Initial Installation and Setup ................. 23
Installing a Circuit ......................... 24
System Test .............................. 26
Operational Test .......................... 28
Chapter 3: STARTING HFJV ON A PATIENT
Connecting a Patient to the LifePulse .......... 31
HFJV Pressure Monitoring .................. 32
7 Steps to Begin HFJV ..................... 33
Adjusting Ventilator Settings ................ 36
Adjusting Ventilator Alarms ................. 36
Vent Inop During Test ...................... 57
Vent Inop While Operating .................. 58
Check Vent .............................. 59
Low Gas Pressure ......................... 60
Battery Alarms ............................ 61
Check Circuit ............................ 62
High Temp - Water ........................ 63
Low Temp - Water ......................... 64
High Temp - Gas .......................... 65
Low Temp - Gas ........................... 66
High or Low Level - Water ................... 67
Chapter 6: HUMIDIFICATION
Humidifier Overview ....................... 69
Patient Circuit Overview .................... 70
Starting the Humidifier ..................... 71
Stopping the Humidifier .................... 71
Water Level Sensing ....................... 72
Gas Flow through Humidifier ................ 73
Gas Flow through Patient Box ............... 74
Understanding Gas Temperature ............. 75
Understanding Water Temperature ........... 76
Temperature Controls ...................... 76
Identifying Proper Humidification ............. 77
Changing the Circuit ....................... 78
Chapter 7: BATTERY
Charging the Battery ....................... 81
Running on Battery ........................ 82
Battery Maintenance ....................... 83
Chapter 8: CLEANING AND STORAGE ....... 85
Chapter 4: PATIENT MANAGEMENT
Objectives ............................... 37
Oxygenation ............................. 38
Oxygenating Overexpanded Lungs ........... 40
Oxygenation Flow Chart .................... 41
Ventilation ............................... 42
Ventilation Flow Chart ...................... 44
Suctioning the Patient ...................... 45
Weaning the Patient ....................... 46
Chapter 5: VENTILATOR ALARMS
Alarms Overview .......................... 47
Alarms Displays and Sounds ................. 48
Adjusting Alarm Limits ..................... 49
Servo Pressure ............................ 50
Mean Airway Pressure ...................... 52
Loss of PIP ............................... 54
Cannot Meet PIP .......................... 55
High PIP ................................. 56
Appendix A
Owner Responsibility for Maintenance ......... 87
Warranty ................................ 88
Appendix B
Ventilator Alarms and Reactions .............. 89
Humidifier Alarms and Reactions ............. 90
Appendix C
Ordering Information ...................... 91
Appendix D
Device Specifications ...................... 93
Appendix E
International Symbols Table ................ 101
Appendix F
Electromagnetic Declarations and Tables ...... 105
Page 4
Page 5
Warnings and Cautions
PROLOGUE
Anyone using or involved with the use of the LifePulse High
Frequency Ventilator (HFV) should study and observe the
Warnings and Cautions listed here.
WARNINGS indicate that there is a possibility of direct injury
to either the operator or patient. The term Warning is also used
where required by law or to meet the requirements of certain
performance standards.
CAUTIONS indicate that there is a possibility of damage to the
instrument, other property, or indirectly to other individuals.
NOTES provide additional information intended to avoid
inconveniences during operation.
The following Warnings and Cautions are organized by
their relevancy to set-up, operation, or maintenance of the
LifePulse.
Set-Up of LifePulse
WARNING: Federal law restricts this device to sale by or on the
order of a licensed medical practitioner.
WARNING :This device is to be operated only by a properly
trained individual in accordance with the User Manual. Refer to
the User Manual for applicable Warnings and Cautions related to
the use of this device.
This device complies with the EMC limits for the Medical
Device Directive 93/42/EEC (EN 55011 Class I and EN 60601-1-2).
These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation.
WARNING: Use of accessories, transducers, and cables other
than those specified, with the exception of transducers and
cables sold by Bunnell Incorporated as replacement parts, may
result in increased emissions or decreased immunity of the
LifePulse HFV.
WARNING: The LifePulse HFJV should not be used adjacent
to or stacked with other equipment; if adjacent or stacked use
is necessary, the LifePulse HFJV should be observed to verify
normal operation in the configuration in which it will be used.
WARNING: Do not modify the LifePulse, Patient Box, or Patient
Circuit (the ventilator system) without authorization from Bunnell
Incorporated.
WARNING: Only use the hospital grade power cord supplied by
Bunnell that locks onto the power cord inlet receptacle on the
back of the LifePulse. Don’t substitute any other power cord.
WARNING: Only connect to a hospital grade power outlet with
a protective earth ground.
WARNING: To avoid the risk of electrical shock, this equipment
must only be connected to a supply mains with protective earth.
WARNING: Do not use the LifePulse in the presence of
flammable gases or anesthetics. There is a risk of explosion. This
Warning applies to hyperbaric chambers due to their oxygen-rich
environment.
WARNING: The ventilator may be prone to malfunction if it has
been transported, stored, or operated outside its recommended
environmental ranges.
WARNING: This equipment may cause interference with radio
communications and other devices in the vicinity if not installed
and used in accordance with the User Manual. If this device does
cause electrical interference with other devices, which can be
determined by turning the equipment off and on, the user is
encouraged to try to correct the problem by one or more of the
following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment to an outlet on a circuit different
from that to which the other device(s) is connected.
WARNING: Only use medical grade oxygen and air that is dry
and free of dust and oil. The gas supply pressure must be 30 – 60
psi (205.85 - 413.70 kPa).
WARNING: Do not operate the ventilator system or components
if they have suffered physical damage or do not appear to be
operating properly.
WARNING: Always use an air/oxygen blender to supply the
mixed gas source to the LifePulse.
WARNING: Do not connect 100% oxygen directly to the “Mixed
Gas Input” on the rear panel of the LifePulse, unless you intend
to only deliver 100% oxygen.
WARNING: Plug the ventilator into an electrical wall outlet at
all times to maintain proper battery charge. Charging must only
occur in a properly ventilated room. The user must ensure the
battery is in good condition and has sufficient capacity prior to
every patient use.
CAUTION: The capacity of the battery must be checked
regularly and the battery replaced if necessary. Avoid deep
discharges, as they lead to premature degradation of the
battery.
Bunnell LifePulse HFV User Manual | 5
Page 6
PROLOGUE
WARNING: Improper replacement of the battery by non-Bunnell
authorized service personnel could result in an unacceptable risk
(e.g., excessive temperature, fire, or explosion).
WARNING: If the battery charge indicator stays red with the
LifePulse plugged into an electrical wall outlet, the battery needs
to be checked by Bunnell-authorized service personnel. The
ventilator and humidifier will function normally when plugged in
but the battery will not provide adequate power, if needed.
WARNING: The LifePulse battery is low and needs to be
charged when the battery fuel gauge is blinking red. Plug the
LifePulse into an electrical wall outlet.
WARNING: When the battery fuel gauge is blinking red with the
BATTERY DEPLETED alarm lit and a high priority alarm active the
battery charge is too low to continue running the LifePulse and it
will soon go to Standby mode.
WARNING: The discharge rate of the battery changes with
the ventilator and humidifier settings as well as with the age of
the battery. This can shorten the overall time the ventilator will
operate on battery power before the battery depletes and the
ventilator stops operating.
WARNING: Use only a LifePort adapter with a standard
endotracheal tube and the Bunnell Patient Circuit when
operating the LifePulse Ventilator.
WARNING: Do not connect any additional tubing or pressure
monitors between the pressure monitoring tube of the LifePort
adapter and the Patient Box. Doing so will degrade the
LifePulse’s ability to measure airway pressures accurately and
may result in inappropriate pressures or volumes being delivered
to the patient.
WARNING: Do not pre-fill the humidifier cartridge of the Patient
Circuit prior to starting the LifePulse. Doing so may result in the
cartridge overfilling and the delivery of a bolus of water to the
patient. The cartridge will fill automatically once the LifePulse
begins operating.
WARNING: The LifePulse should pass two tests to ensure
proper operation. Run the System Test and an Operational Test
using a test lung before starting the LifePulse on a patient. If
the LifePulse displays a monitored PEEP different from 0 ±1 cm
H2O during the Operational Test, the Patient Box may be out of
calibration and should be replaced.
WARNING: Never initiate a test while the LifePulse ventilator
is connected to a patient. Doing so may result in inappropriate
pressures or volumes being delivered to the patient, resulting in
volutrauma.
WARNING: Remove the test lung following the successful
completion of the System and Operational Tests. Place the
LifePulse in Standby mode before connecting it to the patient.
Failure to do so may result in inappropriate pressures or volumes
being delivered to the patient, resulting in volutrauma.
CAUTION: Always keep an extra Patient Box, Patient Circuit,
and LifePort adapters near the patient’s bedside. When
operating lifesaving equipment, it is recommended to have backup equipment available for rapid intervention by properly trained
medical professionals.
WARNING: Do not drop the Patient Box. Always handle the
Patient Box with care; it is a precision device. If the Patient
Box is dropped, it must be tested by running the System and
Operational Tests using a test lung to verify normal function prior
to using it on a patient.
WARNING: Use ONLY sterile water for inhalation, USP, in the
LifePulse humidifier cartridge of the Patient Circuit. The use of
deionized water or saline solutions may cause a malfunction of
the water level sensors in the humidifier cartridge, resulting in
the cartridge overfilling and the delivery of a bolus of water to
the patient.
WARNING: Latch the water inlet tube on the humidifier
cartridge of the Patient Circuit into the water pump housing prior
to connecting to the water supply. Failure to do so may result in
cartridge overfill and delivery of water to the patient by gravity
feed.
WARNING: Clamp the water supply tube prior to opening the
water pump door to prevent cartridge overfill and delivery of
water to the patient by gravity feed. The water supply should be
positioned at or below the level of the humidifier cartridge as an
added precaution.
WARNING: Always use a conventional ventilator in tandem
with the LifePulse. The conventional ventilator is essential for
providing gas for the patient’s spontaneous breathing, PEEP, and
periodic sigh breaths.
CAUTION: Do not use pens, pencils, fingernails, or other
pointed objects to push the buttons on the LifePulse front panel.
Doing so may result in damage to the buttons and may cause
them to fail.
Bunnell LifePulse HFV User Manual | 6
Page 7
PROLOGUE
Operation of LifePulse
WARNING: In case of malfunction of any of the built-in
monitoring or LifePulse functions, the operator must still assume
full responsibility for proper ventilation and patient safety in all
situations.
WARNING: Do not use in conjunction with magnetic resonance
imaging (MRI); patient injury may result.
WARNING: Exposure to known sources of EMI with medical
devices such as diathermy, electrocautery, RFID (Radio
Frequency Identification), and electromagnetic security systems
such as metal detectors could affect the function of the LifePulse
HFV. Use of the LifePulse around such sources should be avoided
if possible until effects are known. If unexplained changes in
LifePulse performance are observed, it may be necessary to
take mitigation measures, such as re-orienting, relocating the
LifePulse, or shielding its location. If necessary, discontinue use
of the interfering device or the LifePulse. Note that the presence
of RFID devices may not be obvious.
WARNING: Do not place open containers of liquid on top of or
above the LifePulse ventilator. Liquids getting into the ventilator
can cause equipment malfunction with a risk of patient injury.
WARNING: Do not block the air intake vents. Air is drawn in
through the air intake vents to cool the ventilator’s components.
Blocking the air intake may result in overheating and possible
malfunction.
WARNING: Do not block the dump valve exhaust port on the
back of the LifePulse. Doing so may interfere with the internal
pressure being safely released during certain alarm conditions,
which could increase the risk of volutrauma.
CAUTION: Do not position the LifePulse so that it is difficult
to access the electrical power cord in case it needs to be
disconnected from the electrical wall outlet, if necessary (e.g.,
isolated from the main power supply).
WARNING: A properly trained medical professional must be
in attendance at all times when the LifePulse is connected to
a patient in order to monitor the patient and detect alarms or
other indications of a problem.
WARNING: There are no airway pressure alarms active while the
LifePulse is in the Standby mode.
WARNING: A negative monitored PEEP may indicate the airway
pressure transducer is out of calibration or faulty.
WARNING: A significant discrepancy (>4 cm H2O) between the
delivered PEEP and the monitored PEEP may indicate a potential
problem and should always be evaluated by a qualified trained
individual.
WARNING: Place the LifePulse in Standby mode if there is any
concern about the ventilator functionality or safety. Support
the patient using the tandem conventional ventilator or other
available ventilation options.
WARNING: Place the LifePulse in Standby mode prior to troubleshooting if the Patient Box pinch valve stops operating while on
a patient. This action prevents the patient from being exposed to
inappropriately large volume delivery and possible volutrauma.
WARNING: Troubleshoot alarms only in accordance with the
guidelines in the User Manual. Troubleshooting must only be
performed by properly trained individuals who have a thorough
understanding of the LifePulse ventilator and its operation.
Always attempt to correct the cause of the alarm before pressing
the ENTER button.
WARNING: Patient connections must only be made in the Standby
mode. Do not connect the LifePulse Patient Circuit to the LifePort
adapter on the patient’s ET tube while the LifePulse is running.
Failure to comply risks high pressures and volumes being delivered
to the patient, which may result in volutrauma.
WARNING: Inspect the Pressure Monitoring Tube of the LifePort
adapter for condensation as condensates may interfere with safe
and effective ventilator functionality. To mitigate condensates
in the Pressure Monitoring Tube, check the LifePort adapter’s
orientation to ensure the Pressure Monitoring Tube is pointed
upward. It may also be necessary to reduce the humidifier
temperature (cartridge/water) to minimize condensation.
Replace the LifePort adapter if necessary.
WARNING: The LifePulse requires at least 20 seconds before
adequate ventilation is delivered. Conventional ventilation
should be provided during this start-up period.
WARNING: Do not ignore audible and visual alarms. Alarms
indicate changing conditions and should never go unheeded.
WARNING: Alarms from different medical devices can sound
similar. Identify and respond to LifePulse alarms based on the
alarm priority lamp, the alarm indicator message, and the audible
alarm tone and rhythms.
WARNING: Press the ENTER button to resume ventilation if
a LOSS OF PIP alarm has occurred with the Ready indicator
off. The Patient Box will be cycling but no gas (breaths) will be
delivered to the patient until the LOSS OF PIP alarm is cleared by
pressing the ENTER button. Always troubleshoot the potential
causes of an alarm following the guidelines in the User Manual.
Bunnell LifePulse HFV User Manual | 7
Page 8
PROLOGUE
WARNING: Do not initiate ventilation with kinks or obstructions
present in the Patient Circuit or LifePort pressure monitoring
tube. Doing so may result in alarms or delays in ventilator
support. Pressures may be generated that are too high or too
low for the patient’s needs.
WARNING: Do not leave the patient’s bedside while the LifePulse
water pump is running during initial start-up or following a circuit
change. A properly trained person must observe the cartridge fill
with sterile water for inhalation, USP, to the second water level
sensing pin and the water pump stops pumping.
WARNING: The LifePulse will not detect extubation of a
patient during start-up (i.e., from the time the ENTER button is
pushed until the Ready indicator is illuminated). During start-up
the ventilator may appear to be functioning normally despite
extubation and will generate no audible or visual alarms. A
properly trained individual must verify the patient is being
ventilated following start-up.
WARNING: There will be no LOSS OF PIP alarm for the first 15
seconds after the ENTER button is pushed. A properly trained
person must observe the LifePulse reach the set PIP and the Ready
indicator illuminate before leaving the patient’s bedside.
WARNING: Do not press the ENTER button during a LOSS
OF PIP alarm with the Ready indicator illuminated if the patient
is stable and the Servo has locked at or near an established
operating value. Doing so will temporarily silence the audible
alarm and unlock the LifePulse Servo. Always troubleshoot the
potential causes of an alarm following the guidelines in the User
Manual prior to pressing the ENTER button.
WARNING: Do not manually adjust the alarm volume to the point
that the audible alarm cannot be recognized (is effectively off).
Doing so will limit the LifePulse’s ability to alert clinicians to alarm
conditions and expose the patient to potentially unsafe conditions
that may result in injury.
WARNING: The conventional ventilator rate should not exceed
10 breaths per minute when running in tandem with the LifePulse
ventilator. Doing so could result in excessive Mean Airway
Pressure or minute volume being delivered to the patient. The
recommended conventional ventilator rate is 0 - 5 breaths per
minute.
NOTE: PIP, MAP, and PEEP are reported as approximate tracheal
pressures and are measured inside the LifePort adapter. They may
not accurately reflect alveolar pressures and, in fact, may be higher
or lower than alveolar pressures.
WARNING: Always evaluate inadvertent PEEP (LifePulse
monitored PEEP greater than CV set PEEP). The patient’s lung
volume should be assessed when inadvertent PEEP is present.
Inadvertent PEEP reduces delta pressure, which may result in
under-ventilation (hypercarbia).
WARNING: Place the LifePulse ventilator in Standby mode prior
to suctioning to avoid fluctuations in the monitored pressures that
may result in alarm conditions or possible injury to the patient.
WARNING: The LifePulse must be in Ready mode prior to
suctioning if suctioning while the LifePulse is running. Failure to do
so may result in inappropriate pressure or volumes being delivered
to the patient.
WARNING: Always troubleshoot a change in Servo that is greater
than 1.0 psi (6.89 kPa) from the previously established baseline,
especially if the LifePulse control settings have not been changed.
Servo changes this large or larger usually represent a mechanical
problem and may pose a risk of volutrauma.
WARNING: The patient is not being ventilated by the LifePulse
during VENT INOP Alarms 01 & 05-09, which place the LifePulse
in Standby mode.
WARNING: The patient is not being ventilated by the LifePulse
during VENT INOP Alarm 10, which places the LifePulse into
Standby mode. Always troubleshoot the potential causes of an
alarm following the guidelines in the User Manual prior to pressing
the ENTER button. Pressing the ENTER button will re-initiate
ventilation immediately.
WARNING: Do not manually adjust alarm limits around MAP and
Servo to the point that they become irrelevant (i.e., are effectively
off). Doing so will negate critical alarms and expose the patient to
potentially unsafe conditions that may result in injury.
WARNING: Place the LifePulse in Standby mode before unkinking
any portion of the Patient Circuit. This action prevents the patient
from being exposed to inappropriately high pressure or large
volume delivery (volutrauma).
WARNING: Do not press the PUSH TO LOAD button on the
Patient Box while the LifePulse is running on a patient. Doing so
could result in the patient being exposed to inappropriately high
pressures or large volume delivery (volutrauma).
WARNING: Do not tip the humidifier cartridge while removing it
from the cartridge housing if the Patient Circuit is connected to
the patient’s ET tube via the LifePort adapter. Doing so may result
in water spilling into the Patient Circuit and a bolus of water being
delivered to the patient.
WARNING: Do not open the water pump door while the pump is
pumping. Doing so may result in a pinch hazard to the operator.
Bunnell LifePulse HFV User Manual | 8
Page 9
PROLOGUE
WARNING: Removal of the humidifier cartridge from the
cartridge housing while the LifePulse is running exposes the heater
plate and cartridge heater cup, which may be in excess of 60ºC.
WARNING: Always attempt to minimize excessive condensation
(rainout) in the Patient Circuit between the ventilator and the
Patient Box. Failure to do so may result in water interfering with
the monitored pressure signal at the LifePort adapter resulting in
fluctuations in the delivered PIP.
WARNING: Do not under-humidify gas being delivered to the
patient. Adjust the cartridge temperature so that condensation
is always present in the green delivery tube between the Patient
Box and the LifePort adapter.
WARNING: External temperature of the heated portion of the
patient circuit may exceed 50°C. Contact with the patient’s skin
should be avoided.
WARNING: Pushing the PAUSE button on the Humidifier shuts
off the heating elements that provide for gas warming and
humidification. A properly trained individual must not leave the
patient’s bedside with the Humidifier in the Pause mode.
