C)
Storage Temp:-20°C to 60°C
Calibration Temp: 21°C ±2°C
Fitting Batteries 7
Switching On 7
Change Pressure Units 8
Voltage and Current Measurements 8
Typical Calibration Set-up (Pressure to Voltage) 9
Zero Display Reading 9
Making Measurements
Basic Mode (Task BASIC)10
Pressure Transmitter (P-I) Task10
Pressure Converter (P-P) Task11
Voltage Output Pressure Transmitter (P-V) Task11
Current to Pressure Converter (I-P) Task12
Pressure Switch Test (P-Switch) Task12
Pressure to Display (P-Display) Task13
Leak Tests (Leak Test) Task14
Transmitter Simulator (TX SIM) Task15
Relief Valve Test (REL VALVE) Task16
Advanced T ask17
General17
Select Input17
Ambient Temperature Measurement17
Process Functions18
Tare Process Function19
Min/Max Process Function20
Filter Process Function20
Flow Function21
% Span21
Select Output22
v
K239 Issue No. 1
Contents (Contd.)
Advanced T ask (contd.)
Electrical Outputs (Loop Power)22
mA Step23
mA Ramp24
mA Value26
Define New Task26
Clear Task26
Memory Operations27
Saving Display or Datalog27
Store Operations27
Recalling Stored Data27
Datalog Operations28
General30
Contrast30
Store Mode30
Calibration31
Settings - Select Set-up Option31
Units
Define Special Units
Language
RS232
Powerdown
31
32
32
32
32
Calibration33
General33
Calibration Check33
Calibration Adjustment34
General Procedures34
Using The Calibration Menu34
Temperature 35
Service
Approved Service Agents36
K239 Issue No. 1
vi
Contents (Contd.)
Appendix 1 - Hydraulic Actuator version
Introduction37
Safety Instructions38
Preparation for Use38
Bleeding the System39
Operation40
Draining the Hydraulic Fluid40
Flushing, replenishing or Changing the Hydraulic Fluid41
Appendix 2 - Baseefa Certificate of Conformity
vii
K239 Issue No. 1
Notes
(This page intentionally blank)
K239 Issue No. 1
viii
INTRODUCTIONSummary of Functions
General
The DPI 610 is a rugged, portable precision instrument available either as an indicator
or calibrator for use in hazardous areas. Additionally, an intrinsically safe hydraulic
calibrator version (HC) is available.
The instrument is primarily used for calibrating instrumentation and systems over the
range -1 to 20 bar, (HC version to 400 bar). Used in conjunction with e xternal transducers,
the range of the standard calibrator/indicator can be extended up to 700 bar.
The instrument is also capable of sinking an externally supplied current loop up to a
maximum current of 24 mA and measuring dc input voltages of ± 30V and dc currents
of ± 55 mA. Ambient temperature measurement is also provided.
The DPI 610 is powered by six alkaline C-cells and has an RS232 interface.
Important Notice
Zinc-carbon and zinc-chloride cells should NOT be used in this instrument.
Use only the battery types as shown in the Table on page 7.
Description of Procedures
In the procedures outlined in this User Guide, hard (fixed function) and soft (var iable
function) key operations are shown in bold type (e.g.) TASK and F1. These statements
mean press the TASK key and press the F1 key. Soft key operations can be allocated
to both the F1 and F2 keys. Where a specific soft function is referred to it is written in
bold italics (e.g.)
This instrument has a number of operating modes which are described in simplified
form in the following sections. Diagrams accompanying the procedures give typical
selection sequences and shaded controls indicate that this control key should be
pressed in the appropriate sequence. Diagrams should be read from left to right, top to
bottom where appropriate. A shaded display soft box indicates that the function key
immediately below that soft box should be pressed (either F1 for the left hand soft box
or F2 for the right). A typical diagram is shown below (e.g.).
PROCESS
.
TASK: BASIC
VOLTAGE
PRESSURE INT
CURRENT
PRESSURE
UNITS
V
bar
+
SELECT UNITS
OF PRESSURE
inHg
bar
hPa
Pa
SELECT UNITS
In the above diagram the following key sequence is indicated.
(a)Press the F2 key (the key immediately below the
UNITS
(b)Use the Up and Down cursor keys (only) to select the required option.
(If all keys shaded, use all these keys to select or enter data).
(c)Press the ENTER key.
K239 Issue No.11
soft box).
INTRODUCTIONSummary of Functions
Using This Guide
The following key symbols are used in the procedure diagrams which follow
Shaded cursor keys indicate that a combination of these four
keys, Up, Down, Left and Right should be used to (e.g.) enter an
alpha numeric value or to select a function.
Indicates the ENTER key. Used to confirm an operation or a
selection. Shading indicates key operation.
Exit key, used to clear current menu selection and return to next
menu level above current level. Used as an escape key from
current operation. Shading indicates key operation.
The following table shows the maximum measurement input ratings of the instrument
which should not be exceeded.
