Druck DPI 610 IS User Manual

DPI 610 IS
PORTABLE PRESSURE CALIBRATOR/INDICATOR
USER GUIDE
K239
Calibrator Version
Hydraulic Calibrator Version
Indicator Version
DPI 610 IS Portable Pressure Calibrator
Software version 1.XX
This document is the property of Druck Limited and may not, either in part or whole, be copied or otherwise reproduced, communicated in any way to third parties nor stored in any Data Processing system, without the express written authority of Druck Limited.
i
K239 Issue No. 1
INTRINSIC SAFETY CONDITIONS OF USE
This instrument is designed to be intrinsically safe when operated
in accordance with the BASEEFA Cer tification document and schedule (see Appendix 2).
BASEEFA Certificate of Conformity No. Ex 99E2002X
BASEEF A being an Approv ed Certification Body , in accordance with Article 14 of the Council Directive of the European Communities of 18th December , 1975 (76/117/EEC) certifies that the apparatus has been found to comply with harmonised European Standards:
EN 50014: 1992
EEx ia IIC T4
EN 50020: 1994
and has successfully met the examination and test requirements recorded in confidential report Number,
98(C)0818 (ERA Report Ref.3627/856), dated January, 1999
NOTE: Attention is drawn to Pages 4/5 of the Certificate
of conformity for electrical connection parameters
SPECIAL CONDITION OF USE
The DPI 610 IS Series Pressure calibrator is NOT capab le of withstanding the 500V r.m.s. electric strength test between the external connectors and the frame of the apparatus as required by Clause 6.4.12 of EN50020 and this must be tak en into account when using the apparatus for input measurements in a system.
K239 Issue No. 1
ii
SAFETY
The Manufacturer has designed this product to be entirely safe when operated correctly.
l Please pay close attention to the Safety Instructions outlined on this page and
elsewhere in this manual. They have been designed to protect the user from personal injury and the equipment from damage.
Potentially hazardous operations are indicated in the text by means of a hazard warning triangle. Specific warnings relating to each section of the
!
l Please observe the installation advice and any operational limits given in this
l This equipment must only be used for the purpose for which it was designed
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories supplied by the manufacturer for use with the instrument.
Test leads
Only use the test leads supplied with this instrument; the test leads must not be used with any other test equipment.
T o xic Materials
No toxic materials are employed in this equipment
manual are given at the beginning of that section. On the instrument, this symbol indicates that the user should refer to the User Manual.
manual.
Repair and Maintenance
The instrument must be maintained, either by the manufacturer or a competent person. Please refer to supplier for details of approved ser vice agents. A list of Druck Subsidiaries who will be able to assist and advise is given on Page 38.
Software Issue
This guide contains operating instructions for instruments with software Version 1.XX.
This product meets the essential protection requirements of the relevant
C E
WARNING This is a Class A product. In a domestic environment this product may
EEC directives. Fur ther details of applied standards may be found in the product specification.
cause radio interference in which case the user may be required to take adequate measures.
iii
K239 Issue No. 1
Specification
Pressure Ranges (Internal Transducer s)
Accuracy
Combined non-linearity, Hysteresis and repeatability
70 mbar to 20 bar (Calibrator): 0.025% F.S. 35 bar to 700 bar (Indicator): 0.025% F.S. 70 bar to 400 bar (Hydraulic) 0.025% F.S.
T emperature Effects
±0.004% of reading/°C (averaged over -10° to -40°C w.r.t. 20°C)
Electrical Parameters
V oltage Inputs
Range: ±50V Accuracy ±0.05% rdg, ±0.004% F.S. Resolution 100mV max
Current Inputs
Range: ±55mA Accuracy ±0.05% rdg, ±0.004% F.S. Resolution 1mA max
Current sink
Range: 24 mA Accuracy ±0.05% rdg, ±0.01% F.S.
Display
Size: 60x60 mm LCD Graphics Reading ±99999, update rate 2 readings/sec
Environmental
Operating Temp.: -10°C to 50°C (Calibrated -10 °C to 40°C) Storage Temp: -20°C to 60°C Calibration Temp: 21°C ±2°C
Sealing
Sealed to IP54
Physical
Size: 300 x170x140 mm Weight: 3 kg
K239 Issue No. 1
iv
Contents
Introduction
General 1
Description of Procedures 1 Using the Guide 2
Summary of Functions 3
OPERATOR CONTROLS 3 DISPLAY 3 HARD KEY FUNCTIONS 4 SOFT KEYS 5 CURSOR KEYS 5 ELECTRICAL CONNECTIONS 6
Getting Started
Fitting Batteries 7 Switching On 7 Change Pressure Units 8 Voltage and Current Measurements 8 Typical Calibration Set-up (Pressure to Voltage) 9 Zero Display Reading 9
Making Measurements
Basic Mode (Task BASIC) 10 Pressure Transmitter (P-I) Task 10 Pressure Converter (P-P) Task 11 Voltage Output Pressure Transmitter (P-V) Task 11 Current to Pressure Converter (I-P) Task 12 Pressure Switch Test (P-Switch) Task 12 Pressure to Display (P-Display) Task 13 Leak Tests (Leak Test) Task 14 Transmitter Simulator (TX SIM) Task 15 Relief Valve Test (REL VALVE) Task 16
Advanced T ask 17
General 17 Select Input 17 Ambient Temperature Measurement 17 Process Functions 18 Tare Process Function 19 Min/Max Process Function 20 Filter Process Function 20 Flow Function 21 % Span 21 Select Output 22
v
K239 Issue No. 1
Contents (Contd.)