WARNING: Raising or lowering the Gas temperature setting
may raise or lower the patient’s temperature. The normal setting
of 40°C is designed to keep the patient from losing heat through
respiration. It should not be changed under normal operating
circumstances.
WARNING: Do not move the LifePulse while it is running on a
patient. Doing so may result in water entering the Patient Circuit
and creating a risk of water delivery to the patient. Monitor the
water level closely under these circumstances
WARNING: Running the LifePulse while the ventilator or patient
is in motion or while high-flow or large tidal volume conditions
exist may result in water entering the Patient Circuit and creating
a risk of aspiration. Please monitor the water level closely.
WARNING: Periodically check the gas filter/water trap. If water
is visible, drain the water by opening the valve. Failure to do so
may result in a ventilator malfunction.
Maintenance/Cleaning of LifePulse
CAUTION: The LifePulse ventilator system must receive
inspections and service at regular intervals. Preventive
maintenance should be performed every 2000 hours or at least
once a year. Overhauls should be performed every 6000 hours of
use. All service and maintenance must be performed by Bunnell
authorized service personnel.
to the cleaning instructions in the User Manual.
WARNING: Do not store LifePulse with the pinch tube of the
circuit loaded in the pinch valve jaws of the Patient Box.
WARNING: Disconnect the battery before disassembly of the
ventilator.
WARNING: Do not remove the LifePulse cover due to possible
shock hazard. The ventilator should only be serviced by Bunnell
authorized service personnel.
WARNING: Severe electrical shock to the user or service
personnel may be delivered by the battery and/or battery
charger if direct contact is made with their electrical connections.
CAUTION: If the battery case is damaged (e.g., from dropping
or impact during servicing or shipping) the battery must be
replaced. Otherwise, it may leak battery acid and cause the
LifePulse to malfunction.
CAUTION: Always plug the LifePulse power cord into an
electrical wall outlet during storage between uses. This action
will help avoid deep discharges of the battery which cause
degradation and decrease battery life.
WARNING: The Patient Circuit and/or Patient Box should be
removed to release any pressurized gas before servicing the
ventilator.
CAUTION: Batteries that are improperly disposed of may cause
harm to the environment.
CAUTION: The LifePulse, Patient Box, and Patient Circuit
contain components made of hazardous materials that need to
be disposed of appropriately. Improper disposal may cause harm
to the environment.
CAUTION: Do not sterilize the LifePulse ventilator or Patient
Box. The internal components are not compatible with
sterilization techniques.
WARNING: Do not reuse or sterilize Patient Circuits or LifePort
adapters. Failure to discard the Patient Circuit after one use or
seven days may result in leaks, improper temperature, and water
level control.
WARNING: The use of non-recommended solvents or cleaners
can damage the enclosure(s) or user interface and damage
equipment causing the LifePulse to malfunction. Refer to the
cleaning instructions in the User Manual.
CAUTION: Do not ship the LifePulse ventilator or Patient Box
in any packaging other than a factory authorized carton or case.
Doing so may result in damage to the ventilator and will void the
warranty.
WARNING: Clean and disinfect the LifePulse and Patient Box
prior to patient applications to avoid the risk of infection. Refer
Bunnell LifePulse HFV User Manual | 9
Page 10
PROLOGUE
Indications for Use
The Bunnell LifePulse High Frequency Ventilator (HFV) is indicated
for use in ventilating critically ill infants with pulmonary interstitial
emphysema (PIE). Infants studied ranged in birth weight from 750
to 3529 grams and in gestational age from 24 to 41 weeks.
The Bunnell LifePulse High Frequency Ventilator is also indicated
for use in ventilating critically ill infants with respiratory distress
syndrome (RDS) complicated by pulmonary air leaks who are, in
the opinion of their physicians, failing on conventional ventilation.
Infants of this description studied ranged in birth weight from 600
to 3660 grams and in gestational age from 24 to 38 weeks.
Contraindications
Use of this device is contraindicated in patients requiring tracheal
tubes smaller than 2.5 mm ID.
This device is not intended for use outside of a device user facility
(21 CFR 821.3 (g)).
Adverse Side Effects
The adverse side effects noted during the use of high frequency jet
ventilation (HFJV) include those commonly found during the use
of conventional positive pressure ventilators. These adverse effects
include:
to sale by or on the order of a licensed medical
practitioner.
WARNING: This device is to be operated only in
accordance with the User Manual. Refer to the
User Manual for applicable Warnings and Cautions
related to the use of this device.
WARNING: Do not use the LifePulse in the
presence of flammable gases or anesthetics.
There is a risk of explosion. This warning applies
to hyperbaric chambers due to their oxygen-rich
environment.
PROLOGUE
Intended User
The LifePulse should be used only by those who
• are professional health care providers,
• have received training in the use of the LifePulse system, and
• have knowledge of pulmonary physiology and mechanical
ventilation
Use Environment
The LifePulse should be used only
WARNING: Do not use in conjunction with
magnetic resonance imaging (MRI); patient injury
may result.
WARNING: Do not modify the LifePulse, Patient
Box, or Patient Circuit (the ventilator system)
without authorization from Bunnell Incorporated.
WARNING: Exposure to known sources of
EMI with medical devices such as diathermy,
electrocautery, RFID (Radio Frequency
Identification), and electromagnetic security
systems such as metal detectors could affect the
function of the LifePulse HFV. Use of the LifePulse
around such sources should be avoided if possible
until effects are known. If unexplained changes
in LifePulse performance are observed, it may be
necessary to take mitigation measures, such as
re-orienting, relocating the LifePulse, or shielding
its location. If necessary, discontinue use of the
interfering device or the LifePulse. Note that the
presence of RFID devices may not be obvious.
• in hospital critical care units (e.g., NICU, PICU)
• during transport of a patient within hospitals or health care
facilities
Servicing Guidelines
Regular Service: The LifePulse must be serviced at regular
intervals by Bunnell Incorporated authorized service personnel who
have received specialized training.
Complete Service Records: All service performed on the LifePulse
must be recorded in a service log in accordance with hospital
procedures and local and national regulations.
Service Contract: We strongly recommend, for optimal care and
performance, that all service on the LifePulse be performed as part
of a service contract with Bunnell Incorporated.
Bunnell LifePulse HFV User Manual | 11
Page 12
WARNING: The ventilator may be prone to
malfunction if it has been transported, stored, or
operated outside its recommended environmental
ranges.
WARNING: Disconnect the battery before
disassembly of the ventilator.
WARNING: Do not remove the LifePulse cover
due to possible shock hazard. The ventilator
should only be serviced by Bunnell authorized
service personnel.
WARNING: The Patient Circuit and/or Patient Box
should be removed to release any pressurized gas
before servicing the ventilator.
PROLOGUE
Disclaimers
Improper Use Environment: Bunnell Incorporated has no
responsibility for the safe operation of the LifePulse if the Intended
Use and User requirements specified in this document are not
followed.
Nonprofessional Servicing: Bunnell Incorporated has no
responsibility for the safe operation of the LifePulse if service or
repairs are performed by persons other than Bunnell Incorporated
authorized service personnel.
The owner/user of this equipment will have the sole responsibility
and liability for any damage or injury to persons or property
(including the equipment itself) resulting from
• Operation not in accordance with the user instructions
• Faulty maintenance not in accordance with the authorized
maintenance instructions
• Repair by anyone other than Bunnell authorized service
personnel
• Modification of the equipment or accessories
• Use of components or accessories that have either been
damaged or not authorized for use with this equipment by
Bunnell Incorporated
The disposal of any Bunnell Incorporated equipment is the
responsibility of the owner of that equipment. Disposal
should always be in accordance with any local and/or national
requirements.
Bunnell LifePulse HFV User Manual | 12
Page 13
1
5
LifePulse
HIGH FREQUENCY VENTILATOR
MONITOR
29.3
PIP
8.2
PEEP
Chapter 1: OVERVIEW
Monitor Section
2
2.8
SERVO
cm H2O
2I.I
cm H2O
Δ
P
cm H2O
II.4
MAP
cm H2O
3
bunnell
INSPIRED INFANT CARE
4
The displays in the Monitor section approximate the pressures at
the tip of the endotracheal tube and the internal servo or drive
pressure of the LifePulse.
The LifePulse’s pressure monitoring system is active in both
Standby and Ready modes. The pressures are updated in the
Monitor displays every 10 seconds in Standby mode and every 2
seconds in Ready mode.
The LifePulse monitors the following:
1. PIP
Peak Inspiratory Pressure, the average maximum airway
pressure measured for each HFJV breath
2. Servo
Internal drive pressure of the LifePulse; indicates how much gas
flow must be produced to meet the set PIP, Rate, and I-Time
requested by the operator for a particular patient
SERVO
Servo is an indication of how much gas flow the
LifePulse must generate to meet the requested
settings. It is regulated by the microprocessor and is
outside the control of the operator. Bigger patients,
or those with more compliant lungs, will require higher
Servo. Infants with smaller and/or less compliant lungs
will require lower Servo.
Changes in Servo above or below the established
operating level for particular settings alert clinicians
that the LifePulse’s performance has changed or that
the patient’s condition (e.g., lung compliance, airway
resistance, or lung volume) may be improving or
worsening.
3. ∆P
PIP minus PEEP; proportional to tidal volume; as ∆P increases,
tidal volume increases and vice versa.
4. MAP
Mean Airway Pressure, an average of all pressures applied
to the airway. MAP includes pressures produced by HFJV,
conventional ventilation, or spontaneous breathing.
5. PEEP
Positive End-Expiratory Pressure, the average minimum
airway pressure
Bunnell LifePulse HFV User Manual | 13
Page 14
1234
READY
+
-
UPPER
LIMIT
SYSTEM
TEST
+
-
UPPER
LIMIT
LOWER
LOWER
LIMIT
LIMIT
ALARMS
LOSS OF PIP
CANNOT MEET PIP
HIGH PIP
BATTERY DEPLETED
LOW GAS PRESS
CHECK VENT
VENT INOP
AUDIO
PAUSED
Chapter 1: OVERVIEW
Alarms Section
The LifePulse alarm system alerts the operator, both audibly and
visually, to changes in the ventilator or the patient. The alarm
statements are not visible until they are lit.
The Alarms section has 6 key features:
1. UPPER and LOWER LIMIT buttons
adjust the limits for the Servo and Mean Airway Pressure (MAP)
alarms
2. READY indicator
when lit, indicates that all internal alarms have been
automatically set
LifePulse
HIGH FREQUENCY VENTILATOR
bunnell
INSPIRED INFANT CARE
567
MORE ON ALARMS
See Ventilator Alarms Chapter for a detailed
description of each alarm, how to interpret and
troubleshoot alarms, and how to adjust MAP and
Servo alarm limits.
WARNING: Do not ignore audible and visual
alarms. Alarms indicate changing conditions and
should never go unheeded.
3. Alarm Priority Lamp
a visual alert for an alarm and its priority, based on color and
flashing sequence
4. AUDIO PAUSED button
silences audible alarms for 60 seconds
5. ALARMS indicators
display various potentially hazardous conditions
6. Battery Fuel Gauge
indicates the charge capacity of the battery
7. SYSTEM TEST button
self-tests the LifePulse’s components
Alarm indicators include the following:
• LOSS OF PIP
• CANNOT MEET PIP
• HIGH PIP
• BATTERY DEPLETED
• LOW GAS PRESS(URE)
• CHECK VENT
• VENT INOP
Bunnell LifePulse HFV User Manual | 14
Page 15
Chapter 1: OVERVIEW
02
VENT INOP
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
I:E RATIO
SYSTEM
TEST
LifePulse
bunnell
INSPIRED INFANT CARE
BEEP
BEEP
BEEP
BEEP
READY
BEEP
ALARMS
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
cm H2O
BPMSECONDS
LifePulse
bunnell
INSPIRED INFANT CARE
BATTERY
READY
o
C
BEEP
BEEP
BEEP
ALARMS
Alarms Terminology
±
High Priority
Alarm Priority Lamp – flashing red
Alarm Sound – 2 bursts of 5 beeps every 5 seconds
• LOSS OF PIP
• HIGH PIP
• VENT INOP w/ Codes (01-10)
• HIGH LEVEL
• CHECK CIRCUIT LEVEL
• BATTERY DEPLETED
• Power Up (no alarm indicator)
• VENT INOP (no code) - Vent stopped or reset via Watchdog
circuit
• Vent reset via software or backup vent alarm detected
• Battery missing, malfunction, or cannot accept a charge
(no alarm indicator)
Power Down
Alarm Sound – 1 beep every 0.5 seconds
Bunnell LifePulse HFV User Manual | 15
Page 16
Chapter 1: OVERVIEW
Controls Section
1
2345
The Controls section contains the parameters you will adjust when
managing a patient during HFJV with the LifePulse. The Controls
section has 7 key features:
1. PIP
displays the set PIP in cm H2O that the LifePulse will deliver to
the patient
CONTROLS
+
-
cm H2O
420
RATEI-TIME
20
PIP
LifePulse
HIGH FREQUENCY VENTILATOR
+
.0206.I
-
BPMSECONDS
+
:
I
-
I:E RATIO
bunnell
INSPIRED INFANT CARE
76
MORE ON CONTROLS
See Patient Management Chapter for a detailed
description of when and how to adjust settings.
2. Rate
displays the set Rate in breaths per minute (bpm) that the
LifePulse will deliver to the patient
3. I-Time
displays the inspiratory time in seconds that will be delivered
each breath.
4. I:E Ratio
displays the ratio of inspiratory time to expiratory time at the
current LifePulse settings
5. STANDBY Button
pauses HFJV
6. ENTER Button
begins HFJV at the set PIP, Rate, and I-Time
7. Adjustment Buttons
increase or decrease PIP, Rate, and I-Time settings; each
parameter has its own adjustment buttons
Bunnell LifePulse HFV User Manual | 16
Page 17
+
-
HIGH FREQUENCY VENTILATOR
HUMIDIFIER
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH TEMP
LOW LEVEL
cm H2O
88.8
LifePulse
bunnell
INSPIRED INFANT CARE
PURGE
PATIENT
BOX
BATTERY
_
+
GAS OUT
o
C
1
2345
6
1098
Chapter 1: OVERVIEW
Humidifier Section
The Humidifier section contains the components required to
heat and humidify the gas that is delivered to the patient by the
LifePulse. The Humidifier section has 10 key features:
1. PAUSE Button
places the humidifier into Standby mode and silences
humidifier alarms
2. Temp Display
displays the current gas temperature, but toggles between
the gas temperature and the set temperatures of both gas
and water
3. Adjustment Buttons
increase or decrease the gas and water temperature settings
4. GAS Temperature Button
when pressed, displays the temperature setting of the gas that
is delivered to the patient
7
5. WATER Temperature Button
displays the temperature setting of the water that humidifies
the gas in the cartridge
6. Humidifier Alarms
display various Gas and Water alarms
7. Battery Charge Indicator
indicates if battery is charging.
8. Water Pump
transfers water from a water supply to the humidifier cartridge
9. WhisperJet Connection
receives the cable connector from the WhisperJet Patient Box
10. Cartridge Receptacle
contains the humidifier cartridge of the Patient Circuit
Patient Circuit
The LifePulse humidifier uses a disposable cartridge-and-tubing
set called the Patient Circuit.
The Patient Circuit heats and humidifies the gas that is delivered
from the LifePulse to the Patient Box where it is broken up into
breaths that are pushed through the jet nozzle in the LifePort
adapter into the patient’s endotracheal tube.
A complete description of the Patient Circuit is in the Humidifier
Chapter.
Bunnell LifePulse HFV User Manual | 17
Page 18
Chapter 1: OVERVIEW
Rear Panel
1
234
The Rear Panel has the following components:
1. Air Intake
pulls air into the LifePulse to cool electronic components
DO NOT BLOCK
2. Ventilator Power
main power switch for turning the LifePulse on and off
3. Alarm Volume
adjusts the volume of the audible alarm
4. Oxygen Sample Switch
turns on and off the flow of oxygen to the oxygen sample port
5. Gas Filter/Water Trap
collects particulates, humidity, and water that may have
inadvertently entered the gas supply
67891011
CHECK GAS FILTER/WATER TRAP
Gas filter/water trap should be checked periodically for
water accumulation. Drain the water as necessary.
WARNING: Do not block the air intake. Air is drawn
in through the air intake vents to cool the ventilator’s
components. Blocking the air intake may result in
overheating and possible malfunction.
WARNING: Do not block the dump valve exhaust
port on the back of the LifePulse. Doing so may
interfere with the internal pressure being safely
released during certain alarm conditions, which could
increase the risk of volutrauma.
6. Mixed Gas Input
5
receives the high-pressure oxygen hose from an air/oxygen
blender
7. Dump Valve & Pressure Relief Valve Exhaust Port
exhausts gas away from the patient in alarm conditionsDO NOT BLOCK
8. Oxygen Sample Port
allows an oxygen sensor to be attached to the LifePulse
for FiO2 measurements (O2 Sensor Flow Diverter)
9. Potential Equalization Conductor
accessible stud provided for easily checking the leakage
current and ground continuity of the device; these checks are
performed periodically by the biomedical engineering staff as
part of their routine safety evaluations (ISO60601-1 3rd Edition
Clause 8.6.7.)
10. Circuit Breaker
interrupts and resets the power to the LifePulse to protect it
11. Electrical Plug Inlet
receives the main power cord connector for the electrical plug
WARNING: Periodically check the gas filter/water
trap. If water is visible, drain the water by opening
the valve. Failure to do so may result in a ventilator
malfunction.
Bunnell LifePulse HFV User Manual | 18
Page 19
Top View
Chapter 1: OVERVIEW
1
Patient Box
The Patient Box makes and measures the high-frequency breaths.
It attaches to the front panel of the LifePulse and is placed near
the patient during operation.
The Patient Box has 5 key features:
1. PUSH TO LOAD Button
pressed to allow insertion of the pinch tube portion of the
Patient Circuit into the jaws of the pinch valve
2
45
3
2. Pressure Monitoring Port
receives a pressure pulse from the patient through the
LifePort’s pressure monitoring tube and sends the signal
to a pressure transducer
3. Cable Connector
connects the Patient Box to the LifePulse
Side View
WARNING: Do not drop the Patient Box. Always
handle the Patient Box with care; it is a precision
device. If the Patient Box is dropped, it must be
tested by running the System and Operational Tests
using a test lung to verify normal function prior to
using it on a patient.
WARNING: Do not place open containers of liquid
on top of or above the LifePulse ventilator. Liquids
getting into the ventilator can cause equipment
malfunction with a risk of patient injury.
4. Pinch Valve
pinches and releases the circuit pinch tube to create
HFJV breaths
5. Purge Port
receives a continuous flow of gas through the circuit purge
tubing; the gas flow is used to maintain the patency of the
LifePort pressure monitoring tube
Patient Box Storage
A patient box storage area is located behind the handle on top of
the LifePulse. This area is intended to hold the patient box when
not in use.
Do not place open fluids in the patient box storage area.
Bunnell LifePulse HFV User Manual | 19
Page 20
Chapter 1: OVERVIEW
LifePort Adapter
The LifePulse is designed to be used with the LifePort adapter
distributed by Bunnell Incorporated. This adapter allows the CV
3
1
2
and the LifePulse to be simultaneously connected to a patient.
The LifePort has three main features:
1. 15 mm Connector
provides the standard connection to the CV circuit
2. Jet Port
connection for the Patient Circuit that delivers the high frequency breaths provided by the LifePulse
3. Pressure Monitoring Tube
connects to the Patient Box in order to measure and the airway
pressures
Replace the standard 15 mm endotracheal (ET) tube adapter with
the Bunnell LifePort adapter appropriate for the patient’s ET tube.