PRESSURE120% FULL SCALE
VOLTAGE30 V d.c.
CURRENT55 mA d.c.
Note 1: The display flashes if the input pressure, voltage or current overrange.
Note 2: Max applied voltage for exter nal loop supply = 30V dc (see Page 8).
K239 Issue No.1
2
INTRODUCTIONSummary of Functions
OPERATOR CONTROLS (Figure 1)
These divide into two groups, the operator/display controls (shown in Figure 1) and the
pressure/vacuum generation components (Shown in Figure 2). The operator controls
and a typical display, common to all instr ument versions, is shown below.
Display
max 30V
DPI 610 IS
TASK: BASIC
Electrical Measurement
Input Sockets
VOLTAGE
PRESSURE INT
CURRENT
V
bar
PRESSURE
UNITS
Cursor Keys
F2F1
Enter Key
Function (Soft) KeysOn/Off KeyHard Keys
Figure 1 - DPI 610 Keypad
DISPLAY
The display section of the instrument basically divides into four distinct sections. The
two main sections of the display are used to display a
a
Source (output parameter).
The remaining sections are used as a status display area
and to define soft key functions. A typical display is shown below (e.g.),
* These key funtions are not available in BASIC mode
K239 Issue No.1
4
INTRODUCTIONSummary of Functions
SOFT KEYS (Fig.1)
Three soft keys, designated F1, EXIT and F2 , are situated immediately below the
display as shown below. These keys have their function allocated by the instrument
software which is indicated in the bottom of the display (Voltage for F1 and Units for F2
in this example). They are used to select menu (program) options and are fully
described under the appropriate section headings.
CURSOR KEYS (Fig.1)
The cursor keys consist of a block of four key s, arranged as shown in Figure 1. They
are designated Up (Ù), Down (Ú), Left (×) and Right (Ø). In programs where options
need to be selected from a list, (e.g.) the TASK selection program, the Up and Down
cursor keys are used to highlight one of the options, from which it can be selected by
the ENTER key. In TASK mode, where more than one page of options are provided, theLeft (×) and Right (Ø) cursor keys will switch between pages.
External Interfaces under flap
Power, RS 232, External Transducer
Electrical Inputs,
V, I, Switch and mA sink
Test port. Connect to unit under test.
G1/8" coupling. Use a sealing washer.
Hard keys
Fine pressure adjustment.
Start with control screwed out
half way to allow fine adjustment
up or down.
Pump, pressurises
test port and internal
sensor. Use to approximately
set output pressure
Figure 2 - DPI 610 Calibrator Controls
Cursor Keys
Soft Keys F1, EXIT, F2
Release Valve.
Releases system
pressure through vent
port. Use to gradually,
or completely reduce
pressure
WARNING: BATTERIES MUST ONLY BE FITTED IN A SAFE AREA.
USE ONLY THE BATTERIES SPECIFIED IN THE TABLE.
Caution:
Old batteries can leak and cause corrosion. Never leave
discharged batteries in the instrument. Old batteries should be
treated as hazardous waste and disposed of accordingly.
Switching On
Press the I/O switch on the front panel and proceed as follows.
TASK: BASIC
I/O
VOLTAGE
PRESSURE INT
CURRENT
V
bar
PRESSURE
UNITS
TASK: BASIC
CURRENT
PRESSURE INT
VOLTAGE
PRESSURE
UNITS
mA
bar
The first time that the instrument is powered up, it will power up in BASIC mode with the
main screen displaying voltage in the measurand display area and pressure in the
source display area. To switch to
F1 to return to
Voltage
.
Current
as measurand, press F1 as shown. Similarly,
Note: No other keys are active in this mode and the instrument can only be
reconfigured by pressing the TASK key and selecting another mode.
K239 Issue No.17
Getting Started
Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the
units required, press TASK and select any task, other than BASIC, press SETUP and
proceed as detailed on page 31. To return to BASIC mode, press TASK and selectBASIC.
TASK: BASIC
VOLTAGE
PRESSURE INT
CURRENT
PRESSURE
UNITS
V
bar
+
SELECT UNITS
OF PRESSURE
inHg
bar
hPa
Pa
SELECT UNITS
In BASIC mode, the unit is configured to carry out basic Pressure to Voltage (P to V) or
Pressure to Current (P to I) tests a typical test procedures follow.
Voltage and Current Measurements
Connect the electrical input sockets as follows for voltage and current measurements.
Use the test leads provided and DO NOT push bare wires into the sockets.
Note: Max applied voltage = 30V dc, Max input current = 55mA dc
Vin
mA Sink
CURRENT
max 30V
mA in
55mA
-
+
max
VOLTAGE
max 30V
(FROM EXTERNAL SOURCE)
+
V
-
Vin
mA Sink
mA in
55mA
max
Note: Max applied voltage for exter nal loop supply = 30V dc
K239 Issue No.1
8
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.