Advanced T ask (contd.)
Electrical Outputs (Loop Power) 22
mA Step 23 mA Ramp 24 mA Value 26 Define New Task 26 Clear Task 26
Memory Operations 27
Saving Display or Datalog 27 Store Operations 27 Recalling Stored Data 27 Datalog Operations 28
Auto Log (Timer) 28 Manual Logging 28 Recall Datalog Files 29 Downloading Datalog Files 29 Delete Datalog Files 30
Using Set-up 30
General 30 Contrast 30 Store Mode 30 Calibration 31 Settings - Select Set-up Option 31
Units Define Special Units Language RS232 Powerdown
31 32 32 32 32
Calibration 33
General 33 Calibration Check 33 Calibration Adjustment 34 General Procedures 34 Using The Calibration Menu 34
Temperature 35
Service
Approved Service Agents 36
K239 Issue No. 1
vi
Contents (Contd.)
Appendix 1 - Hydraulic Actuator version
Introduction 37 Safety Instructions 38 Preparation for Use 38 Bleeding the System 39 Operation 40 Draining the Hydraulic Fluid 40 Flushing, replenishing or Changing the Hydraulic Fluid 41
Appendix 2 - Baseefa Certificate of Conformity
vii
K239 Issue No. 1
Notes
(This page intentionally blank)
K239 Issue No. 1
viii
INTRODUCTION Summary of Functions
General
The DPI 610 is a rugged, portable precision instrument available either as an indicator or calibrator for use in hazardous areas. Additionally, an intrinsically safe hydraulic calibrator version (HC) is available.
The instrument is primarily used for calibrating instrumentation and systems over the range -1 to 20 bar, (HC version to 400 bar). Used in conjunction with e xternal transducers, the range of the standard calibrator/indicator can be extended up to 700 bar.
The instrument is also capable of sinking an externally supplied current loop up to a maximum current of 24 mA and measuring dc input voltages of ± 30V and dc currents of ± 55 mA. Ambient temperature measurement is also provided.
The DPI 610 is powered by six alkaline C-cells and has an RS232 interface.
Important Notice
Zinc-carbon and zinc-chloride cells should NOT be used in this instrument. Use only the battery types as shown in the Table on page 7.
Description of Procedures
In the procedures outlined in this User Guide, hard (fixed function) and soft (var iable function) key operations are shown in bold type (e.g.) TASK and F1. These statements mean press the TASK key and press the F1 key. Soft key operations can be allocated to both the F1 and F2 keys. Where a specific soft function is referred to it is written in bold italics (e.g.)
This instrument has a number of operating modes which are described in simplified form in the following sections. Diagrams accompanying the procedures give typical selection sequences and shaded controls indicate that this control key should be pressed in the appropriate sequence. Diagrams should be read from left to right, top to bottom where appropriate. A shaded display soft box indicates that the function key immediately below that soft box should be pressed (either F1 for the left hand soft box or F2 for the right). A typical diagram is shown below (e.g.).
PROCESS
.
TASK: BASIC
VOLTAGE
PRESSURE INT
CURRENT
PRESSURE
UNITS
V
bar
­+
SELECT UNITS OF PRESSURE
inHg
bar
hPa Pa
SELECT UNITS
In the above diagram the following key sequence is indicated.
(a) Press the F2 key (the key immediately below the
UNITS
(b) Use the Up and Down cursor keys (only) to select the required option.
(If all keys shaded, use all these keys to select or enter data).
(c) Press the ENTER key.
K239 Issue No.11
soft box).
INTRODUCTION Summary of Functions
Using This Guide
The following key symbols are used in the procedure diagrams which follow
Shaded cursor keys indicate that a combination of these four keys, Up, Down, Left and Right should be used to (e.g.) enter an alpha numeric value or to select a function.
Indicates the ENTER key. Used to confirm an operation or a selection. Shading indicates key operation.