1
23
LifePort Cutaway Diagram
WARNING: Inspect the Pressure Monitoring Tube of the
LifePort adapter for condensation as condensates may
interfere with safe and effective ventilator functionality.
To mitigate condensates in the Pressure Monitoring
Tube, check the LifePort adapter’s orientation to ensure
the Pressure Monitoring Tube is pointed upward. It may
also be necessary to reduce the humidifier temperature
(cartridge/water) to minimize condensation. Replace the
LifePort adapter if necessary.
Bunnell LifePulse HFV User Manual | 20
Page 21
LIFEPULSE COMPONENTS
• LifePulse Ventilator
• WhisperJet Patient Box
• Patient Circuit
• LifePort Adapter
• LifePulse Pole Stand
Chapter 1: OVERVIEW
Use with Conventional Ventilators
The LifePulse is used in conjunction with a conventional ventilator
(CV). The CV has four functions:
• provides fresh gas for the patient’s spontaneous breathing
• regulates positive end expiratory pressure (PEEP)
The purpose of the PEEP from the CV is to maintain the inflation
of alveoli for adequate functional residual capacity. CV PEEP is the
primary control for MAP.
The purpose of the supplementary IMV from the CV is to provide
sigh breaths sufficient to recruit atelectatic alveoli.
Bunnell LifePulse HFV User Manual | 21
Page 22
Chapter 1: OVERVIEW
Bunnell LifePulse HFV User Manual | 22
Page 23
Chapter 2: SETUP AND TESTS
WARNING: Do not operate the ventilator system or
components if they have suffered physical damage or
do not appear to be operating properly.
WARNING: Only use medical grade oxygen and air
that is dry and free of dust and oil. The gas supply
pressure must be 30 – 60 psi (205.85 - 413.70 kPa).
WARNING: Always use an air/oxygen blender to
supply the mixed gas source to the LifePulse.
WARNING: Do not connect 100% oxygen directly
to the “Mixed Gas Input” on the rear panel of the
LifePulse, unless you intend to only deliver 100%
oxygen.
WARNING: Only use the hospital grade power cord
supplied by Bunnell that locks onto the power cord
inlet receptacle on the back of the LifePulse. Don’t
substitute any other power cord.
WARNING: Only connect to a hospital grade power
outlet with a protective earth ground.
Initial Installation and Setup
A. Activities to be Performed Upon Receipt of a LifePulse
1. Remove the ventilator from its shipping carton. Save the
carton for return shipping, as needed.
2. Visually inspect the ventilator to make sure there is no
damage as a result of shipping.
3. Contact Bunnell Incorporated to report the receipt of the
ventilator and set up in-service training. Training is provided
at the hospital by a Bunnell Clinical Specialist and training
materials are available online for easy access.
4. Set up the ventilator, install a Patient Circuit, and
perform the tests as described in this chapter.
5. Complete and sign the warranty registration form and return
it to:
Bunnell Incorporated
436 Lawndale Drive
Salt Lake City, UT 84115
WARNING: Plug the ventilator into an electrical wall
outlet at all times to maintain proper battery charge.
Charging must only occur in a properly ventilated
room. The user must ensure the battery is in good
condition and has sufficient capacity prior to every
patient use.
1. Attach an air/oxygen supply from either a low flow blender
(0-30 L/min) or the low flow output (2-100 L/min) of a
standard blender to the Mixed Gas Input connector on the
rear panel.
2. Attach the power cord provided by Bunnell to the inlet
receptacle on the rear panel, lock it in place, and plug the
LifePulse into a hospital grade electrical outlet.
3. Verify that the battery charge indicator is green (blinking
or solid).
4. Attach the Patient Box to its connector on the front panel.
Bunnell LifePulse HFV User Manual | 23
Page 24
HUMIDIFIER
WARNING: Do not pre-fill the humidifier cartridge
of the Patient Circuit prior to starting the LifePulse.
Doing so may result in the cartridge overfilling and
the delivery of a bolus of water to the patient. The
cartridge will fill automatically once the LifePulse
begins operating.
1
PAUSE
88.8
TEMPERATURE
Chapter 2: SETUP AND TESTS
Installing a Circuit
A Patient Circuit must be installed in order to use the LifePulse.
Follow these steps to install the Patient Circuit:
1. Open cartridge door and insert humidifier cartridge into
receptacle.
2. Open water pump door by lifting up. Secure the water inlet
tube inside the water pump and snap the pump door closed.
3. Attach the green gas inlet tube to the green-coded Gas Out
port on the LifePulse.
+
-
GASWATER
Life Pulse
HIGH FREQUENCY VENTILATOR
3
GAS OUT
PURGE
bunnell
INSPIRED INFANT CARE
2
PATIENT
BOX
BATTERY
_
+
(Continued on next page)
Bunnell LifePulse HFV User Manual | 24
Page 25
WARNING: Use ONLY sterile water for inhalation,
USP, in the LifePulse humidifier cartridge of the
Patient Circuit. The use of deionized water or saline
solutions may cause a malfunction of the water level
sensors in the humidifier cartridge, resulting in the
cartridge overfilling and the delivery of a bolus of
water to the patient.
WARNING: Never initiate a test while the LifePulse
ventilator is connected to a patient. Doing so may
result in inappropriate pressures or volumes being
delivered to the patient, resulting in volutrauma.
Chapter 2: SETUP AND TESTS
4. Attach the purge tube to the yellow-coded Purge port on the
LifePulse and to the yellow-coded port on the Patient Box.
5. Press PUSH TO LOAD on top of the Patient Box and gently
stretch the pinch tube into the jaws of the pinch valve.
6. Connect the clear pressure monitoring tube from a LifePort
adapter to its port on the Patient Box.
7. Insert the green delivery tube into the Jet port on the LifePort
adapter.
8. Attach the water transfer tube to the water inlet tube and spike
the water supply. The water supply should be at or below the
level of the humidifier cartridge.
9. Unclamp the water transfer tube.
The circuit is now installed and ready for testing.
4
8
Water Supply
9
5
7
4
6
Bunnell LifePulse HFV User Manual | 25
Page 26
WARNING: Never initiate a test while the LifePulse
ventilator is connected to a patient. Doing so may
result in inappropriate pressures or volumes being
delivered to the patient, resulting in volutrauma.
Chapter 2: SETUP AND TESTS
System Test
Note: Some steps may already have been completed.
1. Plug in the LifePulse to a hospital grade electrical outlet.
WARNING: The LifePulse should pass two tests to
ensure proper operation. Run the System Test and
an Operational Test using a test lung before starting
the LifePulse on a patient. If the LifePulse displays a
monitored PEEP different from 0 ±1 cm H2O during
the Operational Test, the Patient Box may be out of
calibration and should be replaced.
3
2. Attach an air/oxygen supply from either a low flow blender
(0-30 L/min) or the low flow output (2-100 L/min) of a standard
blender to the Mixed Gas Input connector on the rear panel.
3. Turn the Ventilator Power switch on the rear panel to the ON
position and verify that the LEDs and displays in the Humidifier
section light up.
The LifePulse activates into Standby mode with an audible alarm
sounding. The alarm lamp on top of the LifePulse will flash red.
This alarm may be cleared by pushing AUDIO PAUSED. The
STANDBY button will illuminate, the alarm lamp will turn yellow,
and an audible alarm will sound every 18 seconds.
4. Attach a LifePort adapter to an ET tube and test lung. Do not
attach the conventional ventilator. Leave the 15 mm connector
of the LifePort open to room air.
5. Insert the green delivery tube of the Patient Circuit into the
Jet port on the side of the LifePort adapter.
2
1
4
5
Bunnell LifePulse HFV User Manual | 26
Page 27
Chapter 2: SETUP AND TESTS
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
SYSTEM
TEST
LifePulse
bunnell
INSPIRED INFANT CARE
ALARMS
o
C
88
888
.8888.8
:
8
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
ALARMS
+
+
+
-
-
-
+
-
+
-
+
-
AUDIO
PAUSED
STANDBY
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH TEMP
LOW LEVEL
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
cm H2O
BPMSECONDS
88.8
88.8
-8.8
88.8
88.8
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
20.
420
.020
6.I
:
I
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
ALARMS
+
+
+
-
-
-
+
-
+
-
+
-
AUDIO
PAUSED
STANDBY
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH TEMP
LOW LEVEL
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
cm H2O
BPMSECONDS
00.0
00.0
00.0
00.0
00.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
6. Connect the clear pressure monitoring tube of the LifePort
adapter to its port on the Patient Box.
7. Press SYSTEM TEST button in the alarm section of the
LifePulse.
6
An automatic test begins which determines the integrity of all the
ventilator’s electronics and valves. The displays will count from 0 to
9 to assure no LEDs are burned out. An internal test is performed
during the count.
7
Note: The humidifier LEDs and displays do not light up during the
System Test.
If necessary, the system test may be stopped immediately by
pressing STANDBY.
8. Make sure all the ventilator LEDs and displays on the front
panel are lit, and listen to make sure the audible indicators are
activated. Note: The humidifier LEDs and displays do not light
up during the System Test.
If no problems are detected by the LifePulse, the default settings
will be displayed in the Controls section, all zeros will appear
in the Monitor section, and the audible and visual alarms will
activate. The alarm indicator lamp will be flashing red.
8
If problems are detected, refer to Alarms Chapter to troubleshoot
Vent Inop alarms that occur during the System Test.
9. Silence the audible and visual alarms by pressing AUDIO
PAUSED button to the right of the Alarms section.
This action resets the audible and visual alarms.
9
Bunnell LifePulse HFV User Manual | 27
Page 28
Chapter 2: SETUP AND TESTS
+
+
+
+
20.
420
.0206.I
:
I
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
ALARMS
+
+
+
-
-
-
+
-
+
-
+
-
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH TEMP
LOW LEVEL
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
cm H2O
BPMSECONDS
00.0
00.0
0.0
00.0
00.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
20
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PIP
cm H2O
cm H2O
20
LifePulse
bunnell
INSPIRED INFANT CARE
READY
ALARMS
Operational Test
The Operational Test will ensure that the LifePulse is able to
1
ventilate at the set parameters and that the Patient Box is
functioning properly.
1
To perform the test, follow these steps (Note: Some steps may
already have been completed):
1. Attach a LifePort adapter to an ET tube and test lung. Do not
attach the conventional ventilator. Leave the 15 mm connector
of the LifePort open to room air.
2. Connect the LifePort adapter to the Patient Circuit and
Patient Box as described above in the System Test.
3. Press ENTER to activate the default control settings.
4. Make sure the monitored PIP is able to reach the set PIP, the
READY light illuminates, and the PEEP displays 0.0.
2
(Continued on next page)
4
3
Bunnell LifePulse HFV User Manual | 28
Page 29
Chapter 2: SETUP AND TESTS
ALARMS
READY
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PEEP
cm H2O
0.0
LifePulse
bunnell
INSPIRED INFANT CARE
ALARMS
20.
420
.0206.I
:
I
ALARMS
+
-
+
-
+
-
STANDBY
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
PIP
RATEI-TIME
I:E RATIO
cm H2O
cm H2O
cm H2O
cm H2O
cm H2O
BPMSECONDS
0.0
LifePulse
bunnell
INSPIRED INFANT CARE
0.0
0.0
0.0
0.0
READY
ALARMS
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
cm H2O
BPMSECONDS
LifePulse
bunnell
INSPIRED INFANT CARE
BATTERY
READY
o
C
ALARMS
5
5. Verify that the Ready indicator illuminates.
6. Make sure the PEEP displays 0.0 ±1.0 cm H20.
If the PEEP displays a value greater than 1.0 cm H2O or less than -1.0 cm
6
H2O, call the Bunnell Hotline before proceeding or change the Patient Box
and run both tests again.
While the LifePulse is running on the test lung, switch the circuit breaker
on the rear panel to the OFF position (see Rear Panel section in Overview
chapter). This will force the LifePulse to switch to battery power. The
battery indicator light will illuminate “BATTERY” in the Alarms section and
the low level alarm will be activated. Cancel the audible alarm by pressing
the Audio Paused button.
7. If the Battery Fuel Gauge displays one flashing red bar with a medium
priority alarm, or the ventilator stops operating, the LifePulse’s battery
should be further evaluated by calling the Bunnell Hotline before it is
used on a patient.
7
Switch the circuit breaker on the rear panel to the ON position. This will
switch the LifePulse back to the wall power outlet. The Battery indicator
light will turn off in the Alarms section.
8. Press STANDBY after a successful Operational Test.
WARNING: Remove the test lung following the
successful completion of the System and Operational
Tests. Place the LifePulse in Standby mode before
connecting it to the patient. Failure to do so may
result in inappropriate pressures or volumes being
delivered to the patient, resulting in volutrauma.
9. Remove the test lung set-up (LifePort adapter, ET tube, and test
lung) and store it for future use.
The LifePulse system is now ready for patient use.
Note: The LifePulse should be plugged in to an electrical outlet at
all times to maintain battery charge. Be sure to remove the pinch
tubing from the pinch valve in the patient box when storing with a
circuit installed.
8
If a test fails, call the Bunnell Hotline
800-800-4358 (HFJV)
1-801-467-0800 (International)
Bunnell LifePulse HFV User Manual | 29
Page 30
Chapter 2: SETUP AND TESTS
Bunnell LifePulse HFV User Manual | 30
Page 31
Chapter 3: STARTING HFJV ON A PATIENT
Connecting a Patient to the LifePulse
1
The following steps describe how to connect a patient to the
LifePulse in preparation for start-up:
1. Secure the caps on the new LifePort adapter. Use a LifePort
adapter that is either the same size or one size bigger than the
endotracheal (ET) tube.
32
1
4
2. With the LifePulse still in Standby mode after the successful
completion of the System and Operational tests, replace the
standard ET tube adapter with the LifePort adapter.
3. Reconnect the patient to the CV by connecting the
conventional circuit to the 15-mm connector of the LifePort.
4. Once the patient has been stabilized on the CV—and with
the LifePulse still in Standby mode—connect the pressure
monitoring tube of the LifePort adapter to the Patient Box.
5. Connect the Patient Circuit to the Jet port on the side of the
LifePort adapter.
5
WARNING: Patient connections must only be made
in the Standby mode. Do not connect the LifePulse
Patient Circuit to the LifePort adapter on the patient’s
ET tube while the LifePulse is running. Failure to comply
risks high pressures and volumes being delivered to the
patient, which may result in volutrauma.
Bunnell LifePulse HFV User Manual | 31
Page 32
HIGH FREQUENCY VENTILATOR
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
II.4
2I.I
8.2
29.3
2.8
LifePulse
bunnell
INSPIRED INFANT CARE
WARNING: Do not connect any additional tubing or
pressure monitors between the pressure monitoring
tube of the LifePort adapter and the Patient Box.
Doing so will degrade the LifePulse’s ability to
measure airway pressures accurately and may result in
inappropriate pressures or volumes being delivered to
the patient.
WARNING: A properly trained medical professional
must be in attendance at all times when the LifePulse is
connected to a patient in order to monitor the patient
and detect alarms or other indications of a problem.
WARNING: There are no airway pressure alarms active
while the LifePulse is in the Standby mode.
2
3
1
Chapter 3: STARTING HFJV ON A PATIENT
HFJV Pressure Monitoring
The LifePulse, in its Standby mode, will begin to monitor the
pressures being delivered by the CV or other high frequency
ventilator once the pressure monitoring tube of the LifePort
adapter has been connected to the Patient Box. It will take about
90 seconds for the accurate measurements of the PIP, PEEP, and
MAP to be displayed in the Monitor section. The monitored values
will be updated every 10 seconds.
Note: The pressures displayed on the LifePulse may be different
from the pressures displayed on the CV--remember that the
CV displays pressures measured proximally while the LifePulse
displays approximations of distal pressures.
When HFJV is started on a patient, accurate pressure monitoring
allows the LifePulse to provide the patient with appropriate
ventilation and to protect the patient in the event of an alarm
condition after the Ready indicator illuminates. The LifePulse
monitors the following:
1. PIP
Peak Inspiratory Pressure, the average maximum airway
pressure measured for each HFJV breath.
5
4
SERVO - Increases may indicate
• Improved compliance and/or resistance
• Moisture interference in the LifePort
• Air leaks (pulmonary and/or tubing)
• Disconnected or kinked tubes
SERVO - Decreases may indicate
• Worsening compliance and/or resistance
• Patient needs suctioning
• Tension pneumothorax
• ET tube obstruction
• Right mainstem intubation
2. SERVO
The internal drive pressure of the LifePulse; indicates how
much gas flow must be produced to meet the set PIP, Rate, and
I-Time requested by the operator for a particular patient. Servo
tends to be higher for larger patients and lower for smaller
patients.
Servo is a good indicator of changes in the patient or ventilators.
See the information box to the left.
3. ∆P
PIP minus PEEP; ∆P is roughly proportional to tidal volume. As ∆P increases, tidal volume increases and vice versa.
4. MAP
Mean Airway Pressure, an average of pressures measured from
the total pressure waveform. It includes pressures produced by
HFJV, conventional ventilation, and spontaneous breathing.
5. PEEP
Positive End-Expiratory Pressure, the average minimum airway
pressure.
Bunnell LifePulse HFV User Manual | 32
Page 33
+
ALARMS
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
cm H2O
28
LifePulse
bunnell
INSPIRED INFANT CARE
28
PIP
cm H2O
READY
ALARMS
-
ALARMS
:
I
I:E RATIO
ALARMS
bunnell
INSPIRED INFANT CARE
ENTER
MONITOR
CONTROLS
+
-
cm H2O
420
RATEI-TIME
29
PIP
HUMIDIFIER
LifePulse
HIGH FREQUENCY VENTILATOR
READY
+
.0206.I
-
BPMSECONDS
231
WARNING: Do not initiate ventilation with kinks or
obstructions present in the Patient Circuit or LifePort
pressure monitoring tube. Doing so may result in
alarms or delays in ventilator support. Pressures may
be generated that are too high or too low for the
patient’s needs.
Chapter 3: STARTING HFJV ON A PATIENT
7 Steps to Begin HFJV
Initiation of HFJV involves seven steps:
1. Select a starting HFJV PIP that matches the measured CV PIP.
Input your chosen value by using the + or - button next to the
PIP display in the Controls section.
2. Use the default HFJV Rate of 420 bpm; alternatively, for larger
patients or patients prone to or exhibiting gas trapping, lower
the Rate in increments of 60 bpm (to 360, 300, 240 bpm) to
avoid hyperinflation.
Lower Rates create longer expiratory times, which may be
necessary to alleviate gas trapping.
3. After selecting the starting values, press ENTER.
As the LifePulse begins to operate, note that the monitored values
return to zero and new average values are displayed based on the
new Control settings.
The LifePulse’s microprocessor will begin increasing the Servo to
whatever value will produce the set PIP for the set I-Time at the
set Rate on a given patient.
Although it will typically take a short time for the actual PIP to
reach the set PIP, the displayed PIP will equilibrate slower because
of the averaging characteristics of the display (10-second average
updated every 2 seconds). Thus, it may take longer--typically
within a minute--for the monitored PIP to reach the set value.
WARNING: Do not leave the patient’s bedside while
the LifePulse water pump is running during initial
start-up or following a circuit change. A properly
trained person must observe the cartridge fill with
sterile water for inhalation, USP, to the second
water level sensing pin and the water pump stops
pumping.
The READY indicator will illuminate when the monitored PIP is
stable around the set PIP (+2.0 to -1.5 cm H2O) for 20 seconds,
indicating that the LifePulse is providing ventilation at the settings
requested and automatically set alarm conditions have been
established.
Also, the water pump will begin to pump water into the humidifier
cartridge. The pump will stop automatically once the water has
reached the correct level.
Bunnell LifePulse HFV User Manual | 33
Page 34
WARNING: Do not press the PUSH TO LOAD button
on the Patient Box while the LifePulse is running on
a patient. Doing so could result in the patient being
exposed to inappropriately high pressures or large
volume delivery (volutrauma).
