Babytherm 8004/8010
Technical Service Bulletin # 5
Re: Replacing Controller PCB, new Software Version 1.01
Date: April 30, 2001
Reference Doc: Service Bulletin from Dräger Medizintechnik GmbH:
6132.300 Babytherm 8004/8010 # 5, February 16, 2001
Reason: A new EEPROM on the Controller PCB P/N 8290674 requires the new
software version 1.01. If a new Controller PCB is used with the previous
software version error code 15 will be generated.
Solution: When replacing the PCB Controller P/N 8290674, software version 1.01
must be installed.
Devices affected: All units Babytherm 8004/8010 with installed software version 1.00.
When: With next replacement of Controller PCB P/N 8290674.
Reporting: N/A
Additional Info: N/A
Cost: Covered by Customer, post warranty.
Ordering Info: EPROM SW1.01 Babytherm 8004/8010 P/N 2M 30 440
Distribution: Dräger Service Personnel and Authorized Service Organizations.
If you have any questions, please contact Technical Support by
phone at 1-800-543-5047 or by fax at 1-215-721-5789.
Dräger Medical, Inc.
Technical Product Manager
6132.300 Babytherm 8004/8010 TSB No. 5 April 2001 1/1