Exit key, used to clear current menu selection and return to next menu level above current level. Used as an escape key from current operation. Shading indicates key operation.
Hardkey (total 7). Legend beside key symbol indicates function. Shading indicates key operation.
Maximum Instrument Ratings
The following table shows the maximum measurement input ratings of the instrument which should not be exceeded.
PRESSURE 120% FULL SCALE VOLTAGE 30 V d.c. CURRENT 55 mA d.c.
Note 1: The display flashes if the input pressure, voltage or current overrange.
Note 2: Max applied voltage for exter nal loop supply = 30V dc (see Page 8).
K239 Issue No.1
2
INTRODUCTION Summary of Functions
OPERATOR CONTROLS (Figure 1)
These divide into two groups, the operator/display controls (shown in Figure 1) and the pressure/vacuum generation components (Shown in Figure 2). The operator controls and a typical display, common to all instr ument versions, is shown below.
Display
max 30V
DPI 610 IS
TASK: BASIC
Electrical Measurement Input Sockets
VOLTAGE
PRESSURE INT
CURRENT
V
bar
PRESSURE
UNITS
Cursor Keys
F2F1
Enter Key
Function (Soft) KeysOn/Off Key Hard Keys
Figure 1 - DPI 610 Keypad
DISPLAY
The display section of the instrument basically divides into four distinct sections. The two main sections of the display are used to display a a
Source (output parameter).
The remaining sections are used as a status display area
and to define soft key functions. A typical display is shown below (e.g.),
Status display
Measurand display
(INPUT)
Source display
(INPUT)
Soft boxes
TASK: BASIC
VOLTAGE
PRESSURE INT bar
CURRENT
Measurand (input parameter)
­+
V
PRESSURE
UNITS
and
K239 Issue No.13
INTRODUCTION Summary of Functions
HARD KEY FUNCTIONS (Fig. 1)
yeKnoitcnuFegaPfeR
7
13
9
71-01
62-32
23,82
23,82
2
2
,egassemrorrenanitlusertesfforegrala
ni(sneercsyalpsid02otpuerotsotrehtiedesu TOHSPANS otro,)edom
.FFOdnaNOtnemurtsniehtnrutotdesusiyeksihT
,egralootoreZ
nI.elifgolatadroneercsderotsylsuoiverpafotohspanseht EROTS ,edom
.noitcelesnevigatpeccaot,syektfosehthtiwnoitcnujnoc
sserp,levelunemynamorfyletelpmoctiuqoT.tignidecerp TIXE ehtlitnu
.yrevilednosretemaraptluafedniatrecotputeserahcihwsretemarap
orezotstpmettA.orezfo%5nihtiwsignidaeryalpsidehtfi,yalpsidecruos
ytnewteraerehT.noitarbilacecivedlanretxefosepyttnereffidforebmuna
enindnademmargorp-erperahcihwfonevele,elbaliavasnoitarugifnocksat
.yalpsidehtotneercsderotsylsuoiverpallacerotdesusiyeksihT
eht,syekrosrucehtgnisuyB.derotsneercstsalehtsyalpsidnoitceles
.snoitacolyromemhguorhtkcabrodrawrofrehtiellorcsnacrotarepo
yletaidemmilevelehtot,levelunemroneercstnerrucehtmorftixeot
O/I
ehT PUTES noitarugifnoclarenegstnemurtsniehtotsseccasedivorpyek
*PUTES
OREZ
*TUPNI
KSAT
RETNE
TIXE
ehT OREZ rodnarusaemdetcelesehtrehtieorezotdesuebnacyek
ehT TUPNI .deyalpsidsidnarusaemhcihwtcelesotdesusiyek 81
ehT KSAT roftnemurtsniehtgnirugifnocyldiparfosnaemasadesusiyek
.elbanifedresuera
ehT TUPTUO stuptuoecruosstnemurtsniehtfohcihwtcelesotdesusiyek
*TUPTUO
*EROTS
*LLACER
.deyalpsidsi
stnemurtsniehtwohnopugnidnepeD EROTS siyeksiht,putessiedom
nineercsagolyllaunam GOLATAD .edom
ehtnognidnepeD EROTS rehtoesllaceryeksihtfonoitarepo,pu-tesedom
ehT RETNE ni,ro,)atadderetnetpecca(atadretneotrehtiedesusiyek
ehT TIXE syektfosdnadrahrehtoehtllahtiwnoitcnujnocnisetarepoyek
ECRUOS/ERUSAEM .deyalpsidsineercs
* These key funtions are not available in BASIC mode
K239 Issue No.1
4
INTRODUCTION Summary of Functions
SOFT KEYS (Fig.1)
Three soft keys, designated F1, EXIT and F2 , are situated immediately below the display as shown below. These keys have their function allocated by the instrument software which is indicated in the bottom of the display (Voltage for F1 and Units for F2 in this example). They are used to select menu (program) options and are fully described under the appropriate section headings.