Summary of Starting HFJV on a Patient
Chapter 3: STARTING HFJV ON A PATIENT
The LifePulse may pause every time the CV delivers a breath.
The interruptions in the HFJV pulses are caused by the delivery
of conventional breaths at PIP higher than the set PIP on the
LifePulse.
Except in cases of extremely poor lung compliance, it is usually
best to allow the HFJV pulses to continue uninterrupted by
lowering the CV PIP.
4. If desired, lower the CV PIP to just below the threshold of
interruptions; you may need to lower it even more when
treating infants with pulmonary air leaks.
1. Connect LifePulse to patient
2. Monitor CV or HFOV pressures
3. Choose starting LifePulse (HFJV) settings
4. Press ENTER to initiate HFJV
5. Decrease CV rate to 0-5 bpm
6. Adjust PEEP to match monitored MAP and stabilize SaO
2
7. Check blood gas in 30-45 minutes to evaluate settings
WARNING: There will be no LOSS OF PIP alarm for
the first 15 seconds after the ENTER button is pushed.
A properly trained person must observe the LifePulse
reach the set PIP and the Ready indicator illuminate
before leaving the patient’s bedside.
5. Lower the CV Rate to CPAP for pulmonary air leaks and chronic
lung disease, or 1 to 5 bpm for temporary alveolar recruitment.
Generally, the more pulmonary air leaks are a concern, the lower
you will set the background CV Rate, PIP, and I-Time on the CV.
PEEP is a better way to control oxygenation in patients with air
leaks. The more atelectasis is a concern, the higher you will set the
background CV Rate, PIP, PEEP, and I-Time.
Background CV Rates are used only as a temporary recruitment
maneuver to reverse atelectasis while PEEP is optimized. A CV
Rate of 5 bpm can be used temporarily for more aggressive
recruitment. Once optimal PEEP is established, the Rate can be
reduced to 0-2 bpm.
If atelectasis and air leaks are of equal concern after PEEP has
been optimized, start with the lowest number of CV breaths
possible and modest CV PIP and I-Time settings.
WARNING: The LifePulse will not detect extubation
of a patient during start-up (i.e., from the time the
ENTER button is pushed until the Ready indicator is
illuminated). During start-up the ventilator may appear
to be functioning normally despite extubation and
will generate no audible or visual alarms. A properly
trained individual must verify the patient is being
ventilated following start-up.
Bunnell LifePulse HFV User Manual | 34
Page 35
ALARMS
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
LifePulse
bunnell
INSPIRED INFANT CARE
+
-
UPPER
LIMIT
LOWER
LIMIT
II.3
READY
MAP
cm H2O
ALARMS
WARNING: Always use a conventional ventilator
in tandem with the LifePulse. The conventional
ventilator is essential for providing gas for the patient’s
spontaneous breathing, PEEP, and periodic sigh
breaths.
Chapter 3: STARTING HFJV ON A PATIENT
You may find that, once high-frequency jet ventilation is initiated,
the monitored PEEP is different than the set PEEP. This difference
is common because the LifePulse measures PEEP with different
hardware and software than the CV.
The CV PEEP is the primary control of MAP during HFJV. The PEEP
may need to be increased to match the MAP monitored from the
CV prior to start-up.
6. Adjust the CV PEEP to maintain appropriate SaO2. Remember
that average values are displayed, so give the LifePulse 20
seconds between adjustments to display the monitored
PEEP value.
It is important to reevaluate or optimize PEEP periodically based
on the patient’s FiO2 requirements and SaO2. PEEP should be
optimized to obtain the best SaO2 at the lowest FiO2 level.
Be aware of significant changes in Servo and Mean Airway
Pressure (MAP) that may occur when manipulating the CV settings.
For example, a pressure change of 2 cm H2O in the PEEP setting
will cause a change of about 2 cm H2O in the MAP.
7. If changing PEEP produces a MAP or Servo alarm, press ENTER
to incorporate the change and establish new limits.
Alternatively, the MAP or Servo alarm limits may be adjusted
manually.
7
Wait 30-45 minutes to get a blood gas and evaluate HFJV
effectiveness.
Bunnell LifePulse HFV User Manual | 35
Page 36
Chapter 3: STARTING HFJV ON A PATIENT
360
ALARMS
+
-
+
-
+
-
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
LifePulse
bunnell
INSPIRED INFANT CARE
READY
ALARMS
PAUSE
TEMPERATURE
GASWATER
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
ALARMS
+
+
+
-
-
-
+
-
+
-
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
SERVO
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
cm H2O
I2.6
4.7
88.8
LifePulse
bunnell
INSPIRED INFANT CARE
READY
ALARMS
Adjusting Ventilator Settings
To adjust settings in the Controls section,
1. Press the + or - button next to the setting you want to change
(PIP, Rate, or I-Time).
2. Continue to press the + or - button until the desired value is
displayed.
In Ready mode, the display and the ENTER button will continue to
flash until ENTER is pressed. The displays will not flash in Standby
mode.
213
READY INDICATOR
The Ready indicator will turn off when the ENTER
button is pressed. As soon as monitored values have
been stable for 20 seconds, new alarm limits will be set
and the Ready indicator will illuminate indicating the
alarms are again active.
Always wait until the Ready indicator is ON before
leaving the patient’s bedside.
1
32
3. Press ENTER to activate the change.
Adjusting Ventilator Alarms
To adjust Servo and MAP alarm settings while in Ready mode (no
alarm limits are set until Ready indicator is lit):
1. Press the desired UPPER or LOWER LIMIT button next to the
MAP or Servo display.
The current set value for the selected limit will be displayed.
2. Press the + or - button to change the limit to the desired value.
3. Press the UPPER or LOWER LIMIT button to activate the
change and return the display to the monitored value.
If the button is not pressed, the change will still be activated and
the display will return to the monitored value automatically after
10 seconds.
WARNING: Do not use pens, pencils, fingernails,
or other pointed objects to push the buttons on the
LifePulse front panel. Doing so may result in damage
to the buttons and may cause them to fail.
Bunnell LifePulse HFV User Manual | 36
Page 37
Chapter 4: PATIENT MANAGEMENT
CLINICAL OBJECTIVES
Oxygenation
Concerns
Ventilation &
High Pressure
Concerns
Conventional
Ventilator
LifePulse
Ventilator
Objectives
Managing patients on high-frequency jet ventilation is similar to
managing patients on a CV.
The main difference with the LifePulse is you will be using less
pressure and much smaller tidal volumes to meet your clinical
objectives.
Oxygenation - The CV settings will be adjusted most often when
oxygenation of the patient is your primary concern.
Ventilation - The LifePulse settings will be adjusted most often
when ventilation (CO2 removal) and/or the consequences of using
high airway pressures (e.g., pulmonary air leaks) are your greatest
concern.
The following pages will discuss patient management in more
detail.
Bunnell LifePulse HFV User Manual | 37
Page 38
Chapter 4: PATIENT MANAGEMENT
Oxygenation
The main choices for improving oxygenation require increasing
Mean Airway Pressure (MAP) by elevating one or more of the
following:
• CV PEEP
• CV Rate
• CV PIP
• CV I-Time
Exception: If the patient on conventional ventilation has grossly
overexpanded lungs, the lungs will need to deflate considerably
before any improvements in oxygenation will result. Refer to the
next section of this chapter for more information about addressing
overexpanded lungs.
Here are the choices for improving oxygenation due to atelectasis:
1. Increase CV PEEP
2. Increase CV Rate
WARNING: The conventional ventilator rate should not
exceed 10 breaths per minute when running in tandem
with the LifePulse ventilator. Doing so could result in
excessive Mean Airway Pressure or minute volume
being delivered to the patient. The recommended
conventional ventilator rate is 0 - 5 breaths per minute.
1. Increase CV PEEP to stabilize alveoli.
Changes to PEEP are made with the CV because the LifePulse has
no PEEP control. However, the PEEP adjustment will be displayed
on the LifePulse in the Monitor section.
Always optimize PEEP before increasing CV Rate, PIP, or I-Time.
Adequate PEEP levels are essential for avoiding derecruitment
between conventional breaths. CV Rate, PIP, and I-Time can be
reduced or eliminated once optimal PEEP is achieved.
2. Increase the background CV Rate.
Avoid providing more than 5 bpm with the background CV Rate.
If you feel you need more CV breaths to oxygenate, it may be an
indication that PEEP is too low. CV breaths are to be considered a
temporary recruitment maneuver while you recruit lung volume or
optimize PEEP.
Bunnell LifePulse HFV User Manual | 38
Page 39
Chapter 4: PATIENT MANAGEMENT
3. Increase CV PIP
4. Increase CV I-Time
WARNING: Always evaluate inadvertent PEEP
(LifePulse monitored PEEP greater than CV set PEEP).
The patient’s lung volume should be assessed when
inadvertent PEEP is present. Inadvertent PEEP reduces
delta pressure, which may result in under-ventilation
(hypercarbia).
WARNING: A significant discrepancy (>4 cm H2O)
between the delivered PEEP and the monitored PEEP
may indicate a potential problem and should always be
evaluated by a qualified trained individual.
3. Increase CV PIP delivered with the background conventional
breaths to achieve a moderate chest rise.
The CV PIP should not be raised if it is already at a level adequate
to reach the critical opening pressure of atelectactic alveoli,
as determined by breath sounds, chest rise, X-ray, and other
resources.
Also, remember to keep the CV PIP below the HFJV PIP to avoid
interrupting the high-frequency breaths.
4. Increase CV I-Time, in combination with adequate levels of
PEEP and CV PIP, to reverse atelectasis.
Consider carefully the combined effect of PIP and I-Time increases.
Increasing the CV I-Time when the CV PIP is set at a high level
increases the risk of lung injury.
Raising HFJV PIP and Rate would be secondary considerations. If
you increase the HFJV Rate, be sure to watch the HFV-monitored
PEEP level. Inadvertent PEEP may develop as the I:E ratio is
shortened.
Raising HFJV I-Time has occasionally been shown to be effective in
improving oxygenation.
Bunnell LifePulse HFV User Manual | 39
Page 40
EXCEPTION! Overexpanded Lungs
Chapter 4: PATIENT MANAGEMENT
Oxygenating Overexpanded Lungs
There is one major exception to the standard oxygenation
strategies. This exception arises when the patient on conventional
ventilation has grossly overexpanded lungs.
If you observe overexpansion on X-ray, you will need to allow
the lungs to deflate considerably before any improvements in
oxygenation will result.
To accomplish this deflation, make sure that the HFJV Rate is set
low enough to allow adequate exhalation time. Lowering the
HFJV Rate increases expiratory time and reduces the risk of gas
trapping. Also, set the CV Rate near zero once you have started
the LifePulse.
In most cases, DO NOT DECREASE PEEP. Overexpanded lungs
are usually a result of gas trapping, not excessive PEEP. Decreasing
CV support (Rate, PIP, and I-Time) is usually a more effective
strategy. PEEP must be maintained, or even increased, when the
CV Rate is very low to support MAP and maintain oxygenation.
However, beware that even if the patient initially responds well
to this strategy, poor oxygenation may result sometime later due
to atelectasis if the MAP is not adequate. You may then need to
provide 3-5 CV breaths as a temporary recruitment strategy to
reverse atelectasis while PEEP is optimized.
Bunnell LifePulse HFV User Manual | 40
Page 41
Chapter 4: PATIENT MANAGEMENT
No
No
Yes
Yes
No
Yes
Yes
No
No
Yes
Yes
Yes
Yes
To raise MAP & PaO2 try the
5. increase FiO
too Low
To decrease gas trapping & raise
5. increase FiO
too High?
too High?
Pa CO2
too Low?
too Low?
PaO2 too
High?
PaO2 too
High?
Decrease FiO2 until < 0.40
then decrease CV PEEP
No
PaO2 is
Oxygenation Flow Chart
Hypoxemia
Underinflation
or Atelectasis?
following in order:
1. increase CV PEEP
2. increase CV Rate (3 -5 bpm)
3. increase CV PIP
4. increase CV I-time
PaC O2
Go to Ventilation Flow Chart
Overinflation
or P.I.E./Air leak?
PaO2 try the following in order:
1. decrease CV Rate
2. decrease HFJV Rate *
(60 bpm at a time)
3. decrease HFJV PIP **
4. decrease CV PEEP ***
PaCO2
Pa CO2
* Decreasing HFJV Rate decreases minute ventilation. It may also lower PaCO
** Decreasing HFJV PIP decreases
*** Decreasing CV PEEP increases
Maintain Current Settings
by increasing exhalation time.
2
pressure (amplitude) and minute ventilation ; PaCO2may increase..
pressure (amplitude) and decreases MAP ; PaCO2 and PaO2 may decrease.
Bunnell LifePulse HFV User Manual | 41
Page 42
Chapter 4: PATIENT MANAGEMENT
Ventilation
Managing the patient’s PaCO2 is accomplished by adjusting the
LifePulse’s settings.
Studies have shown that ventilation (CO2 elimination) during
high-frequency jet ventilation is proportional to the tidal volume
squared (VT2). Tidal volume on the LifePulse is roughly proportional
to delta P (∆P), the arithmetic difference between PIP and PEEP.
Thus, small changes in PIP or PEEP can produce significant
changes in a patient’s PaCO2.
The main choices for improving ventilation require increasing
minute ventilation by changing one or more of the following:
• HFJV PIP
• HFJV I-Time
• PEEP
• HFJV Rate
1. Increase HFJV PIP
2. Increase HFJV I-Time
1. Increase HFJV PIP by 1 to 2 cm H2O at a time if the patient’s
PaO2 is acceptable but PaCO2 is too high.
If the patient’s PaO2 is low and PaCO2 is high, increasing PIP may
correct both problems at once since raising the HFJV PIP increases
the MAP.
If PEEP has been increased to improve oxygenation, you may need
to increase the HFJV PIP by an equal amount in order to keep ∆P
the same and maintain tidal volume and adequate ventilation.
2. Increase HFJV I-Time by .004 to .006 seconds at a time if the
patient’s pathophysiology involves longer inspiratory time
constants.
You may need to lower the HFJV Rate in increments of 60 bpm to
maintain an adequate I:E ratio (1:5) for effective exhalation.
Bunnell LifePulse HFV User Manual | 42
Page 43
Chapter 4: PATIENT MANAGEMENT
3. Decrease PEEP
4. Increase HFJV Rate
Raising PaCO
2
3. Lower PEEP to increase ∆P if the primary concern is cardiac
compromise and oxygenation is acceptable.
Be careful not to compromise oxygenation when you lower PEEP.
4. Increase the HFJV Rate by 60 bpm at a time to improve
ventilation without further increasing HFJV PIP. This strategy is
only effective if the lungs are not overinflated.
Increasing HFJV Rate is much less effective than increasing delta
P (∆P) for reducing PaCO2.
Raising the CV Rate or PIP seldom improves ventilation, but it
might be helpful in some cases as long as HFJV is not interrupted.
Again, this strategy is only effective when the lungs are not
overinflated.
The main choices for raising PaCO2 involve doing the opposite of
the above suggestions for lowering PaCO2.
Decreasing the HFJV PIP is the most effective way to increase
PaCO2. However, the concomitant drop in MAP may cause PaO2 to
fall. Be prepared to raise PEEP in such cases to maintain adequate
MAP.
Reducing the HFJV Rate will also increase PaCO2, but it may have
the opposite effect if hyperinflation is present; it may lower PaCO2
by lengthening the exhalation time.
Do not hesitate to use the minimum HFJV Rate of 240 bpm when
indicated for hyperinflation. The LifePulse can provide adequate
ventilation over its entire range of HFJV Rates. With the HFJV
I-Time at .020 seconds, the HFJV Rate of 240 bpm produces an I:E
Ratio of 1:12.
If monitored MAP falls when the HFJV Rate is lowered, you may
need to raise PEEP to maintain adequate oxygenation.
Bunnell LifePulse HFV User Manual | 43
Page 44
Chapter 4: PATIENT MANAGEMENT
Yes
Yes
Yes
Yes
Yes
Yes
To raise P aCO2 try the
following in order:
1. decrease HFJV PIP
2. decrease HFJV Rate
3. decrease HFJV I-time
4. increase CV PEEP
PaCO2
Hypocarbia
PaCO2 too
Low?
Hypercarbia
PaCO2 too
High?
To lower PaCO2 try the
following in order:
1. increase HFJV PIP
2. increase HFJV I-time
3. increase HFJV Rate*
4. decrease CV PEEP **
PaCO2
too High?
PaCO2
too Low?
PaO2 too
Low?
PaO2 too
Low?
PaO2 too
High?
PaO2 too
High?
Go to Oxygenation Flow Chart
Decrease FiO2 until < 0.40
then decrease CV PEEP
Maintain Current Settings
Yes
Yes
No
No
No
No
No
No
* Increasing HFJV Rate increases minute ventilation. However, if lungs are hyperinflated, decreasing
HFJV Rate can lower PaCO
2
by increasing exhalation time.
** Decreasing CV PEEP increases Δ pressure (amplitude) and lowers PaCO
2
, but it will also lower MAP, which
may lower PaO
2
.
Ventilation Flow Chart
Bunnell LifePulse HFV User Manual | 44
Page 45
Chapter 4: PATIENT MANAGEMENT
Suctioning the Patient
High-frequency jet ventilation may mobilize and help remove
secretions. Be prepared to suction the patient’s airway soon after
starting the LifePulse on a patient.
Suctioning may need to be performed more frequently in the first
4 to 6 hours after starting HFJV. Suctioning frequency may then
decrease.
You can suction using a closed suction catheter during HFJV.
WARNING: Place the LifePulse ventilator in Standby
mode prior to suctioning to avoid fluctuations in the
monitored pressures that may result in alarm conditions
or possible injury to the patient.
WARNING: The LifePulse must be in Ready mode prior
to suctioning if suctioning while the LifePulse is running.
Failure to do so may result in inappropriate pressure or
volumes being delivered to the patient.
1. Suction according to your normal protocol.
2. If necessary, provide a few manual CV breaths after suctioning
to reestablish lost lung volume and improve oxygenation.
3. Make sure the READY light is illuminated before leaving the
bedside.
4. Call the Bunnell Hotline at 800-800-4358 if you have questions
about suctioning.
Bunnell LifePulse HFV User Manual | 45
Page 46
Wean SLOWLY!
Lower HFJV PIP
Chapter 4: PATIENT MANAGEMENT
Weaning the Patient
Weaning from the LifePulse will need to be performed as the
patient improves. The goal in most cases will be to wean the
patient off the LifePulse to non-invasive ventilation (NIV).
General guidelines for weaning:
• Decrease minute ventilation slowly by lowering PIP on the
LifePulse and the CV.
Lower FiO
2
Maintain Optimal MAP
Adjust PEEP if Necessary
Extubate to NIV
• Lower FiO2 as often as possible to the lowest value that
still provides adequate oxygenation to the patient. Remember
to adjust both blenders so the settings remain equal on the
CV and the LifePulse.
• Adjust PEEP as necessary to maintain a MAP sufficient to
achieve adequate oxygenation. Don’t wean PEEP until FiO2
<0.40.
• Continue decreasing PIP as blood gases allow, weaning slowly
(1-2 cm H2O of PIP), unless hyperventilation is occurring, in
which case PIP should be weaned faster.
• Wean HFJV Rate as tolerated to encourage spontaneous
breathing by the patient.
• When the patient is stable and receiving “extubatable”
settings as determined by a physician, extubate to an NIV
level near the level of the last recorded MAP.
Bunnell LifePulse HFV User Manual | 46
Page 47
Chapter 5: VENTILATOR ALARMS
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
ALARMS
READY
+
+
-
-
AUDIO
PAUSED
HIGH FREQUENCY VENTILATOR
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
LifePulse
bunnell
INSPIRED INFANT CARE
1
2
READY INDICATOR ON
1. Monitored PIP +2.0 to -1.5 cm H2O of set PIP for
20 Seconds.
2. Servo alarm limits vary according to size of patient.
3. MAP alarms are set
3
Alarms Overview
The LifePulse’s alarm system alerts the operator, both audibly
and visually, to changes in the ventilator or the patient. The alarm
statements are not visible until they are lit.
The Alarms area has three key features:
1. Upper and lower alarm limits for Servo and Mean Airway
Pressure (MAP)
2. Alarm indicators for various potentially hazardous conditions
3. AUDIO PAUSED button
Both the upper and lower alarm limits for Servo and MAP are set
automatically and can be adjusted manually. They are set when the
Ready indicator illuminates, which occurs after the monitored PIP
has stabilized within a 3.5 cm H2O (+2.0 to -1.5 cm H2O) window
of the set PIP for at least 20 seconds. Servo increases to bring the
monitored PIP up to the set PIP after ENTER has been pressed.
Servo increases after the ENTER button has been pressed to bring
the monitored PIP up to the set PIP. Servo limits vary according to
the size of the patient; wider limits are set for larger patients and
tighter limits for smaller patients.
WARNING: Troubleshoot alarms only in accordance
with the guidelines in the User Manual. Troubleshooting
must only be performed by properly trained individuals
who have a thorough understanding of the LifePulse
ventilator and its operation. Always attempt to correct
the cause of the alarm before pressing the ENTER
button.
WARNING: Place the LifePulse in Standby mode if
there is any concern about the ventilator functionality
or safety. Support the patient using the tandem
conventional ventilator or other available ventilation
options.
WARNING: Alarms from different medical devices can
sound similar. Identify and respond to LifePulse alarms
based on the alarm priority lamp, the alarm indicator
message, and the audible alarm tone and rhythms.
Bunnell LifePulse HFV User Manual | 47
Page 48
Chapter 5: VENTILATOR ALARMS
High Priority Alarms
• LOSS OF PIP
• HIGH PIP
• VENT INOP w/ Codes (01-10)
• HIGH LEVEL
• CHECK CIRCUIT LEVEL
• BATTERY DEPLETED
• Power Up (no alarm indicator)
• VENT INOP (no code) - Vent stopped or
reset via Watchdog circuit
• Vent reset via software or backup vent alarm
detected (no alarm indicator)
Medium Priority Alarms
• LOW GAS PRESSURE
• CANNOT MEET PIP
• CHECK VENT (w/ Codes) 111, 222, 333)
• MAP UPPER or LOWER LIMIT
• SERVO UPPER or LOWER LIMIT
• CHECK CIRCUIT TEMP
• LOW LEVEL
• LOW TEMP - WATER
• HIGH TEMP - WATER
• LOW TEMP - GAS
• HIGH TEMP - GAS
• BATTERY LOW (Fuel Gauge Blinks Red)
• CHECK VENT (No Code) - Humidifier
stopped or reset via Watchdog circuit, failure
to detect the purge, or valve status feedback
errors
• Backup humidifier alarm detected (no alarm
indicator)
Alarm Displays and Sounds
Ventilator alarms will be identified by an illuminated indicator in
the Alarms section, a yellow or red alarm priority lamp on top of
the LifePulse, and a specific audible alarm.
Each alarm indicator will be accompanied by a specific display
and sound:
• High Priority Alarms - 2 bursts of 5 beeps every 5 seconds and
a flashing red light
• Medium Priority Alarms - 3 longer beeps every 5 seconds and
a flashing yellow light
• Low Priority Alarms - 1 beep every 18 seconds and a solid
yellow light
The Servo and MAP alarms are displayed as a blinking UPPER or
LOWER LIMIT button, depending upon which alarm occurs.
Call the Bunnell Hotline at 800-800-4358 (Int’l: 1-801-467-0800) if
an alarm occurs with no alarm indicator in the Alarms section.
Note: There are six alarms that can be reset by pressing the
AUDIO PAUSED button:
• Power Up
• Check Vent - No Codes (some exceptions)
• Vent Inop - No Codes
• Battery - Running on Battery
• Battery missing, malfunction, or cannot take a charge
(no alarm indicator)
• Power Down
Note: See Alarms Terminology Section for more information
regarding alarm displays and descriptions.
Low Priority Alarms
• STANDBY
• PAUSE
• BATTERY (running on battery)
• Battery missing, malfunction, or cannot
accept a charge (no alarm indicator)
Bunnell LifePulse HFV User Manual | 48
Page 49
PAUSE
TEMPERATURE
GASWATER
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
ALARMS
+
+
+
-
-
-
+
-
+
-
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
SERVO
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
cm H2O
I2.6
4.7
88.8
LifePulse
bunnell
INSPIRED INFANT CARE
READY
ALARMS
WARNING: Do not ignore audible and visual alarms.
Alarms indicate changing conditions and should never
go unheeded.
1
32
Chapter 5: VENTILATOR ALARMS
Adjusting Alarm Limits
The upper and lower limits of the MAP and Servo alarms are set
automatically but can be adjusted manually.
Press the UPPER or LOWER LIMIT buttons any time after the
Ready indicator illuminates to see where the limits have been set.
If the Ready indicator is not lit, the limits have not been set and
the Servo and MAP displays will not change when you press one
of the limit buttons.
Do the following to adjust Servo and MAP alarm limits once the
Ready indicator is lit:
1. Press the desired UPPER or LOWER LIMIT button next to the
MAP or Servo display; the current set value for the selected
limit will be displayed.
2. Press the + or - button to change the limit to the desired value.
WARNING: Do not manually adjust alarm limits
around MAP and Servo to the point that they
become irrelevant (i.e., are effectively off). Doing so
will negate critical alarms and expose the patient
to potentially unsafe conditions that may result in
injury.
WARNING: Do not manually adjust alarm volume
to the point that the audible alarm cannot be
recognized (is effectively off). Doing so will limit
the LifePulse’s ability to alert clinicians to alarm
conditions and expose the patient to potentially
unsafe conditions that may result in injury.
3. Press the selected limit button again to activate the change
and return the display to the monitored value.
If the button is not pressed again, the change will still be activated
and the display will return to the monitored value after 10
seconds.
Bunnell LifePulse HFV User Manual | 49
Page 50
Chapter 5: VENTILATOR ALARMS
ALARMS
+
-
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
SERVO
UPPER
LIMIT
LOWER
LIMIT
4.2
LifePulse
bunnell
INSPIRED INFANT CARE
READY
ALARMS
SERVO
The Servo alarms alert the user to changes in the internal drive
pressure or flow rate required to deliver breaths at the set PIP,
Rate, and I-Time.
The UPPER LIMIT button will blink when a high Servo condition
is reached; the LOWER LIMIT button will blink when a low Servo
condition is met.
The alarm limits are set automatically as follows:
Servo Alarm Set
<1 ±0.2
1-5 ±20%
WARNING: Always troubleshoot a change in Servo
that is greater than 1.0 psi (6.89 kPa) from the
previously established baseline, especially if the
LifePulse control settings have not been changed.
Servo changes this large or larger usually represent
a mechanical problem and may pose a risk of
volutrauma.
WARNING: Place the LifePulse in Standby mode
before unkinking any portion of the Patient Circuit.
This action prevents the patient from being exposed to
inappropriately high pressure or large volume delivery
(volutrauma).
>5 ±1.0
High Servo
Checks:
1. Examine the patient for spontaneous breathing, crying,
coughing, hiccups, seizures, bronchospasm, or the need for
suctioning. Calm, medicate, sedate, or suction the patient.
2. Inspect the LifePort adapter and connections for any leaks or
cracks. If the adapter’s integrity is in question, replace it with a
new one.
3. Inspect the Patient Circuit for leaks, poor connections, or
occlusions. Correct these conditions if present.
4. Inspect the Patient Circuit between the humidifier and Patient
Box for signs of over-humidification (droplets of water). If
necessary, lower the Water temperature setting to reduce
condensation (see Humidification Chapter).
Note: Keep Gas temperature at 40°C.
5. Evaluate the Patient Box for abnormal functioning (odd sounds
or visible problems with pinch valve operation). Change
the Patient Box if necessary.
6. Examine the patient for re-expansion of the lung due to
evacuation of an air leak, such as pneumothorax or bronchopleural fistula, or for improved lung volume following
recruitment.
Bunnell LifePulse HFV User Manual | 50
Page 51
Chapter 5: VENTILATOR ALARMS
ALARMS
+
-
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
SERVO
UPPER
LIMIT
LOWER
LIMIT
I.8
LifePulse
bunnell
INSPIRED INFANT CARE
READY
ALARMS
Low Servo
Checks:
1. Inspect the endotracheal tube for improper positioning,
plugging, or occlusion. Correct these conditions if present.
2. Examine the patient for signs of airway obstruction.
3. Inspect the exhalation limb of the conventional breathing
circuit for kinks.
4. Examine the patient for a tension pneumothorax or atelectasis.
5. Evaluate the Patient Box for abnormal functioning (odd sounds
or visible problems with pinch valve operation). Change
the Patient Box if necessary.
Bunnell LifePulse HFV User Manual | 51
Page 52
Chapter 5: VENTILATOR ALARMS
ALARMS
+
-
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
MAP
UPPER
LIMIT
LOWER
LIMIT
cm H2O
9.4
LifePulse
bunnell
INSPIRED INFANT CARE
READY
Mean Airway Pressure
The MAP alarms alert the user to changes in the patient, CV
settings, or the CV or HFV circuit sufficient to cause an increase or
decrease in measured MAP.
An LED in the UPPER LIMIT button will blink when a high MAP
condition is met; an LED in the LOWER LIMIT button will blink
when a low MAP condition is met.
The upper alarm limit is set automatically by adding 1.5 cm
H2O to the MAP measured at the moment the Ready indicator
illuminates.
High MAP
Checks:
1. Inspect the CV for changes in settings (e.g., PEEP) or
performance.
2. Inspect the exhalation limb of the conventional breathing
circuit and remove any kinks or water.
3. Inspect the pressure monitoring tube of the LifePort adapter
for excess condensation or plugging. If necessary, reduce
the Water temperature setting to reduce condensation.
Note: Keep Gas temperature at 40°C.
4. Inspect the endotracheal tube or airway for obstruction.
5. Check the Patient Box pinch valve to verify it is closing
completely.
Bunnell LifePulse HFV User Manual | 52
Page 53
Chapter 5: VENTILATOR ALARMS
ALARMS
+
-
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
MAP
UPPER
LIMIT
LOWER
LIMIT
cm H2O
4.8
LifePulse
bunnell
INSPIRED INFANT CARE
READYREADY
ALARMS
Mean Airway Pressure
The lower alarm limit is set automatically by subtracting 1.5 cm
H2O from the MAP measured at the moment the Ready indicator
illuminates.
Low MAP
Checks:
1. Inspect the CV for changes in settings (e.g., PEEP) or
performance.
2. Inspect the pressure monitoring tube of the LifePort adapter
for excess condensation or plugging. If necessary, reduce the
Water temperature setting to reduce condensation.
Note: Keep Gas temperature at 40°C.
3. Inspect the endotracheal tube for proper positioning.
4. Inspect the Patient Circuit and humidifier cartridge
for leaks, poor connections, or kinks.
Bunnell LifePulse HFV User Manual | 53
Page 54
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
-
PAUSE
TEMPERATURE
GASWATER
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
88.8
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
LifePulse
bunnell
INSPIRED INFANT CARE
LOSS OF PIP
+
+
-
-
ALARMS
ALARMS
WARNING: Place the LifePulse in Standby mode prior
to troubleshooting if the Patient Box pinch valve stops
operating while on a patient. This action prevents the
patient from being exposed to inappropriately large
volume delivery and possible volutrauma.
WARNING: There will be no LOSS OF PIP alarm
for the first 15 seconds after the ENTER button is
pushed. A properly trained person must observe the
LifePulse reach the set PIP and the Ready indicator
illuminate before leaving the patient’s bedside.
Chapter 5: VENTILATOR ALARMS
LOSS OF PIP
The LOSS OF PIP alarm is activated by one or more of the
following conditions:
• The monitored PIP has dropped below 25% of the set PIP
• The monitored PIP is less than 3 cm H20
• Monitored value for ∆P is ≤ 2 cm H20
Checks:
1. Examine the patient for spontaneous breathing, crying,
coughing, hiccups, seizures, bronchospasm, or the need for
suctioning. Calm, medicate, sedate, or suction the patient.
2. Over-humidification of the Patient Circuit may be causing
water droplets to interfere with pressure monitoring. Check the
LifePort adapter’s orientation to make sure the pressure
monitoring tube is pointed upward. Adjust the humidifier
temperature settings (i.e., lower the Water temperature in
1-degree increments) to minimize excess condensation.
Note: Keep Gas temperature set at 40°C; lower Water
temperature as needed to control humidity and condensation.
WARNING: Press the ENTER button to resume
ventilation if a LOSS OF PIP alarm has occurred
with the Ready indicator off. The Patient Box will be
cycling but no gas (breaths) will be delivered to the
patient until the LOSS OF PIP alarm is cleared by
pressing the ENTER button.
WARNING: Do not press the ENTER button during
a LOSS OF PIP alarm with the Ready indicator
illuminated if the patient is clinically appropriate
and the Servo has locked at or near an established
operating value. Doing so will temporarily silence
the audible alarm and unlock the LifePulse Servo.
Always troubleshoot the potential causes of an alarm
following the guidelines in the User’s Manual prior to
pressing the ENTER button.
WARNING: Place the LifePulse in Standby mode
prior to troubleshooting if the Patient Box pinch
valve stops operating while on a patient. This
action prevents the patient from being exposed to
inappropriately large volume delivery and possible
volutrauma.
WARNING: Place the LifePulse in Standby mode
before unkinking any portion of the Patient Circuit.
This action prevents the patient from being exposed
to inappropriately high pressure or large volume
delivery (volutrauma).
3. Inspect the Patient Circuit and humidifier cartridge for leaks or
poor connections.
4. Check the pinch valve for possible malfunctions resulting in low
PIP. If necessary, replace the Patient Box while the LifePulse is
in the Standby mode.
Note: If a LOSS OF PIP alarm occurs in the non-Ready condition,
the servo control valves are turned off to decrease Servo and
protect the patient. The alarm activates if the PIP is <3 cm H2O or
∆P is ≤ 2 cm H2O 15 seconds after ENTER is pressed. No other
alarms are active in the non-Ready condition. Pressing ENTER will
clear the alarm and allow the LifePulse another attempt to reach
the set PIP.
Bunnell LifePulse HFV User Manual | 54
Page 55
MONITOR
88.8
cm H2O
cm H2O
+
-
cm H2O
SERVO
88.8
Δ
P
88.8
MAP
420
RATEI-TIME
88.8
TEMPERATURE
88.8
PIP
-8.8
PEEP
CONTROLS
20.
PIP
HUMIDIFIER
PAUSE
LifePulse
HIGH FREQUENCY VENTILATOR
+
+
-
-
UPPER
LIMIT
SYSTEM
cm H2O
cm H2O
BPMSECONDS
TEST
+
+
-
-
UPPER
LIMIT
+
.8888.8
-
+
-
GASWATER
LOWER
LIMIT
LOWER
LIMIT
ALARMS
LOSS OF PIP
CANNOT MEET PIP
CANNOT MEET PIP
HIGH PIP
BATTERY DEPLETED
LOW GAS PRESS
CHECK VENT
+
:
8
-
I:E RATIO
ALARMS
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
bunnell
INSPIRED INFANT CARE
Chapter 5: VENTILATOR ALARMS
CANNOT MEET PIP
The CANNOT MEET PIP alarm occurs when one of the following
conditions is met:
• The monitored PIP has not stabilized within +2.0 and -1.5 cm
H20 of the set PIP for 20 seconds within three minutes of
pressing ENTER.
• The monitored PIP has not stabilized within +2.0 and –1.5 cm
H2O of the set PIP for 20 seconds before the Servo has risen
to its maximum output of 20.
Checks:
1. Examine the patient for spontaneous breathing, crying,
coughing, hiccups, seizures, bronchospasm, or the need for
suctioning. Calm, medicate, sedate, or suction the patient.
2. Make sure the Servo hasn’t reached the maximum output
of 20 before reaching the set PIP. The patient may be too large
to be ventilated by the LifePulse at the current settings.
3. Over-humidification of the Patient Circuit may be causing
water droplets to interfere with pressure monitoring. Check
the LifePort adapter’s orientation to make sure the pressure
monitoring tube is pointed upward. Adjust humidifier
temperature settings (i.e., lower the Water temperature in
1-degree increments) to minimize excess condensation.
Note: Keep Gas temperature set at 40°C; lower Water
temperature as needed to control humidity and condensation.
4. Inspect the patient circuit and humidifier cartridge
for leaks or poor connections.
5. Check the pinch valve for possible malfunction resulting in
fluctuating pressures. Replace the Patient Box while the
LifePulse is in the Standby mode.
Bunnell LifePulse HFV User Manual | 55
Page 56
MONITOR
88.8
cm H2O
cm H2O
+
-
cm H2O
SERVO
88.8
Δ
P
88.8
MAP
420
RATEI-TIME
88.8
TEMPERATURE
88.8
PIP
-8.8
PEEP
CONTROLS
20.
PIP
HUMIDIFIER
PAUSE
LifePulse
HIGH FREQUENCY VENTILATOR
+
-
cm H2O
+
-
cm H2O
+
.8888.8
-
BPMSECONDS
+
-
GASWATER
ALARMS
LOSS OF PIP
CANNOT MEET PIP
HIGH PIP
HIGH PIP
BATTERY DEPLETED
LOW GAS PRESS
CHECK VENT
+
:
8
-
I:E RATIO
ALARMS
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
bunnell
INSPIRED INFANT CARE
WARNING: Patient connections must only be
made in the Standby mode. Do not connect the
LifePulse Patient Circuit to the LifePort adapter on
the patient’s ET tube while the LifePulse is running.
Failure to comply risks high pressures and volumes
being delivered to the patient, which may result in
volutrauma.
Chapter 5: VENTILATOR ALARMS
HIGH PIP
The HIGH PIP alarm will occur when any of the following
conditions occur:
• The monitored airway pressure has exceeded the set PIP by
5 cm H20 continuously for 1 second
• The monitored PIP for all breaths delivered in the last 30
seconds has exceeded the set PIP by 10 cm H20
• The monitored PIP for all breathes delivered in the last 0.75
seconds has exceeded the set PIP by 30 cm H20
• The monitored PIP is >65 cm H20
Checks:
1. Is excessive pressure being delivered by the CV due to failure
or an occluded exhalation limb of CV circuit?
2. Inspect the endotracheal tube for improper positioning,
kinking, or occlusion.
3. Inspect the pressure monitoring tube of the LifePort adapter
for excess condensation or kinking. Check the adapter’s
orientation to make sure the pressure monitoring tube
is pointed upward. Replace the LifePort adapter if necessary.
4. Check the pinch valve for possible malfunction resulting in high
PIP. If necessary, replace the Patient Box while the LifePulse is
in the Standby mode.
5. If the alarm persists, it could be a stuck Servo control valve.
If necessary, change out the LifePulse ventilator.
Bunnell LifePulse HFV User Manual | 56
Page 57
02
VENT INOP
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
I:E RATIO
LifePulse
bunnell
INSPIRED INFANT CARE
ALARMS
VENT INOP codes will appear in the I:E Ratio
display in the Controls section.
Chapter 5: VENTILATOR ALARMS
VENT INOP During Test
During System Test: Codes 02, 03, 04
If any of these codes appear during the System Test, record the
code number and call the Bunnell Hotline. Turn the Ventilator
Power switch off and back on to clear the failed test so you can
retry the System Test again after troubleshooting with a Bunnell
clinical specialist.