CURSOR KEYS (Fig.1)
The cursor keys consist of a block of four key s, arranged as shown in Figure 1. They are designated Up (Ù), Down (Ú), Left (×) and Right (Ø). In programs where options need to be selected from a list, (e.g.) the TASK selection program, the Up and Down cursor keys are used to highlight one of the options, from which it can be selected by the ENTER key. In TASK mode, where more than one page of options are provided, the Left (×) and Right (Ø) cursor keys will switch between pages.
External Interfaces under flap Power, RS 232, External Transducer
Electrical Inputs, V, I, Switch and mA sink
Test port. Connect to unit under test. G1/8" coupling. Use a sealing washer.
Hard keys
Fine pressure adjustment. Start with control screwed out half way to allow fine adjustment up or down.
Pump, pressurises test port and internal sensor. Use to approximately set output pressure
Figure 2 - DPI 610 Calibrator Controls
Cursor Keys
Soft Keys F1, EXIT, F2
Release Valve. Releases system pressure through vent port. Use to gradually, or completely reduce pressure
Vent Port
Select positive or negative pressure
K239 Issue No.15
INTRODUCTION Summary of Functions
ELECTRICAL CONNECTIONS
Protective cover
Close when not using connectors
External Transducer
USERS232 INSAFE AREAONLY
RS 232
(Connect and
use in Safe Area
Only)
Temperature Sensor
Figure 3 - Electrical System Connections
Measurement
inputs and
Source
Outputs are made via the control panel sockets as
shown below.
Input Window
max 30V
DPI 610 IS
TASK: BASIC
CAT II
INPUTS
VOLTAGE
V
OUTPUTS
PRESSURE INT
CURRENT
PRESSURE
UNITS
F2F1
bar
Figure 4 - Electrical Measurement Inputs/Source Outputs
K239 Issue No.1
6
Electrical Measurement Input Sockets
Output Window
Getting Started
Fitting Batteries
rerutcafunaM .oNepyT
rezigrenE39NEepyTlairtsudnI
rezigrenE2MA.C.41RL.39E
llecaruD41RL-0041NM
atraV2MA.C.41RLepyT4104.oN
llecorP41RL-0041NM
Low battery indicator
TASK: BASIC
VOLTAGE
PRESSURE INT
CURRENT
PRESSURE
UNITS
-
V
bar
+
Cover fixing screws
Cover fixing screws
Six Alkaline C-cells (Type only as Table)
WARNING: BATTERIES MUST ONLY BE FITTED IN A SAFE AREA.
USE ONLY THE BATTERIES SPECIFIED IN THE TABLE.
Caution:
Old batteries can leak and cause corrosion. Never leave discharged batteries in the instrument. Old batteries should be
treated as hazardous waste and disposed of accordingly.
Switching On
Press the I/O switch on the front panel and proceed as follows.
TASK: BASIC
I/O
VOLTAGE
PRESSURE INT
CURRENT
V
bar
PRESSURE
UNITS
TASK: BASIC
CURRENT
PRESSURE INT
VOLTAGE
PRESSURE
UNITS
mA
bar
The first time that the instrument is powered up, it will power up in BASIC mode with the main screen displaying voltage in the measurand display area and pressure in the source display area. To switch to
F1 to return to
Voltage
.
Current
as measurand, press F1 as shown. Similarly,
Note: No other keys are active in this mode and the instrument can only be
reconfigured by pressing the TASK key and selecting another mode.
K239 Issue No.17
Getting Started
Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the units required, press TASK and select any task, other than BASIC, press SETUP and proceed as detailed on page 31. To return to BASIC mode, press TASK and select BASIC.
TASK: BASIC
VOLTAGE
PRESSURE INT
CURRENT
PRESSURE
UNITS
V
bar
­+
SELECT UNITS OF PRESSURE
inHg
bar
hPa Pa
SELECT UNITS
In BASIC mode, the unit is configured to carry out basic Pressure to Voltage (P to V) or Pressure to Current (P to I) tests a typical test procedures follow.
Voltage and Current Measurements
Connect the electrical input sockets as follows for voltage and current measurements. Use the test leads provided and DO NOT push bare wires into the sockets.
Note: Max applied voltage = 30V dc, Max input current = 55mA dc
Vin
mA Sink
CURRENT
max 30V
mA in
55mA
-
+
max
VOLTAGE
max 30V
(FROM EXTERNAL SOURCE)
+
V
-
Vin
mA Sink
mA in
55mA
max
Note: Max applied voltage for exter nal loop supply = 30V dc
K239 Issue No.1
8
Loading...
+ 36 hidden pages