Code 02 is related to the purge system. Causes can be
• a lack of purge gas supply from the LifePulse
• a failure of the purge valve or pressure transducer in the
Patient Box
• a Patient Circuit being disconnected from the LifePort
adapter on the test lung
• a Patient Box being disconnected from the front panel of the
LifePulse
02 – Test of purge valve/airway transducer
03 – Test of PWM valve/servo transducer
04 – Test of watchdog circuit
Code 03 is related to the servo system. The 03 code will appear
if
• any of the valves do not open or are sluggish during the
System Test
• the servo pressure transducer does not recognize (respond
to) the opening of the servo control valves
• the dump valve is open
• the green gas inlet tube of the Patient Circuit is not
connected to the Gas Out port
Sometimes running the LifePulse on a test lung for 5 minutes
will correct this alarm condition by cycling the servo control
valves.
Code 04 is related to watchdog circuit failure. The watchdog
circuit checks the microprocessor and software in the LifePulse.
Bunnell LifePulse HFV User Manual | 57
Page 58
MONITOR
88.8
-8.8
cm H2O
cm H2O
cm H2O
+
-
SERVO
88.8
Δ
P
88.8
MAP
420
RATEI-TIME
88.8
TEMPERATURE
88.8
PIP
PEEP
CONTROLS
20.
PIP
HUMIDIFIER
PAUSE
LifePulse
HIGH FREQUENCY VENTILATOR
+
-
cm H2O
+
-
cm H2O
+
.8888.8
-
BPMSECONDS
+
-
GASWATER
ALARMS
LOSS OF PIP
CANNOT MEET PIP
HIGH PIP
BATTERY DEPLETED
LOW GAS PRESS
CHECK VENT
VENT INOP
+
:
8
-
I:E RATIO
ALARMS
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
05
bunnell
INSPIRED INFANT CARE
Chapter 5: VENTILATOR ALARMS
VENT INOP While Operating
During Operation: Codes 01 and 05-10
Codes 01 and 05-09 are all related to microprocessor or
electronic problems. These codes place the LifePulse in Standby
mode.
1. Support the patient on CV
2. Turn off the Ventilator Power switch to reset the LifePulse’s
microprocessor
3. Run the System and Operational Tests on a test lung before
restarting on a patient
WARNING: The patient is not being ventilated by
the LifePulse during VENT INOP Alarms 01 & 05-09,
which place the LifePulse in Standby mode.
WARNING: The patient is not being ventilated by
the LifePulse during VENT INOP Alarm 10, which
places the LifePulse into Standby mode. Always
troubleshoot the potential causes of an alarm
following the guidelines in the User’s Manual prior
to pressing the ENTER button.
VENT INOP CODES
01 – A/D converter failure
05 – RAM failure
06 – ROM failure
07 – NMI failure to execute
08 – IRQ failure to execute
Code 10 indicates a significant rise in Servo, which can be caused
by a dramatic change in ventilator settings, such as changing the
Rate from 420 to 240 bpm.
1. Check for mechanical problems such as tubing disconnects or
kinks
2. Correct any disconnects or kinks before pressing ENTER to
restart the LifePulse
VENT INOP NO CODE
There is one VENT INOP alarm that does not generate a code. It
is associated with the Watchdog circuit resetting the ventilator’s
microprocessor. This condition is quite rare. Typically, the LifePulse
continues to run normally following a reset.
There is also a backup ventilator alarm that can be generated if
the primary alarm system fails. There is no alarm indicator for this
condition, only a high priority alarm.
09 – Both IRQ and NMI failure to execute
10 – Significant rise in Servo
Bunnell LifePulse HFV User Manual | 58
Page 59
READY
LOWER
LIMIT
LOWER
LIMIT
ALARMS
LOSS OF PIP
CANNOT MEET PIP
HIGH PIP
BATTERY DEPLETED
LOW GAS PRESS
CHECK VENT
CHECK VENT
+
:
8
-
I:E RATIO
ALARMS
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
bunnell
INSPIRED INFANT CARE
MONITOR
88.8
-8.8
cm H2O
cm H2O
cm H2O
+
-
SERVO
88.8
Δ
P
88.8
MAP
420
RATEI-TIME
88.8
TEMPERATURE
88.8
PIP
PEEP
CONTROLS
20.
PIP
HUMIDIFIER
PAUSE
LifePulse
HIGH FREQUENCY VENTILATOR
+
-
UPPER
LIMIT
SYSTEM
cm H2O
cm H2O
BPMSECONDS
TEST
+
-
UPPER
LIMIT
+
.8888.8
-
+
-
GASWATER
Codes displayed in the temperature display on
the humidifier
111 - RAM Failure
222 - ROM Failure
333 - A/D Converter Failure
Chapter 5: VENTILATOR ALARMS
CHECK VENT
The CHECK VENT alarm indicates a problem (feedback error) with
one of the valves in the LifePulse or Patient Box, or electronic
failures in the Humidifier section. The following are possible
causes:
• The purge gas supply is missing or the purge tube is
disconnected or kinked.
• The pressure monitoring tube is disconnected or kinked at the
Patient Box pressure monitoring tube connector.
• A problem exists with the purge value or pinch valve in the
Patient Box, or the servo valves in the LifePulse.
Checks:
1. Make sure the purge tubing is connected and not kinked at
both the Purge port on the LifePulse’s front panel and the
From Purge port on the Patient Box.
2. Make sure the pressure monitoring tube is connected to the
Patient Box and not kinked.
3. Make sure the purge gas supply is present at the LifePulse
front panel by disconnecting the purge tube. If gas is present,
reconnect the purge tube.
4. Replace the Patient Box to determine if there is a purge valve
failure or a problem with the pinch valve causing the alarm.
5. If the purge gas supply is absent or if the alarm continues
after changing the Patient Box, change out the LifePulse.
6. The Humidifier codes can only be cleared by turning off
and back on the Ventilator Power switch to reset the
Humidifier. The patient will need to be supported during this
process. Once the LifePulse powers up, enter the patient’s
settings and press ENTER to restart the LifePulse.
7. Replace the LifePulse if the alarm cannot be cleared.
Bunnell LifePulse HFV User Manual | 59
Page 60
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
-
PAUSE
TEMPERATURE
GASWATER
CIRCUIT FAULT
HIGH TEMP
LOW LEVEL
88.8
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
LifePulse
bunnell
INSPIRED INFANT CARE
LOW GAS PRESS
+
+
-
-
ALARMS
ALARMS
WARNING: Only use medical grade oxygen and air
that is dry and free of dust and oil. The gas supply
pressure must be 30 – 60 psi (205.85 - 413.70 kPa).
Chapter 5: VENTILATOR ALARMS
LOW GAS PRESSURE
The LOW GAS PRESS alarm indicates that the pressure of the gas
supply to the LifePulse is below 30 psi (205.85 kPa) or the gas
supply pressure switch is faulty.
Checks:
1. Make sure a gas supply is connected to the blender and that
the low flow output from the blender is connected to the
Mixed Gas Input on the back of the LifePulse.
2. Make sure the gas supply pressure is greater than 30 psi
(205.85 kPa). If the supply pressure is less than this level, check
the high pressure supply hoses for leaks. Change the gas
source, blender, or high pressure hoses as needed.
3. If the alarm persists, there may be a problem with the gas
supply pressure switch in the LifePulse, in which case you will
have to switch to another LifePulse ventilator.
Bunnell LifePulse HFV User Manual | 60
Page 61
Chapter 5: VENTILATOR ALARMS
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
CONTROLS
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
BATTERYDEPLETED
READY
AUDIO
MONITOR
PIP
PEEP
CONTROLS
READY
UPPER
SERVO
cm H2O
cm H2O
Δ
P
cm H2O
cm H2O
MAP
LIMIT
SYSTEM
UPPER
LIMIT
TEST
LOWER
LOWER
ALARMS
LIMIT
BATTERY
LIMIT
PAUSED
STANDBY
ENTER
The battery’s capacity can be extended by placing the
Humidifier in Pause mode.
Battery Alarms
The battery alarms are progressive as the capacity of the battery
decreases.
If the LifePulse is running on battery,
• the BATTERY alarm indicator will be lit
• the battery charge indicator lamp will be off
• there will be a low priority alarm, which can be cleared by
pressing AUDIO PAUSED
• the green bars on the battery fuel gauge will indicate the
battery’s remaining capacity
If the LifePulse is plugged in and the battery has malfunctioned,
is missing, or will not take a charge,
• the battery charge indicator lamp will be red continuously
• there will be a low priority alarm
• the battery fuel gauge bars will not be lit
If the battery’s capacity is low,
• the BATTERY alarm indicator will be lit
• there will be a medium priority alarm
• the battery fuel gauge bars will be depleted down to a single
flashing red bar
WARNING: If the battery charge indicator stays red
with the LifePulse plugged into an electrical wall
outlet, the battery needs to be checked by Bunnellauthorized service personnel. The ventilator and
humidifier will function normally when plugged in
but the battery will not provide adequate power, if
needed.
WARNING: The LifePulse battery is low and needs to
be charged when the battery fuel gauge is blinking
red. Plug the LifePulse into an electrical wall outlet.
WARNING: When the battery fuel gauge is blinking
red with the Battery Depleted alarm lit and a high
priority alarm active the battery charge is too low
to continue running the LifePulse and it will go to
Standby mode in a minute or two.
Plug the power cord into an electrical wall outlet to re-establish
the electrical power supply.
If the battery’s capacity is critically low and unable to support
the LifePulse’s continued functioning,
• the BATTERY DEPLETED alarm indicator will be lit
• there will be a high priority alarm
Immediate action must be taken to keep the LifePulse from being
forced into Standby mode.
Checks:
1. Verify that the LifePulse’s power cord is connected securely at
the back panel and at the electrical wall outlet. If it is, try
plugging the power cord into a different outlet.
2. If the battery has failed, switch out the LifePulse so the battery
can be checked and serviced.
3. If the hospital’s electrical power supply has failed, switch out
the LifePulse with one that has a fully charged battery.
Otherwise, support the patient with manual ventilation or other
forms of ventilatory support.
Bunnell LifePulse HFV User Manual | 61
Page 62
Chapter 5: VENTILATOR ALARMS
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH
TEMP
LOW
LEVEL
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH
LOW
LEVEL
LifePulse
bunnell
INSPIRED INFANT CARE
o
C
READY
ALARMS
CHECK CIRCUIT
CHECK CIRCUIT alarms will appear on the right side of the
Humidifier section.
CHECK CIRCUIT - TEMP alarms indicate that the cartridge door
is not properly latched or that an electrical fault is present in the
humidifier cartridge or Patient Circuit.
CHECK CIRCUIT - LEVEL alarms indicate that the humidifier
cartridge did not fill to the normal operating level in 86 seconds
CHECK CIRCUIT - TEMP
Checks:
1. Inspect the humidifier cartridge door to make sure it is latched
properly.
2. Open the cartridge door, leave the cartridge in place, and
check the spring-loaded contact pins by pressing them inward
with a pencil eraser. Make sure they spring in and out and that
all the pins are at about the same level ±1-2 mm. Try pulling
stuck pins back out to the level of the other pins. Close the
cartridge door, press PAUSE again, and see if the alarm has
cleared.
3. If the alarm persists, press STANDBY and replace the Patient
Breathing Circuit. Press ENTER to resume normal operation.
If the alarm does not go away, it is likely that the humidifier
module has failed internally and the LifePulse will need to be
changed out.
CHECK CIRCUIT - LEVEL
Checks:
1. If the alarm occurs within the first 2 minutes after pressing
ENTER, check the water supply to see if it is empty or
disconnected. Also, inspect the water supply tubing coming
from the water supply to see if it is clamped off or
disconnected.
2. Repeat steps 2 and 3 above.
Bunnell LifePulse HFV User Manual | 62
Page 63
Chapter 5: VENTILATOR ALARMS
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGHTEMP
LOW
LEVEL
40.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
HIGH TEMP - WATER
The HIGH TEMP - WATER alarm will occur if one of the following
conditions is met:
• The temperature measured in the humidifier cartridge has
exceeded the Water temperature setting by more than 3°C
and has remained high for 10 minutes
• The measured temperature is ≥45°C, which is the upper
allowable limit
Checks:
1. Place the humidifier in Pause mode. Open the cartridge door
and check the spring-loaded contact pins with a pen or pencil.
Make sure they spring in and out and that all the pins are at
about the same level ±1-2 mm. Try pulling stuck pins back out
to the level of the other pins. Close the cartridge door, press
PAUSE again, and see if the alarm has cleared.
2. If the alarm persists, press STANDBY and replace the Patient
Breathing Circuit. Press ENTER to resume normal operation.
If the alarm does not go away, it is likely that the humidifier
module has failed internally and the LifePulse will need to be
changed out.
Bunnell LifePulse HFV User Manual | 63
Page 64
Chapter 5: VENTILATOR ALARMS
READY
UPPERLIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
CIRCUIT FAULT
HIGH
TEMP
LOW
LEVEL
40.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
LOW TEMP - WATER
The LOW TEMP - WATER alarm will occur when the following
condition is met:
• The temperature measured in the humidifier cartridge has
dropped below the Water temperature setting by 3°C and
remained there more than 30 minutes
Checks:
1. Place the humidifier in Pause mode. Open the cartridge door
and check the spring-loaded contact pins using a pen or pencil.
Make sure they spring in and out and that all the pins are at
about the same level ±1-2 mm. Try pulling stuck pins back out
to the level of the other pins. Close the cartridge door, press
PAUSE, and see if the alarm has cleared.
2. If the alarm persists, press STANDBY and replace the Patient
Breathing Circuit. Press ENTER to resume normal operation.
If the alarm does not go away, it is likely that the humidifier
module has failed internally and the LifePulse will need to be
changed out.
Bunnell LifePulse HFV User Manual | 64
Page 65
Chapter 5: VENTILATOR ALARMS
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH TEMP
LOW LEVEL
44.2
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
HIGH TEMP - GAS
The HIGH TEMP - GAS alarm occurs when one of the following
conditions is met:
• The temperature in the patient circuit, as measured proximal to
the Patient Box, has exceeded the Gas temperature setting by
more than 3°C and remained there for more than 1 minute
• The measured temperature is ≥45°C, which is the upper
allowable limit
Checks:
1. Place the humidifier in Pause mode. Open the cartridge door
and check the spring loaded contact pins using a pen or pencil.
Make sure they spring in and out and that all the pins are at
about the same level ±1-2 mm. Try pulling stuck pins back out
to the level of the other pins. Close the cartridge door, press
the PAUSE button, and see if the alarm has cleared.
2. If the alarm persists, press STANDBY and replace the Patient
Breathing Circuit. Press ENTER to resume normal operation.
If alarm does not go away, it is likely that the humidifier module
has failed internally and the LifePulse will need to be
changed out.
Bunnell LifePulse HFV User Manual | 65
Page 66
Chapter 5: VENTILATOR ALARMS
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH
TEMP
LOW
LEVEL
36.8
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
LOW TEMP - GAS
The LOW TEMP - GAS alarm will occur when the following
condition is met:
• The temperature in the patient circuit, as measured proximal
to the Patient Box, has dropped below the Gas temperature
setting by more than 3°C and has remained there for more
than 3 minutes
Checks:
1. Place the humidifier in Pause mode. Open the cartridge door
and check the spring loaded contact pins using a pen or pencil.
Make sure they spring in and out and that all the pins are at
about the same level ±1-2 mm. Try pulling stuck pins back
out to the level of the other pins. Close the cartridge door,
press PAUSE, and see if the alarm has cleared.
2. If the alarm persists, press STANDBY and replace the Patient
Breathing Circuit. Press ENTER to resume normal operation. If
the alarm does not go away, it is likely that the humidifier
module has failed internally and the LifePulse will need to be
changed out.
Bunnell LifePulse HFV User Manual | 66
Page 67
READY
UPPERLIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPERLIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH
TEMP
LOW
LEVEL
40.0
Life Pulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
MONITOR
88.8
-8.8
cm H2O
cm H2O
cm H2O
+
-
SERVO
88.8
Δ
P
88.8
MAP
420
RATEI-TIME
40.0
TEMP
88.8
PIP
PEEP
CONTROLS
20.
PIP
HUMIDIFIER
PAUSE
LifePulse
HIGH FREQUENCY VENTILATOR
READY
READY
READY
+
-
UPPERLIMIT
SYSTEM
cm H2O
cm H2O
TEST
+
-
UPPER
LIMIT
+
.8888.8
-
BPMSECONDS
+
-
o
GASWATER
C
LOWER
LIMIT
LOWER
LIMIT
ALARMS
LOSS OF PIP
CANNOT MEET PIP
HIGH PIP
BATTERY DEPLETED
LOW GAS PRESS
CHECK VENT
VENT INOP
+
:
8
-
I:E RATIO
CHECK CIRCUIT
HIGHTEMP
LOW
LEVEL
bunnell
INSPIRED INFANT CARE
Chapter 5: VENTILATOR ALARMS
HIGH or LOW LEVEL - WATER
The High Level alarm indicates that the water in the humidifier
cartridge has reached the high water level sensor. The Low Level
alarm indicates the water in the humidifier cartridge has not
reached the low water level sensing pin.
HIGH LEVEL - WATER
Checks:
1. Make sure sterile H2O, not saline, was used to fill the humidifier
cartridge.
2. Make sure the water inlet tube is installed properly in the water
pump housing.
3. Was the humidifier manually filled to the high level sensor?
NOTE: If the circuit fails and has been in use for less
than seven days, call Bunnell Incorporated for a
Return Authorization number and return the circuit for
a possible credit.
4. Was the water supply clamp opened prior to securing the
water inlet tube inside the water pump housing, so that the
humidifier cartridge over-filled by gravity?
5. Open the cartridge door, leave the cartridge in place, and
check the spring-loaded contact pins by pressing them inward
with a pencil eraser. Make sure they spring in and out and that
all the pins are at about the same level ±1-2 mm. Try pulling
stuck pins back out to the level of the other pins. Close the
cartridge door, press PAUSE, and see if the alarm has cleared.
6. Press STANDBY, support the patient with the conventional
ventilator, and replace the Patient Circuit.
7. Press ENTER to resume normal operation. If the alarm does not
go away, it is likely that the humidifier module has failed
internally and the LifePulse will need to be changed out.
LOW LEVEL - WATER
Checks:
1. Inspect the water supply tubing from the water supply to see if
it is clamped off or disconnected. Unclamp or reconnect the
water supply tubing.
2. Inspect the water supply to see if it is empty. Replace the water
supply if necessary.
3. Make sure that sterile H2O, not saline, was used to fill the
humidifier cartridge.
4. Follow 5-6 above.
Bunnell LifePulse HFV User Manual | 67
Page 68
Chapter 5: VENTILATOR ALARMS
Bunnell LifePulse HFV User Manual | 68
Page 69
Chapter 6: HUMIDIFICATION
+
-
HIGH FREQUENCY VENTILATOR
HUMIDIFIER
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
CHECK CIRCUIT
HIGH TEMP
LOW LEVEL
cm H2O
88.8
LifePulse
bunnell
INSPIRED INFANT CARE
PURGE
PATIENT
BOX
BATTERY
_
+
GAS OUT
o
C
ALARMS
Humidifier Overview
The Humidifier section resides in the lower half of the LifePulse
and consists of controls, alarms, a PAUSE button, a receptacle for
the humidifier cartridge, and a water pump.
The Humidifier section is responsible for heating and humidifying
the gas provided to the patient. It requires very little adjustment
and will self regulate the water level, water temperature, and gas
temperature.
The Humidifier alarms will alert you to potential problems with the
Patient Circuit.
Bunnell LifePulse HFV User Manual | 69
Page 70
Chapter 6: HUMIDIFICATION
3
2
4
1
10
7
5
6
Patient Circuit Overview
The LifePulse’s Humidifier section uses a disposable cartridge-andtubing set called the Patient Circuit. This item is often referred
to as “the cartridge,” “the circuit,” or “the humidifier cartridge/
circuit”; all are terms that refer to all or part of the Patient Circuit.
It is pre-assembled and contains the heating wire, thermistors, and
connections needed for operation.
The Patient Circuit includes the following components:
1. Cartridge
contains the water used to humidify the gas delivered to the
patient
2. Gas Inlet Tube
transports blend ed gas to the cartridge
3. Cartridge Thermistor
measures the temperature of the humidified gas before it exits
the cartridge in order to maintain the desired water temperature
4
4. Purge Tube
8
delivers a low flow of dry gas to the Patient Box; the gas will
periodically purge the LifePort pressure monitoring tube
5. Patient Circuit Tube
the primary conduit for transferring heated, humidified gas
from the LifePulse to the patient
The temperature thermistor at the tip of the white
wire in the circuit measures the actual temperature of
the humidified gas in the circuit just before the Patient
Box. The value is displayed in the TEMP window in the
Humidifier display section.
The blue wire is a heating element that is turned on
and off according to feedback from the measured
gas temperature. It also minimizes the amount of
condensation in the tubing.
WARNING: External temperature of the heated
portion of the patient circuit may exceed 50°C.
Contact with the patient’s skin should be avoided.
WARNING: Do not reuse or sterilize Patient Circuits
or LifePort adapters. Failure to discard the Patient
Circuit after one use or seven days may result in
leaks, improper temperature, and water level control.
9
6. Heating Wire
heats the gas being delivered to the patient to maintain the
desired temperature
7. Circuit Thermistor
measures the temperature of the gas before the gas passes
through the Patient Box in order to maintain the desired gas
temperature
8. Pinch Tubing
periodically compressed and released by the Patient Box pinch
valve to create high frequency breaths
9. Circuit Leur Connector
inserts by friction fit into the Jet port of the LifePort adapter
10. Water Inlet Tube
transfers sterile USP inhalation water from a supply bag to the
cartridge
Bunnell LifePulse HFV User Manual | 70
Page 71
Chapter 6: HUMIDIFICATION
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
40.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
Press ENTER to start humidifier
and fill cartridge
Press PAUSE to stop and restart
humidifier functions
WARNING: Do not pre-fill the humidifier cartridge
of the Patient Circuit prior to starting the LifePulse.
Doing so may result in the cartridge overfilling and
the delivery of a bolus of water to the patient. The
cartridge will fill automatically once the LifePulse
begins operating.
WARNING: Do not open the water pump door while
the pump is pumping. Doing so may result in a pinch
hazard to the operator.
WARNING: Always attempt to minimize
condensation/rainout in the Patient Circuit between
the ventilator and the Patient Box. Failure to do so
may result in water interfering with the monitored
pressure signal at the LifePort adapter resulting in
fluctuations in the delivered PIP.
Starting the Humidifier
The humidifier starts when ENTER is pressed to begin HFJV:
• The water pump starts filling the humidifier cartridge with
water.
• Water in the cartridge is heated by the hot plate behind the
cartridge.
• The circuit heater wire in the patient circuit tube begins
heating the gas.
• The water pump stops when the water in the cartridge contacts
the second water level sensing pin (normal operating level).
Stopping the Humidifier
The humidifier will stop operating whenever STANDBY is pressed.
The humidifier will resume operation when ENTER is pressed.
The humidifier will also stop operating when the PAUSE button in
the Humidifier section is pressed. The stoppage will be indicated
by the following conditions:
• flashing blue light in the corner of the PAUSE button
• solid yellow alarm priority lamp
• single-beep alarm every 18 seconds as a reminder that the
humidifier is not operating
Pressing PAUSE again will bring the humidifier back into
operation.
WARNING: Do not under-humidify gas being
delivered to the patient. Adjust the cartridge
temperature so that condensation is always present in
the green delivery tube between the Patient Box and
the LifePort adapter.
WARNING: Pushing the PAUSE button on the
humidifier shuts off the heating elements that provide
for gas warming and humidification. A properly
trained person must not leave the patient’s bedside
with the Humidifier in the PAUSE mode.
Bunnell LifePulse HFV User Manual | 71
Page 72
Chapter 6: HUMIDIFICATION
High water level
sensing pin
Low water level
sensing pin
Normal water level is at
2nd level sensing pin
Water Level Sensing
Whenever installing a new Patient Circuit, watch the cartridge fill
with water until the pump stops to ensure that the pump does not
overfill the cartridge.
The pump stops when the water reaches the second water level
sensing pin inside the cartridge.
The cartridge has a third sensing pin to detect high water
and prevent overfill into the circuit. The water pump stops
automatically if the water level reaches the third pin. A HIGH
WATER LEVEL alarm will be generated and the water pump is
disabled. If the pump does not turn off at the third sensing pin,
press PAUSE to turn off the humidifier while you troubleshoot the
problem.
If any of the water level sensing pins are defective, the water
level can’t be sensed properly. A default maximum fill time of 86
seconds is programmed to turn the pump off in time to prevent
overfill.
Bunnell LifePulse HFV User Manual | 72
Page 73
Chapter 6: HUMIDIFICATION
Gas Flow through Humidifier
The gas flows through the LifePulse humidifier and circuit to the
patient as follows:
1. The gas enters the cartridge through the green gas inlet tube.
2. The gas flows down into the water and then back up through a
Venturi mechanism, which atomizes some of the heated water.
3. The gas flows over the heated reservoir of water and past a
series of baffles which filter out water droplets.
4. The humidified gas passes over the cartridge thermistor.
The LifePulse’s microprocessor uses the measured temperature
to regulate the amount of heat delivered to the cartridge
through the metal heating plate.
5. The black heater panel heats the water based on feedback
control from the cartridge thermistor.
6. Heated and humidified gas leaves the cartridge and enters the
circuit tubing with blue and white wires inside.
4
6
1
2
3
5
(continued on next page)
Bunnell LifePulse HFV User Manual | 73
Page 74
Chapter 6: HUMIDIFICATION
Gas Flow through Patient Box
7. The heated and humidified gas flows into the pinch tubing
portion of the circuit where the pinch valve creates the high frequency breaths.
8. The gas begins to cool as it leaves the pinch valve and
condensation occurs as a result.
Pinch tubingPinch valve
78
Bunnell LifePulse HFV User Manual | 74
Page 75
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
40.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
The actual Gas Temperature should
be 40°C and will be displayed in the
Temp window.
Chapter 6: HUMIDIFICATION
Understanding Gas Temperature
The actual gas temperature is measured in the patient circuit tube
just before the Patient Box and is displayed in the Temp window in
the Humidifier section.
Upon starting HFJV, the temperature in the circuit will approach
the default set value of 40oC; it may overshoot slightly before
settling in right at or near 40oC.
There is a drop in gas temperature of 3oC from the circuit
thermistor to the patient. The default setting for the Gas
temperature is 40oC to compensate for this temperature drop.
If the set gas temperature is 40oC, the actual temperature of the
gas as it enters the patient will be approximately 37oC (i.e., body
temperature).
The lungs are very effective heat exchangers. Therefore, unless
there is a reason for wanting to raise or lower the patient’s core
temperature, do not adjust the set gas temperature to any value
other than 40oC.
WARNING: Raising or lowering the Gas temperature
setting may raise or lower the patient’s temperature.
The normal setting of 40oC is designed to keep the
patient from losing any heat through respiration.
It should not be changed under normal operating
circumstances.
Bunnell LifePulse HFV User Manual | 75
Page 76
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
38.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
READY
UPPER
LIMIT
LOWER
LIMIT
VENT INOP
CHECK VENT
LOW GAS PRESS
BATTERY DEPLETED
HIGH PIP
CANNOT MEET PIP
LOSS OF PIP
+
+
-
-
UPPER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
READY
PIP
PEEP
SERVO
P
Δ
MAP
cm H2O
cm H2O
cm H2O
cm H2O
88.8
88.8
-8.8
88.8
88.8
+
-
+
-
20.
420
.8888.8
:
8
+
-
PIP
RATEI-TIME
I:E RATIO
cm H2O
BPMSECONDS
ALARMS
+
-
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PAUSE
TEMP
GASWATER
40.0
LifePulse
bunnell
INSPIRED INFANT CARE
READY
o
C
ALARMS
Use the + and - buttons to
raise or lower Gas and Water
Temperature.
WARNING: Do not under-humidify gas being
delivered to the patient. Adjust the cartridge
temperature so that condensation is always present in
the green delivery tube between the Patient Box and
the LifePort adapter.
Chapter 6: HUMIDIFICATION
Understanding Water Temperature
The water temperature controls the humidity level of the gas
delivered to the patient. Lowering the set water temperature is
the best way to reduce excess condensation in the circuit. See the
next page for information about identifying proper humidification
levels.
Upon starting HFJV, the water temperature will default to a set
value of 38°C. The set water temperature can be seen by pressing
WATER in the Humidifier section.
Temperature Controls
The temperature in the cartridge (water) and the circuit tubing
(gas) are set and controlled independently. Pressing the WATER or
GAS button in the Humidifier section will display the set value of
the respective temperature.
The gas temperature should be kept at its default setting of 40°C.
Change the water temperature to provide appropriate levels of
humidification if over- or under-humidification occurs. See the
next page for information about identifying proper humidification
levels. To change the water temperature,
12
1. Press WATER to display the set value.
2. Adjust the temperature up or down by using
the + or - adjustment button.
The displayed value is the new set temperature. The ENTER
button does not have to be pressed to activate the new set
temperature value.
See Identifying Proper Humidification for information about
appropriate humidification levels.
Bunnell LifePulse HFV User Manual | 76
Page 77
Fine mist
Proper Humidification
Fine mist in green delivery tube of circuit
Chapter 6: HUMIDIFICATION
Identifying Proper Humidification
Proper humidification of the gas in the circuit will appear as a fine
mist in the green delivery tube between the Patient Box and the
LifePort adapter. It will look similar to the mist you would see when
breathing on a mirror or against a car window on a cold winter
day. This condensation is an indication that the gas has reached
100% relative humidity.
Over-humidification appears as a collection of water droplets
pooling in the clear portion of the circuit tubing between the
Humidifier and the Patient Box. Water in this portion of the circuit
can cause pressure instability and LOSS OF PIP alarms. This
condition can be alleviated by lowering the set water temperature
in small increments, as needed.
Water pooling
Over-Humidification
Water droplets continuously pooling in clear
portion of circuit tube
Tube is dry
Under-Humidification
Green delivery tube of circuit is dry
Under-humidification will be indicated by the green delivery tube
between the Patient Box and the LifePort being dry. This can be
corrected by raising the set water temperature in 0.5oC increments
at 30-minute intervals until condensation is present.
Bunnell LifePulse HFV User Manual | 77
Page 78
Chapter 6: HUMIDIFICATION
Changing the Circuit
1
9
5
8
4
Removing the Old Circuit
A circuit change should be done with the LifePulse in Standby
mode and the patient supported by the CV. The procedure is
best performed with two people. Both people can complete
their tasks simultaneously.
1. Press STANDBY on the LifePulse and increase CV settings
to the minimum level required to support the patient.
2. Disconnect the used circuit from the Jet port on the side
of the LifePort adapter and cap the Jet port.
4
6
Water Supply
7
3
3. Remove the pinch tube from the jaws of the pinch valve on the
Patient Box.
4. Disconnect the purge tube from both the Patient Box and the
Purge port on the LifePulse.
5. Disconnect the green gas inlet tube from the Gas Out port on
the LifePulse.
6. Clamp off the water transfer tube connecting the water supply
to the water inlet tube.
7. Disconnect the water transfer tube from the water inlet tube.
2
WARNING: Clamp the water supply tube prior to
opening the water pump door to prevent cartridge
overfill and delivery of water to the patient by gravity
feed. The water supply should be positioned at or
below the level of the humidifier cartridge as an
added precaution.
WARNING: Do not tip the humidifier cartridge
while removing it from the cartridge housing if the
LifePulse is running. Doing so might result in water
spilling into the Patient Circuit and a bolus of water
being delivered to the patient.
8. Remove the water inlet tube from the water pump.
9. Open the cartridge door and remove the cartridge.
Discard the used Patient Circuit.
You are now ready to install the new circuit.
Bunnell LifePulse HFJV User Manual | 78
Page 79
Chapter 6: HUMIDIFICATION
3
1
5
4
Installing the New Circuit
Once the used circuit has been removed from the bedside and
discarded, the new circuit may be installed.
4
8
Water Supply
2
7
6
3. Attach the green gas inlet tube to the Gas Out port on the
LifePulse.
4. Attach the purge tube to the purge ports on both the
LifePulse and the Patient Box.
1. Slide the new humidifier cartridge into the cartridge door and
close the door.
2. Secure the water inlet tube into the water pump and snap the
pump housing closed.
WARNING: Clamp the water supply tube prior to opening
the water pump door to prevent cartridge overfill and
delivery of water to the patient by gravity feed. The water
supply should be positioned at or below the level of the
humidifier cartridge as an added precaution.
5. Insert the pinch tubing into the jaws of the pinch valve on
the Patient Box.
6. Uncap the Jet port on the LifePort adapter and insert the
green delivery tube into the Jet port.
7. Reattach the water transfer tube to the water inlet tube
and spike the water supply. The water supply should be at
or below the level of the humidifier cartridge.
8. Unclamp the water transfer tube.
Bunnell LifePulse HFV User Manual | 79
Page 80
WARNING: Do not leave the patient’s bedside while
the LifePulse water pump is running during initial startup or following a circuit change. A properly trained
person must observe the cartridge fill with sterile
water for inhalation, USP, to the second water level
sensing pin and the water pump stops pumping.
WARNING: Do not open the water pump door while
the pump is pumping. Doing so may result in a pinch
hazard to the operator.
WARNING: Removal of the humidifier cartridge from
the cartridge housing while the LifePulse is running
exposes the heater plate and cartridge heater cup,
which may be in excess of 60ºC.
WARNING: Do not move the LifePulse while it is
running on a patient. Doing so may result in water
entering the Patient Circuit and creating a risk of
water delivery to the patient. Monitor the water level
closely under these circumstances.
Chapter 6: HUMIDIFICATION
Completing the Circuit Change
To complete a circuit change,
1. Resume HFJV by pressing ENTER and lowering the CV settings
to their original values.
2. Verify that the water pump stops when water fills the cartridge
to normal operating level (approximately 2/3).
3. Verify that in 20-30 minutes humidity (condensation) appears in
the green delivery tube of the circuit between the Patient Box
and the LifePort adapter.
On rare occasions, a defective cartridge might not fill, might over
fill, or might not heat properly. The Patient Circuit will then need to
be replaced.
WARNING: Operating while the LifePulse or patient
is in motion or while in high-flow or large tidal volume
conditions may result in water entering the Patient
Circuit and creating a risk of aspiration. Please monitor
the water level closely.
Bunnell LifePulse HFV User Manual | 80
Page 81
Chapter 7: BATTERY
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
HIGH FREQUENCY VENTILATOR
CONTROLS
HUMIDIFIER
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
PIP
RATEI-TIME
I:E RATIO
PAUSE
TEMP
GASWATER
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
cm H2O
BPMSECONDS
LifePulse
bunnell
INSPIRED INFANT CARE
PURGE
PATIENT
BOX
BATTERY
_
+
GAS OUT
READY
o
C
ALARMS
The LifePulse Model 204 contains a lead-acid battery for
continued operation during power outages, after accidental
disconnects, and while moving a patient within the hospital
setting, if necessary.
Charging the Battery
The battery automatically charges when the LifePulse is plugged
into an electrical wall outlet with the Circuit Breaker switch on
the rear panel in the ON (I) position. Note: The Ventilator Power
switch does not have to be in the ON (I) position for the battery to
charge.
The battery will be fully charged in no more than twelve (12)
hours. This charging time is based on the time needed to
reach a full charge following a full discharge. The charging
time may be significantly longer than 12 hours if the battery
has been discharged several times in succession without being
fully recharged or if the battery has been left uncharged for a
significant amount of time. It is best to leave the LifePulse
plugged in at all times to make sure the battery is fully
charged.
While the battery is charging, the charge indicator lamp below the
water pump indicates the charge status. When the LifePulse is first
plugged in, the charger circuit checks the battery’s condition, the
charge indicator lamp will blink green to indicate charging, turn
solid green to indicate fully charged, turn solid red to indicate a
bad battery, or alternately blink green and red to indicate a severe
discharge. The severe discharge condition can take up to 30
minutes to resolve itself.
If the battery is missing or cannot take a charge, the charge
indicator lamp will turn solid red and the battery fuel gauge will
be off. This low priority alarm can be cleared by pressing AUDIO
PAUSED. Once a charging battery is fully charged, the charge
indicator lamp and the battery fuel gauge will turn from blinking
green to solid green. The charger circuit continues to apply a
Battery Indicators
small, “trickle” current to the battery to maintain the charge level.
Bunnell LifePulse HFV User Manual | 81
Page 82
Chapter 7: BATTERY
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
CONTROLS
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
BATTERY
READY
STANDBY
ENTER
CONTROLS
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
AUDIO
MONITOR
PIP
PEEP
CONTROLS
READY
UPPER
SYSTEM
UPPER
LIMIT
TEST
LIMIT
SERVO
cm H2O
cm H2O
Δ
P
cm H2O
cm H2O
MAP
LOWER
LOWER
ALARMS
LIMIT
BATTERY
LIMIT
PAUSED
STANDBY
ENTER
WARNING: The discharge rate of the battery
changes with the ventilator and humidifier settings as
well as with the age of the battery. This can shorten
the overall time the ventilator will operate on battery
power before the battery depletes and the ventilator
stops operating.
Running on Battery
The LifePulse will automatically switch to battery power if the main
electrical power is lost for any reason. (Note: The LifePulse cannot
be powered up on battery power alone.)
The Battery indicator will light up in the Alarms section of the front
panel if the LifePulse is running on battery. The charge indicator
lamp below the water pump will turn off, and the low priority
audible and visual alarm will be activated. This alarm can be
cleared by pressing AUDIO PAUSED.
The actual battery run time available once the battery is fully
charged depends on a number of variables including the battery
age, condition, and the ambient temperature. The absolute
condition of the battery cannot be determined by the charger
circuit. Therefore, even though the green lamp is on continuously,
the run time of old or defective batteries may be less than the
operating times specified.
As the battery capacity drops the battery fuel gauge will decrease
(turn off) one bar at a time. There are five green bars and the last
green bar turns red and starts flashing when the capacity is very
low. The alarm priority changes to medium at this time (yellow
flashing alarm priority lamp and three beeps every 5 seconds).
WARNING: If the battery charge indicator stays
red with the LifePulse plugged into an electrical wall
outlet, the battery needs to be checked by Bunnellauthorized service personnel. The ventilator and
humidifier will function normally when plugged in
but the battery will not provide adequate power,
if needed.
It is important to note that the fuel gauge reflects the battery
voltage not the rate of discharge in time. The voltage will decrease
differently depending on the LifePulse settings, battery age, and
usage history.
The LifePulse must be plugged into an electrical power supply
once the BATTERY Low alarm is generated; or, the LifePulse must
be switched out with one that has a fully charged battery.
Placing the LifePulse humidifier into Pause mode can extend the
battery run time up to three minutes before Standby mode will
occur.
Bunnell LifePulse HFV User Manual | 82
Page 83
ALARMS
AUDIO
PAUSED
STANDBY
ENTER
CONTROLS
MONITOR
PIP
PEEP
SERVO
P
Δ
MAP
UPPER
LIMIT
UPPER
LIMIT
LOWER
LIMIT
LOWER
LIMIT
SYSTEM
TEST
cm H2O
cm H2O
cm H2O
cm H2O
BATTERYDEPLETED
READY
CAUTION: The capacity of the battery must be
checked regularly and the battery replaced if
necessary. Avoid deep discharges, as they lead to
premature degradation of the battery.
Chapter 7: BATTERY
The final battery alarm is BATTERY DEPLETED. This high priority
alarm is the final alert before the battery fails and the LifePulse is
forced into Standby mode. To avoid being forced into Standby
mode the LifePulse must either be immediately plugged into an
electrical wall outlet or be switched out with one that has a fully
charged battery.
If the LifePulse is forced into Standby mode, support the patient
and immediately plug the LifePulse power cord into an electrical
wall outlet. Once power is restored, press ENTER to re-establish
ventilation.
Battery Maintenance
The battery should always be fully charged, and if possible,
never deeply discharged. No damage to the battery will result
from leaving the LifePulse plugged into an electrical wall outlet
indefinitely. If the LifePulse cannot be plugged into an electrical
wall outlet indefinitely, at a minimum it should be plugged in at
least once every month for 12 hours. The battery may be damaged
as a result of long-term storage without being charged and should
be replaced over time and with extensive use.
Main Power Switch
Circuit Breaker
To evaluate the battery’s capacity/condition, set up the LifePulse
so it can be run on a test lung with the humidifier operational
(refer to Setup and Tests Chapter). Make sure the LifePulse has
been plugged in to an electrical wall outlet for at least 12 hours
and the Battery Charge Indicator is lit solid green (battery fully
charged) before beginning this evaluation.
Turn ON the main power switch on the rear panel. The Battery
Fuel Gauge should display five green bars when the battery is fully
charged. Once the LifePulse is running on default settings, switch
the circuit breaker on the rear panel to the OFF position. This will
force the LifePulse to switch to battery power. Let the LifePulse
run on battery for 15 minutes, then note how many Fuel Gauge
bars are lit.
The battery has adequate capacity if one or more green bars are
lit at the end of 15 minutes. If the last bar turns red and flashes
with a Medium Priority alarm, or the LifePulse stops operating,
the battery capacity/function should be evaluated by a Bunnell
authorized service personnel before the next patient application.
The battery’s capacity should be evaluated once a month.
Switch the circuit breaker on the rear panel to the ON position.
This will switch power to the LifePulse back to the wall outlet. The
LifePulse may now be powered off by turning the main power
switch to the OFF position. Be sure to leave the LifePulse plugged
into the wall outlet for at least 12 hours after the battery capacity
has been evaluated to allow the battery to fully recharge.
The LifePulse will still function normally as long as it is plugged
into electrical power.
Bunnell LifePulse HFV User Manual | 83
Page 84
WARNING: Improper replacement of the battery by
non-Bunnell authorized service personnel could result
in an unacceptable risk (e.g., excessive temperature,
fire, or explosion).
WARNING: Batteries that are improperly disposed of
may cause harm to the environment.
Chapter 7: BATTERY
Physically damaged batteries may leak and damage the ventilator
and/or other equipment. Damaged batteries should be replaced
immediately by Bunnell authorized service personnel.
Batteries should last approximately three years under normal
operating conditions with proper care. The battery should be
replaced every 3 years.
The battery should only be replaced by Bunnell-authorized service
personnel and should only be replaced with the specific battery
supplied by Bunnell Incorporated.
These batteries contain lead and should be recycled. Follow local
or regional standards for proper disposal.
Bunnell LifePulse HFV User Manual | 84
Page 85
Chapter 8: CLEANING AND STORAGE
WARNING: Clean and disinfect the LifePulse and
Patient Box prior to patient applications to avoid the
risk of infection. Refer to cleaning instructions in the
User’s Manual.
WARNING: The use of non-recommended
solvents or cleaners can damage the enclosure(s)
or user interface and damage equipment causing
the LifePulse to malfunction. Refer to the cleaning
instructions in the User’s Manual.
WARNING: Do not reuse or sterilize Patient Circuits
or LifePort adapters. Failure to discard the Patient
Circuit after one use or seven days may result in leaks,
improper temperature and water level control.
WARNING: Plug the ventilator into an electrical wall
outlet at all times to maintain proper battery charge.
Charging must only occur in a properly ventilated
room. The user must ensure the battery is in good
condition and has sufficient capacity prior to every
patient use.
Chapter 8: CLEANING AND STORAGE
The exterior surfaces of the LifePulse and Patient Box may be
cleaned with a mild soap and water solution or hospital grade
disinfectants. Care should be taken not to saturate the device
surfaces or spray cleaning products directly on surfaces.
The disposable Patient Circuit and LifePort adapters are for single
patient use only and must be discarded after use. Do not sterilize
and reuse the Patient Circuits.
During storage, plug the LifePulse into an electrical wall outlet
in a properly ventilated room in order to maintain proper
battery charge.
Ensure the battery is in good condition and has sufficient capacity
prior to every patient use.
Do not store the LifePulse with the pinch tube of the circuit loaded
in the pinch valve jaws of the Patient Box. Doing so may result
in the pinch tube sticking to itself and could create a LOSS OF
PIP alarm on start-up or a spontaneous but brief delivery of an
excess volume of gas to the patient when the pinch tube becomes
unstuck.
WARNING: Do not store LifePulse with the pinch
tube of the circuit loaded in the pinch valve jaws of
the Patient Box.
WARNING: The LifePulse, Patient Box, and Patient
Circuit contain components made of hazardous
materials that need to be disposed of appropriately.
Improper disposal may cause harm to the
environment.
CAUTION: Do not sterilize the LifePulse ventilator
or Patient Box. The internal components are not
compatible with sterilization techniques.
Bunnell LifePulse HFV User Manual | 85
Page 86
Chapter 8: CLEANING AND STORAGE
Bunnell LifePulse HFV User Manual | 86
Page 87
Appendix A
Owner Responsibility for Maintenance
The Bunnell Incorporated equipment and authorized accessories
and supplies are designed to function as specified when operated,
maintained, and serviced in accordance with approved manuals
and instructions.
The LifePulse High Frequency Ventilator and Patient Box(es)
should be returned to the factory every 2,000 hours or every year,
whichever comes first, to be checked, re-calibrated, maintained,
and have components repaired and replaced for the equipment to
operate reliably. Parts that have failed, in whole or in part, exhibit
excessive wear, are contaminated, or are otherwise at the end of
their useful life should not be used. Furthermore, they should be
replaced immediately with parts supplied by Bunnell Incorporated
or parts which are approved by Bunnell Incorporated.
The Life Pulse should be overhauled every 6,000 hours of use. The
battery should be replaced every 3 years.
Equipment that is not functioning correctly or is otherwise in need
of repair or maintenance must not be used until all necessary
repairs and/or maintenance have been completed and Bunnell
authorized service personnel have verified that the equipment
is functioning according to specifications and ready for use. This
equipment, and any accessories or component parts, should not
be modified.
The owner/user of this equipment will have the sole responsibility
and liability for any damage or injury to persons or property
(including the equipment itself) resulting from
• Operation not in accordance with the user instructions,
• Faulty maintenance not in accordance with the authorized
maintenance instructions,
• Repair by anyone other than Bunnell service personnel,
• Modification of the equipment or accessories, or use of
components or accessories that have either been damaged
or not authorized for use with this equipment by Bunnell
Incorporated.
The disposal of any Bunnell Incorporated equipment is the
responsibility of the owner of that equipment. Disposal
should always be in accordance with any local and/or national
requirements.
Bunnell LifePulse HFV User Manual | 87
Page 88
Appendix A
Warranty
The Bunnell Incorporated LifePulse High Frequency Ventilator is
warranted to be free from defects in materials and workmanship
and to meet published specifications under normal use for a
period of one (1) year from the date of in-service. The foregoing
is in lieu of any other warranty, expressed or implied, including
without limitation any warranty of merchantability, except as to
title, and can be amended only in writing by a duly authorized
representative of Bunnell Incorporated. Repair of this equipment
shall be done by a Bunnell certified technician. Buyer’s exclusive
remedy and Bunnell Incorporated’s sole obligation shall be
to replace, repair, or issue credit, at the discretion of Bunnell
Incorporated, for the parts that become defective or fail to meet
published specifications during the warranty period, provided that,
Bunnell Incorporated will not be liable under this warranty unless:
A. Bunnell Incorporated is notified promptly in writing by
buyer upon discovery of defects or failure to meet
specifications.
B. The defective part is returned to Bunnell Incorporated, transportation prepaid by buyer;
C. The defective unit or part is received by Bunnell Incorporated
for adjustment no later than one (1) week following the last day
of the warranty period; and
D. Bunnell Incorporated’s examination of the returned unit
discloses, to Bunnell Incorporated’s satisfaction, that the
defect(s) or failure(s) has not been caused by misuse, neglect,
improper installation or operation, unauthorized repair or
alteration, operator error or accident.
Any authorization of buyer by Bunnell Incorporated for repair of
Buyer’s unit or part by Buyer must be in writing to prevent voiding
warranty.
Products or parts replaced under the terms of this warranty are
warranted only through the terms of the original warranty.
THIS WARRANTY DOES NOT COVER NORMAL MAINTENANCE
SUCH AS CLEANING, ADJUSTMENT, OR LUBRICATION, AND
UPDATING OF EQUIPMENT OR PARTS. THIS WARRANTY
SHALL BE VOID AND SHALL NOT APPLY IF THE EQUIPMENT IS
USED WITH ACCESSORIES OR PARTS NOT MANUFACTURED
BY BUNNELL INCORPORATED OR AUTHORIZED FOR USE IN
WRITING BY BUNNELL INCORPORATED, OR IF THE EQUIPMENT
IS NOT MAINTAINED IN ACCORDANCE WITH A PRESCRIBED
SCHEDULE OF MAINTENANCE.
Visit the Bunnell Incorporated website at www.bunl.com
for a complete list of available products.
Bunnell LifePulse HFV User Manual | 91
Page 92
Appendix C
Bunnell LifePulse HFV User Manual | 92
Page 93
Appendix D
Device Specifications
LifePulse High Frequency Ventilator Model 204
Description of Controls
POWER Turns ventilator on or off; does not affect battery charger circuit
ENTER Initiates ventilation using selected settings; resets MAP and Servo alarm limits
STANDBY Ceases ventilation and places humidifier in Pause
SYSTEM TEST Initiates self test routine
AUDIO PAUSED Silences audible alarm for 60 seconds
Description of Ventilator Controls
PIP Selects new PIP setting
Rate Selects new Rate and changes I:E RATIO setting
I-Time Selects new I-Time and changes I:E RATIO setting
MAP LIMITS Displays and/or adjusts MAP alarm limits
SERVO LIMITS Displays and/or adjusts SERVO alarm limits
Increment (+) / Decrement (-) Adjusts the control settings
Parameters Set and Their Ranges
PIP 8 to 50 cm H2O, 1 cm H2O increments
Rate 240 to 660 bpm, 10 bpm increments
I-Time 0.020 to 0.034 sec, 0.002 sec increments
MAP LIMITS 0.0 to 40.0 cm H2O, 0.1 cm H2O increments
SERVO LIMITS 0.0 to 25.0 psi (0.0 to 172.4 kPa), 0.1 psi (0.689 kPa) increments
SERVO ≤20 psi (137.9 kPa)
Parameters Measured and Displayed and Their Ranges
PIP 0.0 to 92.3 cm H2O, 0.1 cm H2O resolution
PEEP -9.9 to 92 cm H2O, 0.1 cm H2O resolution (PEEP <20), 1 cm H2O (PEEP ≥ 20)
Δ P 0.0 to 92.3 cm H2O, 0.1 cm H2O resolution
MAP 0.0 to 92.3 cm H2O, 0.1 cm H2O resolution
SERVO 0.0 to 40.0 psi (0.0 to 275.8 kPa), 0.1 psi (0.689 kPa) resolution
I:E RATIO 1:1.6 to 1:12, displays VENT INOP codes
Bunnell LifePulse HFV User Manual | 93
Page 94
Appendix D
Audible and Visual Indicators and Alarms
READY PIP stable within +2.0 and -1.5 cm H2O of set value for 20 seconds
LOW GAS PRESSURE Mixed gas inlet supply pressure <30 psi (205.85 kPa)
LOSS OF PIP 15 seconds after ENTER is pressed, the airway pressure is <3 cm H2O, or PIP - PEEP
(∆P) is ≤2 cm H2O, or the airway pressure drops to <25% of the set PIP in the
Ready condition.
HIGH PIP Instantaneous airway pressure > set PIP + 5 cm H2O, or >65 cm H2O for 1 second, or
PIP > set PIP + 10 cm H2O for every breath cycle delivered for 30 seconds, or
PIP > set PIP + 30 cm H2O for every breath cycle delivered for 0.75 second
CANNOT MEET PIP Does not achieve Ready status or the Servo reaches 20 within 3 minutes of pressing the
ENTER button
VENT INOP (without a code) Ventilator reset or stopped via the watchdog circuit
VENT INOP (with a code) 01 A/D converter failure
02 During System Test, failure of the airway pressure system
03 During System Test, failure of servo pressure system
04 During System Test, failure of ventilator microprocessor watchdog circuit
05 RAM failure
06 ROM failure
07 NMI failure to execute
08 IRQ failure to execute
09 Both NMI and IRQ fail to execute
10 Rapid increase in Servo (>3.4 psi; 23.44 kPa)
CHECK VENT Valve status feedback error
Pinch valve status feedback error
Failure to detect purge
UPPER/LOWER MAP LIMIT MAP at onset of Ready condition ±1.5 cm H2O or
adjustable manual setting
UPPER/LOWER SERVO LIMIT Servo at onset of Ready condition ±1 if Servo >5, 20% if Servo between 1 and 5,
and 0.2 if Servo <1, or adjustable manual setting
BATTERY Operating from battery, AC power is interrupted
BATTERY LOW Battery’s useful capacity low during operation
BATTERY DEPLETED Battery’s useful charge expended
Audible and Visual Alarm Only Backup ventilator alarm detected
Ventilator reset via the software
Audible Alarm Only Loss of power to the ventilator
Alarm logic circuit malfunction
BATTERY FUEL GAUGE Shows approximate level of charge remaining on the battery
Bunnell LifePulse HFV User Manual | 94
Page 95
Appendix D
BATTERY CHARGING/FAILURE Indicates battery fails to charge, charging, or fully charged when connected to AC power
STANDBY Ventilator is in the Standby mode
ENTER Ventilator is in the Operating mode
SYSTEM TEST Ventilator is in the Self Test mode
System Sensitivity and Accuracy
Airway pressure sensitivity 0.1 cm H2O
Airway pressure accuracy ±2.0 cm H2O
Servo pressure sensitivity 0.04 psi (0.276 kPa)
Servo pressure accuracy ±1.0 psi (6.895 kPa)
Response Times
Patient airway pressure 0.5 kHz
sampling rate
Time to display pressures within specified accuracies after a step change:
Standby 80 seconds, updated every 10 seconds
Running, not Ready condition 10 seconds, updated every 2 seconds
Running, Ready condition 20 seconds, updated every 2 seconds
Time until PIP is stable <3 minutes
Time required for
compliance chamber to
vent from 20 psi to ambient <1.0 second
Miscellaneous
Gas supply requirements 30 to 60 psi (206.85 – 413.7 kPa)
>0.5 SCFM (14 slpm)
Oxygen Sample Port Flow 1.5 ±0.5 slpm
Power requirements 100-240 VAC, 50-60 Hz, 2A
Battery Operating Time >10 minutes
Battery Recharge Time <12 hours
Battery Qualification Time <30 minutes
Hour Meter 0-99,999 Hours, Non-resettable
High, Medium, and
Low Auditory Alarm 45 – 85 dBA Sound Pressure Range
Bunnell LifePulse HFV User Manual | 95
Page 96
Dimensions
Front height 13.1 in (33.3 cm)
Length 16.3 in (41.4 cm)
Width 12.3 in (31.2 cm)
Weight (with power cord) 26.9 lbs (12.2 kg)
Humidifier
Description of Controls
PAUSE Ceases/initiates humidification system
GAS Displays gas set temperature
WATER Displays water set temperature
Appendix D
Increment (+) / Decrement (-) Adjusts the Water or Gas temperature settings
Parameters Set, Measured, and/or Displayed
Gas temperature setting 32.0 to 42.0°C, 0.1°C increments
Gas temperature measured 20.0 to 50.0°C, 0.1°C resolution
Water temperature setting 32.0 to 42.0°C, 0.1°C increments
Water temperature measured 20.0 to 50.0°C, 0.1°C resolution
Default Settings
Gas Temperature 40.0°C
Water Temperature 38.0°C
Audible and Visual Indicators and Alarms
CHECK CIRCUIT - LEVEL Water level not detected within 86 seconds after exiting Pause mode
CHECK CIRCUIT - TEMP Gas or water temperature sensor determined to be out of range or cartridge door is open
HIGH LEVEL High water level detected
LOW LEVEL Low water detected for 2 minutes
HIGH TEMP - WATER More than 3°C above set point or >45°C for 10 minutes
LOW TEMP - WATER More than 3°C below set point for 30 minutes
Bunnell LifePulse HFV User Manual | 96
Page 97
HIGH TEMP - GAS More than 3°C above set point or >45°C for 1 minute
LOW TEMP - GAS More than 3°C below set point for 3 minutes
CHECK VENT (without a code) Humidifier reset or stopped via the watchdog circuit
CHECK VENT (with a code) 111 RAM failure
222 ROM failure
333 A/D converter failure
Audible and Visual Alarm Only Backup humidifier alarm detected
PAUSE Humidifier in Pause mode
Sensitivity and Accuracy
Temperature sensitivity ~0.1°C
Temperature accuracy ±2.0°C
Appendix D
Servo Response Times
Gas temperature 3 minutes
Water temperature 30 minutes
Pump Rate 0.7 mL/second
Patient Circuit
Humidifier Cartridge
H2O reservoir volume 50 mL
Compressible volume 20 mL
Total internal volume 70 mL
Patient Circuit
Gas Inlet Tube 10 ±0.25 inches (25.4 ±0.63 cm)
Water Inlet Tube 12 ±0.25 inches (30.5 ±0.63 cm)
Patient Circuit Tube 74 ±1.5 inches (188 ±3.8 cm)
Bunnell LifePulse HFV User Manual | 97
Page 98
Patient Box Model 314
Dimensions
Width 3.3 inches (8.4 cm)
Height 3.3 inches (8.4 cm)
Length 4.4 inches (11.2 cm)
Weight 1.8 lbs (0.81 kg)
Output
0 V 100 cm H2O
Appendix D
10.0 V 0 cm H2O
Sensitivity 10 mV / 0.1 cm H2O
Miscellaneous
Sound Pressure Level <45 dBA
External Temperature <44°C
Environmental Conditions
Transportation and Storage
Ambient Temperature -10 to 50°C
Relative Humidity 10 to 80% (non-condensing)
Atmospheric Pressure 500 to 1060 hPa
Operating
Ambient Temperature 10 to 40°C
Relative Humidity 10 to 75% (non-condensing)
Atmospheric Pressure 700 to 1060 hPa
Bunnell LifePulse HFV User Manual | 98
Page 99
Appendix D
Classifications/Applied Parts
• FDA Class III type medical equipment.
• Class I Electrical Equipment per ISO 60601-1.
• Fulfills the requirements of IEC 60601-1:2005 Ed: 3. Conforms to AAMI ES60601-1, IEC STDS 60601-1-6 & 60601-1-8,
Certified to CSA 22.2 NO. 60601-1.
• Type BF Applied Parts. The Patient Box and the gas delivery tube of the Patient Circuit may come in contact with the
patient and are therefore considered applied parts.
• Degree of protection against ingress per IEC 60529: IP21 rating on the LifePulse Ventilator and IP24 on the Patient Box.
• The LifePulse HFV System is categorized as an external communicating device with tissue/bone/dentin contact for
prolonged periods per ISO 10993.
Bunnell LifePulse HFV User Manual | 99
Page 100
Appendix D
Bunnell LifePulse HFV User Manual | 100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.